K Number
K193633

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-03-26

(90 days)

Product Code
Regulation Number
888.3030
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

APTUS® Ankle Trauma System 2.8/3.5 is indicated for fractures, osteotomies, malunions and non-unions of the distal tibia and fibula.

Device Description

The subject device includes various designs for internal fixation of the distal tibia and distal fibula; additional lengths of 3.5 TriLock Screws (locking screws, other lengths cleared in K110908); a new design of 3.5 Cortical Screws (non-locking) in various lengths; and a new design of concave washer for use with Medartis 3.5 Cortical Screws. The subject device plates are provided in multiple anatomic designs that vary in length, width, and thickness. The plat designs include: 3.5 TriLock Distal Tibia L Plates and T Plates; 2.8/3.5 TriLock Distal Tibia Plates Medial; 2.8/3.5 TriLock Distal Tibia Plates Anterolateral; 2.8/3.5 TriLock Distal Fibula Plates Lateral, with and without Flap; 2.8 TriLock Distal Fibula Plates Crossed; and 2.8 TriLock Distal Fibula Plates Straight. The subject device plates are used with TriLock locking screws and non-locking screws (cortical and cancellous), including subject device screws and previously cleared Medartis screws. Compatible TriLock locking screws and non-locking cortical screws have a diameter of 2.8 mm and 3.5 mm, and overall lengths ranging from 8 mm to 60 mm. The 2.8 Cortical Screws (non-locking) and 3.5 Cortical Screws (non-locking) have a double-lead thread design. All TriLock Screws (locking) have a double-lead thread design. The subject device plates also are compatible with Medartis K-Wires cleared in K092038. The subject device screws include screws with the same design as 3.5 TriLock Screws cleared in K110908, with a diameter of 3.5 mm and provided in additional lengths of 10 mm, 12 mm, and 14 mm. The subject device screws also include a new design of 3.5 Cortical Screws provided with a diameter of 3.5 mm and lengths of 10 mm to 60 mm. The subject device plates, washer, and screws are manufactured from unalloyed titanium conforming to ASTM F67 or titanium alloy conforming to ASTM F136. The subject device components are provided non-sterile to the end user.

AI/ML Overview

The provided document is a 510(k) summary for the Medartis AG APTUS® Ankle Trauma System 2.8/3.5. This document describes the device, its intended use, and its substantial equivalence to previously cleared predicate devices. It focuses on the non-clinical performance data (biocompatibility, sterilization, engineering analysis, and mechanical testing) to demonstrate equivalence.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in terms of specific performance metrics with numerical targets. Instead, it relies on demonstrating substantial equivalence to predicate devices through various non-clinical tests and comparisons of technological characteristics. The "reported device performance" is essentially that it met the requirements of these non-clinical tests and is comparable to predicate devices.

Acceptance Criteria (Implied)Reported Device Performance
BiocompatibilityReferenced from previously cleared devices (K091479, K103332, K192297, and K191848). Implied to meet established biocompatibility standards.
Sterilization (for non-sterile components)To be sterilized by moist heat (by the end-user), referenced from previously cleared devices (K091479, K103332, K192297, and K191848). Implied to meet sterilization efficacy.
Sterilization (for sterile components)X-ray beam sterilization, referenced from K191848. Implied to meet sterilization efficacy.
Packaging and Sterile Barrier Shelf LifeReferenced from K191848. Implied to meet standards for maintaining sterility over time.
Engineering Analysis (Structural Integrity/Design Validation)Performed. Details not provided, but implies the design is sound and comparable to predicate devices.
Mechanical Testing (e.g., strength, fatigue)Performed according to ASTM F382 (Standard Specification for Metallic Bone Plates) and ASTM F543 (Standard Specification and Test Methods for Metallic Medical Bone Screws). The document explicitly states the primary predicate device K013248 "is in support of substantial equivalence in terms of comparison mechanical testing" and additional predicate devices are for "comparative mechanical testing."
Material CompositionPlates: Unalloyed titanium conforming to ASTM F67 or titanium alloy conforming to ASTM F136. Screws: Ti-6Al-4V alloy, ASTM F136. Washer: Titanium alloy conforming to ASTM F136. This aligns with predicate and reference devices.
Indications for UseDemonstrated as substantially equivalent to predicate devices, noting minor differences that do not impact substantial equivalence.
Technological Characteristics (Design, Components, Manufacturing)Demonstrated as substantially equivalent to predicate devices, noting similar designs, compatible components, and identical manufacturing processes and materials for Medartis AG devices.

