K Number
K240387

Validate with FDA (Live)

Date Cleared
2024-06-21

(134 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angioplasty (PTA).

The PCI Guide Wires are not intended for use in the neurovasculature.

Device Description

The basic structure of the MINAMO blue / viola consists of a hybrid stainless steel and nitinol (Ni-Ti) core wire, with a stainless steel distal inner rope coil. Surrounding the inner rope coil and the distal core wire is a radiopaque platinum nickel / stainless steel outer coil. The outer and inner coils are soldered to the tapered core wire. A similar coil design is used with other ASAHI guide wires, such as the predicate MINAMO (K190176). In addition, coatings are applied on the surface of the product. The coil portion is coated with a silicone and hydrophilic coating. The proximal side of the core wire is coated with PFA, silicone and PTFE.

The nominal outer diameter of the MINAMO is 0.36 mm. The devices are available in three lengths: 190 cm, 235 cm and 300 cm.

The MINAMO blue / viola radiopaque distal tip enables the user to view the position of the tip under X-ray fluoroscopy.

All sizes are available with a straight, pre-shape, and J-shape design.

AI/ML Overview

The provided text focuses on the 510(k) premarket notification for the MINAMO blue and MINAMO viola guide wires. It describes the device, compares it to predicate and reference devices, and outlines the non-clinical and biocompatibility testing performed to demonstrate substantial equivalence.

However, the text does not contain the specific information requested regarding acceptance criteria and a study proving the device meets these criteria in the context of an AI/ML-enabled medical device.

The document states that the MINAMO blue and MINAMO viola are catheter guide wires, which are physical medical devices, not AI/ML-enabled software devices. Therefore, the questions related to:

  • AI/ML performance metrics (e.g., Sensitivity, Specificity, AUC)
  • Sample sizes for test sets in an AI/ML context
  • Number of experts for ground truth establishment
  • Adjudication methods
  • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies
  • Standalone AI performance
  • Training set sample sizes and ground truth establishment for training data

are not applicable to the submitted document as it describes a traditional medical device (guide wire) and not an AI/ML product.

The document focuses on demonstrating the substantial equivalence of the guide wires to previously cleared predicate devices through:

  • Indications for Use: Same as the predicate.
  • Technological Characteristics: Similar materials, design, operating principle, and sterilization.
  • Non-Clinical Testing/Performance Data: Bench tests following FDA guidance, demonstrating performance similar to predicate devices and meeting acceptance criteria for physical and mechanical properties (e.g., tensile strength, torque strength, tip flexibility, coating integrity, lubricity, corrosion resistance, kink resistance, radiopacity, dimensional verification, simulated use).
  • Biocompatibility: In accordance with ISO 10993.

Based on the provided text, here's what can be extracted and what cannot:


1. A table of acceptance criteria and the reported device performance:

The document states: "The in vitro bench tests demonstrated that the MINAMO blue and MINAMO viola met all acceptance criteria and performed similarly to the predicate devices."
However, the specific numerical acceptance criteria and the quantified reported device performance values are NOT provided in the document. It only lists the types of tests performed.

Types of Performance Tests (Implied Acceptance Criteria for these properties):

  • Tensile Strength
  • Torque Strength
  • Torqueability
  • Tip Flexibility
  • Coating Integrity
  • Lubricity
  • Visual Inspection
  • Corrosion Resistance
  • Kink Resistance
  • Radiopacity
  • Dimensional Verification
  • Simulated use in a clinically relevant mode
  • Coating Integrity/Particulate Evaluation in a clinically relevant mode

Biocompatibility Tests (Implied Acceptance Criteria: Non-cytotoxic, Non-sensitizer, Non-irritant, Non-toxic, Non-pyrogenic, Non-hemolytic, Not an activator, Thromboresistant):

TestConclusion
Cytotoxicity - ISO 10993-5Non-cytotoxic
Sensitization - ISO 10993-10Non-sensitizer
Irritation - ISO 10993-23Non-irritant
Systemic Toxicity - ISO 10993-11 (Acute)Non-toxic
Systemic Toxicity - ISO 10993-11 (Pyrogen)Non-pyrogenic
Hemocompatibility - ISO 10993-4 (Hemolysis)Non-hemolytic
Hemocompatibility - ISO 10993-4 (PTT)Not an activator
Hemocompatibility - ISO 10993-4 (Complement Activation)Not an Activator
Hemocompatibility - ISO 10993-4 (Thrombogenicity)Thromboresistant

For all other points (2-9), the information is NOT applicable to this document as it does not describe an AI/ML-enabled device or an AI/ML study.

