K Number
K241379

Validate with FDA (Live)

Date Cleared
2024-09-16

(124 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Description

The subject device, Powered Wheelchair, mainly powered by battery, motivated by DC motor, driven by user controlling joystick and adjusting speed. The Powered Wheelchairs consist of two foldable armrests, a backrest, a seat cushion, a safety belt, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, two Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller. The NXN20-205 and NXN20-205M Powered Wheelchair is intended to provide mobility to a disabled or elderly person limited to a seated position. The Powered Wheelchair has 7 inch front wheel and 12 inch rear tire. The motor of electric wheelchair is DC24V 200W; the battery is 24V 12AH, Li-ion battery; the charger is 24V/3A. Max. loading can not be over than 110Kgs. Max. distance of travel on the fully charged battery is 16km and Max. speed forward is 5.5km/h. The braking time is about 2s, and the braking distance is ≤790m.

AI/ML Overview

The provided text is a 510(k) Summary for a Powered Wheelchair (NXN20-205, NXN20-205M) and does not contain details about acceptance criteria and a study proving a device meets them in the context of an AI/ML medical device.

The document describes a traditional medical device (a powered wheelchair) and its substantial equivalence to a predicate device based on non-clinical testing against established ISO standards for wheelchairs. There is no mention of AI or machine learning in this report.

Therefore, I cannot provide the requested information, such as:

  • Table of acceptance criteria and reported device performance for an AI/ML device: This document details performance specifications for a physical wheelchair (e.g., speed, range, loading capacity, stability), not AI/ML metrics.
  • Sample size used for the test set and data provenance: No test set or data provenance for an AI/ML model is mentioned.
  • Number of experts used to establish ground truth and qualifications: This is irrelevant for a physical powered wheelchair.
  • Adjudication method: Not applicable.
  • MRMC comparative effectiveness study: Not conducted as no AI assistance is involved.
  • Standalone (algorithm only) performance: Not applicable.
  • Type of ground truth used: Not applicable in the context of an AI/ML device.
  • Sample size for the training set: No training set is mentioned.
  • How the ground truth for the training set was established: Not applicable.

The document indicates that "Software Verification and Validation Testing" was performed in accordance with FDA Guidance for device software functions, but this refers to traditional software engineering validation, not AI/ML model validation with ground truth and performance metrics.

In summary, the provided text does not contain the information required to answer the prompt about acceptance criteria and studies for an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

September 16, 2024

Zhejiang Nysin Medical Co., Ltd. % Boyle Wang Manger Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China

Re: K241379

Trade/Device Name: Powered Wheelchair (NXN20-205, NXN20-205M) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: September 3, 2024 Received: September 3, 2024

Dear Boyle Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Heather L. Dean -S

Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K241379

Device Name

Powered Wheelchair (NXN20-205, NXN20-205M)

Indications for Use (Describe)

The device is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

K241379

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's information

Name: Zhejiang Nysin Medical Co., Ltd. Address: Building 13, No. 305 Jiulong North Road, Dongcheng District, Yongkang, Jinhua City, Zhejiang Province, China Tel: +86-579-87153722/ +86-18969388822 Contact: Zhang Xiaoying

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

Date of Preparation: Sep.02,2024

2.0 Device information

Trade name: Powered Wheelchair Common name: Powered Wheelchair Classification name: Powered Wheelchair Model(s): NXN20-205, NXN20-205M

3.0 Classification

Production code: ITI Regulation number: 21 CFR 890.3860 Classification: Class II Panel: Physical Medicine

4.0 Predicate device information

Manufacturer: Kunshan Aoshida Electric Technology Co., Ltd. Trade/Device: A08 Power Wheelchair

{5}------------------------------------------------

510(k) number: K163204

5.0 Indication for Use Statement

The device is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

6.0 Device description

The subject device, Powered Wheelchair, mainly powered by battery, motivated by DC motor, driven by user controlling joystick and adjusting speed.

The Powered Wheelchairs consist of two foldable armrests, a backrest, a seat cushion, a safety belt, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, two Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller.

