K Number
K242127

Validate with FDA (Live)

Date Cleared
2024-08-16

(28 days)

Product Code
Regulation Number
864.5400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS) are bench top, fully automated, random access analyzers designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

Device Description

The ACL TOP Family 50 Series are fully automated coagulation analyzers that utilize the same intuitive software, the same consumables, reagents, calibrators and controls, and provide the same analytical methodology for routine and specialty assay result reporting as the predicate ACL TOP Family.

The ACL TOP Family 50 Series instrument performs the following types of tests, using the same optical measuring wavelengths and test parameters as the predicate ACL TOP Family:

  • Coagulometric (Turbidimetric) Measurements
  • . Chromogenic (Absorbance) Measurements
  • . Immunological Measurements

The ACL TOP Family 50 Series also offers new pre-analytical features not available on the current ACL TOP Family as described below. These features are not intended to replace laboratory quality policies. The features simply alert the instrument operator to a potential HIL (Hemoglobin, Icteric and Lipemia) interference situation specific to the assays requested for a sample, underfilled sample tubes or a detected clog. The user will determine how to handle these situations (for example, by not reporting the results, or reporting the results with, or without, additional comments).

AI/ML Overview

The provided text does not contain detailed acceptance criteria and a study proving the device meets those criteria in the traditional sense of a clinical or performance study for a diagnostic device.

This document is a 510(k) summary for a Special 510(k) submission, focusing on software changes (new remote features and enhanced cybersecurity measurements) to an existing device, the ACL TOP Family 50 Series. The key statement regarding acceptance criteria and proof of performance is:

"The software verification and validation study results demonstrate that the ACL TOP Family 50 Series with updated nonanalytical features is safe and effective for its intended purpose and equivalent in performance to the predicate device (K150877)."

This indicates that the "acceptance criteria" were related to the software's functionality, security, and the assertion that these non-analytical changes do not impact the analytical performance of the instrument. The "study" mentioned is a "software verification and validation study."

Given this, I will extract and infer the information based on the context of a software-focused 510(k) for a device where analytical performance is already established by a predicate.

Here's the breakdown based on your requested format:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (Inferred from text)Reported Device Performance (Inferred from text)
Remote Control FunctionalityThe added permission-based remote-control function for desktop sharing must operate as intended, providing remote access.Functionality was verified through software verification and validation. The features are available on Windows 10 (SW version 6.5.3) instruments.
Remote Software/OS/Test Parameter DeliveryThe functionality to remotely deliver software, OS updates (patches), and test parameter releases must operate securely and effectively.Functionality was verified through software verification and validation. The features are available on Windows 10 (SW version 6.5.3) instruments.
Security and Privacy ControlsThe new remote features must utilize security and privacy controls by design and installed by default, ensuring data integrity and user confidentiality. Enhanced cybersecurity measures (e.g., MS BitLocker, Digital Signature, MS AppLocker) must be effectively implemented.Security and privacy controls were verified to be in place. New risk mitigation controls (MS BitLocker, Digital Signature, MS AppLocker) for enhanced cybersecurity were added and validated as part of the software verification and validation.
Impact on Analytical PerformanceThe new remote and cybersecurity features, being non-analytical, must not negatively impact the analytical performance of the instrument (i.e., coagulation, chromogenic, and immunological measurements, and pre-analytical HIL check, tube fill height check, and clog detection functions). The overall performance must remain equivalent to the predicate device (K150877).The document explicitly states: "These features do not impact the analytical performance of the instrument." The software verification and validation study results demonstrate the device "is safe and effective for its intended purpose and equivalent in performance to the predicate device (K150877)." This implies that the analytical functions were not degraded by the software updates.
CompatibilityThe new features are intended to be available only for ACL TOP Family 50 Series instruments at Windows 10 (SW version 6.5.3) and not for Windows 7 versions.This compatibility restriction is stated and assumed to be met by the software itself.

2. Sample size used for the test set and the data provenance
The document refers to a "software verification and validation study results." For software changes, the "test set" would typically involve functional and security testing scenarios rather than patient data. The document does not specify a sample size in terms of patient data or the provenance (country of origin, retrospective/prospective) because the changes are non-analytical software updates to an existing, already cleared device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable based directly on the provided text, as the "ground truth" for these software functionalities would be their correct operation and security posture, established by software engineers, cybersecurity experts, and regulatory experts. The document does not detail specific experts or their qualifications for the V&V study.

