K Number
K241970

Validate with FDA (Live)

Date Cleared
2024-08-02

(28 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Powder Free Nitrile Examination Gloves (Blue) , Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The tested chemotherapy drugs are: Carmustine 3.3 mg/ml(3,300 ppm) 27.7 (35.5,27.7,27.7) Cisplatin 1 mg/ml (1,000 ppm) >240min Cyclophosphamide 20 mg/ml (20,000 ppm) >240min Dacarbazine 10mg/ml (10,000 ppm) >240min Doxorubicin HCl 2 mg/ml(2,000 ppm) >240min Etoposide 20 mg/ml (20,000 ppm) >240min Fluorouracil 50mg/ml(50,000 ppm) >240min Gemcitabine HCl 38 mg/ml (38,000 ppm) >240min Ifosfamide 50mg/ml (50,000 ppm) >240min Methotrexate 25mg/ml (25,000 ppm) >240min Mitomycin C 0.5 mg/ml(500 ppm) >240min Mitoxantrone HC1 2 mg/ml (2,000 ppm) >240min Paclitaxel 6 mg/ml (6,000 ppm) >240min Thiotepa 10 mg/ml (10,000 ppm) 59.4(67.5,59.4,65.9) Vincristine Sulfate 1 mg/ml (1,000 ppm) > 240min The tested non-chemotherapy drugs are: Fentanyl Citrate injection 100mcg/2ml >240min Note: Carmustine and Thiotepa have extremely low permeation times of 27.7 and 59.4 minutes respectively. Warning: Do Not Use with Carmustine, Thiotepa

Device Description

The Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the examiner. The gloves are blue color, powder free, nitrile gloves. The gloves are offered in six sizes: XS、S、M、L、XL、XXL, packed in a paper box.

The proposed gloves are designed and manufactured in accordance with the ASTM D6319-19 standard and are tested for use with chemotherapy drugs and Fentanyl per ASTM D6978-05. The proposed device was modification from the legally marketed (existing) device "Powder free Nitrile Examination Gloves (Blue, Violet Blue, White, Cobalt Blue)"/K202356, manufactured by "KANGLONGDA VIETNAM PROTECTION TECHNOLOGY COMPANY LIMITED", same as the sponsor of this submission.

There are two modifications, one is to add a large model XXL, and the other is to add the "Tested for Use with Chemotherapy Drugs and Fentanyl Citrate" indications for use than the legally marketed (existing) device. Testing has been conducted on the proposed device passed as per ASTM D6978-05, ASTM D6319-19, ASTM D5151-19, ASTMD 6124-06.

The materials and the manufacturing process technology are the same.

AI/ML Overview

The provided document (a 510(k) summary for medical gloves) does not contain information about a study that proves a device meets acceptance criteria related to AI/ML or human reader performance. The document is for "Powder Free Nitrile Examination Gloves" and pertains to mechanical and material properties, as well as resistance to chemical permeation.

Therefore, I cannot provide details on:

  1. A table of acceptance criteria and reported device performance for an AI/ML device.
  2. Sample size used for a test set or data provenance for an AI/ML device.
  3. Number of experts used to establish ground truth or their qualifications for an AI/ML device.
  4. Adjudication method for an AI/ML device test set.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study for human readers with and without AI.
  6. Standalone performance of an algorithm.
  7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for an AI/ML device.
  8. Sample size for the training set of an AI/ML device.
  9. How the ground truth for an AI/ML training set was established.

However, based on the document, I can present the acceptance criteria and performance for the glove device as non-clinical testing.


Acceptance Criteria and Study for Powder Free Nitrile Examination Gloves

This document describes the non-clinical testing performed on "Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate" to demonstrate its safety and effectiveness.

