K Number
K182853

Validate with FDA (Live)

Date Cleared
2019-03-15

(156 days)

Product Code
Regulation Number
882.5805
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HORIZON® TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.

Device Description

The HORIZON® TMS Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation.

The HORIZON® TMS Therapy System is a non-invasive tool for the stimulation of cortical neurons for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from antidepressant medication in the current episode. The HORIZON® TMS Therapy System is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient and outpatient settings, including physicians' offices, clinics, and hospitals.

The HORIZON® TMS Therapy System is an integrated system consisting of a combination of hardware, software, and accessories.

AI/ML Overview

This document describes the Magstim HORIZON® TMS Therapy System, specifically focusing on K182853, which adds a new treatment protocol (iTBS) to a previously cleared device (K180907). The document demonstrates substantial equivalence to predicate devices, rather than presenting a de novo study with acceptance criteria and performance data for a novel device. Therefore, many of the requested sections (sample size, expert ground truth, adjudication, MRMC) are not applicable as they relate to clinical studies proving efficacy or diagnostic performance, which were not performed for this submission.

However, I can extract information related to the device's technical specifications and the basis for its safety and effectiveness claims.

Acceptance Criteria and Device Performance

Since this is a substantial equivalence submission for adding a treatment protocol, there aren't traditional "acceptance criteria" for performance that a new device would need to meet in a clinical trial. Instead, the acceptance criteria here are demonstrating that the addition of the iTBS protocol poses no new issues of safety or effectiveness, and that the device remains substantially equivalent to already cleared devices.

Table of "Acceptance Criteria" (demonstrated equivalency) and Reported Device "Performance" (characteristics for comparison):

CriteriaHORIZON® TMS Therapy System (Subject Device - K182853)Mag Vita TMS Therapy System w/Theta Burst Stimulation (Primary Predicate - K173620)HORIZON® TMS Therapy System (Secondary Predicate - K180907)
Intended Use/Indications for UseIdentical: Treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.Identical: Treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.Identical: Treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Technological CharacteristicsIdentical (basic design, mechanism of action, specifications) to K180907, with addition of iTBS protocolCompared for iTBS protocolIdentical (basic design, mechanism of action, specifications)
Principles of OperationEquivalent to predicates (system setup, patient prep, motor threshold determination, coil position, treatment administration)Equivalent for iTBS protocolEquivalent
Electrical SafetyMeets IEC 60601-1YesMeets IEC 60601-1
EMC StandardsMeets EN 60601-1-2YesMeets EN 60601-1-2
iTBS Treatment Protocol - Stimulation Intensity120% of the MT120% of the MTN/A (not applicable to K180907 as a standard protocol)
iTBS Treatment Protocol - Repetition Rate50 Hz (5 pulses per sec)50 Hz (5 pulses per sec)N/A
iTBS Treatment Protocol - Train Duration2 sec2 secN/A
iTBS Treatment Protocol - Inter-train Interval8 sec8 secN/A
iTBS Treatment Protocol - Burst Pulses33N/A
iTBS Treatment Protocol - Bursts200200N/A
iTBS Treatment Protocol - Inter Pulse interval20 msec20 msecN/A
iTBS Treatment Protocol - Number of trains2020N/A
iTBS Treatment Protocol - Number of Pulses per Session600600N/A
iTBS Treatment Protocol - Treatment Session Duration3.09 min3.09 minN/A
iTBS Treatment Protocol - Sessions/week55N/A
iTBS Treatment Protocol - Treatment Schedule5 daily sessions for 6 weeks5 daily sessions for 6 weeksN/A
iTBS Treatment Protocol - Area of brain to be stimulatedLeft Dorsolateral Prefrontal CortexLeft Dorsolateral Prefrontal CortexN/A
Non-clinical performance testing for iTBSDemonstrated that the intensity of individual stimuli in iTBS is equal and kept constant throughout delivery to ensure constant dose, equivalent to K173620.Confirmed this principle is essential for clinical performance.N/A

Study Details (Not Applicable for this Submission Type):

As this is a 510(k) submission based on substantial equivalence for a minor modification (adding a treatment protocol) to an already cleared device, detailed clinical studies with the specific elements requested (sample sizes, expert ground truth, adjudication methods, MRMC studies) were not performed or required for this particular submission. The "study" here is primarily a comparative analysis to existing cleared devices and verification of the technical aspects of the new protocol addition.

