K Number
K233790

Validate with FDA (Live)

Device Name
ACL TOP 970 CL
Date Cleared
2023-12-29

(31 days)

Product Code
Regulation Number
864.5425
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACL TOP 970 CL is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic use by health care professionals in a clinical laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis.

The system provides results for both direct measurements and calculated parameters.

Device Description

The ACL TOP 970 CL is an additional member of the ACL TOP Family 70 Series previously FDA cleared under K231031. This family member consists of two side-by-side test modules:

  • . Main Module (ACL TOP 550 CTS, K150877) the subject of this submission
  • Chemiluminescent (CL) Module previously FDA cleared under K221359 .

The Main Module to the ACL TOP 970 CL instrument performs the following types of tests, using the same optical measuring wavelengths and test parameters as the predicate (ACL TOP Family 50 Series):

  • . Coagulometric (Turbidimetric) Measurements
  • . Chromogenic (Absorbance) Measurements
  • . Immunological Measurements

The ACL TOP 970 CL is an additional member of the ACL TOP Family 70 Series (K231031) and utilizes the same consumables, reagents, calibrators, and controls, and provides the same analytical methodology for routine and specialty assay result reporting as the predicate (ACL TOP Family 50 Series).

The ACL TOP 970 CL also offers the same pre-analytical features available on the ACL TOP Family 50 Series. These features alert the instrument operator to a potential HIL (Hemoglobin, Icteric and Lipemia) interference situation specific to the assays requested for a sample, underfilled sample tubes or a detected clog.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "ACL TOP 970 CL" device. This device is a Multipurpose System For In Vitro Coagulation Studies. Based on the content, it does not appear to be an AI/ML-driven device that would involve the complex ground truthing, expert reads, MRMC studies, or training/test set definitions typically associated with such technologies.

Instead, this submission is for a new hardware configuration (the ACL TOP 970 CL Main Module) that is substantially equivalent to a previously cleared device (ACL TOP Family 50 Series, K150877). The "studies" mentioned are analytical studies (precision and method comparison) to demonstrate that the new configuration performs equivalently to the predicate device for various coagulation assays.

Therefore, many of the requested points related to AI/ML device studies (e.g., number of experts, adjudication methods, MRMC studies, training set details) are not applicable to this type of device submission and are not found in the provided text.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a quantified, pre-defined table format for each test. Instead, it refers to industry-standard guidelines (CLSI EP05-A3, CLSI EP09c, 3rd Ed) and states that "all analytical studies were performed in accordance to established plans and protocols and design control procedures. Testing verified that all acceptance criteria were met and results equivalent to the predicate device."

However, we can infer the performance metrics from the results presented:

Performance MetricAcceptance Criteria (Inferred - based on "results equivalent to the predicate device" and meeting CLSI guidelines)Reported Device Performance (ACL TOP 970 CL Main Module)
PrecisionMeeting CLSI EP05-A3 guidelines for within-run and total %CV and comparability to predicate device performance.(See "Precision" tables below for specific values per assay and material. All deemed acceptable.)
Method ComparisonDemonstrated equivalence (slope near 1, intercept near 0, high correlation 'r') when compared to predicate device (ACL TOP 550 CTS) across the analytical measuring range. Meeting CLSI EP09c, 3rd Ed guidelines.(See "Method Comparison" tables below for specific values per assay. All deemed acceptable.)
Thermal VerificationNo impact on analytical results from structural changes.Confirmed no impact.
Optical Stray Light VerificationNo impact on analytical results from new back wall design.Confirmed no impact.
Environmental VerificationNo impact on analytical results from changes to skin/air intake.Confirmed no impact.