1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical Testing of Gloves):

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319-19Physical Dimensions TestLength (mm): XS/S/M/L/XL/XXL: ≥230Pass
Width (mm): XS: 70±10mm; S: 85±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmPass
Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Pass
ASTM D5151-19Watertightness Test for Detection of HolesBe free from holes when tested in accordance with ASTM D5151Pass
ASTM D412Physical Properties (Before Aging)Tensile Strength: 14MPa, minPass
Ultimate Elongation: 500% minPass
ASTM D412Physical Properties (After Aging)Tensile Strength: 14MPa, minPass
Ultimate Elongation: 400% minPass
ASTM D6124Powder ContentMeet the requirements of Less than 2mg per glovePass
ASTM D6978-05 (2019)Detect the Permeation time by Chemotherapy Drugs and Fentanyl of the gloveMeet the requirements of ASTM D6978-05. (Specific breakthrough times listed elsewhere in document)Pass
ISO 10993-10Irritation and SensitizationNon-irritating; Non-sensitizerUnder the conditions of the study, not an irritant and not a sensitizer
ISO 10993-5CytotoxicityNon CytotoxicUnder conditions of the study, not cytotoxic

2. Sample size used for the test set and the data provenance:

  • The document does not specify the exact sample sizes (e.g., number of gloves tested for each attribute) for the physical and chemical resistance tests. It only states that tests were conducted according to the listed ASTM and ISO standards.
  • Data Provenance: The testing was conducted to verify that the device meets design specifications and complies with international standards. The origin of the testing data is not explicitly stated beyond being "non-clinical tests."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This is not applicable as the device is a medical glove, not an AI/ML diagnostic or prognostic tool. Ground truth is established by objective physical and chemical measurements following established standards, not by expert consensus or physician reads.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable for this type of device testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable as this is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable as this is not an AI/ML device.

7. The type of ground truth used:

  • For the physical properties (dimensions, watertightness, tensile strength, powder content), the ground truth is based on direct physical measurement and standard test methods outlined in ASTM standards.
  • For chemical permeation (chemotherapy drugs and Fentanyl), the ground truth is established by analytical measurement of breakthrough time according to ASTM D6978-05.
  • For biocompatibility (irritation, sensitization, cytotoxicity), the ground truth is established by biological assays according to ISO 10993 standards.

8. The sample size for the training set:

  • Not applicable as this is not an AI/ML device. The "training" for glove manufacturing relates to process control and material specifications, not a data-driven model.

9. How the ground truth for the training set was established:

  • Not applicable as this is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

August 2, 2024

Kanglongda Vietnam Protection Technology Company Limited % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd. FangShan District Beijing, 102401 China

Re: K241970

Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (XS,S,M,L,XL,XXL) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, OPJ, ODO, LZC Dated: July 3, 2024 Received: July 5, 2024

Dear Ray Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241970

Device Name

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentany! Citrate

Indications for Use (Describe) The Powder Free Nitrile Examination Gloves (Blue) , Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The tested chemotherapy drugs are: Carmustine 3.3 mg/ml(3,300 ppm) 27.7 (35.5,27.7,27.7) Cisplatin 1 mg/ml (1,000 ppm) >240min Cyclophosphamide 20 mg/ml (20,000 ppm) >240min Dacarbazine 10mg/ml (10,000 ppm) >240min Doxorubicin HCl 2 mg/ml(2,000 ppm) >240min Etoposide 20 mg/ml (20,000 ppm) >240min Fluorouracil 50mg/ml(50,000 ppm) >240min Gemcitabine HCl 38 mg/ml (38,000 ppm) >240min Ifosfamide 50mg/ml (50,000 ppm) >240min Methotrexate 25mg/ml (25,000 ppm) >240min Mitomycin C 0.5 mg/ml(500 ppm) >240min Mitoxantrone HC1 2 mg/ml (2,000 ppm) >240min Paclitaxel 6 mg/ml (6,000 ppm) >240min Thiotepa 10 mg/ml (10,000 ppm) 59.4(67.5,59.4,65.9) Vincristine Sulfate 1 mg/ml (1,000 ppm) > 240min The tested non-chemotherapy drugs are: Fentanyl Citrate injection 100mcg/2ml >240min Note: Carmustine and Thiotepa have extremely low permeation times of 27.7 and 59.4 minutes respectively. Warning: Do Not Use with Carmustine, Thiotepa

