K Number
K240613

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-06-20

(107 days)

Product Code
Regulation Number
888.3030
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APTUS Elbow Dorsal Olecranon Plates are indicated for fractures and osteotomies, in particular for the ulna.

Device Description

The subject device APTUS Elbow Dorsal Olecranon Plates are available in three (3) designs: Standard, Medium and Extended. The designs differ in the shape of the proximal ends of the Standard Plates are available in 6 lengths; The Medium Plates are available in 2 lenghts and the Extended Plates are available in 3 lenghts, each for left and right. All plates have appropriate for the olecranon. The subject device plates are compatible with screws and k-wires previously cleared in K051567, K09332, K112560, K232144 (for screws) and K092038 (for k-wires).

The subject device plates are manufactured from titanium alloy, conforming to ASTM F136. All plates are provided sterile and selected plates are provided non-sterile aswell.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device (APTUS Elbow Dorsal Olecranon Plates), which focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific performance-based acceptance criteria through clinical studies or AI algorithm validation. Therefore, the information needed to answer many of your questions, particularly those related to AI performance, ground truth establishment, expert adjudication, and MRMC studies, is not present in this document.

However, I can extract information regarding non-clinical testing and general regulatory context, and then explain why other requested details are absent.

Here's a breakdown based on the provided text, and an explanation of what cannot be answered:

1. A table of acceptance criteria and the reported device performance

The document mentions non-clinical (mechanical) testing based on a specific ASTM standard and FDA guidance, but it does not provide a table of acceptance criteria and reported performance values. It only states a general conclusion:

  • Acceptance Criteria Mentioned: Minimum requirements pointed out in the "FDA guidance for Industry and Food and Drug Administration Staff, Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway (2022)" and ASTM F382.
  • Reported Device Performance: "The mechanical test results show that the bending structural stiffness of the Dorsal Olecranon Medartis APTUS Elbow Plates are higher than the acceptance values indicated in the relative FDA guidance..."
  • Specific Values: No specific numerical acceptance criteria or reported device performance values (e.g., in N/mm or similar units) are provided in this regulatory letter.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified for non-clinical (mechanical) tests. For clinical studies, the document explicitly states: "Clinical data were not provided in this submission." This means there was no clinical test set for patient data.
  • Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable as no clinical data or patient test sets were used or provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not applicable. This information would be relevant for studies involving human interpretation or AI algorithm validation using expert-labeled data, which were not part of this 510(k) submission. No clinical ground truth was established from patient data.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods are used to establish ground truth in clinical or imaging studies, which were not conducted for this 510(k) submission.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. The document explicitly states: "Clinical data were not provided in this submission." An MRMC study, especially one involving AI assistance, would fall under clinical data. This submission is for a traditional medical device (bone plates), not an AI/ML-driven diagnostic or therapeutic device that would typically undergo such a study.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • No. This study type is for AI/ML algorithms. The device submitted is a mechanical orthopedic implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the non-clinical mechanical testing, the "ground truth" or reference for evaluating performance was defined by the stated FDA guidance and ASTM standards for mechanical properties (e.g., bending stiffness). There was no "ground truth" concerning patient outcomes, pathology, or expert consensus on clinical imaging, as no clinical data was used.

8. The sample size for the training set

  • Not applicable. This applies to AI/ML algorithms that require training data. This submission is for a physical medical device.

9. How the ground truth for the training set was established

  • Not applicable for the same reason as point 8.

Summary of Device Acceptance & Study

The acceptance of this device by the FDA (K240613) is based on its substantial equivalence to legally marketed predicate devices (K120070, K193633, K193639, K193554) and successful completion of non-clinical (mechanical) testing.

  • Acceptance Criteria for Substantial Equivalence: Similar intended use, similar technological characteristics (design, principle of operation, material), and comparable safety and effectiveness, supported by non-clinical testing.
  • Study Proving Acceptance:
    • Study Type: Non-clinical (mechanical) testing.
    • Objective: To demonstrate that the bending structural stiffness of the APTUS Elbow Dorsal Olecranon Plates is at least "higher than the acceptance values indicated in the relative FDA guidance for Industry and Food and Drug Administration Staff, Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway (2022)" and meets ASTM F382.
    • Results: The mechanical tests confirmed that the device's stiffness was higher than the specified acceptance values.
    • Clinical Data: No clinical data or studies were performed or submitted. The FDA 510(k) process for traditional devices like bone plates often relies on substantial equivalence and non-clinical data if the technological characteristics are well-understood.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

June 20, 2024

Medartis AG Claudia De Santis Head Global Regulatory Affairs Hochbergerstrasse 60E Basel, BS 4057 Switzerland

Re: K240613

Trade/Device Name: APTUS Elbow Dorsal Olecranon Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: March 5, 2024 Received: May 21, 2024

Dear Claudia De Santis:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Thomas Mcnamara -S

