K Number
K233106

Validate with FDA (Live)

Date Cleared
2024-04-19

(205 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AccuCath Ace™ Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This device may be used for adult and pediatric patients, including those patients with difficult intravascular access who may have small, fragile, and/or non-palpable vessels, with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath Ace™ IV Catheter is suitable for use with power injectors.

Device Description

The AccuCath Ace™ Intravascular Catheter system consists of a radiopaque catheter with a valve mechanism delivered over a guidewire with an atraumatic tip design; a flashback chamber to enhance flashback visualization, and a safety container that prevents sharp injuries. The AccuCath Ace IV Catheter is designed to reduce blood exposure during insertion, for use with ultrasound, and for use with the Cue Needle Tracking System.

AI/ML Overview

The information provided indicates that the AccuCath Ace™ Intravascular Catheter did not undergo a new study to prove it meets acceptance criteria for its current submission. Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (AccuCath™ Intravascular Catheter, K162894) based on a modification to its Indications for Use, which now explicitly includes patients with difficult intravascular access (DIVA).

The document states that a risk analysis determined that no verification or validation activities were required because the modifications to the Indications for Use and labeling "do not include any changes to the design, materials, performance, or risk profile of the cited predicate device."

Therefore, the "acceptance criteria" and "device performance" described below are in reference to the previous clearance of the predicate device and the modifications that had been made to it, for which "Verification, sterilization, biocompatibility, and packaging testing was carried out as necessary for each of these changes at the time of the change." The current submission does not detail these specific tests, their criteria, or results, but rather asserts that the changes "were found to be as safe and as effective and introduced no new or modified risks."

The table below summarizes the information provided regarding the comparison between the subject device (AccuCath Ace™ Intravascular Catheter) and its predicate device (AccuCath™ Intravascular Catheter, K162894), highlighting what is considered "met" based on the substantial equivalence argument rather than new primary testing.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria Category (Implied by Comparison)Predicate Device Performance / CharacteristicSubject Device Performance / CharacteristicComparison / Status
Primary Indication for UseInsertion into vascular system for blood sampling, BP monitoring, fluid administration.Insertion into vascular system for blood sampling, BP monitoring, fluid administration.Same
Additional Indication for UseGeneral Use PatientsAdult and pediatric patients, including those with difficult intravascular access (DIVA).Met (New indication supported by external literature, no design change, no new risk)
Catheter DimensionsLength: 1.25 & 2.25 inches; Diameter: 18, 20, 22 gaugeLength: 1.25 & 2.25 inches; Diameter: 18, 20, 22 gaugeSame
Duration of UseShort term (<30 days)Short term (<30 days)Same
Primary Device ComponentsNeedle, Guidewire, CatheterNeedle, Guidewire, CatheterSame
Means of InsertionPercutaneous, over a guidewirePercutaneous, over a guidewireSame
Insertion SitePeripheralPeripheralSame
Primary Device MaterialsCatheter Base: Pebax®, Polyurethane; Needle: Stainless Steel; Guidewire: NitinolCatheter Base: Pebax®, Polyurethane; Needle: Stainless Steel; Guidewire: NitinolSame
Catheter Proximal ConfigurationLuer ConnectionLuer ConnectionSame
Catheter Distal ConfigurationOpen EndedOpen EndedSame
Number of LumensSingle LumenSingle LumenSame
Power Injection Maximum Flow Rate6 mL/s6 mL/sSame
SterilityProvided SterileProvided SterileSame
Available ConfigurationsStandalone, Basic KitStandalone, Basic Kit, Intermediate KitMet (Addition of intermediate kit, no effect on safety/effectiveness, no new risk)
Cue Needle Tracking System CompatibilityNone2.25" AccuCath Ace IV CathetersMet (Addition of compatibility, no effect on safety/effectiveness, no new risk. Verification testing done on aspects related to this change: needle, packaging, ultrasound system/Cue compatibility, tracking accuracy, magnetization. All acceptance criteria met.)
Safety and EffectivenessFound to be safe and effectiveFound to be safe and effectiveMet (Based on predicate's performance and prior verification for incremental changes)
Risk ProfileAcceptable risk profileAcceptable risk profileMet (No new or modified risks identified)

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states: "The results of the risk analysis determined that no verification or validation activities were required because the subject device modifications to the Indications for use and resulting modifications to the instructions for use and labeling do not include any changes to the design, materials, performance, or risk profile of the cited predicate device. Therefore, it is not necessary to conduct additional performance tests including verification and validation."

