K Number
K240871

Validate with FDA (Live)

Date Cleared
2024-11-18

(234 days)

Product Code
Regulation Number
870.1330
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Synxess Neurovascular Guidewire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guidewire is intended for use only in the neuro vasculature.

Device Description

The Synxess Neurovascular Guidewire is available in diameters of 0.014 inch (0.35 mm) and 0.010 inch (0.25 mm) and has 200 cm length. The device is comprised of a core wire, coils, and coatings, and is supplied with a shaping mandrel and a torque device. The guidewire distal tip is straight and shapeable. The device is provided sterile and is for single use.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, the Synxess Neurovascular Guidewire, and details the non-clinical testing performed to establish substantial equivalence to predicate devices. It does not describe an AI/ML-driven device or a study involving human readers and AI assistance. Therefore, many of the requested points in the prompt (e.g., AI performance metrics, MRMC study, expert ground truth establishment for AI, training set details) are not applicable to this document.

However, I can provide information based on the provided text regarding the device's acceptance criteria and the study that proves it meets them, focusing on the non-clinical performance testing conducted for a guidewire.

Here's the breakdown based on the provided document:

Acceptance Criteria and Device Performance for Synxess Neurovascular Guidewire

The acceptance criteria for this non-AI device are based on passing various performance bench tests and biocompatibility tests in accordance with relevant international standards and FDA guidance. The study proving the device meets these criteria is the battery of non-clinical tests performed.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for each test are implicitly "Pass" or "Met acceptance criteria" as defined by the referenced standards and guidance documents. The reported device performance for all listed tests was successful.

TestStandards, Guidance, Test MethodReported Device Performance
Visual InspectionFDA Guidewire Guidance and ISO 11070Pass
Dimensional VerificationFDA Guidewire Guidance and ISO 11070Pass
Simulated UseFDA Guidewire GuidancePass
Tensile Strength and Tip PullFDA Guidewire Guidance and ISO 11070Pass
Torque StrengthFDA Guidewire GuidancePass
TorqueabilityFDA Guidewire GuidancePass
Coating IntegrityFDA Guidewire GuidancePass
Particulate EvaluationFDA Guidewire GuidancePass
LubricityFDA Guidewire GuidancePass
Corrosion ResistanceFDA Guidewire Guidance and ISO 11070Pass
Kink ResistanceFDA Guidewire GuidancePass
Tip FlexibilityFDA Guidewire GuidancePass
RadiopacityFDA Guidewire Guidance and ISO 11070Pass
Fracture TestISO 11070Pass
Flexing TestISO 11070Pass
Shaping Mandrel Visual and Dimensional InspectionThe shaping mandrel was examined for signs of impurities or defects. The mandrel length and outer diameter were measured.Met acceptance criteria
Shaping Mandrel Corrosion ResistanceISO 11070Pass
Shaping Mandrel Tensile StrengthThe tensile strength between the shaping mandrel rod and handle was measured.Met acceptance criteria
Torque Device Visual Inspection and Simulated UseThe torque device was examined for signs of impurities or defects. Simulated use evaluated the torque device compatibility and performance.Met acceptance criteria
Biocompatibility Tests
Cytotoxicity Study Using MTT MethodISO 10993-5Pass
Sensitization Study Guinea Pig Maximization TestISO 10993-10Pass
Intracutaneous Reactivity Test In RabbitsISO 10993-23Pass
Acute Systemic Toxicity Study In MiceISO 10993-11Pass
Pyrogen Test In RabbitsISO 10993-11 / USP<151>Pass
Hemolysis Assay-Direct Contact And Extract MethodsISO 10993-4 / ASTM F756-17Pass
Complement Activation SC5b-9 AssayISO 10993-4Pass
Thromboresistance Evaluation In DogsISO 10993-4Pass
Sterilization Validation
EO Sterilization Process ValidationISO 11135:2014Validated
Sterilant ResidualsISO 10993-7:2008Met acceptance criteria
Shelf-Life Validation
Shelf-Life (3 years)Testing of accelerated aged devicesValidated

