K Number
K240932

Validate with FDA (Live)

Date Cleared
2024-08-05

(122 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indications for Use - VIO 3

The Erbe electrosurgical unit (ESU/Generator) model VIO 3 with instruments and accessories is intended to deliver high frequency (HF) electrical current for the cutting and/or coagulation of tissue, including ablation.

Indications for use - ERBECRYO 2

The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion (destruction) of tissue by the application of extreme cold, and cooling of tissue during radiofrequency ablation.

Indications for use - HybridTherm ablation probes
The HybridTherm probes are intended for ablation and coagulation of soft tissue. The following probes can also be used for cutting tissue: HybridTherm monopolar, eTip and HybridTherm bipolar, eTip.

Device Description

The HybridTherm system consists of the electrosurgical unit (ESU) Model VIO 3, the cryosurgical unit Model ERBECRYO 2, the HybridTherm ablation probes and accessories.

HybridTherm ablation probes
The HybridTherm probes are a product family consisting of 8 probes to offer the user a range of instruments based upon the physician's preference. There are 4 monopolar and 4 bipolar ablation probes, which differ in the function of the distal end: with and without a monopolar cutting tip (eTip), as well as in their shape (i.e. straight or angulated/curved).

The HybridTherm probes combine high-frequency (HF) technology with cryotechnology. The HF technology is used for cutting (only variants with eTip), coagulation, and ablation of tissue. To reduce tissue desiccation and to increase the ablation zone, the probes are cooled. Cooling also minimizes tissue adhesion to the electrode. The primary function of the cooling is to support the ablation process.

VIO 3
The ESU Model VIO 3 delivers high frequency (HF) electrical current for cutting, coagulation and ablation of tissue. The unit can be mounted/secured to a cart/system carrier or on a ceiling mount. Different Footswitches are available for activating the ESU and as applicable to change modes of the ESU (Remode feature). The ESU Model VIO 3 has several clearly defined cutting, coagulation and ablation modes with different electrical waveforms and electrical parameters that are programmed with defined effect levels. Each effect level corresponds to a defined maximum power output and a voltage limitation. The ESU has a touchscreen monitor that provides the user with an onscreen tutorial as well as settings and operational information. It also provides a small number of physical controls, such as the power switch, instrument sockets and a neutral electrode receptacle.

ERBECRYO 2
The cryosurgical unit Model ERBECRYO 2 is connected to an electrical power source and a carbon dioxide (CO2) source. Upon activation, the Unit delivers a regulated CO2flow to the end of the connected instrument for cooling. The CO2 returning from the probe to the ERBECRYO 2 exists via the exhaust gas hose. Appropriate gas cylinder adapters are available for the various international gas cylinder connections.

HybridTherm System
The HybridTherm probes are connected to the units (VIO 3 and ERBECRYO 2) via respective plugs. The units are connected to each other via an Erbe Communication Bus (ECB) cable, whereby the user only interacts with the ESU Model VIO 3 (master-slave technology: the cryosurgical unit is controlled via the VIO 3). The ERBECRYO 2 additionally requires a CO2 gas cylinder. A footswitch is connected to the ESU, with which the activation of the available modes is started. The ESU supplies the HF current which is needed for cutting (only HybridTherm probe variants with eTip), coagulation and ablation of tissue. The ERBECRYO 2 supplies CO2 gas which is needed for cooling via the Joule-Thomson-effect to support the ablation.

AI/ML Overview

The provided text is a 510(k) summary for the HybridTherm System. It focuses on demonstrating substantial equivalence to predicate devices, primarily through comparison of technical characteristics and non-clinical performance testing. It does not include information about AI/ML models, human reader studies, or the establishment of ground truth for such studies.

Therefore, many of the requested details about acceptance criteria, human performance, and ground truth are not present in this document. The information provided heavily emphasizes hardware and traditional medical device testing rather than algorithmic performance.

