K Number
K240072

Validate with FDA (Live)

Date Cleared
2024-02-09

(30 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The glove is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, gloves were tested for use with Chemotherapy drugs and Fentanyl citrate in accordance with ASTM D6978.

Device Description

Powder Free Nitrile Examination Gloves (Blue, Purple-Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes-Extra Small, Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and fentanyl citrate as per ASTM D6978-05(2019).

AI/ML Overview

The provided document describes the acceptance criteria and results of non-clinical testing for "Powder Free Nitrile Examination Gloves (Blue, Purple-Blue), Tested For Use With Chemotherapy Drugs and Fentanyl Citrate" (K240072). This is a physical medical device (gloves), not an AI/ML powered device. Therefore, questions related to AI/ML specific criteria (such as effect size of human readers with AI assistance, sample size for training sets, number of experts for ground truth, adjudication methods) are not applicable.

Here's the relevant information from the document:

1. A table of acceptance criteria and the reported device performance

Referenced StandardsTest Performed/PurposeAcceptance CriteriaReported Device Performance (Results)
ASTM D6319-19Physical Dimensions LengthMinimum 220mm for size XS and S, 230mm for size M, L, XL, XXLPass
ASTM D6319-19Physical Dimensions Palm WidthXS: 70±10mm, S: 80±10mm, M: 95±10mm, L:110±10mm, XL: 120±10mm, XXL: 130±10mmPass
ASTM D6319-19Physical Dimensions ThicknessFinger: 0.05mm (min), Palm: 0.05mm (min)Pass
ASTM D6319-19, ASTM D412-16 (2021)Physical PropertiesTensile Strength (Min 14 MPa), Elongation (Before Aging 500% and after aging 400% min)Pass
ASTM D6319-19, ASTM D5151-19Freedom from holesG-I, AQL 2.5Pass
ASTM D6319-19, ASTM D6124-06 (2017)Powder ResidueMax 2mg/glovePass
ASTM D6978-05 (2019)Chemotherapy Drugs, Opioid Drug PermeationRefer the tables for specific Minimum BDT (Breakthrough Detection Time) for each drug (e.g., >240 minutes for most, with specific lower times for Carmustine and Thiotepa)Pass (Specific BDTs are listed in the document for each drug)
ISO 10993-10:2010Irritation and Skin SensitizationNo Skin sensitization and Skin irritationIs non-sensitization and Non-irritation
ISO 10993-5:2009Cytotoxicity TestUnder the conditions of this study, no cytotoxic potentialUnder the conditions of this study, the test article extract showed potential toxicity
ISO 10993-11:2017Acute systemic toxicity studySubject showed no adverse biological reactionUnder the conditions of this study, there was no evidence of acute systemic toxicity.

Breakthrough Detection Time (BDT) for Chemotherapy Drugs and Fentanyl Citrate:

Please refer to the tables provided in pages 3, 4, 7, and 8 of the document for the detailed BDT for each specific chemotherapy drug and Fentanyl Citrate for both Blue and Purple-Blue gloves. The general acceptance criterion implied is that the gloves provide a certain level of protection (BDT), and the performance reports these measured BDTs. Notably, for Carmustine and Thiotepa, the BDTs are explicitly called out as "extremely low permeation times" with a warning not to use them with these drugs.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample sizes for each specific non-clinical test (e.g., number of gloves tested for physical dimensions, chemotherapy permeation, etc.). However, these tests are generally conducted by manufacturers according to standardized protocols (e.g., ASTM standards) which define the required sample sizes for each test. The provenance specific to the data (e.g., country of origin, retrospective/prospective) is not detailed for these non-clinical bench tests. These are laboratory tests on the product itself.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a physical device submission (nitrile examination gloves) and not an AI/ML-powered device that would require expert-established ground truth from medical imaging or similar data. The acceptance criteria are based on defined physical and chemical properties measured through standardized bench tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are typically associated with human review of data, especially in clinical or AI/ML studies, to establish ground truth. This document pertains to non-clinical bench testing of a physical product.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device, and no MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an AI/ML device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for this device is established through performance against recognized industry standards (e.g., ASTM D6319-19, ASTM D6978-05 (2019), ISO 10993 series). These standards define objective, measurable physical, chemical, and biological properties of the gloves. For example, breakthrough detection time for chemotherapy drugs is a direct measurement based on established laboratory procedures.

8. The sample size for the training set

Not applicable. This is a physical device; there is no "training set" in the context of AI/ML.

