K Number
K241109

Validate with FDA (Live)

Date Cleared
2024-08-22

(122 days)

Product Code
Regulation Number
870.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Single-use Sterile High-pressure Angiographic Syringe and Accessories are intended for the injection of contrast media or saline, they shall be used with U.S. legally marketed angiographic injectors.

Device Description

The proposed device is intended for the injection of contrast media or saline. It includes disposable syringes, connection tube, J shape tube, and spike.

Syringe: The syringes are intended to be used with an U.S. legally marketed angiographic injector.

Compatibility is shown in the Table 1 Compatibility between Syringe and Injectors. (Note: the newly added and the difference between K211564 and the proposed device is highlighted in Red.)

Connection tube: it is used to connect the syringe and the catheter. The tubes (extension tube) are also available in various configurations, which are Type B (used with single shot syringe, highlighted in Red mark), renamed "Straight tube" to "Type B" and added a new model (400204) connecting tube in the Table. The product structure, manufacturing process, product materials have not changed. Type Y and type T tube are used with dual shots syringe. The pressure specification for connection tube is provided in Table 2 Pressure Specifications for Connection Tube. (Note: the newly added and the difference between K211564 and proposed device is highlighted in Red)

J shape tube: it is used to draw contrast media/saline into the syringe barrel before installation.

Spike: it is used to draw contrast media/saline into the syringe barrel before the syringe installation. The pressure specification for the spike is provided in the following table as Table 3 Pressure Specifications for Spike (Note: this model is only used to distinguish whether it is made with or without DEHP). The product materials, manufacturing process and product structure have not changed. Spikes not made with DEHP are marked with Red. The new additions of spike models are: 700204- 1, 700204-2,700205-1, 700205-2, 700206, 700207-1, 700207-2)

AI/ML Overview

This document is an FDA 510(k) summary for a medical device, specifically Single-use Sterile High-pressure Angiographic Syringes and Accessories. It outlines the comparison between the proposed device and a legally marketed predicate device to establish substantial equivalence.

Important Note: This document describes a medical device clearance process for physical hardware (syringes and accessories), not an AI/Software as a Medical Device (SaMD). Therefore, many of the requested categories related to AI/MRMC studies, expert consensus on images, ground truth for training/test sets, etc., are not applicable to this type of submission. The 'acceptance criteria' here refer to meeting engineering and biocompatibility standards, and 'proving the device meets the acceptance criteria' is done through non-clinical laboratory testing.

Here's an attempt to answer your questions based on the provided text, while acknowledging that many are not relevant for this device type:


1. A table of acceptance criteria and the reported device performance

The acceptance criteria are primarily defined by adherence to recognized international and ASTM/USP standards for medical devices and the ability to withstand specified pressures. The reported performance confirms compliance with these standards and pressure specifications.

Acceptance Criteria CategorySpecific Acceptance Criteria (from standards/specifications)Reported Device Performance/Compliance
BiocompatibilityISO 10993-4, -5, -7, -10 standards; USP-NF <151> Pyrogen Test; ASTM F756 Hemolysis AssessmentNo Cytotoxicity, No Irritation, No Sensitization, No Pyrogen, No Acute Toxicity, No Hemolysis
Material SafetyEthylene Oxide sterilization residuals (ISO 10993-7)Complies with ISO 10993-7
Sterility & PackagingSeal strength (ASTM F88/F88M-23), Seal leaks (ASTM F1929-15), Bacterial Endotoxin Limit (USP-NF <85>)Demonstrated compliance with specified standards
Syringe PerformanceISO 7886-1:2017 (manual), ISO 7886-2:2020 (power-driven) Including: Lubricant weight, Graduated capacity tolerance, overall length of scale, syringe dimension, nozzle lumen, dead space, liquid leakage, air leakage, flow characteristics.Complies with ISO 7886 standards; Meets performance for listed characteristics.
Connection IntegritySmall-bore connectors (ISO 80369-7:2021)Complies with ISO 80369-7
Pressure WithstandingSyringes: 300psi, 400psi, 1200psi (depending on model) Connection tube: 400psi, 1200psi (depending on model) Spike: 400psiEach device meets performance under maximum sustained pressure specifications. Test reports indicated compliance.
Shelf Life5-year stability5-year shelf life
CompatibilityCompatibility with U.S. legally marketed angiographic injectors (listed in Table 1)Compatibility tests passed, demonstrating device meets specified performance requirements.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document states "Nonclinical tests were conducted to verify that the proposed device met all design specifications." and refers to compliance with ISO, ASTM, and USP standards. However, specific sample sizes for each non-clinical test are not disclosed in this summary. The data provenance is internal testing conducted by Shenzhen Boon Medical Supply Co., Ltd. (China). This would be prospective, laboratory-based testing of manufactured units, not retrospective analysis of patient data.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

Not Applicable. This is a hardware medical device; no "ground truth" based on expert medical opinion (like for imaging diagnosis) is established for the purpose of these non-clinical tests. The "truth" is whether the device meets engineering specifications and safety standards as measured by laboratory equipment.


