K Number
K163204

Validate with FDA (Live)

Date Cleared
2017-05-05

(171 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Description

The A08 power wheelchair is an indoor/outdoor powered wheelchair that is battery operated. The design of this wheelchair is basically similar to other powered wheelchairs that are already on the market. The A08 power wheelchair is with a 100 kg (220 lbs) weight capacity. It is basic conventional rear wheels drive, rigid frame vehicle that are battery powered. It consists primarily of a foldable welded-aluminum frames, two sealed transaxle motors (250 W, DC 24V) drive system, electromagnetic braking system, electric motor controller (Yisheng Electric Co., Ltd., Model: WS-1) and two Li-ion batteries with an off-board battery charger (2 A). It is powered by two 12 volt Li-ion DC batteries with 20 km (12.5 miles) with 20 Ah which maximum speed upto 7 km/hr (4.4 mph). The controller with a joystick attaches to either armrest and allows the rider to control the movement and velocity of the powered wheelchair. When the joystick is released, the electromagnetic brake will be actuated and the power wheelchair is slow to stop.

AI/ML Overview

The provided document is a 510(k) premarket notification for a power wheelchair (A08) and focuses on establishing substantial equivalence to a predicate device (PL001, K113463). It does not contain information about acceptance criteria and a study proving a device meets these criteria in the context of AI/ML performance.

The document describes a medical device (a power wheelchair) that has undergone non-clinical testing to comply with various ISO and IEC standards. These standards typically relate to the safety, performance, and construction of such devices, not to AI/ML software performance or diagnostic accuracy.

Therefore, I cannot extract the requested information regarding:

  • 1. A table of acceptance criteria and the reported device performance: The document only lists a series of ISO/IEC standards that the device "complied with," but does not provide specific acceptance criteria (e.g., minimum accuracy, sensitivity, specificity) for an AI/ML algorithm or the numerical performance results against those criteria.
  • 2. Sample size used for the test set and the data provenance: Not applicable as no AI/ML performance study is described.
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
  • 8. The sample size for the training set: Not applicable.
  • 9. How the ground truth for the training set was established: Not applicable.

The document is purely a regulatory submission for a physical medical device, demonstrating its equivalence to an existing device based on mechanical, electrical, and safety standards. It does not involve any AI/ML components or associated performance studies.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is depicted in black and white, and the text is in a simple, sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 5, 2017

Kunshan Aoshida Electric Technology Co., Ltd % Mrs. Junnata Chang Official Correspondent IRC China 16f-2(16a), No. 462, Sec. 2, Chongde Rd., Beitun Dist. Taichung, 406 TW

Re: K163204

Trade/Device Name: A08 Power Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: March 30, 2017 Received: April 13, 2017

Dear Mrs. Junnata Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Hoffmann -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163204

Device Name A08 Power Wheelchair

Indications for Use (Describe)

The device is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Applicant

Name: Kunshan Aoshida Electric Technology Co., Ltd

Address: No. 65, Matang Road, Kunshan City, Jiangsu Province, 215333, China

Contact person: Shen Wenhua

Phone: +86-512-57816288

Fax: +86-512-57816388

Date prepared: Jan. 20, 2017

Device identification

  • Trade name : A08 power wheelchair
  • Common name: Powered wheelchair
  • Classification name: Powered wheelchair
  • Medical specialty (Panel): Physical Medicine Device
  • Regulation number: 890.3860
  • Product code: ITI
  • Classification: Class II

Predicate devices

Trade name: PL001 power wheelchair

510(k) number: K113463

Manufacture: SUZHOU KD MEDICAL APPLIANCE CO. LTD.

Regulation number: 890.3860

Product Code: ITI

Classification: Class II

Intend use of device

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device description:

The A08 power wheelchair is an indoor/outdoor powered wheelchair that is battery operated. The design of this wheelchair is basically similar to other powered wheelchairs that are already on the market.

The A08 power wheelchair is with a 100 kg (220 lbs) weight capacity. It is basic

{4}------------------------------------------------

conventional rear wheels drive, rigid frame vehicle that are battery powered. It consists primarily of a foldable welded-aluminum frames, two sealed transaxle motors (250 W, DC 24V) drive system, electromagnetic braking system, electric motor controller (Yisheng Electric Co., Ltd., Model: WS-1) and two Li-ion batteries with an off-board battery charger (2 A). It is powered by two 12 volt Li-ion DC batteries with 20 km (12.5 miles) with 20 Ah which maximum speed upto 7 km/hr (4.4 mph). The controller with a joystick attaches to either armrest and allows the rider to control the movement and velocity of the powered wheelchair. When the joystick is released, the electromagnetic brake will be actuated and the power wheelchair is slow to stop.

