K Number
K232576

Validate with FDA (Live)

Date Cleared
2023-11-21

(88 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Medical nitrile examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

The proposed device, Medical Nitrile Examination Gloves, is a powder free patient examination glove, it is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. The device is provided in blue, purple, white and black colors. The device meets the requirements of ASTM D6319-19: Standard specification for Nitrile Examination Gloves for Medical Application. The proposed gloves are available in four sizes, which are S, M, L, XL, it could be selected by the user depended on size of hand. The different between each size is just in the dimension. The proposed device is provided in non-sterile.

AI/ML Overview

The medical device being discussed is "Medical Nitrile Examination Gloves" (K232576). The provided document is a 510(k) premarket notification, which focuses on demonstrating substantial equivalence to a legally marketed predicate device, not on proving clinical efficacy via extensive studies with human subjects.

Therefore, the requested information regarding multi-reader multi-case studies, human reader improvement with AI assistance, and the number of experts for ground truth establishment are not applicable as this is a physical medical device (gloves) and not an AI/software device.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The non-clinical tests conducted to demonstrate substantial equivalence are summarized in the following table. The "Results (Pass/Fail)" column directly indicates whether the device met the stated acceptance criteria.

Test MethodologyPurposeAcceptance CriteriaSpecificsReported Device Performance (Pass/Fail)
ASTM D6319-19, ASTM D3767-03 (2020)Physical DimensionsWidthS: 80±10mmM: 95±10mmL: 110±10mmXL: 120±10mmPass
LengthS: 220mm minM: 230mm minL: 230mm minXL: 230mm minPass
ThicknessPalm: 0.05mm minFinger: 0.05mm minPass
ASTM D6319-19, ASTM D412-16Physical PropertiesBefore AgingTensile Strength: 14MPa minUltimate Elongation: 500% minPass
After AgingTensile Strength: 14MPa minUltimate Elongation: 400% minPass
ASTM D5151-19Freedom from HolesBe free from holes when tested in accordance with ASTM D5151-Pass
ASTM D6124-06Powder Free ResidueLess than 2mg per glove-Pass
ISO 10993-10:2010IrritationNon-Irritating-Pass
ISO 10993-10:2010SensitizationNon-Sensitizing-Pass
ISO 10993-11:2009Acute Systemic ToxicityNon-toxicity-Pass

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the exact sample sizes (number of gloves or test samples) used for each specific non-clinical test (e.g., how many gloves were tested for dimensions, tensile strength, or holes). However, the tests were conducted according to recognized international and national standards (ASTM and ISO), which typically specify appropriate sampling plans and methodologies.

The data provenance is from non-clinical tests conducted by the manufacturer, Jiangsu Complete Sincerity Medical New Materials Co., Ltd., in China. The tests are retrospective in the sense that they were performed on samples of the manufactured device to demonstrate compliance with standards for the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This question is not applicable. For a physical device like examination gloves, "ground truth" is established by direct measurement and standardized testing against predefined physical and chemical properties as specified in the ASTM and ISO standards. There is no subjective interpretation by medical experts required for these types of tests.

4. Adjudication Method for the Test Set

This question is not applicable. As explained above, the assessment is based on objective, standardized tests with pass/fail criteria. There is no need for human adjudication of results in the traditional sense of medical image interpretation or clinical outcomes.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/software device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/software device.

7. The Type of Ground Truth Used

The ground truth used for this device's evaluation is based on:

  • Standardized measurement values: For dimensions (width, length, thickness), tensile strength, and elongation percentage as defined by ASTM D6319-19, ASTM D3767-03, and ASTM D412-16.
  • Absence of defects/undesirable characteristics: For freedom from holes (ASTM D5151-19) and powder-free residue (ASTM D6124-06).
  • Biological responses/toxicity: Absence of irritation, sensitization, and acute systemic toxicity as defined by ISO 10993-10:2010 and ISO 10993-11:2009.

These are objective, quantifiable, and internationally recognized criteria for medical gloves.

8. The Sample Size for the Training Set

Not applicable. This is a physical device, not an AI/machine learning model. There is no "training set" in the context of this 510(k) submission.

9. How the Ground Truth for the Training Set was Established

Not applicable. There is no training set for this physical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 21, 2023

Jiangsu Complete Sincerity Medical New Materials Co., Ltd. % Diana Hong General Manger Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China

Re: K232576

Trade/Device Name: Medical Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 10, 2023 Received: August 25, 2023

Dear Diana Hong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232576

Device Name Medical nitrile examination gloves

Indications for Use (Describe)

Medical nitrile examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K232576

  • Date of Preparation: 10/20/2023 1.
    1. Sponsor Identification

Jiangsu Complete Sincerity Medical New Materials Co., Ltd.

