Search Results
Found 2 results
510(k) Data Aggregation
(139 days)
The Quantum Perfusion Centrifugal Blood Pump CP22NG with Integrated Sensors indicated for use exclusively with the Quantum Centrifugal Drive is intended to pump the blood through an extracorporeal circuit and to monitor pressure for periods lasting less than 6 hours for the purpose of providing either:
(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or
(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
Device is intended for adult patients.
The Quantum Perfusion Centrifugal Blood Pump CP22NG with Integrated Sensors (CP22V-NG) is a standalone, single-use device intended for use in medical procedures requiring extracorporeal circulation. The device is designed to pump blood via centrifugal force through an extracorporeal circuit for periods lasting up to six (6) hours.
The device is non-toxic, non-pyrogenic, EtO-sterilized, and packaged in a single blister. Blood contact surfaces are coated with a stable, biocompatible compound to reduce platelet activation and adhesion while preserving platelet function.
The furnished document is a 510(k) clearance letter for a medical device which is a physical blood pump, not an AI/software as a medical device (SaMD). Therefore, the questions related to AI/SaMD testing methodologies (such as sample size for training/test sets, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable here.
However, I can extract the relevant information regarding the performance data and the conclusion of the substantial equivalence for this physical device.
1. Table of Acceptance Criteria and Reported Device Performance
The document states that performance testing was conducted according to the special controls of 21 CFR 870.4360 and confirmed the subject device is equivalent to the predicate. While specific numerical acceptance criteria and precise performance data are not explicitly detailed in this summary, the categories of tests performed indicate the areas where the device's performance was evaluated against established benchmarks (likely derived from the predicate device's performance or regulatory standards).
Acceptance Criteria Category | Reported Device Performance (Implied) |
---|---|
Operating Parameters | Confirmed equivalence to predicate. |
Dynamic Blood Damage | Confirmed equivalence to predicate. |
Heat Generation | Confirmed equivalence to predicate. |
Air Entrapment | Confirmed equivalence to predicate. |
Mechanical Integrity | Confirmed equivalence to predicate. |
Durability/Reliability | Confirmed equivalence to predicate. |
Biocompatibility | Validated by predicate device data. |
EtO Sterilization | Validated by predicate device data. |
Product Shelf-Life | Validated by predicate device data. |
Packaging Validation | Validated by predicate device data. |
2. Sample Size Used for the Test Set and Data Provenance
This information is not provided in a quantifiable manner (e.g., number of units tested).
- Sample Size: Not specified. The document states "Performance testing...was performed with the subject device (CP22V-NG)." This implies a sufficient number of devices were tested to draw conclusions, but the exact count is not given.
- Data Provenance: The tests were likely conducted by the manufacturer, Spectrum Medical S.r.l., as part of their 510(k) submission. The document lists the company's address in Mirandola, MO, Italy. It's a retrospective study from the perspective of the submission, as the tests were completed before the submission.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This is not applicable as the device is a physical blood pump, not an AI/SaMD requiring human expert annotation for ground truth. The "ground truth" for this type of device would be established through engineering specifications, validated test methods, and performance against established physiological or mechanical standards as outlined in the special controls (21 CFR 870.4360).
4. Adjudication Method for the Test Set
Not applicable for a physical device. Performance is measured objectively through lab tests and engineering evaluation, not through subjective reviewer adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is a physical medical device, not an AI-assisted diagnostic tool designed to be used by human readers.
6. Standalone Performance
Yes, standalone performance testing was conducted. The non-clinical performance tests listed (Operating Parameters, Dynamic Blood Damage, Heat Generation, Air Entrapment, Mechanical Integrity, and Durability/Reliability) directly evaluate the device's inherent performance characteristics independently. The device's performance was then compared to that of the predicate device to establish substantial equivalence.
7. Type of Ground Truth Used
The "ground truth" for this device's performance evaluation is based on:
- Engineering Specifications: The device is designed to meet specific technical parameters (e.g., flow rate, pressure monitoring accuracy, heat generation limits).
- Regulatory Standards: Performance was measured against the special controls outlined in 21 CFR 870.4360, which define acceptable parameters for non-roller type blood pumps.
- Predicate Device Performance: The primary method for proving substantial equivalence was demonstrating that the subject device's performance was equivalent to that of its predicate device, which has a proven safety and effectiveness record.
8. Sample Size for the Training Set
Not applicable. This is a physical device, not an AI/Machine Learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, for the reason stated above.