2. Sample sized used for the test set and the data provenance: Not applicable (not an AI/ML study). The document refers to "bench testing" and "in vitro" tests, implying physical samples of the guide wire.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable (not an AI/ML study requiring expert review of data).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable (not an AI/ML study).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For the physical device, the "ground truth" would be established by standardized testing protocols and measurements of physical properties. For biocompatibility, it's based on ISO 10993 standards and the biological responses observed in the tests (e.g., no cellular reactivity, no delayed dermal contact sensitization).
8. The sample size for the training set: Not applicable (no AI model).
9. How the ground truth for the training set was established: Not applicable (no AI model).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a caduceus, which is a traditional symbol of medicine. The text on the right is in a bold, sans-serif font, with "FDA" in a larger font size than the rest of the text.

June 21, 2024

Asahi Intecc Co., Ltd % C. Cynthia Valenzuela Director, Quality & Regulatory Asahi Intecc USA, Inc. 3002 Dow Avenue Suite 212 Tustin, California 92780

Re: K240387

Trade/Device Name: MINAMO blue; MINAMO viola Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: February 8, 2024 Received: February 8, 2024

Dear C. Cynthia Valenzuela:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device. or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

{2}------------------------------------------------

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Lydia S. Glaw -S

Digitally signed by Lydia S. Glaw -S Date: 2024.06.21 10:27:46 -04'00'

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240387

Device Name MINAMO blue; MINAMO viola

Indications for Use (Describe)

PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angioplasty (PTA).

The PCI Guide Wires are not intended for use in the neurovasculature.

Type of Use (Select one or both, as applicable)
✖ Prescription Use (Part 21 CFR 801 Subpart D) ❍ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

[as required by 21CFR § 807.92(c)]

^SAHI INTECC CO.,LTD. Global Headquarters and R&D Center 3-100 Akatsuki-cho, Seto-shi, Aichi 489-0071 Japan TEL : +81-561-48-5551 FAX : +81-561-48-5552 http://www.asahi-intecc.co.jp/

MINAMO blue, MINAMO viola

510(k)

DATE PREPARED:31 January 2024
APPLICANT:ASAHI INTECC CO., LTD.
3-100 Akatsuki-cho, Seto
Aichi 489-0071, Japan
PRIMARY CONTACT:Mrs. Cynthia Valenzuela
Director, Regulatory Affairs
ASAHI INTECC USA, INC.
3002 Dow Avenue, Suite 212
Tustin, California 92780
Phone: (714) 442 0575
Fax: (949) 377 3255
Email: cynthiav@asahi-intecc-us.com
ALTERNATE CONTACT:Mr. Hiroshi Obara
Manager, Regulatory, Affairs
ASAHI INTECC CO., INC.
3-100 Akatsuki-cho
Seto, Aichi, Japan 489-0071
Email: hiroshi.obara@asahi-intecc.com
TRADE NAME:MINAMO blue, MINAMI viola
DEVICE CLASSIFICATION:Class II, 21CFR § 870.1330
CLASSIFICATION NAME:Catheter Guide Wire
PRODUCT CODE:DQX
PREDICATE DEVICE(S):MINAMO (K190176)
REFERENCE DEVICE(S):CROSSLEAD Peripheral Guide Wire (K213315)ASAHI PTCA Guide Wire SION blue (K122468, K191464)

Intended Use/Indications for Use:

PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angioplasty (PTA).

The PCI Guide Wires are not intended for use in the neurovasculature.