Digitally signed by Thomas Mcnamara -S Date: 2024.06.20 16:21:01 -04'00

For Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K240613

Device Name

APTUS Elbow Dorsal Olecranon Plates

Indications for Use (Describe)

The APTUS Elbow Dorsal Olecranon Plates are indicated for fractures and osteotomies, in particular for the ulna.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) #: K240613

510(k) Summary

Prepared on: 2024-06-20

Contact Details

21 CFR 807.92(a)(1)

Applicant NameMedartis AG
Applicant AddressHochbergerstrasse 60E Basel BS 4057 Switzerland
Applicant Contact Telephone+41 798835288
Applicant ContactMrs. Claudia De Santis
Applicant Contact Emailclaudia.desantis@medartis.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameAPTUS Elbow Dorsal Olecranon Plates
Common NamePlate, Fixation, Bone; Screw, Fixation, Bone
Classification NameSingle/multiple component metallic bone fixation appliances and accessories (Primary), Smooth or threaded metallic bone fixation fastener
Regulation Number888.3030 (Primary), 888.3040
Product Code(s)HRS (Primary), HWC
Legally Marketed Predicate Devices 21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first) Product Code
K120070SYNTHES VARIABLE ANGLE LCP ELBOW SYSTEM HRS
K193633APTUS Ankle Trauma System 2.8/3.5 HRS
K193639APTUS Foot System HRS
Device Description Summary 21 CFR 807.92(a)(4)

The purpose of this submission is to obtain marketing clearance for new APTUS Elbow Dorsal Olecranon Plates.

The subject device APTUS Elbow Dorsal Olecranon Plates are available in three (3) designs: Standard, Medium and Extended. The designs differ in the shape of the proximal ends of the Standard Plates are available in 6 lengths; The Medium Plates are available in 2 lenghts and the Extended Plates are available in 3 lenghts, each for left and right. All plates have appropriate for the olecranon. The subject device plates are compatible with screws and k-wires previously cleared in K051567, K09332, K112560, K232144 (for screws) and K092038 (for k-wires).

The subject device plates are manufactured from titanium alloy, conforming to ASTM F136. All plates are provided sterile and selected plates are provided non-sterile aswell.

Intended Use/Indications for Use

The APTUS Elbow Dorsal Olecranon Plates are indicated for fractures and osteotomies, in particular for the ulna.

Indications for Use Comparison

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(5)

{4}------------------------------------------------

The Indications for Use Statements (IFUS) for the primary predicate device K120070 include language for fractures, osteotomies, malunions and non-unions of the olecranon and proximal ulna. The Indications for Use Statements (IFUS) for the subject device include language for fractures and osteotomies of the ulna, which is covered by the Indications for Use Statements (IFUS) for the primary predicate device K120070.

The subject device is part of the APTUS Elbow system, which in turn is part of the APTUS fixation system. The intended use of the APTUS fixation system includes fractures, osteotomies and arthrodesis of the hand, forearm, shoulder and foot. This covers the intended use of the primary predicate device, which includes the fixation of fractures of the distal humerus, olecranon and ulna.

Technological Comparison

21 CFR 807.92(a)(6)

The primary predicate device K120070 is in support of substantial equivalence in terms of same technological characteristics, similar design characteristics (including screw holes to accommodate locking screws), similar principle of operation and similar plate material.

The additional predicate devices K193554 are in support of substantial equivalence in terms of same design characteristics regarding screw holes to accommodate Trilock locking and non-locking (cortical) screws.

All subject device final finished components (plates) are manufactured using identical material and same manufacturing steps as used for the additional predicate devices K193639, and therefore, are substantially equivalent to these previously-cleared devices regarding biocompatibility.

The additional predicate K193633 is in support of substantial equivalence in terms of same same moist heat sterilization process for subject devices provided non-sterile to the end user and are packaged using the same material.

The subject device components that are provided sterile to the end user are materials, are sterilized by the same method, and have the same sterile barrier shelf life as the additional predicate K193633.

Clinical data were not provided in this submission.

The subject device plates are compatible with Medatis screws previously cleared in K051567, K091479, K10332144. The subject device plates also are compatible with Medartis K-Wires previously cleared in K092038.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The safety and performance of the plates were established by performing to ASTM F382 and comparing the results to the minimum requirements pointed out in the relative FDA guidance for Industry and Food and Drug Administration Staff, Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway (2022). Clinical data were not provided in this submission.

The mechanical test results show that the bending structural stiffness of the Dorsal Olecranon Medartis APTUS Elbow Plates are higher than the acceptance values indicated in the relative FDA guidance for Industry and Food and Drug Administration Staff, Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway (2022)). Clinical data were not provided in this submission.

The subject device, the primary predicate device, and the additional predicate devices have intended use, have similar technological characteristics, and are made of identical or similar materials. The primary predicate, and the additional predicate devices encompass the same range of physical dimensions, are packaged in similar materials, and are similar methods.

The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.