  1. Sample size used for the test set and the data provenance: Not applicable. No new clinical test set was used for this 510(k) submission. The data provenance is implied by the predicate devices being legally marketed in the US.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical efficacy was established by the predicate devices' historical data or studies, not new data for this submission.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/diagnostic imaging device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm-only device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the original clearance of the predicate devices, clinical efficacy would have been established through clinical trials and outcomes data. For this specific submission, the ground truth for safety and effectiveness of the iTBS protocol is implicitly based on its successful clearance and clinical use in the primary predicate device (Mag Vita TMS Therapy System w/Theta Burst Stimulation, K173620) and the physical and electrical equivalence of the HORIZON system to its prior clearance (K180907).
  7. The sample size for the training set: Not applicable. This is not an AI/machine learning device requiring a training set in the conventional sense.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 15, 2019

Magstim Company Ltd. Tom Campbell Regulatory Manager Spring Gardens, Whitland, Carmarthenshire SA34 OHR. United Kingdom

Re: K182853

Trade/Device Name: HORIZON TMS Therapy System Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: February 5, 2019 Received: February 11, 2019

Dear Tom Campbell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Pamela D. Scott -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182853

Device Name HORIZON TMS Therapy System

Indications for Use (Describe)

The HORIZON TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "magstim" in a bold, sans-serif font. The word is colored in a dark blue, and there is a light blue circle above the "i" in "magstim". There is a registered trademark symbol to the right of the word.

510(k) SUMMARY Magstim's HORIZON®TMS Therapy System

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Magstim Company Limited Spring Gardens, Whitland, Carmarthenshire SA34 OHR, United Kingdom

Phone: +44 (0) 1994 240798 Facsimile: +44 (0) 1994 240061

Contact Person: Tom Campbell

Date Prepared: March 15, 2019

Name of Device

HORIZON® TMS Therapy System

Common or Usual Name/

Repetitive Transcranial Magnetic Stimulation (rTMS) System

Classification

Repetitive Transcranial Magnetic Stimulation (rTMS) System

21 C.F.R. § 882.5805, Class II, product code OBP

Predicate Device

Mag Vita TMS Therapy System w/Theta Burst Stimulation, Tonica Elektronik A/S. (K173620) (Primary Predicate)

HORIZON® TMS Therapy System, The Magstim Company Limited. (K180907) (Secondary Predicate)

Device Description

The HORIZON® TMS Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation.

The HORIZON® TMS Therapy System is a non-invasive tool for the stimulation of cortical neurons for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from antidepressant medication in the current episode. The HORIZON® TMS Therapy System is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient and outpatient settings, including physicians' offices, clinics, and hospitals.

{4}------------------------------------------------

The HORIZON® TMS Therapy System is an integrated system consisting of a combination of hardware, software, and accessories. Its technological characteristics are described in further detail below.

Intended Use / Indications for Use

The HORIZON® TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.

Technological Characteristics

The HORIZON® TMS Therapy System is comprised of the following components:

    1. HORIZON® Stimulator
    • a. HORIZON® User Interface;
    • b. HORIZON® Mainframe;
    • c. HORIZON® Power Supply;
    • d. Accessory Cables;
    • e. Accessory Footswitch.
    1. Coil for MT Determination
    • a. HORIZON® MT Remote Coil.
    1. Coil(s) for Treatment
    • a. HORIZON® AFC:
    • b. HORIZON® E-z Cool Coil.
    1. Accessory Cart(s) and Coil Holding Mechanism(s)
    • a. Magstim®Trollev:
    • b. Magstim® Coil Stand(s);
    • c. HORIZON® E-z Cart;
    • d. HORIZON® E-z Arm.
    1. Accessory Marking Apparatus
    • a. TMS Patient Caps.

