K Number
K233091

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-10-26

(30 days)

Product Code
Regulation Number
870.4360
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Quantum Perfusion Centrifugal Blood Pump CP37 with Integrated Sensors, indicated for use exclusively with Quantum Centrifugal Drive of Spectrum Medical, is intended to pump the blood through an extracorporeal circuit and to monitor pressure for periods lasting less than 6 hours for the purpose of providing either:

(i) Full of partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or

(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.

Devices are intended for adult patients.

Device Description

The Quantum Perfusion Centrifugal Blood Pump with Integrated Sensors is a standalone single use device intended to pump blood into the extracorporeal circuit and to monitor pressure for periods lasting less than 6 hours. The device is equipped with two sensors integrated in the blood inlet and outlet connectors.

The device is non-toxic, non pyrogenic, sterilized by ethylene oxide and packaged in a single blister. Blood contact surfaces of the device are coated with a stable biocompatible surface to reduce platelet activation and adhesion while preserving platelet function.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called the "Quantum Perfusion Centrifugal Blood Pump CP37 with Integrated Sensors." The purpose of a 510(k) is to demonstrate that a new device is substantially equivalent to a predicate device already on the market. In this case, the predicate device is the "Quantum Perfusion Centrifugal Blood Pump CP22 with Integrated Sensors" (K201320), and a reference device is the "Quantum Perfusion Centrifugal Blood Pump CP37" (K202169).

The filing states that the new device, CP37 with Integrated Sensors, is essentially a variant introduced into the existing product family, aligning with the general structure, principle of operation, intended use, manufacturing, and sterilization processes of the previously cleared predicate and reference devices.

Based on the provided text, the device's acceptance criteria and the study that proves it meets those criteria are described as follows:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly present a table of crisp, quantitative acceptance criteria and corresponding reported device performance metrics in the format typically seen for a new algorithmic or diagnostic device. Instead, the approach taken is to demonstrate substantial equivalence to existing cleared devices (CP22 with Integrated Sensors and CP37).

The implicit "acceptance criteria" appear to be that the new device variant (CP37 with Integrated Sensors) performs equivalently to the predicate (CP22 with Integrated Sensors) and reference (CP37) devices for its intended use, without raising new issues of safety or effectiveness.

The reported device performance is conveyed through the statement:
"Technical specifications have been verified through testing activities performed according to the same internal applicable standards/protocols of the original cleared devices. Applicable testing activities demonstrated that the proposed device model does not raise any issues of safety and effectiveness as compared to the currently cleared predicate and reference product."

Therefore, the "performance" is that it met the same standards as the cleared devices.

Acceptance Criteria (Implicit)Reported Device Performance
Maintain the same general structure, principle of operation, intended use, manufacturing, and sterilization processes as predicate and reference devices.The new REF code introduced is aligned with the original submission K201320 in these aspects.
Technical specifications verified according to applicable standards/protocols used for original cleared devices.Technical specifications have been verified through testing activities performed according to the same internal applicable standards/protocols of the original cleared devices.
No new issues of safety and effectiveness are raised compared to the predicate and reference products.Applicable testing activities demonstrated that the proposed device model does not raise any issues of safety and effectiveness as compared to the currently cleared predicate and reference product.
Appropriate for its intended use.The proposed devices have been demonstrated to be appropriate for their intended use.

2. Sample Size Used for the Test Set and Data Provenance:

The document states:
"The following activities were performed to demonstrate product safety and effectiveness, considering the proposed change and related impact:

  • ...
  • Review and Assessment of data available on predicate( K201320) and reference (K202169) devices submission and applicability to new variant introduced in the product family portfolio."

This indicates that the "test set" for demonstrating safety and effectiveness was primarily based on a review and assessment of existing data from the previously cleared predicate (K201320) and reference (K202169) devices. There is no mention of a new, distinct "test set" with a specific sample size.

Data Provenance: The original data for the predicate and reference devices would have been generated during their respective premarket clearances. The current submission is a Special 510(k) for a device variant, often relying on existing data and non-clinical bench testing. The document does not specify the country of origin or whether the original data was retrospective or prospective beyond referring to previous submissions.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not applicable/provided in the document. The submission is for a physical medical device (blood pump with sensors), not an AI/software diagnostic device that requires expert-established ground truth for image or data interpretation. The "ground truth" for this device's performance would be engineering specifications, physical measurements, and biocompatibility testing, not expert consensus on interpretations.

