K Number
K232220

Validate with FDA (Live)

Date Cleared
2023-12-05

(132 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OSSTEM Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.

Device Description

SS Abutment System is made of titanium alloy. SS Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.

AI/ML Overview

The medical device described in the document is the SS Abutment System from Osstem Implant Co., Ltd. This device is an endosseous dental implant abutment, classified as Class II under regulation 21 CFR 872.3630 with product code NHA.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The submission claims substantial equivalence to predicate devices, focusing on design, function, indications for use, material, manufacturing process, and performance characteristics. The acceptance criteria are implicitly aligned with the performance characteristics of the predicate devices, particularly regarding mechanical properties, biocompatibility, sterilization, and MR compatibility.

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Mechanical PropertiesFatigue strength according to ISO 14801 standards and FDA guidance for Root-form Endosseous Dental Implants and Endosseous Dental Abutments. The criteria imply that the subject device's fatigue performance should be equal to or greater than that of the predicate devices, especially when dimensions differ (e.g., larger diameters or specific length variations that might impact stress distribution). The document explicitly mentions "not present a new worst case for performance testing."For ComOcta Plus Abutment: Differing diameters (Ø5.5 ~ Ø7.6) and lengths (L 5, 6, 7, 7.5, 8, 9.5; L 7.5; L 5, 6, 7, 8, 8.5, 10; L 7.5; L 9.5; L 8.5; L 9.5) compared to predicate (Ø5.5 x L 6.5, 7.5, 8.5, 9.5; Ø6.5 x L 6.5, 7.5, 8.5, 9.5). Performance: "larger than predicate device, so it is considered to have equal to or greater performance than that of the predicate device." "fatigue of the proposed device is considered to be equal to or greater than that of the predicate device." Therefore, no additional fatigue testing was conducted.For ComOcta Milling Abutment: Lengths (L 14, L 14.35) not in predicate range, but diameter range is same. Performance: "fatigue of the proposed device is considered to be equal to or greater than that of the predicate device." No additional fatigue testing. Modifiability is within safe limits (min post height 4mm, gingival height 1-3mm, no angulation).For Port Abutment: Larger diameter (Ø6.06) compared to predicate (Ø4.8), same length range. Performance: "added diameter is bigger than reference device and the lengths 3.56.5mm is same as the range of reference device." "not received single load because this is used for making overdenture that means load is dispersed to the full denture. Therefore, we don't consider additional fatigue testing."For O-ring Abutment: Larger diameter (Ø4.3) compared to predicate (Ø3.5), same gingiva height range. Performance: "added diameter is bigger than reference device and the gingiva height 04mm is same as the range of predicate device." "not received single load because this is used for making overdenture that means load is dispersed to the full denture. Therefore, we don't consider additional fatigue testing."For Closing Screw: Longer dimensions than predicate. Performance: "does not affect safety and effectiveness because it is used to protect the exposed platform of the implant during healing period." No additional fatigue testing.
BiocompatibilityCompliance with FDA Guidance Document Use of International Standard ISO 10993-1. The material composition of the device should be safe for biological contact."The SS Abutment System has same materials, manufacturer, manufacturing process etc., as predicate devices. Therefore, we didn't conduct additional biocompatibility test."
Sterilization ValidationDemonstrated sterility and maintenance of sterility throughout the claimed shelf-life. The sterilization process should be validated and consistent with predicate devices or established methods."Validation of the gamma irradiation process was previously conducted for the predicated device. There has been no change to the manufacturing or sterilization processes since then and the subject devices do not present a new worst case for sterilization validation; therefore, additional validation is not required."
Shelf-lifeStability and integrity of the device over its claimed shelf-life."SS Abutment System are made with titanium and titanium alloy we don't consider about shelf life of material by itself because this metal is widely known that it generally has no adversely affect by aging. Therefore we certify that product such like metal has no shelf life." The predicate device likely had an 8-year shelf life for some components (e.g., Closing Screw), and this is maintained where applicable.
MR CompatibilitySafety in the Magnetic Resonance (MR) Environment, including assessment of magnetically induced displacement force and torque, per FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment.""Non-clinical worst-case MRI review was performed to evaluate the Subject device components in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance."

