AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

High-speed Air Turbine Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Dental Low Speed Handpieces including Straight and Geared Angle Handpieces and Air Motors: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.

Device Description

The proposed device is used to hold rotary instruments such as burs and drills to drive movements to achieve dental operations such as cutting, and drilling. The working principle of the proposed device is that the current and compressed air arrive at the nose through each control valve of the circuit system and air circuit system to drive the motor, turbine (or blade) rotation (shaft core rotation), thus driving the needle rotation to achieve the purpose of drilling and tooth cutting, and the other water and air meet the head to achieve the cooling purpose through each control valve.

AI/ML Overview

The provided text is a 510(k) Summary for dental handpieces, which focuses on establishing substantial equivalence to a predicate device rather than presenting a performance study with specific acceptance criteria and detailed device performance metrics.

Therefore, I cannot extract the information required to populate a table of acceptance criteria and reported device performance directly from this document. The document primarily highlights that the proposed device met design specifications and performed as well as the legally marketed predicate device based on non-clinical tests compliant with various ISO and IEC standards. It explicitly states, "No clinical study is included in this submission."

Given this, I am unable to provide a response that directly answers your request for acceptance criteria and device performance from a clinical study, sample sizes for test sets, expert involvement, adjudication methods, MRMC study details, standalone performance, or training set specifics, as these are details typically found in clinical performance studies, not in this type of 510(k) submission.

The document focuses on non-clinical testing for safety and effectiveness, and comparability to a predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.

November 12, 2024

Foshan SOCO Precision Instrument Co.,Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, 102401 China

Re: K241074

Trade/Device Name: High-speed Air Turbine Handpieces (SCHG01, SCHG03, SCHG18, SCHF19, SCHF22), Dental Low Speed Handpieces including Straight and Geared Angle Handpieces and Air Motors (Straight Handpieces: SCHD05-S, SCHD16-S, SCHD17-S, SCHD24-S, SCHD27-S Geared Angle Handpieces: SCHD05-C, SCHD06-C, SCHD16-C, SCHD17-C, SCHD24-C, SCHD27-C, SCHD27-C, SCHD29-S, SCHB4, SCHB5, SCHB20, SCHB64 Air Motors: SCHD16-M, SCHD17-M, SCHD24-M, SCHD27-M) Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB, EGS

Dated: October 14, 2024 Received: October 15, 2024

Dear Ray Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical

{2}------------------------------------------------

devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bobak Shirmohammadi -S

For Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241074

Device Name

High-speed Air Turbine Handpieces (SCHG01, SCHG03, SCHG18, SCHF19, SCHF22); Dental Low Speed Handpieces including Straight and Geared Angle Handpieces and Air Motors (Straight Handpieces: SCHD05-S, SCHD16-S, SCHD17-S, SCHD24-S, SCHD27-S Geared Angle Handpieces: SCHD05-C, SCHD06-C, SCHD16-C, SCHD17-C, SCHD24-C, SCHD27-C, SCHD29-S, SCHB4, SCHB5, SCHB20, SCHB64 Air Motors: SCHD16-M, SCHD17-M, SCHD24-M, SCHD27-M) Indications for Use (Describe) High-speed Air Turbine Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Dental Low Speed Handpieces including Straight and Geared Angle Handpieces and Air Motors: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

The assigned 510(k) Number:

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

  • Date of Preparation: 10/14/2024 1.
    1. Sponsor Identification

Foshan SOCO Precision Instrument Co.,Ltd.

3-5 FL. Bldg 4, District A and 5 FL. Bldg 3, District A Guangdong New Light Source Industrial Base, Luocun, Shishan Town, Nanhai District, Foshan City, 52826, Guangdong, P.R. China Contact Person: Xiao Gou Position: Regulator's Representative/PRRC Tel: +86-13876245168 Fax: +86-0757-81806003 Email: 824283813@qq.com

    1. Designated Submission Correspondent
      Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China, 102401 Contact Person: Ray Wang Position: General Manager Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com
Trade NameModel
High-speed Air Turbine HandpiecesSCHG01, SCHG03, SCHG18, SCHF19, SCHF22
Dental Low Speed Handpiecesincluding Straight and Geared AngleHandpieces and Air MotorsStraight Handpieces:SCHD05-S, SCHD16-S, SCHD17-S, SCHD24-S, SCHD27-SGeared Angle Handpieces:SCHD05-C, SCHD06-C, SCHD16-C, SCHD17-C,SCHD24-C, SCHD27-C, SCHD29-S.

