K Number
K233970

Validate with FDA (Live)

Date Cleared
2024-01-16

(32 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nephron Nitrile™. Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Bleomycin Sulfate (15.0 mg/ml) Busulfan (6.0 mg/ml) Carboplatin (10.0 mg/ml) Cisplatin (1.0 mg/ml) Cyclophosphamide (20.0 mg/ml) Cytarabine (100.0 mg/ml) Dacarbazine (10.0 mg/ml) Daunorubicin HCl (5.0 mg/ml) Docetaxel (10.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Epirubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fludarabine (25.0 mg/ml) Fluorouracil (50.0 mg/ml) Gemcitabine (38.0 mg/ml) Idarubicin HCl (1.0 mg/ml) Ifosfamide (50.0 mg/ml) Irinotecan (20.0 mg/ml) Mechlorethamine HCl (1.0 mg/ml) Melphalan (5.0 mg/ml) Methotrexate (25.0 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone HCl (2.0 mg/ml) Paclitaxel (6.0 mg/ml) Rituximab (10.0 mg/ml) Trisenox (1.0 mg/ml) Vincristine Sulfate (1.0 mg/ml) The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutes Warning: Not for Use with: Carmustine, Thiotepa The tested Opioid is: Fentanyl Citrate Injection (100 mcg/2 mL)

Device Description

Nephron Nitrile™. Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) are Class I patient examination gloves, bearing the product codes LZA, LZC, OPJ, QDO (21 CFR 880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). The powder-free gloves are made from Nitrile (NBR) latex and are blue in color. The product is non-sterile, fingertip textured, ambidextrous with beaded cuff, and single use only.

Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) include glove sizes: X-Small, Small, Medium, Large, X-Large and XX-Large.

AI/ML Overview

This document describes the non-clinical testing performed for the Nephron Nitrile™ Nitrile Powder-Free Examination Blue Gloves to meet performance and safety criteria.