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________ K241970

  • Date of Preparation: 07/31/2024 1.
    1. Submitter

KANGLONGDA Vietnam Protection Technology Company Limited

Lot CN 05, Viglacera - Phong Dien Industrial Park, Phong Hoa Commune, Phong Dien District, Thua Thien Hue Province, Vietnam, 49316

Contact Person: Yuxiang Yao Position: Batching Supervisor Tel: +86-13722853299

Email: yaoyuxiang 1212@126.com

    1. Submission Correspondent

Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401 Contact Person: Ray Wang Position: General Manager Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com

    1. Subject Device Identification
      Trade Name: Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

Common Name: Medical Examination Glove

Regulatory Information:

Classification: I

Product Code: LZA, LZC, OPJ, QDO

Regulation Number: 21 CFR 880.6250

Review Panel: General Hospital

{5}------------------------------------------------

5. Predicate Device Identification

Predicate Device:

510(k) Number: K202356

Product Name: Powder free Nitrile Examination Gloves (Blue, Violet Blue, White, Cobalt Blue)

Manufacturer: Kanglongda Vietnam Protection Technology Company Limited

Regulatory Information:

Classification: I

Product Code: LZA

Regulation Number: 21 CFR 880.6250

Review Panel: General Hospital

Common Name: Nitrile Patient Examination Gloves (Powder Free)

Reference Device:

510(k) Number: K240629

Product Name: Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl

Manufacturer: Xingyu Medical Tech Co., Ltd

Regulatory Information:

Classification: I

Product Code: LZA, LZC, OPJ, QDO

Regulation Number: 21 CFR 880.6250

Review Panel: General Hospital

Common Name: Nitrile Patient Examination Gloves (Powder Free)

Device Description 6.

The Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the examiner. The gloves are blue color, powder free, nitrile gloves. The gloves are offered in six sizes: XS、S、M、L、XL、XXL, packed in a paper box.

The proposed gloves are designed and manufactured in accordance with the ASTM D6319-19 standard and are tested for use with chemotherapy drugs and Fentanyl per ASTM D6978-05. The proposed device was

{6}------------------------------------------------

modification from the legally marketed (existing) device "Powder free Nitrile Examination Gloves (Blue, White, Cobalt Blue)"/K202356, manufactured by "KANGLONGDA VIETNAM Violet Blue, PROTECTION TECHNOLOGY COMPANY LIMITED", same as the sponsor of this submission.

There are two modifications, one is to add a large model XXL, and the other is to add the "Tested for Use with Chemotherapy Drugs and Fentanyl Citrate" indications for use than the legally marketed (existing) device. Testing has been conducted on the proposed device passed as per ASTM D6978-05, ASTM D6319-19, ASTM D5151-19, ASTMD 6124-06.

The materials and the manufacturing process technology are the same.

We selected K240629 as our reference device and K202356 as our predicate device. Detailed comparison information can be found in the 510k Summary.

  1. Indications For Use

The Powder Free Nitrile Examination Gloves(Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

DrugsConcentrationMinimum BreakthroughDetection Time
The tested chemotherapy drugs are:
Carmustine3.3 mg/ml(3,300 ppm)27.7(35.5,27.7,27.7)
Cisplatin1 mg/ml (1,000 ppm)>240min
Cyclophosphamide20 mg/ml (20,000 ppm)>240min
Dacarbazine10mg/ml (10,000 ppm)>240min
Doxorubicin HCl2 mg/ml(2,000 ppm)>240min
Etoposide20 mg/ml (20,000 ppm)>240min
Fluorouracil50mg/ml(50,000 ppm)>240min
Gemcitabine HCl38 mg/ml (38,000 ppm)>240min
Ifosfamide50mg/ml (50,000 ppm)>240min
Methotrexate25mg/ml (25,000 ppm)>240min
Mitomycin C0.5 mg/ml(500 ppm)>240min
Mitoxantrone HCl2 mg/ml (2,000 ppm)>240min
Paclitaxel6 mg/ml (6,000 ppm)>240min
Thiotepa10 mg/ml (10,000 ppm)59.4(67.5,59.4,65.9)
Vincristine Sulfate1 mg/ml (1,000 ppm)>240min

The tested drugs are:

{7}------------------------------------------------

The tested non-chemotherapy drugs are:
FentanylinjectionCitrate100mcg/2ml>240min

Note: Carmustine and Thiotepa have extremely low permeation times of 27.7 and 59.4 minutes respectively.

