(89 days)
The Pen Injector is a re-usable pen injector designed for single patient use by diabetics for the self-injection of a desired dose of insulin. The Pen Injector uses HUMALOG (insulin lispro) injection (U-100) available in 3 mL cartridges, and a single use detachable and disposable insulin pen needle (supplied separately, needle sizes including: 32G4mm, 31G4mm, 31G5mm, and 31G6mm).
The Pen Injector is a reusable mechanical pen-injector capable of injecting a dose of up to 80 units of insulin, in 1 unit increments. The Pen Injector consists of a pen injector body, a push block, a injection button, a dose display window, a dose adjustment knob, a reservoir and a pen injector cap. The intended dose is mechanically set by rotating a dose adjustment knob. The insulin is injected by depressing the knob which mechanical coupling causes the piston in the insulin cartridge to move forward thereby expelling the intended dose. The Pen Injector is intended for single user and provided for non-sterile.
The provided document describes the FDA 510(k) clearance for the Wuxi NEST Biotechnology Co., Ltd. Pen Injector (K240774). The acceptance criteria and supporting studies are primarily focused on performance testing to demonstrate substantial equivalence to the predicate device, NovoPen Echo® (K182387).
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria / Performance Metric | Requirement/Standard | Reported Device Performance (Pen Injector K240774) |
|---|---|---|
| Dose Accuracy | ISO 11608-1:2022 | Meets ISO 11608-1:2022 requirements |
| Injection Force | ISO 11608-1:2022 | Meets ISO 11608-1:2022 requirements |
| Injection Time | ISO 11608-1:2022 | Meets ISO 11608-1:2022 requirements |
| Biocompatibility - Cytotoxicity | ISO 10993-5:2009 | Meets requirements |
| Biocompatibility - Skin Sensitization | ISO 10993-10:2021 | Meets requirements |
| Biocompatibility - Irritation | ISO 10993-23:2021 | Meets requirements |
| Compatible with 3mL HUMALOG (insulin lispro) U-100 cartridges | Demonstrated compatibility | Compatible |
| Compatible with specified pen needles | Demonstrated compatibility | Compatible (listed sizes: 32G4mm, 31G4mm, 31G5mm, 31G6mm) |
| Shelf Life | 5 years (verified by accelerated aging) | 5 years |
| Service Life (reuse cycles) | 3,000 times | 3,000 times |
| Transportation Resistance | ASTM D4169:2022 | Functions as intended after transportation |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample sizes (e.g., number of devices, cartridges, or test repetitions) used for each individual performance test (dose accuracy, injection force, injection time, biocompatibility, shelf life, service life, transportation). It generally states that "corresponding performance tests were conducted" or "aging studies have been conducted."
The provenance of the data is from Wuxi NEST Biotechnology Co., Ltd. in China, and the studies are prospective in nature, conducted specifically to demonstrate the performance of the Pen Injector for this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to the reported studies. The performance tests rely on objective measurements against established international standards (ISO, ASTM) and not on human expert interpretation or ground truth establishment in a medical context.
4. Adjudication Method for the Test Set
Not applicable. The tests are based on direct physical and chemical measurements against pre-defined thresholds from international standards.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a mechanical pen injector, not an AI-powered diagnostic or imaging device, so MRMC studies or human reader performance with AI assistance are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to the core performance testing of the device (dose accuracy, injection force, etc.) which demonstrates the standalone capabilities of the Pen Injector mechanism without human-in-the-loop performance evaluation in a clinical setting. The tests confirm the device's inherent mechanical functionalities meet the specified standards.
7. The Type of Ground Truth Used
The "ground truth" for the performance tests outlined is defined by:
- International Standards: ISO 11608-1:2022 (for injection system performance), ISO 10993 series (for biocompatibility), ASTM F1980-16 (for accelerated aging/shelf life), and ASTM D4169:2022 (for transportation).
- Device Specifications: The inherent design parameters and intended performance of the Pen Injector (e.g., maximum dose, dial increments, service life).
