K Number
K191951

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-04-27

(645 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Avazzia OTC TENS for aesthetics, model BEST-AV1™: EZZI-LIFT™ Device is indicated for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation.

Device Description

Candidate device is a handheld, AA battery-operated portable device for use in the home or clinic. It has pre-set modes and is for over-the-counter use in the home or a clinical setting. The candidate device uses transcutaneous electrical stimulation (TENS) for aesthetic use for facial, neck, and body stimulation. The EZZI-LIFT device has built-in electrodes. Accessories, listed below, are also available to be used with the device.

AI/ML Overview

The Avazzia OTC TENS for Aesthetics, model BEST-AV1: EZZI-LIFT Device did not have a clinical study that proves the device meets acceptance criteria for a specific clinical outcome. Instead, the device's substantial equivalence was established through non-clinical bench testing, performance testing, and a comparison of its technical characteristics to predicate devices.

Here's the breakdown based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not present a formal "acceptance criteria" table with specific quantitative thresholds for clinical performance. Instead, it demonstrates substantial equivalence by comparing the technical specifications of the candidate device (EZZI-LIFT) to its predicate devices (NuFACE Trinity and NuBODY Skin Toning Device) and by confirming compliance with relevant safety standards.

The comparison table provided in the document highlights the following characteristics:

CharacteristicUnitsCandidate (EZZI-LIFT)Primary Predicate (NuFACE Trinity)Secondary Predicate (NuBody)
Device NameEZZI-LIFTNuFACE TrinityNuBody
510KK191951K181008K171588
ManufacturerAvazzia, Inc.Carol Cole Company (dba NuFACE)Carol Cole Company (dba NuFACE)
Indications for UseOver-the-counter aesthetic use including facial and neck stimulation or body skin stimulation.Facial and neck stimulation; over-the-counter cosmetic use.Body skin stimulation; over-the-counter cosmetic use.
Anatomic SitesFace, neck, and bodyFace and neckBody
Number of Output Modesnumber411
Low Battery IndicatorYesYesYes
Automatic shut off(minutes)60205
Compliance with Voluntary StandardsYes/NoIEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, IEC 62366IEC 60601-1, IEC 60601-1-2IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60529, ISO 14971, IEC 60601-1-6, IEC 62366
Compliance with 21 CFR 898Yes / NoYesYesYes
Weightounces5.4910 to 14
Dimensions [W x H x D](in.)2.6" X 4.7" X 1.35"2.8" x 5.1" x 1.3"2" x 6.5" x 6.0"
Housing Materials and ConstructionPCBs inside plastic case housingThermoplasticThermoplastic
User Interface DisplayLEDs and switchesnot publicly availablenot publicly available
Energy typeElectric stimulationElectric stimulationElectric stimulation
Number, Size, and Type of BatteriesTwo 1.5 V AA batteriesInternal rechargeable Lithium-ion batteryInternal rechargeable Lithium-ion battery
Stimulation Output Specifications
Waveform Shapepositive square wave followed by a damped sinusoidal waveform of variable durationpulsed biphasic modulated square wavemonophasic voltage modulated square wave
Max output voltage (+/- 20%) at 500 ΩV-422828
Max output voltage (+/- 20%) at 2,000 ΩV-122not publicly availablenot publicly available
Max output voltage (+/- 20%) at 10,000 ΩV-348not publicly availablenot publicly available
Max output current (+/-20%) at 500 ΩμA363400900
Max output current (+/-20%) at 2,000 ΩμA117not publicly availablenot publicly available
Max output current (+/-20%) at 10,000 ΩμA38not publicly availablenot publicly available
Duration of primary (depolarizing phase)mSec0.56060
Pulse DurationmSec1.16060
FrequencyHz15 to 1218.38.3
Net Charge per pulse at 500 ΩμC4not publicly available54
Max phase chargeμC10not publicly availablenot publicly available
Max current density at 500 Ω (Built-in, Y, Brush)μΑ/cm²800419468
Max current density at 500 Ω (Pencil)μΑ/cm²19,000
Max average power density at 500 Ω (Built-in, Y, Brush)μW/cm²500not publicly available4,180
Max average power density at 500 Ω (Pencil)μW/cm²3,500
Burst mode (a) pulse per burstnumber12020
Burst mode (b) burst per secondnumbern/a8.38.3
Burst mode (c) burst duration(sec)n/a2.42.4
Burst mode (d) duty cyclen/a20.2sec20.2sec
On Time (per second)seconds0.1not publicly availablenot publicly available
Off Time (per second)seconds0.9not publicly availablenot publicly available
Electrodes (a) materialsStainless steel 316Chrome-platedChrome-plated
(b) electroconductive median/aConductive gelConductive gel
(c) electrode-to-skin impedance rangeless than 15not publicly availablenot publicly available
(d) max duration of use (same as device shut off)(minutes)602020
(e) conductive surface area (built in rectangular)cm21.94--
(e) conductive surface area (spherical)cm25.60not publicly availablenot publicly available
(e) conductive surface area (brush/comb)cm20.56not publicly availablenot publicly available
(e) conductive surface area (small circular, pencil-like)cm20.023not publicly availablenot publicly available

