K Number
K240942

Validate with FDA (Live)

Device Name
CINA-CSpine
Manufacturer
Date Cleared
2024-09-12

(160 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CINA-CSpine is a radiological computer aided triage and notification software indicated for use in the analysis of cervical spine CT images.

The device is intended to assist hospital networks and appropriately trained physician specialists by flagging and communication of suspected positive findings compatible with acute cervical spine fractures including non-displaced fracture lines and/or displaced fracture fragments.

CINA-CSpine uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device is not designed to detect vertebral compression fractures.

The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.

The results of CINA-CSpine are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images.

Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

Device Description

CINA-CSpine is a radiological computer-assisted triage and notification software device.

CINA-CSpine runs on a standard "off the shelf" server/workstation and consists of CSpine Image Processing Application, which can be integrated, deployed and used with the CINA Platform (cleared under K200855) or other medical image communications devices. CINA-CSpine receives cervical spine CT scans identified by the CINA Platform or other medical image communications device, processes them using deep learning algorithms involving execution of multiple computational steps to identify the suspected positive findings compatible with acute cervical spine fractures and generates results files to be transferred by CINA Platform or a similar medical image communications device for output to a PACS system or workstation for worklist prioritization.

To identify the suspected presence of cervical fractures, the device uses a deep learning model trained end-to-end on 1,338 cases acquired from US and France, representing a distribution of fracture presentations, locations and acquisition protocols, including multiple scanner models from Siemens, Philips, GE and Canon/Toshiba. Additional deep learning models are used to locate the individual vertebrae to exclude images that do not conform to the expected field of view.

DICOM images are received, recorded and filtered before processing. The series are processed chronologically by running algorithms on each series to detect suspected positive findings of a cervical spine fracture, then active notifications on the flagged series are sent to the worklist application.

The Worklist Application displays the active notification of new studies with suspected findings when they come in. All the cervical spine CT studies which include at least 5 visible cervical vertebrae received by CINA-CSpine device are displayed in the worklist and those on which the algorithms have detected a suspected finding are marked with an icon (i.e., active notification). In addition, a compressed, grayscale, unannotated image that is marked "not for diagnostic use" is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for diagnostic use beyond notification.

Presenting the trained physician specialist with notification facilitates earlier triage by allowing image prioritization in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

The CINA platform is an example of medical image communications platform for integrating and deploying the CINA-CSpine image processing application. The medical image communications device (i.e., the technical platform) provides the necessary requirements for interoperability based on the standardized DICOM protocol and services to communicate with existing systems in the hospital radiology department such as CT modalities or other DICOM nodes (DICOM router or PACS for example). It is responsible for transferring, converting formats, notifying of suspected findings and displaying medical device data such as radiological data. The medical image communications server includes the Worklist client application, which receives notifications from the CINA-CSpine Image Processing application.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Performance Goal)Reported Device Performance (mean [95% CI])Predicate Device Performance (mean [95% CI])
Sensitivity≥ 80%90.3% [84.5% - 94.5%]91.7% [82.7% - 96.9%]
Specificity≥ 80%91.9% [86.8% - 95.5%]88.6% [81.2% - 93.8%]
Time-to-Notification (All Cases)Not specified (Comparable to predicate)2.9 minutes [2.7 - 3.0]Not specified
Time-to-Notification (True Positive Cases)Not specified (Comparable to predicate)2.8 minutes [2.6 - 3.0]3.9 minutes [3.8 - 4.1]

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: 328 clinical anonymized cases.
  • Data Provenance: Retrospective, multicenter, multinational. Data was acquired from:
    • US: 60.4% of cases, including a U.S. teleradiology company with a database from various U.S. hospitals across different territories.
    • OUS: 39.6% of cases.
    • Scanner Manufacturers: GE (31.1%), Philips (21.6%), Siemens (28.7%), Canon (18.3%), and 36 different scanner models.
    • Time Periods: The validation dataset was from independent sites and different time periods compared to the training data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Three.
  • Qualifications of Experts: US-board-certified expert radiologists.