2. Sample Size Used for the Test Set and the Data Provenance

The document explicitly states: "Clinical data were not provided in this submission." This means there was no human-patient-based test set. All data presented (biocompatibility, sterilization, mechanical testing, engineering analysis) are non-clinical.

  • Sample Size for Test Set: Not applicable, as no clinical test set was used.
  • Data Provenance: Non-clinical (laboratory/bench testing, engineering analysis). No country of origin for human data as none was used.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable, as no clinical test set requiring expert ground truth was used.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set requiring adjudication was used.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a metallic bone fixation system, not an AI-powered diagnostic or assistive tool.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical implant, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical performance evaluation is based on established engineering and materials standards (e.g., ASTM F382, ASTM F543, ASTM F67, ASTM F136), and documented performance of legally marketed predicate devices.

8. The Sample Size for the Training Set

Not applicable, as no machine learning algorithm requiring a training set was used.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no machine learning algorithm requiring a training set was used.

{0}------------------------------------------------

March 26, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medartis AG % Kevin A. Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 EL Camino Real. Suite 400 San Diego, California 92130

Re: K193633

Trade/Device Name: APTUS® Ankle Trauma System 2.8/3.5 Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC, HTN Dated: December 27, 2019 Received: December 27, 2019

Dear Kevin Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali. MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K193633

Device Name

APTUS® Ankle Trauma System 2.8/3.5

Indications for Use (Describe)

APTUS® Ankle Trauma System 2.8/3.5 is indicated for fractures, malunions and non-unions of the distal tibia and fibula.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

{3}------------------------------------------------

510(k) Summary

K193633

Medartis AG

APTUS® Ankle Trauma System 2.8/3.5

February 24, 2020

ADMINISTRATIVE INFORMATION

Manufacturer NameMedartis AGHochbergerstrasse 60ECH-4057 Basel, SwitzerlandTelephone: +41 61 633 34 34Fax: +41 61 633 34 00
Official ContactAndrea Kiefer-SchweizerHead of Quality Management and Regulatory Affairs
Representative/ConsultantKevin A. Thomas, PhDFloyd G. Larson, MS, MBAPaxMed International, LLC12264 El Camino Real, Suite 400San Diego, CA 92130Telephone: +1 858-792-1235Fax: +1 858-792-1236

Email:

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameAPTUS® Ankle Trauma System 2.8/3.5
Common NamePlate, fixation, bone
Classification NameSingle/multiple component metallic bone fixation appliancesand accessories
Classification Regulation21 CFR 888.3030
Product CodeHRS
Classification PanelOrthopedic
Reviewing OfficeOffice of Health Technology 6 (Orthopedic Devices)
Reviewing DivisionDivision of Health Technology 6 C(Restorative, Repair and Trauma Devices)

kthomas@paxmed.com flarson@paxmed.com

PREDICATE DEVICE INFORMATION

Primary Predicate Device K013248, Synthes (USA) LCP Distal Tibia Plates, Synthes (USA)

Additional Predicate Devices K092812, Synthes (USA) 2.7mm / 3.5mm LCP Anterolateral Distal Tibia Plates, Synthes (USA) K011335, Synthes One-third Tubular DCL Plate, Synthes (USA) K083213, Synthes 2.7 mm/3.5 mm LCP Distal Fibula Plates, Synthes (USA) K000684, Small Fragment Dynamic Compression Locking (DCL) System, Synthes (USA) K091479, APTUS® Foot System, Medartis AG

{4}------------------------------------------------

Reference Devices K103332, APTUS® Ulna Plates, Medartis AG K 192297. APTUS® Wrist Arthrodesis Plates, Medartis AG K191848, APTUS® Wrist Spanning Plates 2.5, Medartis AG.

INDICATIONS FOR USE STATEMENT

APTUS® Ankle Trauma System 2.8/3.5 is indicated for fractures, osteotomies, malunions and non-unions of the distal tibia and fibula.

SUBJECT DEVICE DESCRIPTION

The subject device includes various designs for internal fixation of the distal tibia and distal fibula; additional lengths of 3.5 TriLock Screws (locking screws, other lengths cleared in K110908); a new design of 3.5 Cortical Screws (non-locking) in various lengths; and a new design of concave washer for use with Medartis 3.5 Cortical Screws.