9. How the ground truth for the training set was established

Not applicable. There is no training set for this physical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 9, 2024

Luliang Hongruida Health Protection Technology Co., Ltd. Guocai Zhang General Manager Zhaokua Area, Luliang Industrial Zone, Qujing, Yunnan 655606 China

Re: K240072

Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue, Purple-Blue), Tested For Use With Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO, OPJ Dated: January 10, 2024 Received: January 10, 2024

Dear Guocai Zhang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

llan G

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices

{2}------------------------------------------------

OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240072

Device Name

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

Indications for Use (Describe)

The glove is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, gloves were tested for use with Chemotherapy drugs and Fentanyl citrate in accordance with ASTM D6978.

Blue gloves:
Chemotherapy DrugMinimum BDT (Minutes)
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240
Carmustine 3.3 mg/ml (3,300 ppm)25.5
Chloroquine 50mg/ml (50,000ppm)>240
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cyclosporin 100 mg/ml (100,000 ppm)>240
Cytarabine HCL, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel HCL, 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin/Carboplatin, 10mg/ml (10,000ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)66.8
Topotecan, 1mg/ml (1,000ppm)>240
Trisenox, 1mg/ml (1,000ppm)>240
Velcade (Bortezomib) , 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240
Fentanyl CitrateMinimum BDT (Minutes)
Fentanyl Citrate Injection (100mcg/2mL)>240

{4}------------------------------------------------

  • Please note that the following drugs have extremely low permeation times: Carmustine: 25.5 minutes, Thiotepa: 66.8 minutes

Warning: Do not use with Carmustine and Thiotepa.

Purple-Blue gloves:
Chemotherapy DrugMinimum BDT (Minutes)
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240
Carmustine 3.3 mg/ml (3,300 ppm)21.2
Chloroquine 50mg/ml (50,000ppm)>240
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cyclosporin 100 mg/ml (100,000 ppm)>240
Cytarabine HCL, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel HCL, 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin/Carboplatin, 10mg/ml (10,000ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)36.3
Topotecan, 1mg/ml (1,000ppm)>240
Trisenox, 1mg/ml (1,000ppm)>240
Velcade (Bortezomib), 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240
Fentanyl CitrateMinimum BDT (Minutes)
Fentanyl Citrate Injection (100mcg/2mL)>240
* Please note that the following drugs have extremely low permeation times:

Carmustine: 21.2 minutes, Thiotepa: 36.3 minutes

Warning: Do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

{5}------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

Zhaokua Area, Luliang Industrial Zone, Qujing City, Yunnan Province, 655606 China

510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR §807.92.

The assigned 510(K) numbers: K240072 Date Prepared: February 06, 2024

1. Owner's Identification:

Mr. Zhang Guocai Luliang Hongruida Health Protection Technology Co., Ltd. Zhaokua Area, Luliang Industrial Zone, Qujing City, Yunnan Province, 655606 China Tel: 86-311-66179668

Contact: Mr. Zhang Guocai Luliang Hongruida Health Protection Technology Co., Ltd. Zhaokua Area, Luliang Industrial Zone, Qujing City, Yunnan Province, 655606 China Tel: 909-590-1611 Email: fdareg(@hongray.com.cn or janicema(@)hongrayusa.com

2. Name of the Device:

Trade / Product Name: Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Common Name: Examination Gloves Classification Name: Patient Examination Glove Specialty Classification Regulation: 21 CFR 880.6250 Product Code: LZA LZC QDO OPJ Classification Panel: General Hospital and Personal use Device Class: Class I

3. Predicate Device Information:

Luliang Hongruida Health Protection Technology Co., Ltd. Powder Free Nitrile Examination Gloves (Blue, Purple-Blue), Tested for Use with Chemotherapy Drugs (K213440) Device Class: Class I Product Code: LZA LZC

4. Device Description:

Powder Free Nitrile Examination Gloves (Blue, Purple-Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes-Extra Small, Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and fentanyl citrate as per ASTM D6978-05(2019).

5. Device Modification

The proposed modification to the predicate device is to add the claim that the subject device was tested for use with fentanyl citrate besides the claim of use with Chemotherapy drugs. The proposed device (exactly same as predicate device) was tested for use with fentanyl citrate as per ASTM D6978. There are no differences with these two gloves in materials, manufacturing process and bench performance.

{7}------------------------------------------------

Zhaokua Area, Luliang Industrial Zone, Qujing City, Yunnan Province, 655606 China

6. Indications for Use:

The glove is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, gloves were tested for use with Chemotherapy drugs and fentanyl citrate in accordance with ASTM D6978.

