K Number
K173761

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-08-17

(249 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Penumbra Reperfusion Catheters and Separators: As part of the Penumbra System, the Reperfusion Catheters and Separators are in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra 3D Revascularization Device: As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra Aspiration Tubing: As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Description

The Penumbra System Reperfusion Catheter JET 7 is a component to the currently available Penumbra System. The Reperfusion Catheter JET 7 component provides a larger lumen to assist in the efficient removal of thrombus from the neurovasculature. The devices are provided sterile, non-pyrogenic, and intended for single use only.

AI/ML Overview

The Penumbra System Reperfusion Catheter JET 7 is a medical device for revascularization in acute ischemic stroke. It was compared to a predicate device, the Penumbra System ACE 68 Reperfusion Catheter, to demonstrate substantial equivalence through non-clinical testing.

1. Table of Acceptance Criteria and Reported Device Performance

The device underwent various non-clinical tests (biocompatibility, bench-top, and animal studies) to confirm its safety and performance. The tables below summarize the acceptance criteria and results for key tests mentioned in the provided text.

Biocompatibility Testing

TestAcceptance CriteriaResultsPass/Fail
In Vitro CytotoxicitySample extracts must yield cell lysis grade 2 or lowerGrade 1: SlightPass
SensitizationTest Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control Grade < 1)Grade 0: No visible changePass
Acute Intracutaneous Reactivity (Irritation)The difference in the mean test article and mean control score must be grade 1.0 or lowerGrade ≤ 1.0 difference between mean test article and mean control scorePass
Acute Systemic ToxicityNo evidence of systemic toxicity (e.g., > 10% weight loss in ≥3 animals, ≥2 mortalities, ≥2 abnormal behaviors)No evidence of systemic toxicity (no weight loss, no death, all animals appeared normal)Pass
Rabbit Pyrogen StudySample extracts must not cause a total rise in body temperature of ≥0.5°CNon-pyrogenic: No single animal had a total body temperature rise of ≥0.5°CPass
In Vitro HemolysisSample extracts must be non-hemolytic (≤ 2% hemolytic index)Non-hemolytic: Hemolytic Index = 0.70%, Corrected Hemolytic Index = 0.00%Pass
Coagulation (Prothrombin Time)Clotting times must be similar to negative control valuesTest article coagulation times were statistically lower than negative controlPass
Coagulation (Partial Thromboplastin Time)Clotting times must be similar to predicate (negative control) values using analysis of varianceTest article coagulation times were statistically similar to the predicatePass
Complement ActivationConcentrations of C3a and SC5b-9 in test samples statistically similar to predicate and lower than positive controlTest sample concentrations of C3a and SC5b-9 were statistically similar or lower than predicate, and statistically lower than positive controlPass
Dog ThrombogenicityDevice must be non-thrombogenic after 4 hours in vivo compared to a control deviceNo significant thrombosis with a Grade of 0 observed in 2 out of 2 test and 2 out of 2 control sitesPass

Bench-top Testing

AttributeSpecificationResults
Dimensional / Visual InspectionUnits meet all product specifications.Pass
Simulated Use (Intracranial Access, Vessel Access Entry Performance, Delivery/Retrieval Forces & Clot Removal)Effectiveness of clot removal and catheter does not collapse under vacuum in an anatomical model.Pass
Physician Evaluation (Deliverability & Clot Removal)Performance evaluated in a simulated neurovascular tortuosity model with predicate as baseline.Pass
Kink Resistance (Distal, Midshaft, Proximal)No kinking when formed in 2.5 mm, 10 mm, and 25 mm radius.Pass
Particulate testing≥ 10 μm ≤ 6000 particles; ≥ 25 μm ≤ 600 particlesPass (for mandated sizes)
Coating IntegrityCoating has not delaminated, peeled, or flaked prior to or after simulated use particulate testing.Pass
Markerband VisibilityThe markerband is fluoroscopically visible.Pass
Hub Air AspirationNo leaks detected when vacuum is pulled on the injection lumen.Pass
Pressure Test45 psi for 30 sec minimum.Pass
JET 7 / Sheath or 8F Guide Catheter Friction ForceMaximum value per specification.Pass
JET 7 / 0.014 in. Guidewire Friction ForceMaximum value per specification.Pass
Joint sections bond strengthMinimum value per specification.Pass
Markerband Section Bond StrengthMinimum value per specification.Pass
Hub to Shaft Bond StrengthMinimum value per specification.Pass
Hub to Hypotube Bond StrengthMinimum value per specification.Pass
Elongation to failureElongation ≥ 5%.Pass
CorrosionNo visible corrosion immediately after corrosion testing procedure.Pass

