Intended Use

The Broadway 8 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system.

Device Description

The Broadway 8 Catheter is a single lumen, flexible, variable stiffness catheter designed for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. It has a radiopaque marker band on the distal end and a luer hub at the proximal end. The Broadway 8 Catheter shaft has a lubricious hydrophilic coating at the distal end to reduce friction during use. It is packaged with a FastPass Delivery Assist Catheter, two rotating hemostasis valves (RHVs) and two peel-away introducer sheaths.

The FastPass Delivery Assist Catheter may be used in conjunction with the Broadway 8 Catheter to facilitate the introduction of the Broadway 8 Catheter into distal vasculature and aid in navigation to distal anatomy. The FastPass Delivery Assist Catheter has a lubricious hydrophilic coating at the distal end to reduce friction during use.

AI/ML Overview

The provided text is a 510(k) summary for the Stryker Neurovascular Broadway 8 Catheter. It outlines the device's characteristics and compares it to a predicate device to demonstrate substantial equivalence. However, this document does not describe acceptance criteria and a study that proves the device meets those criteria in the context of an AI/ML device.

The acceptance criteria tables and study details typically requested for AI/ML device performance (e.g., sensitivity, specificity, AUC, human reader improvement with AI assistance, ground truth establishment by multiple experts, sample sizes for training and test sets, etc.) are not applicable to this 510(k) summary, as the Broadway 8 Catheter is a physical medical device (a percutaneous catheter), not an AI/ML software device.

Therefore, I cannot fulfill your request for:

  • A table of acceptance criteria and the reported device performance related to AI/ML metrics. The performance data provided is for physical bench testing (e.g., tensile strength, particulate, kink resistance) and an animal study assessing vascular safety, which are relevant to a catheter but not to AI/ML performance.
  • Sample sized used for the test set and the data provenance, number of experts for ground truth, adjudication method, MRMC comparative effectiveness study, standalone performance, type of ground truth, training set sample size, and how training ground truth was established. These details are specific to the development and validation of AI/ML algorithms and are not found in the provided document for this physical device.

The document focuses on demonstrating substantial equivalence of the catheter to a predicate device through:

  • Bench Testing: Verifying physical properties, mechanical integrity, and material characteristics (e.g., Dimensional Verification, Tensile Strength, Particulate, Coating Integrity, Kink Resistance, Burst Pressure, Simulated Use Testing).
  • Animal Study: Assessing vascular safety and navigation performance in a swine model compared to the predicate device.
  • Biocompatibility Testing: Ensuring the materials are safe for biological interaction.

The "Conclusions" section reiterates that the device is substantially equivalent based on these aspects, not on AI/ML performance.

{0}------------------------------------------------

December 19, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Stryker Neurovascular Shazia Hakim Principal Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, California 94538

Re: K241768

Trade/Device Name: Broadway 8 Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: November 19, 2024 Received: November 20, 2024

Dear Shazia Hakim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Sara S. Thompson -S

for Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241768

Device Name Broadway 8 Catheter

Indications for Use (Describe)

The Broadway 8 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

SubmitterStryker Neurovascular47900 Bayside ParkwayFremont, CA 94538Phone: 510-413-2500
ContactShazia Hakim
Date PreparedJune 19, 2024
Device NameBroadway 8 Catheter
Common NamePercutaneous Catheter
Classification NameCatheter, Percutaneous, Neurovasculature (21 CFR 870.1250)
Product CodeQJP
Device ClassificationClass II
Predicate DeviceAXS Vecta Intermediate Catheter (AXS Vecta 74), K200206

Device Description

The Broadway 8 Catheter is a single lumen, flexible, variable stiffness catheter designed for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. It has a radiopaque marker band on the distal end and a luer hub at the proximal end. The Broadway 8 Catheter shaft has a lubricious hydrophilic coating at the distal end to reduce friction during use. It is packaged with a FastPass Delivery Assist Catheter, two rotating hemostasis valves (RHVs) and two peel-away introducer sheaths.

The FastPass Delivery Assist Catheter may be used in conjunction with the Broadway 8 Catheter to facilitate the introduction of the Broadway 8 Catheter into distal vasculature and aid in navigation to distal anatomy. The FastPass Delivery Assist Catheter has a lubricious hydrophilic coating at the distal end to reduce friction during use.

Indications for Use

The Broadway 8 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system.