4. Adjudication Method for the Test Set

The ground truth was established by the consensus of the three US-board-certified expert radiologists. This implies a 3-expert consensus (e.g., all 3 agree, or majority vote if specific rules were defined for disagreement, which is not further detailed).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not reported. The study focused on the standalone performance of the AI device and compared its performance metrics (Sensitivity, Specificity, Time-to-Notification) to those reported for the predicate device. There is no mention of human readers improving with AI assistance.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop)

Yes, a standalone performance testing was performed. The described study evaluated the software's performance (Sensitivity and Specificity) in detecting cervical spine fractures on non-contrast CT scans without human intervention in the initial detection process.

7. Type of Ground Truth Used

Expert Consensus: The ground truth was established by the consensus of three US-board-certified expert radiologists.

8. Sample Size for the Training Set

The deep learning model was trained end-to-end on 1,338 cases.

9. How the Ground Truth for the Training Set Was Established

The document states that the training data was acquired from US and France, representing a distribution of fracture presentations, locations, and acquisition protocols. However, it does not explicitly detail how the ground truth was established for this training set (e.g., if it was also expert consensus, based on pathology reports, or other methods). It can be inferred that it would likely follow a similar rigorous annotation process to establish "true" fracture presence, but specific details are not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

Avicenna.AI % John J. Smith, M.D., J.D. Partner Hogan Lovells, US LLP Columbia Square 555 Thirteenth Street, NW Washington, District of Columbia 20004

Re: K240942

September 12, 2024

Trade/Device Name: CINA-CSpine Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological Computer Aided Triage And Notification Software Regulatory Class: Class II Product Code: QAS Dated: April 5, 2024 Received: August 16, 2024

Dear John J. Smith, M.D., J.D.:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240942

Device Name CINA-CSpine

Indications for Use (Describe)

CINA-CSpine is a radiological computer aided triage and notification software indicated for use in the analysis of cervical spine CT images.

The device is intended to assist hospital networks and appropriately trained physician specialists by flagging and communication of suspected positive findings compatible with acute cervical spine fractures including non-displaced fracture lines and/or displaced fracture fragments.

CINA-CSpine uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device is not designed to detect vertebral compression fractures.

The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.

The results of CINA-CSpine are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images.

Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY

AVICENNA.AI's CINA-CSPINE

Submitter

Applicant:

AVICENNA.AI Espace Mistral B, 375 avenue du Mistral, 13600 LA CIOTAT France

Contact Person:

Stephane Berger Regulatory Manager Phone: +33 6 12 12 28 13 E-mail: stephane.berger@avicenna.ai

Date prepared: September 12, 2024

Device Identification

Name of Device:CINA-CSpine
Classification Name:Radiological computer aided triage andnotification software
Regulation No:21 CFR § 892.2080
Product Code:QAS
Regulatory Class:Class II
Classification Panel:Radiology

Predicate Device

The CINA-CSpine device is substantially equivalent to the following predicate device with regard to indications for use, performance, and technological characteristics:

510(k):K190896
Trade Name:BriefCase for CSF Triage
Manufacturer:Aidoc Medical, Ltd.
Classification Name:Radiological computer aided triage and notification software
Regulation No:21 CFR § 892.2080
Product Code:QAS
Regulatory Class:Class II

{5}------------------------------------------------

Device Description

CINA-CSpine is a radiological computer-assisted triage and notification software device.

CINA-CSpine runs on a standard "off the shelf" server/workstation and consists of CSpine Image Processing Application, which can be integrated, deployed and used with the CINA Platform (cleared under K200855) or other medical image communications devices. CINA-CSpine receives cervical spine CT scans identified by the CINA Platform or other medical image communications device, processes them using deep learning algorithms involving execution of multiple computational steps to identify the suspected positive findings compatible with acute cervical spine fractures and generates results files to be transferred by CINA Platform or a similar medical image communications device for output to a PACS system or workstation for worklist prioritization.

To identify the suspected presence of cervical fractures, the device uses a deep learning model trained end-to-end on 1,338 cases acquired from US and France, representing a distribution of fracture presentations, locations and acquisition protocols, including multiple scanner models from Siemens, Philips, GE and Canon/Toshiba. Additional deep learning models are used to locate the individual vertebrae to exclude images that do not conform to the expected field of view.