Warning: Do Not Use with Carmustine, Thiotepa

{8}------------------------------------------------

510k Summary

449S
400441Sep 1 the proper property and the comments of the careerA
441September 19, 2017 11:30œ13
mnari--------------œ
C)
111
***toﻟD
4041ro39321
. .Page 18 -13)1
11
5SCConcession and Concession Comparison5. .1
ITEM8. Technological Characteristic Comparison SummaryTable 1 Comparison of Technology CharacteristicsRemark
Product CodeProposed Device (K241970)Powder Free Nitrile ExaminationGloves (Blue), Tested for Use withChemotherapy Drugs and FentanylCitratePredicate Device(K202356)Powder free Nitrile Examination Gloves(Blue, Violet Blue, White, Cobalt Blue)Reference Device (K240629)Nitrile Disposable Examination Gloves,Tested for Use with Chemotherapy Drugsand Fentanyl
Regulation No.LZA, LZC, OPJ, QDO21 CFR 880.6250LZA21 CFR 880.6250LZA, LZC, OPJ, QDO21 CFR 880.6250SAME
ClassIIISAME
Intended UseThe Powder Free Nitrile ExaminationGloves (Blue), Tested for Use withChemotherapy Drugs and FentanylCitrate is a disposable device intendedfor medical purposes that is worn on theexaminer's hand or finger to preventcontamination between patient andexaminer.The Powder free Nitrile ExaminationGloves (Blue, Violet Blue, White,Cobalt Blue) is a disposable deviceintended for medical purposes that isworn on the examiner's hands to preventcontamination between patient andexaminer.The Nitrile Disposable ExaminationGloves, Tested for Use withChemotherapy Drugs and Fentanyl, is adisposable device intended for medicalpurposes that is worn on the examiner'shand or finger to prevent contaminationbetween patient and examiner. The testedchemotherapy drugs are:SAMESimilar 1
The tested chemotherapy drugs are:
DrugsConcentrationDrugsConcentrationMinimum BreakthroughDetection Time
The tested chemotherapy drugs are:
Bleomycin sulfateBleomycin sulfate(15.0 mg/ml 15,000 ppm)>240 min
CarboplatineCarboplatine(10.0 mg/ml 10,000 ppm)>240 min
CarmustineCarmustine (BCNU)(3.3mg/ml 3,300-ppm)26.7
Cisplatin1mg/ml (1,000ppm)Cisplatin(1.0-mg/ml. 1,000 ppm)>240 min
Cyclophosphamide(20.0 mg/ml. 20,000 ppm)Cyclophosphamide(20.0 mg/ml. 20,000 ppm)>240 min
Cytarabine, (Cytosine)(100.0 mg/mL 100,000 ppm)
Dacarbazine(10.0-mg/mL 10,000 ppm)Dacarbazine10mg ml (10,000 ppm)>240 min
Doxorubicin HCL(2.0-mg/ml 2,000 ppm)Doxorubicin HCl2mg/ml(2,000 ppm)>240 min
Etoposide (Toposar)+(20.0 mg/mL 20,000 ppm)Etoposide20mg/ml (20,000 ppm)>240 min
Fluorouracil (5-Flu)(50.0-mg/mL-50,000-ppm)Fluorouracil50mg/ml(50,000ppm)>240 min
Idarubicin(1.0-mg/ml 1,000 ppm)
Gemcitabine HCIGemcitabine HCI38mg/ml (38,000 ppm)>240 min
Ifosfamide(50.0 mg/mL 50,000 ppm)Ifosfamide50mg/ml (50,000 ppm)>240 min
Mechlorethamine HCL(1.0 mg/ml 1,000 ppm)
Methotrexate(25.0 mg/mL 25,000 ppm)Methotrexate25mg/ml (25,000 ppm)>240 min
Mitomycin CMitomycin C0.5 mg/ml(500 ppm)>240 min
Mitoxantrone HCL(2.0 mg/ml. 2,000 ppmeMitoxantrone HCL2 mg/ml (2,000 ppm)>240 min
Paclitaxel(6.0 mg/mL 6,000 ppm)Paclitaxel6 mg/ml (6,000 ppm)>240 min
Thiotepa(10.0 mg/mL 10,000 ppm)Thiotepa10mg/ml (10,000-ppm)37.6
Vincristine Sulfate(1.0 mg/mL 1,000 ppmVincristine Sulfate1 mg ml (1,000ppm)>240 min
The tested-non-chemotherapy drugs are:The tested non-chemotherapy drugs are:
MESNA(100.0 mg/mL 100,000 ppm)
Trisenox (Arsenic Trioxide)(1.0 mg/ml 1,000 ppm)
Fentanyl Citrate Injection(100 mcg/2mL)+
Note: Carmustine and Thiotep haveextremely low permeation times of 26.7and 37.6 minutes respectively.Note: Carmustine and Thiotepa haveextremely low permeation times of 27.7and 59.4 minutes respectively.