The operator controls the HORIZON® TMS Therapy System via the HORIZON® User Interface, using a graphic LCD panel with touchscreen technology. The operator instructions, given through the HORIZON® User Interface, direct the HORIZON® Mainframe in charging and discharging the device's high voltage discharge capacitor. The discharge is delivered to the patient via the stimulating coil. Motor threshold level can be determined using the HORIZON® MT Remote Coil. Treatment is delivered to the treatment area via either the HORIZON® AFC or the HORIZON® E-z Cool Coil, which is positioned above the treatment area. Positioning, and fixation, of the coil over the treatment area is accomplished using the Coil Holding Mechanism(s). The HORIZON® Power Supply provides power to charge the high voltage capacitor in the HORIZON® Mainframe.

Non-Clinical Performance Testing

The non-clinical performance testing of the HORIZON® TMS Therapy System including electrical safety, electromagnetic compatibility, biocompatibility, human factors and magnetic field characteristics has been tested as required and cleared by the FDA earlier in K180907.

{5}------------------------------------------------

Additional verification testing has demonstrated that the intensity of the individual stimuli in iTBS is equal and kept constant throughout the delivery of the full treatment. As stated by the primary predicate device (K173620), the clinical performance of iTBS is dependent on the stimuli being delivered at equal intensity, to ensure that a constant dose of stimuli is delivered during treatment.

This demonstrates that the HORIZON® TMS Therapy System is safe and effective for use in treatment of Major Depressive Disorder.

Substantial Equivalence

The HORIZON® TMS Therapy System is substantially equivalent to its predicate devices, the HORIZON® TMS Therapy System (K180907) and the Mag Vita TMS Therapy System w/Theta Burst Stimulation (K173620).

The intended use and indications for use for the HORIZON® TMS Therapy System and the secondary predicate device, the HORIZON® TMS Therapy System (K180907) are identical.

The key technological characteristics including basic design, mechanism of action, specifications are also identical for both the HORIZON® TMS Therapy System and the secondary predicate device (K180907).

The principles of operation of the HORIZON® TMS Therapy System is equivalent to the secondary predicate device, the HORIZON® TMS Therapy System (K180907). The only change to the HORIZON® TMS Therapy System is a modification to the labelling to include a recommendation for an additional treatment protocol known as iTBS. This treatment protocol, is identical to the treatment protocol recommended by the primary predicate device, the Maq Vita TMS Therapy System w/Theta Burst Stimulation (K173620). The design of the HORIZON® TMS Therapy System is unchanged from that previously cleared under K180907.

Transcranial magnetic stimulation is enabled in the HORIZON® TMS Therapy System and in the predicate devices, as all have the same key system components, consisting of electromagnetic coils, a coil holding mechanism, a TMS stimulator and software. The operation procedure consisting of system setup, patient preparation, determination of patients' motor threshold, coil position, and administration of treatment at pre-defined treatment stimulation parameters is unchanged in the HORIZON® TMS Therapy System to that cleared earlier in (K180907).

The basic software capabilities related to treatment administration are the same as the secondary predicate (K180907)

The HORIZON® TMS Therapy System meets the same electrical and mechanical safety standards (IEC 60601-1) and the same EMC standards (EN 60601-1-2).

The similarities and minor differences between the HORIZON® TMS Therapy System, the Mag Vita TMS Therapy System w/Theta Burst Stimulation (K173620) (Primary Predicate) and the HORIZON® TMS Therapy System (K180907) (Secondary Predicate) are described in Table 1.

{6}------------------------------------------------

Conclusions

In summary, the intended use and indications for use for the HORIZON® TMS Therapy System and the secondary predicate device, the HORIZON® TMS Therapy System (K180907) are identical.

Furthermore, the key technological characteristics including basic design, mechanism of action, specifications are also identical for both the HORIZON® TMS Therapy System and the secondary predicate device (K180907).

The principles of operation, including the method for determining coil position, the treatment procedure and the relevant protocol parameters are substantially equivalent to the predicate devices.

Non-clinical performance testing has demonstrated that the change to the HORIZON® TMS Therapy System labelling to add a recommendation for an additional treatment protocol known as iTBS, raises no new issues of safety or effectiveness. Furthermore, the intended use, as described in the labeling, has not changed as a result of this modification.

Thus, the HORIZON® TMS Therapy System is considered to be substantially equivalent.