4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not Applicable. As this is a hardware device undergoing non-clinical engineering and biological safety tests, there is no adjudication process involving multiple human reviewers for a "test set" in the context of diagnostic accuracy. Test results are objective measurements against defined standards.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not Applicable. This device is a physical syringe and accessories, not an AI software or system that assists human readers in diagnosis. No MRMC study was performed.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not Applicable. This is not an algorithm or software device.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context is defined by international, national, and industry-recognized standards and specifications (e.g., ISO, ASTM, USP) for medical device performance, safety, and biocompatibility. Compliance is determined by objective laboratory measurements and tests against these established benchmarks.


8. The sample size for the training set

Not Applicable. As this is a hardware device, there is no "training set" in the context of machine learning or AI.


9. How the ground truth for the training set was established

Not Applicable. There is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 22, 2024

Shenzhen Boon Medical Supply Co., Ltd. Shengyu Fan International Regulation Commissioner No.18 Jirong Road, Shenkeng, Henggang Street Longgang District Shenzhen, Guangdong 518173, China

Re: K241109

Trade/Device Name: Single-use Sterile High-pressure Angiographic Syringes and Accessories Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: July 20, 2024 Received: July 22, 2024

Dear Shengyu Fan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Shruti N. Mistry -S

Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K241109

Device Name

Single-use Sterile High-pressure Angiographic Syringes and Accessories

Indications for Use (Describe)

Single-use Sterile High-pressure Angiographic Syringe and Accessories are intended for the injection of contrast media or saline, they shall be used with U.S. legally marketed angiographic injectors.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

    1. Date of Preparation: March 3, 2024

2. Applicant Identification: Shenzhen Boon Medical Supply Co., Ltd. No.18 Jirong Road, Shenkeng, Henggang Street, Longgang District Shenzhen Guangdong, China 518173

Establishment Registration Number: 3012395857 Contract Person: Gui Wenjuan

Position: Management Representative Tel: +86-755-28638515

Fax: +86-755-28638033 E-mail: guiwenjuan@szboon.com

Subject Device Information:

The assigned 510(k) Number: K241109 Trade Name: Single-use Sterile High-pressure Angiographic Syringes and Accessories Common Name: Disposable angiographic syringe Models

Regulatory Information:

Classification Name: Injector And Syringe, Angiographic Classification: II Product Code: DXT Regulation Number: 21 CFR 870.1650 Review Panel: General Hospital

3. Identification of Predicate Device

510(k) Number: K211564 Product Name: Sterile High-pressure Angiographic Syringes for Single-use Manufacturer: Shenzhen Boon Medical Supply Co., Ltd.

    1. Designated Submission Correspondent Shengyu Fan (Primary Contact Person) E-mail:

faguibu07@szboon.com

DARCO International Co., Ltd. 510 Shotgun Rd. Suite 520, Sunrise, FI 33326

Tel: 954-380-8595

Fax: 877-804-4505

{4}------------------------------------------------

E-mail: ahr@darcointernational.net

{5}------------------------------------------------

5. Description of Proposed Device

Product NameModels
Syringe100101, 100103, 100113, 200101, 200102, 300101, 300105, 100104,100114, 200103, 300103, 100111, 100121, 100123, 100124, 100125,200104, 100108, 100118, 100129, 100130, 100109, 100119, 200107,300102, 300108, 300110, 300111, 300112, 100110, 100120, 100112,100122, 300107, 200105, 200113, 200110, 200111
Connection Tube400101, 400102, 400103, 600101, 600102, 500105, 500106, 500107,500108, 400201, 400202, 400203, 400204, 600201, 600202, 500101,500102, 500103, 500104, 500201, 500202, 500203, 500204, 500205,500206, 500207, 500208
J Shape tube700103
Spike700101, 700102, 700104-1, 700104-2,700105-1, 700105-2, 700106, 700107-1, 700107-2,700204-1, 700204-2,700205-1, 700205-2, 700206, 700207-1, 700207-2

Indications for Use:

Single-use Sterile High-pressure Angiographic Syringe and Accessories are intended for the injection of contrast media or saline, they shall be used with U.S. legally marketed angiographic injectors.