The NXN20-205 and NXN20-205M Powered Wheelchair is intended to provide mobility to a disabled or elderly person limited to a seated position.

The Powered Wheelchair has 7 inch front wheel and 12 inch rear tire.

The motor of electric wheelchair is DC24V 200W; the battery is 24V 12AH, Li-ion battery; the charger is 24V/3A.

Max. loading can not be over than 110Kgs.

Max. distance of travel on the fully charged battery is 16km and Max. speed forward is 5.5km/h.

The braking time is about 2s, and the braking distance is ≤790m.

7.0 Summary of Non-Clinical Testing

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability

ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electrically powered wheelchairs

ISO 7176-3 : 2012 Wheelchairs - Part 3: Determination of effectiveness of brakes

ISO 7176-4 : 2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range

ISO 7176-5 : 2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space

ISO 7176-6: 2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs

ISO 7176-7 : 1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions

{6}------------------------------------------------

ISO 7176-8 : 2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths

ISO 7176-9: 2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs ISO 7176-10 : 2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs

ISO 7176-11 : 2012 Wheelchairs - Part 11: Test dummies

ISO 7176-13 : 1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces

ISO 7176-14 : 2008 Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods

ISO 7176-15: 1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling

ISO 7176-21 : 2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers

ISO 7176-22 : 2014 Wheelchairs - Part 22: Set-up procedures

ISO 7176-25:2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs

ISO 16840-10: 2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method

IEC 60601-1-2:2020 Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests

Software Verification and Validation Testing

Software documentation including verification & validation was provided in accordance with FDA Guidance: Content of Premarket Submissions for Device Software Functions for software.

The Software Validation is in compliance with FDA Guidance.

8.0 Summary of Clinical Testing

No clinical study implemented for the electric wheelchair.

9.0 Technological Characteristic Comparison Table

{7}------------------------------------------------

ItemProposed devicePredicate deviceRemark
Product CodeITIITISame
Regulation No.21 CFR 890.386021 CFR 890.3860Same
ClassIIIISame
Product namePowered WheelchairA08 Power Wheelchair-
510(k) No.K241379K163204-
ModelsNXN20-205NXN20-205MA08-
Intended UseThe device is a motor driven, indoor and outdoor transportation vehicle with the intended use toIt is a motor driven, indoor and outdoor transportation vehicle with the intended useSame
provide mobility to a disabled or elderly person limited to a seated position.to provide mobility to a disabled or elderly person limited to a seated position.
Use environmentIndoor and outdoor useIndoor and outdoor useSame
PatientPopulationThis product is suitable fordisabled people with mobilitydifficulties and elderly people.This product is suitable fordisabled people with mobilitydifficulties and elderly people.Same
Product structureConsist of two foldable armrests,a backrest, a seat cushion, asafety belt, a foldable frame, tworear driving wheels with hubmotor/electromagnetic brakeassemblies, two pivoting casters,two Li-ion batteries, an off-boardbattery charger, a control panel,and an electric motor controller.Consists primarily of afoldable welded-aluminumframes, two sealed transaxlemotors drive system,electromagnetic brakingsystem, electric motorcontroller and two Li-ionbatteries with an off-boardbattery charger.Similar
Driving systemDirect drive on the rear wheelsDirect drive on the rearwheelsSame
Movement controlmethodBy Joystick controlBy Joystick controlSame
Number ofwheels44Same
Brake systemAutomatic electromagneticbrake systemElectromagneticbrake systemSame
Main framematerialModel NXN20-205 consistsof Aluminum alloy, andmodel NXN20-205M consistwith Magnesium alloy.Welded-aluminumSame
MotorBrushless motor,DC24V* 200W*2pcsBrushless motor,24 VDC *250W * 2 pcsMinor differences in the
BatteryDC 24V 12Ah Lithium-ion, 2 pcsLithium-ion20 Ah x 24 VDCdimensions will notimpact the safety and
Off-board chargerOff-board chargereffectiveness of the
Battery chargerInput: 100-240 VACOutput: DC 24V, 3AInput: 110-240 VACOutput: DC 24V, 2 Ampsubstantial equivalence.

