K Number
K233010

Validate with FDA (Live)

Device Name
Beagank 4T Plus
Manufacturer
Date Cleared
2023-11-21

(60 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation.

Device Description

The BEAGANK 4T PLUS is a cosmetic device which is intended for use on the face and body skin. The device contains a rechargeable lithium ion battery and is provided with a charger. The device has four electrode metal pads on the head that deliver electrical impulses to the face and body skin and provide four different operational modes. The device operates by applying Transcutaneous Electrical Nerve Stimulation to the face and body skin. The electrodes that are positioned on the skin must be covered with a wet cotton pad. Purified water or commercially available cosmetic formulation liquid can be used as a conductive media to generate the electric pulses. Two attachments are provided with the BEAGANK 4T PLUS: Scalp attachment, Small attachment.

AI/ML Overview

The provided document is a 510(k) summary for the Beagank 4T Plus device. It does not contain acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of clinical performance or diagnostic accuracy.

Instead, the document focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and compliance with safety standards. It explicitly states: "No new clinical studies have been submitted as part of this Premarket Notification."

Therefore, I cannot provide the requested information regarding acceptance criteria for clinical performance and a study proving those criteria are met, as such information is not present in the provided text.

However, I can extract information related to non-clinical performance and safety testing which serves as the basis for the FDA clearance in this case.

Here's what can be extracted:

1. A table of acceptance criteria and the reported device performance:

Based on the "Summary of Non-Clinical / Test Data," the device performance is demonstrated by compliance with specific safety standards. The "acceptance criteria" here are essentially passing these standards.

Acceptance Criteria (Standard)Reported Device Performance (Compliance)
IEC 60601-1 (Electrical Safety and Essential Performance)Complies
IEC 60601-1-2 (Electromagnetic Compatibility)Complies
ISO 10993-5 (Biocompatibility - Cytotoxicity)Complies
ISO 10993-10 (Biocompatibility - Irritation and Sensitization)Complies
ISO 10993-23 (Biocompatibility - Irritation for medical devices and in vitro test method for skin irritation)Complies

2. Sample size used for the test set and the data provenance:
Not applicable for this type of submission. The tests are for device safety and electrical performance, typically conducted on a sample of the manufactured device. Details on the specific number of units tested and data provenance (country of origin, retrospective/prospective) are not provided in this 510(k) summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This is a non-clinical safety and performance evaluation, not a clinical study involving expert ground truth for interpretation. The "ground truth" is adherence to established engineering and biocompatibility standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. This is not a clinical study.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No. The document explicitly states "No new clinical studies have been submitted." This is not an AI-assisted diagnostic device, and thus no MRMC study would be relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a physical electrical stimulation device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable for clinical ground truth. For the non-clinical tests, the "ground truth" is the established requirements and limits defined within the cited IEC and ISO standards.

8. The sample size for the training set:
Not applicable. There is no machine learning or AI component that would require a training set.

9. How the ground truth for the training set was established:
Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 21, 2023

Belega Inc. % Nana Nagashita Regulatory Consultant Ken Block Consulting Co., Ltd. 1-9-9 Tsukiji Chuo-ku, Tokyo 104-0045 Japan

Re: K233010

Trade/Device Name: Beagank 4T Plus Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: September 15, 2023 Received: September 22, 2023

Dear Nana Nagashita:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Heather L. Dean -S

Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices

{2}------------------------------------------------

OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233010

Device Name BEAGANK 4T PLUS

Indications for Use (Describe)