The subject device plates are provided in multiple anatomic designs that vary in length, width, and thickness. The plat designs include: 3.5 TriLock Distal Tibia L Plates and T Plates; 2.8/3.5 TriLock Distal Tibia Plates Medial; 2.8/3.5 TriLock Distal Tibia Plates Anterolateral; 2.8/3.5 TriLock Distal Fibula Plates Lateral, with and without Flap; 2.8 TriLock Distal Fibula Plates Crossed; and 2.8 TriLock Distal Fibula Plates Straight.

The subject device plates are used with TriLock locking screws and non-locking screws (cortical and cancellous), including subject device screws and previously cleared Medartis screws. Compatible TriLock locking screws and non-locking cortical screws have a diameter of 2.8 mm and 3.5 mm, and overall lengths ranging from 8 mm to 60 mm. The 2.8 Cortical Screws (non-locking) and 3.5 Cortical Screws (non-locking) have a double-lead thread design. All TriLock Screws (locking) have a double-lead thread design. The subject device plates also are compatible with Medartis K-Wires cleared in K092038.

The subject device screws include screws with the same design as 3.5 TriLock Screws cleared in K110908, with a diameter of 3.5 mm and provided in additional lengths of 10 mm, 12 mm, and 14 mm. The subject device screws also include a new design of 3.5 Cortical Screws provided with a diameter of 3.5 mm and lengths of 10 mm to 60 mm.

The subject device plates, washer, and screws are manufactured from unalloyed titanium conforming to ASTM F67 or titanium alloy conforming to ASTM F136.

The subject device components are provided non-sterile to the end user.

PERFORMANCE DATA

Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: biocompatibility referenced from K091479, K103332, K192297, and K191848; moist heat sterilization (to be performed by the end user) also referenced from K091479, K103332, K192297, and K191848; X-ray beam sterilization, packaging, and sterile barrier shelf life referenced from K191848; engineering analysis; and mechanical testing according to ASTM F382 and ASTM F543. Clinical data were not provided in this submission.

{5}------------------------------------------------

EQUIVALENCE TO MARKETED DEVICE

The subject device is substantially equivalent in indications and design principles to the primary predicate device, the additional predicate devices, and the reference devices listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements (IFUS) and the technological characteristics of the subject device, the primary predicate device, the additional predicate devices, and the reference devices.

The primary predicate device K013248 is in support of substantial equivalence in terms of comparable IFUS, plate designs (including use with locking and non-locking screws), screw designs, and materials. The primary predicate device K013248 also is in support of substantial equivalence in terms of comparison mechanical testing.

The additional predicate devices K092812, K011335, K083213, K000684 are in support of substantial equivalence in terms of comparable plate designs (including use with locking and non-locking screws), and for comparative mechanical testing.

The additional predicate device K091479 is in support of substantial equivalence in terms of comparable plate designs (including use with locking and non-locking screws), compatible screws and screw designs, and identical materials (unalloyed titanium conforming to ASTM F67 and titanium alloy conforming to ASTM F136).

The reference device K103332 is in support of substantial equivalence in terms of comparison mechanical testing.

The reference device K192297 is in support of substantial equivalence in terms of identical plate material (ASTM F67), and identical materials for the Class II accessories and instruments.

The reference device K191848 is in support of substantial equivalence in terms of identical plate material (ASTM F136), and for sterilization and shelf life for product provided sterile to the end user.

The subject device, the primary predicate device, the additional predicate devices, and the reference devices have the same intended use for internal fixation of bones including the lower extremity. The IFUS for K013248, K092812, K083213, K000684 include specific language referencing the tibia, fibula, or both.

Differences among the IFUS for the subject device, the primary predicate device, the additional predicate devices, and the reference devices include specific language not in the subject device IFUS referring to internal fixation of bones other than the lower extremity, or the use in osteopenic bone (K013248, K092812, K083213, K000684). The IFUS for the reference devices K103332, K192297, and K191848 include language referring to internal fixation of the upper extremity. These minor differences do not impact substantial equivalence because all IFUS express equivalent intended use for internal fixation of various bones.

The plates from the subject device and the plates cleared in K091479, K103332, K192297, and K191848 have the same technological characteristics, have similar design characteristics, include screw holes to accommodate locking and non-locking screws, and are made of identical materials.

The plates from the subject device, the primary predicate device K013248, and the additional predicate devices K092812, K011335, K083213, and K000684 have similar designs for specific anatomic locations and encompass a similar range of physical dimensions (overall length, and thickness).

