Search Results
Found 3 results
510(k) Data Aggregation
(87 days)
CINA-VCF is a radiological computer aided triage and notification software indicated for use in patients aged 50 years and over undergoing non-enhanced or contrast-enhanced CT scans which include the chest and/or abdomen.
The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCF) findings.
CINA-VCF uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.
The results of CINA-VCF are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
CINA-VCF is a radiological computer-assisted triage and notification software device.
CINA-VCF runs on a standard "off the shelf" server/workstation and consists of VCF Image Processing Application, which can be integrated, deployed and used with the CINA Platform (cleared under K200855) or other compatible medical image communications devices. CINA-VCF receives nonenhanced or contrast-enhanced CT scans (which include the chest and/or abdomen) identified by the CINA Platform or other compatible medical image communications device, processes them using algorithmic methods involving execution of multiple computational steps to identify suspected presence of Vertebral Compression Fractures (VCF) findings and generates results files to be transferred by CINA Platform or a similar medical image communications device for output to a PACS system or workstation for worklist prioritization.
DICOM images are received, recorded and filtered before processing. The series are processed chronologically by running algorithms on each series to detect suspected positive findings of a vertebral compressions fracture (VCF).
The device uses deep learning models to detect VCF at the T1-L5 level. The models were trained endto-end on a dataset of 886 series collected from multiple centers in the USA and France satisfying the device protocol and representing a large distribution of scanner models from Siemens, Philips, GE and Canon (formerly Toshiba), acquisition protocols, spine presentation and fracture location and severity. Additional models, trained on subsets of this dataset, are used to locate the spine, identify the vertebra bodies and exclude vertebra which have been subjected to vertebroplasty or contains orthopedic material.
The Worklist Application displays all incoming suspect cases, each notified case is marked with an icon. In addition, compressed, grayscale, unannotated images that are captioned "not for diagnostic use" is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for diagnostic use beyond notification.
Presenting the specialist with worklist prioritization facilitates earlier triage by allowing prioritization of images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
The CINA Platform is an example of medical image communications platform for integrating and deploying the CINA-VCF image processing applications. It provides the necessary requirements for interoperability based on the standardized DICOM protocol and services to communicate with existing systems in the hospital radiology department such as CT modalities or other DICOM nodes (DICOM router or PACS for example). It is responsible for transferring, converting formats, notifying of suspected findings and displaying medical device data such as radiological data. The CINA Platform server includes the Worklist client application which receives notifications from the CINA-VCF Image Processing application.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criterion | Reported Device Performance (CINA-VCF) |
---|---|
Primary Endpoint: ROC AUC | 0.974 [95% CI: 0.962 - 0.986] (Exceeded the 0.95 performance goal) |
Sensitivity | 95.2% [95% CI: 90.7% - 97.9%] |
Specificity | 92.9% [95% CI: 89.4% - 96.5%] |
Accuracy (Overall Agreement) | 93.7% [95% CI: 91.1% - 95.7%] |
Time-to-Notification (All cases, Mean ± SD) | 23.4 ± 8.4 seconds (Median: 21.0 seconds, 95% CI: [22.7 - 24.2], Min: 9.0, Max: 60.0) |
Time-to-Notification (True Positive cases, Mean ± SD) | 21.7 ± 7.5 seconds (Median: 20.0 seconds, 95% CI: [20.5 - 22.8], Min: 9.0, Max: 45.0) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: 474 clinical anonymized cases.
- Data Provenance: Retrospective, multinational study. Data provided from multiple US (66.9%) and OUS (33.1%) clinical sites. The data included 180 (37.9%) positive cases (CT with VCF) and 294 (62.1%) negative cases.
- Patient Demographics: Mean age 72.1 ± 10.1 years old (MIN = 50 yo and MAX = 100 yo), 50.8% female. Data accounted for race/ethnicity in the intended US patient population.
- Image Acquisition: Acquired by 4 different scanner makers and 38 different scanner models. Various scanner parameters were considered, including slice thickness, number of detector rows, kVp ranges, contrast vs. non-contrast, imaging protocol (chest and/or abdomen), and reconstruction kernel (soft/standard).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Three.
- Qualifications: US-board-certified expert radiologists.
4. Adjudication Method for the Test Set
- Method: Consensus of three US-board-certified expert radiologists. A case was considered positive if at least one moderate or severe vertebral compression fracture located within the thoracic or lumbar spine was identified by the experts.
