K Number
K232764

Validate with FDA (Live)

Date Cleared
2024-02-07

(152 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentany1 Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-the- counter use.

In addition, these gloves were tested for use with Chemotherapy Citrate in accordance with ASTM D6978-05, Standards Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

The subject device in this 510(k) Notification is a Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate. The subject device is a patient examination glove made from nitrile rubber compound, Cornflower Blue color, powder-free, and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-19, the Standard Specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs and fentanyl citrate.

AI/ML Overview

This document is a 510(k) Premarket Notification from ECO Medi Glove Sdn. Bhd. for their Cornflower Blue Powder Free Nitrile Examination Glove, which has been tested for use with chemotherapy drugs and Fentanyl Citrate. The document summarizes the device's characteristics, its intended use, and the performance criteria it meets based on non-clinical testing.

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance CriteriaReported Device Performance
Chemotherapy & Fentanyl Permeation (ASTM D6978-05)
Carmustine (3.3mg/ml)Minimum Breakthrough detection time: Not specified as a strict acceptance criteria, but predicate's performance was 7.28 minutes for comparison. The subject device should demonstrate comparable or better performance for substantial equivalence.15.4 minutes (Note: A warning is issued not for use with Carmustine due to low permeation time)
Cyclophosphamide (20mg/ml)Minimum Breakthrough detection time: Not specified, but predicate's performance was >240 minutes.>240 minutes
Cytarabine (100mg/ml)Minimum Breakthrough detection time: Not specified, but predicate's performance was >240 minutes.>240 minutes
Doxorubicin HCl (2.0mg/ml)Minimum Breakthrough detection time: Not specified, but predicate's performance was >240 minutes.>240 minutes
Etoposide (20mg/ml)Minimum Breakthrough detection time: Not specified, but predicate's performance was >240 minutes.>240 minutes
Fluorouracil (50mg/ml)Minimum Breakthrough detection time: Not specified, but predicate's performance was >240 minutes.>240 minutes
Methotrexate (25mg/ml)Minimum Breakthrough detection time: Not specified, but predicate's performance was >240 minutes.>240 minutes
Paclitaxel (6mg/ml)Minimum Breakthrough detection time: Not specified, but predicate's performance was >240 minutes.>240 minutes
Thiotepa (10mg/ml)Minimum Breakthrough detection time: Not specified as a strict acceptance criteria, but predicate's performance was 2.67 minutes for comparison. The subject device should demonstrate comparable or better performance for substantial equivalence.56.2 minutes (Note: A warning is issued not for use with ThioTepa due to low permeation time)
Fentanyl Citrate Injection (100mg/2ml)Minimum Breakthrough detection time: Not specified, but reference device's performance was No breakthrough time at 240 minutes.>240 minutes
Physical (ASTM D6319-19)
Overall Length (mm)XS, S: Min 220mm; M, L, XL, XXL: Min 230mmMeets ASTM D6319-19
Width (±10mm)XXS = 65mm; XS = 70mm; S = 80mm; M = 95mm; L = 110mm; XL = 120mm; XXL = 130mmISO 2859-1 / S2/AQL 4.0
Thickness at Palm (mm)Min: 0.05 mmISO 2859-1 / S2/AQL 4.0
Thickness at Finger Tip (mm)Min: 0.05 mmMeets ASTM D6319-19
Tensile Strength (MPa) - Before Aging14minMeets ASTM D6319-19
Ultimate Elongation (%) - Before Aging500minMeets ASTM D6319-19
Tensile Strength (MPa) - After Aging (70°C for 168 hrs @ 100°C for 22 hrs)14minMeets ASTM D6319-19
Ultimate Elongation (%) - After Aging400minMeets ASTM D6319-19
Freedom from Pinholes (ASTM D5151-19)AQL 2.5, Inspection Level G-1Meets ASTM D5151-19
Powder-free Residue (ASTM D6124-06(2022))≤2.0 mg/pcMeets ASTM D6124-06(2022)
Biocompatibility
Irritation (ISO 10993-23:2021)Not an irritantPass (test article was a non-irritant)
Sensitization (ISO 10993-10:2021)Not a sensitizerPass (test article was a non-sensitizer)
Acute Systemic Toxicity (ISO 10993-11:2017)Not induce acute systemic toxicityPass (did not induce any acute systemic toxicity)

