K Number
K240045

Validate with FDA (Live)

Device Name
QStat Cartridge
Manufacturer
Date Cleared
2024-03-27

(82 days)

Product Code
Regulation Number
864.5430
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous or arterial whole blood sample using the Quantra Hemostasis Analyzer. The QStat Cartridge includes tests to assess coagulation via the intrinsic and extrinsic pathways and includes a test with tranexamic acid to evaluate clot lysis characteristics.

The OStat Cartridge is intended for in vitro diagnostic use by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS), and Clot Stability to Lysis (CSL).

The QStat Cartridge is indicated for the evaluation of blood coagulation and clot lysis in patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in trauma and liver transplantation procedures.

Results obtained with the QStat Cartridge should not be the sole basis for patient diagnosis.

For prescription use only.

Device Description

The QStat Cartridge is a single-use, multi-channel disposable plastic cartridge used with the Quantra Hemostasis Analyzer for the evaluation of blood coagulation and clot lysis. The measurements are performed in four test channels of the disposable cartridge which enable differential testing with different sets of reagents without the need for any reagent preparation or controlled pipetting. The cartridge utilizes a citrated evacuated blood collection tube filled with a patient whole blood sample The proprietary technology SEER Sonorheometry measures the evolution of shear modulus (i.e., clot stiffness) in all four channels as a function of time. The QStat Cartridge is intended for use in patients 18 years or older by professionals in a hospital setting (point of care or laboratory) to assess possible hypocoagulable and hypercoagulable conditions in trauma and liver transplantation procedures.

Each QStat Cartridge is pre-filled with lyophilized reagent beads individually sealed in an airtight pouch. After a QStat Cartridge is removed from its primary packaging, it is inserted into the instrument dock. A whole blood sample, collected in a 3.2% sodium citrate anticoagulant blood collection tube (minimum volume 2.7 mL), is attached directly to the cartridge and the test is initiated using the touch screen interface on the Quantra Hemostasis Analyzer. The cartridge is the only component of the Quantra System that is in direct contact with blood. The fluidic system within the instrument draws the sample into the cartridge where it is warmed to 37°C, aliquoted, introduced and mixed with the lyophilized reagents, and analyzed. When the test is complete, the cartridge is released from the dock to be disposed of in an appropriate biosafety sharps container.

AI/ML Overview

The provided text is a 510(k) Premarket Notification from the FDA regarding the HemoSonics QStat Cartridge. The purpose of this submission is to expand the approved sample matrices to include arterial whole blood, in addition to the previously cleared venous whole blood.

The document does not detail a study proving the device meets acceptance criteria in the manner of an AI/ML device, as it concerns an in vitro diagnostic (IVD) device for measuring blood viscoelastic properties. Therefore, many of the requested items (e.g., number of experts, adjudication methods, MRMC study, training sets, ground truth establishment for AI/ML) are not applicable or described in this type of submission.

However, based on the information provided, we can infer some aspects relevant to device performance and the expansion of its indications for use.

Here's an analysis based on the provided text, addressing the applicable points and explaining why others are not relevant in this context:

1. A table of acceptance criteria and the reported device performance

The document defines the device's function and indications for use, but it does not provide a table of quantitative acceptance criteria and reported performance metrics for the device itself (e.g., accuracy, precision) in the context of an AI/ML algorithm evaluation. This is a 510(k) submission for a modification to an existing IVD, primarily focused on expanding the sample type. The "acceptance criteria" here refer to demonstrating substantial equivalence for the new claim (arterial blood).

What is reported is the intent to demonstrate equivalency between arterial and venous whole blood samples.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document states: "HemoSonics is submitting this Special 510(k) to demonstrate the equivalency of arterial and venous whole blood samples analyzed on the Quantra Hemostasis Analyzer with the QStat Cartridge in order to expand the sample matrices accepted on this system."