5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study Was Done
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted to evaluate human readers with and without AI assistance for effect size. The study focused on the standalone performance of the AI device and compared its time-to-notification to a predicate device, not directly to human reader performance with or without the AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, a standalone performance testing was performed. The study describes "Avicenna.Al conducted a retrospective, multinational and blinded study with the CINA-VCF application... to evaluate the software's performance."
7. The Type of Ground Truth Used
- Type of Ground Truth: Expert consensus. Specifically, "ground truth established by consensus of three US-board-certified expert radiologists."
8. The Sample Size for the Training Set
- Sample Size: 886 series.
9. How the Ground Truth for the Training Set Was Established
- The device uses deep learning models that "were trained end-to-end on a dataset of 886 series collected from multiple centers in the USA and France satisfying the device protocol and representing a large distribution of scanner models... and fracture location and severity." While the text describes the dataset, it does not explicitly state how the ground truth for this training set was established. It implies that the "device protocol" guided the selection of data, likely with some form of expert labeling or pre-existing clinical reports classifying the fractures, similar to the test set ground truth, but this is not directly stated.
Ask a specific question about this device
(89 days)
CINA-iPE is a radiological computer-aided triage and notification software indicated for use in patients undergoing contrast-enhanced CT scans (not dedicated CTPA protocol) for other clinical indications than pulmonary embolism suspicion, including at least a part of the lung. The device is intended to assist hospital networks and appropriately trained radiologists in workflow triage by flagging and communicating suspected positive findings for incidental Pulmonary Embolism (iPE). The device is indicated for adults and transitional adolescents (18 to 21 years old but treated as adults).
CINA-iPE uses an artificial intelligence algorithm to analyze images and highlight cases with detected incidental PE on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected incidental PE findings. The device is not designed to detect PE in subsegmental arteries.
Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.
The results of CINA-iPE are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
CINA-iPE is a radiological computer aided triage and notification software device.
CINA-iPE runs on a standard "off the shelf" server/workstation and consists of an Image Processing Application, which can be integrated, deployed, and used with the CINA Platform (cleared under K200855) or other medical image communications devices. CINA-iPE receives contrast-enhanced CT scans (not dedicated CTPA protocol) including at least a part of the lung identified by the CINA Platform or other medical image communications device, processes them using deep learning algorithms involving the execution of multiple computational steps to identify the suspected presence of an incidental pulmonary embolism and generates results files to be transferred by CINA Platform or a similar medical image communications device for output to a PACS system or worklist prioritization.
To identify the suspected presence of pulmonary embolisms, the device uses a deep learning model trained end-to-end on 5.429 cases acquired from US and France, representing a distribution of PE sizes, locations and acquisition protocols, including multiple scanner models from Siemens, Philips, GE and Canon/Toshiba. Additional models are used to locate the aorta and main pulmonary artery, enabling assessment of the contrast timing. The lung's parenchyma is segmented to evaluate both the presence of the lungs in the field of view and to limit the region of interest for detecting the presence of pulmonary embolisms.
DICOM images are received, recorded and filtered before processing. The series are processed chronologically by running algorithms on each series to detect suspected positive findings of an incidental Pulmonary Embolism (iPE), then active notifications on the flagged series are sent to the Worklist Application.
The Worklist Application displays the active notification of new studies with suspected findings when they come in. All the contrast-enhanced CT studies received by CINA-iPE device are displayed in the worklist and those on which the algorithms have detected finding are marked with an icon (i.e., passive notification). In addition, a compressed, grayscale, unannotated image that is marked "not for diagnostic use" is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for diagnostic use beyond notification.
Presenting the radiologist with notification facilitates earlier triage by allowing prioritization of images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care image interpretation practice alone.
The CINA platform is an example of medical image communications platform for integrating and deploying the CINA-iPE image processing application. The medical image communications device (i.e., the technical platform) provides the necessary requirements for interoperability based on the standardized DICOM protocol and services to communicate with existing systems in the hospital radiology department such as CT modalities or other DICOM nodes (DICOM router or PACS for example). It is responsible for transferring, storing, converting formats, notifying of suspected findings and displaying medical device data such as radiological data. The medical image communications server includes the Worklist client application in which notifications from the CINA-iPE Image Processing application are received.
The provided text describes the acceptance criteria and the study conducted to prove that the CINA-iPE device meets these criteria.
Here's an organized breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The primary acceptance criteria for the CINA-iPE device were its Sensitivity and Specificity in identifying incidental Pulmonary Embolism (iPE), measured against a performance goal of 80%.
