AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Horizon® 3.0 TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

Device Description

The Horizon® 3.0 TMS Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation. ("TMS").

The Horizon® 3.0 TMS Therapy System is a non-invasive tool for the stimulation of cortical neurons for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

The Horizon® 3.0 TMS Therapy System is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient and outpatient settings, including physicians' offices, clinics, and hospitals.

The Horizon® 3.0 TMS Therapy System configurations are an integrated system consisting of a combination of hardware, software, and configurations:

  • Horizon 3.0 Inspire (Subject of this submission).
  • Horizon 3.0 (Previously cleared under K232235, K223154, K222171 and K211389).
  • Horizon 3.0 with StimGuide Pro (Cleared under K232235).

All three configurations, including the subject Horizon 3.0 Inspire configuration, have identical intended use/indications for use, common specifications, equivalent performance and equivalent composition. All three devices share equivalent technological characteristics and principles of operation.

All configurations are composed from the following main components:

  • Stimulating Unit & Power Supply
  • User Interface
  • Applicating Coil for Motor Threshold
  • Applicating Coil for Treatment Delivery
  • System and Applicating Coil Cart and Holding Arm

The Horizon 3.0 with StimGuide Pro specifically includes a stereotactic infrared tracking system for aiding coil positioning.

AI/ML Overview

The provided text is a 510(k) summary for the Horizon 3.0 TMS Therapy System, specifically introducing the "Horizon 3.0 Inspire" configuration. This document primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device (K232235, Horizon 3.0 TMS Therapy System) rather than presenting a standalone study with defined acceptance criteria and performance metrics against a specific clinical outcome.

Therefore, the requested information, particularly regarding acceptance criteria for clinical performance, sample sizes for test sets where ground truth was established by experts, and effect sizes for MRMC studies, is not present in this document. The document describes non-clinical testing to demonstrate that the new configuration is as safe and effective as the predicate device.

Here's a breakdown of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance:

The document describes compliance with various safety and performance standards, but does not present a table of acceptance criteria with corresponding performance results in a way that would typically be seen for a new clinical efficacy study. Instead, it refers to demonstrating equivalence to a predicate device through non-clinical testing.

Acceptance Criteria CategoryDescription of Compliance/Performance
Electromagnetic Compatibility (EMC)Compliant "to the requirements of IEC 60601-1-2 to demonstrate an equivalent level of Electromagnetic Compatibility as the primary predicate device." Test plan utilized the same performance criteria as the primary predicate.
Electrical & Mechanical SafetyCompliant "to the requirements of IEC 60601-1 (including IEC 60601-1-8) to demonstrate an equivalent level of Electrical and Mechanical safety as the primary predicate device."
Thermal SafetyCompliant "to the requirements of IEC 60601-1 (including IEC 60601-1-6 and IEC 60601-1-8) to demonstrate an equivalent level of Thermal safety as the primary predicate device." Additional testing ensured the Horizon Air Film Coil could execute OCD, iTBS, and rTMS protocols at worst-case ambient conditions without excessive temperatures. The system can perform recommended protocols safely and effectively under worst-case simulated ambient conditions (80%-100% machine output, 30°C ambient).
Software Verification & ValidationSoftware developed following IEC 62304. Documentation demonstrates appropriate function and that software cannot contribute to unacceptable risk. Changes for "Inspire" configuration verified for correctness and no negative impact on predicate configurations. Cybersecurity testing performed to ensure "Inspire" configuration is as secure as the primary predicate.
Usability/Human Factors EngineeringEvaluation demonstrates appropriate human factors and usability for intended use, free from unacceptable use-related risks (IEC 60601-1-6, IEC 62366-1, ANSI/AAMI HE75). Consistency of workflows, UIs, and components from predicate contribute to a healthy usability profile.
Magnetic Pulse Output Testing and Magnetic & Electrical Field TestingElectric field distribution measured in a human head phantom filled with physiologic saline solution. Values of Magnetic and Electric field obtained by measuring induced voltage. Results demonstrated "equivalent Power Outputs, Electric Fields between the subject device configuration applicating coil and the primary predicate device configuration applicating coil." Key characteristics (stimulator output, field spatial distribution, E-Field decay, output waveforms, rate of change) were equivalent, especially at clinically relevant depths.
Safety Feature TestingIEC 60601-1 testing of fault scenarios to ensure appropriate safety features. System-level testing (software verification) exercised various interlocks. Testing introduced failure modes to confirm hardware backup interlocks. Demonstrates substantial equivalence of safety features.
Acoustic TestingIn simulated use scenarios (OCD Protocol at maximum machine output), the Horizon 3.0 "Inspire" configuration demonstrated equivalent acoustic properties to the predicate and fell below safe exposure limits.