4. Identification of Proposed Device

{5}------------------------------------------------

SCHB4, SCHB5, SCHB20, SCHB64
Air Motors:
SCHD16-M, SCHD17-M, SCHD24-M, SCHD27-M

Common Name: Handpiece, Air-Powered, Dental

Regulatory Information

Classification Name: Handpiece, Air-Powered, Dental Classification: I Product Code: EFB; EGS Regulation Number: 21 CFR 872.4200 Regulation Name: Dental handpiece and accessories Regulation Medical Specialty: Dental Review Panel: Dental

Indication For Use Statement:

High-speed Air Turbine Handpieces:

It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.

Dental Low Speed Handpieces including Straight and Geared Angle Handpieces and Air Motors:

It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.

న. Device Description

The proposed device is used to hold rotary instruments such as burs and drills to drive movements to achieve dental operations such as cutting, and drilling. The working principle of the proposed device is that the current and compressed air arrive at the nose through each control valve of the circuit system and air circuit system to drive the motor, turbine (or blade) rotation (shaft core rotation), thus driving the needle rotation to achieve the purpose of drilling and tooth cutting, and the other water and air meet the head to achieve the cooling purpose through each control valve.

6. Identification of Predicate Device(s)

Predicate Device 510(k) Number: K220179 Product Name: High-Speed Air Turbine Handpieces / Straight Handpieces / Geared Angle Handpieces / Air Motors Manufacturer: Foshan COXO Medical Instrument Co., Ltd.

{6}------------------------------------------------

  • Non-Clinical Test Conclusion 7.
    Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • A ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

  • ISO 10993-10: 2010 Biological evaluation of medical devices Part 10: Tests for irritation > and skin sensitization

  • A ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

  • Pyrogen Study in Rabbits using USP General Chapter <151> Pyrogen Test A

  • IEC 60601-1: 2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

  • ISO 60601-1-2: 2020 Medical electrical equipment Part 1-2: General requirements for A basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests

  • IEC 80601-2-60: 2019 Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment

  • ISO 14457: 2017 Dentistry - Handpieces and motors

  • A ISO 17664: 2017 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices

  • ISO 17665-1: 2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices

  • A ISO 9168: 2009 Dentistry - Hose connectors for air driven dental handpieces

  • Clinical Test Conclusion 8.

No clinical study is included in this submission.

  • Substantially Equivalent (SE) Comparison 9.

Table 1 General Comparison

ItemProposed Device (K241074)Predicate Device (K220179)Remark
ManufacturerFoshan SOCO Precision Instrument Co.,Ltd.Foshan COXO Medical Instrument Co., Ltd./
Device NameHigh-speed Air Turbine Handpieces / Dental LowSpeed Handpieces including Straight and GearedAngle Handpieces and Air MotorsHigh-Speed Air Turbine Handpieces/ StraightHandpieces/ Geared Angle Handpieces/ Air Motors/

