Search Results
Found 1 results
510(k) Data Aggregation
Electrodes Handpieces (BT, JET) are indicated for coagulation and ablation of soft tissue when used in conjunction with Mygen V-1000 radio frequency generator.
Electrodes Handpieces (BT, JET) are sterile, single-use electrosurgical electordes intended to be use in conjunction with Mygen V-1000 RF generator (K180999).
Electrodes Handpieces (BT, JET) consist of electrode tip, insulation part, handle.
Cooling of the electrode is provided by saline which is pumped (using the peristaltic pump) through the inflow tubing, the electrode and out through the outflow tubing. This is an enclosed system within the electrode and saline does not contact the patient. For the electrodes (BT, JET), dispersive electrodes are provided so that electrical current flows through the tip of the electrodes, through the target tissue and to the dispersive electrodes.
This document (K193355/S002) is not for an AI/ML medical device, but rather for "Electrode Handpieces (BT, JET)" which are electrosurgical electrodes used for coagulation and ablation of soft tissue. Therefore, the questions related to AI/ML device acceptance criteria, study details, and human-in-the-loop performance are not applicable.
However, I can extract the acceptance criteria and study details that were performed for this specific medical device from the provided text.
Here's a breakdown of the information found, specifically addressing the device's performance data and acceptance criteria:
1. Table of Acceptance Criteria and Reported Device Performance
Test/Criteria | Acceptance Criteria | Reported Device Performance |
---|---|---|
Biocompatibility (ISO10993-1;2009) | Patient contact components and materials must meet ISO10993-1;2009 standards. | Tested and validated according to ISO10993-1;2009. Patient contact components and materials are identical to the predicate device. |
Shelf-Life Testing | Packaging seal strength, packaging dye penetration, appearance, conduction, and sterility must meet acceptance criteria (ASTM F 1980 Accelerated Aging). | All testing results passed the acceptance criteria for packaging seal strength, packaging dye penetration, appearance, conduction, and sterility, in accordance with ASTM F 1980 Accelerated Aging method for a valid shelf life. |
Medical Electrical Equipment Safety (IEC 60601-1) | General Requirements for basic safety and essential performance must be fulfilled. | The requirements of specified standards (IEC 60601-1) were fulfilled. |
Electromagnetic Compatibility (IEC 60601-1-2) | General Requirements for basic safety and essential performance (collateral standards: electromagnetic compatibility) must be fulfilled. | The requirements of specified standards (IEC 60601-1-2) were fulfilled. |
High Frequency Surgical Equipment Safety (IEC 60601-2-2:2009) | Particular requirements for the basic safety and essential performance must be fulfilled. | The requirements of specified standards (IEC 60601-2-2:2009) were fulfilled. |
Temperature Sensor Accuracy (of V-1000 generator) | Measured values by the temperature sensor of V-1000 and the temperature calibrator must be within ±3°. | The test results confirmed that the values measured by the temperature sensor of V-1000 and the temperature calibrator were within the acceptance criteria (±3°). Tested in multiple power modes, three times for each power setting. |
Bacterial Endotoxin Concentration (LAL Test) | Bacterial endotoxin concentration should be |
Ask a specific question about this device
Page 1 of 1