K Number
K201202

Validate with FDA (Live)

Device Name
MI View&GO
Date Cleared
2020-06-02

(29 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

Device Description

MI View&GO is a software-only medical device which will be delivered with the new generation of Siemens SPECT/CT and PET/CT scanners.

MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)

MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study information for the MI View&GO VA10A device:

Important Note: The provided document is a 510(k) summary for the MI View&GO VA10A device. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed de novo clinical study with specific acceptance criteria for novel performance claims. As such, many of the typical clinical study elements like "effect size," "sample size for test set," "number of experts," etc., are not explicitly described in the same way they would be for a new device with novel performance claims requiring rigorous clinical validation. The testing described is primarily for verification and validation of the software’s functions, performance requirements, and hazard mitigations against predetermined acceptance values, in line with established industry standards for medical device software.


1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of the submission (510(k) for a software update with substantial equivalence), the acceptance criteria are not in the form of clinical performance metrics like sensitivity/specificity for a diagnostic task. Instead, they relate to software engineering and regulatory compliance.

Acceptance Criterion TypeReported Device PerformanceComments
Software FunctionalityFunctions work as designed"Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed."
Performance RequirementsPerformance requirements and specifications have been met"performance requirements and specifications have been met"
Hazard MitigationAll hazard mitigations have been fully implemented"all hazard mitigations have been fully implemented."
Testing OutcomesAll testing has met the predetermined acceptance values"All testing has met the predetermined acceptance values."
Risk ManagementRisk analyses in compliance with ISO 14971"Risk Management has been ensured via risk analyses in compliance with ISO 14971"
Quality System ComplianceDesigned and manufactured in accordance with Quality System Regulations (21 CFR 820)"The device is designed and manufactured in accordance with Quality System Regulations as outlined in 21 CFR 820."
CybersecuritySpecific cybersecurity controls implemented to prevent unauthorized access, modifications, misuse, or denial of use; controls for information storage and transfer"The software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the software and external devices."
TraceabilityTraceability of requirements ensured during component integration, software validation, and system testing"Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing."
Compliance with StandardsAdherence to recognized and established industry standards (EN ISO 13485, IEC 62304, ISO 14971, IEC 62366-1, ISO 15223-1, NEMA PS 3.1 – 3.20)"Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards for development including EN ISO 13485 and IEC 62304." (Specific standards listed later)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as a separate "test set" in the context of image data for diagnostic performance evaluation. The "test set" here refers to the software testing environment. The document states "Verification and Validation activities have been successfully performed on the software package." This implies internal testing with various inputs.
  • Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not specified. Since this is primarily a software update and substantial equivalence claim, the focus is on functional verification rather than a de novo clinical study with specific patient data sets to prove diagnostic accuracy.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not specified. For this type of submission, "ground truth" refers more to established software requirements and specifications rather than a diagnostic standard established by human experts on a clinical dataset.
  • Qualifications of Experts: Not specified.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable or described. Adjudication is typically used in clinical studies when multiple human readers evaluate data, and their disagreements need to be resolved to establish a ground truth. Here, "ground truth" pertains to software functionality meeting its design specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study was not performed or described. The MI View&GO VA10A is described as an image processing software for viewing, manipulation, quantification, analysis, and comparison, not specifically an AI-driven diagnostic aid that would improve human reader performance in terms of diagnostic accuracy. Its features are largely related to workflow, display, and basic image processing.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable in the sense of a diagnostic algorithm's standalone performance (e.g., sensitivity/specificity for a disease). MI View&GO is a "medical diagnostic application" intended to be "utilized by appropriately trained health care professionals to aid in the management of diseases." It's a tool for professionals, not an autonomous diagnostic algorithm. Its performance is linked to the functionality described (viewing, manipulation, etc.), which were verified to meet specifications, not generate diagnostic outputs independently.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" for this submission are the predetermined acceptance values and functional specifications as defined by the software's design and requirements. It also includes compliance with recognized industry standards (ISO 14971, IEC 62304, etc.) and regulatory requirements (21 CFR 820, cybersecurity guidance). This is a software engineering "ground truth" rather than a clinical diagnostic ground truth like pathology or outcome data.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. MI View&GO VA10A is not described as an AI/Machine Learning device that undergoes a "training phase" on a dataset. It is image processing software.

9. How the Ground Truth for the Training Set was Established

  • Ground Truth for Training Set: Not applicable. As it's not an AI/ML device requiring a training set, the establishment of ground truth for such a set is not relevant.