{6}------------------------------------------------

The plates from the subject device are compatible with subject device screws and with Medartis AG screws cleared in K091479, and also are compatible with Medartis AG K-Wires cleared in K092038.

All of the subject device final finished components are manufactured in the same facilities using identical materials and identical manufacturing processes as used for the Medartis AG devices cleared in K091479. K103332, K192297, and K191848, and, therefore, are substantially equivalent to these devices regarding biocompatibility.

The subject device includes components that are provided non-sterile and components that are provided sterile. The subject device components that are provided non-sterile are packaged using the same materials, and are to be sterilized by the same methods as the Medartis AG devices cleared in K091479. K 103332, and K 192297. The subject device components that are provided sterile to the end user are packaged using the same materials, are to be sterilized by the same method, and have the same sterile barrier shelf life as the Medartis AG devices cleared in K191848.

The differences among the plates from the subject device, the primary predicate device, the additional predicate devices, and the reference devices include variations in the designs of the plates (number of designs, overall dimensions, placement of screw holes), and variations in the sizes of the compatible screws. The plates from the primary predicate device K013248 and from the additional predicate devices K092812, K011335, K083213, and K000684 are made of a different titanium alloy material compared to the subject device. These slight differences among the subject device, the primary predicate, and reference devices do not impact safety or effectiveness.

The subject device screws have similar designs and are made of the identical as the Medartis AG devices cleared in K091479 and K103332.

The subject device washer is made of the identical titanium alloy conforming to ASTM F136 as the subject device plates and screws and has a design similar to the Medartis AG washers cleared in K112560.

The subject device Class II accessories (trays and device-specific instruments) have similar designs and are made of identical materials as the Medartis AG Class II accessories and instruments previously cleared in K192297.

CONCLUSION

The subject device, the primary predicate device, and the reference device have the same intended use, have similar technological characteristics, and encompass a similar range of physical dimensions appropriate to the anatomy. The subject device and the primary predicate device are made of the identical material. The data included in this submission demonstrate substantial equivalence to the primary predicate device, the additional predicate devices, and the reference devices listed above.

{7}------------------------------------------------

Substantial Equivalence – Indications for Use Statements

Subject DeviceIndications for Use Statement
K193633APTUS® Ankle Trauma System 2.8/3.5Medartis AGAPTUS® Ankle Trauma System 2.8/3.5 is indicated for fractures, osteotomies, malunions and non-unions of the distal tibia and fibula
Primary Predicate Device
K013248Synthes (USA) LCP Distal Tibia PlatesSynthes (USA)The Synthes Locking Compression Plate (LCP) System is intended for fixation of fractures, osteotomies, and non-unions of the claviclefibula, particularly in osteopenic bone. The Synthes LCP Distal Tibia Plates are intended for fixation of complex intra- and extra-articularas a part of the Synthes Small Fragment LCP System.
Additional Predicate Devices
K092812Synthes (USA) 2.7mm / 3.5mm LCP Anterolateral Distal Tibia PlatesSynthes (USA)Synthes 2.7mm / 3.5mm LCP Anterolateral Distal Tibia Plates are indicated for fractures, osteotomies, and non-unions of the distal tibia
K011335Synthes One-third Tubular DCL PlateSynthes (USA)No 510(k) Summary or Indications for Use Statement are available on the FDA 510(k) database website
K083213Synthes 2.7 mm/3.5 mm LCP Distal Fibula PlatesSynthes (USA)The Synthes 2.7 mm/3.5 mm LCP Distal Fibula Plates are indicated for fractures, osteotomies, and non-unions of the metaphyseal area
K000684Small Fragment Dynamic Compression Locking (DCL) SystemSynthes (USA)Fixation of fractures, osteotomies, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula
K091479APTUS® Foot SystemMedartis AGThe APTUS® Foot System is intended for use in small bones, in particular in fractures, osteotomies and arthrodesis of the tarsals, metatarsals
Reference Devices
K103332APTUS® Ulna PlatesMedartis AGAPTUS® Ulna Plates are indicated for fractures and osteotomies, in particular for the ulna.
K192297APTUS® Wrist Arthrodesis PlatesMedartis AGAPTUS® Wrist Arthrodesis Plates are indicated for wrist arthrodesis.
K191848APTUS® Wrist Spanning Plates 2.5Medartis AGAPTUS® Wrist Spanning Plates 2.5 are intended for use in forearm fractures.
a.
cle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, andcular fractures and osteotomies of the distal tibia and other small bones
bia, especially in osteopenic bone.
d diaphyseal region of the distal fibula, especially in osteopenic bone.
a, particularly in osteopenic bone.
tatarsals and phalanges.