While it explicitly states the intent to perform a study for equivalency, the sample size and data provenance (country, retrospective/prospective) are not provided in this document. This information would typically be detailed in a separate study report submitted as part of the 510(k) supporting data, which is not included in this high-level summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This question is typically relevant for interpretative devices, especially those using AI/ML where human expert interpretation forms the ground truth. For an IVD like the QStat Cartridge, which measures objective viscoelastic properties of blood, the "ground truth" is established by the accuracy and precision of the measurement itself against a reference method or known values, not by expert consensus on image interpretation. Therefore, this concept of "experts" to establish ground truth does not directly apply here.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Similar to point 3, adjudication methods are used in AI/ML performance studies where multiple human readers might disagree, and a consensus process is needed to establish ground truth. This is not applicable to a laboratory device that performs quantitative measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC study was not done, nor would it be relevant for this type of device. MRMC studies are specific to evaluating the impact of AI assistance on human reader performance in tasks like image interpretation (e.g., radiology for diagnostic accuracy). The QStat Cartridge is an IVD that provides quantitative measurements of blood coagulation; it does not involve human "readers" interpreting cases in the sense of an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is also more applicable to AI/ML software. The QStat Cartridge is a physical device that performs measurements. Its performance is evaluated fundamentally in a standalone manner (the device's ability to accurately measure blood properties), but not in the context of an "algorithm only" as understood for AI/ML. The device generates quantitative parameters (Clot Time, Clot Stiffness, etc.). An "algorithm" within the device would process sensor data to yield these parameters, and the evaluation would focus on the accuracy of these output parameters.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For an IVD like the QStat Cartridge, the "ground truth" for its measurements would be established through:

  • Reference Methods: Comparison of the QStat Cartridge's measurements (Clot Time, Clot Stiffness, etc.) against established, validated reference methods for measuring blood viscoelastic properties or related coagulation parameters.
  • Known Samples: Testing with samples having known or characterized coagulation properties (e.g., calibrated controls, samples spiked with known substances).
  • Clinical Correlation: While not a direct "ground truth" for the measurement itself, the clinical utility of the device's outputs is validated by correlating its results with patient clinical status or outcomes.

The document states the device provides "semi-quantitative indications" and its results "should not be the sole basis for patient diagnosis," indicating its role as a supportive diagnostic tool. The specific ground truth used for the arterial vs. venous equivalency study is not detailed here, but it would involve comparing measurements from paired arterial and venous samples from the same patient.

8. The sample size for the training set

This question is applicable to AI/ML models. Since this is an IVD device measuring physiological parameters, the concept of a "training set" for an AI model as typically distinguished from a "test set" does not apply in the same way. The device's measurement principles are based on established biophysical principles (SEER Sonorheometry), not on training a machine learning algorithm on a dataset. Any internal "algorithms" would be deterministic or model-based, not learned from data in the AI sense.

9. How the ground truth for the training set was established

Not applicable, as explained in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 27, 2024

HemoSonics LLC Deborah Winegar Vice-President, Clinical Affairs 4020 Stirrup Creek Drive Suite 105 Durham, North Carolina 27703

Re: K240045

Trade/Device Name: QStat Cartridge Regulation Number: 21 CFR 864.5430 Regulation Name: Coagulation System For The Measurement Of Whole Blood Viscoelastic Properties In Perioperative Patients Regulatory Class: Class II Product Code: QFR Dated: March 19, 2024 Received: March 19, 2024

Dear Deborah Winegar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding

{1}------------------------------------------------

When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Min Wu

Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K240045

Device Name

QStat Cartridge

Indications for Use (Describe)

The QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coaqulation and clot lysis state of a 3.2% citrated venous or arterial whole blood sample using the Quantra Hemostasis Analyzer. The QStat Cartridge includes tests to assess coagulation via the intrinsic and extrinsic pathways and includes a test with tranexamic acid to evaluate clot lysis characteristics.

The OStat Cartridge is intended for in vitro diagnostic use by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS), and Clot Stability to Lysis (CSL).

The QStat Cartridge is indicated for the evaluation of blood coagulation and clot lysis in patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in trauma and liver transplantation procedures.

Results obtained with the QStat Cartridge should not be the sole basis for patient diagnosis.

For prescription use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SPECIAL 510(K) SUMMARY

APPLICANT INFORMATION

Submission Date:January 2024
Submitter Information:HemoSonics, LLC4020 Stirrup Creek Drive, Suite 105Durham, NC 27703Phone: 919-244-6990
Contact Person:Deborah Winegar, PhDEmail: dwinegar@hemosonics.comPhone: 919-244-6990

PROPRIETARY AND ESTABLISHED NAMES

QStat® Cartridge

REGULATORY INFORMATION

Trade/Device Name:QStat Cartridge
Regulation Number:21 CFR 864.5430
Regulation Name:Coagulation system for the measurement of whole bloodviscoelastic properties in perioperative patients
Regulatory Classification:Class II
Product Code:QFR

PURPOSE OF SUBMISSION

To add a sample matrix to the intended use and indications for use statement for the QStat Cartridge. This change does not alter the device's fundamental scientific technology.