The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

Owner/Submitter:Belega Co., Ltd.1-4-20 Sonezaki Shinchi, Kita-kuOsaka-shi, Osaka, Japan 530-0002
Contact Person:Ken Block Consulting Co., Ltd.Ms. Nana NagashitaRegulatory consultant1-9-9 Tsukiji, Chuo-kuTokyo, Japan 104-0045TEL: +81-3-6260-6688FAX: +81-3-6260-6689nagashita@kenblockconsulting.com
Date Prepared:September 20, 2023
Submission Type:Traditional 510(k)
Proposed DeviceManufacturer: BELEGA Co., Ltd.Trade Name: BEAGANK 4T PLUSCommon Name: Aesthetic TENS DeviceClassification Name: Transcutaneous Electrical Nerve Stimulator for Pain ReliefProduct Code: NFORegulatory Standard: 21 CFR 882.5890
Predicate Device:510(k) Number: K191951Manufacturer: Avazzia, Inc.Trade Name: Avazzia OTC TENS for Aesthetics, model BEST-AV1: EZZI LIFT DeviceCommon Name: Aesthetic TENS DeviceClassification Name: Transcutaneous Electrical Nerve Stimulator for Pain ReliefProduct Code: NFORegulatory Standard: 21 CFR 882.5890
Reference Device:510(k) Number: K011935Manufacturer: Salton Inc.Trade Name: Rejuvenique® system, Model RJV-10Common Name: Aesthetic TENS DeviceClassification Name: Transcutaneous Electrical Nerve Stimulator for Pain ReliefProduct Code: NFO, GYB (Media, Conductive)Regulatory Standard: 21 CFR 882.5890, 21 CFR 882.1275 (Media, Conductive)
Reference Device:510(k) Number: K181008Manufacturer: Carol Cole Company dba NuFACETrade Name: NuFACE TrinityCommon Name: Aesthetic TENS DeviceClassification Name: Transcutaneous Electrical Nerve Stimulator for Pain ReliefProduct Code: NFO

{5}------------------------------------------------

BEAGANK 4T PLUS is a cosmetic device which is intended for use on the face and body Device Description: skin. The device contains a rechargeable lithium ion battery and is provided with a charger. The device has four electrode metal pads on the head that deliver electrical impulses to the face and body skin and provide four different operational modes.

The device operates by applying Transcutaneous Electrical Nerve Stimulation to the face and body skin. The electrodes that are positioned on the skin must be covered with a wet cotton pad. Purified water or commercially available cosmetic formulation liquid can be used as a conductive media to generate the electric pulses. Two attachments are provided with the BEAGANK 4T PLUS:

  • Scalp attachment -
  • Small attachment

Statement of The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic Intended Use: use including facial and neck stimulation or body skin stimulation.

Summary of The major differences between the proposed and predicate devices are summarized in the Technological tables below: Characteristics:

Table 1: Basic characteristics
New DevicePredicate DeviceReference DeviceReference Device
Trade NameBEAGANK 4TPLUSEZZI-LIFT DeviceRejuveniqueNuFACE Trinity
510(k)K233010K191951K011935K181008
Indication for UseThe BEAGANK4T PLUS is ahandheld portabledevice for over-the-counteraesthetic useincluding facialand neckstimulation orbody skinstimulation.The Avazzia OTCTENS foraesthetics, modelBEST-AV1™:EZZI-LIFT™Device is indicatedfor over-the-counter aestheticuse includingfacial and neckstimulation orbody skinstimulation.Rejuveniquesystem is indicatedfor cosmetic use.The NuFACETrinity is intendedfor facial and neckstimulation and isindicated for over-the-countercosmetic use.
Device ClassClass IIClass IIClass IIClass II
Operating PowerInternalrechargeableLithium-ionbatteryTwo 1.5 V AAbatteries9 V batteryInternalrechargeableLithium-ionbattery
Method of Line CurrentIsolationType BFUnknownUnknownType BF
PatientLeakageCurrentNormalDC: <1μΑAC: <1μΑTested per IEC60601-1UnknownUnknownN/A - Batteryoperated
SinglefaultDC: <1μΑAC: <1μΑTested per IEC60601-1UnknownUnknownN/A - Batteryoperated
Ave. DC currentthrough electrodeswhen device is on butno pulse are beingapplied (μΑ)<1μΑUnknownUnknownUnknown
OutputChannelsMethodofChannelIsolationN/A - 1 outputchannelUnknownElectrode groupselected by relaysN/A - 1 outputchannel
Regulated Current orRegulated VoltageVoltageUnknownVoltageBoth
Automatic OverloadTripNot required dueto circuit designUnknownNoNot required dueto circuit design
Automatic No-LoadTripNot required dueto circuit designUnknownNoYes
Automatic Shut OffYesYesYesYes
Patient OverrideControlYesUnknownYesYes
Weight136g153g(5.4 ounces)80 g9 oz. withoutcharging base
Number of outputmodes4 modes:maximum 5minutes each4 modes: 60minutes1Unknown
Low Battery IndicatorYesYesYesYes
SterilityNon-SterileNon-SterileNon-SterileNon-Sterile
Electrical safetyIEC 60601-1Medicalelectricalequipment -Part 1 Generalrequirementsfor basicsafety andessentialperformanceIEC 60601-1-2MedicalelectricalIEC 60601-1Medicalelectricalequipment -Part 1 Generalrequirementsfor safetyIEC 60601-1-2Medicalelectricalequipment -Part 1-2:GeneralUnknownIEC 60601-1IEC 60601-1-2
Part 1-2:Generalrequirementsfor basicsafety andessentialperformance –CollateralStandard:Electromagnetic disturbances–Requirementsand testsfor basicsafety andessentialperformance –CollateralStandard:Electromagnetic disturbances–Requirementsand testsIEC 60601-2-10 medicalelectricalequipment –Part 2:10;Particularrequirementsfor the safetyof nerve andmusclestimulators
Housing Materials andConstructionABSPCBs insideplastic casehousingABS Plastic, snaplatch assemblyThermoplastic
AccessoriesSmall and ScalpattachmentBuilt-in, Y, Brush,PencilUnknownUnknown