{8}------------------------------------------------

Substantial Equivalence – Technological Characteristics

FeatureSubject DevicePrimary Predicate DeviceAdditional Predicate DevicesAdditional Predicate DeviceReference Device
K193633APTUS® Ankle Trauma System 2.8/3.5K013248Synthes (USA) LCP Distal Tibia PlatesK092812Synthes (USA) 2.7mm / 3.5mm LCPAnterolateralDistal Tibia PlatesK011335Synthes One-third Tubular DCL PlateK083213Synthes 2.7 mm/3.5 mm LCP DistalFibula PlatesK000684Small Fragment Dynamic CompressionLocking (DCL) SystemK091479APTUS® Foot SystemK103332APTUS® Ulna Plates
Medartis AGSynthes USASynthes USAMedartis AGMedartis AG
Product CodeHRSHRSHRS, HWC, KTTHRS, HWCHRS, HWC
Intended UseInternal fixation of the upper extremityInternal fixation, including the lowerextremityInternal fixation, including the lowerextremityInternal fixation of the lower extremityInternal fixation of the upper extremity
Reason forPredicate/ReferenceDeviceNot applicablePlate designs, including use with lockingand non-locking screws;Comparison mechanical testingPlate designs, including use with lockingand non-locking screws;Comparison mechanical testingPlate designs, including use with lockingand non-locking screws;Compatible screws;Materials ASTM F67 and ASTM F136Reference device for comparisonmechanical testing
PlatesAnatomic plate designsMultiple designs and sizesScrew holes accommodate locking and non-locking (cortex) screwsAnatomic plate designsMultiple designs and sizesScrew holes accommodate locking and non-locking (cortex) screwsAnatomic plate designsMultiple designs and sizesScrew holes accommodate locking and non-locking (cortex) screwsAnatomic plate designsStraight, T-Shaped, Grid, Wavy and WingMultiple designs and sizesScrew holes accommodate locking and non-locking (cortex) screwsVarious overall dimensionsThickness 1.3 mm to 1.6 mmAnatomic plate designsMultiple designs and sizesScrew holes accommodate locking and non-locking (cortex) screwsVarious overall dimensionsWidths 7-36 mmLengths 7-184 mmThickness 0.6 mm to 3.2 mm
2.8/3.5 TriLock Distal Tibia PlatesMedialLocking technology- TriLock multidirectional ( $\pm$ 15°) andangular stable locking holes (locking / non-locking)- Oblong holes (non-locking)- One compression hole (non-locking)Implant dimensions- Screw holes total: 11, 13, 15, 17, 19, 21,23, 25- Left and Right- Plate length: 91-260 mm- Plate thickness: 3.25 mmDePuy Synthes 3.5 mm LCP Low BendMedial Distal Tibia PlateK013248Locking technology- LCP fixed-angle locking holes- LCP combi-holes (locking / non-locking)Implant dimensionsScrew holes in shaft: 4, 6, 8, 10, 12 and 14hole lengths- Left and Right- Plate length: 109-239 mm- Plate thickness: Not stated
2.8/3.5 TriLock Distal Tibia PlatesAnterolateralLocking technology- TriLock multidirectional ( $\pm$ 15°) andangular stable locking holes (locking / non-locking)- Oblong holes (non-locking)Implant dimensions- Screw holes total: 11, 13, 15, 17, 19, 21,23, 25- Left and Right- Plate length: 91-260 mm- Plate thickness: 3.0mmDePuy Synthes 3.5 mm LCPAnterolateral Distal Tibia PlateK092812, K000684Locking technology- LCP fixed-angle locking holes- LCP combi-holes (locking / non-locking)Implant dimensions-Screw holes in shaft: 4, 6, 8, 10, 12 and 14hole lengths- Left and Right- Plate length: 109-239 mm- Plate thickness: 2.0 mm in the metaphysealregion, 3.6 mm in the diaphyseal region
FeatureSubject DevicePrimary Predicate DeviceAdditional Predicate DevicesAdditional Predicate DeviceReference Device
