K Number
K233587

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-06-03

(208 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.

Device Description

The LOCATOR Angled Abutment is designed to be used with LOCATOR FIXED and LOCATOR Attachment Systems for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants in the mandible or maxilla. The LOCATOR Angled Abutment consists of various height abutment bodies with an integrated abutment screw. The attachment features are identical compared to LOCATOR Abutments of the LOCATOR High Retention Attachment System (LOCATOR FIXED), cleared in K213391. The LOCATOR Angled Abutment is compatible with Straumann BLX Implant System cleared in K173961. The LOCATOR Angled Abutment will be used with the accessories of the LOCATOR Implant Attachment System (retention inserts, denture attachment housing, and ancillary processing parts) and LOCATOR FIXED Attachment System (fixed inserts, denture attachment housing) for the attachment of a restoration. The LOCATOR Angled Abutment uses identical attachment features as the LOCATOR Abutments, but instead of the attachment features being aligned coaxially, the interfacing features are provided at a 15 degree angle to allow for angle correction, substantially equivalent to the device of K190040. The LOCATOR Angled Abutments and integrated abutment screw are manufactured from titanium (Ti-6Al-4V). The LOCATOR Angled Abutment body is titanium nitride (TiN) coated, identical to LOCATOR Abutments.

AI/ML Overview

The provided FDA 510(k) summary for the "LOCATOR Angled Abutment" does not describe a study involving an AI/Machine Learning (ML) device or any specific performance metrics for such a device like sensitivity, specificity, or AUC, or a comparative effectiveness study involving human readers.

This submission is for a medical device (dental implant abutment) and focuses on demonstrating substantial equivalence to a predicate device through physical and mechanical testing, material compatibility, and intended use.

Therefore, I cannot provide a detailed answer to your request based on the provided text, as the information required (acceptance criteria for an AI/ML device, details of an AI/ML study, ground truth establishment, expert adjudication, MRMC studies, etc.) is not present.

In summary, the document does not contain the information needed to answer your questions regarding acceptance criteria and performance of an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zest Anchors, LLC Maleata Hall Director Regulatory Affairs 2875 Loker Ave E Carlsbad, California 92010

June 3, 2024

Re: K233587

Trade/Device Name: LOCATOR Angled Abutment (Various) Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 3, 2023 Received: May 6, 2023

Dear Maleata Hall:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

INSTRUCTIONS

Submission Number (if known)

K233587

Device Name

LOCATOR Angled Abutment (Various)

Indications for Use (Describe)

The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.

IMPLANT COMPATIBILITY

Implant MfgImplant Diameters (Ø) mmImplant SystemNameImplantPlatform NamePlatformDiameter (Ø)mmConnectionType
Straumann3.5, 3.75, 4.0, 4.5BLXRegular Base2.9Bone Level
5.0, 5.5, 6.5BLXWide Base2.9Bone Level

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo features the letters "ZD" in a bold, blue font on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a similar blue font, with the word "SOLUTIONS" appearing below in a smaller font. The logo is clean and professional, suggesting a company that provides dental solutions.

510(K) Summary – K233587 LOCATOR Angled Abutment

i. General Information on Submitter

Applicant:Zest Anchors, LLC
Address:2875 Loker Avenue, EastCarlsbad, CA 92010 USA
Telephone:800-487-1357
Contact Person:David Lin
Contact Title:Sr. Regulatory Affairs Specialist
Email:regulatoryaffairs@zestdent.com
Date Prepared:May 31, 2024

ii. General Information on Device

Proprietary Name:LOCATOR Angled Abutment (Various)
Common Name:Dental Implant Abutment
Classification Name:Endosseous dental implant abutment(21 CFR 872.3630)
Regulatory Class:II
Product Code:NHA (Abutment, Implant, Dental, Endosseous)

iii. Predicate Device

Predicate Device510(k) Number
Locator High Retention Attachment System (LOCATOR FIXED): Zest Anchors,LLCK213391
Reference Devices510(k) Number
Straumann BLX Line Extension - New Abutments: Institut Straumann AGK190040
LOCATOR Implant Anchor Abutment: Zest Anchors, LLCK072878
Straumann® BLX Implant SystemK173961K191256K181703K210855K212533

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo consists of the letters "ZD" in a bold, blue font on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a smaller, bold, blue font, with the word "SOLUTIONS" written below in a slightly smaller, lighter blue font.