2. Sample size used for the test set and the data provenance:

  • Test Set: No patient-specific test set for clinical performance is mentioned. The testing described is non-clinical/bench testing.
  • Data Provenance: The data provenance is from laboratory testing performed by the manufacturer, comparing the new configuration's performance to the primary predicate device and relevant standards. This is retrospective in the sense that it relies on existing standards and predicate device data, but the tests themselves were conducted on the new device configuration.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

This information is not applicable as the document describes non-clinical testing against engineering standards and comparison to a predicate device, not a human reader study where experts establish ground truth for a clinical dataset.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable for the same reasons as point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC comparative effectiveness study was done or described. This is a device for Transcranial Magnetic Stimulation, not typically an imaging AI device that aids human readers in interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device itself is a standalone medical device for delivering TMS therapy. The described testing is of the device's technical specifications and safety profile, not of an "algorithm only" in the context of an AI diagnostic. The software components were verified and validated as part of the overall device system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

For the non-clinical tests described, the "ground truth" (or reference) used were:

  • International standards: (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 62366-1, ANSI/AAMI HE75, ISO 10993-1, ISO 10993-10, ISO 14971, IEC 62304, AAMI TRS7, AAMI TIR97).
  • Performance of the legally marketed predicate device (K232235): The goal was to demonstrate "equivalent" safety and effectiveness.
  • Phantom head model: For Magnetic and Electrical Field Testing, a human head phantom model filled with physiologic saline solution was used to measure induced voltage.

8. The sample size for the training set:

This information is not applicable. This document does not describe the development or training of a machine learning algorithm; it describes a hardware and software system for TMS therapy.

9. How the ground truth for the training set was established:

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 30, 2024

The Magstim Company Limited Daniel Gregory Principal Systems Engineer Spring Gardens Whitland Carmarthenshire, SA34 0HR United Kingdom

Re: K241518

Trade/Device Name: Horizon® 3.0 TMS Therapy System (Horizon 3.0 Inspire); Horizon® 3.0 TMS Therapy System (Horizon 3.0 with StimGuide Pro); Horizon® 3.0 TMS Therapy System (Horizon 3.0) Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP; QCI Dated: May 29, 2024 Received: May 29, 2024

Dear Daniel Gregory:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vivek J. Pinto -S

Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241518

Device Name

Horizon® 3.0 TMS Therapy System (Horizon 3.0 Inspire);

Indications for Use (Describe)

Horizon 3.0 TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Prepared on: 2024-05-29

Contact Details21 CFR 807.92(a)(1)
Applicant NameThe Magstim Company Limited
Applicant AddressSpring Gardens Whitland Carmarthenshire SA34 0HR United Kingdom
Applicant Contact Telephone+441994240798
Applicant ContactMr. Daniel Gregory
Applicant Contact Emaildaniel.gregory@magstim.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameHorizon® 3.0 TMS Therapy System (Horizon 3.0 Inspire);Horizon® 3.0 TMS Therapy System (Horizon 3.0 with StimGuide Pro);Horizon® 3.0 TMS Therapy System (Horizon 3.0)
Common NameRepetitive transcranial magnetic stimulation system
Classification NameTranscranial Magnetic Stimulator
Regulation Number882.5805
Product Code(s)OBP, QCI - 21 C.F.R. § 882.5802
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K232235Horizon 3.0 TMS Therapy SystemOBP
K182853Horizon TMS Therapy SystemOBP
Device Description Summary21 CFR 807.92(a)(4)
The Horizon® 3.0 TMS Therapy System is a computerized, electromechanical medical device thatproduces and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, orpulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of thecerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to thehead is known as Transcranial Magnetic Stimulation. ("TMS").
The Horizon® 3.0 TMS Therapy System is a non-invasive tool for the stimulation of cortical neuronsfor the treatment of Major Depressive Disorder in adult patients who have failed to achievesatisfactory improvement from prior antidepressant medication in the current episode, as well as anadjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).
The Horizon® 3.0 TMS Therapy System is used for patient treatment by prescription only under thesupervision of a licensed physician. It can be used in both inpatient and outpatient settings, includingphysicians' offices, clinics, and hospitals.
The Horizon® 3.0 TMS Therapy System configurations are an integrated system consisting of a combination of hardware, software, and

The Horizon 3.0 TMS Therapy System Conngurations are an integrated system configurations:

{5}------------------------------------------------

  • Horizon 3.0 Inspire (Subject of this submission).