{6}------------------------------------------------

{7}------------------------------------------------

Table 2: Technological Characteristics

Output Specifications: Mode 1
New DevicePredicate DeviceReference DeviceReference Device
Trade NameBEAGANK 4TPLUSEZZI-LIFT DeviceRejuveniqueNuFACE Trinity
Indications for UseThe BEAGANK 4TPLUS is a handheldportable device forover-the-counteraestheticuseincluding facial andneck stimulation orbodyskinstimulation.The Avazzia OTCTENS for aesthetics,modelBEST-AV1TM:EZZI-LIFTTM Device isindicated for over-the-counter aestheticuse including facialand neck stimulationor bodyskinstimulation.Rejuvenique systemis indicatedforcosmetic use.The NuFACETrinity is intendedfor facial and neckstimulation and isindicated for over-the-counter cosmeticuse.

{8}------------------------------------------------

Waveform ShapeRectangle, biphasic asymmetricPositive square wave followed by a damped sinusoidal waveform of variable duration depending on damping and body loadingPulsed biphasic, Rectangular (+phase), Spike (- phase)Pulsed biphasic modulated square wave
Max output voltage(+/-15%)(+/-20%)(+/-10%)28VDC
@ 500Ω127mV (0.12V)-42V18.8Vnot publicly available
@2kΩ515mV (0.51V)-122V24.8Vnot publicly available
@ 10kΩ1.83V-348V28.0Vnot publicly available
Max output current(+/-15%)(+/-20%)(+/-10%)400μA (0.4mA)
@ 500Ω0.19mA363μΑ(0.363 mA)37.6mAnot publicly available
@2kΩ0.18mA117μΑ (0.117 mA)12.4mAnot publicly available
@ 10kΩ0.16mA38μΑ (0.038 mA)2.8mAnot publicly available
Duration of primary pulse at 500ΩPositive: 130µs (0.13msec)Negative: 130µs (0.13msec)0.5msecnot publicly available60msec
Pulse Duration (Cycle) at 500Ω265µs (0.265msec)1.1msec300msec (+phase)124.7msec (-phase, exponential)60msec
Frequency3.80kHz15-121Hz8Hz fixed8.3Hz
Net Charge per pulse at 500Ω0.025µC4µC0µCnot publicly available
Max phase charge0.059µC10µC11.3µCnot publicly available
Max current density at 500Ω- Standard: 0.15mA/cm²- Scalp: 0.53mA/cm²- Small:Built-in, Y, Brush: 800μA/cm²(0.80mA /cm²)46.4mA/cm²not publicly available
1.36mA/cm2Pencil:
19,000μA/cm2(19mA/cm2)
With a wet cottonpad (Purified water)
- Standard:0.15mA/cm2
- Scalp:0.53mA/cm2Small: 1.36mA/cm2
Max power densityat 500Ω (Smallestelectrode)8.51μW/cm2Built-in, Y, Brush:Average500μW/cm22.31mW/cm2(2,310μW/cm2)not publiclyavailable
- With a wetcotton pad11.39μW/cm2Pencil: Average3,500μW/cm2
Max. AverageCurrent (mA)0.105not publiclyavailablenot publiclyavailablenot publiclyavailable
Burst mode
Pulse per burstN/AN/A16020
Burst per secondN/AN/A1/240 (per electrodegroup)8.3
Burst durationN/AN/A20sec2.4
Duty cycleN/AN/A1/1220.2sec
On Time (seconds)N/AN/A20sec/electrodegroupnot publiclyavailable
Off Time (seconds)N/AN/Anot publiclyavailablenot publiclyavailable
Output Specifications: Mode 2
New DevicePredicate DeviceReference DeviceReference Device
Trade NameBEAGANK 4TPLUSEZZI-LIFT DeviceRejuveniqueNuFACE Trinity
Indications for UseThe BEAGANK 4TPLUS is a handheldportable device forover-the-counteraesthetic useincluding facial andThe Avazzia OTCTENS for aesthetics,model BEST-AV1TM: EZZI-LIFTTM Device isindicated for over-Rejuvenique systemis indicated forcosmetic use.The NuFACETrinity is intendedfor facial and neckstimulation and isindicated for over-

