K Number
K233520

Validate with FDA (Live)

Date Cleared
2024-01-12

(72 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl, is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The tested chemotherapy drugs are: Carmustine ( 3.3mg/mL 3,300 ppm) 11.1 min Cisplatin ( 1.0 mg/mL 1,000 ppm) >240 min Cyclophosphamide (20 mg/mL 20,000 ppm) >240 min Cytarabine (Cytosine) (100 mg/mL 100,000 ppm) >240 min Dacarbazine (10 mg/mL 10,000 ppm) >240 min Doxorubicin (Adriamycin) (2.0 mg/mL 2,000 ppm) >240 min Etoposide (Toposar) (20.0 mg/mL 20,000 ppm) >240 min Fluorouracil (5 Flu) (50.0 mg/mL 50,000 ppm) >240 min Ifosfamide (50.0 mg/mL 50,000 ppm) >240 min Methotrexate (25.0 mg/mL 25,000 ppm) >240 min Mitomycin C (0.5 mg/mL 500 ppm) >240 min Mitoxantrone HCL (2.0 mg/mL 2,000 ppm >240 min Paclitaxel (6.0 mg/mL 6,000 ppm) >240 min Thiotepa (10.0 mg/mL 10.000 ppm) 34.8 min Vincristine Sulfate (1.0 mg/mL 1,000 ppm >240 min Fentanyl Citrate Injection (100 mcg/2mL) >240 min Note: Carmustine and Thiotepa have extremely low permeation times of 11.1 and 34.8 minutes respectively. Warning: Do Not Use with Carmustine. Thiotepa

Device Description

The Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue color, powder free, nitrile ambidextrous gloves are offered in five sizes, extra small, small, medium, large, and extra-large, packed in a paper box. The gloves are designed and manufactured in accordance with the ASTM D6319 standard and are tested for use with chemotherapy drugs and Fentanyl per ASTM D6978 standard.

AI/ML Overview

The provided document describes the acceptance criteria and results for the "Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl" device.

Here's an analysis of the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance (Proposed Device)
DimensionsPass
Length (XS, S)220mm min
Length (M, L, XL)230mm min
Width (XS)70±10mm
Width (S)80±10mm
Width (M)95±10mm
Width (L)110±10mm
Width (XL)120±10mm
Thickness (Finger)≥0.05mm
Thickness (Palm)≥0.05mm
Physical properties (ASTM D6319-19)Pass
Tensile Strength (Before Aging)≥14 MPa, min.
Tensile Strength (After Aging)≥14 MPa, min.
Elongation (Before Aging)500%, min.
Elongation (After Aging)400%, min.
Freedom from holes (ASTM D5151-19)Pass
No leakage at sampling level of G-1, AQL 2.5Complies
Residual Powder (ASTM D6124-06)Pass
<2mg per gloveComplies
Permeation by Chemotherapy Drugs and Fentanyl (ASTM D6978-05)Pass (All listed drugs met or exceeded predicate, except Carmustine and Thiotepa which showed differences but were analyzed as not raising new safety concerns, with a warning to not use with Carmustine or Thiotepa)
Carmustine (3.3mg/mL 3,300 ppm)11.1 min
Cisplatin (1.0 mg/mL 1,000 ppm)>240 min
Cyclophosphamide (20 mg/mL 20,000 ppm)>240 min
Cytarabine (Cytosine) (100 mg/mL 100,000 ppm)>240 min
Dacarbazine (10 mg/mL 10,000 ppm)>240 min
Doxorubicin (Adriamycin) (2.0 mg/mL 2,000 ppm)>240 min
Etoposide (Toposar) (20.0 mg/mL 20,000 ppm)>240 min
Fluorouracil (5 Flu) (50.0 mg/mL 50,000 ppm)>240 min
Ifosfamide (50.0 mg/mL 50,000 ppm)>240 min
Methotrexate (25.0 mg/mL 25,000 ppm)>240 min
Mitomycin C (0.5 mg/mL 500 ppm)>240 min
Mitoxantrone HCL (2.0 mg/mL 2,000 ppm)>240 min
Paclitaxel (6.0 mg/mL 6,000 ppm)>240 min
Thiotepa (10.0 mg/mL 10,000 ppm)34.8 min
Vincristine Sulfate (1.0 mg/mL 1,000 ppm)>240 min
Fentanyl Citrate Injection (100 mcg/2mL)>240 min
Biocompatibility (ISO 10993-1, -5, -10, -11, -23)Pass
Irritation (ISO 10993-23)Not an irritant
Sensitization (ISO 10993-10)Not a sensitizer
In Vitro Cytotoxicity (ISO 10993-5)Showed potential toxicity (Note: See point 7 for context)
Acute Systemic Toxicity (ISO 10993-11)Does not induce acute systemic toxicity response