2. Sample size used for the test set and the data provenance

The document does not describe a traditional "test set" in the context of a clinical performance study. Instead, performance is established through non-clinical testing and comparison to predicate devices, often leveraging the fact that the subject device either has:

  • Identical characteristics to a predicate.
  • Dimensions that are considered to offer equivalent or superior performance (e.g., larger diameter on an implant component).
  • A different intended use/loading condition (e.g., overdenture dispersing load) that mitigates concerns about increased stress from dimensional changes.

Therefore, there is no sample size for a "test set" from patients or images. The "samples" would be the configurations of the devices tested in laboratory settings (e.g., for fatigue), or the material samples for biocompatibility, or the specific device designs analyzed for MR compatibility. The document does not specify the number of units tested for any of these non-clinical evaluations, but refers to "worst-case scenarios" for mechanical testing, which implies selection of specific configurations deemed most challenging.

Data Provenance: Not applicable as no human or image data is used for performance testing. The "data" comes from laboratory physical testing, material standards, and scientific literature.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. As this is a non-clinical submission based on substantial equivalence to predicate devices and established standards (e.g., ISO, ASTM, FDA guidance), there is no "ground truth" derived from expert consensus on patient data. The "truth" is based on adherence to engineering and material science principles, and regulatory guidelines.

4. Adjudication method for the test set

Not applicable for the reasons outlined in point 3.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical dental implant abutment, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not a software algorithm.

7. The type of ground truth used

The "ground truth" for this submission is based on:

  • Standards and Guidance Documents: ISO 14801 for fatigue testing, ISO 10993-1 for biocompatibility, and FDA guidance documents for dental implants and MR compatibility.
  • Material Science: Established properties of Titanium Alloy (ASTM F136), Titanium Gr 3 & 4 (ASTM F67), and POM (Polyoxymethylene).
  • Existing Validations: Relying on previous biocompatibility, sterilization, and shelf-life validations performed for predicate devices with identical materials, manufacturing processes, or for which the subject device does not represent a new worst-case.
  • Scientific Rationale: For MR compatibility, scientific literature and rationale are used.

8. The sample size for the training set

Not applicable. This is a physical medical device, not a machine learning model requiring a training set.

9. How the ground truth for the training set was established

Not applicable for the reasons outlined in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

December 5, 2023

Osstem Implant Co., Ltd. % Peter Lee RA/OA Manager HioSSEN Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030

Re: K232220

Trade/Device Name: SS Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 31, 2023 Received: November 2, 2023

Dear Peter Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232220

Device Name SS Abutment System

Indications for Use (Describe)

The OSSTEM Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Implant Co., Lt g-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

510(k) Summary K232220

Date: December 5, 2023

1. Company and Correspondent making the submission

-Submitter's Name:Osstem Implant Co., Ltd.
-Address:66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan,48002, Republic of Korea
-Contact:Ms. Seungju Kang
-Phone:+82-51-850-2500
-Correspondent's Name:Hiossen Inc.
-Address:85 Ben Fairless Dr. Fairless Hills, PA 19030
-Contact:Mr. Peter Lee
-Phone:+1-267-759-7031

2. Proposed Device

- Trade or (Proprietary) Name: SS Abutment System
- Classification Name: Endosseous dental implant Abutment
- Regulation Number: 21 CFR 872.3630
- Devce Classification: Class II
- Classification Product Code: NHA

3. Predicated and Reference Devices

Primary Predicate
K160670US/SS/GS System
Reference Device
K182091Osstem Abutment System
K062051SS System

4. Indication for use

The OSSTEM Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.

5. Device Description

SS Abutment System is made of titanium alloy. SS Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a stylized font, with the word "IMPLANT" written in a smaller font below it. The logo is simple and clean, and it is likely used to represent the company and its products.

Osstem Implant Co., Ltd. 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

SS Abutment System is similar to other commercially available products based on the intended use, technology used, claims, material composition employed and performance characteristics.