Description of Device iv.

The purpose of this submission is to obtain marketing clearance for the LOCATOR Angled Abutment. The LOCATOR Angled Abutment is designed to be used with LOCATOR FIXED and LOCATOR Attachment Systems for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants in the mandible or maxilla.

The LOCATOR Angled Abutment consists of various height abutment bodies with an integrated abutment screw. The attachment features are identical compared to LOCATOR Abutments of the LOCATOR High Retention Attachment System (LOCATOR FIXED), cleared in K213391. The LOCATOR Angled Abutment is compatible with Straumann BLX Implant System cleared in K173961. The LOCATOR Angled Abutment will be used with the accessories of the LOCATOR Implant Attachment System (retention inserts, denture attachment housing, and ancillary processing parts) and LOCATOR FIXED Attachment System (fixed inserts, denture attachment housing) for the attachment of a restoration. The LOCATOR Angled Abutment uses identical attachment features as the LOCATOR Abutments, but instead of the attachment features being aligned coaxially, the interfacing features are provided at a 15 degree angle to allow for angle correction, substantially equivalent to the device of K190040. The LOCATOR Angled Abutments and integrated abutment screw are manufactured from titanium (Ti-6Al-4V). The LOCATOR Angled Abutment body is titanium nitride (TiN) coated, identical to LOCATOR Abutments.

Indication for Use

The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.

Implant MfgImplant Diameters (Ø) mmImplant SystemNameImplantPlatform NamePlatformDiameter (Ø)mmConnectionType
Straumann3.5, 3.75, 4.0, 4.5BLXRegular Base2.9Bone Level
Straumann5.0, 5.5, 6.5BLXWide Base2.9Bone Level

IMPLANT COMPATIBILITY

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo features the letters "ZD" in a bold, blue font on the left side. To the right of the letters, the words "ZEST DENTAL" are stacked on top of the word "SOLUTIONS" in a smaller, blue font. The logo is clean and modern, and the use of blue gives it a professional and trustworthy feel.

Predicate Device Comparison V.

The following table compares the Indications for Use and key technological characteristics of the subject and predicate device:

Characteristic/ FeatureLOCATORAngled Abutment(Subject Device –K233587)Locator HighRetentionAttachmentSystem(Predicate Device– K213391)Straumann BLXLine Extension –New abutments(ReferenceDevice #1 –K190040)LOCATORImplant AnchorAbutment(ReferenceDevice #2 –K072878)Comparison
Reason forpredicate orreferencen/aFixed archrestoration indicationAbutmentangulation featureDesign of abutmentattachment featuresPredicate isunchanged andreference device isadded for angulationfeature.
Indicationfor useThe LOCATORAngled Abutment isindicated for theattachment of full orpartial, fixed andremovable,restorations retainedby endosseousimplants to restoremasticatory functionfor the patient. TheLOCATOR AngledAbutment iscompatible withStraumann BLXImplant SystemRB/WB Ø 3.5, 3.75,4.0, 4.5, 5.0, 5.5, 6.5The High RetentionAttachment System isdesigned to supportfixed, partial or full archrestorations onendosseous dentalimplants in the mandibleor maxilla for thepurpose of restoringmasticatory function. It isused in fixed hybridrestorations that can beattached with a snap-insystem.The StraumannRetentive Systemis indicated for theattachment of fullor partial dentureson Straumanndental implants.The Locator ImplantAnchor Abutment forEndosseous DentalImplants is appropriatefor use withoverdentures or partialdentures retained inwhole or in part byendosseous implants inthe mandible or maxilla.Similar: The subjectand predicate deviceshave a similarindication for use.
RestorationTypeFixed and removableFixedFixed andremovableRemovableSimilar: In addition toa Fixed restorationtype as indicated forthe predicate device,the subject device isalso indicated for aremovable restorationtype.
FDAProductCodeNHA (Abutment, Implant,Dental, Endosseous) 21CFR 872.3630NHA (Abutment,Implant, Dental,Endosseous) 21CFR 872.3630NHA(Abutment,Implant, Dental,Endosseous)21 CFR872.3630NHA(Abutment,Implant, Dental,Endosseous)21 CFR872.3630Same: Thesubject andpredicate deviceshave the sameFDA product code
DESIGN
AbutmentHeight2.5 mm to 7.5 mmN/A1.5mm to6.5mm0 - 6.0 mmSimilar.Predicate,reference, andsubject abutmentheight are similar.
AbutmentTypeAngledN/AStraight andAngledStraightSimilar.Predicate,reference, andsubject abutmentangles are similar.
AbutmentAngled15°N/A0° and 15°Similar.Predicate,reference, andsubject abutmentangles aresimilar.
AbutmentScrewM1.6 x 0.35 Thread,.069" Thread Depth,Ø.062" Screw BodyDiameter,90° Screw Head TaperN/AM1.6 x 0.35 Thread,.069" Thread Depth,Ø.062" Screw BodyDiameter,90° Screw HeadTaperN/ASame.Predicatereference of criticalabutment screwfeatures are thesame.
Materials
AbutmentAndAbutmentScrewMaterialTi-6Al-4V ELIN/ATAV (Ti-6Al-4V)Ti-6Al-4V ELISame: subject,predicate, andreference devicesare made from thesame material -titanium (Ti-6Al-4V)
DeviceMaterialSurfaceTreatmentTiNN/ATiNTiNSame. Predicateand subject deviceuse identicalsurface treatmentprocess
Sterilization
SterileMoist heat end usersterilizationMoist heat enduser sterilizationMoist heat enduser sterilizationMoist heat enduser sterilizationSame. Predicateand subject deviceare both suppliednon-sterile

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo has the letters "ZD" in a bold, dark blue font on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a smaller, dark blue font. Below that, the word "SOLUTIONS" is written in a light blue font with a trademark symbol next to it.

vi. Summary of Non-Clinical Performance Testing

The critical features were identified on the OEM components (implant bodies, abutments, and abutment fixation screws), which are required for proper function. Using calibrated equipment with the appropriate accuracy the critical features were measured and documented. Using the data collected, including the variations, the specifications for the design were created. These specifications were reviewed for manufacturability. All critical tolerances were verified functionally in OEM implants.

Fatigue testing according to ISO 14801: 2016 was performed for the tallest abutment cuff height LOCATOR Angled Abutment along with the smallest diameter OEM implant.

The LOCATOR Angled Abutments are TiN coated. TiN coating performance was tested per ASTM F1044 and ASTM F1147.

The packaging of the LOCATOR Angled Abutments is equivalent to the packaging of the predicate device. Packaging and shipping validation testing was completed where the LOCATOR Angled Abutment worst case device and packaging were undamaged after the test, as desired.

Cleaning and sterilization are leverage in K072878 and this 510(k).

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo features the letters "ZD" in a stylized, connected design on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a bold, sans-serif font, with the word "SOLUTIONS" appearing in a smaller font size below "DENTAL".

MR compatibility testing was conducted per ASTM F2052-15, ASTM F2213-17, ASTM F2182-19, ASTM F2119-07, and FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment". The tests that were conducted are Force: static magnetic field induced displacement force, Torque: static magnetic field induced torque, Heating: Radiofrequency field (RF) induced heating, Image Quality: susceptibility induced image artifacts, Heating: Gradient field induced heating, and Vibration: Gradient field induced vibration.

An assessment for biocompatibility per ISO 10993-1 was conducted using testing from K072878 and additional cytotoxicity testing per ISO 10993-5 was provided in this submission.

No other new testing was performed as a part of this submission for the determination of substantial equivalence.

vii. Substantial Equivalence

The risk management activities and results of the testing described above provide reasonable assurance that the subject devices have demonstrated substantial equivalence to the predicate devices in the that they utilize the same materials and fundamental designs and also have the same intended use and principles of operation.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)