Device Description:

The proposed device is intended for the injection of contrast media or saline. It includes disposable syringes, connection tube, J shape tube, and spike.

Syringe: The syringes are intended to be used with an U.S. legally marketed angiographic injector.

Compatibility is shown in the Table 1 Compatibility between Syringe and Injectors. (Note: the newly added and the difference between K211564 and the proposed device is highlighted in Red.)

{6}------------------------------------------------

510(K) Summary Table 1 Compatibility between Syringe and Injectors

Model(Syringe)Volume (ml)TypeResistantliquid leakpressure (psi)Injector
100101200mlSingle shot400MCT&MCT Plus CT, K924116Vistron CT, K991557EnVision CT, K934086
100103200mlSingle shot400Stellant-S, K182273
100113200/200mlDual shots400Stellant-D, K182273
20010165/65mlDual shots300Spectris, K935668
20010265/115mlDual shots300Solaris MRI, K033247
300101150mlSingle shot1200Mark V, K822536
300105130mlSingle shot1200Mark III&Mark IV, K822536
100104200mlSingle shot400CT 9000&CT9000 ADV, K912944CT OptiVantage, K063503
100114200/200mlDual shots400CT 9000&CT9000 ADV, K912944CT OptiVantage, K063503
20010360/60mlDual shots300Optistar LF, Elite, K073592
300103150mlSingle shot1200ILLUMENA, K963071
100111200mlSingle shot400Empower CT, K071378
100121200/200mlDual shots400Empower CT, K071378
100123100/200mlDual shots400Dual Shots, K052633Dual shots Alpha, K062168
10012460/100mlDual shots400Dual Shots, K052633Dual shots Alpha, K062168
10012560/200mlDual shots400Dual Shots, K052633Dual shots Alpha, K062168
20010460/60mlDual shots300Sonic Shot, K091743
100108200mlSingle shot400Dual shots CT, K062168Dual shots Alpha, K062168
100118200/200mlDual shots400Dual shots CT, K133189Dual shots Alpha, K062168
100129125mlSingle shot400CT Optione, K152361 CTOptiVantage, K063503 CT 9000&CT9000ADV,K912944
100130125/125mlDual shots400CT Optione, K152361 CTOptiVantage, K063503 CT 9000&CT9000ADV,K912944
100109100mlSingle shot400Dual Shots, K062168Dual shots Alpha, K062168
100119100/100mlDual shots400Dual Shots, K062168Dual shots Alpha, K062168
200107100/100mlDual shots300EZEM Empower MR, K062449
300102125mlSingle shot1200120S, K092896
300108150mlSingle shot1200Mark VII, K122086
300110150mlSingle shot1200Rempress, K092896
300111150mlSingle shot1200Angiomat 6000, K944875
300112200mlSingle shot1200Illumena, K963071
100110200mlSingle shot400Medrad V Plus, K903390
100120200/200mlDual shots400Medrad V Plus, K903390
100112200mlSingle shot400Accutron CT& Accutron CT-D
100122200/200mlDual shots400Accutron CT-D
300107200mlSingle shot1200Accutron HP& Accutron HP-D
20010565/65mlDual shots300Accutron MR,
20011365/200mlDual shots300Accutron MR,
20011065/115mlDual shots300Medrad MRXperion,K182276
200111115mlSingle shot300Spectris&Solaris, K033247

{7}------------------------------------------------

Connection tube: it is used to connect the syringe and the catheter. The tubes (extension tube) are also available in various configurations, which are Type B (used with single shot syringe, highlighted in Red mark), renamed "Straight tube" to "Type B" and added a new model (400204) connecting tube in the Table. The product structure, manufacturing process, product materials have not changed. Type Y and type T tube are used with dual shots syringe. The pressure specification for connection tube is provided in Table 2 Pressure Specifications for Connection Tube. (Note: the newly added and the difference between K211564 and proposed device is highlighted in Red)

ModelMaximum Withstanding Pressure(psi)Type
400101400Type B
400102400Type Y
400103400Type T
600101400Type T
600102400Type Y
5001051200Type B

Table 2 Pressure Specifications for Connection Tube

{8}------------------------------------------------

510(K) Summary
5001061200Type B
5001071200Type B
5001081200Type B
400201400Type B
400202400Type Y
400203400Type T
400204400Type B
600201400Type T
600202400Type Y
5001011200Type B
5001021200Type B
5001031200Type B
5001041200Type B
5002011200Type B
5002021200Type B
5002031200Type B
5002041200Type B
5002051200Type B
5002061200Type B
5002071200Type B
5002081200Type B

J shape tube: it is used to draw contrast media/saline into the syringe barrel before installation.