DeviceContent
ComOcta Plus AbutmentDescriptionIt is used for making general cement- retained/combination prosthesis and used for thick gingiva or deeply placed implant.
MaterialTi-6Al-4V (ASTM F136), TiN Coating
Diameter (D) and Length (L)Ø5.5 x L 5, 6, 7, 7.5, 8, 9.5
Ø6.0 x L 7.5
Ø6.5 x L 5, 6, 7, 8, 8.5, 10
Ø6.8 x L 7.5
Ø7.0 x L 9.5
Ø7.2 x L 8.5Ø7.6 x L 9.5
ComOcta Milling AbutmentDescriptionIt is used as cement/combination-retained prosthesis and it is used when the abutment's margin shape needs corrections. This device is intended to be modified chairside by the clinician.
MaterialTi-6Al-4V (ASTM F136)
Diameter (D) and Length (L)Ø5.5 x L 14
Ø6.5 x L 14.35
ComOcta Protect CapDescriptionIt is made of plastic material and formed with a cover cap to protect porous upper structure.
MaterialPOM
Diameter (D) and Length (L)Ø5.3 x L 6.35, 7.85, 9.35
Port AbutmentDescriptionPort Abutment is used for prosthetic restoration. It is for implant retained overdenture at maxilla/mandible in case of the patient has no teeth.
MaterialTi-6Al-4V (ASTM F136), TiN Coating
Diameter (D) and Length (L)Ø6.06 x L 3.5, 4.5, 5.5, 6.5
O-ring AbutmentDescriptionIt is used for overdenture with O-ring attachment.
MaterialTi-6Al-4V (ASTM F136)
Diameter (D) and Gingiva Height (G/H)Ø4.3 x G/H 0, 2, 4
MaterialTitanium Gr 3 (ASTM F67)
Diameter (D) and Length (L)Ø4.8 x L 8.2Ø6.0 x L 8.2
Closing ScrewDescriptionIt is used to protect the exposed platform of the implant during healing period
MaterialTitanium Gr 4 (ASTM F67)
Diameter (D) andLength (L)$\text{\O}3.5$ x L 6.0$\text{\O}4.3$ x L 6.4

The specifications of the proposed device are as follow;

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, lighter font. The logo is simple and modern, and it is likely used to represent the company's brand.

Osstem Implant Co., Ltd

66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

6. Substantial Equivalence Matrix

These subject devices are cleared in past 510(k); therefore, indication for use, shape, connection structure, material, surface treatment, manufacturer and etc. are the same with predicated devices except dimension of additional products.

Subject DevicePrimary Predicate DeviceRemark
Device NameComOcta Plus AbutmentComOcta Plus AbutmentSame
510(k)Number-K160670-
ManufacturerOsstem Implant Co., LtdOsstem Implant Co., LtdSame
DesignImage: Two ComOcta Plus AbutmentsImage: Two ComOcta Plus AbutmentsSame
Indication forUseThe OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.The OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.Same
Principle ofOperationAs general cement retainedrestoration, it is connected withfixture and cemented crown onthe abutmentAs general cement retainedrestoration, it is connected withfixture and cemented crown onthe abutmentSame
Diameter (D)and Length(L)Ø5.5 x L 5, 6, 7, 7.5, 8, 9.5Ø6.0 x L 7.5Ø6.5 x L 5, 6, 7, 8, 8.5, 10Ø6.8 x L 7.5Ø7.0 x L 9.5Ø7.2 x L 8.5Ø7.6 x L 9.5Ø5.5 x L 6.5, 7.5, 8.5, 9.5Ø6.5 x L 6.5, 7.5, 8.5, 9.5Different
MaterialTitanium Alloy (ASTM F136)Titanium Alloy (ASTM F136)Same
S.E.SimilaritiesProposed ComOcta Plus Abutment has same design, function and indication foruse; and is made with same material with same manufacturing process by samemanufacturer compared to that of the predicate ComOcta Plus Abutment,K160670.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, lighter font. The logo is simple and modern.

Osstem Implant Co., Ltd.