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319-19 (Nitrile Examination Gloves for Medical Application)Length of the glovesExtra-Small: 220 mm min; Small: 220 mm min; Medium: 230 mm min; Large: 230 mm min; X-Large: 230 mm min; XX-Large: 230 mm minExtra-Small: 242 mm; Small: 243 mm; Medium: 235 mm; Large: 237 mm; X-Large: 250 mm; XX-Large: 238 mm
Width of the glovesXS: 70±10 mm; S: 80±10 mm; M: 95±10 mm; L: 110±10 mm; XL: 120±10 mm; XXL: 130±10 mmExtra-Small: 74 mm; Small: 85 mm; Medium: 95 mm; Large: 113 mm; X-Large: 121 mm; XX-Large: 129 mm
Thickness of the glovesPalm: 0.05 mm min for all sizes; Finger: 0.05 mm min for all sizesExtra-Small: Palm 0.112 mm, Finger 0.145 mm; Small: Palm 0.095 mm, Finger 0.115 mm; Medium: Palm 0.077 mm, Finger 0.111 mm; Large: Palm 0.106 mm, Finger 0.109 mm; X-Large: Palm 0.089 mm, Finger 0.115 mm; XX-Large: Palm 0.113 mm, Finger 0.107 mm
Physical properties - Tensile strength (Before aging)14MPa min for all sizesExtra-Small: 33.74MPa; Small: 32.35MPa; (Medium, Large, X-Large, XX-Large results are the same as predicate and meet criteria)
Physical properties - Tensile strength (After aging)14MPa min for all sizesExtra-Small: 35.99MPa; Small: 32.14MPa; (Medium, Large, X-Large, XX-Large results are the same as predicate and meet criteria)
Physical properties - Ultimate Elongation (Before aging)500% min for all sizesExtra-Small: 565%; Small: 538%; (Medium, Large, X-Large, XX-Large results are the same as predicate and meet criteria)
Physical properties - Ultimate Elongation (After aging)400% min for all sizesExtra-Small: 514%; Small: 435%; (Medium, Large, X-Large, XX-Large results are the same as predicate and meet criteria)
ASTM D5151-19 (Detection of Holes in Medical Gloves)Holes in the glovesAQL 2.5Gloves Pass AQL 2.5
ASTM D6124-06 (Reapproved 2017) (Residual Powder on Medical Gloves)Residual powder in the gloves≤ 2 mg/gloveExtra-Small: 0.822 mg/glove; Small: 0.147 mg/glove; (Medium, Large, X-Large, XX-Large results are the same as predicate and meet criteria)
ASTM D6978-05 (Reapproved 2019) (Permeation by Chemotherapy Drugs)Breakthrough detection time of chemotherapy drugsBleomycin Sulfate (15 mg/ml) >240 Minutes; Busulfan (6 mg/ml) >240 Minutes; Carboplatin (10 mg/ml) >240 Minutes; Cisplatin (1 mg/ml) >240 Minutes; Cyclophosphamide (20 mg/ml) >240 Minutes; Cytarabine (100 mg/ml) >240 Minutes; Dacarbazine (10 mg/ml) >240 Minutes; Daunorubicin HCl (5 mg/ml) >240 Minutes; Docetaxel (10 mg/ml) >240 Minutes; Doxorubicin HCl (2 mg/ml) >240 Minutes; Epirubicin HCl (2 mg/ml) >240 Minutes; Etoposide (20 mg/ml) >240 Minutes; Fludarabine (25 mg/ml) >240 Minutes; Fluorouracil (50 mg/ml) >240 Minutes; Gemcitabine (38 mg/ml) >240 Minutes; Idarubicin HCl (1 mg/ml) >240 Minutes; Ifosfamide (50 mg/ml) >240 Minutes; Irinotecan (20 mg/ml) >240 Minutes; Mechlorethamine HCl (1 mg/ml) >240 Minutes; Melphalan (5 mg/ml) >240 Minutes; Methotrexate (25 mg/ml) >240 Minutes; Mitomycin C (0.5 mg/ml) >240 Minutes; Mitoxantrone HCl (2 mg/ml) >240 Minutes; Paclitaxel (6 mg/ml) >240 Minutes; Rituximab (10 mg/ml) >240 Minutes; Thiotepa (10 mg/ml) >240 Minutes (Note: Product warning for Carmustine and Thiotepa based on previous data indicated low permeation times); Trisenox (1 mg/ml) >240 Minutes; Vincristine Sulfate (1 mg/ml) >240 MinutesAll listed chemotherapy drugs met the acceptance criteria of >240 minutes for breakthrough detection, except Carmustine (33.8 mins) and Thiotepa (128.1 mins) for which there is a warning (results are identical to the predicate).
Breakthrough detection time of Opioid drugsFentanyl Citrate Injection (100mcg/2mL) >240 MinutesFentanyl Citrate Injection (100mcg/2mL) >240 Minutes (results are identical to the predicate)
ISO 10993-23 (Tests for Irritation)Local dermal irritation of a test article extractUnder the condition of study not an irritantUnder the conditions of the study, the test article met the requirements of the test.
ISO 10993-10 (Tests for Skin Sensitization)Skin sensitization in Guinea pigs by maximization testUnder the conditions of the study, not a sensitizerUnder the conditions of the study, the test article was not considered a sensitizer.
ISO 10993-5 (Tests for In Vitro Cytotoxicity)Potential of a test article to cause cytotoxicityUnder the conditions of the study, non-cytotoxicThe undiluted test articles extract and 50% test article extract dilution did not meet the requirements of the test and the 25%, 12.5%, 6.25%, and 3.13% test article extract dilutions met the requirements of the test. Cytotoxicity concern was addressed by acute systemic toxicity testing.
ISO 10993-11 (Tests for Systemic Toxicity)Acute systemic toxicity of a test article, extract following injectionUnder the conditions of study, the device extracts do not pose a systemic toxicity concernUnder the conditions of study, there was no mortality or evidence of systemic toxicity.

2. Sample size used for the test set and the data provenance:

The document explicitly states "The results for Medium, Large, X-Large and XX-Large are the same as the predicate," indicating that for these sizes, existing data from the previous 510(k) submission (K231349) were leveraged. New testing was conducted for the added "Extra-Small" and "Small" sizes.

  • Extra-Small and Small Sizes: Specific values are provided for these sizes, indicating new testing was performed for them. The document does not specify the exact number of gloves or samples tested for each physical property (length, width, thickness, tensile strength, ultimate elongation, residual powder, holes) for these new sizes, nor the number of replicates for the chemotherapy drug and opioid permeation tests for the subject device. However, these tests would typically follow the methodologies outlined in the referenced ASTM and ISO standards, which define appropriate sample sizes and testing protocols.
  • Medium, Large, X-Large, and XX-Large Sizes: For these sizes, the results are stated to be "the same as the predicate," which implies that the data provenance for these sizes is from the previous 510(k) submission (K231349), and likely involved testing conducted per the same ASTM and ISO standards.

