K Number
K202575

Validate with FDA (Live)

Date Cleared
2021-01-22

(140 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ClearPoin® Array System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint® Array System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insettion including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices.

Device Description

The ClearPoint Array System is comprised of a workstation laptop with workstation software, the SMARTGrid™ MRI-Guided Planning Grid, the SMARTFrame™ Array MRI-Guided Trajectory Frame, SmartFrame Array Reducer Tube Kit, the ClearPoint™ Accessory Kit, the SMARTFrame™ Array Thumb Wheel Extension Set, and the MRI Neuro Procedure Drape.

A pre-alignment may be used when it is desired to obtain an approximate alignment using a Surgical Navigation System prior to performing final alignment and device placement in the MR Scanner using real-time MR Images. Pre-alignment using an SNS is not intended to provide accurate stereotactic placement. The final alignment and insertion must be performed using real-time MR images with ClearPoint Array software prior to inserting a device in the brain. The SMARTGrid™ MRI-Guided Planning Grid and associated Marking Tool are designed to assist the physician to precisely position the entry hole as called out in the trajectory planning software.

The SMARTFrame™ Array MRI-Guided Trajectory Frame is an adjustable trajectory frame that provides the quidance and fixation for neurosurgical tools. MRI visible fluids in the Targeting Stem along with the fiducial markers in the frame allows for trajectory feedback when the physician views the MR images, makes changes and confirms the subsequent MR images. The ClearPoint Array System can be used with any MRI-compatible head fixation frame to immobilize the patient's head with respect to the scanner table, as well as with any imaging coil(s) (supplied by scanner manufacturers) that meet the physician's desired imaging quality. ClearPoint Neuro also supplies an optional head fixation frame that can be used with the ClearPoint Array System.

The ClearPoint System Array Workstation includes the following:

  1. ClearPoint Workstation Software (for trajectory planning and monitoring)
  2. Laptop Computer

The hardware components of the ClearPoint Array System are the SMARTFrame Array and accessories. Some of the accessories are common components with the ClearPoint System (K200079). They are all single use devices that are provided sterile. They include the following:
1 SMARTFrame Array Pack
a. SMARTFrame Array (adjustable trajectory frame to guide and hold the neurosurgical tools, includes Probe Adapter and Tracker Rod)
b. SMARTFrame Array Scalp Mount Base (includes fiducials, titanium screws, and support pins)
c. Entry Point Locator
d. Targeting Stem
e. Centering Device
f. Dock
q. Device Lock (2 different diameters)
h. Screwdriver
i. 2.1-mm Guide Tube
j. Center Drill Guide
k. Offset Drill Guide

  1. 3.4-mm Drill Reducer Tube
    m. Center Insertion Guide
    n. Offset Insertion Guide
    2 SmartFrame Array Thumb Wheel Extension Set for the trajectory frame
    3 SmartFrame Array Guide Tube Kit
    a. 1.7-mm Guide Tube
    b. 2.5-mm Guide Tube and Device Lock
    c. 3.2-mm Guide Tube and Device Lock

Common components to ClearPoint System include:

  1. SMARTGrid Pack (interacts with the Software to determine the desired location of the burr hole)
    a. Marking Grid
    b. Marking Tool
  2. Accessory pack
    a. Peel away sheath
    b. Stylet
    c. Depth Stop
    d. Ruler
  3. MRI Neuro Procedure Drape
AI/ML Overview

The provided document is a 510(k) premarket notification summary for the ClearPoint Array System. This type of submission focuses on demonstrating "substantial equivalence" to a predicate device, rather than proving a device "meets acceptance criteria" in the way one might for a novel AI/ML device or a device with new performance claims.

Therefore, the document does not contain the typical information required to answer all parts of your request, such as a multi-reader multi-case study, detailed ground truth establishment for a test set, or separate training set details, as these are more common for AI/ML device submissions with explicit performance claims against real-world clinical outcomes.

However, based on the information available, I can extract what is relevant to performance and "acceptance":

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in the format of a pre-defined performance goal that the device must achieve (e.g., "sensitivity must be >X%"). Instead, it provides performance validation values for key navigational accuracy metrics, which likely represent measurements taken to demonstrate the device's capabilities and support its substantial equivalence claim. The predicate device's targeting accuracy serves as a benchmark for comparison.

Metric (ClearPoint Array System)Reported PerformancePredicate Device Performance (Acceptance Benchmark/Comparison)
Positional Error (Mean)0.78 mm (x,y,z)≤ 1.5 mm
Positional Error (99% CI)1.14 mm (x,y,z)-
Angular Error (Mean)0.67°-
Angular Error (99% CI)0.85°-

(Note: The second row for positional error (1.52, 0.46, 1.94, and -1.41, 0.73, -2.08) appears to be additional data points for positional error (potentially x, y, z components or separate measurements), but the document does not clarify their exact meaning beyond being under 'Positional Error (mm)' and 'Mean', 'Std. Dev', '99% CI'. For clarity, I'm using the first row as the primary 'overall' mean positional error provided.)