OSSTEM Implant Co., Ltd., Ltd., Ltd.,
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

DifferencesThe diameter Ø5.5 ~ Ø7.6 of proposed device differs from predicate device; but isincluded in the dimension range of the predicate device, except for the largerdiameters Ø6.8 ~ Ø7.6. But they are larger than predicate device, so it isconsidered to have equal to or greater performance than that of the predicatedevice.
The diameter Ø5.5 and Ø6.5 of proposed device are same and there is onlydifference in the length 5mm and 6mm which is smaller than the predicate device.However, proposed device has same design, function, connection and platform etcas predicate deivce. In addition, since proposed device is straight type and thedistance between the embedding plane and the centre of the hemispherical hoadingmember always is 11mm according to the ISO 14801, it has same moment arm.
As a result, the fatigue of the proposed device is considered to be equal to orgreater than that of the predicate device. Therefore, we don't conduct additionalfatigue testing.
: Proposed ComOcta Plus Abutment and the predicate device have common indesign, function, indications for use, material, manufacturing process,manufacturer, etc.; therefore, the proposed device is substantially equivalent to thepredicate device.
Subject DevicePredicate DeviceRemark
Device NameComOcta Milling AbutmentComOcta Plus AbutmentDifferent
510(k)Number-K160670-
ManufacturerOsstem Implant Co., LtdOsstem Implant Co., LtdSame
DesignImage: ComOcta Milling AbutmentImage: ComOcta Plus AbutmentDifferent
Indication forUseThe OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.The OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.Same
Principle ofOperationAs general cement retainedrestoration, it is connected withAs general cement retainedrestoration, it is connected withSame

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern.

Osstem Implant Co., Ltd.

Osstem Implant Co., و Ltd.
66-16, Bansong-ro 513beon-gil, Haundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

fixture and cemented crown onthe abutmentfixture and cemented crown onthe abutment
Diameter (D)and Length(L)Ø5.5 x L 14Ø6.5 x L 14.35Ø5.5 x L 6.5, 7.5, 8.5, 9.5Ø6.5 x L 6.5, 7.5, 8.5, 9.5Different
MaterialTitanium Alloy (ASTM F136)Titanium Alloy (ASTM F136)Same
S.E.SimilaritiesProposed ComOcta Milling Abutment has same function and indication for use;and is made with same material with same manufacturing process by samemanufacturer compared to that of the predicated ComOcta Plus Abutment,K160670.DifferencesThe length 14mm and 14.35mm of proposed device is not included in dimensionrange of predicate device. However, the diameter range of proposed device issame as predicate device, and since the proposed device is compatible with thesame implant as predicate device, it has same design, function, connection andplatform etc as predicate deivce. In addition, since proposed device is straight typeand the distance between the embedding plane and the centre of the hemisphericalhoading member always is 11mm according to the ISO 14801, it has samemoment arm.As a result, the fatigue of the proposed device is considered to be equal to orgreater than that of the predicate device. Therefore, we don't conduct additionalfatigue testing.The ComOcta Milling Abutment can be modified chairside by hand, to aminimum post height 4mm, and a gingival height between 1mm and 3mm, withno angulation allowed. These dimensions are not significantly different comparedto the predicate device and do not present a new worst case for performancetesting. In addition, proposed ComOcta Milling Abutment and the predicateComOcta Plus Abutment have common in design, function, indications for use,material, manufacturing process, manufacturer, etc.; therefore, the proposed
Subject DevicePredicate DeviceRemark
Device NameComOcta Protect CapComOcta Retraction CapDifferent
510(k)Number-K160670-
ManufacturerOsstem Implant Co., LtdOsstem Implant Co., LtdSame

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, and the word "IMPLANT" is in a smaller, gray font below it. There is a small circle to the right of the word "OSSTEM".

sstem Implant Co., Ltd.