K193633APTUS® Ankle Trauma System 2.8/3.5K013248Synthes (USA) LCP Distal Tibia PlatesK092812Synthes (USA) 2.7mm / 3.5mm LCPAnterolateralDistal Tibia PlatesK011335Synthes One-third Tubular DCL PlateK083213Synthes 2.7 mm/3.5 mm LCP DistalFibula PlatesK000684Small Fragment Dynamic CompressionLocking (DCL) SystemK091479APTUS® Foot SystemK103332APTUS® Ulna Plates
Medartis AGSynthes USASynthes USAMedartis AGMedartis AG
3.5 TriLock Distal Tibia L Plates andT PlatesLocking technology- TriLock multidirectional (±15°) andangular stable locking holes (locking / non-locking)- Oblong holes (non-locking)Implant dimensions- Screw holes total: 7- Left and Right in L design- Plate lengths: 50 mm- Plate thickness: 2.5 mm2.8/3.5 Distal Fibula Plate Lateral withand without FlapLocking technology- TriLock multidirectional (±15°) andangular stable locking holes (locking / non-locking)- Oblong holes (non-locking)Implant dimensions- Screw holes total: 13, 15, 17, 19, and 21With Flap14,and 16-Left and Right- Plate length: 93-192mmLCP One-third Tubular Plates 3.5K011335Locking technology- LCP fixed-angle locking holesImplant dimensions- Screw holes: 2-12 holes- Plate length: 28-148 mm- Plate thickness: no information availableLCP Distal Fibula PlatesK083213Locking technology- LCP fixed-angle locking holes- LCP combi-holes (locking / non-locking)Implant dimensions- Screw holes in shaft 3,4,5,6,7,9,11,13 and15- Left and right- Plate thickness: no information available- Plate length: 73-229 mm
- Plate thickness: 2.5 mm2.8 Distal Fibula Plate Crossed2.8 Distal Fibula Plate StraightLocking technology- TriLock multidirectional (±15°) andangular stable locking holes (locking / non-locking)- Oblong holes (non-locking)Implant dimensions- Screw holes total: 9,11,13,15 and 17- Plate lengths: 63-119 mm- Plate thickness: 2.5 mmLCP One-third Tubular Plates 3.5K011335Locking technology- LCP fixed-angle locking holesImplant dimensions- Screw holes: 2-12 holes- Plate length: 28-148 mm- Plate thickness: no information available
FeatureSubject DevicePrimary Predicate DeviceAdditional Predicate DevicesAdditional Predicate DeviceReference Device
K193633APTUS® Ankle Trauma System 2.8/3.5K013248Synthes (USA) LCP Distal Tibia PlatesK092812Synthes (USA) 2.7mm / 3.5mm LCPAnterolateralDistal Tibia PlatesK091479APTUS® Foot SystemK103332APTUS® Ulna Plates
K011335Synthes One-third Tubular DCL Plate
K083213Synthes 2.7 mm/3.5 mm LCP DistalFibula Plates
K000684Small Fragment Dynamic CompressionLocking (DCL) System
Medartis AGSynthes USASynthes USAMedartis AGMedartis AG
Plate MaterialsTi-6Al-4V alloy, ASTM F136Titanium alloy, Ti-6Al-7NbTitanium alloy, Ti-6Al-7NbUnalloyed titanium, ASTM F67Unalloyed titanium, ASTM F67
Screws
Screw DesignsSubject device screws and previouslycleared Medartis APTUS Screws(K091479, K112560, K110908):Locking screws: cortical thread form,Compatible screws:Locking screws- 3.5 mm locking screws, length 10 - 65 mmCortical screwsCompatible screws:Locking screws- 3.5 mm locking screws, length 10 - 65 mmCortical screwsSelf-tappingconventional (cortex) and lockingLocking screws:Cortical thread form, double-lead thread,self-drilling, self-tapping, fully threadedNon-locking screws:
double-lead thread,self-tapping, fully threaded- 3.5 mm Cortex screws, length 10 - 60 mm- 3.5 mm Cortex screws, length 10 - 60 mmCortical thread form, single-lead thread,self-drilling, self-tapping, fully threaded
Non-locking screws: cortical thread form,single-leadthread, self-tapping, fully threadedCancellous screws- 4.0 mm Cancellous screws, length 10 - 60mmCancellous screws- 4.0 mm Cancellous screws, length 10 - 60mmPreviously cleared Medartis APTUS Screws(K051567, K103332):Locking screws:Cortical thread form, double-lead thread,self-tapping, fully threadedNon-locking screws:Cortical thread form, single-lead thread,