  • Horizon 3.0 (Previously cleared under K232235, K223154, K222171 and K211389).

  • Horizon 3.0 with StimGuide Pro (Cleared under K232235).

The three tiers of system offer equivalent safety and effectiveness with the main purpose allowing for physician offices, clinics and hospitals to choose a configuration that suits the organizational needs and provide different entry levels to promote the accessibility of TMS Therapy Treatments to health delivery organizations.

All three configurations, including the subject Horizon 3.0 Inspire configuration, have identical intended use/indications for use, common specifications, equivalent performance and equivalent composition. All three devices share equivalent technological characteristics and principles of operation.

All configurations are composed from the following main components:

- Stimulating Unit & Power Supply

  • User Interface
  • Applicating Coil for Motor Threshold
  • Applicating Coil for Treatment Delivery
  • System and Applicating Coil Cart and Holding Arm

The Horizon 3.0 with StimGuide Pro specifically includes a stereotactic infrared tracking system for aiding coil positioning.

Intended Use/Indications for Use

Horizon 3.0 TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patieve to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

Indications for Use Comparison

The indications for use for the subject device and the primary predicate device (K232235) are identical.

The reference predicate, K182853, where some of the subject device components have been re-used from was previously cleared for the treatment of Major Depressive Disorder in adult patieve satisfactory improvement from prior antidepressant medication in the current episode. The performance data within this submission, such as the applicating coil comparison to the primary predicate, demonstrate why these components do not raise any differing questions of safety or effectiveness when also used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

Technological Comparison

Introduction:

The Horizon® 3.0 TMS Therapy System is offered in three system configurations, the third of which is the subject of this submission:

Horizon 3.0, Horizon 3.0 with StimGuide Pro and Horizon 3.0 Inspire.

All system configurations, including the subject Horizon 3.0 Inspire configuration, feature the same core components and operating principles that drive the device. All three configurations in subject Horizon 3.0 Inspire configuration, have identical intended use/indications for use, common specifications, equivalent composition. All three devices share equivalent technological characteristics and principles of operation, substantiated below.

Principles of Operation:

Horizon 3.0 Inspire is identical to the primary principles of operation. Horizon 3.0 Inspire is capable of delivering and provides the same recommended treatment protocols (OCD, TTMS and iTBS) as the primary predicate device.

The method for coll positioning is also identical to the primary predicate, with the Horizon 3.0 "Inspire" following the same principle of the statistical distance from the Motor Threshold (MT) Hotspot to the chosen treatment (4cm from MT Hotspot, Dorsomedial Prefrontal Cortex, for OCD and 5.5cm from MT Hotspot, Left Dorsolateral Prefrontal Cortex, for MDD).

The principles of operation with respect to MT Determination remain identical to the primary predicate device - either by using visual qualitative monitoring for APB response or quantitative data provided by EMG.

In summary, the principles of operation of the subject Horizon 3.0 Inspire configuration remain identical to the primary predicate device configurations - Horizon 3.0 and Horizon 3.0 with StimGuide Pro.

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

{6}------------------------------------------------

Technological Characteristics:

The differences between Horizon 3.0 Inspire and the previously cleared device configurations from the primary predicate device are listed below. The intent of which is to reduce the cost and footprint for the end user whilst also providing an equivalent level of safety and effectiveness as the predicate device configurations. The three tiers of system offectiveness with the main purpose allowing for physician offices, clinics and hospitals to choose a configuration that suits the organizational needs and provide different entry levels to promote the accessibility of Therapy Treatments to health delivery organizations.