Metric | Acceptance Criteria (Performance Goal) | Reported Device Performance [95% CI] |
---|---|---|
Sensitivity | ≥ 80% | 87.8% [82.2% - 92.2%] |
Specificity | ≥ 80% | 92.0% [87.3% - 95.4%] |
Additional Performance Data (Sub-group Analysis):
Arterial Segment | Sensitivity [95% CI] |
---|---|
Main (N = 55) | 96.3% [87.5% - 99.6%] |
Interlobar (N = 73) | 94.5% [86.6% - 98.5%] |
Lobar (N = 127) | 92.9% [87.0% - 96.7%] |
Segmental (N = 179) | 88.3% [82.6% - 92.6%] |
Time-to-Notification:
Metric | MEAN ± SD | MEDIAN | 95% CI | MIN | MAX |
---|---|---|---|---|---|
CINA-iPE All cases | |||||
(N = 381) | 1.5 ± 0.5 | 1.4 | [1.4 - 1.5] | 0.3 | 2.7 |
CINA-iPE True Positive cases | |||||
(N = 159) | 1.5 ± 0.4 | 1.5 | [1.4 - 1.6] | 0.7 | 3.1 |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 381 clinical anonymized cases.
- Data Provenance: Retrospective, multinational study.
- Country of Origin: Data was acquired from multiple U.S. and OUS (Outside US) clinical sites. Specifically, 56.4% (215) of cases came from U.S. clinical sources.
- Retrospective/Prospective: Retrospective.
- Independence: The independence of the standalone validation dataset from the training data was ensured using data from independent sites and different time periods.
- Patient Demographics: 53.5% Male and 46.7% Female. Mean ± SD age: 64.5 ± 15.8 years (range: 18 - 99 years).
- Scanner Diversity: Acquired primarily by 4 different scanner makers (GE-31.5%, Philips-28.3%, Siemens-26%, and Canon-13.9%) and 39 different scanner models.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Three (3)
- Qualifications of Experts: US-board-certified expert radiologists.
4. Adjudication Method for the Test Set
The ground truth was established by consensus of the three US-board-certified expert radiologists. While the specific mechanism of reaching consensus (e.g., 2 majority, discussion, etc.) is not detailed, the term "consensus" implies agreement among the experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? The provided text does not explicitly state that a multi-reader multi-case (MRMC) comparative effectiveness study was done to evaluate how human readers improve with AI vs. without AI assistance. The study focuses on the standalone performance of the algorithm and its time-to-notification effectiveness for triage/prioritization.
- Effect Size of Human Improvement: Not applicable, as an MRMC comparative effectiveness study was not described.
6. Standalone (Algorithm Only) Performance Study
- Was a standalone study done? Yes, a standalone performance testing study was conducted.
- Details: The study evaluated the CINA-iPE application's performance in identifying incidental pulmonary embolisms (iPE) on contrast-enhanced CT images. The primary endpoint was the device's Sensitivity and Specificity.
7. Type of Ground Truth Used
- Ground Truth Type: Expert Consensus. The ground truth was established by the consensus of three US-board-certified expert radiologists.
8. Sample Size for the Training Set
- Sample Size for Training Set: 5,429 cases.
9. How the Ground Truth for the Training Set Was Established
- Method: The deep learning model was "trained end-to-end on 5.429 cases acquired from US and France, representing a distribution of PE sizes, locations and acquisition protocols." The precise method for establishing ground truth for training is not explicitly detailed but it's implied that these cases were labeled with PE sizes and locations, likely through expert review similar to the test set, but this is not explicitly stated.
Ask a specific question about this device
(111 days)
CINA CHEST is a radiological computer aided triage and notification software indicated for use in the analysis of Chest and Thoraco-abdominal CT angiography. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communicating suspected positive findings of (1) Chest CT angiography for Pulmonary Embolism (PE) and (2) Chest or Thoraco-abdominal CT angiography for Aortic Dissection (AD).
CINA CHEST uses an artificial intelligence algorithm to analyze images and highlight cases with detected PE and AD on a standalone Web application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected PE or AD findings. Notifications include compressed preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.
The results of CINA CHEST are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
CINA CHEST is a radiological computer-assisted triage and notification software device.
The software system is based on algorithm-programmed components and is comprised of a standard off-the-shelf operating system and additional image processing applications.
DICOM images are received, recorded and filtered before processing. The series are processed chronologically by running algorithms on each series to detect suspected positive findings of a pulmonary embolism (PE) or an aortic dissection (AD), then notifications on the flagged series are sent to the Worklist Application.