Here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

The document mentions "Side-by-side tissue testing against the respective predicate device(s) shows that the performance (i.e. cutting, coagulation and ablation) is as good as the performance of the predicate device(s)." This indicates a comparative performance acceptance criterion, but specific numerical thresholds or a formal table are not provided in this summary. The acceptance criterion is qualitative: performance is "as good as" the predicate.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. This was "side-by-side tissue testing," likely in a lab setting, not a clinical study with patients or image-based data. No information on sample size (e.g., number of tissue samples), country of origin, or retrospective/prospective nature is given.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. Ground truth for the "tissue testing" was likely established through objective measurements of ablation zone size or tissue characteristics, not expert consensus (as there are no images or clinical interpretations involved).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. There's no indication of adjudication for "tissue testing."

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. The document explicitly states "Side-by-side tissue testing" and discusses engineering safety and performance standards (e.g., electrical safety, EMC, biocompatibility, sterilization). There is no mention of a human reader study or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This device is a physical electrosurgical system, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the "side-by-side tissue testing," the ground truth would be physical measurements of the effect on tissue (e.g., ablation zone size, tissue integrity) rather than expert consensus or pathology from human subjects. The document does not specify the exact methods but implies objective performance metrics.

8. The sample size for the training set

Not applicable. There is no AI/ML model for which a training set would be defined.

9. How the ground truth for the training set was established

Not applicable, as there is no AI/ML model or training set.


Summary of available and missing information based on the provided text:

Available Information:

  • Acceptance Criteria (Qualitative): Device performance (cutting, coagulation, ablation) is "as good as" the predicate device(s) based on side-by-side tissue testing.
  • Study Type: Non-clinical (bench) testing, including:
    • Side-by-side tissue testing
    • Functional testing and design controls (21 CFR 820.30)
    • Electromagnetic compatibility (IEC 60601-1-2)
    • Electrical safety (IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18)
    • Software verification and cybersecurity (IEC 62304, FDA Guidance) - Note: This relates to the software controlling the electrosurgical unit, not an AI/ML component.
    • Biocompatibility (ISO 10993-1)
    • Sterilization validation (ISO 11135)
    • Packaging and shelf-life validation (ISO 11607-1, ISO 11607-2, ASTM F 1980)
  • Ground Truth Type: Physical/objective measurements of tissue effects for performance testing.

Missing Information (Not contained in the provided 510(k) summary, as it pertains to AI/ML or human reader studies):

  • Specific numerical acceptance criteria for tissue testing.
  • Sample size for tissue testing (e.g., number of tissue samples).
  • Data provenance (country of origin, retrospective/prospective).
  • Number and qualifications of experts for ground truth establishment.
  • Adjudication method.
  • Multi-reader multi-case (MRMC) comparative effectiveness study results or effect sizes.
  • Standalone (algorithm-only) performance.
  • Sample size or ground truth establishment for a training set (as no AI/ML component is discussed).

This document is a standard 510(k) for a physical electrosurgical device, not a submission for an AI/ML-driven device. Therefore, the detailed questions about AI model validation and human performance studies are not addressed within this text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 5, 2024

Erbe Elektromedizin GmbH Matthias Kollek Regulatory Affairs Specialist Waldhoernlestrasse 17 Tuebingen, 72072 Germany

Re: K240932

Trade/Device Name: HybridTherm System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI, JOS, GEH Dated: July 22, 2024 Received: July 22, 2024

Dear Matthias Kollek:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long H. Chen -> Digitally signed by Long H. Chen-S

Date: 2024.08.05 13:39:20 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240932

Device Name HybridTherm System

Indications for Use (Describe)

Indications for Use - VIO 3

The Erbe electrosurgical unit (ESU/Generator) model VIO 3 with instruments and accessories is intended to deliver high frequency (HF) electrical current for the cutting and/or coagulation of tissue, including ablation.

Indications for use - ERBECRYO 2

The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion (destruction) of tissue by the application of extreme cold, and cooling of tissue during radiofrequency ablation.