Note on Biocompatibility: The document states that despite the in vitro cytotoxicity showing "potential toxicity," the overall biocompatibility assessment was deemed acceptable, asserting it "does not raise different safety or performance questions since the subject device has acceptable biocompatibility per the biocompatibility endpoint assessment."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the exact sample sizes used for each performance test (e.g., how many gloves were tested for dimensions, holes, tensile strength, or chemotherapy permeation). It references standards like ASTM D6319-19, ASTM D5151-19 (G-1, AQL 2.5), ASTM D6124-06, and ASTM D6978-05, which would prescribe the methodology and sample sizes.

The data provenance regarding country of origin or retrospective/prospective nature is not explicitly mentioned. The applicant, Anhui Powerguard Technology Co., Ltd., is based in China, suggesting the testing may have been conducted there or through contract labs globally.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is a physical medical device (examination gloves), not an AI/software device that requires expert interpretation of images or data to establish ground truth. Therefore, the concept of "experts used to establish ground truth" in the context of medical imaging or AI algorithms is not applicable here. The "ground truth" for this device is established by physical and chemical testing against established engineering and safety standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Similarly, as this is a physical device and not an AI/software for diagnostic interpretation, adjudication methods like 2+1 or 3+1 are not applicable. Performance is measured against quantitative criteria defined by international standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic tools, not for medical gloves.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, this question is not applicable. The device is a physical product (gloves), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the device's performance is established by:

  • International Standards: ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application), ASTM D6124-06 (Standard Test Method for Residual Powder on Medical Gloves), ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves), and ASTM D6978-05 (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs).
  • Biocompatibility Standards: ISO 10993-1, ISO 10993-5 (In Vitro Cytotoxicity), ISO 10993-10 (Dermal Sensitization), ISO 10993-11 (Acute Systemic Toxicity), and ISO 10993-23 (Primary Skin Irritation).

These standards define the methodologies and acceptance limits for evaluating the physical, chemical, and biological properties of the gloves.

Regarding In Vitro Cytotoxicity: The report notes that the "test article extract showed potential toxicity." However, the analysis in the 510(K) Summary explicitly states: "Analysis: In vitro cytotoxicity and acute systemic toxicity information for the predicate device is not publicly available. This does not raise different safety or performance questions since the subject device has acceptable biocompatibility per the biocompatibility endpoint assessment." This indicates that while an in vitro test showed some indication, the overall biocompatibility assessment, likely considering other in vivo tests like skin irritation and sensitization which passed, led to the conclusion of acceptable biocompatibility. This implicitly uses the wider expert consensus and risk assessment methodology embedded in the ISO 10993 series of standards as the 'ground truth' for biocompatibility.

8. The sample size for the training set

This question is not applicable as this is a physical medical device, not an AI/software device that requires a training set.