Table1-General Comparison

{8}------------------------------------------------

Table2 Performance Comparison

ItemSubject DeviceSubject DevicePredicate DeviceRemark
ModelNXN20-205NXN20-205MA08--
Overall length980mm960mm890 mm
Overall width580mm580mm603 mmMinor differences in the
Stowagelength690mm650mm324 mmdimensions will notimpact the safety and
Stowage width250mm250mm603 mmeffectiveness of thesubstantial equivalence.
Stowageheight720mm740mm670 mm
Weight, w/Battery42.5 lbs. /19.3 kg40.33 lbs. /18.3 kg61.7 lbs. /28 kgThe difference willnot raise any new safetyand effectivenessconcerns.
Frontwheel(inch)7(PU solid tire)8(PU solid tire)Smaller sizes of font wheels
Rear tire (inch)12(PU solid tire)10(PU solid tire)Larger sizes of rear wheelsbring steadier pivotingfunction than predicatedevice.
CruisingRange(km)1620The subject device complieswith ISO 7176-4: 2008Wheelchairs - Part 4: Energyconsumption of electricwheelchairs and scooters fordetermination of theoreticaldistance range, thesedifferences do not affectsafety and effectiveness.
Obstacleclimbing(mm)2540The smaller height in theobstacle climbing will notimpact the safety andeffectiveness of the subjectdevice.
Static stabilityNot publicly availableBoth of the devices are
forward
Static stabilityrearwardevaluated according to standard ISO 7176-1:2014, so the different static stability will not impact the safety and effectiveness
Static stabilitysideways
Max. loading(kg)242.5lbs(110kg)220 lbs (100 kg)More loading weight means more convenient for the transportation
Min. Turningradium810mm800mmSame
Minimumbrakingdistance0.79m1mSimilar
Max SpeedForwards1.53m/s (5.5 km/h)1.94m/s (7 km/h)The devices are evaluated according to standard ISO 7176-6:2018, so the different will not impact the safety and effectiveness
Max. SpeedBackward0.8m/s (2.88 km/h)Not publicly availableThe devices are evaluated according to standard ISO 7176-6:2018, so the different will not impact the safety and effectiveness
ControllerYanteon Mechanical & Electronic Technology (Shanghai) Co. Ltd.Joystick Y2450MController Y2430MYisheng Electric Co. Ltd ,WS-1, 40ADifferentAlthough different controller is used, both the control system, including the joystick controller, the electromagnetic brakes and the user interface are similar. The joystick controls the directions and speed of movement, and when the joystick is released, the powered wheelchair will slow down to stop and the brakes will automatically re-engage. The controller also provides the battery status displaying and abnormal condition displaying. Both of the control systems are evaluated according to standard ISO 7176-14:2008 and software
validation requirement andthere are no new safety andeffectiveness concerns dueto the difference.
Speed controlmethodJoystick control methodJoystick control methodSame

{9}------------------------------------------------

{10}------------------------------------------------

Table3 Safety Comparison

ItemProposed DevicePredicate DeviceRemark
Materialscontacting userArmrest: PU;Backrest/Seat: Sandwich meshfabric (polyester)Seat belt: NylonController Housing:ABSJoystick: SiliconeButton:SiliconeArmrest: PU;Backrest/seat cushion: PU foamcovered by nylon fabric clothBiocompatibility evaluationhas been carried out perISO 10993-1. There are nonew safety andeffectiveness concerns dueto the difference.
Biocompatibilityof materialscontacting userComply with FDA GuidanceComply with ISO 10993-1, FDAGuidance, Tests includedCytotoxicity (ISO 10993-5:2009),Sensitization and IntracutaneousReactivity (ISO 10993-10:2010)Same

Summary of substantial equivalence discussion:

The technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

10.0 Conclusion

The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device under K163204.

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).