4. Adjudication method for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for clinical studies involving human interpretation of medical images or tests. For software verification and validation of non-analytical features, testing protocols and bug reporting/resolution processes would be used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. This device is a coagulation analyzer, not an AI-powered diagnostic imaging device involving "human readers." The changes specifically "do not impact the analytical performance of the instrument."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is a device that runs algorithms for coagulation measurements, and the "standalone" performance of these measurement algorithms was established by the predicate device (K150877). The new features are remote control and cybersecurity updates, not new analytical algorithms. The software verification and validation focused on these new non-analytical features operating correctly without human intervention (e.g., remote update deployment), but this is not an "algorithm only" performance study in the typical sense for clinical impact.

7. The type of ground truth used
For the software verification and validation related to the new remote and cybersecurity features, the "ground truth" would be:

  • Functional Specification Adherence: The software correctly performs the defined remote control and update delivery functions according to its design specifications.
  • Security Standard Compliance: The cybersecurity features meet established security standards and mitigate identified risks.
  • Non-Interference: The new features do not interfere with the validated analytical performance of the device.

8. The sample size for the training set
Not applicable. This is a software update for an existing medical device, not a machine learning or AI algorithm development that requires a "training set" of data.

9. How the ground truth for the training set was established
Not applicable, as there is no training set for the software changes described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

August 16, 2024

Instrumentation Laboratory (IL) Co. Sneh Pingle Regulatory Affairs Manager I 180 Hartwell Road Bedford, Massachusetts 01730

Re: K242127

Trade/Device Name: ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS) Regulation Number: 21 CFR 864.5400 Regulation Name: Coagulation Instrument Regulatory Class: Class II Product Code: GKP Dated: July 18, 2024 Received: July 19, 2024

Dear Sneh Pingle:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding

{1}------------------------------------------------

When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min Wu-S

Min Wu. Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K242127

Device Name

ACL TOP Family 50 Series (ACL TOP 750 CTS; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 350 CTS; ACL TOP 350 CTS)

Indications for Use (Describe)

The ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS) are bench top, fully automated, random access analyzers designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "werfen" in a sans-serif font. The word is written in a dark blue color. The letters are evenly spaced and the word is horizontally aligned.

510(k) Summary

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.

Submitter's InformationInstrumentation Laboratory Company180 Hartwell RoadBedford, MA 01730-2443 (USA)
------------------------------------------------------------------------------------------------------------------
Contact PersonSneh PingleRegulatory Affairs Manager IPhone: 781-674-3266Fax: 781-861-4207Email: spingle@werfen.com
--------------------------------------------------------------------------------------------------------------------------------------
Preparation DateJuly 15, 2024
---------------------------------
Device Trade NameACL TOP Family 50 Series Models:
ACL TOP 350 CTS ACL TOP 550 CTS ACL TOP 750 ACL TOP 750 CTS ACL TOP 750 LAS
Predicate DeviceACL TOP Family 50 SeriesK150877
-----------------------------------------------------
Regulatory InformationRegulation Section No.21 CFR 864.5400
Regulation DescriptionCoagulation instrument
ClassificationClass II
Product CodeGKP
PanelHematology (81)

{4}------------------------------------------------

wert

Device Description

The ACL TOP Family 50 Series are fully automated coagulation analyzers that utilize the same intuitive software, the same consumables, reagents, calibrators and controls, and provide the same analytical methodology for routine and specialty assay result reporting as the predicate ACL TOP Family.

The ACL TOP Family 50 Series instrument performs the following types of tests, using the same optical measuring wavelengths and test parameters as the predicate ACL TOP Family:

  • Coagulometric (Turbidimetric) Measurements
  • . Chromogenic (Absorbance) Measurements
  • . Immunological Measurements

The ACL TOP Family 50 Series also offers new pre-analytical features not available on the current ACL TOP Family as described below. These features are not intended to replace laboratory quality policies. The features simply alert the instrument operator to a potential HIL (Hemoglobin, Icteric and Lipemia) interference situation specific to the assays requested for a sample, underfilled sample tubes or a detected clog. The user will determine how to handle these situations (for example, by not reporting the results, or reporting the results with, or without, additional comments).

Intended Use / Indications for Use

The ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS) are bench top, fully automated, random access analyzers designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

Special Conditions for Use Statement

· For prescription use only.