Blue gloves
Chemotherapy DrugMinimum BDT (Minutes)
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240
Carmustine 3.3 mg/ml (3,300 ppm)25.5
Chloroquine 50mg/ml (50,000ppm)>240
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cyclosporin 100 mg/ml (100,000 ppm)>240
Cytarabine HCL, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel HCL, 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin/Carboplatin, 10mg/ml (10,000ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)66.8
Topotecan, 1mg/ml (1,000ppm)>240
Trisenox, 1mg/ml (1,000ppm)>240
Velcade (Bortezomib), 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240
Fentanyl CitrateMinimum BDT (Minutes)
Fentanyl Citrate Injection (100mcg/2mL)>240

{8}------------------------------------------------

Zhaokua Area, Luliang Industrial Zone, Qujing City, Yunnan Province, 655606 China

  • Please note that the following drugs have extremely low permeation times: Carmustine: 25.5 minutes, Thiotepa: 66.8 minutes Warning: Do not use with Carmustine and Thiotepa. Purple-Blue gloves

Minimum BDT (Minutes) Chemotherapy Drug Bleomycin Sulfate 15mg/ml (15000 ppm) >240 Busulfan 6mg/ml (6,000 ppm) >240 Carboplatin 10mg/ml (10,000 ppm) >240 Carmustine 3.3 mg/ml (3,300 ppm) 21.2 Chloroquine 50mg/ml (50,000ppm) >240 >240 Cisplatin 1mg/ml (1,000 ppm) Cyclophosphamide 20mg/ml (20,000 ppm) >240 Cyclosporin 100 mg/ml (100,000 ppm) >240 Cytarabine HCL, 100 mg/ml (100,000 ppm) >240 Dacarbazine 10 mg/ml (10,000 ppm) >240 Daunorubicin HCL, 5 mg/ml (5,000 ppm) >240 Docetaxel HCL, 10 mg/ml (10,000 ppm) >240 Doxorubicin HCL, 2 mg/ml (2,000 ppm) >240 Epirubicin HCL, 2 mg/ml (2,000 ppm) >240 Etoposide, 20 mg/ml (20,000 ppm) >240 Fludarabine, 25 mg/ml (25,000 ppm) >240 Fluorouracil, 50mg/ml (50,000ppm) >240 Gemcitabine, 38mg/ml (38,000ppm) >240 Idarubicin HCL, 1mg/ml (1,000ppm) >240 Ifosfamide, 50mg/ml (50,000ppm) >240 >240 Irinotecan, 20mg/ml (20,000ppm) Mechlorethamine HCI, 1mg/ml (1,000ppm) >240 Melphalan, 5mg/ml (5,000ppm) >240 Methotrexate, 25mg/ml (25,000ppm) >240 >240 Mitomycin, 0.5mg/ml (500ppm) Mitoxantrone HCL, 2mg/ml (2,000ppm) >240 Oxaliplatin, 5mg/ml (5,000ppm) >240 Paclitaxel, 6mg/ml (6,000ppm) >240 Paraplatin/Carboplatin, 10mg/ml (10,000ppm) >240 Retrovir, 10mg/ml (10,000ppm) >240 Rituximab, 10mg/ml (10,000ppm) >240 Thiotepa, 10mg/ml (10,000ppm) 36.3 Topotecan, 1mg/ml (1,000ppm) >240 Trisenox, 1mg/ml (1,000ppm) >240 Velcade (Bortezomib) , 1mg/ml (1,000ppm) >240 Vincristine Sulfate, 1mg/ml (1,000ppm) >240

Fentanyl CitrateMinimum BDT (Minutes)
Fentanyl Citrate Injection (100mcg/2mL)>240
  • Please note that the following drugs have extremely low permeation times:

Carmustine: 21.2 minutes, Thiotepa: 36.3 minutes

Warning: Do not use with Carmustine and Thiotepa.

