K Number
K232102

Validate with FDA (Live)

Date Cleared
2023-09-15

(63 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ClearPoint Array System (Version 1.2) is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI in conjunction with MR imaging. The ClearPoint Array System (Version 1.2) is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices.

Device Description

The ClearPoint Array System is comprised of a workstation laptop with workstation software, the SMARTGrid™ MRI-Guided Planning Grid, the SMARTFrame™ Array MRI-Guided Trajectory Frame, SmartFrame Array Reducer Tube Kit, the ClearPoint™ Accessory Kit, the SMARTFrame™ Array Thumb Wheel Extension Set, and the MRI Neuro Procedure Drape. The ClearPoint Array Workstation includes the following:

  1. ClearPoint Array Workstation Software (for trajectory planning and monitoring)
  2. Laptop Computer
    The hardware components of the ClearPoint Array System are the SMARTFrame Array and accessories, and are listed below. They are all single use devices provided sterile. Beyond the changes described above, there have been no modifications to the hardware compared to the last cleared version of the device (K214040).
  3. SMARTFrame Array Pack
    a. SMARTFrame Array (adjustable trajectory frame to guide and hold the neurosurgical tools, includes Probe Adapter, Tracker Rod)
    b. SMARTFrame Array Scalp Mount Base (includes fiducials, titanium screws, and titanium standoff pins)
    c. Entry Point Locator
    d. Targeting Stem
    e. Centering Device
    f. Dock
    g. Device Lock (2 different diameters)
    h. Screwdriver
    i. 2.1-mm Guide Tube
    j. 4.5 Center Drill Guide
    k. 4.5 Offset Drill Guide
    l. 3.4-mm Drill Reducer Tube
    m. Center Insertion Guide
    n. Offset Insertion Guide
  4. SmartFrame Array Thumb Wheel Extension Set for the trajectory frame.
  5. SmartFrame Array Guide Tube Kit
    a. 1.7-mm Guide Tube
    b. 0.5-mm Guide Tube and Device Lock
    c. 3.1-mm Guide Tube and Device Lock
    d. SmartFrame Array Guide Tubes (sold separately)
    e. 7.9mm Center and Offset Device Guides
    f. 5.4mm Center and Offset Device Guides
    Common Components to the ClearPoint System are listed below. These components have not been modified since their clearance (K214040, K200097, K100836, K091343).
  6. SMARTGrid Pack (interacts with the Software to determine the desired location of the burr hole) (K100836)
    a. Marking Grid
    b. Marking Tool
  7. Accessory Pack (K200097)
    a. Peel away sheath
    b. Stylet
    c. Depth Stop
    d. Ruler
  8. MRI Neuro Procedure Drape (K091343)
AI/ML Overview

The provided text describes the ClearPoint Array System (Version 1.2) and its non-clinical testing for substantial equivalence to a predicate device. It primarily focuses on software modifications and accuracy specifications.

Here's an analysis of the acceptance criteria and study proving the device meets them, based only on the provided text:

Key Takeaway: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical testing. It does not describe a clinical study comparing human reader performance with and without AI, or a standalone AI performance study. The "device meets acceptance criteria" refers to the non-clinical performance benchmarks.


1. A table of acceptance criteria and the reported device performance

The document presents performance validation data as part of the non-clinical testing, which serves as the acceptance criteria for the accuracy of the system.

Performance ValidationAcceptance Criteria (Implicit from Reported Performance)Reported Device Performance (Mean)Reported Device Performance (99% CI)
Positional Error (mm)
X, Y, ZNot explicitly stated as a separate "acceptance criteria" column, but the reported 99% CI implies the acceptable range.X: 0.78, Y: 1.52, Z: -1.41X: 1.14, Y: 1.94, Z: -2.08
Angular Error (deg.)
Not explicitly stated as a separate "acceptance criteria" column, but the reported 99% CI implies the acceptable range.0.67°0.85°

Note on Acceptance Criteria: The document states, "The results of all testing met the acceptance criteria and demonstrated that the modified ClearPoint Array Software complies with all design specifications and performs as expected." However, the specific numerical acceptance criteria thresholds (e.g., "positional error must be less than X mm") are not explicitly listed in a separate column from the reported performance. Instead, the reported performance (especially the 99% CI) is the validation against which the "acceptance criteria" were met. The implication is that these measured values fell within the pre-defined acceptable limits for each parameter.