9. How the ground truth for the training set was established

This question is not applicable as this is a physical medical device, not an AI/software device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 12, 2024

Anhui Powerguard Technology Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K233520

Trade/Device Name: Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: November 1, 2023 Received: November 1, 2023

Dear Ivy Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

llan v

For Bifeng Qian, M.D., Ph.D. Assistant Director

{2}------------------------------------------------

DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233520

Device Name

Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl

Indications for Use (Describe)

The Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentany1, is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The tested chemotherapy drugs are: Chemotherapy drug Concentration Minimum Breakthrough Detection Time Carmustine ( 3.3mg/mL 3,300 ppm) 11.1 min Cisplatin ( 1.0 mg/mL 1,000 ppm) >240 min Cyclophosphamide (20 mg/mL 20,000 ppm) >240 min Cytarabine (Cytosine) (100 mg/mL 100,000 ppm) >240 min Dacarbazine (10 mg/mL 10,000 ppm) >240 min Doxorubicin (Adriamycin) (2.0 mg/mL 2,000 ppm) >240 min Etoposide (Toposar) (20.0 mg/mL 20,000 ppm) >240 min Fluorouracil (5 Flu) (50.0 mg/mL 50,000 ppm) >240 min Ifosfamide (50.0 mg/mL 50,000 ppm) >240 min Methotrexate (25.0 mg/mL 25,000 ppm) >240 min Mitomycin C (0.5 mg/mL 500 ppm) >240 min Mitoxantrone HCL (2.0 mg/mL 2,000 ppm >240 min Paclitaxel (6.0 mg/mL 6,000 ppm) >240 min Thiotepa (10.0 mg/mL 10.000 ppm) 34.8 min Vincristine Sulfate (1.0 mg/mL 1,000 ppm >240 min Fentanyl Citrate Injection (100 mcg/2mL) >240 min Note: Carmustine and Thiotepa have extremely low permeation times of 11.1 and 34.8 minutes respectively. Warning: Do Not Use with Carmustine. Thiotepa

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary

K233520

(As required by 21 CFR 807.92)

Date prepared: 2024-01-12

A. Applicant:

Name: Anhui Powerguard Technology Co., Ltd. Address: Lingbi North Development Zone, Suzhou City, Anhui Province, China Contact: Mr. Nick Dai Title: Vice General Manager Tel: +86-13851555057 Email: nick.dai@powerguard.cc

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com

B. Device:

Trade Name: Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl Common Name: Nitrile Patient Examination Gloves (Powder Free) Size(s): XS, S, M, L, XL

Regulatory Information

Classification Name: Polymer Patient Examination Glove Classification: Class I Product code: LZA, LZC, OPJ, QDO Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

C. Predicate device:

K223903 Non-Sterile, Single use, Powder-free examination glove, Blue, tested for use with Chemotherapy drugs and Fentanyl SEMPERIT INVESTMENTS ASIA PTE. LTD.

D. Indications for use of the device:

The Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and

{5}------------------------------------------------

Fentanyl, is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

The tested chemotherapy drugs are:

Chemotherapy drugConcentrationMinimum Breakthrough Detection Time
Carmustine(3.3mg/mL 3,300 ppm)11.1 min
Cisplatin(1.0 mg/mL 1,000 ppm)>240 min
Cyclophosphamide(20 mg/mL 20,000 ppm)>240 min
Cytarabine (Cytosine)(100 mg/mL 100,000 ppm)>240 min
Dacarbazine(10 mg/mL 10,000 ppm)>240 min
Doxorubicin (Adriamycin)(2.0 mg/mL 2,000 ppm)>240 min
Etoposide (Toposar)(20.0 mg/mL 20,000 ppm)>240 min
Fluorouracil (5 Flu)(50.0 mg/mL 50,000 ppm)>240 min
Ifosfamide(50.0 mg/mL 50,000 ppm)>240 min
Methotrexate(25.0 mg/mL 25,000 ppm)>240 min
Mitomycin C(0.5 mg/mL 500 ppm)>240 min
Mitoxantrone HCL(2.0 mg/mL 2,000 ppm)>240 min
Paclitaxel(6.0 mg/mL 6,000 ppm)>240 min
Thiotepa(10.0 mg/mL 10,000 ppm)34.8 min
Vincristine Sulfate(1.0 mg/mL 1,000 ppm)>240 min
Fentanyl Citrate Injection(100 mcg/2mL)>240 min

Note: Carmustine and Thiotepa have extremely low permeation times of 11.1 and 34.8 minutes respectively. Warning: Do Not Use with Carmustine, Thiotepa

E. Device Description:

The Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue color, powder free, nitrile ambidextrous gloves are offered in five sizes, extra small, small, medium, large, and extra-large, packed in a paper box.