2. Sample Sizes Used for the Test Set and Data Provenance

  • Biocompatibility: The specific number of samples for each biocompatibility test (e.g., number of extracts, animals) is not explicitly stated in the summary, but it implies standard practices for these types of tests conducted under GLP. The data provenance is from previously performed studies (leveraged from the predicate device ACE 68, K161640 cleared on July 12, 2016). These studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices (GLP). The location of these previous studies is not specified but is assumed to be within a regulated, GLP-compliant environment.
  • Bench-Top Testing: The text does not specify exact sample sizes for each bench-top test, but it states that "the units used in this Design Verification testing meet all product specifications," suggesting multiple samples were tested for each attribute.
  • Animal Study: The study used porcine subjects. "One side of each swine was treated with the Reperfusion Catheter JET 7 and the contralateral side was treated with predicate device." The exact number of swine is not stated, but the Dog Thrombogenicity test (part of biocompatibility) mentions "2 out of 2 test site and 2 out of 2 control sites," suggesting a small number of animals for that specific test.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Biocompatibility: Not directly applicable in the sense of expert consensus for ground truth on performance. Biocompatibility results are based on objective laboratory measurements against established standards (EN ISO 10993-1:2009/AC:2010). The animal study involved a "Sponsor Pathologist" to assess vascular response by gross necropsy and histopathology. No specific number or additional qualifications beyond "Sponsor Pathologist" are provided.
  • Bench-Top Testing: A "Physician Evaluation" was performed for deliverability and clot removal. The number of physicians and their specific qualifications are not detailed beyond "Multiple Physician."

4. Adjudication Method for the Test Set

  • Biocompatibility: For tests like Systemic Toxicity, the results were assessed based on objective criteria (e.g., weight loss, mortality, abnormal behavior). For the Dog Thrombogenicity study, assessment of thrombosis was made by the Sponsor Pathologist. No formal adjudication (e.g., 2+1, 3+1) is mentioned as it's typically not explicitly used for these types of non-clinical tests unless there are subjective interpretations needing consensus, which isn't detailed here.
  • Bench-Top Testing (Physician Evaluation): The method of physician evaluation for deliverability and clot removal is not described as having an adjudication process beyond "multiple physician performance evaluation."

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No MRMC comparative effectiveness study was done. The submission states, "No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes." The comparison was primarily against the predicate device based on non-clinical data and leveraging clinical outcomes from existing literature on similar devices.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

This device (Penumbra System Reperfusion Catheter JET 7) is a physical medical device, not an AI algorithm. Therefore, "standalone" performance in the context of an algorithm does not apply. The device's performance is inherently "standalone" in vitro and in vivo animal testing, with human involvement primarily in operating the device and evaluating the results.

7. The Type of Ground Truth Used

  • Biocompatibility: Ground truth was established by adherence to recognized international standards (EN ISO 10993-1:2009/AC:2010) and objective measurements defined within these standards. For the animal study, pathology (gross necropsy and histopathology by a Sponsor Pathologist) served as the ground truth for vascular response.
  • Bench-Top Testing: Ground truth was established by engineering specifications, physical measurements, and qualitative assessments (e.g., "no kinking," "no leaks," "fluoroscopically visible") based on predefined acceptance criteria. For the "Physician Evaluation," the ground truth was expert opinion/observation on deliverability and clot removal compared to a baseline (predicate device).