Summary of the Study:

The "study" or rather, the evidence presented in this 510(k) submission, is a demonstration of Verification and Validation (V&V) activities. This included ensuring that:

  1. The software's functions worked as designed.
  2. Performance requirements and specifications were met.
  3. All hazard mitigations were fully implemented according to risk analyses (ISO 14971).
  4. All testing met predetermined acceptance values.
  5. The device adheres to recognized industry standards (e.g., IEC 62304, ISO 13485) and Quality System Regulations (21 CFR 820).
  6. Cybersecurity controls are in place.

The overall approach is a comparison to predicate devices (syngo.via MI Workflows, syngo.CT View&GO, syngo.CT Extended Functionality) to demonstrate substantial equivalence in terms of intended use, indications for use, and technological characteristics, especially for the new or modified features. The study proves that the device meets its acceptance criteria by stating that "All testing has met the predetermined acceptance values" and that V&V activities confirmed functional and performance requirements were met as per software development and regulatory standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

Siemens Medical Solutions USA, Inc. % Clayton Ginn Regulatory Technical Specialist 2501 N Barrington Road HOFFMAN ESTATES IL 60192

Re: K201202

Trade/Device Name: MI View&GO VA10A Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: May 1, 2020 Received: May 4, 2020

Dear Clayton Ginn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

June 2, 2020

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K201202

Device Name MIView&GO VA10A

Indications for Use (Describe)

MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, neurology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."

Form Approved: 0MB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

Section 05

510(k) Summary

K201202

as required by 21 CFR Part 807.87(h)

ldentification of the Submitter

Primary Contact:Alternate Contact:
Submitter:Clayton GinnRegulatory TechnicalSpecialistSiemens Medical SolutionsUSA, Inc.810 Innovation DriveKnoxville, TN 37932Veronica PadhariaRegulatory TechnicalSpecialistSiemens Medical SolutionsUSA, Inc.2501 N. Barrington RoadHoffman Estates, IL 60192
Telephone Number:(865) 898-2692(630) 877-5761
Fax Number:(865) 218-3227(847) 304-6023
Name / Address ofManufacturerSiemens Medical Solutions USA, IncMolecular Imaging2501 N. Barrington RoadHoffman Estates, IL 60192USA
Date of Submission:May 1st, 2020
Identification of the product
Device Proprietary Name:MI View&GO VA10A
Common Name:Image Processing Software
Classification Name:Picture Archiving and Communication System per 21 CFR892.2050
Product Code:LLZ
Classification Panel:Radiology
Device Class:Class II
Device Proprietary Name:syngo.via MI Workflows VB40A
Common Name:Image Processing Software
Classification Name:Picture Archiving and Communication System per 21 CFR892.2050
Product Code:LLZ
Classification Panel:Radiology
Device Class:Class II
Manufacturer:Siemens Medical SolutionsUSA, inc.
510(k) Number:K191309 (July 2019)
Reference Predicate Device
Device Proprietary Name:syngo.CT View&GO VB20A
Common Name:Image Processing Software
Classification Name:Picture Archiving and Communication System per 21 CFR892.2050
Product Code:LLZ
Classification Panel:Radiology
Device Class:Class II
Manufacturer:Siemens Medical SolutionsUSA, Inc.
510(k) Number:K170952 (April 2017)
Reference Predicate Device
Device Proprietary Name:syngo.CT Extended Functionality VB40A
Common Name:Image Processing Software
Classification Name:System, X-ray, Tomography, Computed per 21 CFR
Product Code:JAK
Classification Panel:Radiology
Device Class:Class II
Manufacturer:Siemens Medical SolutionsUSA, Inc.
510(k) Number:K192402 (September 2019)

{4}------------------------------------------------

Primary Predicate Device

Classification Name:

892.1750


{5}------------------------------------------------

syngo.via MI Workflows is deemed the primary predicate device due to it being the most similar to the device under review of this submission with respect to indications for use and technical characteristics. syngo.CT View&GO and syngo.CT Extended Functionality are considered reference predicate devices because MI View&GO has integrated technical characteristics initially cleared within these devices.

{6}------------------------------------------------

Device Description

MI View&GO is a software-only medical device which will be delivered with the new generation of Siemens SPECT/CT and PET/CT scanners.

MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)

MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function.