4. Adjudication Method for the Test Set:

This information is not applicable/provided. Since the "test set" primarily involved a review of existing data and non-clinical testing for a physical device, there isn't an "adjudication method" in the sense of reconciling differences in expert interpretations of data as would be found in diagnostic studies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that are interpreted by human readers, often with and without AI assistance, to measure the impact of AI on reader performance. This submission is for a physical device, not an AI diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done:

No, a standalone study was not done in the context of an algorithm's performance without human intervention. This device is a physical centrifugal blood pump with integrated sensors, not an AI algorithm. Its performance is measured by its mechanical and biocompatibility characteristics, not by an algorithm's diagnostic accuracy.

7. The Type of Ground Truth Used:

The "ground truth" for this device relates to its engineering specifications, functional performance (e.g., pumping efficiency, pressure monitoring accuracy), material properties, and biocompatibility. This is established through:

  • Engineering specifications and design documents.
  • Bench testing results measuring flow rates, pressure accuracy, material strength, etc.
  • Biocompatibility testing (e.g., ISO 10993 standards).
  • Sterilization validation data.
  • Compliance with applicable standards (e.g., ISO 15223-1:2021 for labeling).

The document explicitly states: "Technical specifications have been verified through testing activities performed according to the same internal applicable standards/protocols of the original cleared devices." This implies that the "ground truth" and verification methods align with the established standards and protocols used for similar devices.

8. The Sample Size for the Training Set:

This information is not applicable/provided. This device is a physical medical device, not an AI/machine learning model that undergoes "training." Therefore, there is no training set sample size to report.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable/provided. As the device is not an AI/machine learning model, there is no training set or corresponding ground truth establishment process in that context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 26, 2023

Qura S.r.1. Raffaella Tommasini QA&RA Director Via di Mezzo, 23 Mirandola, MO 41037 Italy

Re: K233091

Trade/Device Name: Quantum Perfusion Centrifugal Blood Pump CP37 with Integrated Sensors Regulation Number: 21 CFR 870.4360 Regulation Name: Nonroller-type blood pump Regulatory Class: Class II Product Code: KFM Dated: September 26, 2023 Received: September 26, 2023

Dear Raffaella Tommasini:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Eric E. Richardson -S

for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233091

Device Name

Quantum Perfusion Centrifugal Blood Pump CP37 with Integrated Sensors

Indications for Use (Describe)

The Quantum Perfusion Centrifugal Blood Pump CP37 with Integrated Sensors, indicated for use exclusively with Quantum Centrifugal Drive of Spectrum Medical, is intended to pump the blood through an extracorporeal circuit and to monitor pressure for periods lasting less than 6 hours for the purpose of providing either:

(i) Full of partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or

(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.

Devices are intended for adult patients.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a partial blue circle. To the right of the figure, the word "QURA" is written in a bold, blue font. Below "QURA", the phrase "advanced medical solutions" is written in a smaller, lighter blue font.

510(K) SUMMARY

SUBMITTER l.

Submitter Name:Qura S.r.l.
Submitter Address:Via di Mezzo, 23 41037 Mirandola (MO) Italy
Contact Person:Raffaella Tommasini, QA&RA Director – Qura s.r.l.
Phone:+39 0535 1803050
e-mail:raffaella.tommasini@quramed.com
Fax:+39 0535 1803051
Date Summary Prepared:September 26th, 2023

DEVICE II.

Proprietary Name:Quantum Perfusion Centrifugal Blood Pump CP37 withIntegrated Sensors
Common Name:Centrifugal Blood Pump with Integrated Sensors
Classification Name:Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Regulatory Class:II
Product Code:KFM
Panel:Cardiovascular Devices, Office of Health Technology 2(OHT2) / Division of Health Technology 2 B (CirculatorySupport, Structural and Vascular Devices)

III. PREDICATE AND REFERENCE DEVICES

Predicate device:

Proprietary Name:Quantum Perfusion Centrifugal Blood Pump CP22 withIntegrated Sensors
Common Name:Centrifugal Blood Pump CP22 with Integrated Sensors
Classification Name:Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Regulatory Class:II
Product Code:KFM
Panel:Cardiovascular Devices, Office of Health Technology 2(OHT2) / Division of Health Technology 2 B (CirculatorySupport, Structural and Vascular Devices)
510(k) Number:K201320

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for Qura advanced medical solutions. The logo features a stylized human figure with arms outstretched, enclosed within a curved line. The text "QURA" is written in a bold, sans-serif font, with the words "advanced medical solutions" in a smaller font below it. The logo is blue.