8. The Sample Size for the Training Set

This product is a physical medical device and not an AI/machine learning algorithm. Therefore, the concept of a "training set" in the context of AI is not applicable. The device's design and manufacturing processes are developed based on engineering principles and prior knowledge from predicate devices.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, the concept of a "training set" does not apply to this physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 17, 2018

Penumbra, Inc. Michaela Mahl, MSBE Senior Manager Regulatory Affairs One Penumbra Place Alameda, California 94502

Re: K173761

Trade/Device Name: Penumbra System (Reperfusion Catheter JET 7) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: July 17, 2018 Received: July 18, 2018

Dear Michaela Mahl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Xiaolin Zheng -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173761

Device Name

Penumbra System (Reperfusion Catheter JET 7)

Indications for Use (Describe)

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra 3D Revascularization Device

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo features the word "Penumbra" in a red sans-serif font. To the right of the name is a red circular graphic with a white "P" inside, which is the company's symbol.

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Penumbra System® Reperfusion Catheter JET 7.

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Michaela Mahl Senior Manager, Regulatory Affairs Phone: (510) 748-3288 FAX: (510) 217-6414 Email: mmahl@penumbrainc.com

1.3 Date of Preparation of 510(k) Summary

July 17, 2018

1.4 Device Trade or Proprietary Name

Penumbra System® (Reperfusion Catheter JET 7)

1.5 Device Classification

Regulatory Class: II Classification Panel: Neurology Classification Name: Percutaneous Catheter Regulation Number: 21 CFR §870.1250 Product Code: NRY (Catheter, Thrombus Removal)

1.6 Predicate Devices

510(k) Number / Clearance DateName of DeviceName of Manufacturer
Primary Predicate Device
K161640 cleared on July 12, 2016Penumbra System – ACE 68Reperfusion CatheterPenumbra, Inc.One Penumbra PlaceAlameda, CA 94502 USA
Additional Predicate Device
K162901 (applicable for indicationonly) cleared on April 20, 2017Penumbra System – 3DRevascularization DevicePenumbra, Inc.One Penumbra PlaceAlameda, CA 94502 USA

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Penumbra" in a red font, followed by a red circle with a white "P" inside. The "P" is stylized with a horizontal line through the middle. The logo is simple and clean, with the company name clearly displayed next to its icon.

