K Number
K241708

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-08-06

(54 days)

Product Code
Regulation Number
870.1340
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Abiomed 14Fr Low Profile Introducer Kit is intended to facilitate access to the vascular system for the introduction and removal of the Impella CP Catheter and ancillary devices.

Device Description

The Abiomed 14Fr Low Profile Introducer Kit (Gen 2) is a sterile, single-use, prescription device. The 14Fr Low Profile Introducer Kit consists of an introducer sheath and a tapered sheath dilator which is compatible with an 0.035" guidewire. The 14Fr Low Profile Introducer Kit is kitted with an 0.035" access guidewire, supplemental dilators and a luer adapter for convenience to help facilitate insertion. The 14Fr Low Profile Introducer Sheath consists of a sheath hub with three-way stopcock and flush port at its proximal end a sheath body at its distal end. The sheath hub features an introducer cap, hemostasis valve, side-port with three-way stopcock and flush port, a butterfly (suture pad), and connects to the dilator hub. The coil reinforced polymer sheath body has an insertion profile of 14Fr to allow the insertion and removal of the Impella CP Catheter and ancillary devices. The 14Fr Low Profile Dilator consists of a dilator body, a tapered tip at the distal end, and a hub at the proximal end, which connects with the sheath hub. Additionally, the 14Fr dilator has a hydrophilic coating to aid in insertion of the device into the vasculature.

AI/ML Overview

This looks like a 510(k) summary for a medical device, specifically an introducer kit. The document describes the device, its intended use, and comparison to a predicate device. However, it does not contain the level of detail regarding acceptance criteria, study design, and performance metrics that would be expected for a typical AI/ML-based medical device described in the prompt's request.

Here's why and what information can be extracted:

This is NOT an AI/ML Medical Device.
The device described, the "Abiomed 14Fr Low Profile Introducer Kit," is a physical medical device (a catheter introducer) and not an AI/ML software device. This means the requested information about AI model performance, training sets, ground truth establishment methods, human reader studies, etc., is not applicable to this document. The "performance testing" section refers to traditional bench tests for physical device characteristics (e.g., visual inspection, mechanical testing, coating integrity, sterility, biocompatibility).

Therefore, I cannot populate most of the requested fields accurately as they pertain to AI/ML models.

However, I can extract the relevant information regarding the device's acceptance criteria and the type of study performed to show it met these criteria, as presented in the document for this physical device.

Information Extracted (Interpreted for a non-AI/ML device):

  1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria for a Physical Device: The document states that performance testing "met the acceptance criteria." While specific numerical criteria are not listed, the implied criteria would be successful completion of each bench test without failure.
    • Reported Device Performance: "showed that the subject device, Abiomed 14Fr Low Profile Introducer Kit (Gen2), met the acceptance criteria and demonstrated substantial equivalence for its intended use."
    Acceptance Criteria (Implied for physical device)Reported Device Performance
    Device passes specified bench testsMet acceptance criteria
    Demonstrates substantial equivalenceDemonstrated substantial equivalence for intended use
    Biocompatibility: Safe for patient contactPassed ISO 10993-1
    No new safety or performance issues identifiedNone identified
  2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for individual tests.
    • Data Provenance: Not applicable as these are bench tests on manufactured device samples, not patient data.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth for a physical device is typically established through engineering specifications, material science, and regulatory standards, not expert consensus on patient data.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-assisted device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI algorithm. The device performance (standalone) was assessed through bench testing.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For a physical device, ground truth is based on engineering specifications, material properties, and established regulatory standards (e.g., ISO 10993 for biocompatibility, defined mechanical test parameters).
  8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device.
  9. How the ground truth for the training set was established:

    • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 6, 2024

Abiomed Inc. Anshul Shah Director, Regulatory Affairs 22 Cherry Hill Drive Danvers, Massachusetts 01923

Re: K241708

Trade/Device Name: Abiomed 14Fr Low Profile Introducer Kit Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: June 6, 2024 Received: June 13, 2024

Dear Anshul Shah:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Page

2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Nicole M. Gillette -S

Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K241708

Device Name Abiomed 14Fr Low Profile Introducer Kit

Indications for Use (Describe)

The Abiomed 14Fr Low Profile Introducer Kit is intended to facilitate access to the vascular system for the introduction and removal of the Impella CP Catheter and ancillary devices.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo with a grid of four squares. The top-left and bottom-left squares are blue, while the top-right and bottom-right squares are maroon. The squares are arranged in a 2x2 grid, creating a simple and abstract design.

Abiomed 14Fr Low Profile Introducer Kit 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807.92.

A. Application Information:

Date Prepared:July 30, 2024
Submitter's Name & Address:ABIOMED, Inc.22 Cherry Hill DriveDanvers, MA 01923
Contact Person:Anshul ShahDirector, Regulatory AffairsPh: 323-541-5318E-mail: anshah@abiomed.com

B. Device Information:

Trade or Proprietary Name:Abiomed 14Fr Low Profile Introducer Kit
Common or Usual Name:Introducer, Catheter
FDA Classification:Class II, DYB, 21 CFR 870.1340
Regulation Description:Catheter Introducer

ﻥ Predicate Device:

The primary predicate device was the Abiomed 14Fr Low Profile Introducer Set, which is cleared under K222113.

D. Device Description:

The Abiomed 14Fr Low Profile Introducer Kit (Gen 2) is a sterile, single-use, prescription device. The 14Fr Low Profile Introducer Kit consists of an introducer sheath and a tapered sheath dilator which is compatible with an 0.035" guidewire. The 14Fr Low Profile Introducer Kit is kitted with an 0.035" access guidewire, supplemental dilators and a luer adapter for convenience to help facilitate insertion.