{9}------------------------------------------------

Zhaokua Area, Luliang Industrial Zone, Qujing City, Yunnan Province, 655606 China

ItemsSubject DeviceK240072Predicate DeviceK213440Result ofcomparison
Trade NamePowder Free Nitrile ExaminationGloves (Blue, Purple-Blue),Tested for Use withChemotherapy Drugs andFentanyl CitratePowder Free NitrileExamination Gloves (Blue,Purple-Blue), Tested for Usewith Chemotherapy DrugsDifferent*
Product CodeLZA, LZC, QDO, OPJLZA, LZCDifferent*
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassIISame
Indications for UseThe glove is a disposable deviceintended for medical purposes thatis worn on the examiner's hand toprevent contamination betweenpatient and examiner. Inadditiongloves were tested for usewith Chemotherapy drugs andfentanyl citrate in accordance withASTM D6978.The glove is a disposable deviceintended for medical purposesthat is worn on the examiner'shand to prevent contaminationbetween patient and examiner. Inaddition these gloves were testedfor use with Chemotherapy drugsin accordance with ASTMD6978.Different*
Powder or PowderFreePowder FreePowder FreeSame
Design FeatureAmbidextrousAmbidextrousSame
Material useNitrileNitrileSame
ColorBlue, Purple-BlueBlue, Purple-BlueSame
SterilityNon-SterileNon-SterileSame
Single useSingle useSingle useSame
Chemotherapy Drugand Fentanyl CitratePermeation ClaimSee below comparison tableSee below comparison table/

7. Comparison of Technological Characteristics Between the Subject Device and Predicate Device

  • The subject device and predicate device are exactly the same. The subject device was tested for use with Chemotherapy drugs and fentanyl citrate, but the predicate device was tested for use with Chemotherapy drugs only.

Dimensions and Performance Comparison Table:

TechnologicalCharacteristicsSubject DeviceK240072Predicate DeviceK213440Result ofcomparison
LengthMinimum 220mm for sizeXS and S, 230mm for sizeM, L, XL, XXLMinimum 220mm for size XSand S, 230mm for size M, L,XL, XXLSame
Palm Width (size) (mm)
XS$70\pm10$$70\pm10$Same
S$80\pm10$$80\pm10$Same
M$95\pm10$$95\pm10$Same

{10}------------------------------------------------

Zhaokua Area, Luliang Industrial Zone, Qujing City,
Yunnan Province, 655606 China

L110±10110±10Same
XL120±10120±10Same
XXL130±10130±10Same
Thickness(mm)
FingerMinimum 0.05Minimum 0.05Same
PalmMinimum 0.05Minimum 0.05Same
Tensile Strength, BeforeAging14MPa, min14MPa, minSame
Ultimate Elongation,Before Aging500%, min500%, minSame
Tensile Strength, AfterAccelerated Aging14MPa, min14MPa, minSame
Ultimate Elongation,After Accelerated Aging400%, min400%, minSame
Freedom from holesIn accordance withASTM D 5151-19,following ASTM D6319- 19, G-I, AQL 2.5In accordance with ASTMD 5151-19,following ASTM D6319- 19,G-I, AQL 2.5Same
Powder-Content≤ 2 mg per glove≤ 2 mg per gloveSame
ISO 10993-10:2010Skin Irritation StudyUnder the conditions of thestudy, not an irritantUnder the conditions of thestudy, not an irritantSame
ISO 10993-10:2010MaximizationSensitization StudyUnder the conditions of thestudy, not a sensitizerUnder the conditions of thestudy, not a sensitizerSame
ISO 10993-5:2009Cytotoxicity TestUnder the conditions of this study,the test article extract showedpotential toxicityUnder the conditions of this study,the test article extract showedpotential toxicitySame
ISO 10993 Part 11Systemic toxicityUnder the conditions of this study,there was no evidence of acutesystemic toxicityUnder the conditions of this study,there was no evidence of acutesystemic toxicitySame

Chemotherapy Permeation Comparison:

Tested Chemotherapy Drug and ConcentrationMinimum BDT (Minutes)Result ofcomparison
Subject DeviceK240072Predicate DeviceK213440
BluePurple-BlueBluePurple-Blue
Bleomycin Sulfate 15mg/ml (15000 ppm)>240>240>240>240Same
Busulfan 6mg/ml (6,000 ppm)>240>240>240>240Same
Carboplatin 10mg/ml (10,000 ppm)>240>240>240>240Same
Carmustine 3.3 mg/ml (3,300 ppm)25.521.225.521.2Same
Chloroquine 50mg/ml (50,000ppm)>240>240>240>240Same
Cisplatin 1mg/ml (1,000 ppm)>240>240>240>240Same

{11}------------------------------------------------

Zhaokua Area, Luliang Industrial Zone, Qujing City, Yunnan Province, 655606 China