Spike: it is used to draw contrast media/saline into the syringe barrel before the syringe installation. The pressure specification for the spike is provided in the following table as Table 3 Pressure Specifications for Spike (Note: this model is only used to distinguish whether it is made with or without DEHP). The product materials, manufacturing process and product structure have not changed. Spikes not made with DEHP are marked with Red. The new additions of spike models are: 700204- 1, 700204-2,700205-1, 700205-2, 700206, 700207-1, 700207-2)

Table 3 Pressure Specifications for Spike
Model (containsDEHP)OriginalModel (notmade with DEHP)NewaddedMaximumWithstandingPressure (psi)Type
700101//Long Spike
700102//Short Spike
700104-1700204-1400Single AirChamber Transfer Set

{9}------------------------------------------------

700104-2700204-2400Single Air ChamberTransfer Set withExtensionTube
700105-1700205-1400Dual Air ChamberTransfer Set
700105-2700205-2400Dual Air ChamberTransfer Set withExtension Tube
700106700206/Transfer Set withC-Clamp
700107-1700207-1/Transfer Set withClave Connector
700107-2700207-2/Clave connectoransfer set with Checkvalve

6. Non-Clinical Test Conclusion

Nonclinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-7:2008 Biological evaluation of Medical device- Part 7: Ethylene oxide sterilization residuals;

ASTM F88/F88M-23 Standard test method for seal strength of flexible barrier materials: ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration;

USP-NF<85> Bacterial Endotoxin Limit;

ISO 7886-1: 2017 Sterile hypodemic syringes for Single-use-Part 1: Syringes for manual use; ISO 7886-2: 2020 Sterile hypodemic syringes for Single-use-Part 2: Syringes for use with power-driven syringe pumps;

  • The test items include Lubricant weight, Graduated capacity tolerance, overall length of । scale, syringe dimension, nozzle lumen, dead space, liquid leakage, air leakage and flow characteristics;
    ISO 80369-7: 2021 Small-bore connectors for liquids and gases in healthcare applications -Part 7: Connectors for intravascular or hypodermic applications ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for In Vitro Cytotoxicity;

ISO 10993-10:2021 Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization;

ISO 10993-4: 2017 Biological Evaluation of Medical Device -Part 4: Selection of tests for interactions with blood:

ASTM F756: 2017, Standard Practice for Assessment of Hemolytic Properties of Materials USP-NF <151> Pyrogen Test

Compatibility Test Report includes information between injectors, syringe, connection tube and spike/J shape tube. The compatibility test demonstrated that each device meets performance

{10}------------------------------------------------

under maximum sustained pressure specifications.