self-tapping, fully threaded
Screw DiametersSubject device screws:3.5 mm locking and non-lockingLocking screws: 2.3, 2.7, 3.5 mmCortical screws: 2.3, 2.7, 3.0, 3.5 mmCancellous screws: 4.0 mmLocking screws: 2.3, 2.7, 3.5 mmCortical screws: 2.3, 2.7, 3.0, 3.5 mmCancellous screws: 4.0 mm2.0 to 2.8 mmLocking screws: 2.5 mmCortical screws: 1.5 mm, 2.5 mm
Previously cleared:2.8 mm (locking and non-locking)3.5 lockingPreviously cleared:1.5 mm and 2.5 mm
Screw LengthsSubject device screws:3.5 mm locking: 10 mm to 14 mm3.5 mm non-locking: 10 mm to 60 mmPreviously cleared:2.8 mm locking:8 mm to 60 mm2.8 mm non-locking): 8 mm to 60 mm3.5 mm locking:16 mm to 60 mmLocking screws8-26 mm (2.3 diameter)8-65 mm (2.7 diameter)6-65 mm (3.5 mm diameter)Cortical screw8-26 mm (2.3 diameter);8-65 mm (2.7 diameter)8-26 mm (3.0 diameter);6-65 mm(3.5 mm diameter)Cancellous screws12-60 mm (4.0 mm diameter)Locking screws8-26 mm (2.3 diameter)8-65 mm (2.7 diameter)6-65 mm (3.5 mm diameter)Cortical screw8-26 mm (2.3 diameter);8-65 mm (2.7 diameter)8-26 mm (3.0 diameter):6-65 mm(3.5 mm diameter)Cancellous screws12-60 mm (4.0 mm diameter)8 to 45 mmLocking screws: 8 mm to 24 mmCortical screws: 1.5 mm Ø, 8 mm to 15 mm2.5 mm Ø, 8 mm to 24 mmPreviously cleared:Various lengths
Screw MaterialsSubject device and previously clearedscrews:Ti-6Al-4V alloy, ASTM F136Compatible screws:Ti-6Al-4V alloy, ASTM F136; Stainless steelCompatible screws:Ti-6Al-4V alloy, ASTM F136; Stainless steelTi-6Al-4V alloy, ASTM F136Ti-6Al-4V alloy, ASTM F136
How Provided
SterilityProvided non-sterile andProvided sterileProvided non-sterile andProvided sterileProvided non-sterile andProvided sterileProvided non-sterileProvided non-sterile
SterilizationNon-sterile: to be sterilized by moist heatSterile: X-ray beam irradiationNon-sterile: to be sterilized by moist heatSterile: method not statedNon-sterile: to be sterilized by moist heatSterile: method not statedTo be sterilized by moist heatTo be sterilized by moist heat
UsageSingle-patient, single-useSingle-patient, single-useSingle-patient, single-useSingle-patient, single-useSingle-patient, single-use
Reference DeviceReference Device
K192297APTUS® Wrist Arthrodesis PlatesK191848APTUS® Wrist Spanning Plates 2.5
Medartis AGMedartis AG
HRS, HWCHRS, HWC
Internal fixation of the upper extremityInternal fixation of the upper extremity
Reference device for ASTM F67 material;Reference device for Class II accessory andinstrument materialsReference device for ASTM F136 material;Reference device for sterilization and shelflife for product provided sterile

{9}------------------------------------------------

Substantial Equivalence – Technological Characteristics

Reference DeviceK192297APTUS® Wrist Arthrodesis Plates
Medartis AG
Reference DeviceK191848APTUS® Wrist Spanning Plates 2.5
Medartis AG

{10}------------------------------------------------

Substantial Equivalence – Technological Characteristics

Reference DeviceReference Device
K192297APTUS® Wrist Arthrodesis PlatesK191848APTUS® Wrist Spanning Plates 2.5
Medartis AGMedartis AG
Unalloyed titanium, ASTM F67Ti-6Al-4V alloy, ASTM F136
Previously cleared:Previously cleared:
Ti-6Al-4V alloy, ASTM F136Ti-6Al-4V alloy, ASTM F136
Provided non-sterileProvided sterile
To be sterilized by moist heatX-ray beam irradiation
Single-patient, single-useSingle-patient, single-use

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.