{7}------------------------------------------------

ModelHigh-speed Air Turbine Handpieces:SCHG01, SCHG03, SCHG18, SCHF19, SCHF22Dental Low Speed Handpieces including Straightand Geared Angle Handpieces and Air Motors:Straight Handpieces:SCHD05-S, SCHD16-S, SCHD17-S, SCHD24-S,SCHD27-SGeared Angle Handpieces:SCHD05-C, SCHD06-C, SCHD16-C, SCHD17-C,SCHD24-C, SCHD27-C, SCHD29-S,SCHB4, SCHB5, SCHB20, SCHB64Air Motors:SCHD16-M, SCHD17-M, SCHD24-M, SCHD27-MHigh-Speed Air Turbine Handpieces:CX207, CX207-G, CX207-2, CX207-A, CX207-A-2,CX207-B, CX207-B-2, CX207-C, CX207-C-2,CX207-W, CX207-W-2Straight Handpieces:CX235-2, CX235-2A, CX235-2B, CX235-2F,CX235-2G, CX235-2C, CX235-2S2Geared Angle Handpieces:CX235-1B, CX235-1C, CX235-1F, CX235-1G,CX235C1, CX235C2, CX235C3, CX235C4,CX235C5, CX235C6, CX235C7, CX235C8,CX235-2S, CX235-2S1Air Motors:CX235-3B, CX235-3F, CX235-3CAnalysis 1ProposedPredicateProposedPredicateProposedPredicateProposedPredicateProposedPredicate
Product CodeEFB, EGSEFB, EGSSAMEDeviceDeviceDeviceDeviceDeviceDeviceDeviceDeviceDeviceDevice
Regulation No.21 CFR 872.420021 CFR 872.4200SAMEItem(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)Remark
ClassIISAMEModel:Model:Model:Model:Model:Model:Model:Model:Model:Model:
IndicationFor UseHigh-speed Air Turbine Handpieces:It is intended for removing carious material, excessfilling material, cavity and crown preparation,finishing tooth preparations and restorations, rootcanal preparations and polishing teeth.Dental Low Speed Handpieces including Straightand Geared Angle Handpieces and Air Motors:It is intended for removing carious material, excessfilling material, cavity and crown preparation,finishing tooth preparations and restorations, rootcanal preparations and polishing teeth.High-Speed Air Turbine Handpieces:It is intended for removing carious material, excessfilling material, cavity and crown preparation,finishing tooth preparations and restorations, rootcanal preparations and polishing teeth.Straight Handpieces:It is intended for removing carious material, excessfilling material, cavity and crown preparation,finishing tooth preparations and restorations, rootcanal preparations and polishing teeth.Geared Angle Handpieces:It is intended for removing carious material, excessfilling material, cavity and crown preparation,finishing tooth preparations and restorations, rootcanal preparations and polishing teeth.Air Motors:It is intended for removing carious material, excessfilling material, cavity and crown preparation,finishing tooth preparations and restorations, rootcanal preparations and polishing teeth.SAMESCHG01CX207SCHG03CX207-WSCHG18CX207-ASCHF19CX207-GSCHF22CX207-W
Air/water ports4/6 ports2/4/6 portsSAMEAngle20°20°20°20°20°20°20°20°45°20°Analysis 3
Fiber optic GlassRodSCHB5, SCHD24-S, SCHD24-CCX207-G, CX235-1C, CX235C7Analysis 1CartridgeCollet/OpenCollet/OpenOpenOpenOpenOpenOpenOpenCollet/OpenOpenSAME
Composition ofmaterialCopper (chromium plating), aluminum,rubber and stainless steelCopper (chromium plating), aluminum,rubber and stainless steelSAMEOperational modesAir-poweredAir-poweredAir-poweredAir-poweredAir-poweredAir-poweredAir-poweredAir-poweredAir-poweredAir-poweredSAME
ChuckPush button, Latch, WrenchPush button, ScrewAnalysis 2Chuck TypePush buttonPush buttonPush buttonPush buttonPush buttonPush buttonPush buttonPush buttonPush buttonPush buttonSAME
Spray1 way spray1 way spray3 way spray3 way spray1 way spray1 way spray3 way spray3 way spray3way spray3 way spraySAME
Maximum airpressureHigh-speed Air Turbine Handpieces: 0.2 0.25MpaDental Low Speed Handpieces including Straight andGeared Angle Handpieces and Air Motors: 0.245MPa 0.3925MPaHigh-speed Air Turbine Handpieces: 0.2 ~0.25MpaStraight handpiece & Geared Angle handpiece & AirMotors: 0.245MPa ~ 0.3925MPaSAMEHole4 hole4 hole4 hole4 hole4 hole4 hole6 hole6 hole4 hole4 holeSAME
Maximum waterpressureHigh-speed Air Turbine Handpieces: 0.2 0.25MpaDental Low Speed Handpieces including Straight andGeared Angle Handpieces and Air Motors: 0.245MPa 0.3925MPaHigh-Speed Air Turbine Handpieces: 0.2 ~0.25MpaStraight handpiece & Geared Angle handpiece & AirMotors: 0.245MPa ~ 0.3925MPaSAMELED//////LEDLEDLED/Analysis 3
Speed in rpmsHigh-speed Air Turbine Handpieces:280000 ~ 450000rpmDental Low Speed Handpieces including Straight andGeared Angle Handpieces and Air Motors:< 40000rpmHigh-speed Air Turbine Handpieces:280000 ~ 450000rpmStraight handpiece & Geared Angle handpiece& Air Motors: < 40000rpmSAMELED Voltage//////2.9-3.2V2.9-3.2V//SAME
Conformancewith standards forshanksType 3 per ISO 1797-1Type 3 per ISO 1797-1SAMELight Intensity//////>7000 lux>7000 lux>7000 lux/Analysis 3
CouplingdimensionsComplied with ISO 3964Complied with ISO 3964SAMEFiber//////////SAME
Hose connectionsType 2 for 4- hole modelType 3 for 6-hole modelType 1 for 2-hole model / Type 2 for 4- holeModel / Type 3 for 6-hole modelSAMEProposed Device (K241074)Predicate Device (K220179)Remark
LubricantThe lubricants approved by FDA, such as DO-ALLDental Handpiece Lubricant (K073353)The lubricants approved by FDA, such as DO-ALLDental Handpiece Lubricant (K073353)SAMEHigh-speed Air Turbine HandpiecesHigh-Speed Air Turbine Handpieces
ComplianceStandardsComplied with ISO 10993- 5, ISO 10993-10, ISO10993-11, USP <151> Pyrogen Testing, IEC 60601-1,IEC 60601-1-2, IEC 80601- 2-60, ISO 14457, ISO17664, ISO 17665-1, ISO 9168Complied with ISO 10993- 5, ISO 10993-10, ISO10993-23, ISO 10993-11, USP <151> PyrogenTesting, IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60, ISO 14457, ISO 17664, ISO 17665-1, ISO 9168SAMEGripComplied with ISO 1797-1Complied with ISO 1797-1SAME
SizeComplied with ISO 1797-1Complied with ISO 1797-1SAME