Predicate Comparison 1.7

System NamePenumbra System®
Device NameACETM 68JET 7
510(k) No.K161640 (applicable for ACE 68)K162901 (applicable for indication only)K173761
ClassificationClass II, NRYSAME
IndicationPenumbra Reperfusion Catheters and SeparatorsAs part of the Penumbra System, theReperfusion Catheters and Separators areindicated for use in the revascularization ofpatients with acute ischemic stroke secondary tointracranial large vessel occlusive disease(within the internal carotid, middle cerebral –M1 and M2 segments, basilar, and vertebralarteries) within 8 hours of symptom onset.Patients who are ineligible for intravenous tissueplasminogen activator (IV t-PA) or who fail IVt-PA therapy are candidates for treatment.Penumbra 3D Revascularization DeviceAs part of the Penumbra System, the Penumbra3D Revascularization Device is indicated for usein the revascularization of patients with acuteischemic stroke secondary to intracranial largevessel occlusive disease (within the internalcarotid, middle cerebral – M1 and M2 segments)within 8 hours of symptom onset. Patients whoare ineligible for intravenous tissue plasminogenactivator (IV t-PA) or who fail IV t-PA therapyare candidates for treatment.Penumbra Aspiration TubingAs part of the Penumbra System, the PenumbraSterile Aspiration Tubing is indicated to connectthe Penumbra Reperfusion Catheters to thePenumbra Aspiration Pump.Penumbra Aspiration PumpThe Penumbra Aspiration Pump is indicated as avacuum source for Penumbra AspirationSystems.SAME
System NamePenumbra System®JET 7
Device NameACETM 68
Materials
Proximal hubGrilamid (TR55-LX)SAME
Strain Relief [Hub Sleeve]Grilamid (TR55)SAME
Strain Relief304 Stainless Steel (SS)SAME
ID BandPolyolefin, PET yellow [black foil]Polyolefin, PET black [white foil]
LinerPTFESAME
Catheter Shaft
ExtrusionsPolyurethane, Polyether Block Amide, Nylon12SAME
Distal Coil ReinforcementNiTi wireSAME
Proximal CoilReinforcementSS wire and NiTi wireSAME
Extrusion ColorantsClear/ Natural or PurpleSAME
Tip ShapeStraightSAME
MarkerbandPlatinum/Iridium (90% Pt, 10% Ir)SAME
CoatingHydrophilic (proprietary)SAME
Dimensions
Proximal OD0.084 in Max0.085 in Max
Proximal ID0.068 in Min0.072 in Min
Distal OD0.084 in Max0.085 in Max
Distal ID0.068 in Min0.072 in Min
Effective Length115, 120, 125, 127, 132 cmSAME
Distal Flex Length30 cmSAME
Coating Length30 cmSAME
Accessories
Peelable SheathPTFESAME
Rotating Hemostasis ValvePolycarbonate, silicone o-ringSAME
Shaping Mandrel0.038in OD stainless steelSAME
Packaging Materials
PouchPolyester/Polyethylene/TyvekSAME
Packaging HoopPolyethyleneSAME
Packaging Tray(Kit Configuration)Polyethylene terephthalate, PolystyreneSAME
Packaging CardPolyethyleneSAME
Display CartonSBS PaperboardSAME
Hoop: Hoop/Packaging Card/Pouch/BoxSAME
Packaging ConfigurationKit: Tray/Retainer/Lid/AspirationTubing/Accessory Pouch/Pouch/BoxSAME
SterilizationEOSAME
Shelf-Life36 Months6 Months
System NamePenumbra System®
Device NameACE™ 68JET 7
UseSingle use, disposableSAME

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Penumbra" in a red font, followed by a red circle with a white "P" inside. The word "Penumbra" is written in a sans-serif font. The "P" in the circle is also in a sans-serif font. The logo is simple and modern.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the word "Penumbra" in red font, followed by a red circular logo with a white "P" inside. The logo is positioned to the right of the word "Penumbra". The font is sans-serif and the logo is simple and modern.

Device Description 1.8

The Penumbra System Reperfusion Catheter JET 7 is a component to the currently available Penumbra System. The Reperfusion Catheter JET 7 component provides a larger lumen to assist in the efficient removal of thrombus from the neurovasculature. The devices are provided sterile, non-pyrogenic, and intended for single use only.

1.9 Indications for Use

Penumbra Reperfusion Catheters and Separators

As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra 3D Revascularization Device

As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Penumbra Aspiration Tubing

As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.

1.10 Summary of Non-Clinical Data

As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the device follows.

Included in this section are summary descriptions of the testing, which substantiates the substantially equivalent performance of the subject Penumbra System JET 7 device to the predicate device:

  • Biocompatibility .
  • Design Verification (Bench-Top Testing) .
  • Design Validation (GLP Animal Testing) ●

The subject Penumbra System Reperfusion Catheter JET 7 met all established requirements.