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the exact sample sizes (e.g., number of guidewires) used for each individual performance bench test or biocompatibility test. It only states that the device "passed all performance bench testing." For biocompatibility, some tests specify animal models (e.g., guinea pig, rabbits, mice, dogs), but the specific number of animals is not provided.
  • Data Provenance: The document generally indicates that the tests were performed in accordance with "internal requirements, referenced guidance and international standards." The country of origin for the data is implied to be from the manufacturer, Enlight Medical Technologies (Shenzhen) Co., Ltd., located in China. The data would be considered prospective as it was generated specifically for this 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • This question is not applicable as this is a physical medical device (guidewire), not an AI/ML-driven diagnostic device requiring expert clinical interpretation for ground truth. The "ground truth" for these tests is defined by the physical and chemical properties measured against established engineering and biological standards.

4. Adjudication Method for the Test Set

  • This question is not applicable for the reasons stated above. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where human interpretation of medical images or patient data requires consensus, which is not relevant here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No. An MRMC study was not performed because this is a physical medical device, not an AI/ML-driven system that assists human readers in diagnostic tasks. The document explicitly states: "No clinical testing was deemed necessary to support the substantial equivalence of the subject device."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an algorithm-only device.

7. The Type of Ground Truth Used

  • The "ground truth" for this device's performance is established by engineering standards (e.g., measurements of tensile strength, torqueability, flexibility, dimensions) and biological/chemical standards for biocompatibility (e.g., cytotoxicity assays, pyrogen tests, hemolysis assays). These are objective, quantifiable measures against predefined acceptance criteria from international standards (e.g., ISO 11070, ISO 10993 series) and FDA guidance.

8. The Sample Size for the Training Set

  • Not applicable. This is a physical device, not an AI/ML model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. This is a physical device, not an AI/ML model.

{0}------------------------------------------------

November 18, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract design, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line. The logo is clean and professional, reflecting the organization's role in regulating food and drugs.

Enlight Medical Technologies (Shenzhen) Co., Ltd. Kason Gui RA Manager 5-6/F, Building A, Xinghui Technology Industrial Park 52 Huaning Road, Xinshi Community, Dalang Street, Longhua District Shenzhen, 518109 China

Re: K240871

Trade/Device Name: Synxess Neurovascular Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: MOF Dated: March 30, 2024 Received: October 15, 2024

Dear Kason Gui:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The "510(k) Premarket Notification" database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note that the Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the quality system (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the QS regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the medical device report (MDR) regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-devicesafety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Xiaolin Zheng -S

For Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240871

Device Name Synxess Neurovascular Guidewire

Indications for Use (Describe)

Synxess Neurovascular Guidewire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guidewire is intended for use only in the neuro vasculature.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Synxess Neurovascular Guidewire K240871

Submission Sponsor:Enlight Medical Technologies (Shenzhen) Co., Ltd.5-6/F, Building A, Xinghui Technology Industrial Park, 52Huaning Road, Xinshi Community, Dalang Street, LonghuaDistrict, Shenzhen, 518109, P.R.ChinaTel: +86-755-2100-4622
Applicant Contact:Eason Zhaomin_zhao@enlight-medical.com

Correspondent Contact: Kason Gui jinpeng gui@enlight-medical.com

Date Prepared: November 14, 2024

Subject Device
Trade Name:Synxess Neurovascular Guidewire
Common Name:Catheter Guide Wire
Classification Name:Guide, Wire, Catheter, Neurovasculature
Regulation Number:21 CFR 870.1330
Product Code:MOF
Classification:Class II

Predicate Devices:

Predicate: ASAHI CHIKAI Neurovascular Guide Wire (K110584, product code MOF) Predicate: ASAHI CHIKAI 10 Neurovascular Guide Wire (K112979, product code MOF)

Indications for Use Statement

Synxess Neurovascular Guidewire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guidewire is intended for use only in the neuro vasculature.