The gloves are designed and manufactured in accordance with the ASTM D6319 standard and are tested for use with chemotherapy drugs and Fentanyl per ASTM D6978 standard.

F. Summary of Technological Characteristics

Table 1 General Comparison of Proposed and Predicate Devices

DeviceProposed DevicePredicate DeviceResult
---------------------------------------------------

{6}------------------------------------------------

510K #K223903-
Product NamePowder free, Non-sterile, Blue NitrileExamination Gloves, Tested for Use withChemotherapy Drugs and FentanylNon-Sterile, Single use, Powder-freeexamination glove, Blue, tested foruse with Chemotherapy drugs andFentanyl-
Product CodeLZA (primary), LZC, OPJ, QDOLZA (primary), LZC, OPJ, QDOSame
ClassificationClass IClass ISame
RegulationNumber21 CFR 880.625021 CFR 880.6250Same
Indications foruseThe Powder free, Non-sterile, Blue NitrileExamination Glove, Tested for Use withChemotherapy Drugs and Fentanyl, is adisposable device intended for medicalpurposes that is worn on the examiner's handto prevent contamination between patientand examiner.The tested chemotherapy drugs are:Chemotherapy drug Concentration MinimumBreakthrough Detection TimeCarmustine (3.3mg/mL 3,300 ppm) 11.1Cisplatin (1.0 mg/mL 1,000 ppm) >240 minCyclophosphamide (20 mg/mL 20,000 ppm)> 240 minCytarabine (Cytosine) (100 mg/mL 100,000ppm) >240 minDacarbazine (10 mg/mL 10,000 ppm) >240 minDoxorubicin (Adriamycin) (2.0 mg/mL 2,000ppm) >240 minEtoposide (Toposar) (20.0 mg/mL 20,000ppm) >240 minFluorouracil (5 Flu) (50.0 mg/mL 50,000ppm) >240 minIfosfamide (50.0 mg/mL 50,000 ppm) >240 minMethotrexate (25.0 mg/mL 25,000 ppm)> 240 minMitomycin C (0.5 mg/mL 500 ppm) >240minMitoxantrone HCL (2.0 mg/mL 2,000 ppm >240 minPaclitaxel (6.0 mg/mL 6,000 ppm) >240minThis device is an ambidextrous patientexamination glove that is a non-sterile,single use, disposable device intendedfor medical purposes, worn on theexaminer's hand or finger to preventcontamination between patient andexaminer.The tested chemotherapy drugs are:Carmustine (BCNU) (3.3 mg/ml).Permeation time: Carmustine (BCNU)has extremely low permeation times of14.7 minutes.Cisplatin (1.0 mg/ml). Permeation time:no breakthrough up to 240 minutesCyclophosphamide (Cytoxan) (20.0 mg/ml). Permeation time: no breakthroughup to 240 minutesCytarabine (100 mg/ml). Permeationtime: no breakthroughup to 240 minutes Dacarbazine (DTIC)(10.0 mg/ml). Permeation time: nobreakthrough up to 240 minutesDoxorubicin Hydrochloride (2.0 mg/ml).Permeation time: no breakthrough up to240 minutesEtoposide (20.0 mg/ml). Permeationtime: no breakthrough up to 240minutesFluorouracil (50.0 mg/ml).Permeation time: no breakthrough up to240 minutesIfosfamide (50.0 mg/ml). Permeationtime: no breakthrough up to 240minutesSimilar
Thiotepa (10.0 mg/mL 10,000 ppm) 34.8Methotrexate (25.0
Vincristine Sulfate (1.0 mg/mL 1,000 ppmmg/ml). Permeation time: no
>240 minbreakthrough up to 240 minutes
Fentanyl Citrate Injection (100 mcg/2mL) >Mitomycin C (0.5 mg/ml). Permeation
240 mintime: no breakthrough up to 240
Note: Carmustine and Thiotepa haveminutes
extremely low permeation times of 11.1 andMitoxantrone (2.0 mg/ml).
34.8 minutes respectively.Permeation time: no breakthrough up to
Warning: Do Not Use with Carmustine,240 minutes Paclitaxel (Taxol) (6.0
Thiotepamg/ml). Permeation time: no
breakthrough up to 240 minutes
Thiotepa (10.0 mg/ml). Permeation
time: Thiotepa has extremely low
permeation times of 13.6 minutes
Vincristine Sulfate (1.0 mg/ml).
Permeation time: no Breakthrough up
to 240 minutes.
The tested Opiod is:
Fentanyl Citrate Injection
(100mcg/2mL). Permeation: no
breakthrough up to 240 minutes
Please note that the following drugs
have extremely low permeation times:
Carmustine: 14.7 minutes
Thiotepa: 13.6 minutes
Warning: DO NOT USE WITH
CARMUSTINE OR THIOTEPA
Powder freeYesYesSame
Design featureAmbidextrousAmbidextrousSame
MaterialNitrileNitrileSame
ColorBlueBlueSame
OTC useYesYesSame
SterilityNon-sterileNon-sterileSame
UseSinge useSingle useSame
Labeling claimTested for Use with Chemotherapy Drugsand FentanylTested for Use with Chemotherapy Drugsand FentanylSame