2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The document mentions "Accuracy testing was performed," but does not specify the sample size (e.g., number of measurements, number of trajectories, or number of cases) used for this testing.
  • Data Provenance: Not specified. Given it's non-clinical testing for a medical device 510(k), it's typically laboratory-based simulation/phantom data, not patient data from a specific country or collected retrospectively/prospectively.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. This was non-clinical accuracy testing of a stereotaxic guidance system, not a study involving human interpretation of medical images or expert adjudication of clinical outcomes. The "ground truth" for positional and angular accuracy would have been established by engineering measurements against known physical references.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable. This was non-clinical accuracy testing, not a study requiring human adjudication of results.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not conducted or described. The document focuses on the mechanical/software accuracy of the stereotactic guidance system, not on AI assistance for human image readers. The "AI" concept as typically understood in medical image analysis (e.g., for detection or diagnosis) is not relevant to this device's described function as a stereotaxic instrument.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • A form of "standalone" testing was done, but it's for the software components of a stereotaxic guidance system, not a diagnostic AI algorithm. The document states:
    • "ClearPoint Neuro performed extensive Non-Clinical Verification Testing to evaluate the safety and performance of the software components of ClearPoint Array System (Version 1.2)."
    • Specific tests included: "Automated Verification," "Integrated System Verification," "Localization Verification," and "Regression Test Verification."
    • "Accuracy testing was performed to confirm that modifications included in ClearPoint Array 1.2 did not cause any unexpected changes in the accuracy specifications of the software, with successful results."
    • This "accuracy testing" (Table 5-1) represents the "algorithm only" performance for positional and angular accuracy of the guidance system's calculations.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The ground truth for the non-clinical accuracy testing would have been physical measurements or known engineering specifications from a controlled test environment (e.g., phantom studies with precisely known target locations and trajectories). It would not be expert consensus, pathology, or outcomes data, as this is a device for guidance during neurological procedures, not for diagnosis.

8. The sample size for the training set

  • Not applicable. The document describes non-clinical verification testing of a stereotaxic guidance system's software and hardware, not a machine learning model that requires a "training set."

9. How the ground truth for the training set was established

  • Not applicable. As no training set was described, there's no ground truth establishment for it.

{0}------------------------------------------------

September 15, 2023

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

ClearPoint Neuro, Inc. Brennan Sullivan Regulatory Affairs Manager 5 Musick Irvine, California 92618

Re: K232102

Trade/Device Name: ClearPoint Array System (Version 1.2) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: August 18, 2023 Received: August 18, 2023

Dear Brennan Sullivan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the name "Adam D. Pierce -S" on the left side of the image. On the right side of the image, it says "Digitally signed by Adam D. Pierce -S" followed by the date "2023.09.15" and the time "16:20:45 -04'00'". The text on the right side of the image indicates that the document was digitally signed by Adam D. Pierce on September 15, 2023.

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232102

Device Name ClearPoint Array System (Version 1.2)

Indications for Use (Describe)

The ClearPoint Array System (Version 1.2) is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI in conjunction with MR imaging. The ClearPoint Array System (Version 1.2) is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices.

Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary for the ClearPoint Neuro ClearPoint Array System (Version 1.2)

1. SUBMITTER/510(K) HOLDER

ClearPoint Neuro, Inc. 5 Musick Irvine, CA 92618 Contact Person: Mary McNamara-Cullinane Telephone: 508-446-1830

Date Prepared: August 18, 2023

2. DEVICE INFORMATION

Name of Device:ClearPoint Array System (Version 1.2)
Common or Usual Name:ClearPoint Array System
Classification:Neurological Stereotaxic Instrument, 21CFR 882.4560
Regulatory Class:Class II
Product CodeHAW

3. PREDICATE DEVICES

• ClearPoint Array System 1.1 SoftwareK214040
-------------------------------------------------
  • ClearPoint Array System 1.0 Software K202575

4. DEVICE DESCRIPTION

The ClearPoint Array System is comprised of a workstation laptop with workstation software, the SMARTGrid™ MRI-Guided Planning Grid, the SMARTFrame™ Array MRI-Guided Trajectory Frame, SmartFrame Array Reducer Tube Kit, the ClearPoint™ Accessory Kit, the SMARTFrame™ Array Thumb Wheel Extension Set, and the MRI Neuro Procedure Drape. The ClearPoint Array Workstation includes the following:

    1. ClearPoint Array Workstation Software (for trajectory planning and monitoring)
    1. Laptop Computer

The hardware components of the ClearPoint Array System are the SMARTFrame Array and accessories, and are listed below. They are all single use devices provided sterile. Beyond the changes described above, there have been no modifications to the hardware compared to the last cleared version of the device (K214040).