{5}------------------------------------------------

Technological Characteristics and Product Feature Comparison

Stryker Neurovascular has demonstrated the Broadway 8 Catheter is substantially equivalent to the Predicate Device, AXS Vecta Intermediate Catheter based on the same or similar materials, similar design, and the same fundamental operating principles. A comparison of the Subject Device with the Predicate Device is summarized in Table 1 below.

DetailBroadway 8 Catheter(Subject Device)AXS Vecta Intermediate Catheter(AXS Vecta 74)(Predicate Device)
Regulation Number21 CFR 870.1250Same
ClassificationIISame
FDA Product CodeQJPQJP, DQY
510(k) NumberK241768K200206
Indications for UseThe Broadway 8 Catheter is indicatedfor use in facilitating the insertion andguidance of appropriately sizedinterventional devices into a selectedblood vessel in the neurovascularsystem.The AXS Vecta Intermediate Catheter isindicated for use in facilitating theinsertion and guidance of appropriatelysized interventional devices into aselected blood vessel in the peripheraland neurovascular systems. The AXSVecta Intermediate Catheter is alsoindicated for use as a conduit forretrieval devices.
MaterialsCommonly used medical grade plastics& metals with hydrophilic coating.Same
Strain ReliefNonePolyolefin
ReinforcementStainless SteelStainless Steel/Nitinol
CoatingHydrophilic CoatingHydrophilic Coating
Outer Diameter0.098 in.Distal OD:AXS Vecta 71: 0.082 in.AXS Vecta 74: 0.083 in.Proximal OD:AXS Vecta 71: 0.085 in.AXS Vecta 74: 0.087 in.
Inner Diameter0.084 in.AXS Vecta 71: 0.071 in.AXS Vecta 74: 0.074 in.
Effective Length(s)132 cm115, 125, 132 cm
Sterilization MethodEO SterilizationSame
How SuppliedSingle Use/SterileSame
DetailBroadway 8 Catheter(Subject Device)AXS Vecta Intermediate Catheter(AXS Vecta 74)(Predicate Device)
Accessory Devices Provided2 Hemostasis Valves, 2 Peel-AwayIntroducersFastPass Delivery Assist Catheter(ID: 0.016"; Distal OD: 0.033"; BulbOD: 0.075", Length: 156cm)Hemostasis Valve, 2 Peel-AwayIntroducersScout Introducer

Table 1: Substantial Equivalence Comparison

{6}------------------------------------------------

Risk Assessment

Risk assessment of the Broadway 8 Catheter has been conducted in accordance with EN ISO 14971 and Stryker Neurovascular risk management procedures. Stryker Neurovascular has determined that the Broadway 8 Catheter raises no new questions of safety or effectiveness. The risk assessment conducted on the Subject Device did not result in any new hazards or harms, and no significant changes to the occurrences of harm were identified when compared with the Predicate Device.

Testing Summary

Performance Data - Bench

The results of design verification and design validation testing conducted on the Broadway 8 Catheter demonstrate that it performs as designed, is suitable for the indication for use, and is substantially equivalent to the legally marketed Predicate Device. The design verification and design validation bench testing are summarized in Table 2 below:

TestTest Method SummaryConclusions
Dimensional VerificationVerification of critical dimensions.Pass - Samples met establishedacceptance criteria.
Tensile StrengthTensile strength tested per ISO 10555-1.Pass - Samples met establishedacceptance criteria.
Torsional Bond StrengthThe test specimens were rotated to evaluateintegrity after rotation.Pass - Samples met establishedacceptance criteria
Table 2: Performance Data - Bench Testing
TestTest Method SummaryConclusions
ParticulateParticulate was collected from simulated usein a tortuous model, characterized by sizeand particulate type, and compared tocommercially available product.Pass - Device comparable topredicate.
Coating IntegrityHydrophilic coating was inspected beforeand after simulated use in a tortuous model.Pass – Device comparable topredicate
Liquid LeakLiquid leak tested per ISO 10555-1.Pass - Samples met establishedacceptance criteria.
Air LeakAir leak tested per ISO 10555-1.Pass - Samples met establishedacceptance criteria.
Kink ResistanceMeasurement of catheter kink radiusPass - Samples met establishedacceptance criteria
Burst PressureBurst Pressure was tested per ISO 10555-1.Pass - Samples met establishedacceptance criteria
Liner LaminationVerification of inner liner adherence followingsimulated use.Pass - Samples met establishedacceptance criteria
Chemical CompatibilityCatheter was measured and visuallyinspected before and after exposure tovarious clinically relevant chemicals.Pass - Samples met establishedacceptance criteria
Tip BucklingMeasurement of distal catheter bucklingforce.Pass - Samples met establishedacceptance criteria
Product Integrity Post-RemovalCatheter and accessories were inspectedfollowing removal from packaging.Pass - Samples met establishedacceptance criteria.
Simulated Use TestingSimulated use testing in a bench anatomicalmodel using worst-case interventional deviceswith femoral artery access.Pass - Samples met establishedacceptance criteria.
Friction Force TestingCoating lubricity measured by a pinch test andcompared to predicate device.Pass – Device comparable topredicate