NO.7 Wenzhou Road, Economic Development Zone, Shuyang County, Suqian, Jiangsu Province, China, 223600

Contact Person: Shangjun Xie Position: General Manager Tel: +86-527-81229099 Email: 392421042@qq.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Xingqi Wang (Alternative Contact Person)

Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-2281-5850, Fax: 360-925-3199 Email: info@mid-link.net

    1. Identification of Proposed Device
      Trade Name: Medical Nitrile Examination Gloves Common Name: Medical Nitrile Examination Gloves

Regulatory Information

{5}------------------------------------------------

Classification Name: Non-powdered patient examination glove Classification: I Product Code: LZA; Regulation Number: 21CFR 880.6250 Review Panel: General Hospital

Indications for use:

Medical nitrile examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

The proposed device, Medical Nitrile Examination Gloves, is a powder free patient examination glove, it is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. The device is provided in blue, purple, white and black colors. The device meets the requirements of ASTM D6319-19: Standard specification for Nitrile Examination Gloves for Medical Application. The proposed gloves are available in four sizes, which are S, M, L, XL, it could be selected by the user depended on size of hand. The different between each size is just in the dimension. The proposed device is provided in non-sterile.

న్. Identification of Predicate Device

510(k) Number: K220442 Product Name: Synguard Nitrile Exam Glove Manufacturer: Tianjin Aoshang Outdoor Equipment Co., Ltd.

{6}------------------------------------------------

6. Technological Characteristics Comparison

Table 1 Comparison of Technology Characteristics

ITEMProposed DevicePredicate Device K220442Remark
Product CodeLZALZASame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassIISame
Indication for useMedical nitrile examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.Synguard Nitrile Exam Glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.Same
MaterialNATEX CNBR LATEXAcrylonitrile, ButadieneDifferent1
ColorBlue, White, Black, PurpleBlueDifferent2
SterilityNon-sterileNon-sterileSame
Single-useYesYesSame
SizeS, M, L, XLS, M, L, XL
Dimensions(ASTM D6319-19)WidthDifferent3
S$80 \pm 10mm$
M$95 \pm 10mm$
L$110 \pm 10mm$
XL$120 \pm 10mm$
Length
S220mm min
M230mm min
L230mm min
XL230mm min
Thickness
Palm0.05mm min
Finger0.05mm min
Physical Properties(ASTM D6319-19and ASTMD412-16)Before AgingSame
Tensile Strength14MPa min
Ultimate Elongation500% min

{7}------------------------------------------------

Tensile Strength14MPa minTensile Strength14MPa min
UltimateElongation400% minUltimate Elongation400% min
Powder free residue(ASTM D6319-19and ASTMD6124-06)Less than 2mg per gloveLess than 2mg per gloveSame
Freedom fromHoles (ASTMD5151-19)No water leakage occurs. AQL2.5No water leakage occurs. AQL2.5Same
Biocompatibility
Skin IrritationNo IrritationNo Irritation
SensitizationNo SensitizationNo SensitizationSame
Acute SystemToxicityNo Acute Systemic ToxicityNo Acute Systemic Toxicity

Different 1- Material

The material of the proposed device is different from the predicate device. However, the biocompatibility and specification performance tests have been performed on each of the proposed devices and the test results show no significant concerns. The difference will not affect the safety and effectiveness of the proposed device.

Different 2- Color

The color of the proposed device is different from the predicate device. However, the biocompatibility and specification performance tests have been performed on each of the proposed devices and the test results show no significant concerns. Therefore, the difference will not affect the safety and effectiveness of the proposed device.

Different 3- Size& Dimensions

The size and dimension of the proposed device is not exactly same as the predicate device. However, the size and dimension of the proposed device has been covered by ASTM D6319-19. The user can select appropriate model depended on size of user's hand. In addition, its dimension has been tested and met the requirement of ASTM D6319-19. Therefore, the difference will not affect the safety and effectiveness of the proposed device.

7. Summary of Non-Clinical Testing

Non clinical tests were conducted to verify that the performance of the proposed device met all design specifications of the applicable and current standards. The test results demonstrated that the proposed device complies with the following standards:

{8}------------------------------------------------

  • ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application

  • ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves

  • ASTM D3767-03 (2020) Standard Practice for Rubber-Measurement of Dimensions

  • ASTM D412-16 Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers-Tension

  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves

  • ISO 10993-10:2010 Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization.

  • ISO 10993-11:2017 Biological evaluation of medical devices-Part 11: Tests for systemic toxicity;

Test MethodologyPurposeAcceptance CriteriaResults(Pass/Fail)
ASTM D6319-19,ASTM D3767-03(2020)PhysicalDimensionsWidthSMLXL80±10mm95±10mm110±10mm120±10mm
LengthSMLXL220mm min230mm min230mm min230mm min
ThicknessPalmFinger0.05mm min0.05mm min
ASTM D6319-19,ASTM D412-16Physical PropertiesBefore AgingTensileStrengthUltimateElongation14MPa min500% min
After AgingTensileStrengthUltimateElongation14MPa min400% min
ASTM D5151-19Freedom fromHolesBe free from holes when tested in accordance withASTM D5151
ASTM D6124-06Powder free residueLess than 2mg per glove
ISO 10993-10:2010IrritationNon-Irritating

Table 2 Summary of non-clinical performance testing

{9}------------------------------------------------

ISO 10993-10:2010SensitizationNon-SensitizingPass
ISO 10993-11:2009Acute SystemictoxicityNon-toxicityPass

8. Summary of Clinical Testing

No clinical study is included in this submission.

    1. Conclusion:
      The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device K220442.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.