FeatureNewFunctionalityModification Description
Auto ViewingYesStream data into the application immediatelyfollowing acquisition.
Layout Gallery withDynamic Summing,Automatic Layouts, andRemove ImagesYesGenerate Automatic Layouts at runtime,Automatically sum frames for review in thelayout gallery, and remove images from adisplay
Configurable User PresetsYesConfiguration UI for preferred presets andlayouts
Tool GalleryNoFunctionality Cleared in K170952
Organ ProcessingNoFunctionality Cleared in K191309
MI Cardiology Reading andThird Party Software(Cedars 2017, Corridor4DM 2018)YesRestart applications and filter the cardiologydata passed to an application.
Series PanelNoFunctionality Cleared in K170952, K191309
Basic Operation in theViewerYesCorrect slice positions, panning, rotation onsynchronized segments
Automatic Reconstruction,Auto Ranges, and CardiacAuto RangesYesGenerate parallel ranges with Auto Ranges andgenerate a series of images in cardiac planesCardiac Auto Ranges
Display Types and ImageFusionNoFunctionality Cleared in K170952, K191309
3D Visualization andOrientationNoFunctionality Cleared in K170952, K191309
Restricting Volumes andCardiac MaskingYesMask the heart and exclude background regionsthat can have an impact on statisticalcalculations in an examination with CardiacMasking.
Image Optimization andPreparationNoFunctionality Cleared in K170952, K191309
Toggle Image OverlayYesConfigure customized image text
Spine and Rib LabelingNoFunctionality Cleared in K191309

MI View&GO includes the following features:

{7}------------------------------------------------

Neuro DSANoFunctionality Cleared in K192402
Osteo ExtensionNoFunctionality Cleared in K192402
Oncology ExtensionNoFunctionality Cleared in K192402
Vessel ExtensionNoFunctionality Cleared in K192402
ROI HU ThresholdNoFunctionality Cleared in K192402
Dual Energy ROINoFunctionality Cleared in K192402
Endoscopic ViewNoFunctionality Cleared in K192402
Printing of ImagesNoFunctionality Cleared in K170952

Technological Characteristics

The MI View&GO software modifications are based on the commercially available syngo.via MI Workflows software (K191309), syngo.CT View&GO software (K170952), and syngo.CT Extended Functionality (K192402). The features into this software device had no impact on the technological characteristics already present in the commercially available predicate system.

Intended Use

An individual software program, or group of programs, routines, or algorithms that add specific image processing and/or analysis capabilities to a positron emission tomography (PET) and Single Photon Emission Computed Tomography (SPECT) imaging system configuration. A basic set of application programs and routines is included with such computer controlled imaging systems and they can be upgraded to correct programming errors or to add new system capabilities. Some application software routines or groups of routines (packages) must be combined with specific hardware or firmware accessories or configurations in order to function as intended. Application program packages are typically identified by a proprietary name and "version" or "upgrade" number.

The intended use for MI View&GO compared to the primary and reference device has not changed.

Indications for Use

MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)

MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

{8}------------------------------------------------

Performance Testing / Safety and Effectiveness

The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.

Risk Management has been ensured via risk analyses in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards for development including EN ISO 13485 and IEC 62304.

Cybersecurity information in accordance with FDA Guidance documents issued October 2, 2014 has been provided. The software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the software and external devices.

Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.

The device is designed and manufactured in accordance with Quality System Regulations as outlined in 21 CFR 820. The FDA recognized standards are listed as follows:

  • . Recognition Number 13-79: IEC 62304 Edition 1.1 2015-06
  • . Recognition Number 12-300: NEMA PS 3.1 – 3.20 (2016)
  • Recognition Number 5-40: ISO 14971:2007 Second Edition ●
  • . Recognition Number 5-114: IEC 62366-1 Edition 1.0 2015
  • . Recognition Number 5-117: ISO 15223-1 Third Edition 2016

Statement Regarding Substantial Equivalence:

There are no differences in the Intended Use or Fundamental Technological Characteristics of the MI View&GO software as compared to the currently commercially available syngo.via MI Workflows software (K191309). The Indications for Use for MI View&GO is a condensed version of the Indications for Use of syngo.via MI Workflows and does not expand the intended use from that of the predicate device. Both the current and predicate devices are used for viewing, manipulation, 3D- visualization and comparison of medical images from multiple imaging modalities and/or multiple time-points

Additionally, there have been no changes that raise any new issues of safety and effectiveness as compared to the predicate device. Based on this information, as well as the documentation in support of the modifications, it is Siemens opinion that the MI View&GO software-with the modifications outlined in this application-is substantially equivalent to the predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).