Reported Device Performance Tables (from the document):

HemosIL D-Dimer HS 500 (K172903) – D-dimer ng/mL FEU - Precision

MaterialMeanWithin Run %CVTotal %CV
Low Control7334.34.5
High Control26642.52.8
Cut-off Plasma Pool5325.26.0
High Plasma Pool24352.42.4

HemosIL Factor VIII deficient plasma (K034007) – Factor VIII % Activity - Precision

MaterialMeanWithin Run %CVTotal %CV
Normal Control93.33.74.6
Abnormal Control26.53.76.5
Plasma Pool 141.56.37.3
Plasma Pool 25.84.45.4

HemosIL RecombiPlasTin 2G (K070005) – Prothrombin Time Seconds - Precision

MaterialMeanWithin Run %CVTotal %CV
Normal Control11.50.51.2
Abnormal Pool26.30.82.3
Low Abn Control22.91.62.1
High Abn Control38.71.12.6

HemosIL RecombiPlasTin 2G (K070005) – Fibrinogen mg/dL - Precision

MaterialMeanWithin Run %CVTotal %CV
Normal Control3870.91.4
Low Fibrinogen Control1786.16.3
Normal Pool3921.32.0
Abnormal Pool1091.82.4

HemosIL Liquid Anti-Xa (K213464) – Heparin IU/mL - Precision

MaterialMeanWithin Run %CVTotal %CV
UF Low Control0.351.822.81
UF High Control0.651.432.36
UF Pool0.551.692.27
LMW High Control1.571.182.15
LMW Low Control0.642.552.81
LMW Pool0.711.492.05

Method Comparison Results (ACL TOP 970 CL vs. ACL TOP 550 CTS):

HemosIL D-Dimer HS 500 (K172903) – D-dimer ng/mL FEU
N: 136, Slope: 0.939, Intercept: 27.0, r: 0.996

HemosIL Factor VIII deficient plasma (K034007) – Factor VIII % Activity
N: 105, Slope: 1.045, Intercept: 0.0, r: 0.993

HemosIL RecombiPlasTin 2G (K070005) – Prothrombin Time Seconds
N: 118, Slope: 1.000, Intercept: 0.25, r: 0.998

HemosIL RecombiPlasTin 2G (K070005) – Fibrinogen mg/dL
N: 123, Slope: 0.991, Intercept: 5.1, r: 0.998

HemosIL Liquid Anti-Xa (K213464) – Heparin IU/mL
N: 139, Slope: 0.989, Intercept: 0.015, r: 0.997

2. Sample size used for the test set and the data provenance

  • Precision Test Set Sample Size: For precision studies, samples for each material were run for 20 days, two runs per day, 2 replicates per run (n=80). This applies to each of the multiple materials tested for each assay (e.g., Low Control, High Control, etc.).

  • Method Comparison Test Set Sample Size:

    • HemosIL D-Dimer HS 500: N=136 clinical samples
    • HemosIL Factor VIII deficient plasma: N=105 clinical samples
    • HemosIL RecombiPlasTin 2G (Prothrombin Time): N=118 clinical samples
    • HemosIL RecombiPlasTin 2G (Fibrinogen): N=123 clinical samples
    • HemosIL Liquid Anti-Xa: N=139 clinical samples
  • Data Provenance: The document does not specify the country of origin for the data or explicitly state whether the samples were retrospective or prospective. It mentions "clinical samples" for method comparison and "material" (controls/plasma pools) for precision. Typically, such studies for IVD devices are conducted in a controlled laboratory setting (prospective testing) using a mix of manufactured controls/calibrators and patient samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

N/A. This is not an AI/ML device requiring expert interpretation for ground truth. The "ground truth" for this in-vitro diagnostic device is the actual measurement of analytes, established by reference methods or validated predicate devices. Proficiency of technical staff operating the instruments would be presumed as per standard laboratory practices.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

N/A. Not applicable to a measurement device; no human interpretation or adjudication beyond standard laboratory quality control and data review.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

N/A. This is not an AI/ML device that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

N/A. This is a standalone instrument for in-vitro diagnostic testing, not an algorithm. Its performance is based on its ability to accurately and precisely measure analytes. The "performance" tables provided are essentially the standalone performance of the device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The ground truth for an in-vitro diagnostic coagulation system like this is based on:

  • Reference Materials: For precision, known concentration control materials and plasma pools with established values are used.
  • Comparative Measurements: For method comparison, results from the subject device are compared against a legally marketed predicate device (ACL TOP 550 CTS) which serves as the established reference. The assumption is that the predicate device's measurements are equivalent to the "ground truth" for the test.