Ask a specific question about this device
(422 days)
The Quantum Perfusion Centrifugal Blood Pump CP20, indicated for use exclusively with Quantum Centrifugal Drive of Spectrum Medical, is intended to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:
(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or
(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava
Device is intended for adolescent patients (i.e., patients greater than 12 through 21 years of age)
The Quantum Perfusion Centrifugal Blood Pump CP20 with Integrated Sensors, indicated for use exclusively with Quantum Centrifugal Drive of Spectrum Medical, is intended to pump the blood through an extracorporeal circuit and to monitor pressure for periods lasting less than 6 hours for the purpose of providing either:
Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on (i) the heart or great vessels; or
Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway (ii) necessary for open surgical procedures on the aorta or vena cava.
Device is intended for adolescent patients (i.e., patients greater than 12 through 21 years of age)
Quantum Perfusion Centrifugal Blood Pump CP20 and Quantum Perfusion Centrifugal Blood Pump CP20 with Integrated Sensors are standalone single use device intended to pump blood into the extracorporeal circulation circuit and to monitor pressure for periods lasting less than 6 hours.
Quantum Perfusion Centrifugal Blood Pump CP20 with Integrated Sensors is equipped with two sensors integrated in the blood inlet and outlet connectors.
Devices are non-toxic, non-pyrogenic, sterilized by ethylene oxide and packaged in a single blister. Blood contact surfaces of the device are coated with a stable biocompatible surface to reduce platelet activation and adhesion while preserving platelet function.
This document describes the Qura S.r.l. Quantum Perfusion Centrifugal Blood Pump CP20 and CP20 with Integrated Sensors (K220842).
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state "acceptance criteria" with numerical targets in a table format for each performance test. Instead, it lists the types of performance tests conducted to demonstrate substantial equivalence to the predicate and reference devices, implying that the acceptance criterion for each test was to perform comparably or within expected safety limits as per relevant standards and guidance.
Below is a table summarizing the reported device characteristics relevant to performance and the tests conducted. The "Acceptance Criteria" column will be inferred as "Compliance with applicable standards and demonstration of substantial equivalence to predicate/reference devices."
Performance Characteristic/Test Area | Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|---|
Operating Parameters | Compliance with applicable special controls per 21 CFR 870.4360 and ISO 18242; substantial equivalence to predicate/reference. | In-vitro performance tests performed. Device designed to pump blood, with CP20 with Integrated Sensors also monitoring pressure. Maximum blood flow rate: 1.5 L/min. Maximum pump operating pressure: 550 mmHg. (Compared favorably to predicate and reference) |
Dynamic Blood Damage | Compliance with applicable special controls per 21 CFR 870.4360 and ISO 18242; substantial equivalence to predicate/reference. | In-vitro performance tests performed. Blood contact surfaces coated with phosphorylcholine-based biocompatible surface to reduce platelet activation and adhesion while preserving platelet function. |
Heat Generation | Compliance with applicable special controls per 21 CFR 870.4360 and ISO 18242; substantial equivalence to predicate/reference. | In-vitro performance tests performed. |
Air Entrapment | Compliance with applicable special controls per 21 CFR 870.4360 and ISO 18242; substantial equivalence to predicate/reference. | In-vitro performance tests performed. |
Mechanical Integrity | Compliance with applicable special controls per 21 CFR 870.4360 and ISO 18242; substantial equivalence to predicate/reference. | In-vitro performance tests performed. |
Durability/Reliability | Compliance with applicable special controls per 21 CFR 870.4360 and ISO 18242; substantial equivalence to predicate/reference. | In-vitro performance tests performed. Intended for periods lasting less than 6 hours. |
Product Shelf Life | Compliance with applicable standards and substantial equivalence to predicate/reference. | Evaluation performed. |
Sterilization Process (EtO) | Validation per ISO 11135 (Recognition Nr.: 14-529). | Validation performed, device is EtO sterile. |
Packaging Validation | Validation per ISO 11607-1 (Recognition Nr. 14-530). | Validation performed. |
Biocompatibility | Compliance with ISO 10993-1:2018 (Recognition Nr. 2-258) and FDA Guidance "Use of International Standard ISO 10993-1..." | Evaluation performed, including mandatory animal tests. Devices are non-toxic. |
Integrated Sensors (for CP20 with S.) | Ability to measure pressure at inlet and outlet; substantial equivalence to reference device (CP22 with Integrated Sensors). | Equipped with two sensors positioned on the blood inlet and outlet connectors, both able to measure pressure at the same time. |
Priming Volume | Comparable to predicate device (14 mL for PediMag); appropriate for intended patient population. | 20 mL. |
Intended Use Duration | Less than 6 hours for cardiopulmonary bypass or temporary circulatory bypass. (Matches predicate and reference.) |
Ask a specific question about this device
Page 1 of 1