{5}------------------------------------------------

wert

Reason for Special 510(k) Submission

The new remote features introduced on the ACL TOP Family 50 Series are limited to adding permissionbased remote-control function for desktop sharing and remotely delivering software, OS updates (patches) and test parameter releases, utilizing security and privacy controls by design and installed by default. These features do not impact the analytical performance of the instrument and are already available on the ACL TOP Family 70 Series, FDA cleared under K231031.

The new remote features will be available to customers as an "optional add-on service" software tool called ProDx. This service software tool (ProDx) is being expanded beyond consistent monitoring, data gathering and capturing of quality control logs to now also include the following new remote features that are the subject of this Special 510(k):

  • Added permission-based remote-control function for desktop sharing. ●
  • Added functionality to remotely deliver software, OS updates (patches) and test parameter ● releases.

The above features are managed through cloud architecture for seamless and secure deployment; the same cloud architecture as already used on the ACL TOP Family 70 Series.

These new features will be available only for ACL TOP Family 50 Series instruments at Windows 10 (SW version 6.5.3). The features will not be available for ACL TOP Family 50 Series instruments at Windows 7.

{6}------------------------------------------------

werfen

Summary Comparison of Technological Characteristics (Predicate)
ItemPredicate Device (K150877)Subject Device
Similarities
Trade NameACL TOP Family 50 SeriesACL TOP Family 50 Series
Model Trade NamesACL TOP 350 CTS ACL TOP 550 CTS ACL TOP 750 CTS ACL TOP 750 ACL TOP 750 LASSame
ManufacturerInstrumentation Laboratory Co.Same
Indications for Use /Intended UseThe ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS) are bench top, fully automated, random access analyzers designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis.The systems provide results for both direct hemostasis measurements and calculated parameters.Same
Test MethodologyCoagulometric measurement Chromogenic measurement Immunological measurementSame
Wavelengths405 nm 535 nm 671 nmSame
Test MenuClotting Assays Chromogenic Assays Immunological AssaysSame
Test ParametersAssay volumes, rinse and clean cycles, timing, optical parameters, data algorithms, material definitionSame
Summary Comparison of Technological Characteristics (Predicate) (Cont.)
ItemPredicate Device (K150877)Subject Device
Similarities (Cont.)
Reagents, Controls andCalibratorsPackaging, formulation, performance claims in labelingSame
Fluidic HandlingAspiration, dispense, mixing, rinse, clean, temperature control, bulk fluidsSame
Sample HandlingCap piercing, onboard storageSame
Onboard Reagent StorageStirring, temperature controlSame
Reaction and DetectionOptics, temperature controlSame
System SoftwareHardware control, user interface except as noted in the Differences section belowSame
Quality ControlAutomated QCSame
Pre-Analytical HIL CheckStandard for all models:Third measurement wavelength @535 nmand an additional emitter control channelSame
Tube Fill Height CheckStandard for all modelsSame
Clog DetectionStandard for all modelsSame
External Skins and ChassisExternal Skins and ChassisSame
On-market instrumentappearanceOn-market instrument appearanceSame

{7}------------------------------------------------

werfen

{8}------------------------------------------------

Differences
ItemPredicate Device (K150877)Subject Device
Operating SystemWindows 7 and Windows 10Same.
Software/ CybersecurityOn-market instrument software andcybersecurityAdded permission-based remote-controlfunction for desktop sharing andremotely delivering software, OSupdates (patches) and test parameterreleases, utilizing security and privacycontrols by design and installed bydefault.These new features will be availableonly for ACL TOP Family 50 Seriesinstruments at Windows 10 (SWversion 6.5.3). The features will not beavailable for ACL TOP Family 50Series instruments at Windows 7.Added new risk mitigation controls(Example MS BitLocker, DigitalSignature, MS AppLocker) for enhancedcybersecurity.

Conclusion

The technological and functional characteristics of the ACL TOP Family 50 Series (subject device) as described herein are substantially equivalent to that of the ACL TOP Family 50 Series (predicate device). The software verification and validation study results demonstrate that the ACL TOP Family 50 Series with updated nonanalytical features is safe and effective for its intended purpose and equivalent in performance to the predicate device (K150877).

§ 864.5400 Coagulation instrument.

(a)
Identification. A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.(b)
Classification. Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.