The document does not provide information on the country of origin of the data or whether it was retrospective or prospective, beyond indicating that newly generated data for XS and S sizes would be prospective for this submission, while data for other sizes are effectively retrospective from the predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable to the type of device and testing performed. The "ground truth" here is defined by objective, standardized measurements against established ASTM and ISO performance criteria for medical gloves and their resistance to chemicals, not by expert interpretation of images or clinical outcomes.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable. Adjudication methods are typically used in clinical or imaging studies where subjective expert review is involved. The tests for medical gloves are objective and rely on physical measurements and chemical analysis conforming to recognized standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. MRMC studies are relevant for AI/ML-enabled diagnostic devices that rely on human-in-the-loop interpretation. The device in question is a medical glove, which is a physical barrier device, not a diagnostic tool assisted by AI/ML.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable, as it pertains to AI/ML software performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the performance of these medical gloves is based on:

  • Standardized Physical and Chemical Measurements: This includes quantitative measurements like length, width, thickness, tensile strength, ultimate elongation, residual powder, and breakthrough detection times for chemicals.
  • Compliance with Recognized Standards: The acceptance criteria themselves are derived from established industry standards such as ASTM D6319-19, ASTM D5151-19, ASTM D6124-06 (Reapproved 2017), ASTM D6978-05 (Reapproved 2019), and ISO 10993 series for biocompatibility.

8. The sample size for the training set:

This information is not applicable. The device is a physical product (medical glove), and its performance is evaluated through non-clinical bench testing against established standards, not through training and testing of an AI/ML algorithm.

9. How the ground truth for the training set was established:

This information is not applicable, as there is no "training set" in the context of this device's non-clinical evaluation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 16, 2024

Nephron Nitrile, LLC. Lou Kennedy CEO/President/Manager 4777 12th Street Extension West Columbia, South Carolina 29172

Re: K233970

Trade/Device Name: Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: December 15, 2023 Received: December 15, 2023

Dear Lou Kennedy:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233970

Device Name

Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentany()

Indications for Use (Describe)

Nephron Nitrile™. Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Bleomycin Sulfate (15.0 mg/ml) Busulfan (6.0 mg/ml) Carboplatin (10.0 mg/ml) Cisplatin (1.0 mg/ml) Cyclophosphamide (20.0 mg/ml) Cytarabine (100.0 mg/ml) Dacarbazine (10.0 mg/ml) Daunorubicin HCl (5.0 mg/ml) Docetaxel (10.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Epirubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fludarabine (25.0 mg/ml) Fluorouracil (50.0 mg/ml) Gemcitabine (38.0 mg/ml) Idarubicin HCl (1.0 mg/ml) Ifosfamide (50.0 mg/ml) Irinotecan (20.0 mg/ml) Mechlorethamine HCl (1.0 mg/ml) Melphalan (5.0 mg/ml) Methotrexate (25.0 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone HCl (2.0 mg/ml) Paclitaxel (6.0 mg/ml) Rituximab (10.0 mg/ml) Trisenox (1.0 mg/ml) Vincristine Sulfate (1.0 mg/ml) The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutes Warning: Not for Use with: Carmustine, Thiotepa The tested Opioid is: Fentanyl Citrate Injection (100 mcg/2 mL)

{4}------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

  • 区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

A. APPLICANT INFORMATION

510(K) Owner's NameNephron Nitrile, LLC.
Address4777 12th Street Extension, West Columbia, SC29172.
Phone844-937-3888
Fax1-803-926-9853
E-maillkennedy@nephronpharm.comNitrileRegulatory@nephronnitrile.com
Contact PersonLou Kennedy
DesignationChief Executive Officer
Contact Number1-803-569-3110
Contact Emaillkennedy@nephronpharm.comNitrileRegulatory@nephronnitrile.com
Preparation date28 September 2023
Date Submitted15 December 2023

B. DEVICE IDENTIFICATION

Name of the deviceNephron Nitrile™, Nitrile Powder-FreeExamination Blue Gloves (Tested For Use WithChemotherapy Drugs and Fentanyl)
510(k) NumberK233970
Product proprietary or trade nameNephron Nitrile™
Common or usual nameNitrile Examination Glove (Tested For Use WithChemotherapy Drugs and Fentanyl)
Classification nameNon-Powdered Patient Examination Glove,Specialty
Device ClassificationClass-1
Product CodeLZA, LZC, OPJ, QDO
Regulation Number21 CFR 880.6250
Review PanelGeneral Hospital

C. PREDICATE DEVICE

Predicate DeviceNephron Nitrile Powder-Free Nitrile ExaminationGloves (Tested For Use With ChemotherapyDrugs and Fentanyl)
510(k) NumberK231349
Regulatory ClassClass-1
Product codeLZA, LZC, OPJ, QDO
ManufacturerNephron Nitrile, LLC.

{6}------------------------------------------------

D. DESCRIPTION OF THE DEVICE:

Nephron Nitrile™. Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) are Class I patient examination gloves, bearing the product codes LZA, LZC, OPJ, QDO (21 CFR 880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). The powder-free gloves are made from Nitrile (NBR) latex and are blue in color. The product is non-sterile, fingertip textured, ambidextrous with beaded cuff, and single use only.

Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) include glove sizes: X-Small, Small, Medium, Large, X-Large and XX-Large.

E. INDICATION FOR USE OF THE DEVICE:

Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up toThe tested
240 minutes:Opioid is:
Bleomycin Sulfate (15.0 mg/ml)Busulfan (6.0 mg/ml)Carboplatin (10.0 mg/ml)Cisplatin (1.0 mg/ml)Cyclophosphamide (20.0 mg/ml)Cytarabine (100.0 mg/ml)Dacarbazine (10.0 mg/ml)Daunorubicin HCl (5.0 mg/ml)Docetaxel (10.0 mg/ml)Doxorubicin HCl (2.0 mg/ml)Epirubicin HCl (2.0 mg/ml)Etoposide (20.0 mg/ml)Fludarabine (25.0 mg/ml)Fluorouracil (50.0 mg/ml)Gemcitabine (38.0 mg/ml)Idarubicin HCl (1.0 mg/ml)Ifosfamide (50.0 mg/ml)Irinotecan (20.0 mg/ml)Mechlorethamine HCl (1.0 mg/ml)Melphalan (5.0 mg/ml)Methotrexate (25.0 mg/ml)Mitomycin C (0.5 mg/ml)Mitoxantrone HCl (2.0 mg/ml)Paclitaxel (6.0 mg/ml)Rituximab (10.0 mg/ml)Trisenox (1.0 mg/ml)Vincristine Sulfate (1.0 mg/ml)FentanylCitrateInjection(100 mcg/2 mL)Permeation: nobreakthroughup to 240minutes
The following chemotherapy drugs have low permeation times:Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutesWarning: Not for Use with: Carmustine, Thiotepa