Device Description

The Synxess Neurovascular Guidewire is available in diameters of 0.014 inch (0.35 mm)

{5}------------------------------------------------

and 0.010 inch (0.25 mm) and has 200 cm length. The device is comprised of a core wire, coils, and coatings, and is supplied with a shaping mandrel and a torque device. The guidewire distal tip is straight and shapeable. The device is provided sterile and is for single use.

Comparison of Technological Characteristics

The subject device, Synxess Neurovascular Guidewire, is substantially equivalent to the predicate devices in terms of:

  • indications for use; .
  • . materials;
  • technological characteristics; .
  • packaging and sterilization of devices. .

The differences in technological characteristics do not raise new questions of safety and effectiveness compared to the predicate devices as outlined in the comparison table below.

ItemPredicate DeviceASAHI CHIKAIPredicate DeviceASAHI CHIKAI 10Subject DeviceSynxess NeurovascularGuidewire
510(k) NumberK110584K112979K240871
Regulation Number21 CFR 870.1330Same
Regulation ClassIISame
Product CodeMOFSame
Indications forUse/Intended UseASAHI Neurovascular Guide Wire is intendedto be used in the neuro vasculature to facilitatethe placement and exchange of therapeuticdevices such as cerebral catheters duringintravascular therapy. This guide wire isintended for use only in the neuro vasculature.ASAHI Neurovascular Guide Wire is intendedto be used in the neuro vasculature to facilitatethe placement and exchange of therapeuticdevices such as cerebral catheters duringintravascular therapy. This guide wire isintended for use only in the neuro vasculature.Synxess NeurovascularGuidewire is intended to beused in the neurovasculature to facilitate theplacement and exchange oftherapeutic devices such ascerebral catheters duringintravascular therapy. Thisguidewire is intended foruse only in the neurovasculature.
Anatomical SiteNeuro vasculatureNeuro vasculatureSame
FunctionThe guidewire is used to facilitate the placementand exchange of therapeutic devices.The guidewire is used to facilitate the placementand exchange of therapeutic devices.Same
Proximal Diameter0.014 inch0.010 inchGW1410: 0.014 inchGW1010: 0.010 inch

{6}------------------------------------------------

ItemPredicate DeviceASAHI CHIKAIPredicate DeviceASAHI CHIKAI 10Subject DeviceSynxess NeurovascularGuidewire
Distal Diameter0.014 inch0.010 inch0.010 inch
Guidewire Length200 cm, 300 cm200 cm
Core WireConfigurationStainless Steel core wireSame
Coil Length (TipLength)30 cm9.5 cm30 cm
Coil ConfigurationPlatinum/Nickel and Stainless Steel coilsPlatinum/Tungsten andStainless Steel coils
Tip Type and ShapeStraight, shapeableSame
Tip FlexibilityStandardSame
Radiopaque CoilLocation5 cm from tip3 cm from tip3 cm from tip
Core Wire MaterialStainless SteelSame
Coils MaterialPlatinum/Nickel and Stainless SteelPlatinum/Tungsten andStainless Steel
Coating MaterialDistal: HydrophilicProximal: PTFE (only 300 cm guide wire)Same
Bonding MethodSolderingSame
AccessoryInserter, Shaping Device, and Torque DeviceShaping Tool, TorqueDevice
PackagingConfigurationGuidewire is inserted in a tube, placed inpeelable pack, and then in a packaging box.Same
Sterilization MethodEthylene OxideSame
Method of SupplySterile/Single UseSame
Shelf Life3 yearsSame

Testing Summary

The Synxess Neurovascular Guidewire passed all performance bench testing in accordance with internal requirements, referenced guidance and international standards as shown in the table below to support substantial equivalence of the device.