{9}------------------------------------------------

{10}------------------------------------------------

neck stimulation orbody skin stimulation.the-counter aestheticuse including facialand neck stimulationor body skin stimulation.the-counter cosmeticuse.
Waveform ShapeRectangle, biphasicasymmetricPositive squarewave followed by adamped sinusoidalwaveform ofvariable durationdepending ondamping and bodyloadingPulsed biphasic,Rectangular(+phase), Spike (-phase)Pulsed biphasicmodulated squarewave
Max output voltage(+/-15%)(+/-20%)(+/-10%)28VDC
@ 500Ω110mV (0.11V)-42V18.8Vnot publiclyavailable
@2kΩ507mV (0.50V)-122V24.8Vnot publiclyavailable
@ 10kΩ1.87V-348V28.0Vnot publiclyavailable
Max output current(+/-15%)(+/-20%)(+/-10%)400μA (0.4mA)
@ 500Ω0.45mA363μA (0.363mA)37.6mAnot publiclyavailable
@2kΩ0.39mA117μA (0.117mA)12.4mAnot publiclyavailable
@ 10kΩ0.24mA38μA (0.038mA)2.8mAnot publiclyavailable
Duration of primarypulse at 500ΩPositive: 130µs(0.13msec)Negative: 130µs(0.13msec)0.5msecnot publiclyavailable60msec
Pulse Duration(Cycle) at 500Ω265µs (0.265msec)1.1msec300msec (+phase)124.7msec (-phase,exponential)60msec
Frequency3.80kHz15-121Hz8Hz fixed8.3Hz
Net Charge per pulseat 500Ω0.025μC4μC0μCnot publiclyavailable
Max phase charge$0.025\mu C$$10\mu C$$11.3\mu C$not publicly available
Max current density at 500ΩStandard:$0.36mA/cm^2$Scalp:$1.25mA/cm^2$Small:$3.21mA/cm^2$Built-in, Y, Brush:$800\mu A/cm^2$( $0.8mA/cm^2$ )Pencil:$19,000\mu A/cm^2$( $19mA/cm^2$ )$46.4mA/cm^2$not publicly available
With a wet cotton pad (Purified water)Standard:$0.10mA/cm^2$Scalp: $0.36 mA/cm^2$Small: $0.92 mA/cm^2$
Max power density at 500Ω (Smallest electrode)$17.46\mu W/cm^2$Built-in, Y, Brush:Average $500\mu W/cm^2$$2.31mW/cm^2$( $2,310\mu W/cm^2$ )not publicly available
With a wet cotton pad$33.87\mu W/cm^2$Pencil: Average $3,500\mu W/cm^2$
Max. Average Current (mA)0.105not publicly availablenot publicly availablenot publicly available
Burst mode
Pulse per burstN/AN/A16020
Burst per secondN/AN/A1/240 (per electrode group)8.3
Burst durationN/AN/A20sec2.4
Duty cycleN/AN/A1/1220.2sec
On Time (seconds)N/AN/A20sec/electrode groupnot publicly available
Off Time (seconds)N/AN/Anot publicly availablenot publicly available
Output Specifications: Mode 3
New DevicePredicate DeviceReference DeviceReference Device