    1. Clinical Test Conclusion
      No clinical study is included in this submission.

{11}------------------------------------------------

8. Substantially Equivalent (SE) Comparison

Table 4 Comparison of Technology Characteristics

ItemProposed Device K241109Predicate Device K211564Comments
Product CodeDXTDXTSame
RegulationNumberCFR 870.1650CFR 870.1650Same
Indications for UseThe proposed device, Single-useSterile High-pressure AngiographicSyringes and Accessories areintended for the injection of contrastmedia or saline, they shall be usedwith U.S. legally marketedangiographic injectors.The predicate device, Sterile High-pressure Angiographic Syringesfor Single-use are intended for theinjection of contrast media or saline;they shall be used with an US legallymarketed angiographic injectorsDifferent 1
Prescription only(Rx)Prescription only (Rx)Prescription only (Rx)Same
Mode of operationPower-driven operation, single-usePower-driven operation, single-useSame
ConfigurationAngiographic SyringeAngiographic SyringeSame
Connection tubeConnection tubeSame
J shape tube/ SpikeJ shape tube/ Spike
SterilityEO SterilizedEO SterilizedSame
Single UseYesYesSame
Syringe300psi, 400psi, 1200psi300psi, 400psi, 1200psiSame
Connection tube300psi, 400psi, 1200psiNew model added (400204).300psi, 400psi, 1200psiDifferent 2
J shape tube//
Spike400psi400psiSame
Syringe Specification(Volume/ ml)200, 150, 125, 130, 100, 200/200,60/100, 125/125, 100/100, 65/65, 65/115, 60/60, 50/50, 100/200, 115,65/20011 new models added.200, 150, 125, 130, 100, 200/200, 60/100,125/125,100/100,65/65,65/115,60/60,50/50Different 3
Connection tube Specification(overall length,mm)2002500, 15002500, 1500, 1800,2000, 2500, 500, 750, 1000, 12002002500, 15002500, 1500, 1800, 2000,2500, 500, 750, 1000, 1200Same
J shape tube Specification(overall length, mm)240240Same
Spike Specification (overalllength,mm)58.8,47.3, 1000,2800, 1200,2900,180,260,340,420,500,450,550,60058.8,47.3, 1000,2800, 1200,2900, 180,260,340,420,500,450,550,600Same
Performance
SyringeISO 7886ISO 7886Same
Luer connectorISO 80369-7ISO 80369-7Same
CompatibilityPassPassSame
Patient-Contact Material
SyringeBarrelPP (Polypropylene) or PET(Polyethylene terephthalate)PP (Polypropylene) or PET (Polyethyleneterephthalate)Same
SyringePistonPolyisoprene rubberPolyisoprene rubberSame
SyringeLubricantPolydimethylsiloxanePolydimethylsiloxaneSame
Connection tubeTubingPVC (Polyvinylchloride with DEHP)or PVC(Polyvinylchloride without DEHP)or PU (Polyurethane)PVC (Polyvinylchlori de with DEHP) orPVC(Polyvinylchloride without DEHP) orPU (Polyurethane)Same
Connection tubeLuerconnectorsPC (Polycarbonate)PC (Polycarbonate)Same
Connection tubeUVadhesiveUltraviolet adhesiveUltraviolet adhesiveSame
SpikeClosure-piercingdeviceABS (acrylonitrile-butadiene-styrene)ABS (acrylonitrile-butadiene-styrene)Same
SpikeFiltermembranePP(polypropylene)PP(polypropylene)Same
SpikeTubePolyvinyl chloride (PVC)Polyvinyl chloride (PVC)Same
SpikeLuerconnectorcarbonate (PC)carbonate (PC)Same
SpikeProtectivecapAcrylonitrile-butadiene-styrene (ABS)Acrylonitrile-butadiene-styrene (ABS)Same
J shape tubePE (Polyethylene)PE (Polyethylene)Same
No CytotoxicityNo CytotoxicitySame
No IrritationNo IrritationSame
BiocompatibilityNo SensitizationNo SensitizationSame
No PyrogenNo PyrogenSame
No Acute ToxicityNo Acute ToxicitySame
No HemolysisNo HemolysisSame
Endotoxin Limit20 EU per device20 EU per deviceSame
Shelf Life5-year5-yearSame

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

    1. Equivalence Discussion:
      Different #1: Compared to the predicate device, the proposed device changed the product name from Sterile High-pressure Angiographic Syringes for Single-use to: Single-use Sterile Highpressure Angiographic Syringes and Accessories. The indications for use of the product share the same contents.

Different #2: Compared to the predicate device, a new connection tube model 400204 is introduced to the proposed device. Model 400204 is almost identical to 400202, the only difference is that the former has one more check valve than the latter. And the materials, manufacturing process, pressure rating (up to 400 psi, see compatibility test report), wall thickness (0.70±0.05mm) and length (200mm - 2500 mm) of 400204 are the same with the existing models of 400201, 400202 and 400203. The results of performance tests completed on the proposed device demonstrate that the proposed devices comply with FDA recognized standards, which the predicate device also complied with. The results of biocompatibility studies performed on the proposed device demonstrate that the patient materials used in the proposed device are biocompatible

Different #3: Compared the predicated device, the specifications of syringes have added 115ml, 100/200ml and 65/200ml in the proposed device, and the compatibility table between the proposed device and predicate device are shown in Table 1. The difference and the newly added devices are marked in Red.

The results of performance tests completed on the proposed device demonstrate that the proposed devices comply with FDA recognized standards, which the predicate device also complied with. The results of biocompatibility studies performed on the proposed device demonstrate that the patient materials used in the proposed device are biocompatible.

    1. Substantial Equivalence Conclusion:
      Based on the comparison above, the differences between the subject and predicate devices do not raise any new or different questions of safety and effectiveness when compared to the predicate device. The proposed device, Single-use Sterile High Pressure Angiographic Syringes and Accessories is determined to be Substantially Equivalent (SE) to the predicate device.

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.