MEASURAND

The combination of clot time and clot stiffness parameters measured from the four channels of the cartridge provides information about the functional role of coagulation factors, fibrinogen, and platelets in the sample.

TYPE OF TEST

The QStat Cartridge is an in vitro diagnostic device used with the Quantra Hemostasis Analyzer to assess a patient's coagulation system by measuring the viscoelastic properties of a blood sample during clot formation and clot lysis. The QStat Cartridge is used with the Quantra Hemostasis Analyzer (instrument) and QStat Controls (external Quality Control materials).

{4}------------------------------------------------

INTENDED USE AND INDICATIONS FOR USE STATEMENT

The OStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous or arterial whole blood sample using the Quantra Hemostasis Analyzer. The QStat Cartridge includes tests to assess coagulation via the intrinsic and extrinsic pathways and includes a test with tranexamic acid to evaluate clot lysis characteristics.

The OStat Cartridge is intended for in vitro diagnostic use by trained professionals at the pointof-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS), and Clot Stability to Lysis (CSL).

The OStat Cartridge is indicated for the evaluation of blood coagulation and clot lysis in patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in trauma and liver transplantation procedures.

Results obtained with the QStat Cartridge should not be the sole basis for patient diagnosis.

For prescription use only.

DEVICE DESCRIPTION

The QStat Cartridge is a single-use, multi-channel disposable plastic cartridge used with the Quantra Hemostasis Analyzer for the evaluation of blood coagulation and clot lysis. The measurements are performed in four test channels of the disposable cartridge which enable differential testing with different sets of reagents without the need for any reagent preparation or controlled pipetting. The cartridge utilizes a citrated evacuated blood collection tube filled with a patient whole blood sample The proprietary technology SEER Sonorheometry measures the evolution of shear modulus (i.e., clot stiffness) in all four channels as a function of time. The QStat Cartridge is intended for use in patients 18 years or older by professionals in a hospital setting (point of care or laboratory) to assess possible hypocoagulable and hypercoagulable conditions in trauma and liver transplantation procedures.

Each QStat Cartridge is pre-filled with lyophilized reagent beads individually sealed in an airtight pouch. After a QStat Cartridge is removed from its primary packaging, it is inserted into the instrument dock. A whole blood sample, collected in a 3.2% sodium citrate anticoagulant blood collection tube (minimum volume 2.7 mL), is attached directly to the cartridge and the test is initiated using the touch screen interface on the Quantra Hemostasis Analyzer. The cartridge is the only component of the Quantra System that is in direct contact with blood. The fluidic system within the instrument draws the sample into the cartridge where it is warmed to 37°C, aliquoted, introduced and mixed with the lyophilized reagents, and analyzed. When the test is complete, the cartridge is released from the dock to be disposed of in an appropriate biosafety sharps container.

{5}------------------------------------------------

Table 1 summarizes the lyophilized reagents contained in each cartridge channel of the QStat Cartridge and the output parameter reported. Clot times and clot stiffness values obtained from the measurements performed by the QStat Cartridge are combined to form parameters that depict the functional status of the patient's coagulation system. Four (4) of the parameters are measured and two (2) are calculated. The assay provides the following information for each patient sample: Clot Time (CT), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS) and Clot Stability to Lysis (CSL).

ChannelReagentsQStat Cartridge Output Parameter (units of measure)
1Kaolin, calcium buffers&stabilizersClot Time (CT) (seconds)
2Thromboplastin, tranexamic acid(TXA), polybrene, calcium,buffers, and stabilizersNo direct output (see calculated parameters)
3Thromboplastin, polybrene,calcium, buffers & stabilizersClot Stiffness (CS) (hectoPascals)
4Thromboplastin, polybrene,abciximab, calcium, buffers &stabilizersFibrinogen Contribution (FCS) (hectoPascals)
Calculated Parameters
2&3See aboveClot Stability to Lysis (CSL) (percent)
3&4See abovePlatelet Contribution to Clot Stiffness (PCS) (hectoPascals)

Table 1. QStat Cartridge Output Parameters

The analyzer displays the test results (n=5) in three different views: dial display screen, stiffness curves data, and trend screen. The dial display screen is the primary viewing screen and has a dial for each of the five output parameters. Each dial shows the reference range, assay measurement range, parameter abbreviation, and the numerical result for the corresponding parameter. The stiffness curves are a graphical display of shear modulus measurements over time that enable the user to view the development of clot stiffness over time. The trends screen displays results from a patient for up to six time points.