Study Proving Device Meets Acceptance Criteria:

The document indicates that the device meets the acceptance criteria through non-clinical testing. These tests were conducted according to established international and national standards.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: The document does not explicitly state the specific sample sizes used for each individual test (e.g., number of gloves tested for permeation, number of samples for physical properties, number of animals for biocompatibility). It references the standards used, which typically define the required sample sizes for such tests. For example, for "Freedom from Pinholes," it mentions "AQL 2.5, Inspection Level G-1" which implies a sampling plan based on the number of units in a lot. For "Width" and "Thickness" it references "ISO 2859-1 / S2/AQL 4.0," also indicating a sampling scheme.
  • Data Provenance: The document does not explicitly state the country of origin of the data. However, given that the manufacturer is ECO Medi Glove Sdn. Bhd. located in Taiping, Perak Darul Ridzuan, Malaysia, it is highly probable that the testing was conducted in Malaysia or contracted to a testing facility recognized in the region. The data is retrospective, as it's a summary of tests already performed for the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This information is not applicable in this context. The "ground truth" here is defined by objective physical and chemical measurements according to standardized test methods (ASTM, ISO). There is no "expert consensus" or "clinical ground truth" involved in evaluating the performance of examination gloves against these specific criteria. The standards themselves, developed by expert committees, define the acceptable levels.

4. Adjudication Method for the Test Set:

This information is not applicable. Since the evaluation involves objective measurements against predefined criteria in recognized standards, there is no need for an adjudication method by human readers or experts in the way it would be for a diagnostic medical device reading images or interpreting clinical data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. This device is a medical glove, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study related to human readers and AI assistance is entirely irrelevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. As stated above, this device is a medical glove and does not involve any algorithms or AI for standalone performance evaluation.

7. The Type of Ground Truth Used:

The "ground truth" in this context is established by:

  • Standardized Test Methods: Adherence to widely accepted national and international standards such as ASTM D6978-05 (for chemotherapy drug permeation), ASTM D6319-19 (for physical properties and dimensions), ASTM D5151-19 (for pinholes), ASTM D6124-06(2022) (for powder-free residue), and ISO 10993 series (for biocompatibility).
  • Objective Measurement Data: The results of the tests conducted in accordance with these standards provide the "ground truth" for whether the glove meets the defined performance characteristics. For breakthrough times, it's the measured time; for physical properties, it's the quantitative values compared to minimum/maximum thresholds; for biocompatibility, it's the qualitative assessment (Pass/Fail) based on biological response.

8. The Sample Size for the Training Set:

This information is not applicable. There is no "training set" in the context of physical product testing for a medical glove. The testing performed is to verify the manufactured product meets defined specifications.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable for the same reason as in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 7, 2024

ECO Medi Glove Sdn. Bhd Suresh Kumar QA Manager Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate Taiping, Perak Darul Ridzuan 34600 Malaysia

Re: K232764

Trade/Device Name: Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: December 13, 2023 Received: December 18, 2023

Dear Suresh Kumar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Allan G

For Bifeng Qian, M.D., Ph.D. Assistant Director

{2}------------------------------------------------

DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232764

Device Name

Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentany1 Citrate

Indications for Use (Describe)

Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentany1 Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-the- counter use.

In addition, these gloves were tested for use with Chemotherapy Citrate in accordance with ASTM D6978-05, Standards Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Drug ConcentrationMinimum Breakthrough detection time in Minutes
Carmustine (3.3mg/ml or 3000ppm)15.4 minutes
Cyclophosphamide (20mg/ml or 20,000ppm)>240 minutes
Cytarabine (100mg/ml)>240 minutes
Doxorubicin Hydrochloride (2.0mg/ml or 2000ppm)>240 minutes
Etoposide (20mg/ml or 20,000ppm)>240 minutes
Flourouracil (50mg/ml or 50,000)>240 minutes
Methorexate (25mg/ml or 25,000ppm)>240 minutes
Paclitaxel (6mg/ml or 6,000ppm)>240 minutes
Thiotepa (10mg/ml or 10,000ppm)56.2 minutes
Fentanyl Citrate Injection (100mg/2ml)>240 minutes

The maximum testing time is 240 minutes.