{8}------------------------------------------------

Analysis 1

The comparison between the proposed device and each model of the predicate device is shown in Tables 2- Table 5.

Analysis 2

The proposed device is different in Chuck from the predicate device. However, the different not affect the safety and effectiveness of proposed device. Based on the nonclinical tests performed, the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device.

{9}------------------------------------------------

Table 2 Technical comparison for High-speed Air Turbine Handpieces

{10}------------------------------------------------

Rotation Speed280000 ~ 450000rpm280,000 rpm ~ 450,000 rpmSAME
Coolant Air Supply> 1.5L/min (at 200kPa)>1.5 L/min (at 200kPa)SAME
Coolant Water Supply>50 mL/min (at 200kPa)>50 mL/min (at 200kPa)SAME
Rated Torque>0.0005N .m>0.0005N.m (6 hole and 4 hole at 0.28 MPa, 2 hole at 0.22 MPa)SAME
Air Pressure0.2~0.25Mpa0.2 MPa ~ 0.25MPaSAME
Water Pressure0.2 MPa~0.25MPa0.2 MPa ~ 0.25MPaSAME
Bur extraction force22N22NSAME
Bur typeShaft diameter: 1.59 to 1.60 mm, Overall length: max. 23 mm, Shaft clampinglength: at least 11 mm, Blade diameter: max.2 mmISO 1797-1 Type 3, diameter: 1.6mm, fitting length: min. 11mm, overall length: max.23mm, working diameter: max. 2mm.SAME
Flow rate<80 NL/min(at a pressure of (300±100) kpa (3,0±1,0) bar)<80 NL/min(at a pressure of (300±100) kpa (3,0±1,0) bar)SAME

Analysis 3

The SCHF22 model of proposed device is different in Angle and LED from the CX207-W model of predicate dests performed, the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device.