{7}------------------------------------------------

F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATESUBJECT
510(K) Number---K231349K23970
Name of device---Nephron Nitrile Powder-FreeNitrile Examination Gloves(Tested For Use WithChemotherapy Drugs andFentanyl)Nephron Nitrile™, NitrilePowder-Free ExaminationBlue Gloves (Tested For UseWith Chemotherapy Drugsand Fentanyl)Similar
Product Code---LZA, LZC, OPJ, QDOLZA, LZC, OPJ, QDOIdentical
Indication for use---Nephron Nitrile Powder-FreeNitrile Examination Gloves(Tested For Use WithChemotherapy Drugs andFentanyl) is a disposabledevice intended for medicalpurpose that is worn on theexaminer's hand to preventcontamination betweenpatient and examiner. Inaddition, these gloves weretested for use withchemotherapy drugs andfentanyl in accordance withASTM D6978-05 (2019)Standard Practice forAssessment of Resistance ofMedical Gloves toPermeation byChemotherapy Drugs.Nephron Nitrile™, NitrilePowder-Free ExaminationBlue Gloves (Tested For UseWith Chemotherapy Drugsand Fentanyl) is a disposabledevice intended for medicalpurpose that is worn on theexaminer's hand to preventcontamination betweenpatient and examiner. Inaddition, these gloves weretested for use withchemotherapy drugs andfentanyl in accordance withASTM D6978-05 (2019)Standard Practice forAssessment of Resistance ofMedical Gloves toPermeation byChemotherapy Drugs.Identical
Regulation Number---21 CFR 880.625021 CFR 880.6250Identical
Material---NitrileNitrileIdentical
Color---Blue GlovesBlueIdentical
Size---M, L, XL, XXLXS, S, M, L, XL, XXLSimilar
Adding size XS and S
Single Use---Single-useSingle-useIdentical
Sterile/Non-sterile---Non-SterileNon-SterileIdentical
Rx Only or OTC---OTCOTCIdentical
CHARACTERISTICSDEVICE PERFORMANCEComparison
510(K) NumberSTANDARDSPREDICATESUBJECT
K231349---K231349K23970
Dimensions - LengthASTM D6319-19Minimum 230 mm(sizes M – XXL)Minimum 220 mm(Sizes XS-S)Minimum 230 mm(sizes M – XXL)SimilarAdding size XS and S
Dimensions - WidthASTM D6319-19M: 95±10 mmL: 110±10 mmXL: 120±10 mmXXL: 130±10 mmXS: 70±10 mmS: 80±10 mmM: 95±10 mmL: 110±10 mmXL: 120±10 mmXXL: 130±10 mmSimilarAdding size XS and S
Physical Properties-Tensile StrengthASTM D6319-19Before aging14MPa, minAfter aging14MPa, minBefore aging14MPa, minAfter aging14MPa, minIdentical
Physical Properties-Ultimate ElongationASTM D6319-19Before aging500%, minAfter aging400%, minBefore aging500%, minAfter aging400%, minIdentical
ThicknessASTM D6319-19Palm: Minimum0.05 mmFinger: Minimum0.05 mmPalm: Minimum0.05 mmFinger: Minimum0.05 mmIdentical
Powder Free ResidueASTM D6319-19≤2 mg per glove≤ 2 mg per gloveIdentical
Freedom from holesASTM D5151-2019In accordance withASTM D 5151-19,following ASTM D6319-19,G-I, AQL 2.5In accordance withASTM D 5151-19,following ASTM D6319-19,G-I, AQL 2.5Identical
Chemotherapy Drugs Tested with Minimum Breakthrough Detection TimeASTM D6978-05 (2019)Bleomycin Sulfate15 mg/ml (15,000 ppm)>240 MinutesBusulfan6 mg/ml (6,000 ppm)>240 MinutesCarboplatin10 mg/ml (10,000 ppm)>240 MinutesCarmustine3.3 mg/ml (3,300 ppm)33.8 MinutesCisplatin1 mg/ml (1,000 ppm)>240 MinutesCyclophosphamide20 mg/ml (20,000 ppm)>240 MinutesBleomycin Sulfate15 mg/ml (15,000 ppm)>240 MinutesBusulfan6 mg/ml (6,000 ppm)>240 MinutesCarboplatin10 mg/ml (10,000 ppm)>240 MinutesCarmustine3.3 mg/ml (3,300 ppm)33.8 MinutesCisplatin1 mg/ml (1,000 ppm)>240 MinutesCyclophosphamide20 mg/ml (20,000 ppm)>240 MinutesIdentical
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCEComparison
510(K) Number---PREDICATESUBJECT
K231349K23970
Cytarabine100 mg/ml (100,000 ppm)>240 MinutesCytarabine100 mg/ml (100,000 ppm)>240 MinutesIdentical
Dacarbazine10 mg/ml (10,000 ppm)>240 MinutesDacarbazine10 mg/ml (10,000 ppm)>240 MinutesIdentical
Daunorubicin HCl5 mg/ml (5,000 ppm)>240 MinutesDaunorubicin HCl5 mg/ml (5,000 ppm)>240 MinutesIdentical
Docetaxel10 mg/ml (10,000 ppm)>240 MinutesDocetaxel10 mg/ml (10,000 ppm)>240 MinutesIdentical
Doxorubicin HCl2 mg/ml (2,000 ppm)>240 MinutesDoxorubicin HCl2 mg/ml (2,000 ppm)>240 MinutesIdentical
Epirubicin HCl2 mg/ml (2,000 ppm)>240 MinutesEpirubicin HCl2 mg/ml (2,000 ppm)>240 MinutesIdentical
Etoposide20 mg/ml (20,000 ppm)>240 MinutesEtoposide20 mg/ml (20,000 ppm)>240 MinutesIdentical
Fludarabine25 mg/ml (25,000 ppm)>240 MinutesFludarabine25 mg/ml (25,000 ppm)>240 MinutesIdentical
Fluorouracil50 mg/ml (50,000 ppm)>240 MinutesFluorouracil50 mg/ml (50,000 ppm)>240 MinutesIdentical
Gemcitabine38 mg/ml (38,000 ppm)>240 MinutesGemcitabine38 mg/ml (38,000 ppm)>240 MinutesIdentical
Idarubicin HCl1 mg/ml (1,000 ppm)>240 MinutesIdarubicin HCl1 mg/ml (1,000 ppm)>240 MinutesIdentical
Ifosfamide50 mg/ml (50,000 ppm)>240 MinutesIfosfamide50 mg/ml (50,000 ppm)>240 MinutesIdentical
Irinotecan20 mg/ml (20,000 ppm)>240 MinutesIrinotecan20 mg/ml (20,000 ppm)>240 MinutesIdentical
Mechlorethamine HCl1 mg/ml (1,000 ppm)>240 MinutesMechlorethamine HCl1 mg/ml (1,000 ppm)>240 MinutesIdentical
Melphalan5 mg/ml (5,000 ppm)>240 MinutesMelphalan5 mg/ml (5,000 ppm)>240 MinutesIdentical
Methotrexate25 mg/ml (25,000 ppm)>240 MinutesMethotrexate25 mg/ml (25,000 ppm)>240 MinutesIdentical
Mitomycin C0.5 mg/ml (500 ppm)>240 MinutesMitomycin C0.5 mg/ml (500 ppm)>240 MinutesIdentical
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATESUBJECT
510(K) NumberK231349K23970
Mitoxantrone HCl2 mg/ml (2,000 ppm)>240 MinutesMitoxantrone HCl2 mg/ml (2,000 ppm)>240 MinutesIdentical
Paclitaxel6 mg/ml (6,000 ppm)>240 MinutesPaclitaxel6 mg/ml (6,000 ppm)>240 MinutesIdentical
Rituximab 10 mg/ml(10,000 ppm)>240 MinutesRituximab 10 mg/ml(10,000 ppm)>240 MinutesIdentical
Thiotepa10 mg/ml (10,000 ppm)128.1 MinutesThiotepa10 mg/ml (10,000 ppm)128.1 MinutesIdentical
Trisenox1 mg/ml (1,000ppm) >240MinutesTrisenox1 mg/ml (1,000ppm)>240 MinutesIdentical
Vincristine Sulfate1 mg/ml (1,000 ppm)>240 MinutesVincristine Sulfate1 mg/ml (1,000 ppm)>240 MinutesIdentical
Opioid DrugsTested with MinimumBreakthrough DetectionTimeASTM D6978-05 (2019)Fentanyl Citrate Injection(100mcg/2mL)>240 MinutesFentanyl Citrate Injection(100mcg/2mL)>240 MinutesIdentical
BiocompatibilityPrimary SkinIrritation- ISO10993-23:First Edition2021-01Under the conditions of thestudy, the test articlemet the requirements ofthe test.Under the conditions of thestudy, the test articlemet the requirements ofthe test.Identical
DermalSensitization-ISO 10993-10:Fourth Edition2021-11Under the conditions of thestudy, the test article was notconsidered a sensitizer.Under the conditions of thestudy, the test article was notconsidered a sensitizer.Identical
In vitrocytotoxicity-ISO 10993-5:Third Edition2009-06-01Under the Conditions of thestudy, theundiluted test article extractand 50% test article extractdilution didnot meet the requirements ofthe test and the 25%, 12.5%,6.25%, and3.13% test article extractdilutions metthe requirements of the test.Under the conditions of thestudy, theundiluted test articleextract and 50% test articleextract dilution didnot meet the requirements ofthe test and the 25%,12.5%, 6.25%, and3.13% test article extractdilutions metthe requirements of the test.Identical
AcuteSystemicToxicity- ISO10993-11:Third Edition2017-09Under the conditions of thestudy, there was no mortalityor evidence of systemictoxicityUnder the conditions of thestudy, there was no mortalityor evidence of systemictoxicityIdentical