{9}------------------------------------------------

510k Summary

Warning: Do Not Use with Carmustine, ThiotepaPowder freeMaterialColorOTC useSterilityUseLabeling claimSize and Dimensions (mm)Thickness (mm)PhysicalPropertiesFreedom from HolesPowder ContentWarning: Do Not Use with Carmustine, Thiotepa.Powder freeMaterialColorOTC useSterilityUse
YesNitrileBlueYesNon-sterileSingle useTested for Use with Chemotherapy Drugs and FentanylLength:Min 230mm for size XS-XXLWidth:XS:70±10S: 80±10M: 95±10L: 110±10XL: 120±10XXL:130±10Finger:0.08-0.14±0.01Palm:0.06-0.10 ±0.01Before AgingAfter AgingBe free from holes when tested in accordance with ASTMD5151Meet the requirements of ASTM D6124YesNitrileBlue, Violet Blue, White, Cobalt BlueYesNon-sterileSingle use
Not Tested for use with Chemotherapy Drugs and FentanylLength:Min 230mm for size XS-XLWidth:XS:70±10S: 80±10M: 95±10L: 110±10XL: 120±10Finger:0.08-0.14±0.01Palm:0.06-0.10 ±0.01Tensile Strength: 14MPa, minUltimate Elongation:500% minTensile Strength: 14MPa, minUltimate Elongation:400% minBe free from holes when tested in accordance with ASTMD5151Meet the requirements of ASTM D6124
Tested for Use with Chemotherapy Drugs and FentanylLength:Min 220mm for size SMin 230mm for size M-XLWidth:S: 80±10M: 95±10L: 110±10XL: 120±10Finger: ≥0.05; Palm: ≥0.05
Tensile Strength: 14MPa, minUltimate Elongation:500% minTensile Strength: 14MPa, minUltimate Elongation:400% min