ItemProposedDevice(K241074)Model:SCHD05-SPredicateDevice(K220179)Model:CX235-2AProposedDevice(K241074)Model:SCHD16-SPredicateDevice(K220179)Model:CX235-2FProposedDevice(K241074)Model:SCHD17-SPredicateDevice(K220179)Model:CX235-2BProposedDevice(K241074)Model:SCHD24-SPredicateDevice(K220179)Model:CX235-2CProposedDevice(K241074)Model:SCHD27-SPredicateDevice(K220179)Model:CX235-2GRemark
AngleStraightStraightStraightStraightStraightStraightStraightStraightStraightStraight
Table 3 Technical comparison for Straight Handpieces

{11}------------------------------------------------

Transmission Ratio1:11:11:11:11:11:11:11:11:11:1SAME
Speed< 40000 rpm< 40000 rpm< 40000 rpm< 40000 rpm< 40000 rpm< 40000 rpm< 40000 rpm< 40000 rpm< 40000 rpm< 40000 rpmSAME
Spray typeExternalExternalExternalExternalInnerInnerInnerInnerInnerInnerSAME
Fiber//////Glass rod opticGlass rod optic//SAME
LED//////////SAME
Light intensity//////>7000Lux>7000Lux//SAME
Water cooling (ml/min)////> 50 (200kPa)> 50 (200kPa)//> 50 (200kPa)> 50 (200kPa)SAME
Air cooling (L/min)////> 1.5 (200kPa)> 1.5 (200kPa)> 1.5 (200kPa)> 1.5 (200kPa)> 1.5 (200kPa)> 1.5 (200kPa)SAME
/
ItemProposed Device (K241074)Straight HandpiecesPredicate Device (K220179)Straight HandpiecesRemark
Coulping dimensionsComply with ISO 3964Comply with ISO 3964SAME
Bur extraction force>32 N>32 NSAME
Bur typeISO 1797, type 1, diameter 2.35mm, shaft cmalping length: Minimal 11mm, overalllength: max 23mm, blade diameter: max 2mmISO 1797, type 2, diameter 2.35mm, shaft cmalping length: Minimal 33mm, overalllength: max 50mm, blade diameter: max 2mm.ISO 1797-1 Type 1, diameter: 2.35mm, fiting length: min.11mm, overall length: max.23mm,working diameter: max.2mm.ISO 1797-1 Type 2, diameter: 2.35mm, fiting length: min.33mm, overall length:max.50mm, working diameter: max.2mm.SAME

Table 4-1 Technical comparison for Geared Angle Handpieces

ProposedDevice(K241074)PredicateDevice(K220179)ProposedDevice(K241074)PredicateDevice(K220179)ProposedDevice(K241074)PredicateDevice(K220179)
ItemRemark

{12}------------------------------------------------

Model:SCHD05-CModel:CX235C1Model:SCHD06-CModel:CX235C2Model:SCHD16-CModel:CX235-1FModel:SCHD17-CModel:CX235-1BModel:SCHD24-CModel:CX235-1CModel:SCHD27-CModel:CX235-1G
AngleContra angleContra angleContra angleContra angleContra angleContra angleContra angleContra angleContra angleContra angleContra angleContra angleSAME
TransmissionRatio1:11:11:11:11:11:11:11:11:11:11:11:1SAME
Speed (rpm)< 20000< 40000< 20000< 40000< 20000< 40000< 20000< 40000< 20000< 40000< 20000< 40000Analysis 4
Spray type/ExternalExternalExternalExternalExternalExternalInnerInnerInnerInnerInnerInnerSAME
Water cooling(ml/min)//////> 50(200kPa)> 50(200kPa)> 50(200kPa)> 50(200kPa)> 50(200kPa)> 50(200kPa)SAME
Air cooling(L/min)//////> 1.5(200kPa)> 1.5(200kPa)> 1.5(200kPa)> 1.5(200kPa)> 1.5(200kPa)> 1.5(200kPa)SAME
LED////////////SAME
Fiber////////Glass rodopticGlass rodoptic//SAME