8. The sample size for the training set

N/A. This is not an AI/ML device that requires a "training set" in the machine learning sense. The device is a hardware instrument with validated analytical capabilities.

9. How the ground truth for the training set was established

N/A. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 29, 2023

Instrumentation Laboratory Nikita Malladi Regulatory Affairs Manager I 180 Hartwell Road Bedford, Massachusetts 01730

Re: K233790

Trade/Device Name: ACL TOP 970 CL Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose System For In Vitro Coagulation Studies Regulatory Class: Class II Product Code: JPA Dated: November 28, 2023 Received: November 28, 2023

Dear Nikita Malladi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min W

Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233790

Device Name ACL TOP 970 CL

Indications for Use (Describe)

The ACL TOP 970 CL is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic use by health care professionals in a clinical laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis.

The system provides results for both direct measurements and calculated parameters.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "werfen" in a sans-serif font. The word is written in a dark blue color. The background is white, which makes the word stand out.

510(k) Summary

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.

Submitter's InformationInstrumentation Laboratory Company180 Hartwell RoadBedford, MA 01730-2443 (USA)
------------------------------------------------------------------------------------------------------------------
Contact PersonNikita MalladiRegulatory Affairs Manager IPhone: 781-353-1486Fax: 781-861-4207Email: nmalladi@werfen.com
------------------------------------------------------------------------------------------------------------------------------------------
Preparation DateDecember 29, 2023
-------------------------------------
Device Trade NameACL TOP 970 CL
-----------------------------------
Predicate DeviceACL TOP Family 50 SeriesK150877
-----------------------------------------------------
Regulatory InformationRegulation Section No.21 CFR 864.5425
Regulation DescriptionMultipurpose system for in vitro coagulation studies
ClassificationClass II
Product CodeJPA
PanelHematology (81)

{4}------------------------------------------------

Device Description

The ACL TOP 970 CL is an additional member of the ACL TOP Family 70 Series previously FDA cleared under K231031. This family member consists of two side-by-side test modules:

  • . Main Module (ACL TOP 550 CTS, K150877) the subject of this submission
  • Chemiluminescent (CL) Module previously FDA cleared under K221359 .

The Main Module to the ACL TOP 970 CL instrument performs the following types of tests, using the same optical measuring wavelengths and test parameters as the predicate (ACL TOP Family 50 Series):

  • . Coagulometric (Turbidimetric) Measurements
  • . Chromogenic (Absorbance) Measurements
  • . Immunological Measurements

The ACL TOP 970 CL is an additional member of the ACL TOP Family 70 Series (K231031) and utilizes the same consumables, reagents, calibrators, and controls, and provides the same analytical methodology for routine and specialty assay result reporting as the predicate (ACL TOP Family 50 Series).

The ACL TOP 970 CL also offers the same pre-analytical features available on the ACL TOP Family 50 Series. These features alert the instrument operator to a potential HIL (Hemoglobin, Icteric and Lipemia) interference situation specific to the assays requested for a sample, underfilled sample tubes or a detected clog.

Intended Use / Indications for Use

The ACL TOP 970 CL is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic use by health care professionals in a clinical laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis.

The system provides results for both direct measurements and calculated parameters.