Indications for use - HybridTherm ablation probes The HybridTherm probes are intended for ablation and coagulation of soft tissue. The following probes can also be used for cutting tissue: HybridTherm monopolar, eTip and HybridTherm bipolar, eTip.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------


Traditional 510(k) for HybridTherm System

510(k) SUMMARY

ApplicantErbe Elektromedizin GmbHWaldhoernlestrasse 1772072 TuebingenGermanyTel: 0049-7071-755-0Fax: 0049-7071-755-179
Contact PersonDr. Matthias Kollek
Regulatory Affairs Specialist
E-Mail: Matthias.Kollek@erbe-med.com
Date PreparedJuly 22, 2024
Device Information
Trade/Proprietary Name:HybridTherm System
Common Name:Electrosurgical cutting and coagulation device
and accessories
Classification NameElectrosurgical cutting and coagulation device
and accessories
Regulation Number:21 CFR 878.4400,
21 CFR 878.4350
Class:II
Product Code:GEI & JOS
GEH
Legally Marketed Predicate DevicesPrimary
EUSRA RF Electrode - K181249
Additional predicate devices
VIVA combo RF System - K163450
Erbe ESU Model VIO 3 - K190823
Single Use Electrosurgical pencil - K192542
ERBECRYO 2 - K190651

{5}------------------------------------------------

Device Description

The HybridTherm system consists of the electrosurgical unit (ESU) Model VIO 3, the cryosurgical unit Model ERBECRYO 2, the HybridTherm ablation probes and accessories.

HybridTherm ablation probes

The HybridTherm probes are a product family consisting of 8 probes to offer the user a range of instruments based upon the physician's preference. There are 4 monopolar and 4 bipolar ablation probes, which differ in the function of the distal end: with and without a monopolar cutting tip (eTip), as well as in their shape (i.e. straight or angulated/curved).

The HybridTherm probes combine high-frequency (HF) technology with cryotechnology. The HF technology is used for cutting (only variants with eTip), coagulation, and ablation of tissue. To reduce tissue desiccation and to increase the ablation zone, the probes are cooled. Cooling also minimizes tissue adhesion to the electrode. The primary function of the cooling is to support the ablation process.

VIO 3

The ESU Model VIO 3 delivers high frequency (HF) electrical current for cutting, coagulation and ablation of tissue. The unit can be mounted/secured to a cart/system carrier or on a ceiling mount. Different Footswitches are available for activating the ESU and as applicable to change modes of the ESU (Remode feature). The ESU Model VIO 3 has several clearly defined cutting, coagulation and ablation modes with different electrical waveforms and electrical parameters that are programmed with defined effect levels. Each effect level corresponds to a defined maximum power output and a voltage limitation. The ESU has a touchscreen monitor that provides the user with an onscreen tutorial as well as settings and operational information. It also provides a small number of physical controls, such as the power switch, instrument sockets and a neutral electrode receptacle.

ERBECRYO 2

The cryosurgical unit Model ERBECRYO 2 is connected to an electrical power source and a carbon dioxide (CO2) source. Upon activation, the Unit delivers a regulated CO2flow to the end of the connected instrument for cooling. The CO2 returning from the probe to the ERBECRYO 2 exists via the exhaust gas hose. Appropriate gas cylinder adapters are available for the various international gas cylinder connections.

HybridTherm System

The HybridTherm probes are connected to the units (VIO 3 and ERBECRYO 2) via respective plugs. The units are connected to each other via an Erbe Communication Bus (ECB) cable, whereby the user only interacts with the ESU Model VIO 3 (master-slave technology: the cryosurgical unit is controlled via the VIO 3). The ERBECRYO 2 additionally requires a CO2 gas cylinder. A footswitch is connected to the ESU, with which the activation of the available modes is started. The ESU supplies the HF current which is needed for cutting (only HybridTherm probe variants with eTip), coagulation and ablation of tissue. The ERBECRYO 2 supplies CO2 gas which is needed for cooling via the Joule-Thomson-effect to support the ablation.