Cyclophosphamide 20mg/ml (20,000 ppm)>240>240>240>240Same
Cyclosporin 100 mg/ml (100,000 ppm)>240>240>240>240Same
Cytarabine HCL, 100 mg/ml (100,000 ppm)>240>240>240>240Same
Dacarbazine 10 mg/ml (10,000 ppm)>240>240>240>240Same
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240>240>240>240Same
Docetaxel HCL, 10 mg/ml (10,000 ppm)>240>240>240>240Same
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240>240>240>240Same
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240>240>240>240Same
Etoposide, 20 mg/ml (20,000 ppm)>240>240>240>240Same
Fludarabine, 25 mg/ml (25,000 ppm)>240>240>240>240Same
Fluorouracil, 50mg/ml (50,000ppm)>240>240>240>240Same
Gemcitabine, 38mg/ml (38,000ppm)>240>240>240>240Same
Idarubicin HCL, 1mg/ml (1,000ppm)>240>240>240>240Same
Ifosfamide, 50mg/ml (50,000ppm)>240>240>240>240Same
Irinotecan, 20mg/ml (20,000ppm)>240>240>240>240Same
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240>240>240>240Same
Melphalan, 5mg/ml (5,000ppm)>240>240>240>240Same
Methotrexate, 25mg/ml (25,000ppm)>240>240>240>240Same
Mitomycin, 0.5mg/ml (500ppm)>240>240>240>240Same
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240>240>240>240Same
Oxaliplatin, 5mg/ml (5,000ppm)>240>240>240>240Same
Paclitaxel, 6mg/ml (6,000ppm)>240>240>240>240Same
Paraplatin/Carboplatin, 10mg/ml (10,000ppm)>240>240>240>240Same
Retrovir, 10mg/ml (10,000ppm)>240>240>240>240Same
Rituximab, 10mg/ml (10,000ppm)>240>240>240>240Same
Thiotepa, 10mg/ml (10,000ppm)66.836.366.836.3Same
Topotecan, 1mg/ml (1,000ppm)>240>240>240>240Same
Trisenox, 1mg/ml (1,000ppm)>240>240>240>240Same
Velcade (Bortezomib) , 1mg/ml (1,000ppm)>240>240>240>240Same
Vincristine Sulfate, 1mg/ml (1,000ppm)>240>240>240>240Same

Fentanyl Citrate Permeation Comparison:

Minimum BDT (Minutes)
Fentanyl Citrate and ConcentrationSubject DeviceK240072Predicate DeviceK213440Result ofcomparison
BluePurple-BlueBluePurple-Blue
Fentanyl Citrate Injection (100mcg/2mL)>240>240//Different*
  • The subject device and predicate device are exactly the same. The subject device was tested for use with Chemotherapy drugs and fentanyl citrate, but the predicate device was tested for use with Chemotherapy drugs only.

{12}------------------------------------------------

Zhaokua Area, Luliang Industrial Zone, Qujing City, Yunnan Province, 655606 China

8. Summary of Non-Clinical Testing

Non-clinical tests were performed to verify that the subject device will meet the acceptance criteria of the performance test shown below:

Referenced standardsTest Performed/PurposeAcceptance CriteriaResults
ASTM D6319-19Physical Dimensions LengthMinimum 220mm for size XSand S, 230mm for size M, L,XL, XXLPass
ASTM D6319-19Physical Dimensions PalmWidthXS: 70±10mmS: 80±10mmM: 95±10mmL:110±10mmXL: 120±10mmXXL: 130±10mmPass
ASTM D6319-19Physical Dimensions ThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)Pass
ASTM D6319-19ASTM D412-16 (2021)Physical PropertiesTensile Strength (Min14 Mpa)Elongation (Before Aging500% and after aging 400%)MinPass
ASTM D6319-19ASTM D5151-19Freedom from holesG-I, AQL 2.5Pass
ASTM D6319-19ASTM D6124-06 (2017)Powder ResidueMax 2mg/glovePass
ASTM D6978-05 (2019)Chemotherapy Drugs, OpioidDrug PermeationRefer the above tablesPass
ISO 10993-10:2010Irritation and Skin SensitizationNo Skin sensitization and SkinirritationIs non-sensitization andNon-irritation
ISO 10993-5:2009Cytotoxicity TestCytotoxicity TestUnder the conditions ofthis study, no cytotoxicpotentialUnder theconditions of thisstudy, the testarticle extractshowed potentialtoxicity
ISO 10993-11:2017systemic toxicityAcute systemic toxicitystudySubject showed no adversebiological reactionUnder theconditions of thisstudy, there was noevidence of acutesystemic toxicity.

9. Summary of Clinical Testing:

No clinical testing was submitted for the subject device.

10. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.