Osstem Implant Co., و Ltd.
66-16, Bansong-ro 513beon-gil, Haundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

DesignImage: capImage: capDifferent
Indication forUseThe OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.The OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.Same
Principle ofOperationComOcta Protect Cap istemporary connected withComOcta abutment for minimizeforeign body sensation by thepatientComOcta Retraction Cap istemporary connected withComOcta abutment for minimizeforeign body sensation by thepatient and accurate marginimpression function when takingimpression directly from aComOcta abutment.Same
Diameter (D)and Length(L)Ø5.3 x L 6.35, 7.85, 9.35Ø6.4 x L 6.25, 7.75, 9.25Ø7.7 x L 6.1, 7.6, 9.1Different
MaterialPOMPOMSame
SimilaritiesProposed ComOcta Protect Cap has same function, indication for use; and is madewith same material with same manufacturing process by same manufacturercompared to that of the predicated ComOcta Retraction Cap, K160670.DifferencesThe diameter Ø5.3 of proposed device is smaller than that of predicate device andthe length 6.35 ~ 9.35mm of proposed device differs from predicate device.Although there are differences in dimension range and feature, proposed devicehas same compatible device, ComOcta Abutment, and principle of operation withpredicate device. In addition, proposed device is made of POM(Polyoxymethylene) which is same raw material of predicate device.... Proposed ComOcta Protect Cap and the predicated device have common indesign, function, indications for use, material, manufacturing process,manufacturer, etc.; therefore, the proposed device is substantially equivalent to thepredicated device.
Subject DeviceReference DeviceRemark
Device NamePort AbutmentPort AbutmentSame

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a stylized font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern.

Osstem Implant Co., Ltd..
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea

Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

510(k)Number-K182091-
ManufacturerOsstem Implant Co., LtdOsstem Implant Co., LtdSame
DesignImage: Prosthetic systemImage: Abutment SystemSame
Indication forUseThe OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.The Osstem Abutment System isintended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or overdentures.Same
Principle ofOperationUsing making implant retainedoverdenture at maxilla/mandible.Using making implant retainedoverdenture at maxilla/mandible.Same
Diameter (D)and Length(L)$Ø$ 6.06 x L 3.5, 4.5, 5.5, 6.5$Ø$ 4.8 x L 2.5, 3.5, 4.5, 5.5, 6.5Different
MaterialTitanium Alloy (ASTM F136)Titanium Alloy (ASTM F136)Same
S.E.SimilaritiesProposed Port Abutment has same design, function and indication for use; and ismade with same material with same manufacturing process by same manufacturercompared to that of the reference device Port Abutment, K182091.DifferencesThe only difference is that the proposed device has a diameter $Ø$ 6.06 compared tothe reference device diameter $Ø$ 4.8. However, added diameter is bigger thanreference device and the lengths 3.5~6.5mm is same as the range of referencedevice. In addition, proposed device is straight type and is not received single loadbecause this is used for making overdenture that means load is dispersed to thefull denture. Therefore, we don't consider additional fatigue testing.. . Proposed Port Abutment and the predicate device have common in design,function, indications for use, material, manufacturing process, manufacturer, etc.;therefore, the proposed device is substantially equivalent to the predicate device.
Subject DeviceReference DeviceRemark
Device NameO-ring AbutmentO-ring AbutmentSame
510(k) Number-K062051-
ManufacturerOsstem Implant Co., LtdOsstem Implant Co., LtdSame

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Osstem Implant. The logo is in orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, lighter font. The logo is simple and modern.

Osstem Implant Co., Ltd.

Osstem Implant Co., و Ltd.
66-16, Bansong-ro 513beon-gil, Haundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