DICOM images are received, recorded and filtered before processing. The series are processed chronologically by running algorithms on each series to detect suspected positive findings of a cervical spine fracture, then active notifications on the flagged series are sent to the worklist application.

The Worklist Application displays the active notification of new studies with suspected findings when they come in. All the cervical spine CT studies which include at least 5 visible cervical vertebrae received by CINA-CSpine device are displayed in the worklist and those on which the algorithms have detected a suspected finding are marked with an icon (i.e., active notification). In addition, a compressed, grayscale, unannotated image that is marked "not for diagnostic use" is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for diagnostic use beyond notification.

Presenting the trained physician specialist with notification facilitates earlier triage by allowing image prioritization in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

The CINA platform is an example of medical image communications platform for integrating and deploying the CINA-CSpine image processing application. The medical image communications device (i.e., the technical platform) provides the necessary requirements for interoperability based on the standardized DICOM protocol and services to communicate with existing systems in the hospital radiology department such as CT modalities or other DICOM nodes (DICOM router or PACS for example). It is responsible for transferring, converting formats, notifying of suspected findings and displaying medical device data such as radiological data. The medical image communications server includes the Worklist client application, which receives notifications from the CINA-CSpine Image Processing application.

{6}------------------------------------------------

Intended Use / Indications for Use

CINA-CSpine is a radiological computer aided triage and notification software indicated for use in the analysis of cervical spine CT images.

The device is intended to assist hospital networks and appropriately trained physician specialists by flagging and communication of suspected positive findings compatible with acute cervical spine fractures including non-displaced fracture lines and/or displaced fracture fragments.

CINA-CSpine uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device is not designed to detect vertebral compression fractures.

The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.

The results of CINA-CSpine are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

Summary of Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Software Verification and Validation Testing

CINA-CSpine complies with DICOM (Digital Imaging and Communications in Medicine) -Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20.

Avicenna.Al conducted extensive performance validation testing and software verification and validation testing of the CINA-CSpine device as standalone software. CINA-CSpine is tested against its user needs and intended use by the successful execution of planned software verification and validation testing included in this submission.

Software performance, validation and verification testing demonstrated that the CINA-CSPINE met all design requirements and specifications associated with the intended use of the software.

Standalone Performance Testing

Avicenna.Al conducted a retrospective, multicenter, multinational and blinded study with the CINA-CSpine application with the primary endpoint to evaluate the software's performance (Sensitivity and Specificity) in detecting cervical spine (CSpine) fractures on non-contrast CT (NCCT) scans images, in 328 clinical anonymized cases. The dataset included 59.5% Male and 36.3% Female. Mean ± SD age of patients included in the study was 55.3 ± 5.1 yo (range: 18 - 101 yo).

{7}------------------------------------------------

Multiple subgroups of interest were considered in the analysis. In fact, the validation data was acquired from 3 different clinical sources, with one representing an U.S teleradiology company which has access to a database of medical images from various U.S. hospitals distributed throughout the U.S. territories, to account for race/ethnicity in the intended U.S. patient population. Additional scanner parameters considered were slice thickness, number of detector rows, and kVp ranges. Detailed subgroup analysis was reported in the labeling.

The data was provided from multiple US (60.4%) and OUS (39.6%) clinical sites. The independence of the standalone validation dataset from the training data was ensured using data from independent sites and different time periods. There were 155 (47.3%) positive cases (CT with CSpine) and 173 (52.7%) negative cases included in the analysis. The data was acquired primarily by 4 different scanner makers (GE-31.1%, Philips-21.6%, Siemens-28.7%, and Canon-18.3%) and 36 different scanner models.

Device Sensitivity [95% Cl] and Specificity [95% Cl] were computed against the ground truth established by consensus of three US-board-certified expert radiologists.