  • The Horizon 3.0 Inspire Mainframe features a different "interface" panel on the mainframe. This is to support the connection to the Horizon 3.0 Inspire Touchscreen. The mainframe supplies power and communications via this interface. The fundamental principles of the Horizon 3.0 Inspire Main identical to the predicate Horizon 3.0 Mainframe. The maximum output voltage, frequency, biphasic waveform, charging circuity and protocol performance remain identical across all configurations. The only change was interface related to mount a Ul touchscreen on the Mainframe.

  • The Horizon 3.0 Interface Unit from the primary predicate device configurations is not required for the Horizon 3.0 Inspire configuration. It's functions have been into other components. The PC has been moved from the Horizon 3.0 Interface Unit into the Horizon 3.0 Inspire Touchscreen, the EMG Amplifier has been moved into the Horizon 3.0 MEP Pod and some device control mechanisms previously performed via Horizon 3.0 Interface Unit now reside in the Horizon 3.0 Inspire Mainframe. In summary, the functions provided remain identical to the primary predicate and the technological changes were purely a reorganization of these functions to reduce the footprint of the system. Components such as the EMG Amplifier within the new Horizon 3.0 MEP Pod is the same module that is used on the primary predicate configurations.

  • The Horizon 3.0 Inspire Touchscreen as described above now contains the PC that previously resided in the Horizon 3.0 Interface Unit. The technical specification is equivalent to the Horizon 3.0 Interface Unit. Further, the Operating System and U Software executing on the touchscreen are the same as that which executed on the Horizon 3.0 Interface Unit. The software was modified sightly to detect a Horizon 3.0 Inspire configuration and set up device communications accordingly. In summary, the functions provided remain identical to the primary predicate and the technological changes were purely a reorganization of the footprint of the system.

  • The Horizon 3.0 Inspire Mains Filters slightly from the standard Horizon 3.0 Mains Filter of the primary predicate device to ensure the Horizon 3.0 Inspire system as a whole remains compliant to IEC 60601-1-2 with respect to leakage currents, emissions and immunity.

  • The treatment coil is different between the subject Horizon 3.0 Inspire compared to that used on the primary predicate device. The Horizon Air Film Coil is the designated treatment coil for the Horizon 3.0 Inspire configuration. The Horizon Air Film Coil has received prior clearance under K182853, K18097 and K171051 and was previously a predicate for introducing the E-Z Cool Coil range to market. The coil magnetic and electric field output properties were compared to the primary predicate Horizon 3.0 E-z Cool Coil in accordance to the FDA's guidance "Class II Special Control Guidance Document: Repetitive Transcranial Magnetic Stimulation (TMS)" and special controls 21 C.F.R. § 882.5802 and 21 C.F.R. § 882.5805 to reaffirm it's suitability for treatment in accordance with the indications for use and were found to be substantially equivalent.

  • The remaining minor technological changes are for the Cart, Coil Support and Accessory Cables. The Cart and Coil Support mechanism for the Horizon 3.0 Inspire configuration are the Magstim Coll Stand respectively. Both components are of known provenance for supporting the weight of the system components and applicating coil as remaining stable during treatment delivery as these components have been previously cleared under K182853 and K180907. Most of the same cables have been re-used with the exception of the Mainframe communications and power USB leads. These cables have been testing as part of IEC 60601-1 type testing.

As can be seen from the summary above - technologically the Horizon 3.0 Inspire configuration which is the submission is very similar, and in some cases identical, to the primary predicate device configurations cleared under K232235. The technological changes can be considered more of a re-organizational exercise to provide a system of reduced footprint and cost to the encluser whilst still maintaining the same core principles, technology, safety and effectiveness of the primary predicate device.

This demonstrates all three configurations including the subject Horizon 3.0 Inspire configuration, have identical intended use/ indications for use, common specifications, equivalent performance and equivalent composition.

All changes have been appropriately verified and conforms to the same safety standards as the primary predicate device such as IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 62366-1, AAM/ANSI HE75, ISO 10993-1, ISO 10993-10, ISO 14971 for electrical safety, thermal safety, mechanical safety, alarm systems and biocompatibility. Including IEC 62304, AAM TRS7 and AAM TIR97 for software and security testing of the Horizon 3.0 Inspire configuration such as Security Requirement, Threat Mitigation and Vulnerability testing was performed and documented in accordance to demonstrate the Horizon 3.0 Inspire configuration is as secure as the primary predicate device and demonstrate the continued security of the Horizon® 3.0 TMS Therapy System family.