{8}------------------------------------------------

AS REQUIRED BY: 21CFR§807.92

{9}------------------------------------------------

AS REQUIRED BY: 21CFR§807.92

{10}------------------------------------------------

AS REQUIRED BY: 21CFR§807.92

  • Identical except for the expansion of glove sizes to include XS and S while maintaining existent M to XXL, which is the subject of this submission.

{11}------------------------------------------------

AS REQUIRED BY: 21CFR§807.92

There are no significant differences between the products and are identical in terms of intended use, materials, design and manufacturing methods. The devices meet the ASTM standard D6319-19 and D6978-05 (2019).

G. NON-CLINICAL TESTING SUMMARY

TEST METHODPURPOSEACCEPTANCE CRITERIARESULT
ASTM D6319-19Standard Specification forNitrile ExaminationGloves for MedicalApplication.To determine thelength of the glovesExtra-Small: 220 mm minSmall: 220 mm minMedium: 230 mm minLarge : 230 mm minX-Large : 230 mm minXX-Large : 230 mm minExtra-Small: 242 mmSmall: 243 mmMedium: 235 mm*Large: 237 mm*X-Large: 250 mm*XX-Large: 238 mm*
ASTM D6319-19Standard Specification forNitrile ExaminationGloves for MedicalApplication.To determine thewidth of the glovesXS: 70±10 mmS: 80±10 mmM: 95±10 mmL: 110±10 mmXL: 120±10 mmXXL: 130±10 mmExtra-Small: 74 mmSmall: 85 mmMedium: 95 mm*Large: 113 mm*X-Large: 121 mm*XX-Large: 129 mm*
ASTM D6319-19Standard Specification forNitrile ExaminationGloves for MedicalApplication.To determine thethickness of theglovesPalm: 0.05 mm minfor all sizesFinger: 0.05 mm minfor all sizesSize Palm Finger Extra-Small 0.112 mm 0.145 mm Small 0.095 mm 0.115 mm Medium* 0.077 mm 0.111 mm Large* 0.106 mm 0.109 mm X-Large* 0.089 mm 0.115 mm XX-Large* 0.113 mm 0.107 mm
ASTM D6319-19Standard Specification forNitrile ExaminationGloves for MedicalApplication.To determine thephysical properties-Tensile strengthBefore AgeingTensile Strength14MPa min for all sizesAfter AgeingTensile Strength14MPa min for all sizesSize Beforeageing Afterageing Extra-Small 33.74MPa 35.99MPa Small 32.35MPa 32.14MPa The results for Medium, Large, X-Large and XX-Large are the same asthe predicate.
To determine thephysical properties-Ultimate ElongationBefore AgeingUltimate Elongation500% min for all sizesAfter AgeingUltimate Elongation400% min for all sizesSize Beforeageing Afterageing Extra-Small 565% 514% Small 538% 435% The results for Medium, Large, X-Large and XX-Large are the same asthe predicate.