{6}------------------------------------------------

Indications for Use

HybridTherm probes

The HybridTherm probes are intended for ablation and coagulation of soft tissue. The following probes can also be used for cutting tissue: HybridTherm monopolar, eTip and HybridTherm bipolar, eTip.

VIO 3

The Erbe electrosurgical unit (ESU/Generator) Model VIO 3 with instruments and accessories is intended to deliver high frequency (HF) electrical current for the cutting and/or coagulation of tissue, including ablation.

ERBECRYO 2

The ERBECRYO 2 cryosurgical unit and accessories are intended for cryoadhesion, devitalization (destruction) of tissue by the application of extreme cold, and cooling of tissue during radiofrequency ablation.

Compared to the predicate device(s), the VIO 3 and the HybridTherm probes have the same intended use. All devices are intended for the application of high frequency (HF) electrical energy in interventional/surgical procedures to thermally effect or destroy tissue.

Compared to the previously cleared cryosurgical unit (ERBECRYO 2), the ERBECRYO 2 that is part of the HybridTherm System has an extended Indications for Use statement (i.e. cooling of tissue during radiofrequency ablation). However, the intended use of both devices is the supply of CO2 to a connected instrument for cooling via the Joule-Thomson effect. This function is the same for the predicate system. The VIVA Combo RF Generator in combination with the EUSRA RF Electrode uses a peristaltic pump with chilled water to cool the probe/tissue during the application. Therefore, the intended use is essentially the same.

{7}------------------------------------------------

Comparison to predicate devices

VIO 3

CharacteristicsSubject DeviceHybridTherm System (VIO 3)Primary PredicateErbe ESU Model VIO 3 withAccessoriesSecondary PredicateVIVA Combo RF SystemDiscussion
ManufacturerErbe Elektromedizin GmbH(Germany)Erbe Elektromedizin GmbH(Germany)STARmed Co., Ltd.N/A
510(k)K240932K190823K163450N/A
Trade/Proprietary NameErbe ESU Model VIO® 3Erbe ESU Model VIO® 3 withAccessoriesVIVA combo RF SystemN/A
Classification NameElectrosurgical, Cutting &Coagulation & AccessoriesElectrosurgical, Cutting &Coagulation & AccessoriesElectrosurgical, Cutting &Coagulation & AccessoriesSame
Classification Regulation21 CFR 878.440021 CFR 878.440021 CFR 878.4400Same
Product CodeGEIGEIGEISame
Device ClassIIIIIISame
Indications for UseThe Erbe electrosurgical unit(ESU/Generator) model VIO 3with instruments andaccessories is intended todeliver high frequency (HF)electrical current for the cuttingand/or coagulation of tissue,including ablation.The Erbe Electrosurgical Unit(ESU/Generator) Model VIO® 3with Accessories is intended todeliver High Frequency (HF)electrical current for the cuttingand/or coagulation of tissue.The VIVA combo RF AblationSystem is intended for use inpercutaneous andintraoperative coagulation andablation of tissue.Same
Prescription or OTCPrescriptionPrescriptionPrescriptionSame
DimensionsWidth x Height x Depth415 x 215 x 375 mm16.3" x 8.5" x 14.8"Width x Height x Depth415 x 215 x 375 mm16.3" x 8.5" x 14.8"Width x Height x Depth260 x 115 x 348 mm10.2" x 4.5" x 13.7"Same as PrimaryPredicate
CharacteristicsSubject DeviceHybridTherm System (VIO 3)Primary PredicateErbe ESU Model VIO 3 withAccessoriesSecondary PredicateVIVA Combo RF SystemDiscussion
Weight12 kg/26.7 lbs12 kg/26.7 lbs6 kg / 13.2 lbsSame as PrimaryPredicate
Mains voltage 100 - 120 VAC(±10%) / 220 - 240 VAC (±10%)Mains voltage 100 - 120 VAC(±10%) / 220 - 240 VAC (±10%)Mains voltage 100 - 240 VSame as PrimaryPredicate
Mains current max. 6.3AMains current max. 6.3ANot knownSame as PrimaryPredicate
Technical/Performance DataMains frequency 50 Hz / 60 HzMains frequency 50 Hz / 60 HzMains frequency 50 Hz / 60 HzSame
Frequency 350 kHzFrequency 350 kHzFrequency 480 kHzSame as PrimaryPredicate
Maximum Power Output 400wattsMaximum Power Output 400wattsMaximum Power Output 200wattsSame as PrimaryPredicate
WiFi connectionAvailable(deactivated by default,no HF-activation possible if WiFiis enabled)Available(deactivated by default,no HF-activation possible if WiFiis enabled)Not availableSame as PrimaryPredicate
Output Modules/Receptacles4 HF instrument sockets1 Neutral electrode socket2 Footswitch sockets4 HF instrument sockets1 Neutral electrode socket2 Footswitch sockets1 HF Cable socket2 Neutral electrode sockets2 Footswitch socketsSame as PrimaryPredicate
Software VersionV1.1.xV2.0.x1.10Different SoftwareVersion
AccessoriesOne- and two- pedal FootswitchesOne- and two- pedal FootswitchesOne- and two- pedal FootswitchesSame
Connection cablesConnection cablesConnection cablesSame
Neutral ElectrodesNeutral ElectrodesNeutral ElectrodesSame
VIO-Cart, System Carriercompact and System Carriercompact plusVIO-CartN/ANew systemCarriers
CharacteristicsSubject DeviceHybridTherm System (VIO 3)Primary PredicateErbe ESU Model VIO 3 withAccessoriesSecondary PredicateVIVA Combo RF SystemDiscussion
N/AN/APeristaltic PumpSecondary predicatedevice uses aperistaltic pump forcooling of theconnected ablationelectrode.
Condition provided / UseconditionNon-sterile / reusableNon-sterile / reusableNon-sterile / reusableSame