The acceptance criterion, though not explicitly stated as a pass/fail threshold, is implied to be that the device's technical characteristics, particularly regarding stimulation output and safety features, fall within the range of equivalence to predicate devices and comply with accepted safety standards (e.g., ANSI/AAMI NS4:2013/(R)2017). The document explicitly states: "The candidate devices fall within the same range of equivalence for performance and energy output for meeting the ANSI/AAMI NS4:2013/(R)2017 standard requirements for safety; and therefore, they do not raise new questions of safety."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily describes non-clinical bench testing and usability studies, not traditional clinical trials with patient test sets. Therefore, terms like "sample size for the test set" (in a clinical context) and "data provenance" (country of origin, retrospective/prospective) are not directly applicable.

  • Test Set (Non-clinical): The "test set" for performance testing consisted of "a production equivalent of the candidate device." The specific number of devices tested is not mentioned.
  • Usability Study: A usability study was performed, but the number of participants or their demographics are not provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the clearance was based on substantial equivalence to predicate devices and non-clinical testing, rather than a study requiring expert consensus on clinical ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as the clearance was based on non-clinical testing and comparison to predicate devices, which do not involve adjudication by multiple experts on a clinical test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. This device is a Transcutaneous Electrical Nerve Stimulator (TENS) for aesthetic use, not an AI-assisted diagnostic or imaging device that would typically involve human readers or MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable as the device is not an algorithm or AI-based system. It's a hardware device (TENS stimulator). Its "standalone performance" refers to its electrical output, safety, and functionality as evaluated in non-clinical bench tests.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For this type of device and submission, the "ground truth" is primarily established by compliance with recognized electrical and medical device safety standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, IEC 62366, ANSI/AAMI NS4:2013/(R)2017) and the established safety and effectiveness of the legally marketed predicate devices for their intended uses. The document states that "The Avazzia device technological specifications are similar; therefore, they do not pose new questions regarding safety and effectiveness." The usability study also confirmed that the device and instructions "met requirements for usability factors for over-the-counter use."

8. The sample size for the training set

This information is not applicable as the device is a hardware TENS stimulator and does not involve AI/machine learning models with training sets.