As a primary endpoint, the global Sensitivity and Specificity were found to be 90.3% [95%C1: 84.5% - 94.5%] and 91.9% [95%C1: 86.8% - 95.5%], respectively. These findings achieved the 80% performance goal of the study and are similar to the results reported for the predicate BriefCase device (Aidoc Medical) which demonstrated: Sensitivity and Specificity of 91.7% [95%C1: 82.7% -96.9%] and 88.6% [95%CI: 81.2% - 93.8%], respectively.

Additionally, the CINA-CSpine prioritization and triage effectiveness were evaluated by the standalone per-case processing time of the device (time-to-notification), which corresponds to the time between the end of the DICOM reception (made available for CINA-CSpine image processing) and the end of processing (positive or negative identification). Finally, for better comparison with the predicate device, the statistics for only True Positive cases (i.e., identified as positive both by the ground truthers as well as the device) are also provided.

The results are presented in Table below:

Time-to-Notification(minutes)MEAN ± SDMEDIAN95% ClMINMAX
CINA-CSpineAll cases(N = 328)2.9 ± 1.12.6[2.7 - 3.0]1.06.8
CINA-CSpineTrue Positive cases(N = 140 )2.8 ± 1.22.6[2.6 - 3.0]1.06.8

Time-to-Notification for CINA-CSpine Image Processing Application

The mean [95% C1] time-to-notification for all included cases (n = 328) was estimated to be 2.9 [95% CI: 2.7 - 3.0] minutes for CINA-CSpine.

When taking into account only true positive cases (n = 140), the mean [95% Cl] time-to-notification was 2.8 [95%C1: 2.6 - 3.0] minutes for CINA-CSpine and 3.9 [95%C1: 3.8 - 4.1] minutes for BriefCase, the defined predicate device.

{8}------------------------------------------------

The performance testing of the CINA-CSpine device establishes that the subject device is as safe and effective as the predicate device and compatible with the same clinical use, since the performances demonstrated the clinical effectiveness of the subject device and its ability to provide effective prioritization and triage, which is substantially equivalent to that of the predicate device. This establishes that the CINA-CSpine device achieves its intended use and is substantially equivalent to the predicate device.

Substantial Equivalence

The subject CINA-CSpine and the predicate BriefCase for CSF Triage device are both intended to aid in prioritization and triage of radiological images of time sensitive findings for patient detection and diagnosis (i.e., cervical spine fracture) based on the analysis of medical images acquired from radiological signal acquisition systems. The CINA device provides the CINA Platform in which the subject CINA-CSpine prioritization and triage application can be integrated, deployed and used. The labeling of the subject and the predicate devices clearly states that the devices are not for diagnostic use. The subject CINA-CSpine device and the predicate BriefCase for CSF Triage are software packages with similar technological characteristics and principles of operation, and incorporate deep learning AI algorithms that process images, and software to send notifications and to display unannotated preview images. In both devices, the labeling instructs the user to further evaluate and diagnose based only on the original images in the local PACS.

The subject CINA-CSpine, the predicate BriefCase for CSF Triage device operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, and do not remove images from the standard of care FIFO queue; thus, not disturbing standard interpretation of the ittending trained physician specialists. The subject and predicate devices achieve performance of the time-to-notification metric in similar ranges of time, and thus contribute similarly to effective triage and early involvement of the trained physician specialists in evaluating suspected images of cervical spine fracture.

The standalone performance and effectiveness assessment studies demonstrated that the CINA-CSpine device performs as intended and substantially equivalent to the BriefCase for CSF Triage predicate device.

The table below compares the key features of the subject and the predicate devices.