{8}------------------------------------------------

{9}------------------------------------------------

HybridTherm Probes

CharacteristicsSubject DeviceHybridTherm probesPrimary PredicateEUSRA RF ElectrodeSecondary PredicateSingle Use ElectrosurgicalpencilDiscussion
510(k) NumberK240932K181249K192542N/A
Trade/Proprietary NameHybridTherm ProbesEUSRA RF ElectrodeSingle Use Electrosurgicalpencil with non-coated andnon-stick electrode (Non-sterileand sterile)N/A
ManufacturerErbe Elektromedizin GmbHSTARmed Co., Ltd.Modern Medical EquipmentManufacturing, LTD.N/A
Classification NameElectrosurgical Cutting andCoagulation Device andAccessoriesElectrosurgical Cutting andCoagulation Device andAccessoriesElectrosurgical Cutting andCoagulation Device andAccessoriesSame
Classification PanelGeneral and Plastic SurgeryGeneral and Plastic SurgeryGeneral and Plastic SurgerySame
Classification Regulation21 CFR 878.440021 CFR 878.440021 CFR 878.4400Same
Product CodeGEI, JOSGEI, JOSGEISame
CharacteristicsSubject DeviceHybridTherm probesPrimary PredicateEUSRA RF ElectrodeSecondary PredicateSingle Use ElectrosurgicalpencilDiscussion
Device ClassClass IIClass IIClass IISame
Indications for UseThe HybridTherm probes areintended for ablation andcoagulation of soft tissue. Thefollowing probes can also beused for cutting tissue:HybridTherm monopolar, eTipand HybridTherm bipolar, eTip.The EUSRA RF Electrode isindicated for coagulation andablation of soft tissue when usedin conjunction with compatibleradio frequency generator.The devices are used to cutand coagulate soft tissue bymeans of high frequencyelectrical current during anelectrosurgical procedure.Same
Energy UsedHF electrical currentHF electrical currentHF electrical currentSame
Operating PrincipleMonopolar and BipolarMonopolarMonopolarSimilar
Cooling TechnologyCO2 (Joule-Thompson Effect)Cooled WaterN/ASimilar
PhysicalSpecificationsTip LengthMonopolar variants: 9.05mm(active electrode)Bipolar variants: 2x 9,05mm +1,2mm Isolator = 19,3mm5, 7, 10, 15, 20mm69mm, 102mm, 152mm (bladeelectrode)Within dimensions ofprimary predicate
TipDiameters1.9mm19 Gauge (1.06mm)2.36mmWithin dimensions ofprimary andsecondary predicate
FlexibleShaft1250mm1393mmN/ASimilar to primarypredicate
Patient Contacting MaterialsStainless Steel, Gold, PlasticsStainless Steel, PlasticsPlastics, Stainless steel, Teflonone coatSimilar
Condition provided / UseconditionSterile, single useSterile, single useSterile, single useSame
Sterilization methodEO SterilizationEO SterilizationEO SterilizationSame
Extensible Tip LengthNot extensibleExtensible Tip LengthNot extensibleSame as secondarypredicate