{7}------------------------------------------------

Analysis: The proposed device and the predicate device have same indications, only minor differences on the penetration time of carmustine and thiotepa between the proposed device and the predicate device, which will not raise any new safety or performance concerns.

Technological characteristicsProposed DevicePredicate Device (K223903)Result
Length
XS, SMinimum 220mmMinimum 220mmSame

{8}------------------------------------------------

M, L, XLMinimum 230mmMinimum 230mmSame
Palm width (size) (mm)
XS$70\pm10$$70\pm10$Same
S$80\pm10$$80\pm10$Same
M$95\pm10$$95\pm10$Same
L$110\pm10$$110\pm10$same
XL$120\pm10$$120\pm10$Same
Thickness (mm)
Finger$\ge0.05$$\ge0.05$Same
Palm$\ge0.05$$\ge0.05$Same

Table 3 Performance comparison

ItemProposed devicePredicate (K223903)deviceResult
Physical propertiesBefore agingTensile strength14MPa, min14MPa, minSame
Ultimate elongation500%, min500%, minSame
After agingTensile strength14MPa, min14MPa, minSame
Ultimate elongation400%, min400%, minSame
Comply with ASTM D6319Comply with ASTM D6319Same
Freedom from holesBe free from holes when tested inaccordance with ASTM D5151G-1, AQL 2.5Be free from holes when tested in accordance withASTM D5151G-1, AQL 2.5Same
Residual PowderMeet the requirements of ASTMD6124Meet the requirements ofASTM D6124Same

Table 4 Chemotherapy Permeation and Fentanyl Citrate Comparison Claim:

Tested Chemotherapy Drug andConcentrationMinimum breakthrough detection time in minutesResult
Proposed devicePredicate device (K223903)
Carmustine (3.3mg/mL 3,300 ppm)11.114.7Similar
Cisplatin (1.0 mg/mL 1,000 ppm)>240>240Same
Cyclophosphamide (20 mg/mL 20,000 ppm)>240>240Same
Cytarabine (Cytosine) (100 mg/mL 100,000 ppm)>240>240Same
Dacarbazine (10 mg/mL 10,000 ppm)>240>240Same
Doxorubicin (Adriamycin) (2.0 mg/mL 2,000 ppm)>240>240Same
Etoposide (Toposar) (20.0 mg/mL 20,000 ppm)>240>240Same