Please note that the following drugs have extremely low permeation time.

  • Carmustine (BCNU) (3.3mg/ml)- Minimum Breakthrough detection time 15.4 minutes.
  • Thiotepa (10ug/ml) Minimum Breakthrough detection time 56.2 minutes.

WARNING: Not for use with Carmustine and ThioTepa

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration

{4}------------------------------------------------

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

EF PSC Publishing Services (301) 443-6740

{5}------------------------------------------------

ECO Medi Glove Sdn. Bhd. (815262-D)

Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

s@ecomediglove.com.my Website: www.ecomediglove.com.
06 2216
FAX: +60 5 895 3660

TEL +60-3-806 2326
SST NO.: A11-1808-21015730

FAX +60-5-895 7668

K232764 510(K) Summary

Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

1.0 Submitter:

Company Name:ECO Medi Glove Sdn. Bhd.
Company Address:Lot 23826, Jalan Tembaga Kuning,Kamunting Raya Industrial Estate,34600 Taiping, Perak Darul Ridzuan,Malaysia.
Contact Person:Mr. Suresh Kumar
Telephone:+605-806 2316
Fax:+605-806 2315
Email:qa1@riverstone.com.my
  • 2.0 7th February 2024 Preparation Date :

3.0 Name of the Device

Trade Name / Proprietary Name: Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

Device Name: Nitrile Patient Examination gloves.

Device Classification Name: Patient Examination gloves (21 CFR 880.6250).

Device Class: Class I.

Product Code: LZA, LZC, OPJ and QDO

{6}------------------------------------------------

Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

E-mail : sales@ecomediglove.com.my Website: www.ecomediglove.com.r

TEL +60-5-806 2316 FAX +60-5-895 7668

SST NO.: A11-1808-21015730

4.0 Identification of The Legally Marketed Device:

Predicate Device: K103249, Cornflower Blue Powder-Free Exam Gloves with Tested for Use with Chemotherapy Drug Labeling Claim

Reference Device: K220375, Intercept Free, Nitrile Two Toned White/Green, Powder- Free, Textured Fingertips, Non-sterile, Ambidextrous, Beaded Cuff, Medical Examination Gloves, Tested for Use with the Opioids Fentanyl Citrate, Heroin, and both Opioids in simulated Gastric Acid (Vomit)

5.0 Device Description

The subject device in this 510(k) Notification is a Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate. The subject device is a patient examination glove made from nitrile rubber compound, Cornflower Blue color, powder-free, and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-19, the Standard Specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs and fentanyl citrate.

6.0 Indications for Use

Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-thecounter use.

In addition, these gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate in accordance with ASTM D6978-05, Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and Fentanyl CitrateConcentrationMinimum Breakthroughdetection time in Minutes
Carmustine (3.3mg/ml or 3000ppm),15.4 minutes
Cyclophosphamide (20mg/ml or 20,000ppm),>240 minutes
Cytarabine (100mg/ml)>240 minutes
Doxorubicin Hydrochloride (2.0mg/ml or 2000ppm),>240 minutes
Etoposide (20mg/ml or 20,000ppm),>240 minutes
Flourouracil (50mg/ml or 50,000),>240 minutes

{7}------------------------------------------------

Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

Methorexate (25mg/ml or 25,000ppm),>240 minutes
Paclitaxel (6mg/ml or 6,000ppm),>240 minutes
Thiotepa (10mg/ml or 10,000ppm),56.2 minutes
Fentanyl Citrate Injection 100mg/2ml>240 minutes

The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time.

Carmustine (BCNU) (3.3mg/ml)- Minimum Breakthrough detection time 15.4 minutes. Thiotepa (10ug/ml) - Minimum Breakthrough detection time 56.2 minutes.