There are two levels of external OStat Controls (QSL1 and OSL2) that are supplied separately (required but not provided materials) for testing on the Quantra System when changing cartridge lots, changing control lots, or after significant changes are made to the Quantra instrument (e.g., software update).

DEVICE MODIFICATION DESCRIPTION

The QStat Cartridge was previously cleared under K213917 with a sample matrix claim of venous whole blood. HemoSonics is submitting this Special 510(k) to demonstrate the equivalency of arterial and venous whole blood samples analyzed on the Quantra Hemostasis Analyzer with the QStat Cartridge in order to expand the sample matrices accepted on this system.

{6}------------------------------------------------

SUBSTANTIAL EQUIVALENCE INFORMATION

Predicate Device Name: QStat Cartridge

Predicate 510(k) Number: K213917

Comparison with the Predicate:

Table 2 provides an overall comparison of the modified QStat Cartridge with the previously cleared QStat Cartridge.

Modified DevicePredicate Device
QStat Cartridge(Subject of Special 510(k))QStat Cartridge(K213917)
Similarities
ManufacturerHemoSonics, LLCSame
Trade NameQStat CartridgeSame
Common NameWhole Blood Hemostasis SystemSame
Classification NameCoagulation system for the measurement of whole blood viscoelastic properties in perioperative patientsSame
Regulation Number21 CFR 864.5430Same
Product CodeQFRSame
Device ClassIISame
Location of UsePoint of care and laboratory settingsSame
DisposablesQStat Cartridge (multichannel cartridge)Quantra Quality Controls (Level 1 and Level 2)Same
Analyzer HardwareQuantra Hemostasis Analyzer HS-002Same
Differences
Indications for useThe QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous or arterial whole blood sample using the Quantra Hemostasis Analyzer. The QStat Cartridge includes tests to assess coagulation via the intrinsic and extrinsic pathways and includes a test with tranexamic acid to evaluate clot lysis characteristics.The QStat Cartridge is intended for in vitro diagnostic use by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), ClotThe QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous whole blood sample using the Quantra Hemostasis Analyzer. The QStat Cartridge includes tests to assess coagulation via the intrinsic and extrinsic pathways and includes a test with tranexamic acid to evaluate clot lysis characteristics.The QStat Cartridge is intended for in vitro diagnostic use by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot
Modified DevicePredicate Device
QStat CartridgeQStat Cartridge
(Subject of Special 510(k))(K213917)
Stiffness (CS), Fibrinogen Contribution toClot Stiffness (FCS), Platelet Contributionto Clot Stiffness (PCS), and Clot Stabilityto Lysis (CSL).Stiffness (CS), Fibrinogen Contribution toClot Stiffness (FCS), Platelet Contributionto Clot Stiffness (PCS), and Clot Stabilityto Lysis (CSL).
The QStat Cartridge is indicated for theevaluation of blood coagulation and clotlysis in patients age 18 years and older toassess possible hypocoagulable andhypercoagulable conditions in trauma andliver transplantation procedures.The QStat Cartridge is indicated for theevaluation of blood coagulation and clotlysis in patients age 18 years and older toassess possible hypocoagulable andhypercoagulable conditions in trauma andliver transplantation procedures.
Results obtained with the QStat Cartridgeshould not be the sole basis for patientdiagnosis.Results obtained with the QStat Cartridgeshould not be the sole basis for patientdiagnosis.
For prescription use only.For prescription use only.
Sample Type(s)Arterial and venous whole bloodVenous whole blood
Analyzer SoftwareVer 2.4.1 (cleared in K232215)Ver 2.0.36

Table 2: Comparison between K213917 and Modified QStat Cartridge

{7}------------------------------------------------

§ 864.5430 Coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients.