This means there was no new test set specifically for this 510(k) submission to assess the AccuCath Ace™ in the context of the DIVA indication. The claim of equivalence is based on the previously cleared predicate device and earlier incremental changes.

However, the document mentions: "Verification testing was carried out on all changed aspects: the needle, packaging, ultrasound system/Cue compatibility, tracking accuracy, and magnetization. All acceptance criteria was met".

  • Sample Size: Not specified.
  • Data Provenance: Not specified, but these prior tests would likely have been internal validation tests conducted by the manufacturer.

The submission also cites "a body of literary evidence demonstrating that the AccuCath Ace Intravascular Catheter has equivalent or better outcomes in DIVA patients compared to catheters with no guidewire." This refers to clinical literature, not a new test set conducted by the manufacturer for this submission. The specifics of this literature (e.g., sample sizes, countries of origin, retrospective/prospective) are not provided within this document.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable, as a new test set requiring expert ground truth establishment was not conducted for this 510(k) submission. For the "body of literary evidence," the ground truth would be established by the clinicians and researchers involved in those studies, but their specific numbers and qualifications are not detailed here.

4. Adjudication Method for the Test Set

Not applicable, as a new test set was not conducted for this 510(k) submission.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done as part of this submission. The submission relies on existing clinical literature (not a new study conducted for this submission) that suggests improved outcomes for DIVA patients with the AccuCath Ace™ Intravascular Catheter compared to catheters without a guidewire. No "effect size of how much human readers improve with AI vs without AI assistance" is relevant or provided, as this is a medical device, not an AI-assisted diagnostic tool for human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable, as this is a medical device (intravascular catheter), not an algorithm or AI system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the previously conducted "verification testing" for incremental changes (e.g., Cue compatibility), the ground truth would have been established through objective measurement and engineering standards.

For the claim regarding DIVA patients, the submission refers to "a body of literary evidence demonstrating that the AccuCath Ace Intravascular Catheter has equivalent or better outcomes in DIVA patients compared to catheters with no guidewire. These outcomes include improved first attempt success, reduction of insertion complications, improved completion of therapy, increased dwell time of the catheter, and overall patient and clinician satisfaction." This implies that the ground truth for these claims comes from outcomes data reported in clinical literature.

8. The Sample Size for the Training Set

Not applicable, as this is a medical device, not a machine learning model requiring a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as this is a medical device, not a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 19, 2024

Bard Access Systems, Inc. (C.R. Bard, Inc.) Samira Saeed Regulatory Affairs Specialist I 605 N 5600 W Salt Lake City, Utah 84116

Re: K233106

Trade/Device Name: AccuCath Ace™ Intravascular Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: Class II Product Code: FOZ Dated: April 4, 2024 Received: April 16, 2024

Dear Samira Saeed:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

David Walloschek

David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices,

{2}------------------------------------------------

and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233106

Device Name AccuCath Ace™ Intravascular Catheter

Indications for Use (Describe)

The AccuCath Ace™ Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This device may be used for adult and pediatric patients, including those patients with difficult intravascular access who may have small, fragile, and/or non-palpable vessels, with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath Ace™ IV Catheter is suitable for use with power injectors.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Bard Access Systems, Inc. 605 North 5600 West
Salt Lake City, UT 84116 USA
Phone: +1-801-5222-50000
Fax: +1-801-5222-50000
Fax: +1-801-522-4948

Image /page/4/Picture/1 description: The image shows the logos of two companies, BD and BARD. The BD logo is on the left and consists of an orange sun-like symbol and the letters BD in blue. The BARD logo is on the right and consists of the word BARD in green, with the text "has joined BD" below it.