WARNING: Not for use with Carmustine and ThioTepa

7.0 Specification for Nitrile exam gloves (ASTM D6319-19):

Dimension and Thickness of Gloves 7.1

DimensionSizeXXSSize XSSize SSize MSize LSize XLSizeXXL
Overall Length (mm)N/A220min220min230min230min230min230min
Width (± 10mm)N/A708095110120130
Thickness at Palm (mm)N/A0.05min0.05min0.05min0.05min0.05min0.05min
Thickness at Finger Tip(mm)N/A0.05min0.05min0.05min0.05min0.05min0.05min

7.2 Gloves Physical Properties and Holes

MeasurementBefore AgeingAfter Aging at 70°C for168 hrs @ 100°C for 22 hrs
Tensile Strength (MPa)14min14 Min
Ultimate Elongation (%)500min400min
Pin-hole LevelAQL 2.5Inspection Level G-1AQL 2.5Inspection Level G-1

Gloves meet all the specifications listed in ASTM D 6319-19

{8}------------------------------------------------

ECO Medi Glove Sd

Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping, Perak Darul Ridzuan. MALAYSIA

Plant 2 : Lot 23826, Jalan Tembaga Kuning Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

E-mail : sales@ecomediglove.com.my Website: www.ecomediglove.com.my

TEL +60-5-806 2316 FAX +60-5-895 7668

SST NO. : A11-1808-21015730

Comparison of Proposed and Predicate Device

CharacteristicsAcceptance CriteriaCornflower BluePowder Free NitrileExamination GloveTested for Use withChemotherapyDrugs and FentanylCitrate(subject device)Predicate DeviceCornflower BluePowder-Free ExamGloves with Testedfor Use withChemotherapy DrugLabeling ClaimK103249Reference DeviceIntercept Free, NitrileTwo TonedWhite/Green, Powder-Free, TexturedFingertips, Non-sterile,Ambidextrous, BeadedCuff, MedicalExamination Gloves,Tested for Use with theOpioids FentanylCitrate, Heroin, andboth Opioids insimulated Gastric Acid(Vomit)K220375AssessmentSimilarities andDifferences
Product CodeLZA, LZC, OPJ andQDOLZA, LZCLZA, LZC, QDOSame
Intended useA powder-free patientexamination glove is adisposable deviceintended for medicalpurposes that is wornon the examiner's handor finger to preventcontamination betweenpatient and examiner.The device is for over-the-counter use.A powder-free patientexamination glove is adisposable deviceintended for medicalpurposes that is wornon the examiner'shand or finger toprevent contaminationbetween patient andexaminer. The deviceis for over-the-counteruse.A Nitrile powder freemedical examinationglove is a disposabledevice, worn on the handor finger to preventcontamination betweenexaminer and patient orvictim. This specialtyglove has also beentested for use with theOpioid drugs fentanylcitrate, Heroin, and bothOpioid in simulatedGastric Acid.Similar Except thereference deviceglove intended usestatement wasextended toaddress the opioiddrug Heroin andgastric acid
Material useNitrile compoundNitrile compoundNitrile compoundNitrile compoundSame
ColorCornflower BlueCornflower BlueWhiteOutside/LimeGreen insideSame
SterilityNon sterileNon sterileNon sterileNon sterileSame
Single useSingle useSingle useSingle useSingle useSame
SterilityNon-SterileNon-SterileNon-SterileNon-SterileSame
SurfaceFinger TexturedFinger TexturedFinger TexturedFinger TexturedSame
CuffingBeadingRolled BeadingRolled BeadingRolled BeadingRolled BeadingSame
DesignAmbidextrousAmbidextrousAmbidextrousAmbidextrousSame

{9}------------------------------------------------

ECO Medi Glove Sdn. Bhd. (815262-D)

Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

Plant 2 : Lot 23826, Jalan Tembaga Kuning Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