9. How the ground truth for the training set was established

This information is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 27, 2021

Avazzia, Inc Ms. Tammy Lahutsky Regulatory Affairs 13140 Coit Road, Suite 515 Dallas, Texas 75240

Re: K191951

Trade/Device Name: Avazzia OTC TENS for Aesthetics, model BEST-AV1: EZZI-LIFT Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: March 25, 2021 Received: March 26, 2021

Dear Ms. Lahutsky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jitendra Virani Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191951

Device Name

Avazzia OTC TENS for Aesthetics, model BEST-AV1™: EZZI-LIFT™ Device

Indications for Use (Describe)

The Avazzia OTC TENS for aesthetics, model BEST-AV1™: EZZI-LIFT™ Device is indicated for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitted by:
Manufacturer and Sponsor:Avazzia, Inc.13140 Coit RoadSuite 515Dallas, TX 75240 USA
Telephone:214-575-2820
Contact Person:Tammy Lahutsky
Date of Summary Preparation:April 26, 2021
Trade Names of Candidate Devices:Avazzia OTC TENS for aesthetics, model BEST-AV1™: EZZI-LIFT™Device
Device names:
Device Trade Proprietary name:Device Common or Usual Name:Classification Name:Regulation Number:Regulation ClassAvazzia OTC TENS for aesthetics, model BEST-AV1™: EZZI-LIFT™ DeviceAesthetic TENS DeviceTranscutaneous Electrical Nerve Stimulator for Pain Relief21 CFR 882.5890Class II
Product code:NEO

Predicate Devices

  • . NuFACE Trinity Facial and Neck Skin Toning Device, K181008, 08/14/2018, TENS Class II 21 CFR 882.5890 NFO (primary predicate)
  • . NuBODY Skin Toning Device, K171588, 07/27/2017, TENS Class II 21 CFR 882.5890 NFO (secondary predicate)

Reference Device used to support scientific methodology

  • o Avazzia OTC TENS Model Best-AV1™: Med-Best™, Med-Sport™, Avazzia Blue™, Avazzia Star™, K162392, 05/12/2017, TENS Class II 21 CFR 882.5890 NUH is identical in manufacturing to the candidate device and is used to support biocompatibility, electromagnetic compatibility and electrical safety testing.

Indications for Use

The Avazzia OTC TENS for aesthetics, model BEST-AV1™: EZZI-LIFT™ Device is indicated for over-thecounter aesthetic use including facial and neck stimulation or body skin stimulation.

Description of Candidate Device

Candidate device is a handheld, AA battery-operated portable device for use in the home or clinic. It has pre-set modes and is for over-the-counter use in the home or a clinical setting.

{4}------------------------------------------------

510k Aesthetic TENS, K191951

The candidate device uses transcutaneous electrical stimulation (TENS) for aesthetic use for facial, neck, and body stimulation.

How the Device Works

Candidate device uses TENS applied to the skin surface with cutaneous electrodes to stimulate the skin of the face, neck and body for aesthetic use.

The EZZI-LIFT device has built-in electrodes. Accessories, listed below, are also available to be used with the device.

Available Accessories

  • . Cutaneous electrodes:
    • o Spherical Soft tissue electrodes with common name: Y electrodes
    • o Brush electrode
    • Small Circular Electrode with common name: Pencil electrode o
  • Lead wires .

The user can apply the built-in electrodes or place accessory electrodes where indicated and apply stimulation for a recommended period of time. The user manual contains recommended treatment methods and times.

The user controls the output by selecting the mode and power setting.

Scientific Concepts

Various modes in the candidate device are suited for TENS applications. Very short duration high voltage pulsed current (HVPC) stimulates the skin of the face, neck and body for aesthetic use.

Predicate Comparison

Significant physical and performance characteristics of the device including Stimulation Output Specifications & Summary of the Technological Characteristics Compared to the Predicates are shown in the table below.

The candidate device has similar technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate devices as summarized in the technological characteristics.