Summary of non-clinical testing

The A08 power wheelchair complied with the requirements of the ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9. ISO 7176-10. ISO 7176-11. ISO 7176-13. ISO 7176-14. ISO 7176-15. ISO 7176-16, ISO 7176-21, ISO 7176-25, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 14971 and IEC 62133.

Statement of substantial equivalence

The A08 power wheelchair is substantially equivalent to the PL001 (K113463) manufactured by SUZHOU KD MEDICAL APPLIANCE CO. LTD.. They have same intended use of a motor driven, indoor and outdoor transportation vehicle to provide mobility to a disabled or elderly person limited to a seated position.

The design and technological characteristics of this device is basically similar to the predicate device. They are 4-wheeled power wheelchair with basic conventional rear wheel drive which with free-wheel mode. The motor of both devices is equipped with differential mechanism. The anti-tip wheels are equipped behind the rigid frame of both devices. The power supply of both devices is Li-ion batteries to provide power that can be recharged by an off-board charger that can be plugged into an AC outlet when the devices are not in use. The horn is also equipped on both devices have the same user interface, the patient uses the joystick to engage and disengage the power wheelchair motion in both the forward and reverse directions. When the joystick is released, the electromagnetic brake will be actuated and the power wheelchair is slow to stop.

Both seats of the devices are equipped with removable flip-up armrest and there is no headrest on backrest. Seat cushions and backrests of both devices used the same materials, which is PU foam covered by nylon fabric cloth. The seat, seat cushion, backrest, joystick of the device will contact to the patient. The materials of the patient contacting parts used for the A08 power wheelchair and predicate device are exactly

{5}------------------------------------------------

same and no additional processing different from predicate device, so no new safety and efficacy concerning.

The main difference for the two devices is the rated power of the motor, maximum curb height and the maximum speed. While there are minor differences between the devices including max load capacity, dimension, wheels size, braking distance and pre-charging distance, do not alter the intended use function and use of the device. Moreover, the non-clinical tests and the predicate comparisons demonstrate that these differences in their technological characteristics do not raise any questions as to the safety and effectiveness.

Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Kunshan Aoshida Electric Technology Co., Ltd concludes that, A08 power wheelchair is substantially equivalent to predicate devices as described herein.