8. The Sample Size for the Training Set
Not applicable. This device is a mechanical medical device, not an AI/machine learning model, so there is no concept of a "training set."
9. How the Ground Truth for the Training Set was Established
Not applicable, as no training set is involved for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
June 18, 2024
Wuxi NEST Biotechnology Co., Ltd. % Ryan Li, Consultant ICAS Group No.155 Pingbei Road, Minhang District Shanghai, 201109, China
Re: K240774
Trade/Device Name: Pen Injector Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: March 19, 2024 Received: March 21, 2024
Dear Ryan Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shruti N. Mistry -S
Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K240774
Device Name Pen Injector
Indications for Use (Describe)
The Pen Injector is a re-usable pen injector designed for single patient use by diabetics for the self-injection of a desired dose of insulin. The Pen Injector uses HUMALOG (insulin lispro) injection (U-100) available in 3 mL cartridges, and a single use detachable and disposable insulin pen needle (supplied separately, needle sizes including: 32G4mm, 31G4mm, 31G5mm, and 31G6mm).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary - K240774
I. Submitter Information
Submitter Name: Wuxi NEST Biotechnology Co., Ltd. Submitter Address: No. 530, Xida Road, Meicun Industrial Park, Xinwu District, Wuxi, Jiangsu, China Contact Person: Ryan Li Position: Consultant Tel.: + 86 13701581791 Email: IryryanO211@gmail.com Date of Preparation: May 22, 2024
II. Proposed Device
Device Trade Name: Pen Injector Common Name: Pen-injector Classification Name: Piston Syringe Regulation Number: 21 CFR 880.5860 Regulatory Class: Class II Product Code: FMF Review Panel: General Hospital
III. Predicate Device
Primary predicate device 510(k) Number: K182387 Trade Name: NovoPen Echo®
IV. Device Description
The Pen Injector is a reusable mechanical pen-injector capable of injecting a dose of up to 80 units of insulin, in 1 unit increments. The Pen Injector consists of a pen injector body, a push block, a injection button, a dose display window, a dose adjustment knob, a reservoir and a pen injector cap. The intended dose is mechanically set by rotating a dose adjustment knob. The insulin is injected by depressing the knob which mechanical coupling causes the piston in the insulin cartridge to move forward thereby expelling the intended dose. The Pen Injector is intended for single user and provided for non-sterile.
V. Indication for use
The Pen Injector is a re-usable pen injector designed for single patient use by diabetics for the self-injection of a desired dose of insulin. The Pen Injector uses HUMALOG (insulin lispro) injection (U-100) available in 3 mL cartridges, and a single use detachable and disposable insulin pen needle (supplied separately, needle sizes including: 32G4mm, 31G4mm, 31G5mm, and 31G6mm).
{4}------------------------------------------------
VI. Comparison of technological characteristics with the predicate devices
Table 1 summarizes the proposed device technological characteristics with compared to the predicate device under K182387.