Reason for Special 510(k) Submission

This Special 510(k) is submitted for the following reasons:

  • . To obtain FDA clearance for the Main Module (ACL TOP 550 CTS side) of the ACL TOP 970 CL instrument
  • Following FDA clearance, to extend the current FDA cleared on-market hemostasis ● (non-chemiluminescent) assay menu for the ACL TOP Family 50 Series onto the ACL TOP 970 CL instrument through CLIA Categorization Requests

{5}------------------------------------------------

Summary Comparison of Technological Characteristics (Predicate)
ItemPredicate Device (K150877)Subject Device
Trade NameACL TOP Family 50 SeriesACL TOP 970 CL(Main Module: ACL TOP 550 CTS)
Similarities
ManufacturerInstrumentation Laboratory Co.Same
Indications for Use /Intended UseThe ACL TOP Family 50 Series (ACL TOP750; ACL TOP 750 CTS; ACL TOP 750 LAS;ACL TOP 550 CTS; ACL TOP 350 CTS) arebench top, fully automated, random accessanalyzers designed specifically for in vitrodiagnostic clinical use in the hemostasislaboratory for coagulation and/or fibrinolysistesting in the assessment of thrombosis and/orhemostasis.The systems provide results for both directhemostasis measurements and calculatedparameters.The ACL TOP 970 CL is a bench top, fullyautomated, random access analyzerdesigned specifically for in vitro diagnosticuse by health care professionals in a clinicallaboratory for coagulation and/orfibrinolysis testing in the assessment ofthrombosis and/or hemostasis.The system provides results for both directmeasurements and calculated parameters.
Test MethodologyCoagulometric measurement Chromogenic measurement Immunological measurementSame*
Wavelengths405 nm 535 nm 671 nmSame
Test MenuClotting Assays Chromogenic Assays Immunological AssaysSame
Test ParametersAssay volumes, rinse and clean cycles, timing,optical parameters, data algorithms, materialdefinitionSame
Reagents, Controls andCalibratorsPackaging, formulation, performance claims inlabelingSame
Summary Comparison of Technological Characteristics (Predicate) (Cont.)
ItemPredicate Device (K150877)Subject Device
Trade NameACL TOP Family 50 SeriesACL TOP 970 CL(Main Module; ACL TOP 550 CTS)
Similarities (Cont.)
Fluidic HandlingAspiration, dispense, mixing, rinse, clean,temperature control, bulk fluidsSame
Sample HandlingCap piercing, onboard storageSame
Onboard Reagent StorageStirring, temperature controlSame
Reaction and DetectionOptics, temperature controlSame
System SoftwareHardware control, user interface except asnoted in the Differences section belowSame
Quality ControlAutomated QCSame
Pre-Analytical HIL CheckStandard for all models:Third measurement wavelength @535 nmand an additional emitter control channelSame
Tube Fill Height CheckStandard for all modelsSame
Clog DetectionStandard for all modelsSame

*Note: Chemiluminescent measurement on the ACL TOP 970 CL previously FDA cleared under K221359.

{6}------------------------------------------------

ItemPredicate Device (K150877)Subject Device
Differences
ChemiluminescenceMeasurementNot ApplicableChemiluminescent (CL) Modulepreviously FDA cleared under K221359

{7}------------------------------------------------

Performance Summary

Analytical studies (specifically internal precision and method comparison) were performed on one lot of selected representative assays to establish that the labeled performance data of the current menu of FDAcleared (non-chemiluminescent) assays claimed for the ACL TOP Family 50 Series are applicable to the Main Module ACL TOP 970 CL. These studies followed recognized guidelines:

  • CLSI EP05-A3
  • . CLSI EP09c, 3rd Ed

All analytical studies were performed in accordance to established plans and protocols and design control procedures. Testing verified that all acceptance criteria were met and results equivalent to the predicate device.

Precision

Precision studies were performed using one lot of the following selected representative assays tested on an ACL TOP 970 CL instrument. These studies used samples with each material run for 20 days at two runs per day, 2 replicates per run (n=80).