{10}------------------------------------------------

{11}------------------------------------------------

ERBECRYO 2

CharacteristicsSubject DeviceHybridTherm System (ERBECRYO 2)Predicate DeviceERBECRYO 2 Cryosurgical Unit andAccessoriesDiscussion
510(k) NumberK240932K190651Different
Model NameERBECRYO® 2 Cryosurgical Unitand AccessoriesERBECRYO® 2 Cryosurgical Unitand AccessoriesSame
ManufacturerErbe Elektromedizin GmbHErbe Elektromedizin GmbHSame
Common NameCryosurgical Unit and AccessoriesCryosurgical Unit and AccessoriesSame
Classification Regulation21 CFR 878.435021 CFR 878.4350Same
Product CodeGEHGEHSame
Device ClassIIIISame
Indications for UseThe ERBECRYO 2 cryosurgical unit andaccessories are intended for cryoadhesion,devitalization (destruction) of tissue by theapplication of extreme cold, and cooling oftissue during radiofrequency ablation.The ERBECRYO® 2 cryosurgical unit andaccessories are intended for cryoadhesionand devitalization (destruction) of tissue bythe application of extreme cold.Equivalent, extended Indicationsfor Use fall within the intendeduse
Intended UseSupply of CO2 to a connected instrumentfor cooling via the Joule-Thompson effect.Supply of CO2 to a connected instrumentfor cooling via the Joule-Thompson effect.Same
Prescription or OTCPrescriptionPrescriptionSame
MaterialsMetal sheet, aluminum, copper pipes,plastics and wiring, electronic componentsMetal sheet, aluminum, copper pipes,plastics and wiring, electronic componentsSame
Type, Dimensional, and Physical Attributes as well as Performance and Other Specifications
Dimensionsof Unit16.1"x5.1"x14.6" 410x130x370mm)16.1"x5.1"x14.6" 410x130x370mm)Same
Weight of Unit14.8lbs (6.7kg)14.8lbs (6.7kg)Same
CharacteristicsSubject DeviceHybridTherm System (ERBECRYO 2)Predicate DeviceERBECRYO 2 Cryosurgical Unit andAccessoriesDiscussion
Technical Data of UnitRated supply voltage:100 - 240V (±10%)Rated supply voltage:100 - 240V (±10%)Same
Rated supply frequency: 50 - 60HzRated supply frequency: 50 - 60HzSame
Line current: 0.4 - 0.8ALine current: 0.4 - 0.8ASame
Software Version2.0.01.0.3Different
Remote control by VIO 3YesNoDifferent
Performance Data of UnitNo. of effect levels: 1 - 5 (depending oninstrument)No. of effect levels: 1 - 5 (depending oninstrument)Same
Activation: via footswitchActivation: via footswitchSame
Cooling gas used: CO2Cooling gas used: CO2Same
Input Pressure:653 - 943psi (45 - 65bar)Input Pressure:653 - 943psi (45 - 65bar)Same
Other CompatibleEquipmentERBECRYO 2 Cart, VIO Cart, cable, gashoses, gas bottle adapterERBECRYO 2 Cart, VIO Cart, cable, gashoses, gas bottle adapterSame
Provided ConditionNon-sterile, reusableNon-sterile, reusableSame