TestStandards, Guidance, Test MethodResults
Visual InspectionFDA Guidance for Industry and Food and DrugAdministration Staff: Coronary, Peripheral, andNeurovascular Guidewires - Performance Tests andRecommended Labeling, October 2019(thereinafter, “FDA Guidewire Guidance”) and ISOPass

{7}------------------------------------------------

11070
Dimensional VerificationFDA Guidewire Guidance and ISO 11070Pass
Simulated UseFDA Guidewire GuidancePass
Tensile Strength and Tip PullFDA Guidewire Guidance and ISO 11070Pass
Torque StrengthFDA Guidewire GuidancePass
TorqueabilityFDA Guidewire GuidancePass
Coating IntegrityFDA Guidewire GuidancePass
Particulate EvaluationFDA Guidewire GuidancePass
LubricityFDA Guidewire GuidancePass
Corrosion ResistanceFDA Guidewire Guidance and ISO 11070Pass
Kink ResistanceFDA Guidewire GuidancePass
Tip FlexibilityFDA Guidewire GuidancePass
RadiopacityFDA Guidewire Guidance and ISO 11070Pass
Fracture TestISO 11070Pass
Flexing TestISO 11070Pass
Shaping Mandrel Visual andDimensional InspectionThe shaping mandrel was examined for signs of impurities or defects. The mandrel length and outer diameter were measured.Met acceptance criteria
Shaping Mandrel CorrosionResistanceISO 11070Pass
Shaping Mandrel TensileStrengthThe tensile strength between the shaping mandrel rod and handle was measured.Met acceptance criteria
Torque Device VisualInspection and Simulated UseThe torque device was examined for signs of impurities or defects. Simulated use evaluated the torque device compatibility and performance.Met acceptance criteria

Biocompatibility Testing

The Synxess Neurovascular Guidewire was assessed according to the International Standard ISO 10993-1, "Biological evaluation of medical devices. Part 1. Evaluation and testing within a risk management process." The subject device is classified as an external communicating device with circulating blood contact for a limited duration (<24 hours). The following tests were performed:

TestStandardResults
Cytotoxicity Study Using MTT MethodISO 10993-5Pass
Sensitization Study Guinea Pig Maximization TestISO 10993-10Pass
Intracutaneous Reactivity Test In RabbitsISO 10993-23Pass
Acute Systemic Toxicity Study In MiceISO 10993-11Pass

{8}------------------------------------------------

TestStandardResults
Pyrogen Test In RabbitsISO 10993-11/USP<151>Pass
Hemolysis Assay-Direct Contact And ExtractMethodsISO 10993-4/ASTM F756-17Pass
Complement Activation SC5b-9 AssayISO 10993-4Pass
Thromboresistance Evaluation In DogsISO 10993-4Pass

Sterilization Validation

The Synxess Neurovascular Guidewire is sterilized using Ethylene Oxide (EO). The EO sterilization process of the device has been validated per ISO 11135:2014. The sterilant residuals have also been tested and met the acceptance criteria per ISO 10993-7:2008. The device is labeled as "non-pyrogenic".

Shelf-Life Validation

The shelf-life of the Synxess Neurovascular Guidewire is 3 years and is validated by testing of the accelerated aged devices.

Clinical Data

No clinical testing was deemed necessary to support the substantial equivalence of the subject device.

Conclusion

The Synxess Neurovascular Guidewire has the same intended use and indications for use, and similar technological characteristics compared to the predicate ASAHI CHIKAI Neurovascular Guide Wires. The differences in technological characteristics do not raise different questions of safety and effectiveness compared to the predicate, and the nonclinical testing discussed above demonstrates that the subject device performs as intended. Therefore, the subject device is substantially equivalent to the legally marketed predicate devices ASAHI CHIKAI Neurovascular Guide Wire, K110584, and ASAHI CHIKAI 10 Neurovascular Guide Wire, K112979.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.