1.10.1 Leveraged Biocompatibility Testing

The materials used in Reperfusion Catheter JET 7 are currently utilized in the currently available predicate device, Reperfusion Catheter ACE 68. The JET 7 device is manufactured using the same processes and in the same environment as ACE 68. The predicate. ACE 68 is classified as an externally communicating device with circulating blood contact and limited exposure less than 24 hours, which is the same proposed classification for JET 7. All biocompatibility risks have been mitigated through successful completion of prior biocompatibility testing per

EN ISO 10993-1:2009/AC: 2010 which will be leveraged for the JET 7 device. Therefore, no further biocompatibility testing is required.

Evidence of the biocompatibility of the Reperfusion Catheter JET 7 is derived from a series of previously performed studies listed and summarized in the table below. The studies were selected in accordance with EN ISO 10993-11 guidelines (Biological Evaluation of Medical Devices) for a limited exposure (< 24 hours), externally

1 EN ISO 10993-1: 2009/AC:2010 Biological Evaluation of Medical Devices, Part 1: Evaluation and testing within a risk management process

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Penumbra company logo. The word "Penumbra" is written in a blue serif font. To the right of the word is a red circle with a white "P" inside.

communicating device with circulating blood contact. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices (GLP).

TestAcceptance CriteriaResultsPass / Fail
In Vitro CytotoxicitySample extracts must yield cell lysis grade 2 or lowerGrade 1: SlightPass
SensitizationTest Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control Grade < 1)Grade 0: No visible changePass
Acute Intracutaneous Reactivity (Irritation)The difference in the mean test article and mean control score must be grade 1.0 or lowerGrade ≤ 1.0 difference between mean test article and mean control scorePass
Systemic Toxicity
Acute Systemic ToxicitySample extracts must not cause the following:• > 10% weight loss in 3 or more test animals• Mortality of 2 or more test animals• Abnormal behavior in 2 or more test animalsNo evidence of systemic toxicity from sample extracts• No weight loss (all gained weight)• No death• All test animals appeared normalPass
Rabbit Pyrogen StudySample Extracts must not cause a total rise in body temperature of ≥0.5°CNon-pyrogenic: No evidence of material-mediated pyrogenicity; no single animal had a total body temperature rise of ≥0.5°CPass
Hemocompatibility
In Vitro HemolysisSample extracts must be non-hemolytic (≤ 2% hemolytic index)Non-hemolytic:Hemolytic Index = 0.70%Corrected Hemolytic index = 0.00%Pass
Coagulation(Prothrombin Time)Clotting times must be similar to negative control valuesTest article coagulation times were statistically lower than negative controlPass
Coagulation(Partial Thromboplastin Time)Clotting times must be similar to predicate (negative control) values using analysis of varianceTest article coagulation times were statistically similar to the predicatePass
TestAcceptance CriteriaResultsPass / Fail
Complement ActivationThe concentrations of C3a andSC5b-9 in the test samples arestatistically similar to thepredicate (Exposure Control &Ref Material) control andstatistically lower than thepositive control for allexposure timesThe test sampleconcentrations of C3a andSC5b-9 were statisticallysimilar or lower than thepredicate control sampleconcentrations, andstatistically lower than thepositive control sampleconcentrations at all threeexposure timesPass
Dog ThrombogenicityThe device must be non-thrombogenic after 4 hours invivo when compared to acontrol device (BostonScientific Excelsior SL-10microcatheter)No significant thrombosiswith a Grade of 0 wasobserved in 2 out 2 test siteand 2 out of 2 control sites.Based on the evaluationcriteria, the amount ofthrombosis was notconsidered significantPass

EN ISO-10993 GLP Testing Summary

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "Penumbra" in a red font. To the right of the word is a red circle with a white "P" inside. The logo appears to be for a company or organization called Penumbra.

In summary non-clinical testing substantiates that the Penumbra System Reperfusion Catheter JET 7 device is non-cytotoxic, non-sensitizing, non-irritating, non-toxic, nonpyrogenic, non-hemolytic, and non-thrombogenic.