{11}------------------------------------------------

{12}------------------------------------------------

Trade NameBEAGANK 4TPLUSEZZI-LIFT DeviceRejuveniqueNuFACE Trinity
Indications for UseThe BEAGANK 4TPLUS is a handheldportable device forover-the-counteraestheticuseincluding facial andneck stimulation orbodyskinstimulation.The Avazzia OTCTENS for aesthetics,model BEST-AV1TM. EZZI-LIFT™ Device isindicated for over-the-counter aestheticuse including facialand neck stimulationor body skinstimulation.Rejuvenique systemis indicated forcosmetic use.The NuFACETrinity is intendedfor facial and neckstimulation and isindicated for over-the-counter cosmeticuse.
Waveform ShapeRectangle, biphasicsymmetricalPositive squarewave followed by adamped sinusoidalwaveform ofvariable durationdepending ondamping and bodyloadingPulsed biphasic,Rectangular(+phase), Spike (-phase)Pulsed biphasicmodulated squarewave
Max output voltage(+/-15%)(+/-20%)(+/-10%)28VDC
@ 500Ω16.6V-4218.8Vnot publiclyavailable
@2kΩ21.0V-12224.8Vnot publiclyavailable
@10kΩ22.5V-34828.0Vnot publiclyavailable
Max output current(+/-15%)(+/-20%)(+/-10%)400μA
@ 500Ω30.8mA363μA(0.363 mA)37.6mAnot publiclyavailable
@2kΩ8.52mA117μA (0.117 mA)12.4mAnot publiclyavailable
@ 10kΩ2.26mA38μA (0.038 mA)2.8mAnot publiclyavailable
Duration of primarypulse at 500ΩPositive: 80μs(0.08msec)Negative: 80μs(0.08msec)0.5msecnot publiclyavailable60msec
Pulse Duration(Cycle) at 500Ω274μs (0.247msec)1.1msec300msec (+phase)124.7msec (-phase,exponential)60msec
Frequency3.65kHz15-121Hz8Hz fixed8.3Hz
Net Charge per pulseat 500Ω-1.3μC4μC0μCnot publiclyavailable
Max phase charge2.2μC10μC11.3μCnot publiclyavailable
Max current densityat 500Ω- Standard:24.4mA/cm²- Scalp:85.6mA/cm²- Small: 220mA/cm²With a wet cottonpad (Purified water)- Standard:4.26mA/cm²- Scalp:28.21mA/cm²Small:38.43mA/cm²Built-in, Y, Brush:800μA/cm²(0.8mA/cm²)Pencil:19,000μA/cm²(19mA/cm²)46.4mA/cm²not publiclyavailable
Max power densityat 500Ω(Smallestelectrode)106638.40μW/cm²With a wetcotton pad1124.0µW/cm²Built-in, Y, Brush:Average500μW/cm²Pencil: Average3,500μW/cm²2.31mW/cm²(2,310µW/cm²)not publiclyavailable
Max. AverageCurrent (mA)0not publiclyavailablenot publiclyavailablenot publiclyavailable
Burst mode
Pulse per burstN/AN/A16020
Burst per secondN/AN/A1/240 (per electrodegroup)8.3
Burst durationN/AN/A20sec2.4
Duty cycleN/AN/A1/1220.2sec
On Time (seconds)N/AN/A20sec/electrodegroupnot publiclyavailable
Off Time (seconds)N/AN/Anot publiclyavailablenot publiclyavailable
Output Specifications: Mode 4
New DevicePredicate DeviceReference DeviceReference Device
Trade NameBEAGANK 4TPLUSEZZI-LIFT DeviceRejuveniqueNuFACE Trinity
Indications for UseThe BEAGANK 4TPLUS is a handheldportable device forover-the-counteraesthetic useincluding facial andneck stimulation orbody skinstimulation.The Avazzia OTCTENS for aesthetics,model BEST-AV1TM: EZZI-LIFTTM Device isindicated for over-the-counter aestheticuse including facialand neck stimulationor body skinstimulation.Rejuvenique systemis indicated forcosmetic use.The NuFACETrinity is intendedfor facial and neckstimulation and isindicated for over-the-counter cosmeticuse.