{9}------------------------------------------------

Fluorouracil (5 Flu) (50.0 mg/mL 50,000 ppm)>240>240Same
Ifosfamide (50.0 mg/mL 50,000 ppm)>240>240Same
Methotrexate (25.0 mg/mL 25,000 ppm)>240>240Same
Mitomycin C (0.5 mg/mL 500 ppm)>240>240Same
Mitoxantrone HCL (2.0 mg/mL 2,000 ppm)>240>240Same
Paclitaxel (6.0 mg/mL 6,000 ppm)>240>240Same
Thiotepa (10.0 mg/mL 10,000 ppm)34.813.6Similar
Vincristine Sulfate (1.0 mg/mL 1,000 ppm)>240>240Same
Fentanyl Citrate Injection (100 mcg/2mL)>240>240Same

Analysis: Minor differences on the penetration time of carmustine and thiotepa between the proposed device and the predicate device, which will not raise any new safety or performance concerns.

Table 5 Biocompatibility comparison
-----------------------------------------
ItemProposed devicePredicate device (K223903)Result
MaterialNitrileNitrileSame
Biocompatibility
IrritationISO 10993-23Under the conditions of the study, not an irritant.Comply with ISO 10993-10Same
SensitizationISO 10993-10Under the conditions of the study, not a sensitizer.
In Vitro CytotoxicityISO 10993-5Under the conditions of the study, the test article extract showed potential toxicity.Not availableDifferent
Acute Systemic ToxicityISO 10993-11Under the conditions of the study, the device extract does notNot availableDifferent

{10}------------------------------------------------

induceacute
systemictoxicity
response.

Analysis: In vitro cytotoxicity and acute systemic toxicity information for the predicate device is not publicly available. This does not raise different safety or performance questions since the subject device has acceptable biocompatibility per the biocompatibility endpoint assessment.

G. Summary of Non-Clinical Testing

A Biocompatibility

Biocompatibility Testing according to ISO 10993-1:2018, the nature of body contact for the subject device is Surface Device category, Skin Contact and duration of contact is Limited (≤24h). The following tests for the subject device were conducted to evaluate the biocompatibility:

  • ISO 10993-5: In Vitro Cytotoxicity
  • ISO 10993-23: Primary Skin Irritation
  • ISO 10993-10: Dermal Sensitization
  • ISO 10993-11: Acute Systemic Toxicity

△ Performance Testing

Physical performance testing of the proposed device was conducted as per ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application and are tested for use with chemotherapy drugs and Fentanyl per ASTM D6978 standard.

To summarize, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
  • ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs
Test MethodPurposeAcceptance CriteriaResults
Dimensions (width)(thickness)ASTM D6319-19The purpose of the test is toevaluate the physicaldimension of the gloveLength220mm min for XS, S230mm min for M, L, XLPass
WidthXS: 70±10mmS: 80±10mmM: 95±10mmL:110±10mmXL: 120±10mmPass
Palm - 0.05mm min.Pass
Finger-0.05mm min.
Physical propertiesASTM D6319-19The purpose of the test is toevaluate the tensile strength andultimate elongationbefore and after agingTensile Strength:Before Aging ≥14 MPa, min.After Aging ≥14 MPa, min.Elongation:Pass

{11}------------------------------------------------

Before Aging 500%, min.After Aging 400%, min.
Freedom from holesASTM D5151-19The purpose of the test is todetect holes in the glovesNo leakage at sampling level of G-1,AQL 2.5Pass
Residual PowderASTM D6124-06(Reapproved2017)The purpose of the test is todetect the powder residue inthe glove<2mg per glovePass
Permeation byChemotherapyDrugsASTM D6978-05(2019)The purpose of the test is todetect the Permeation timeby Chemotherapy Drugs andFentanyl of the gloveRefer the above table 4Pass

H. Summary of Clinical Testing

No clinical study is included in this submission.

I. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device Powder free, Nonsterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K223903.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.