The 14Fr Low Profile Introducer Sheath consists of a sheath hub with three-way stopcock and flush port at its proximal end a sheath body at its distal end. The sheath hub features an introducer cap, hemostasis valve, side-port with three-way stopcock and flush port, a butterfly (suture pad), and connects to the dilator hub. The coil reinforced polymer sheath body has an insertion profile of 14Fr to allow the insertion and removal of the Impella CP Catheter and ancillary devices. The 14Fr Low Profile Dilator consists of a dilator body, a tapered tip at the distal end,

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo with a grid-like pattern. The logo consists of four squares arranged in a 2x2 grid. The top-left and bottom-left squares are blue, while the top-right and bottom-right squares are red. The squares have rounded corners and are slightly separated from each other.

and a hub at the proximal end, which connects with the sheath hub. Additionally, the 14Fr dilator has a hydrophilic coating to aid in insertion of the device into the vasculature.

E. Indications for Use:

INDICATIONS FOR USE:

The Abiomed 14Fr Low Profile Introducer Kit is intended to facilitate access to the vascular system for the introduction and removal of the Impella CP Catheter and ancillary devices.

F. Technological Characteristics Comparison of Subject and Predicate Devices:

The subject device, 14Fr Low Profile Introducer (Gen2), is identical to the predicate device in Intended Use, general system components, sterilization, guidewire compatibility, insertion profile, and general mechanism of action. Differences between the subject device and the predicate devices were all determined to be minor, have no adverse impact on safety or effectiveness, and raise no different questions of safety or effectiveness compared to the predicate devices. These minor differences include sheath materials, sidearm design, and hub component.

PropertyPrimary PredicateSubject Device
Manufacturer/Model Name/510k ClearanceAbiomed 14Fr Low Profile Introducer Set(Gen1)/K222113Abiomed 14Fr Low Profile Introducer Kit(Gen2)/this submission
Intended UseThe 14Fr Low Profile Introducer Set isintended to facilitate access into thevascular system.The 14Fr Low Profile Introducer Kit isintended to facilitate access into thevascular system.
Indications forUseThe 14Fr Low Profile Introducer Set isintended to facilitate access to the vascularsystem for the introduction and removal ofthe Impella CP Catheter and ancillarydevices.The 14Fr Low Profile Introducer Kit isintended to facilitate access to the vascularsystem for the introduction and removal ofthe Impella CP Catheter and ancillarydevices.
General SystemComponentsA sheath consisting of hub with hemostasis valve, sidearm with stopcock, and a dilator withluer.
SterilizationEthylene Oxide
Guidewire0.035" (or smaller) compatible guidewire
Length13, 25 cm13, 25 cm
Insertion Profile14Fr14Fr
SheathStainless steel reinforced Pebax tubeStainless steel reinforced Pebax tube
HubCap (polymer) w/ hemostasis valve(polymer),sidearm (polymer) w/ stopcock(polymer), suture pad (polymer)Cap (polymer) w/ hemostasis valve(polymer),sidearm (polymer) w/ stopcock(polymer), suture pad (polymer)

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo consisting of eight circles arranged in a 2x4 grid. The circles are colored in two different shades: blue and red. The four circles on the left side of the logo are blue, while the four circles on the right side are red. The circles are arranged in a way that creates a visually appealing and balanced design.

PropertyPrimary PredicateSubject Device
Manufacturer/Model Name/510k ClearanceAbiomed 14Fr Low Profile Introducer Set(Gen1)/K222113Abiomed 14Fr Low Profile Introducer Kit(Gen2)/this submission
DilatorTapered dilator (polymer) w/ Luer (polymer)with hydrophilic coating at dilator tipTapered dilator (polymer) w/ Luer (polymer)with hydrophilic coating at dilator tip
CoatingHydrophilic coating at dilator tipHydrophilic coating at dilator tip
Mechanism ofAction, GeneralInserted manually using standardtechniques. Sheath left indwelling (in vessel)after insertion and dilator removal.Inserted manually using standardtechniques. Sheath left indwelling (invessel) after insertion and dilator removal.
RadiopacityRadiopaque (sheath has internal metal coil)Radiopaque (sheath has internal metal coil)

G. Performance Testing:

The following performance testing was conducted on the Abiomed 14Fr Low Profile Introducer Kit (Gen2).

Bench Testing:

  • Visual Inspection and Dimensional Verification ●
  • . Sheath System Verification
  • Simulated Use Testing Assessment .
  • Packaging Validation Assessment .
  • Biocompatibility Testing Assessment .
  • . Sterilization Assessment
  • . Leak Testing
  • . Mechanical Testing
  • Coating Integrity ●
  • Particulate ●
  • Radiopacity Assessment .

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a logo with four squares arranged in a 2x2 grid. The top-left and bottom-left squares are blue, while the top-right and bottom-right squares are red. There is a small trademark symbol in the top-right corner of the logo.

H. Conclusions:

Performance testing (bench) was completed and showed that the subject device, Abiomed 14Fr Low Profile Introducer Kit (Gen2), met the acceptance criteria and demonstrated substantial equivalence for its intended use. Biocompatibility safety testing conducted in accordance with ISO 10993-1 demonstrates that the device is safe for its patient contact and duration. No clinical data were required to demonstrate substantial equivalence. No new safety or performance issues were identified during the testing; therefore, the subject device may be considered substantially equivalent to the predicate device.

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).