AccuCath Ace™ Intravascular Catheter Traditional 510(k) Premarket Notification

GeneralProvisionsSubmitter Name:Bard Access Systems, Inc. (Bard has joinedBD)
Submitter Address:605 North 5600 WestSalt Lake City, Utah 84116
Contact Person:Samira SaeedRegulatory Affairs Specialist
Telephone Number:801.522.5000
Date of Preparation:April 19th 2024
Trade Name:AccuCath Ace™ Intravascular Catheter
Common Name:catheter, intravascular, therapeutic, short-term less than 30 days
SubjectDevice:Classification Name:Intravascular catheter
Class:Class II
Regulation Number:880.5200
Product Code:FOZ
Classification Panel:General Hospital
PredicateDeviceTrade Name:AccuCath™ Intravascular Catheter
Common Namecatheter, intravascular, therapeutic, short-term less than 30 days
Classification Name:Intravascular catheter
Class:Class II
Regulation Number:880.5200
Product Code:FOZ
Premarket Notification Number:K162894
DeviceDescriptionThe AccuCath Ace™ Intravascular Catheter system consists of a radiopaque catheter witha valve mechanism delivered over a guidewire with an atraumatic tip design; a flashbackchamber to enhance flashback visualization, and a safety container that prevents sharpinjuries. The AccuCath Ace IV Catheter is designed to reduce blood exposure duringinsertion, for use with ultrasound, and for use with the Cue Needle Tracking System.
Indications forUseThe AccuCath Ace™ Intravascular Catheter is inserted into a patient's vascular system tosample blood, monitor blood pressure, or administer fluids intravenously. This devicemay be used for adult and pediatric patients, including those patients with difficultintravascular access who may have small, fragile, and/or non-palpable vessels, withconsideration given to adequacy of vascular anatomy, appropriateness of the solutionbeing infused, and duration of therapy. The AccuCath Ace™ IV Catheter is suitable for usewith power injectors.
TechnologicalCharacteristicsTechnological characteristics of the subject AccuCath Ace™ Intravascular Catheter aresubstantially equivalent with respect to design and function to those of the citedpredicate device.

K233106 - 510(k) Summary 21 CFR 807.92(a)

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows two company logos. The first logo consists of an orange circular graphic with radiating lines, followed by the blue letters "BD". The second logo features the green letters "BARD" in a stylized font, with the text "has joined BD" underneath in a smaller font.

AccuCath Ace™ Intravascular Catheter Traditional 510(k) Premarket Notification

The following table provides a comparison between the subject and predicate device:

AttributePredicate DeviceAccuCath™IntravascularCatheter (K162894)Subject DeviceAccuCath Ace™Intravascular CatheterComparison
OwnerBard AccessSystems, Inc.Bard Access Systems,Inc.Same
ClassificationClass IIClass IISame
510(k) StatusK162894Subject of thisPremarket Notification-
Indications forUseThe AccuCath™IntravascularCatheter is insertedinto a patient'svascular system tosample blood,monitor bloodpressure, oradminister fluidsintravenously. Thisdevice may be usedwith considerationgiven to adequacy ofvascular anatomy,appropriateness ofthe solution beinginfused, andduration of therapy.The AccuCath IVCatheter is suitablefor use with powerinjectors.The AccuCath Ace™Intravascular Catheter isinserted into a patient'svascular system tosample blood, monitorblood pressure, oradminister fluidsintravenously. Thisdevice may be used foradult and pediatricpatients, includingthose patients withdifficult intravascularaccess who may havesmall, fragile, and/ornon-palpable vessels,with considerationgiven to adequacy ofvascular anatomy,appropriateness of thesolution being infused,and duration of therapy.The AccuCath Ace™ IVCatheter is suitable foruse with powerinjectors.DifferentAddedindication forpatients withDIVA which isa subset ofgeneral usepopulation -no effect ondevice safetyoreffectivenessor change inrisk.
CommercialNameAccuCath™IntravascularCatheterAccuCath Ace™Intravascular CatheterDifferentThe minorname changehas no effecton devicesafety oreffectiveness
and results inno new ormodified risks
CatheterDimensionsLength: 1.25 & 2.25 inchesDiameter: 18, 20, 22 gaugeLength: 1.25 & 2.25 inchesDiameter: 18, 20, 22 gaugeSame
Durations ofUseShort term (<30 days)Short term (<30 days)Same
PrimaryDeviceComponentsNeedleGuidewireCatheterNeedleGuidewireCatheterSame
Means ofInsertionPercutaneous, overa guidewirePercutaneous, over aguidewireSame
Insertion SitePeripheralPeripheralSame
PrimaryDeviceMaterialsCatheter BaseMaterials:• Shaft Tubing: Pebax®• Luer Connector:PolyurethaneNeedle: Stainless SteelGuidewire: NitinolCatheter BaseMaterials:• Shaft Tubing: Pebax®• Luer Connector:PolyurethaneNeedle: Stainless SteelGuidewire: NitinolSame
CatheterProximalConfigurationLuer ConnectionLuer ConnectionSame
CatheterDistalConfigurationOpen EndedOpen EndedSame
Number ofLumensSingle LumenSingle LumenSame
PowerInjectionMaximumFlow Rate6 mL/s6 mL/sSame
SterilityProvided SterileProvided SterileSame
AvailableConfigurationsStandaloneBasic KitStandaloneBasic KitIntermediate KitDifferentThe additionof theintermediatekitconfigurationhas no effect
Cue NeedleTrackingSystemCompatibilityNone2.25" AccuCath Ace IVCatheterson devicesafety oreffectivenessand results inno new ormodified risks.DifferentThe additionof Cuecompatibilityhas no effecton devicesafety oreffectivenessand results inno new ormodified risks.Verificationtesting wascarried out onall changedaspects: theneedle,packaging,ultrasoundsystem/Cuecompatibility,trackingaccuracy, andmagnetization.All acceptancecriteria wasmet