{5}------------------------------------------------

Description:

The basic structure of the MINAMO blue / viola consists of a hybrid stainless steel and nitinol (Ni-Ti) core wire, with a stainless steel distal inner rope coil. Surrounding the inner rope coil and the distal core wire is a radiopaque platinum nickel / stainless steel outer coil. The outer and inner coils are soldered to the tapered core wire. A similar coil design is used with other ASAHI guide wires, such as the predicate MINAMO (K190176). In addition, coatings are applied on the surface of the product. The coil portion is coated with a silicone and hydrophilic coating. The proximal side of the core wire is coated with PFA, silicone and PTFE.

The nominal outer diameter of the MINAMO is 0.36 mm. The devices are available in three lengths: 190 cm, 235 cm and 300 cm.

The MINAMO blue / viola radiopaque distal tip enables the user to view the position of the tip under X-ray fluoroscopy.

All sizes are available with a straight, pre-shape, and J-shape design.

Accessory

Some model of the MINAMO blue / viola are packaged with a shaping device. The shaping device can be used to facilitate shaping distal end of the guide wire.

Comparison with Predicate Device and Reference Device:

PredicateDevice Name510(k) Number
Primary PredicateMINAMOK190176
Reference DeviceCROSSLEAD Peripheral Guide WireK213315
ASAHI PTCA Guide Wire SION blueK122468K191464

The subject device has the following similarities to those which previously received 510(k) clearance.

  • Have the similar intended use and indications for use ●
  • Use the same operating principle
  • . Incorporate the same basic designs
  • Incorporate the same materials

{6}------------------------------------------------

Comparison with Predicate Device and Reference Device
---------------------------------------------------------------
MINAMO blueMINAMO violaSubjectMINAMOPredicateCROSSLEADPeripheral Guide WireReferenceASAHI PTCA GuideWire SION blueReferenceComparison
510(k)TBDK190176K213315K122468, K191464--
ManufacturerASAHI INTECCASAHI INTECCASAHI INTECCASAHI INTECC--
Regulation Number21 CFR 830.133021 CFR 830.133021 CFR 830.133021 CFR 830.1330Same
Regulation NameCatheter Guide WireCatheter Guide WireCatheter Guide WireCatheter Guide WireSame
Regulatory ClassIIIIIIIISame
Product codeDQXDQXDQXDQXSame
Indications for UsePCI Guide Wires areintended to facilitate theplacement of balloondilatation catheterduring percutaneouscoronary intervention(PCI) and percutaneoustransluminal angioplasty(PTA).The PCI Guide Wiresare not intended for usein the neurovasculature.PCI Guide Wires areintended to facilitate theplacement of balloondilatation cathetersduring percutaneouscoronary intervention(PCI) and percutaneoustransluminal angioplasty(PTA).The PCI Guide Wiresare not intended for usein the neurovasculature.This product isdesigned to direct acatheter to the desiredanatomical location inthe peripheralvasculature duringdiagnostic orinterventionalprocedures.ASAHI PTCA GuideWires are intended tofacilitate the placementof balloon dilatationcatheters duringpercutaneoustransluminal coronaryangioplasty (PTCA) andpercutaneoustransluminal angioplasty(PTA). The ASAHIPTCA Guide wires arenot to be used in theneurovasculatureSame to Predicate

{7}------------------------------------------------

MINAMO blueMINAMO violaMINAMOCROSSLEADPeripheral Guide WireASAHI PTCA GuideWire SION blueComparison
Nominal OuterDiameter0.36 mm (0.014 inch)0.36 mm (0.014 inch)0.89 mm (0.035 inch)0.36 mm (0.014 inch)Same to Predicate
Overall Length190 cm, 235 cm, 300cm190 cm, 300 cm200 cm, 300 cm180 cm, 190 cm, 300cmWithin the range ofPredicate
Outer CoilPt-NiStainless steelPt-NiStainless steelStainless SteelPt-NiStainless steelSame to Predicate
Tapered Core WireNi-TiStainless steelNi-TiStainless steelNi-TiStainless steelSame to Predicate
Inner CoilStainless steelStainless steelStainless steelStainless steelSame to Predicate
Tip ShapeStraight, Pre-shape,J-shapeStraight, Pre-shape,J-shapeStraight, AngledStraight, Pre-shape,J-shapeSame to Predicate
CoatingHydrophilicHydrophobicHydrophilicHydrophobicHydrophilicHydrophobicHydrophilicHydrophobicSame to Predicate orReference
SolderAu-SnAg-SnAg-SnAg-SnAu-SnAg-SnSame to Predicate orReference
Pipe joint sectionstructureNi-Ti pipeStainless steel coilNi-Ti pipeStainless steel coil--Same to Predicate
SterilizationEthylene oxideSAL 10-6Ethylene oxideSAL 10-6Ethylene oxideSAL 10-6Ethylene oxideSAL 10-6Same