{7}------------------------------------------------

In conclusion - the Horizon® 3.0 TMS Therapy System with the addition of the Horizon 3.0 Inspire configuration and its technological differences from the previously cleared predicate device configurations do not raise any new or differing questions of safety or effectiveness and are substantially equivalent.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Performance testing of the Horizon® 3.0 TMS Therapy System with the addition of the Horizon 3.0 Inspire configuration demonstrates that the device performs as intended for the safe and effective treatment of MDD and as an adjunct for the treatment of OCD and is therefore substantially equivalent to the primary predicate device configurations - Horizon 3.0 with StimGuide Pro. The following nonclinical testing within this submission is relied upon for a determination of substantial equivalence:

Electromagnetic Compatibility in accordance with IEC 60601-1-2 - The Horizon 3.0 "Inspire" configuration has been tested and found compliant to the requirements of IEC 60601-1-2 to demonstrate an equivalent level of Electromagnetic Compatibility as the primary predicate device.

Electrical & Mechanical Safety in accordance with IEC 60601-1 - The Horizon 3.0 "Inspire" configuration has been tested and found compliant to the requirements of IEC 60601-1 (including IEC 60601-1-8) to demonstrate an equivalent level of Electrical and Mechanical safety as the primary predicate device.

Thermal Safety in accordance with IEC 60601-1 - The Horizon 3.0 "Inspire" configuration has been tested and found compliant to the requirements of EC 60601-1 (including IEC 60601-1-6 and IEC 60601-1-8) to demonstrate an equivalent level of Thermal safety as the primary predicate device. Additional testing was performed to ensure that the Horizon Air Film Coil can execute OCD, iTBS and rTMS protocols at worst-case ambient conditions without resulting in excessive temperatures.

Software Verification & Validation (IEC 62304, FDA Software has been developed in a structured manner following IEC 62304. Software lifecycle documentation demonstrates appropriate function of software and demonstrates the software cannot contribute to any unacceptable risk. Changes made to inspire" configuration have been verified for correctness and also do not negatively impact the Horizon 3.0 with StimGuide Pro configurations.

Usability/Human Factors Engineering (IEC 60601-1-6, IEC 62366-1, ANSI/AAMI HE75) - A human factors evaluation demonstrates appropriate human factors and usability of the Horizon® 3.0 TMS Therapy System device configurations for its intended use and demonstrates that the device is free from any unacceptable use related risks.

Magnetic Pulse Output Testing and Magnetic and Electrical Field Testing (Special Controls: 21 C.F.R. § 882.5805 and FDA Guidance "Repetitive Transcranial Magnetic Stimulation (rTMS) Systems – Class II Special Controls Guidance for Industry and FDA Staff" – July 26, 2011) - The electric field distribution produced by the subject coil in a human head phantom model filled with physiologic saline solution were measured for comparison. Values of the Magnetic and Electric field were obtained by measuring induced voltage into a measuring probe inside a phantom head model filled with a physiologic saline solution. The results demonstrated equivalent Power Outputs, Electric Fields between the subject device configuration applicating coll and the primary predicate device configuration applicating coil.

Safety Feature Testing (IEC 60601-1) - IEC 60601-1 testing exercises many fault scenarios to ensure appropriate safety features are maintained. System level testing documented as part of the software verification also exercises the various interlocks with respect to safety features. Testing introducing failure modes to test multiple fault conditions, such as disabling a software check and ensuring a hardware backup interlock is active.

Acoustic Testing (IEC 60601-1) - Acoustic testing of the system under a simulated use scenario demonstrates the Horizon 3.0 "Inspire" configuration with the Horizon Air Film Coil does not reach excessive/unacceptable noise levels.

Clinical Testing - Not Applicable

Based on the above nonclinical testing – it can be concluded that the subject Horizon® 3.0 TMS Therapy System device introducing the Horizon 3.0 "Inspire" configuration is substantially equivalent to the primary predicate device cleared under K232235. The Horizon 3.0 "Inspire" offers an equivalent level of safety and effectiveness as the primary predicate Horizon 3.0 with StimGuide Pro configurations, whilst meeting its goal of offering an alternative system to meet such as footprint and cost to ensure that as many patients as possible can gain access to safe and effective TMS Therapy Treatments.