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows two company logos. The first logo is for BD, with an orange sun-like symbol to the left of the blue letters "BD". The second logo is for BARD, with the green letters "BARD" stacked above the phrase "has joined BD" in a smaller font.

AccuCath Ace™ Intravascular Catheter Traditional 510(k) Premarket Notification

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logos of BD and BARD. The BD logo is on the left and consists of an orange sun-like symbol and the letters "BD" in blue. The BARD logo is on the right and consists of the word "BARD" in green. Below the BARD logo, it says "has joined BD" in a smaller font.

AccuCath Ace™ Intravascular Catheter Traditional 510(k) Premarket Notification

The key modification made to the subject device when compared to the predicate device is the specification of a difficult intravascular access (DIVA) indication to the indications for use. As a result, the indications for use in the product instructions for use were updated to reflect the modification. Previously, several design, packaging, and labeling changes were made to the AccuCath Ace IV Catheter. These changes are currently implemented in the marketed predicate devices.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logos of BD and BARD. The BD logo is in blue, and the BARD logo is in green. Underneath the BARD logo, the text "has joined BD" is written in a smaller font. The logos are placed side by side.

AccuCath Ace™ Intravascular Catheter Traditional 510(k) Premarket Notification

PerformanceTestsAs part of Bard Access Systems, Inc.'s design controls, a risk analysis wasconducted to assess the impact of the proposed subject device indications for usemodifications. The results of the risk analysis determined that no verification orvalidation activities were required because the subject device modifications to theIndications for use and resulting modifications to the instructions for use and labelingdo not include any changes to the design, materials, performance, or risk profile ofthe cited predicate device. Therefore, it is not necessary to conduct additionalperformance tests including verification and validation.Since clearance of the predicate, several additional changes were made to the deviceand documented via Letter to File. Verification, sterilization, biocompatibility, andpackaging testing was carried out as necessary for each of these changes at the timeof the change. In each case, the changed device was found to be as safe and aseffective and introduced no new or modified risks. These historic changes arecurrently implemented in the marketed predicate devices.
ClinicalLiteratureThere is a body of literary evidence demonstrating that the AccuCath Ace IntravascularCatheter has equivalent or better outcomes in DIVA patients compared to catheters withno guidewire. These outcomes include improved first attempt success, reduction ofinsertion complications, improved completion of therapy, increased dwell time of thecatheter, and overall patient and clinician satisfaction. AccuCath Ace includes featuresthat improve ease of insertion and limit vessel damage making the AccuCath AceIntravascular Catheter an ideal option for DIVA patients.
Summary ofSubstantialEquivalenceThe modification to the indications for use and resulting modifications to the productinstructions for use and labeling has no impact on the intended use, technologicalcharacteristics, or risk profile of the subject device because there are no changes to thedesign or performance of the predicate device.Previous minor changes made to the device have been documented and the necessarytesting carried out. The changes were found to have no impact on the safety andeffectiveness of the device, nor did they introduce new risks or modify existing risks.Therefore, the subject AccuCath Ace™ Intravascular Catheter is substantially equivalent tothe predicate AccuCath™ Intravascular Catheter.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).