Reference device:

Proprietary Name:Quantum Perfusion Centrifugal Blood Pump CP37
Common Name:Centrifugal Blood Pump CP37
Classification Name:Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Regulatory Class:II
Product Code:KFM
Panel:Cardiovascular Devices, Office of Health Technology 2(OHT2) / Division of Health Technology 2 B (CirculatorySupport, Structural and Vascular Devices)
510(k) Number:K202169

DEVICE DESCRIPTION IV.

The Quantum Perfusion Centrifugal Blood Pump with Integrated Sensors is a standalone single use device intended to pump blood into the extracorporeal circuit and to monitor pressure for periods lasting less than 6 hours. The device is equipped with two sensors integrated in the blood inlet and outlet connectors.

The device is non-toxic, non pyrogenic, sterilized by ethylene oxide and packaged in a single blister. Blood contact surfaces of the device are coated with a stable biocompatible surface to reduce platelet activation and adhesion while preserving platelet function.

V. INTENDED USE / INDICATIONS FOR USE

The Quantum Perfusion Centrifugal Blood Pump CP37 with Integrated Sensors, indicated for use exclusively with Quantum Centrifugal Drive of Spectrum Medical, are intended to pump the blood through an extracorporeal circuit and to monitor pressure for periods lasting less than 6 hours for the purpose of providing either:

  • (i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or
  • (ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.

Devices are intended for adult patients.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed in a partial orbit shape. To the right of the figure is the word "QURA" in a bold, sans-serif font. Below "QURA" is the text "advanced medical solutions" in a smaller, lighter font.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

One additional device model has been introduced in the Centrifugal Blood Pump with Integrated Sensors product family in order to provide to clinicians/users an additional option to perform cardiopulmonary bypass procedures.

New REF code introduced in the portfolio for Quantum Perfusion Centrifugal Blood Pump with Integrated Sensors product family is aligned with information provided in original submission K201320 in terms of general structure (including packaging configuration), principle of operation, intended use , manufacturing and sterilization processes.

Technical specifications have been verified through testing activities performed according to the same internal applicable standards/protocols of the original cleared devices.

Applicable testing activities demonstrated that the proposed device model does not raise any issues of safety and effectiveness as compared to the currently cleared predicate and reference product.

VII. PERFORMANCE DATA

NON-CLINICAL TESTING

The following activities were performed to demonstrate product safety and effectiveness, considering the proposed change and related impact:

  • Update of labeling and Instructions for Use (IFU) according to ISO 15223-1:2021 -Medical Devices - Symbols to Be Used with Medical Device Labels, Labelling, And Information to Be Supplied - Part 1: General Requirements [Recognition Nr. 5-134];
  • Review and Assessment of data available on predicate( K201320) and reference (K202169) devices submission and applicability to new variant introduced in the product family portfolio.

Animal Study

No animal studies have been performed to support changes object of the present Special 510(k).

CLINICAL TESTING

No clinical data have been included in the current 510(k) submission to support substantial equivalence to legally marketed predicate devices.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for Qura Advanced Medical Solutions. The logo features a stylized human figure inside of a partial circle. The text "QURA" is to the right of the figure, and below that is the text "advanced medical solutions" in a smaller font.

VIII. CONCLUSIONS

Considering all changes performed on original devices cleared K201320 as described in Section "Device Description" of current submission, it could be stated that devices under evaluation are identical in terms of intended use and applicable medical technique.

Based on the testing activities, technological characteristics and the indications for use, the proposed devices have been demonstrated to be appropriate for their intended use and are considered substantially equivalent to Qura's own original devices.

§ 870.4360 Nonroller-type blood pump.

(a)
Nonroller-type cardiopulmonary and circulatory bypass blood pump —(1)Identification. A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (
i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
(2)
Classification —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
Nonroller-type temporary ventricular support blood pump —(1)Identification. A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
Classification. Class III (premarket approval).(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.