{10}------------------------------------------------

. Summary
0k
1ﺎ
U>
Chemotherapy and non-chemo therapy Drugs Tested with Minimum Breakthrough Detection Time as tested per ASTM D6978NASimilar 2Biocompatibility
Minimum Breakthrough Detection TimeDrugsConcentrationMinimum Breakthrough Detection Time
27.7(35.5,27.7,27.7)The tested chemotherapy drugs are:Carmustine3.3 mg/ml(3,300 ppm)>240 minIrritation and Sensitization:Under the conditions of the study, not an irritant and not a sensitizer
>240minCisplatin1 mg/ml (1,000ppm)>240 min
>240minCyclophosphamide20 mg/ml (20,000 ppm)>240 min
>240minDacarbazine10ng/ml (10,000 ppm)>240 min
>240minDoxorubicin HCL2 mg/ml(2,000 ppm)>240 min
>240rainEtoposide20 mag/ml (20,000 ppm)>240 min
>240minFluorouracil50ng/ml (50,000 ppm)>240 min
>240minGemcitabine HCL38mg/ml (38,000 ppm)>240 min
>240minIfosfamide50ng/ml (50,000 ppm)>240 min
>240minMethotrexate25mg/ml (25,000 ppm)>240 min
>240rainMitomycin-C0.5 mg ml(500 ppm)>240 minIrritation and Sensitization:Under the conditions of the study, not an irritant and not a sensitizer
>240minMitoxantrone HCl2mg/ml (2,000 ppm)>240 min
>240rainPaclitaxel6 mag/ml (6,000 ppm)>240 min
59.4(675,59.4.65.9)Thiotepa10 mag/ml (10,000 ppm)37.6
>240rainVincristine Sulfate1 mg/ml (1,000ppm)>240 min
>240minThe tested non-chemotherapy drugs are:
>240minFentanyl Citrate injection100mcg/2ml>240 min
Note: Carmustine and Thiotepa have extremely low permeation times of 27.7 and 59.4 minutes respectively.Warning: Do Not Use with Carmustine, Thiotepa
Note: Carmustine and Thiotepa have extremely low permeation times of 26.7 and 37.6 minutes respectively.Warning: Do Not Use with Carmustine, Thiotepa.ISO 10993-10, under the conditions of the study, not an irritant or a sensitizer
ISO 10993-11, under the condition of acute systemic toxicity test, the test article did not show acute systemic toxicity in vivo.
Cytotoxicity:Under conditions of the study, not cytotoxicCytotoxicity:Under conditions of the study, not cytotoxic

{11}------------------------------------------------

Similar 1-Indications for use

The proposed device and the reference device have same indications, only minor different drugs and the penetration time of carmustine and thiotepa between the proposed device and the predicate device, which will not raise any new safety or performance.

Different 1-Size and dimension, Thickness

The size and dimension, Thickness of the proposed device is not exactly same as the predicate and reference device. However, the size and dimension, thickness of the proposed device has been covered by ASTM D6319-19. The user can select appropriate model depended on size of user's hand. In addition, its dimension and thickness have been tested and met the requirement of ASTM D6319-19. Therefore, the difference will not affectiveness of the proposed device.

Similar 2-Chemotherapy and non-chemo therapy Drugs

Only minor differences on different drugs and the penetration time of carmustine and thiotepa between the proposed device and the reference device. And met the requirement of ASTM D6978-05. Therefore, the difference will not affect the safety and effectiveness of the proposed device.

  1. Summary of Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10: 2021 Biological Evaluation Of Medical Devices - Part 10: Tests For Skin Sensitization

ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.

ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application

ASTM D6124-06 (Reapproved 2022), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs

Test MethodPurposeAcceptance CriteriaResults
ASTM D6319-19Physical Dimensions TestLength (mm):XS/S/M/L/XL/XXL: ≥230Width(mm):XS: 70±10mm; S: 85±10mm;M: 95±10mm; L: 110±10 mm;XL: 120±10 mmXXL: 130±10 mmPass

Table 2 Summary of non clinical performance testing

{12}------------------------------------------------

Thickness (mm):Finger: ≥0.05; Palm: ≥0.05Pass
ASTM D5151-19Watertightness Test for Detection of HolesBe free from holes when tested in accordance with ASTM D5151Pass
ASTM D412Physical PropertiesBefore AgingTensile Strength:14MPa, minPass
Ultimate Elongation:500% minPass
After AgingTensile Strength:14MPa, minPass
Ultimate Elongation:400% minPass
ASTM D6124Powder ContentMeet the requirements of Less than 2mg per glovePass
ASTMD 6978-05 (2019)Detect the Permeation time by Chemotherapy Drugs and Fentanyl of the gloveMeet the requirements of ASTMD 6978-05.Pass
ISO 10993-10Irritation and SensitizationNon-irritating; Non- sensitizerUnder the conditions of the study, not an irritant and not a sensitizer
ISO 10993-5CytotoxicityNon CytotoxicUnder conditions of the study, not cytotoxic

10. Summary of Clinical Testing Not applicable

11. Conclusions

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed device, K202356.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.