{6}------------------------------------------------

PL001 (K113463)A08Comparison to thepredicate device
Indications for useThey are motor driven, indoor and outdoor transportation vehiclewith the intended use to provide mobility to a disabled or elderlyperson limited to a seated position.Same indications for use
Maximum load capacity114 kg (251 lbs)100 kg (220 lbs)Smaller load capacity
Overall length900 mm (35.4")890 mm (35")Smaller overall length
Overall width600 mm (23.6")603 mm (23.7")Larger overall width
Stowage length106 cm(41.7")324 mm(12.8")Larger stowage length
Stowage width510 mm(20.1")603 mm(23.7")Larger stowage width
Stowage height450 mm(17.7")670 mm(26.4")Larger stowage height
Rising150 mm(5.9")40 mm(1.57")Smaller rising
Total mass:with battery : 26 kg (57.2 lbs)without battery : 23 kg(50.6 lbs)with battery : 28 kg(61.7 lbs)without battery : 24 kg(53 lbs)Heavier mass
Mass of heaviest part23 kg (50.6 lbs) body framewith motor24 kg(53 lbs) body frame withmotorHeavier mass of heaviest part
Pivot width250 cm(98.4")150.5 cm(59.3")Smaller pivot width
Reversing width160 cm(63")233 cm(91.7")Larger reversing width
Turning diameter160 cm(63")Same turning diameter
Ground clearance60 mm(2.4")63.5 mm(2.5")Larger ground clearance
Required width of angled corridor120 cm(47.2")100 cm(39.4")Smaller required width ofangled corridor
Required doorway entry depth106 cm(41.7")160 cm(63")Larger required doorway entrydepth
Required corridor width for sideopening190 cm(74.8")120 cm(47.2")Smaller required corridorwidth for side opening
(Continuous)The substantial equivalence comparison of the A08 and PL001 (K113463)
PL001 (K113463)A08Comparison to thepredicate device
Seat plane angleSmaller seat plane angle
Effective seat depth350 mm(13.8")430 mm (17")Larger effective seat depth
Seat width350 mm(13.8")450 mm (17.7")Larger seat width
Effective seat width307 mm ~ 430 mm(12" ~ 17")450 mm(17.7")Larger effective seat width
Seat surface height at front edge440 mm~516 mm(17.3"~20.3")470 mm(18.5")Larger seat surface height atfront edge
Back support angle12°20°Larger back support angle
Back support height395 mm(15.6")470 mm(18.5")Smaller back support height
Back support width350 mm(13.8")440 mm (17.3")Larger back support width
Foot support to seat distance345 ~ 421 mm(13.6" ~ 16.6")400 mm(15.7")Predicate device is adjustable
Foot support clearance70 mm(2.8")55 mm(2.2")Smaller foot support clearance
Arm support height155 mm(6.1")200 mm(7.9")Larger arm support height
Front of arm support to backsupport250 mm(9.8")350 mm(13.8")Larger front of arm support toback support
Arm support length260 mm(10.2")335 mm(13.2")Larger arm support length
Arm support width50 mm(2")Same arm support width
Arm support angle12°Larger arm support angle
Distance between arm supports420 mm(16.5")480 mm(18.9")Larger distance
Horizontal location of the wheelaxle219 mm(8.6")120 mm(4.7")Smaller horizontal location ofthe wheel axle
Vertical location of the wheel axle284 mm(11.2")320 mm(12.6")Larger vertical location of thewheel axle
Caster wheel diameter200 mm(8")176 mm(7")Smaller diameter
(Continuous)The substantial equivalence comparison of the A08 and PL001 (K113463)
PL001 (K113463)A08Comparison to thepredicate device
Material
SeatSeat cushionBackrestPU foam covered by nylon fabric clothSame material
Differential mechanismDifferential rate: 22 : 1Same differential rate
Free-wheel modeYesBoth devices have free- wheelmode
Lights
Head/Tail lightsNoBoth devices do not have lights
Signal lightNo
Warning lightNo
HornYesBoth devices have horn
Anti-tip wheelsYesBoth devices have anti-tipwheels
ArmrestRemovable flip-up armrestBoth devices have adjustablearmrest
HeadrestNoBoth devices do not haveheadrest
Motor
Rated power150 W250 WLarger rated power
Input voltageDC 24VSame input voltage
Amount2 PcsSame amount
Brake systemIntelligent regenerative electromagnetic brakeSame brake system
(Continuous)The substantial equivalence comparison of the A08 and PL001 (K113463)
PL001 (K113463)A08Comparison to the predicate device
ControllerYisheng Electric Co. Ltd , WS-1, 40ASame controller
Rear wheel driveSealed transaxle direct driveSame drive mode
Braking distanceForward:1.5 m (59") at max speedForward: 1 m (39.4") at max speedSmaller braking distance
Per-charge distanceUp to 20 km (12.5 miles)Same pre-charging distance
Maximum speedUp to 6 km/h (3.75 mph) , variableUp to 7 km/hr (4.4 mph), variableLarger max. speed
Maximum curb height30 mm (1.2")40 mm(1.57")Larger curb height
Battery
TypeLi-ion, RechargeableSame type
Rated power20 AhSame rated power
Output voltage24 VDCSame output voltage
AmountDC 12V x 2 PcsSame amount
Battery level indicatorYesBoth devices have battery level indicator
Charger
Typeoff-board, Automatic TypeSame type
Main plugStandard US-plug, 2 pinsSame type
Input power requirement110-220 V / 50-60 HzSame input power requirement
Output voltage / currentDC 24V / 2ASame output voltage / current
(Continuous)The substantial equivalence comparison of the A08 and PL001 (K113463)
PL001 (K113463)A08Comparison to the predicate device
Front wheels
Diameter8" × 2"7" × 2"Smaller diameter
typenon-pneumatic tiresnon-pneumatic tiresSame type
MaterialPolyurethane (PU)Polyurethane (PU)Same materials
Amount22Same amount
Rear wheels
Diameter8" × 2.4"10" × 2.35"Larger diameter
typenon-pneumatic tiresnon-pneumatic tiresSame type
MaterialPolyurethane (PU)Polyurethane (PU)Same materials
Amount22Same amount

The substantial equivalence comparison of the A08 and PL001 (K113463)

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).