    1. SMARTFrame Array Pack

{4}------------------------------------------------

  • SMARTFrame Array (adjustable trajectory frame to guide and hold the a. neurosurgical tools, includes Probe Adapter, Tracker Rod)
  • SMARTFrame Array Scalp Mount Base (includes fiducials, titanium b. screws, and titanium standoff pins)
  • Entry Point Locator C.
  • Targeting Stem d.
  • Centering Device e.
  • t. Dock
  • Device Lock (2 different diameters) g.
  • Screwdriver h.
  • i. 2.1-mm Guide Tube
  • i. 4.5 Center Drill Guide
  • k. 4.5 Offset Drill Guide
  • 3.4-mm Drill Reducer Tube 1.
  • m. Center Insertion Guide
  • n. Offset Insertion Guide
    1. SmartFrame Array Thumb Wheel Extension Set for the trajectory frame.
  • SmartFrame Array Guide Tube Kit 3.
    • a. 1.7-mm Guide Tube
    • b. 0.5-mm Guide Tube and Device Lock
    • 3.1-mm Guide Tube and Device Lock C.
    • d. SmartFrame Array Guide Tubes (sold separately)
    • 7.9mm Center and Offset Device Guides e.
    • f. 5.4mm Center and Offset Device Guides

Common Components to the ClearPoint System are listed below. These components have not been modified since their clearance (K214040, K200097, K100836, K091343).

    1. SMARTGrid Pack (interacts with the Software to determine the desired location of the burr hole) (K100836)
    • a. Marking Grid
    • b. Marking Tool
    1. Accessory Pack (K200097)
    • a. Peel away sheath
    • b. Stylet
    • c. Depth Stop
    • d. Ruler
      1. MRI Neuro Procedure Drape (K091343)

న. INTENDED USE

The ClearPoint® Array System (Version 1.2) is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint® Array System (Version 1.2) is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion including deep brain stimulation (DBS) lead

{5}------------------------------------------------

placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices.

NON-CLINICAL TESTING 6.

ClearPoint Neuro performed extensive Non-Clinical Verification Testing to evaluate the safety and performance of the software components of ClearPoint Array System (Version 1.2). The following software verification testing was performed:

  • Automated Verification ●
  • Integrated System Verification ●
  • Localization Verification
  • Regression Test Verification

The results of all testing met the acceptance criteria and demonstrated that the modified ClearPoint Array Software complies with all design specifications and performs as expected.

Accuracy testing was performed to confirm that modifications included in ClearPoint Array 1.2 did not cause any unexpected changes in the accuracy specifications of the software, with successful results. Table 5-1 outlines the demonstrated accuracy specifications of ClearPoint Array System.

PerformanceValidationPositional Error (mm)Angular Error (deg.)
ClearPointArray SystemMean(x,y,z)Std. Dev.99% CIMeanStd. Dev.99% CI
0.780.391.140.67°0.14°0.85°
1.520.461.94
-1.410.73-2.08

Table 5-1: ClearPoint Array Accuracy Specifications

Cl = confidence interval

7. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The technological characteristics of the Array 1.2 Software are essentially identical to the predicates. The purpose of this Special 510(k) is to incorporate an update to the ClearPoint Array software. In addition to the functions of the predicate device, the ClearPoint Array System (Version 1.2) software includes the following modifications:

  • . Support for parallel trajectory planning and realization using the guide channels of the ClearPoint Array® frame, such that devices can be inserted using guide channel positions without the need to adjust the frame angulation between insertions. This feature set includes:
    • Ability to define one or more parallel trajectory paths corresponding to o physical guide channels on the frame.