| Item | Proposed Device | Primary Predicate Device(K182387) | Discussion |
|---|---|---|---|
| Product Name | Pen Injector | NovoPen Echo® | N/A |
| Product Code | FMF | FMF | Same |
| Regulation No. | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Class | Class II | Class II | Same |
| Indications for Use | The Pen Injector is a re-usablepen injector designed for singlepatient use by diabetics for theself-injection of a desired doseof insulin. The Pen Injectoruses HUMALOG (insulinlispro) injection (U-100)available in 3 mL cartridges,and a single use detachable anddisposable insulin pen needle(supplied separately, needlesizes including: 32G4mm,31G4mm, 31G5mm, and31G6mm). | The NovoPen Echo® is are-useable pen injectordesigned for single patient useby diabetics for theself-injection of a desireddose of insulin. The peninjector uses PenFill® 3 mLcartridge of NovoLog®, 100units/mL (U-100) [insulinaspart injection], Fiasp® 100units/mL (U-100) [insulinaspart injection], and a singleuse detachable and disposablepen needle (suppliedseparately). The pen injectorallows the user to dial thedesired dose from 0.5 to 30units in 0.5 unitincrements | Same |
| UserEnvironment | Professional healthcarefacilities and homeenvironment | Unknown | DifferentNote 1 |
| Reusable Device | Yes | Yes | Same |
| CartridgeVolume | 3 mL (300 units of U-100insulin) | 3 mL (300 units of U-100insulin) | Same |
| Dose Accuracy | Meets ISO 11608-1:2022requirements | Meets ISO 11608-1:2014requirements | Same |
| Dial Increments | 0.01 mL per incrementproviding one unit (1U) doseincrements | 0.005 mL per incrementproviding half unit (0.5U)dose increments | DifferentNote 2 |
| MaximumDelivered Dose | 80 Units | 30 Units | DifferentNote 3 |
| Shelf Life | 5 years | 5 years | Same |
| Service Life | 3,000 times of reuse | Unknown | Different |
| Note 4 | |||
| Biocompatibility | CytotoxicitySkin SensitizationIrritation | Unknown | DifferentNote 5 |
| Sterility | Not a sterile device | Not a sterile device | Same |
{5}------------------------------------------------
Discussion
Note 1
The user environment for the predicate device is unknown. However, this does not raise different questions of safety and effectiveness as both device are prescription use devices.
Note 2
Although the dial increments between the proposed device and the predicate device are different, corresponding performance tests were conducted on the proposed device to demonstrate its comply with the ISO 11608-1 standard. Therefore, this difference would not affect the substantial equivalence.
Note 3
Although the maximum delivered dose between the proposed device and the predicate device are different, corresponding performance tests were conducted on the proposed device to demonstrate the dose accuracy performance comply with the ISO 11608-1 standard. Therefore, this difference would not affect the substantial equivalence.
Note 4
A service life verification test was conducted to demonstrate that the proposed device is valid during its service life. Therefore, this difference would not affect the substantial equivalence.
Note 5
The biocompatibility tests were conducted to demonstrate the proposed device is biocompatible with intended users. Therefore, this difference would not affect the substantial equivalence.
VII. Performance Data
Non-Clinical Performance Test Conclusion
Biocompatibility
The proposed device, Pen Injector is categorized as skin contact with a duration of category A-limited(≤24hrs) according to FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process". The Biocompatibility tests were conducted to verify that the proposed devices are not adverse to human tissue based on the following standards:
- ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation .
{6}------------------------------------------------
and testing within a risk management process
- ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for ● in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests ● for irritation
Performance Testing
Device Verification Tests confirm that the Pen Injector meets the requirements of ISO 11608-1:2022 Needle-based injection systems for medical use - Requirement and test methods - Part 1: Needle-based injection systems.
The Pen Injector is intended to be used with 3mL cartridges of HUMALOG (insulin lispro) injection (U-100). Therefore, the dose accuracy, injection force and injection time performance tests of the 3mL cartridges of HUMALOG (insulin lispro) injection (U-100) with Pen Injector was performed which meets the requirements of ISO 11608-1:2022. In addition, corresponding tests are conducted between the Pen Injector and the compatible pen needles to demonstrate the compatibility.
Shelf Life and Service Life
The Pen Injector has a five-year shelf life, as well as 3000 times of reuse service life. For shelf life, in accordance with ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, aging studies have been conducted to verify a five-year shelf life of the proposed device and ensure that its functionality is successfully maintained throughout the duration of this shelf life. In addition, a service life verification test was conducted on the proposed device to ensure its validity within its service life.
Simulated Transportation
A test is conducted according to ASTM D4169:2022 to demonstrate the proposed device could function as intended after transportation.
Clinical Test Conclusion
No clinical study is included in this submission.
VIII. Conclusion
The conclusion drawn from the non-clinical tests demonstrates that the proposed device, Pen Injector is as safe, as effective, and performs as well as the legally marketed predicate device NovoPen Echo® (K182387).
§ 880.5860 Piston syringe.
(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).