BENCH TEST DATA

{12}------------------------------------------------

ASTM D5151-19Standard Test Method forDetection of Holes inMedical GlovesTo determine theholes in the glovesAQL 2.5Gloves Passes AQL 2.5
ASTM D6124-06(Reapproved 2017)Standard Test Method forResidual Powder onMedical GlovesTo determine theresidual powder inthe gloves$\u2264$ 2 mg/gloveExtra-Small: 0.822 mg/gloveSmall: 0.147 mg/gloveThe results for Medium, Large, X-Large and XX-Large are the same asthe predicate.
TEST METHODPURPOSEACCEPTANCE CRITERIARESULT
ASTM D6978-05(Reapproved 2019)Standard Practice forAssessment ofResistance ofMedical Gloves toPermeation byChemotherapyDrugs.To determine thebreakthroughdetection time ofchemotherapydrugsBleomycin Sulfate 15mg/ml(15,000 ppm) >240 MinutesBleomycin Sulfate 15 mg/ml(15,000 ppm) >240 Minutes
Busulfan 6 mg/ml(6,000 ppm) >240 MinutesBusulfan 6 mg/ml(6,000 ppm) >240 Minutes
Carboplatin 10 mg/ml(10,000 ppm)>240 MinutesCarboplatin 10 mg/ml(10,000 ppm)>240 Minutes
Carmustine 3.3 mg/ml(3,300 ppm) >240 MinutesCarmustine 3.3 mg/ml(3,300 ppm) >240 Minutes
Cisplatin 1 mg/ml(1,000 ppm) >240 MinutesCisplatin 1 mg/ml(1,000 ppm) >240 Minutes
Cyclophosphamide 20 mg/ml(20,000 ppm) >240 MinutesCyclophosphamide 20 mg/ml(20,000 ppm) >240 Minutes
Cytarabine 100 mg/ml(100,000 ppm) >240 MinutesCytarabine 100 mg/ml(100,000 ppm) >240 Minutes
Dacarbazine 10 mg/ml(10,000 ppm)>240 MinutesDacarbazine 10 mg/ml(10,000 ppm) >240 Minutes
Daunorubicin HCl 5 mg/ml(5,000 ppm) >240 MinutesDaunorubicin HCl 5 mg/ml(5,000 ppm) >240 Minutes
Docetaxel 10 mg/ml(10,000 ppm)>240 MinutesDocetaxel 10 mg/ml(10,000 ppm) >240 Minutes
Doxorubicin HCl 2 mg/ml(2,000 ppm) >240 MinutesDoxorubicin HCl 2 mg/ml(2,000 ppm) >240 Minutes
Epirubicin HCl 2 mg/ml(2,000 ppm) >240 MinutesEpirubicin HCl 2 mg/ml(2,000 ppm) >240 Minutes
Etoposide 20 mg/ml(20,000 ppm) >240 MinutesEtoposide 20 mg/ml(20,000 ppm) >240 Minutes
Fludarabine 25 mg/ml(25,000 ppm) >240 MinutesFludarabine 25 mg/ml(25,000 ppm) >240 Minutes
Fluorouracil 50 mg/ml(50,000 ppm) >240 MinutesFluorouracil 50 mg/ml(50,000 ppm) >240 Minutes
ASTM D6978-05(Reapproved 2019)Standard Practice forAssessment ofResistance ofMedical Gloves toPermeation byChemotherapyDrugsTo determine thebreakthroughdetection time ofOpioiddrugsGemcitabine 38 mg/ml(38,000 ppm) >240 MinutesIdarubicin HCl 1 mg/ml(1,000 ppm) >240 MinutesIfosfamide 50 mg/ml(50,000 ppm) >240 MinutesIrinotecan 20 mg/ml(20,000 ppm) >240 MinutesMechlorethamine HCl 1 mg/ml(1,000 ppm) >240 MinutesMelphalan 5 mg/ml(5,000 ppm) >240 MinutesMethotrexate 25 mg/ml(25,000 ppm) >240 MinutesMitomycin C 0.5 mg/ml(500 ppm) >240 MinutesMitoxantrone HCl 2 mg/ml(2,000 ppm) >240 MinutesPaclitaxel 6 mg/ml(6,000 ppm) >240 MinutesRituximab 10 mg/ml(10,000 ppm) >240 MinutesThiotepa 10 mg/ml(10,000 ppm) >240 MinutesTrisenox 1 mg/ml(1,000 ppm) >240 MinutesVincristine Sulfate 1 mg/ml(1,000 ppm) >240 MinutesFentanyl Citrate Injection(100mcg/2mL) >240 MinutesGemcitabine 38 mg/ml(38,000 ppm) >240 MinutesIdarubicin HCl 1 mg/ml(1,000 ppm) >240 MinutesIfosfamide 50 mg/ml(50,000 ppm) >240 MinutesIrinotecan 20 mg/ml(20,000 ppm) >240 MinutesMechlorethamine HCl 1 mg/ml(1,000 ppm) >240 MinutesMelphalan 5 mg/ml(5,000 ppm) >240 MinutesMethotrexate 25 mg/ml(25,000 ppm) >240 MinutesMitomycin C 0.5 mg/ml(500 ppm) >240 MinutesMitoxantrone HCl 2 mg/ml(2,000 ppm) >240 MinutesPaclitaxel 6 mg/ml(6,000 ppm) >240 MinutesRituximab 10 mg/ml(10,000 ppm) >240 MinutesThiotepa 10 mg/ml(10,000 ppm) >240 MinutesTrisenox 1 mg/ml(1,000 ppm) >240 MinutesVincristine Sulfate 1 mg/ml(1,000 ppm) >240 MinutesFentanyl Citrate Injection(100mcg/2mL) >240 Minutes