{7}------------------------------------------------

{8}------------------------------------------------

Performance Data - Animal Study

An animal study was conducted in compliance with FDA Good Laboratory Practice (GLP) Regulation (21 CFR Part 58) at 3-day and 30-day timepoints to assess the navigation to the target artery (3 passes in each artery), angiography, necropsy, and histopathology of the Subject Device, Broadway 8 Catheter, compared to the Predicate Device, AXS Vecta 74 Catheter. The vessels evaluated in the swine model included the ascending pharyngeal artery (APA), internal maxillary artery (IMAX), internal mammary artery (IMA), superficial cervical artery (SCA), and axillary artery. All study endpoints were met and the Broadway 8 Catheter (Test Article) performed as well as or better than the AXS Vecta 74 catheter (Control Article), with no vessel damage observed at the treatment or terminal procedures. The study results demonstrated vascular safety of the Broadway 8 Catheter. This study supports substantial equivalence of the Broadway 8 Catheter to the Predicate Device as evidenced by the absence of vascular damage at the treatment sites during the GLP safety study.

Performance Data - Clinical

No clinical study was conducted as bench testing was determined sufficient for verification and validation purposes.

Sterilization and Shelf Life

The Broadway 8 Catheter and all device accessories are sterilized with 100% Ethylene Oxide to a sterility assurance level (SAL) of 106. The Subject Device and accessories meet the requirements for EO residuals per EN ISO 10993-7 for a limited contact delivery system - externally communicating. The Broadway 8 Catheter and accessories are provided for single use only.

The labeled shelf life for the Broadway 8 Catheter is 1-year. Shelf-life testing (product and packaging) and Distribution Shipping Challenge Conditioning testing were performed on the Subject Device, and the results met established criteria.

Biocompatibility

The Broadway 8 Catheter is categorized as a limited exposure (≤24 hours), externally communicating device with circulating blood contact in accordance with ISO 10993-1 and FDA Guidance, "Use of International ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation of testing within a risk management process." All patient contacting components used in the Broadway 8 Catheter have a history of safe clinical use throughout the medical device industry and/or within Stryker Neurovascular and have supporting biocompatibility data. The results of the biocompatibility testing are summarized in Table 3 below.

{9}------------------------------------------------

TestTest Method SummaryConclusion
Extract and Direct HemolysisISO 10993-4Non-hemolytic
ThromboresistanceISO 10993-4Non-thrombogenic
SC5b-9 Complement Activation AssayISO 10993-4Non-activator
ASTM Partial Thromboplastin TimeISO 10993-4No enhancement or inhibitionof coagulation system
Cytotoxicity Study Using the ISO ElutionMethodISO 10993-5Non-cytotoxic
ISO Maximization Sensitization StudyISO 10993-10No evidence of sensitization
ISO Acute Systemic Toxicity Study –ExtractISO 10993-11No mortality or evidence ofacute systemic toxicity
Material Mediated PyrogenicityUSP <151>Non-pyrogenic reactivity
ISO Intracutaneous Reactivity Study –ExtractISO 10993-23Non-irritant
Table 3: Biocompatibility Tests and Results
-------------------------------------------------------

Conclusions

The Broadway 8 Catheter is substantially equivalent to the Predicate Device based on similar intended use / indications for use, same or similar materials, same fundamental design, and the same operating principles. The conclusions drawn from risk assessments, the bench testing conducted using the Subject Device compared to the Predicate Device, and an animal safety study demonstrate that the Subject Device is suitable for the indication for use with the associated QJP Product Code. Additionally, the testing results summarized above along with the risk assessment demonstrate that the benefits of the device outweigh any residual risks when used in accordance with device Instructions for Use. Stryker Neurovascular has demonstrated that the Broadway 8 Catheter is substantially equivalent to the legally marketed Predicate Device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).