The subject MINAMO blue / MINAMO viola has very similar technological characteristics to the predicate, with minor differences in the construction of the distal tip and in the addition of a hydrophobic coating in the guidewire. The biocompatibility and mechanical testing shows the devices to have equivalent performance.

{8}------------------------------------------------

Non Clinical Testing / Performance Data:

The substantial equivalence of the MINAMO blue and MINAMO viola was evaluated in bench testing that followed the recommendations in the FDA guidance document; Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling, 10, October 2019.

  • Tensile Strength ●
  • Torque Strength ●
  • Torqueability ●
  • Tip Flexibility ●
  • . Coating Integrity
  • Lubricity .
  • . Visual Inspection
  • Corrosion Resistance
  • Kink Resistance ●
  • Radiopacity ●
  • Dimensional Verification .
  • Simulated use in a clinically relevant mode
  • Coating Integrity/Particulate Evaluation in a clinically relevant mode ●

The in vitro bench tests demonstrated that the MINAMO blue and MINAMO viola met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate device.

{9}------------------------------------------------

BIOCOMPATIBILITY:

The MINAMO blue and MINAMO viola was tested in accordance with ISO 10993, and found to be biocompatible. The following tests were performed:

TestTest SummaryConclusion
Cytotoxicity - ISO 10993-5L929 cells/MEM Elution TestThe test system is considered suitable if nosigns of cellular reactivity (Grade 0) are notedfor both the negative control article and themedium control.Non-cytotoxic
Sensitization - ISO 10993-10KLIGMAN Maximization TestThe extracts should show no evidence ofcausing delayed dermal contact sensitizationin the guinea pig.Non-sensitizer
Irritation - ISO 10993-23Intracutaneous Injection TestThe test extract and the negative control mustexhibit similar edema and erythema scores.Non-irritant
Systemic Toxicity - ISO 10993-11Acute System Toxicity TestThe test article must not show significantlygreater biological activity than the control.Non-toxic
Systemic Toxicity - ISO 10993-11Rabbit Pyrogen Test (materialmediated)The test article should not increase the rectaltemperature of any of the animals by morethan 0.5 degrees Celsius.Non-pyrogenic
Hemocompatibility - ISO 10993-4Rabbit Blood Hemolysis TestTest article in direct contact with blood andtest article extract must be non-hemolytic.Non-hemolytic
Hemocompatibility - ISO 10993-4Partial Thromboplastin Time TestThe PTT of the plasma exposed to test articleextract should not significantly decreasedwhen compared to untreated and negativecontrols.Not an activator
Hemocompatibility - ISO 10993-4Complement Activation Assay(SC5b-9)The plasma exposed to test article mustexhibit no significant increase in SC5b-9 whencompared to activated NHS and negativecontrol after 60 minutes exposure.Not an Activator
Hemocompatibility - ISO 10993-4Thrombogenicity Study in DogsCompare results of test article to predicatecontrol for Thrombogenici response.Determine acceptability of results as part ofrisk management.Thromboresistant

Biocompatibility Results (brief)

CONCLUSION:

The MINAMO blue and MINAMO viola have similar intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate and reference device. Performance data demonstrates that the device functions as intended.

Therefore, the MINAMO blue and MINAMO viola are substantially equivalent to the predicate device.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.