Waveform ShapeRectangle,Biphasic symmetricPositive squarewave followed by adamped sinusoidalwaveform ofvariable durationdepending ondamping and bodyloadingPulsed biphasic,Rectangular(+phase), Spike (-phase)Pulsed biphasicmodulated squarewave
Max output voltage(+/-15%)(+/-20%)(+/-10%)28VDC
@ 500Ω15.3V-42V18.8Vnot publiclyavailable
@ 2kΩ21.2V-122V24.8Vnot publiclyavailable
@ 10kΩ22.9V-348V28.0Vnot publiclyavailable
Max output current(+/-15%)(+/-20%)(+/-10%)400μA
@ 500Ω31.4mA363μA(0.363 mA)37.6mAnot publiclyavailable
@ 2kΩ8.60mA117μA (0.117 mA)12.4mAnot publiclyavailable
@ 10kΩ2.15mA38μA (0.038 mA)2.8mAnot publiclyavailable
Duration of primarypulse at 500ΩPositive: 92µs(0.092msec)Negative: 92µs(0.092msec)0.5msecnot publiclyavailable60msec
PulseDuration(Cycle) at 500Ω610µs (0.61msec)1.1msec300msec (+phase)124.7msec (-phase,exponential)60msec
Frequency1.64kHz15 to 121 Hz8 Hz fixed8.3Hz
Net Charge per pulseat 500Ω0µC (biphasicsymmetrical)4μC0μCnot publiclyavailable
Max phase charge0.29µC10µC11.3µCnot publiclyavailable
Max current densityat 500Ω-Standard:25.0mA/cm²Head:87.2mA/cm²Small:224.3mA/cm²With a wet cottonpad (Purified water)-Standard:3.13mA/cm²Scalp:10.97mA/cm²Small:28.21mW/cm²Built-in, Y, Brush:800μA/cm²(0.8mA/cm²)Pencil: 19,000μA/cm²(19mA/cm²)46.4 mA/cm²not publiclyavailable
Max power densityat 500Ω (Smallestelectrode)-51775.54μW/cm²With a wetcotton pad556.96μW/cm²Built-in, Y, Brush:Average500μW/cm²Pencil: Average3,500μW/cm²2.31mW/cm²(2,310µW/cm²)not publiclyavailable
Max. AverageCurrent (mA)0not publiclyavailablenot publiclyavailablenot publiclyavailable
Burst mode
Pulse per burstN/AN/A16020
Burst per secondN/AN/A1/240 (per electrodegroup)8.3
Burst durationN/AN/A20sec2.4
Duty cycleN/AN/A1/1220.2sec
On Time (seconds)N/AN/A20sec/electrodegroupnot publiclyavailable
Off Time (seconds)N/AN/Anot publiclyavailablenot publiclyavailable

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

Clinical Test Data: No new clinical studies have been submitted as part of this Premarket Notification.

Tests were performed on the device which demonstrated that the device is safe and Summary of Non-Clinical / effective, performs comparably to and is substantially equivalent to the predicate device. Test Data: Testing confirmed that the device complies with safety standards for Electrical Safety and Electromagnetic Compatibility testing, specifically IEC standards 60601-1 and 60601-1-2.

Testing confirmed that the device electrodes complies with safety standards for Biocompatibility testing, specifically ISO standards 10993-5, 10993-10, and 10993-23.

Company considers the BEAGANK 4T PLUS device to be substantially equivalent to the Conclusion: predicate device and reference devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).