{5}------------------------------------------------

page3
---------
Device Comparison Table(i.e., design, material, chemical composition, energy source)
CharacteristicunitsCandidatePrimary PredicateSecondary Predicate
Device NameEZZI-LIFTNuFACE TrinityNuBody
510KK191951K181008K171588
ManufacturerAvazzia, Inc.Carol Cole Company(dba NuFACE)Carol Cole Company(dba NuFACE)
Indications for UseThe Avazzia OTC TENSfor aesthetics, modelBEST-AV1™: EZZI-LIFT™ Device isindicated for over-the-counter aesthetic useincluding facial andneck stimulation orbody skin stimulation.The NuFACE TrinityDevice is intended forfacial and neckstimulation and isindicated for over-the-counter cosmeticuse.NuBODY Skin ToningDevice is intended forbody skin stimulationand is indicated forover-the-countercosmetic use.
Anatomic SitesFace, neck, and bodyFace and neckBody
Number of OutputModesnumber411
Low Battery IndicatorYesYesYes
Automatic shut off(minutes)60205
Compliance withVoluntary StandardsYes/NoIEC 60601-1IEC 60601-1-2IEC 60601-2-10ISO 14971IEC 62366IEC 60601-1IEC 60601-1-2IEC 60601-1IEC 60601-1-2IEC 60601-2-10IEC 60529, ISO 14971IEC 60601-1-6IEC 62366
Compliance with 21CFR 898Yes / NoYesYesYes
Weightounces5.4910 to 14
Dimensions[W x H x D](in.)2.6" X 4.7" X 1.35"2.8" x 5.1" x 1.3"2" x 6.5" x 6.0"
Housing Materials andConstructionPCBs inside plasticcase housingThermoplasticThermoplastic
User Interface DisplayLEDs and switchesnot publicly availablenot publicly available
Energy typeElectric stimulationElectric stimulationElectric stimulation
Number, Size, and Typeof BatteriesTwo 1.5 V AA batteriesInternal rechargeableLithium-ion batteryInternal rechargeableLithium-ion battery
Stimulation Output Specifications
CharacteristicunitsCandidatePredicatePredicate
Device NameEZZI-LIFTNuFACE TrinityNuBody
Waveform Shapepositive square wavefollowed by a dampedsinusoidal waveformof variable durationdepending ondamping and bodyloadingpulsed biphasicmodulated squarewavemonophasic voltagemodulated squarewave
Max output voltage(+/- 20%)
- at 500 ΩV-422828
- at 2,000 ΩV-122not publicly availablenot publicly available
- at 10,000 ΩV-348not publicly availablenot publicly available
Max output current(+/-20%)500μA
- at 500 ΩμA363400900
- at 2,000 ΩμΑ117not publicly availablenot publicly available
- at 10,000 ΩμΑ38not publicly availablenot publicly available
Duration of primary(depolarizing phase)mSec0.56060
Pulse DurationmSec1.16060
FrequencyHz15 to 1218.38.3
Net Charge per pulse at500 ΩμC4not publicly available54
Max phase chargeμC10not publicly availablenot publicly available
Max current density at500 ΩμΑ/cm²Built-in, Y, Brush: 800Pencil: 19,000419468
Max average powerdensity at 500 ΩμW/cm²Built-in, Y, Brush: 500Pencil: 3,500not publicly available4,180
Burst mode
(a) pulse per burstnumber12020
(b) burst per secondnumbern/a8.38.3
(c) burst duration(sec)n/a2.42.4
(d) duty cyclen/a20.2sec20.2sec
On Time (per second)seconds0.1not publicly availablenot publicly available
Off Time (per second)seconds0.9not publicly availablenot publicly available
Electrodes

{6}------------------------------------------------

{7}------------------------------------------------

œpageD1A
----------------------------
Stimulation Output Specifications
CharacteristicunitsCandidatePredicatePredicate
Device NameEZZI-LIFTNuFACE TrinityNuBody
(a) materialsStainless steel 316Chrome-platedChrome-plated
(b) electroconductivemedian/aConductive gelConductive gel
(c) electrode-to-skinimpedance rangeless than 15not publicly availablenot publicly available
(d) max duration ofuse (same as deviceshut off)(minutes)602020
(e) conductivesurface area
built in rectangularcm21.94--
sphericalcm25.60not publicly availablenot publicly available
brush/combcm20.56not publicly availablenot publicly available
small circular,pencil-likecm20.023not publicly availablenot publicly available

Comparison of technical characteristics with the predicate devices

Transcutaneous electric nerve stimulation (TENS) is the technological principle for both the subject and predicate devices. It is based on the use of electric stimulation to the skin for aesthetic use.