{6}------------------------------------------------

  • Capabilities to visualize the pre-mount orientation of the frame base(s) o needed to achieve the parallel trajectories corresponding to the guide channel positions.
  • Prescription of adjustments needed to align the frame(s) along any parallel o traiectory path.
  • Instructions for inserting device(s) along one or more parallel trajectory o paths.
  • . Improved trajectory definition and review tools, including the ability to visualize pre-operatively planned trajectories in a three-dimensional scene, ability to modify target/entry points using coordinate values and ability to copy previously defined trajectory paths.
  • Image visualization and rendering performance improvements. ●
  • . Improved automatic fusion of image sets that are acquired in different frames of reference.
  • A new set of interactive viewport tools which provide a richer functional toolset . offering.
  • Translating the software user interface to support nine additional languages. ●

There have been no modifications to any of the hardware components of the predicate device. In addition, the minor differences in the newer software version do not alter the fundamental clinical purpose or present different questions of safety or effectiveness as compared to the predicate. The final software version has been fully validated. Thus, the ClearPoint Array System (version 1.2) is substantially equivalent to the ClearPoint Array System (K214040).The ClearPoint Array System (version 1.2) is substantially equivalent to the predicate The ClearPoint Array System (K214040) because the subject device raises no new issues of safety and effectiveness, meets all test specifications, and the non-clinical testing performed demonstrates that the subject devices is as effective, and performs as well as the predicate device. A side-by-side comparison of the subject device to the predicate can be found below in Table 5-2.

CharacteristicProposed ClearPoint Array System (v1.2)ClearPoint Array System (v1.1)K214040
Classification21 C.F.R. § 882.456021 C.F.R. § 882.4560
Product CodeHAWHAW
Indications for UseThe ClearPoint® Array System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint® Array System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners andIdentical
MR Conditional implants and devices.
EnvironmentMRI Suite or ORIdentical
SMARTGrid PackMRI Planning Grid & Marking tool.Identical
SMARTFrame PackSMARTFrame Array Adjustable Trajectory Frame, ScalpMount Base, Entry Point Locator, Targeting Stem, CenteringDevice, Dock, Device Lock,screwdriver, 2.1-mm Guide Tube, Center Drill Guide, OffsetDrill Guide, 3.4-mm Drill Reducer Tube, Center InsertionGuide, Offset Insertion GuideIdentical
Hand ControllerThumbwheel ExtensionIdentical
Accessory packPeel away sheath, Stylet, Depth stop, rulerIdentical
Drill GuidesDrill Guide, 4.5-mm (included in SMARTFrame Pack)Identical
Targeting Cannula IDNo ID, Targeting Stem is completely fluid-filledIdentical
Targeting Cannula MaterialPEEKIdentical
Guide Tube/Device Guide/Drill Guide IDDrill Guide ID: 4.5 mm (included inSmartFrame Pack)Drill Guide Tube ID: 3.4 mm (included inSmartFrame Pack)Guide Tube ID: 3.2 mmGuide Tube ID: 2.5 mmGuide Tube ID: 2.1 mmGuide Tube ID: 1.7 mmIdentical
Targeting Accuracy≤ 2.0 mmIdentical
PackagingSMARTFrame Array Pack: PETG Tray with sealed TyvekLid inside of a Sealed Mylar-Tyvek PouchSMARTFrame Array ThumbwheelExtension:PETG Tray sealed in a Double Mylar-Tyvek PouchSMARTFrame Array Guide Tube Kit:Sealed Double Mylar-Tyvek PouchesIdentical
SoftwareVersion 1.2Version 1.1
Operating SystemWindows 10, Windows 11Windows 10
Programming LanguagesVisual C#Visual C++Identical
Visualization Software ToolkitFovia HDVR®Identical
DICOM ToolkitMergeCOM-3 Dicom Toolkit®Identical
Image Registration FrameworkInsight Toolkit (ITK)CCRNeuroTargetingRegistrationLibrary

Table 5-2: Side-by-side comparison of ClearPoint Array 1.2 Software with Parent Device

{7}------------------------------------------------

{8}------------------------------------------------

8. CONCLUSIONS

The ClearPoint Array System (Version 1.2) is as safe and effective as the predicate device. The ClearPoint Array System (Version 1.2) has the same indications for use, technological characteristics, and principles of operation as its predicate device. In addition, the minor technological differences between the subject device and its predicate devices raise no new issues of safety or effectiveness. Thus, the ClearPoint Array System (Version 1.2) is substantially equivalent to the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).