{7}------------------------------------------------

Table 1: Substantial Equivalence Summary

CriteriaHORIZON® TMS Therapy System(Subject of this submission)Mag Vita TMS Therapy Systemw/Theta Burst Stimulation(K173620)(Primary Predicate)HORIZON® TMS Therapy System(K180907)(Secondary Predicate)
ManufacturerMagstim Company LimitedTonica Elektronik A/SMagstim Company Limited
Device NameHORIZON® TMS Therapy SystemMag Vita TMS Therapy Systemw/Theta Burst StimulationHORIZON® TMS Therapy System
Clearance date08/14/201808/03/2018
510(k) numberK173620K180907
Device codeOBPOBPOBP
Intended Use/Indications for UseThe HORIZON® TMS Therapy System is indicated forthe treatment of Major Depressive Disorder in adultpatients who have failed to achieve satisfactoryimprovement from prior antidepressant medication inthe current episode.The Mag Vita TMS Therapy System w/ThetaBurst Stimulation is indicated for thetreatment of Major Depressive Disorder inadult patients who have failed to receivesatisfactory improvement from priorantidepressant medication in the currentepisode.The HORIZON® TMS Therapy System is indicated for thetreatment of Major Depressive Disorder in adultpatients who have failed to achieve satisfactoryimprovement from prior antidepressant medication inthe current episode.
Standard Treatment Protocol
Magnetic FieldIntensity120% of the MT-120% of the MT
Stimulus Frequency10 Hz-10 Hz
Stimulus Trainduration4 sec-4 sec
Inter-train interval11-26 sec-11-26 sec
Number of trains75-75
Magnetic Pulses perSession3000-3000
Treatment SessionDuration18.8 min–37.5 min-18.8 min–37.5 min
Sessions/week5-5
Treatment Schedule5 daily sessions for 6 weeks-5 daily sessions for 6 weeks
Area of brain to bestimulatedLeft Dorsolateral Prefrontal Cortex-Left Dorsolateral Prefrontal Cortex
iTBS Treatment Protocol
Stimulation Intensity120% of the MT120% of the MT-
Repetition Rate50 Hz (5 pulses per sec)50 Hz (5 pulses per sec)-
Train Duration2 sec2 sec-
Inter-train Interval8 sec8 sec-
Burst Pulses33-
Bursts200200-
Inter Pulse interval20 msec20 msec-
Number of trains2020-
Number of Pulses perSession600600-
Treatment SessionDuration3.09 min3.09 min-
Sessions/week55-
Treatment Schedule5 daily sessions for 6 weeks5 daily sessions for 6 weeks-
Area of brain to bestimulatedLeft Dorsolateral Prefrontal CortexLeft Dorsolateral Prefrontal Cortex-

{8}------------------------------------------------

{9}------------------------------------------------

HORIZON®MT RemoteCoilHORIZON®E-z Cool CoilHORIZON®AFCCool-B70 CoilHORIZON®MT RemoteCoilHORIZON®E-z Cool CoilHORIZON®AFC
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
Core MaterialAirAirAirAirAirAirAir
ShapeFigure-of-eightFigure-of-eightFigure-of-eightFigure-of-eightFigure-of-eightFigure-of-eightFigure-of-eight
Coil WindingConfigurationInner Diameter (ID)Outer Diameter (OD)Winding Height (H)Number of Windings (NW)ID: 55mmOD: 92mmH: 6mmNW: 2x9ID: 37.5mmOD: 97mmH: 11mmNW: 2x(3x23)ID: 43mmOD: 92mmH: 11mmNW: 2x(3x19)ID: 23mmOD: 96mmH: 12mmNW: 2x11ID: 55mmOD: 92mmH: 6mmNW: 2x9ID: 37.5mmOD: 97mmH: 11mmNW: 2x(3x23)ID: 43mmOD: 92mmH: 11mmNW: 2x(3x19)
Pulse Width330μs340μs300μs280μs330μs340μs300μs
Amplitude in SMTunits(Standard Motor Threshold)0.28 - 1.90 - 1.70.28 - 1.9
Frequency range (Hz)at 100%1 - 201 - 20

§ 882.5805 Repetitive transcranial magnetic stimulation system.

(a)
Identification. A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.(b)
Classification. Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.