At a high level, the subject and predicate devices are based on the following same technological elements:

  • . Transcutaneous electric nerve stimulation (TENS) to the skin for aesthetic use
  • . Quantity of electric stimulation applied to the skin resulting from short, pulsed electrical current that results in only microamperage applications per second, also known as microcurrent.
  • . Meet the same medical device general safety requirements IEC 60601-1 and requirements EMC: IEC 60601-1-2.

There are variations between the subject and predicate devices for specific waveform shape, max voltage, phase charge, pulses per second, duty cycle and current.

Rationale that differences in technical performance specifications are within range of equivalence, and therefore, do not raise new questions.

The variations in technical characteristics fall within the range of Agency accepted differences for devices with the same principal of operation and the same intended use. This is consistent with the Agency accepted standard, ANSI/AAMI NS4:2013/ (R)2017 Transcutaneous electrical nerve stimulators, that references combination of current or charge over time for determining safety for Transcutaneous electrical stimulators.

The ANSI/AAMI document established a maximum charge to prevent serious injury or death.

{8}------------------------------------------------

page 6

Peak instantaneous maximum voltage and pulse width are considered and included in determination of charge per pulse and the resulting current flow and determining safety.

The candidate devices fall within the same range of equivalence for performance and energy output for meeting the ANSI/AAMI NS4:2013/(R)2017 standard requirements for safety; and therefore, they do not raise new questions of safety.

Nonclinical Tests

Nonclinical bench testing was performed to demonstrate that the candidate device met design specifications and accepted safety standards and requirements.

Performance testing was conducted using a production equivalent of the candidate device. The testing consisted of the evaluation of output waveform and energy characteristics. Results were compared to predicate device performance characteristics.

Additionally, output energy, electrical and constructional safety and EMC testing of the candidate device was conducted in accordance with Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, IEC 60601-1:2005 MOD, Medical electrical equipment - Part 1-2 : General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests, IEC 60601-1-2, Edition 4.0 2014-02, and Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators, IEC 60601-2-10, 2016.

The software development process and software verification is conducted according to Medical device software – Software life cycle processes, IEC 62304, 2015 and the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices – Guidance for Industry and FDA Staff – May 11, 2005.

Usability study was performed to assess the candidate device for aesthetic use. The candidate device and accompanying instructions for use met requirements for usability factors for over-the-counter use. No new questions of safety or efficacy were raised. No new risks were raised. The instructions for use support the safe and effective use of the EZZI-LIFT device and accessories for aesthetic use. Reference FDA guidance, "Applying Human Factors and Usability Engineering to Medical Devices, Guidance for Industry and Food and Drug Administration Staff", 2016 and IEC 62366, Medical devices - Part 1: Application of usability engineering to medical devices, 2015.

Results from bench testing and performance tests are compared to that of predicate devices in a comparison table format.

Nonclinical test results and studies, including the performance bench testing, product safety testing, software, usability studies, and risk management activities in conformance with Medical devices — Application of risk management to medical devices, ISO 14971, 2012 were reviewed and evaluated.

The candidate devices fall within the same range of equivalence for output energy characteristics and construction for meeting accepted safety standards for transcutaneous electrical nerve and muscle stimulators; and therefore, they do not raise new questions of safety.

Results confirmed that design and safety requirements were met, and that no new questions of safety or effectiveness were raised. This data was used to determine substantial equivalence.

{9}------------------------------------------------

Safety and Effectiveness

The Avazzia device technological specifications are similar; therefore, they do not pose new questions regarding safety and effectiveness.

Indications for use for TENS OTC for aesthetic use have been established, therefore, indications for use do not pose new questions regarding safety and effectiveness.

Conclusion

The Avazzia EZZI-LIFT device has intended use and technological characteristics that are substantially equivalent to the predicate devices.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).