(a)
Identification. A coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients is an in vitro diagnostic device used to evaluate blood coagulation, fibrinolysis, or both, in perioperative patients, as an aid in the assessment of coagulopathies when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include detailed documentation of, and results from, the following:
(i) A study assessing precision using protocols determined to be acceptable by FDA, to cover the measurement range for each reported parameter (test output). Testing must include native specimens with coagulation profiles representative of the intended use population. In order to cover the measuring range, testing may include a limited number of contrived specimens, not to exceed 10 to 20 percent, or as otherwise deemed appropriate by FDA. The contrived specimens must be prepared to resemble clinical specimens. This testing must evaluate repeatability and reproducibility and provide assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system;
(ii) Studies that demonstrate the performance of each parameter (test output) throughout the claimed measurement range, to include linearity studies or dose-response studies, as applicable to the parameter (test output);
(iii) Potential interferent study that includes evaluation of hemolyzed and lipemic samples as potential interferents; exogenous and endogenous interferents associated with each patient population intended for use with the device, and which might be expected to affect assay performance, must be evaluated; and potential interferents that are specific for, or related to, the technology or methodology of the device. Evaluation of all potential interferents must be performed using a protocol determined to be acceptable to the FDA (
e.g., an FDA-recognized standard) and include both normal and abnormal specimens covering coagulation profiles representative of the intended use population;(iv) A study that evaluates specimen stability under the intended conditions for specimen collection, handling, and storage, using samples that cover the coagulation profiles representative of the intended use population, and using protocols determined to be acceptable by FDA;
(v) A multisite clinical study, determined to be acceptable by FDA, demonstrating performance, relative to clinically relevant and clinically validated laboratory test(s) for each parameter (test output). Further, the study must meet all of the following criteria:
(A) The study must be performed in the intended use population and include representation from all patient populations for whom the device is intended to be used. Potential endogenous and exogenous interferents for each target patient population must be evaluated or known prior to the study;
(B) The study must be conducted at a minimum of three external sites representative of the intended use setting by the intended operators;
(C) Test samples must be collected at time intervals relevant to the device's use in the intended use population;
(D) Clinical specimens, which cover coagulation profiles representative of the intended use population, must be evaluated at each of the three clinical sites in the study;
(E) Analysis of the concordance of clinical interpretation of patient coagulation status made from individual test parameter (test output) results as compared to clinical interpretation of coagulation status from a clinically relevant laboratory test or tests (
e.g., a comparative viscoelastic device or standard laboratory tests) must be conducted; and(F) Expected (reference) values for each parameter (test output) must be demonstrated by testing a statistically appropriate number of samples from apparently healthy normal individuals;
(vi) For a device with a user interface that has information that needs to be interpreted by the user in correctly using the device to achieve the intended test results or a device that does not provide a final output that is a comprehensive interpretation of all parameter (test output) results, a study evaluating the ability of device users to correctly interpret results;
(vii) For any device indicated to guide blood product use, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding blood product use; and
(viii) For any device indicated to guide use of medication, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding use of medication.
(2) The labeling required under § 809.10(b) of this chapter must include the following:
(i) A summary of results from the study required by paragraph (b)(1)(i) of this section, including repeatability, reproducibility, and assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system.
(ii) The claimed measurement range of each parameter (test output), as supported by demonstrated performance of the parameter (test output) throughout the claimed measurement range, including studies required by paragraphs (b)(1)(i) through (iii) and (v) of this section, and, if applicable, paragraphs (b)(1)(vii) and (viii) of this section.
(iii) Identification of known interferents, including all endogenous, exogenous, technology-specific, and patient population-specific interferents, specific to each parameter (test output). The information must include the concentration(s) or level(s) at which interference was found to occur and the concentration range or levels at which interference was not found to occur.
(iv) Information regarding the multisite clinical study required by paragraph (b)(1)(v) of this section, including:
(A) Each patient population evaluated;
(B) Each intended use setting and the operators;
(C) A summary of the results, including the concordance analysis to clinically relevant laboratory test(s); and
(D) Demonstrated expected (reference) values for each parameter (test output).
(3) The labeling required under § 809.10 of this chapter must include the following:
(i) A limiting statement that the result(s) from the device is(are) not intended to be used as the sole basis for a patient diagnosis.
(ii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(vii) of this section that specifically validate an indication for the device's use in guiding blood product use, a limiting statement that the device has not been evaluated to guide blood product use.
(iii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(viii) of this section that specifically validate an indication for the device's use in guiding use of medication, a limiting statement that the device has not been evaluated to guide use of medication.