E-mail : sales@ecomediglove.com.my Website: www.ecomediglove.com.my

TEL +60-5-806 2316

FAX +60-5-895 7668

SST NO. : A11-1808-21015730
DimensionsOverall Length (mm)XS, S: Min 220mmM, L, XL, XXL:Min 230mmWidth (±10mm)Overall Length (mm)Min 230mmWidth (± 10mm)Meets sizesspecified in ASTMD6319.Meets sizes specifiedin ASTM D6319.Different: Subjectdevice sizes meetASTM D6319-19 withaddition of XXS.
Sizes :Extra Extra Small(XXS) = 65mmExtra Small (XS) =70mmSmall (S) = 80mmMedium (M) = 95mmLarge (L) = 110mmExtra Large (XL)= 120mmDouble Extra Large(XXL) = 130mmSizes:Extra Extra Small(XXS) = 65mmExtra Small (XS)=70mmSmall (S) = 80mmMedium = 95mmLarge (L) = 110mmExtra Large (XL) =120mmDouble Extra Large(XXL) = 130mm
Thickness atPalm (mm) Min;0.05 mm Thicknessat Finger Tip (mm)Min 0.05 mmThickness atPalm (mm) Min;0.05 mmThickness atFinger Tip (mm)Min 0.05 mm

{10}------------------------------------------------

ECO Medi Glove Sdn. Bhd. (815262-D)

Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

Plant 2 : Lot 23826, Jalan Tembaga Kuning Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

E-mail : sales@ecomediglove.com.my Website: www.ecomediglove.com.my

TEL +60-5-806 2316

FAX +60-5-895 7668

SST NO. : A11-1808-21015730

PhysicalpropertiesBefore AgeingTensile Strength(MPa)= 14minUltimateElongation (%)= 500minAfter Aging at70°C for168 hrs @ 100°Cfor 22 hrsTensile Strength(MPa)= 14minUltimateElongation (%)= 400minMeets ASTM D6319-19Meets ASTM D6319-19Meets ASTM D6319-19Same
Freedom frompinholesAQL 2.5Inspection LevelG-1MeetsASTM D5151-19MeetsASTM D5151-19MeetsASTM D5151-19Same
Residual Powder≤2.0 mg/pcMeetsASTM D6124-06(2022)MeetsASTM D6124-06(2022)MeetsASTM D6124-06(2022)Same
BiocompatibilityISO 10993-23:2021Biologicalevaluation ofmedical devices-Part 23: Test forirritationUnder the conditions ofthis study, the testarticle was a non-irritant.Under the conditionsof this study, the testarticle was a non-irritant.Under the conditions ofthis study, the test articlewas a non-irritant.Same
ISO 10993-10:2021 BiologicalEvaluation onMedical Device –Part 10:Test for SkinSensitizationUnder the conditions ofthis study, the testarticle was a non-sensitizer.Under the conditionsof this study, the testarticle was a non-sensitizer.Under the conditions ofthis study, the test articlewas a non-sensitizer.Same
ISO 10993-11Biologicalevaluation onmedical device Part11 – Test forsystemic toxicityNot induce any acutesystemic toxicityNot induce anyacute systemictoxicityNot induce any acutesystemic toxicitySame

{11}------------------------------------------------

O Medi Glove Sdn. Bhd.

600 Taiping. Perak Darul Ridzuan, MALAYSIA

amunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA + 73876 lalan Temna

les@ecomedielove.com site: www.ecomedialove.com.my

60-5-806 2316

Resistance

Drugs and

against

5

FAX +60-5-895 7668

.ST NO. : A11-1808-21015730
10 m . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " . " .

Standards Practice 1) Carmustine 1) Carmustine Tested with: Fentanyl Similar: Subject for Assessment of (BCNU) 3.3mg/ml (BCNU) 3.3mg/ml citrate (Injectable): Device was tested Chemotherany resistance of 15.4 minutes 7.28 minutes 100mg/2mil: No with the same Medical Glove to 2) Cyclophosphamide 2) Cisplatin breakthrough time at 240 chemicals as the (Cytoxan) 1.0mg/mL, >240 Fentanyl Citrate Permeation by minutes predicate except Chemotherapy (20mg/ml) minutes with the addition of drugs ASTM >240 minutes 3) Cyclophosphamide Heroin saturated in fentanyl citrate D6978-05(2019) 3) Cytarabine HCI (Cytoxan) injectable Fentanyl No (reference device). (100mg/ml) (20mg/ml) breakthrough time at >240 minutes >240 minutes 240 minutes 4) Doxorubicin HCI 4) Doxorubicin HCI (2.0mg/ml) (2.0mg/ml) And both opioid drugs >240 minutes >240 minutes Fentanyl and Heroin 5) Etoposide 5) Etoposide mixed with simulated (20.0mg/ml (20.0mg/ml gastric acid: represents >240 minutes >240 minutes opioid-contaminated 6) Fluorouracil 6) 5-Fluorouracil vomit from overdose (50.0mg/ml) (50.0mg/ml) victims. >240 minutes >240 minutes No breakthrough time at 7) 7) Methotrexate 240 minutes (25mg/ml) (25mg/ml) >240 minutes >240 minutes 8) Paclitaxel 8) Paclitaxel (6.0mg/ml) (6.0mg/ml) >240 minutes >240 minutes 9) ThioTepa 9) ThioTepa (10.0mg/ml) (10.0mg/ml) 56.2 minutes 2.67 minutes Tested with: Fentanyl Warning : Not for use Citrate (Injection), with Carmustine or 100mg/2ml: no Thio Tepa breakthrough time at 240 minutes 240 minutes

Warning: Not for use with Carmustine or

Thio Tepa

{12}------------------------------------------------

ECO Medi Glove Sdn. Bh

Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping, Perak Darul Ridzuan. MALAYSIA Plant 2 : Lot 23826, Jalan Tembaga Kuning Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA

E-mail : sales@ecomediglove.com.my Website: www.ecomediglove.com.my

TEL +60-5-806 2316 SST NO. : A11-1808-21015730 FAX +60-5-895 7668

8.0 Summary of Non-Clinical Testing

CharacteristicsTest StandardAcceptance CriteriaTest Result
Freedom fromPin holesASTM D5151 -19(Re-approved 2011)AQL 2.5Inspection Level G-1MeetsASTM D5151-19
DimensionsASTM D6319 -19LengthXS, X: Min 220mmM, L, XL, XXL: Min 230mmMeetsASTM D6319-19
ASTM D6319 -19WidthSize XXS = 65mm±10(InternalRequirement)Size XS = 70mm±10Size S = 80mm±10Size M = 95mm±10Size L = 110mm±10Size XL = 120mm±10Size XXL =130mm±10ISO 2859-1 /S2/AQL 4.0
ASTM D6319 -19Thickness at Palm (mm)Min: 0.05 mmISO 2859-1 /S2/AQL 4.0
Thickness at Finger Tip(mm) Min: 0.05 mm
PhysicalpropertiesASTM D6319 -19ASTM D412-16(2021)Before AgeingTensile Strength (MPa)= 14minUltimate Elongation (%)= 500minMeets ASTMD6319-19
ASTM D6319 -19 andASTM D412-06(2021)After Aging at 70°C for168 hrs @ 100°C for 22 hrsTensile Strength (MPa)= 14minUltimate Elongation (%)= 400min
Powder-freeresidueASTM D6124-06(2022)≤2.0 mg/pcMeetsASTM D6124-06(2022)
IrritationISO 10993-23:2021Not an irritantPass
SensitizationISO 10993-10:2021Not a sensitizerPass
Acute SystemicToxicityISO 10993-11:2017Not induce acute systemictoxicityPass

{13}------------------------------------------------

aiping. Perak Darul Ridzuan. MALAYSIA

iping. Perak Darul Ridzuan. MALAYSIA

NO. : A11-1808-210157

8.0 Summary of Clinical Testing

N/A.

9.0 Conclusion

The Conclusions drawn from the non-clinical tests demonstrate that the subject device, Cornflower Blue Powder Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate, is as safe, as effective, and performs as well as or better than the legally marketed Predicate device cleared under K103249.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.