{13}------------------------------------------------

AS REQUIRED BY: 21CFR§807.92

{14}------------------------------------------------

BIOCOMPATIBILITY DATA

TEST METHODPURPOSEACCEPTANCECRITERIARESULT
ISO 10993-23 First edition2021-01 BiologicalEvaluation of MedicalDevices - Part 23, Tests forIrritation.To evaluate the localdermal irritation of a testarticle extract followingintracutaneous injection inrabbits.Under the condition ofstudy not an irritantUnder the conditions of thestudy, the test article met therequirements of the test
10993-10 Fourth edition2021-11 BiologicalEvaluation of MedicalDevices - Part 10, Tests forSkin Sensitization.To evaluate the test item,for the skin sensitizationin Guinea pigs bymaximization test.Under the conditions ofthe study, not asensitizerUnder the conditions of thestudy, the test article was notconsidered a sensitizer
ISO 10993-5 Third Edition2009-06-01 BiologicalEvaluation of MedicalDevices - Part 5, Tests forIn Vitro Cytotoxicity.To determine the potentialof a test article to causecytotoxicityUnder the conditions ofthe study, non-cytotoxicThe undiluted test articlesextract and 50% test articleextract dilution did not meetthe requirements of the testand the 25%, 12.5%, 6.25%,and 3.13% test article extractdilutions met therequirements of the test.Cytotoxicity concern wasaddressed by acute systemictoxicity testing.
ISO 10993-11 Third edition2017-09 BiologicalEvaluation of MedicalDevices - Part 11, Tests forSystemic Toxicity.To evaluate the acutesystemic toxicity of a testarticle, extract followinginjection in mice.Under the conditions ofstudy, the device extractsdo not pose a systemictoxicity concernUnder the conditions of study,there was no mortality orevidence of systemic toxicity

The following non-clinical tests were performed:

  • ASTM D6319-19 Standard Specification for Nitrile examination Gloves for Medical Application.
  • ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves. ●
  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical ● Gloves.
  • ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of ● Medical Gloves to Permeation by Chemotherapy Drugs.
  • ISO 10993-23 First Edition 2021-01 Biological Evaluation of Medical Devices - Part 23, Tests for Irritation.
  • . ISO 10993-10 Fourth Edition 2021-11 Biological Evaluation of Medical Devices - Part 10, Tests for Skin Sensitization.
  • ISO 10993-5 Third Edition 2009-06-01 Biological Evaluation of Medical Devices Part 5, ● Tests for In Vitro Cytotoxicity.
  • ISO 10993-11 Third Edition 2017-09 Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity.

{15}------------------------------------------------

H. CLINICAL TESTING SUMMARY

Not applicable - Clinical data is not needed for gloves.

I. CONCLUSION

The conclusions drawn from the non-clinical test demonstrate that the subject device in 510(K) submission K233970, Nephron Nitrile™, Nitrile Powder-Free Examination Blue Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl), is as safe, as effective, and performs as well as or better than the legally marketed predicate device K231349.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.