1.10.2 Bench-top Testing

The physical and mechanical properties of the Reperfusion Catheter JET 7 device was assessed using standard test methods and pre-determined acceptance criteria. The following tests were performed and all tests passed successfully:

AttributeSpecificationResults
Dimensional / VisualInspectionThese evaluations confirm that the units used in this DesignVerification testing meet all product specifications.Pass
Simulated Use[Intracranial Access, VesselAccess Entry Performance,Delivery/Retrieval Forces &Clot Removal]Simulated use testing of the Penumbra System Reperfusion CatheterJET 7 was performed with accessory devices in an anatomical modelwhich simulated the tortuosity of the neurovasculature. Devices weredelivered through the tortuous anatomical model to evaluate theeffectiveness of the devices to remove clots and that the ReperfusionCatheter does not collapse under vacuum.Pass
Physician Evaluation[Deliverability & ClotRemoval]Multiple Physician performance evaluation of the Penumbra SystemReperfusion Catheter JET 7 in a simulated neurovascular tortuositymodel with the predicate Penumbra System Reperfusion CatheterACE 68 used as a baseline.Pass
Kink Resistance(Distal, Midshaft, Proximal)No kinking when formed in a 2.5 mm, 10 mm, and 25 mm radiusPass

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "Penumbra" in a red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and clean, with the red color providing a sense of energy and the white "P" adding a touch of sophistication.

AttributeSpecificationResults
Particulate testing≥ 10 μm will be ≤ 6000 particlesPass
≥ 25 μm will be ≤ 600 particlesPass
≥ 75 μm will be measured for informational purposes only (FIPO)FIPO
≥ 125 μm will be measured for informational purposes only (FIPO)FIPO
Coating IntegrityCoating has not delaminated, peeled, or flaked prior to or aftersimulated use particulate testingPass
Markerband VisibilityThe markerband is fluoroscopically visiblePass
Hub Air AspirationNo leaks detected when vacuum is pulled on the injection lumenPass
Pressure Test45 psi for 30 sec minimumPass
JET 7 / Sheath or 8F GuideCatheter Friction ForceMaximum value per specificationPass
JET 7 / 0.014 in. GuidewireFriction ForceMaximum value per specificationPass
Joint sections bond strengthMinimum value per specificationPass
Markerband Section BondStrengthMinimum value per specificationPass
Hub to Shaft Bond StrengthMinimum value per specificationPass
Hub to Hypotube BondStrengthMinimum value per specificationPass
Elongation to failureElongation ≥ 5%Pass
TorsionNumber of turns will be recorded for informational purposes only.FIPO
CorrosionNo visible corrosion on Reperfusion Catheter immediately aftercorrosion testing procedurePass

1.10.3 Animal Study

The safety and performance of the Reperfusion Catheter JET 7 compared to the predicate Reperfusion Catheter ACE 68, when "wedged" in a vessel and using maximum aspiration, was evaluated in the accepted porcine vascular model. One side of each swine was treated with the Reperfusion Catheter JET 7 and the contralateral side was treated with predicate device. The purpose of this study was to evaluate the Reperfusion Catheter JET 7 aspiration vascular response when "wedged" within a swine artery compared to the predicate device. Safety of the test articles was assessed by consideration of the acute and chronic vascular response. Vascular response was assessed by contrast angiography and by gross necropsy and histopathology of associated vasculature performed by the Sponsor Pathologist. As testing included the predicate device, results were compared to demonstrate substantial equivalence.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in a red sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.

1.11 Performance Data – Clinical:

No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured the Predicate Device and other devices with similar dimensions used for direct aspiration. The literature review was used to support the determination of substantial equivalence by leveraging clinical outcomes from devices that are considered technologically equivalent.

1.12 Summary of Substantial Equivalence

The subject Penumbra System Reperfusion Catheter JET 7 device is substantially equivalent to the predicate device with regard to intended use, operating principle, design concept, materials, shelf-life, packaging and sterilization processes.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).