Electromagnetic Compatibility was demonstrated to be the primary predicate device through compliance to IEC 60601-1-2. The test plan executed to evaluate the subject Horizon 3.0 "Inspire" configuration utilized the same performance criteria as that used for the primary predicate device configurations. The test plan was composed of the same device specific criteria that was used to evaluate the primary predicate device configurations (Horizon 3.0 with StimGuide Pro) and therefore directly demonstrates a substantial equivalence of the subject Horizon® 3.0 TMS Therapy System introducing the Horizon 3.0 "Inspire" configuration.

{8}------------------------------------------------

Electrical, Mechanical and Thermal safety was demonstrated through the application of IEC 60601-1. The subject Horizon 3.0 "Inspire" configuration was found to be compliant to the standard. Further, in-house device specific protocol testing determined that for TMS, iTBS and OCD treatment protocols under reasonable worst-case simulated ambient conditions (80% - 100% of machine output, at 30°C ambient room temperature) - the Horizon 3.0 "Inspire" configuration with the Horizon Air Film coil could perform the recommended protocols safely and effectively - equivalent to the primary predicate device.

Software Verification and Validation was performed on the subject Horizon 3.0 "Inspire" device configuration. The test procedure used to verify the subject Horizon 3.0 "Inspire" configuration was comprised of test cases used to verify the primary predicate device and its configurations: Horizon 3.0 and Horizon 3.0 with StimGuide Pro. This supports substantial equivale behavior is consistent and compatible across all configurations of the Therapy System. The testing also included cybersecurity testing to ensure the subject Horizon 3.0 "Inspire" device configuration is as secure as the primary predicate device configurations.

Human Factors and Usability Engineering processes demonstrate the Horizon is substantially equivalent to the primary predicate device configurations for its interms of human factors and usability, and demonstrates that the device is free from any unacceptable use related risks. The consistency of workflows, User Interfaces and components of known provenance between all configurations of the device promote a healthy usability profile for all system configurations and the introduction of the Horizon 3.0 "Inspire" configuration of safety or effectiveness with respect to usability and human factors.

Magnetic and Electric Field Testing demonstrated that the subject device configuration - the Horizon Air Film Col, is substantially equivalent to the applicating coil of the primary predicate device configuration. Values of the Magnetic and Electric feld were obtained by measuring induced voltage into a measuring probe inside a phantom head model filled with a physiologic saline solution, with a total volume of 4cm x 3cm x 3cm and a resolution of 5mm.

Key characteristics of the fleld output were equivalent between both coils; such as the required stimulator outputy of age, field spatial distribution, E-Field decay, output waveforms and rate of change. This is especially apparent at the clinically relevant depths of stimulation - where the values of the characteristics of both coils were virtually identical.

The Horizon Air Film Coil has been on the market since K143531 and is used by reference predicate device cleared under K182853. The Horizon Air Film Coil was previously used as the predicate for introducing the Horizon E-z Cool Coil range to market. The Horizon Air Film Coll has a known safety and effectiveness profile, especially for MDD treatments, and the magnetic and electric fields of the Horizon Air Film Coil performed as part of this submission also demonstrates an equivalent level of safety and effectiveness for use as an adjunct for Obsessive Compulsive Disorder. As a result this demonstrates substantial equivalence between the subject Horizon 3.0 "Inspire" device configuration coil and the primary predicate device configurations/applicating coil.

Safety Feature Testing ensured all fault scenarios and safety features tested utilized similar (in the vast majority of cases, the same) procedures that were used to evaluate the primary predicate device, including compliance to IEC 60601-1. The testing provided within this submission demonstrates that there is substantial equivalence of safety features amongst all device configurations – importantly, including the subject Horizon 3.0 "Inspire" configuration.

Acoustic Testing in simulated use conditions (OCD Protocol at maximum machine output) for the subject Horizon 3.0 "Inspire" device configuration demonstrated equivalent acoustic properties to the values recorded for the subject device configuration fall below safe exposure limits demonstrating that Horizon 3.0 "Inspire" is as safe as the primary predicate device.

In conclusion - the subject Horizon® 3.0 TMS Therapy System with the addition of the Horizon 3.0 Inspire configuration is as safe and effective and is substantially equivalent to the primary predicate device cleared under K232235.

§ 882.5805 Repetitive transcranial magnetic stimulation system.

(a)
Identification. A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.(b)
Classification. Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.