DesignImage: Prosthetic system design 1Image: Prosthetic system design 2Same
Indication forUseThe OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.SS system in indicated for use inpartially or fully edentulousmandibles and maxillae, insupport of single or multiple-unitrestorations including; cementedretained, screw retained, oroverdenture restorations, andterminal or intermediate abutmentsupport for fixed bridgework. SSSystem is for single stage surgicalproceduresSame
Principle ofOperationUsing making stud typeoverdenture prosthetics.Using making stud typeoverdenture prosthetics.Same
Diameter (D)and GingivaHeight (G/H)$Ø4.3$ x G/H 0, 2, 4$Ø3.5$ x G/H 0, 2, 4Different
MaterialTitanium Alloy (ASTM F136)Titanium Alloy (ASTM F136)Same
S.E.SimilaritiesProposed O-ring Abutment has same design, function and indication for use; andis made with same material with same manufacturing process by samemanufacturer compared to that of the reference O-ring Abutment, K062051.DifferencesThe only difference is that the proposed device has a diameter $Ø4.3$ compared tothe reference device diameter $Ø3.5$ . However, added diameter is bigger thanreference device and the gingiva height 0~4mm is same as the range of predicatedevice. In addition, proposed device is straight type and is not received single loadbecause this is used for making overdenture that means load is dispersed to thefull denture. Therefore, we don't consider additional fatigue testing.. . Proposed O-ring Abutment and the predicate device have common in design,function, indications for use, material, manufacturing process, manufacturer, etc.;therefore, the proposed device is substantially equivalent to the predicate device.
Subject DeviceReference DeviceRemark
Device NameClosing ScrewClosing ScrewSame
510(k) Number-K062051-
ManufacturerOsstem Implant Co., LtdOsstem Implant Co., LtdSame
DesignImage: dental implantImage: dental implantDifferent
Indication forUseThe OSSTEM Prosthetic systemis intended for use with a dentalimplant to provide support forprosthetic restorations such ascrowns, bridges, or over-dentures.SS system in indicated for use inpartially or fully edentulousmandibles and maxillae, insupport of single or multiple-unitrestorations including; cementedretained, screw retained, oroverdenture restorations, andterminal or intermediate abutmentsupport for fixed bridgework. SSSystem is for single stage surgicalproceduresSame
Principle ofOperationIt is connected with placedimplant to protect othersubstances from ingressing intothe fixture after the surgicaloperation.It is connected with placedimplant to protect othersubstances from ingressing intothe fixture after the surgicaloperation.Same
Diameter (D)and Length (L)Ø3.5 x L 6.0Ø4.3 x L 6.4Ø3.5 x L 5.5Ø4.3 x L 5.85Different
MaterialTitanium Gr 4Titanium Gr 4Same
SterilizationRadiation SterileRadiation SterileSame
Shelf life8 years8 yearsSame
S.E.SimilaritiesProposed Closing Screw has same diameter dimension, design, function andindication for use; and is made with same material with same manufacturingprocess by same manufacturer compared to that of the predicated Closing Screw,K062051.DifferencesAlthough there are differences in dimensions that the proposed device is longerthan predicate device, it does not affect safety and effectiveness because it is usedto protect the exposed platform of the implant during healing period.:. Proposed Closing Screw and the predicate device have common in design,

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a stylized font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern.

Osstem Implant Co., Ltd.

Osstem Implant Co., Ltd..
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

{13}------------------------------------------------

OSTEM
IMPLANT

Dsstem Implant Co.. Ltd

-16. Bansong-ro 513beon-gil. Haeundae-gu. Busan. Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com

therefore, the proposed device is substantially equivalent to the predicate device. Summary of Non-clinical Performance Testing 7.

Non-clinical testing data are submitted to demonstrate substantial equivalence.

Biocompatibility Evaluation

Biocompatibility testing was considered followed the FDA Guidance Document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, " The SS Abutment System has same materials, manufacturer, manufacturing process etc., as predicate devices. Therefore, we didn't conduct additional biocompatibility test.

Sterilization Validation and Shelf-life

Validation of the gamma irradiation process was previously conducted for the predicated device. There has been no change to the manufacturing or sterilization processes since then and the subject devices do not present a new worst case for sterilization validation; therefore, additional validation is not required. In addition, SS Abutment System are made with titanium and titanium alloy we don't consider about shelf life of material by itself because this metal is widely known that it generally has no adversely affect by aging. Therefore we certify that product such like metal has no shelf life.

Mechanical Properties

Fatigue testing was considered according to the FDA Guidance Document Guidance for Industry and FDA Staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment and ISO 14801 standard with the worst case scenario. The subject devices do not provide a new worst case for fatigue testing because they are all straight.

MR Compatibility

Non-clinical worst-case MRI review was performed to evaluate the Subject device components in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.

8. Summary of Clinical Testing

No clinical studies are submitted.

9. Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, Osstem Implant Co., Ltd. concludes that SS Abutment System is substantially equivalent to the predicated devices as herein.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)