Subject device: CINA-CSpineSoftwarePredicate device: BriefCase for CSFTriage Software (K190896)
Intended Use/ Indicationsfor UseCINA-CSpine is a radiologicalcomputer aided triage andnotification software indicated foruse in the analysis of cervicalspine CT images.BriefCase is a radiological computer aidedtriage and notification software indicatedfor use in the analysis of cervical spine CTimages.The device is intended to assist hospitalnetworks and trained radiologists inworkflow triage by flagging and
Subject device: CINA-CSpineSoftwarePredicate device: BriefCase for CSFTriage Software (K190896)
The device is intended to assisthospital networks andappropriately trained physicianspecialists by flagging andcommunication of suspectedpositive findings compatible withacute cervical spine fracturesincluding non-displaced fracturelines and/or displaced fracturefragments.communication of suspected positivefindings of linear lucencies in the cervicalspine bone in patterns compatible withfractures.
CINA-CSpine uses an artificialintelligence algorithm to analyzeimages and highlight cases withdetected findings on a standaloneapplication in parallel to theongoing standard of care imageinterpretation. The device is notdesigned to detect vertebralcompression fractures.BriefCase uses an artificial intelligencealgorithm to analyse images and highlightcases with detected findings on astandalone desktop application in parallelto the ongoing standard of care imageinterpretation. The user is presented withnotifications for cases with suspectedfindings.
The user is presented withnotifications for cases withsuspected findings. Notificationsinclude compressed previewimages that are meant forinformational purposes only, andare not intended for diagnosticuse beyond notification.The device does not alter theoriginal medical image, and it isnot intended to be used as adiagnostic device.The results of CINA-CSpine areintended to be used inconjunction with other patientinformation and based onprofessional judgment to assistwith triage/prioritization ofmedical images.Notified clinicians are ultimatelyresponsible for reviewing fullimages per the standard of care.Notifications include compressed previewimages that are meant for informationalpurposes only and not intended fordiagnostic use beyond notification.The device does not alter the originalmedical image and is not intended to beused as a diagnostic device.The results of BriefCase are intended tobe used in conjunction with other patientinformation an based on their professionaljudgment, to assist withtriage/prioritization of medical images.Notified clinicians are responsible forviewing full images per the standard ofcare.
UserpopulationTrained physician specialistRadiologist
Anatomicalregion ofinterestCervical SpineCervical Spine
Subject device: CINA-CSpineSoftwarePredicate device: BriefCase for CSFTriage Software (K190896)
DataacquisitionprotocolNon-contrast cervical spine CT scansNon-contrast cervical spine CT scans
View DICOMdataDICOM information about the patient, study and current imageDICOM information about the patient, study and current image
Segmentationof region ofinterestNo; device does not mark, highlight, or direct users' attention to a specific location in the original imageNo; device does not mark, highlight, or direct users' attention to a specific location in the original image
AlgorithmArtificial intelligence algorithm with database of imagesArtificial intelligence algorithm with database of images
Notification /PrioritizationYesYes
PreviewimagesPresentation of a compressed, grayscale, unannotated image that is marked "not for diagnostic use".Presentation of a small, compressed, black and white preview image that is labeled "Not for diagnostic use".
Alteration oforiginal imageNoNo
Removal ofcases fromworklistqueueNo. The device operates in parallel with the standard of care, which remains the default option for all cases.No. The device operates in parallel with the standard of care, which remains the default option for all cases.
Structure-CSpine image processing application- Compatibility of use with the CINA Platform (worklist and Image Viewer) or other medical image communications device.- AHS module (orchestrator, image acquisition)- ACS module (image processing).- Aidoc Worklist application for workflow integration (worklist and non-diagnostic basic Image Viewer).

Comparison of key features between CINA-CSpine and predicate device (Aidoc Medical Ltd)

{9}------------------------------------------------

{10}------------------------------------------------

Conclusion

CINA-CSpine and BriefCase have similar intended use and substantially similar indications, technological characteristics, and principles of operation. The following is a summary of the substantial equivalence comparison:

  • । The predicate device is legally marketed.
    CINA-CSpine has the same intended use as the predicate device, i.e., AIDoc Medical, -Ltd.'s BriefCase, and therefore it may be found substantially equivalent.

  • । CINA-CSpine and the predicate device have very similar indications for use.

{11}------------------------------------------------

CINA-CSpine has similar technological characteristics as the predicate and reference i devices, i.e., the deployment of an artificial intelligence algorithm with a database of images.

-Standalone performance study demonstrates that the CINA-CSpine and the predicate BriefCase raise the same types of safety and effectiveness questions, namely, accurate detection of findings within the processed study.

Accordingly, the subject CINA-CSpine device is substantially equivalent to the predicate BriefCase device

§ 892.2080 Radiological computer aided triage and notification software.

(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.