The Worklist Application (on premise) displays the pop-up notifications of new studies with suspected findings when they come in, and provides both active and passive notifications. Active notifications are in the form of a small pop-up containing patient name, accession number and the type of suspected findings (PE or AD). All the chest and thoraco-abdominal CT angiography studies received by CINA CHEST device are displayed in the worklist and those on which the algorithms have detected a suspected finding (PE or AD) are marked with an icon (i.e., passive notification). In addition, a compressed, small black and white image that is marked "not for diagnostic use" is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification. Presenting the radiologist with notification facilitates earlier triage by allowing one to prioritize images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
Here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for CINA CHEST:
Acceptance Criteria and Reported Device Performance
Parameter | Acceptance Criteria (Performance Goal) | Reported Device Performance (CINA CHEST) | Comparison to Predicate (BriefCase) |
---|---|---|---|
Pulmonary Embolism (PE) Detection | |||
Sensitivity | ≥ 80% | 91.1% [95% CI: 86.1% - 94.7%] | Predicate: 90.6% [95% CI: 82.2% - 95.9%] |
Specificity | ≥ 80% | 91.8% [95% CI: 87.1% - 95.1%] | Predicate: 89.9% [95% CI: 82.2% - 95.1%] |
Accuracy | Not explicitly stated as a minimum goal, but reported. | 91.4% | Not explicitly stated for predicate. |
Time-to-Notification (PE) | Not explicitly stated as a minimum/maximum goal, but comparable to predicate. | 63 ± 16.1 seconds (Mean) | |
60.8 seconds (Median) | |||
[95% CI: 61.5 – 64.6] seconds | Predicate: 3.9 [95% CI: 3.7 - 4.1] minutes (234 seconds) | ||
Aortic Dissection (AD) Detection | |||
Sensitivity | ≥ 80% | 96.4% [95% CI: 91.7% - 98.8%] | Not applicable (Predicate is for PE/ICH, not AD) |
Specificity | ≥ 80% | 97.5% [95% CI: 93.8% - 99.3%] | Not applicable |
Accuracy | Not explicitly stated as a minimum goal, but reported. | 97% | Not applicable |
Time-to-Notification (AD) | Not explicitly stated as a minimum/maximum goal, but comparable to reference. | 36.5 ± 9.1 seconds (Mean) | |
34.1 seconds (Median) | |||
[95% CI: 35.4 – 37.5] seconds | Reference (CINA, ICH/LVO): 21.6 ± 4.4 sec (ICH), 34.7 ± 10.7 sec (LVO) |
Study Details
-
Sample sizes used for the test set and the data provenance:
- Pulmonary Embolism (PE): 396 clinical anonymized cases.
- Aortic Dissection (AD): 298 clinical anonymized cases.
- Data Provenance: Retrospective, multicenter study. Data was provided from multiple US clinical sites (230 US cities for PE, and 200 US cities for AD).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: "Several US-board-certified radiologist readers." The exact number is not specified beyond "several".
- Qualifications: US-board-certified radiologists. No specific years of experience are mentioned.
-
Adjudication method for the test set:
- The ground truth was established by "concurrence of several US-board-certified radiologist readers." This implies a consensus-based adjudication, but the specific method (e.g., majority vote, unanimous agreement, or an independent adjudicator in case of disagreement) is not explicitly detailed.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not reported. The study described is a standalone performance evaluation of the CINA CHEST software against a ground truth. It assesses the device's ability to identify PE and AD cases for triage, not the improvement of human readers with AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone study was done. The document explicitly states: "Avicenna.Al conducted a retrospective, multicenter and blinded study with the CINA CHEST software with the primary endpoint to evaluate the software's performance..." and later refers to "The results of the standalone assessment study demonstrated an overall agreement (Accuracy)..." This confirms the study evaluated the algorithm's performance in isolation.
-
The type of ground truth used:
- Expert Consensus. The ground truth was "established by concurrence of several US-board-certified radiologist readers."
-
The sample size for the training set:
- The document does not specify the sample size for the training set. It only details the test set used for performance evaluation.
-
How the ground truth for the training set was established:
- Since the training set sample size is not provided, the method for establishing its ground truth is also not detailed in this document. It is common for AI algorithms to be trained on data with ground truth established by expert radiologists or pathology, but this specific information is absent here.
Ask a specific question about this device
Page 1 of 1