{12}------------------------------------------------

{13}------------------------------------------------

Comparison of Technological Characteristics

Differences of the ESU Model VIO 3 compared to the previously cleared device is an updated software mainly to introduce new modes for cutting, coagulation and ablation of tissue, remote control of the cryosurgical unit ERBECRYO 2 and operation/control of the HybridTherm probes.

Compared to the primary predicate device EUSRA RF Electrode the HybridTherm probes have an additional bipolar variant, a monopolar cutting tip (only variants with eTip) and a lateral and distal coagulation feature. The HybridTherm probes are cooled via CO2 compared to cooling via chilled water which is employed by the predicate probes. In addition, the HybridTherm probes have different materials and different physical specifications which are, however, within the specifications of the respective predicate devices.

Difference of the ERBECRYO 2 compared to the predicate device is an updated software for remote control via the VIO 3 and operation/control of the HybridTherm probes.

The subject device system and the predicate device(s) use the same enerqy (HF electrical current) and principles of operation. Both the monopolar and bipolar technology are established and routinely adopted in interventional procedures for many years. Compared to the monopolar variant, the bipolar variant does not raise different issues of safety and effectiveness. The cooling technology of the subject device by means of CQ2 is the same compared to the previously cleared ERBECRYO 2. Although the predicate system uses a different cooling technology (chilled water), the purpose of the same and the cooling is only used to support the ablation process.

Side-by-side tissue testing against the respective predicate device(s) shows that the performance (i.e. cutting, coagulation and ablation) is as good as the performance of the predicate device(s). The non-clinical testing described below shows that the different technological characteristics do not negatively affect the safety and effectiveness of the subject device.

Non-clinical performance testing

Verification/validation activities from non-clinical testing as described below demonstrate that the differences do not raise any new or different issues of safety or effectiveness of the subject device compared to the predicate devices.

Functional testing and design controls to verify both safety and performance of the subject device was performed in compliance with 21 CFR 820.30 to ensure that the subject device performs as intended and meets design specifications.

Side-by-side tissue testing was performed in compliance with FDA Guidance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", to validate the cutting, coagulation and ablation performance compared to the predicate device(s).

Electromagnetic compatibility (EMC) was tested in compliance with IEC 60601-1-2 and FDA Guidance "Electromagnetic Compatibility (EMC) of Medical Devices".

{14}------------------------------------------------

Electrical safety of the subject device was tested in compliance with, IEC 60601-1; IEC 60601-2-2 and IEC 60601-2-18, as applicable.

Software verification and documentation for cybersecurity was provided in compliance with IEC 62304 and FDA Guidance Documents "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions".

Biocompatibility testing was performed in compliance with ISO 10993-1 and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" to demonstrate biocompatibility of the materials with tissue contact.

Sterilization validation was performed in compliance with ISO 11135 and documentation was provided according to FDA Guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile" showing an SAL of 10-6. EO residual testing and limits comply with ISO 10993-7.

Packaging and shelf-life validation was performed in compliance with ISO 11607-1, ISO 11607-2 and accelerated aged devices (ASTM F 1980).

Conclusion

The subject device has the same intended use, the same fundamental design, substantially equivalent performance characteristics, and the same energy source as the predicate devices. The subject device was tested as described above and the minor differences in technological characteristics were assessed with regards to safety and effectiveness. Taken together, the subject device does not raise new or different questions of safety and effectiveness, and the subject device is substantially equivalent to the predicate devices for the requested indications for use.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.