Summary results are shown in the table below:

HemosIL D-Dimer HS 500 (K172903) – D-dimer ng/mL FEU

MaterialMeanWithin Run %CVTotal %CV
Low Control7334.34.5
High Control26642.52.8
Cut-off Plasma Pool5325.26.0
High Plasma Pool24352.42.4

HemosIL Factor VIII deficient plasma (K034007) – Factor VIII % Activity

MaterialMeanWithin Run %CVTotal %CV
Normal Control93.33.74.6
Abnormal Control26.53.76.5
Plasma Pool 141.56.37.3
Plasma Pool 25.84.45.4
Precision
HemosIL RecombiPlasTin 2G (K070005) – Prothrombin Time Seconds
MaterialMeanWithin Run %CVTotal %CV
Normal Control11.50.51.2
Abnormal Pool26.30.82.3
Low Abn Control22.91.62.1
High Abn Control38.71.12.6
HemosIL RecombiPlasTin 2G (K070005) – Fibrinogen mg/dL
MaterialMeanWithin Run %CVTotal %CV
Normal Control3870.91.4
Low Fibrinogen Control1786.16.3
Normal Pool3921.32.0
Abnormal Pool1091.82.4
HemosIL Liquid Anti-Xa (K213464) – Heparin IU/mL
MaterialMeanWithin Run %CVTotal %CV
UF Low Control0.351.822.81
UF High Control0.651.432.36
UF Pool0.551.692.27
LMW High Control1.571.182.15
LMW Low Control0.642.552.81
LMW Pool0.711.492.05

{8}------------------------------------------------

{9}------------------------------------------------

Method Comparison

Method comparison studies were performed using one lot of the following selected representative assays to compare the performance on the Main Module side of ACL TOP 970 CL (subject device) to the ACL TOP 550 CTS (predicate device). The studies included clinical samples spanning each assay's analytical measuring range (AMR) to demonstrate the equivalent performance between the predicate and subject devices.

Summary results for are shown in the table below:

HemosIL D-Dimer HS 500 (K172903) – D-dimer ng/mL FEU
Subject SystemNSlopeInterceptrPredicate System
ACL TOP 970 CL1360.93927.00.996ACL TOP 550 CTS
HemosIL Factor VIII deficient plasma (K034007) – Factor VIII % Activity
Subject SystemNSlopeInterceptrPredicate System
ACL TOP 970 CL1051.0450.00.993ACL TOP 550 CTS
HemosIL RecombiPlasTin 2G (K070005) – Prothrombin Time Seconds
Subject SystemNSlopeInterceptrPredicate System
ACL TOP 970 CL1181.0000.250.998ACL TOP 550 CTS
HemosIL RecombiPlasTin 2G (K070005) – Fibrinogen mg/dL
Subject SystemNSlopeInterceptrPredicate System
ACL TOP 970 CL1230.9915.10.998ACL TOP 550 CTS
HemosIL Liquid Anti-Xa (K213464) – Heparin IU/mL
Subject SystemNSlopeInterceptrPredicate System
ACL TOP 970 CL1390.9890.0150.997ACL TOP 550 CTS

{10}------------------------------------------------

Main System Verification for HemosIL CL On-Market Reagents

A risk review of the Main System on the ACL TOP 970 CL, which has only minimal differences in hardware and software from the ACL TOP Family 50 Series to support the new Chemiluminescent (CL) System, was performed to identify the necessary verification testing. The following testing supports that there is no analytical impact on the existing HemosIL (non-chemiluminescent) test menu. Therefore, the current FDA-cleared assays can be extended onto the ACL TOP 970 CL without further testing.

ThermalVerificationVerification testing was performed that confirmed the structural changes to the instrument (new skins, modified wall geometry) does not impact analytical results.
Optical StrayLight VerificationVerification testing was performed that a new back wall design does not impact analytical results.
EnvironmentalVerificationVerification testing was performed that a change to the skin of the instrument and air intake modification does not impact analytical results.

Conclusion

Based on the substantial equivalence comparison and the results of the verification testing that was conducted, it was concluded that the indications for use and technological characteristics of the Main Module on the ACL TOP 970 CL are substantially equivalent to the cleared and currently marketed predicate device, ACL TOP Family 50 Series (K150877), with shared performance claims for the FDA cleared on-market hemostasis (non-chemiluminescent) assay menu. The differences between the subject and predicate devices do not impact safety and effectiveness.

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.