Table 4-2 Technical comparison for Geared Angle Handpieces

ItemProposed DevicePredicate DeviceProposed DevicePredicate DeviceProposed DevicePredicate DeviceProposed DevicePredicate DeviceProposed DevicePredicate DeviceRemark
(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)(K241074)(K220179)
Model:Model:Model:Model:Model:Model:Model:Model:Model:Model:
SCHD29-SCX235-2SSCHB4CX235C3SCHB5CX235C7SCHB20CX235C6SCHB64CX235C8
AngleContra angleContra angleContra angleContra angleContra angleContra angleContra angleContra angleContra angleContra angleSAME

{13}------------------------------------------------

Transmission Ratio1:11:14:14:11:51:520:120:164:164:1SAME
Speed (rpm)< 20000< 40000< 5000< 10000< 100000< 200000< 1000< 2000< 313< 625Analysis 4
Spray typeExternalExternalExternalExternalInnerInnerExternalExternalExternalExternalSAME
Water cooling(ml/min)> 50 (200kPa)> 50 (200kPa)//> 50 (200kPa)> 50 (200kPa)> 50 (200kPa)> 50 (200kPa)//SAME
Air cooling (L/min)> 1.5 (200kPa)> 1.5 (200kPa)//> 1.5 (200kPa)> 1.5 (200kPa)////SAME
LED//////////SAME
Fiber////Glass rod opticGlass rod optic////SAME

Analysis 4

The proposed device is different in Speed from the predicate device. However, we tested as ISO 14457 and based on the nonchinical tests performed, the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device.

ItemProposed Device (K241074)Geared Angle HandpiecesPredicate Device (K220179)Geared Angle HandpiecesRemark
CouplingdimensionsComply with ISO 3964Comply with ISO 3964SAME
Bur extraction force>32 N>32 NSAME
Bur typeISO 1797, type 1, diameter 2.35mm, shaft clamping length: Minimal 11mm,overall length: max 23mm, blade diameter: max 2mm.ISO 1797, type 2, diameter 2.35mm, shaft clamping length: Minimal 33mmoverall length: max 50mm, blade diameter: max 2mm.ISO 1797, type 1, diameter 2.35mm, shaft clamping length: Minimal 11mm,overall length: max 23mm, blade diameter: max 2mm.ISO 1797, type 2, diameter 2.35mm, shaft clamping length: Minimal 33mm,overall length: max 50mm, blade diameter: max 2mm.SAME

Table 4-3 Technical comparison for Geared Angle Handpieces

{14}------------------------------------------------

ISO 1797, type 3, diameter1.6mm, shaft cmalping length: Minimal 11mm, overall length: max 23mm, blade diameter: max 2mm.ISO 1797, type 3, diameter1.6mm, shaft cmalping length: Minimal 11mm, overall length: max 23mm, blade diameter: max 2mm.
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ItemProposedDevice(K241074)Model:SCHD16-MPredicateDevice(K220179)Model:CX235-3FProposedDevice(K241074)Model:SCHD17-MPredicateDevice(K220179)Model:CX235-3BProposedDevice(K241074)Model:SCHD24-MPredicateDevice(K220179)Model:CX235-3CProposedDevice(K241074)Model:SCHD27-MPredicateDevice(K220179)Model:CX235-3BRemark
Spray typeExternalExternalInnerInnerInnerInnerInnerInnerSAME
Connector4 hole4 hole or 2 hole4 hole4 hole or 2 hole6 hole6 hole4 hole4 hole or 2 holeSAME
LED////LED (3.3±0.1V)LED (3.3±0.1V)//SAME
Light intensity////> 7000Lux> 7000Lux//SAME
Fiber///////SAME
ItemProposed Device (K241074)Air MotorsPredicate Device (K220179)Air MotorsRemark
RotationspeedForward: 20000Reverse20000(pm/min at 250 kPa)///12000-20000rpm at 250 kPaSAME
Cooling waterflow>50 ml/min at 250 kPa////>50 ml/min at 250 kPaSAME

Table 5 Technical comparison for Air Motors

{15}------------------------------------------------

Spray airsource>1.5 L/min at 250 kPa>1.5 L/min at 250 kPaSAME
CouplingdimensionsComply with ISO 3964Comply with ISO 3964

10. Substantially Equivalent (SE) Conclusion

The conclusions drawn from the nonomical that the device is as safe, as effective, and performs as well as the legally marketed device (K20179).

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.