The primary "acceptance criterion" implicit in this type of submission is that the performance of the new device is substantially equivalent to the predicate device, especially considering the predicate's stated "Targeting Accuracy ≤ 1.5 mm." The reported mean positional errors (0.78mm and other values) are well within this 1.5 mm threshold, indicating that this aspect of performance is considered acceptable and equivalent or improved.

2. Sample size used for the test set and the data provenance

The document states: "Design Verification and Validation was performed relative to these specifications and with acceptable results. These tests included verification of physical, performance, and safety requirements, as well as benchtop accuracy testing."

  • Sample Size: The document does not explicitly state the sample size (number of measurements or trials) used for the benchtop accuracy testing.
  • Data Provenance: The data is described as "benchtop accuracy testing." This implies controlled laboratory conditions, not human subject data from a specific country or clinical setting. It is not retrospective or prospective clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. For a stereotactic guidance system, "ground truth" for accuracy is typically established by precise measurement tools and phantoms in a controlled engineering or physics lab setting, rather than by expert human interpretation of images. The device's performance is measured against known physical coordinates or angles.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as ground truth is established via physical measurements and not human adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This type of study is typically performed for diagnostic AI/ML devices where the AI provides assistance to a human reader, to demonstrate improvement in reader performance. The ClearPoint Array System is a surgical guidance device, not a diagnostic AI/ML tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, in a sense. The "Performance Data" table ("Positional Error" and "Angular Error") represents the standalone performance of the device in a controlled, benchtop setting, measuring its inherent accuracy in guiding and positioning. It's the device's mechanical and software accuracy demonstrated without a human actively interpreting images or making a diagnostic decision in a clinical context.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The ground truth for the "Performance Data" (positional and angular accuracy) would be based on physical precision measurements using highly accurate instruments (e.g., optical trackers, coordinate measuring machines) in a controlled laboratory environment with phantoms that have known, precise target locations.

8. The sample size for the training set

Not applicable. This is not an AI/ML device that requires a training set in the typical sense for machine learning. The software modifications mentioned are likely re-engineering and feature additions for workflow and UI, not learning algorithms.

9. How the ground truth for the training set was established

Not applicable, as there is no mention of a training set for machine learning.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 22, 2021

ClearPoint Neuro, Inc. % John Smith Partner Hogan Lovells US LLP 555 Thirteenth St. NW Washington, District of Columbia 20004

Re: K202575

Trade/Device Name: ClearPoint Array System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: December 23, 2020 Received: December 23, 2020

Dear John Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K202575

Device Name

ClearPoint Array System

Indications for Use (Describe)

The ClearPoin® Array System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint® Array System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insettion including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conductor sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

{4}------------------------------------------------

510(k) SUMMARY

ClearPoint Neuro Inc.'s ClearPoint Array System

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared ClearPoint Neuro Inc.

5 Musick

Irvine, CA 92618

Phone: (949) 900-6833 Facsimile: (49) 900-6834 Contact Person: Peter Piferi, C.O.O Date Prepared: January 22, 2021

Name of Device: ClearPoint Array System

Common or Usual Name: ClearPoint Array System

Classification Name: Neurological Stereotaxic Instrument, 21 CFR 882.4560

Regulatory Class: Class II

Product Code: HAW

Predicate Devices

ClearPoint System and Accessories (K200079)

Intended Use / Indications for Use

The ClearPoint® Array System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint® Array System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices.

Device Description

The ClearPoint Array System is comprised of a workstation laptop with workstation software, the SMARTGrid™ MRI-Guided Planning Grid, the SMARTFrame™ Array MRI-Guided Trajectory Frame, SmartFrame Array Reducer Tube Kit, the ClearPoint™ Accessory Kit, the SMARTFrame™ Array Thumb Wheel Extension Set, and the MRI Neuro Procedure Drape.

A pre-alignment may be used when it is desired to obtain an approximate alignment using a Surgical Navigation System prior to performing final alignment and device placement in the MR Scanner using real-time MR Images. Pre-alignment using an SNS is not intended to provide accurate stereotactic placement. The final alignment and insertion must be performed using real-time MR images with ClearPoint Array software prior to inserting a device in the brain. The SMARTGrid™ MRI-Guided Planning Grid and associated Marking Tool are designed to assist the physician to precisely position the entry hole as called out in the trajectory planning software.

{5}------------------------------------------------

The SMARTFrame™ Array MRI-Guided Trajectory Frame is an adjustable trajectory frame that provides the quidance and fixation for neurosurgical tools. MRI visible fluids in the Targeting Stem along with the fiducial markers in the frame allows for trajectory feedback when the physician views the MR images, makes changes and confirms the subsequent MR images. The ClearPoint Array System can be used with any MRI-compatible head fixation frame to immobilize the patient's head with respect to the scanner table, as well as with any imaging coil(s) (supplied by scanner manufacturers) that meet the physician's desired imaging quality. ClearPoint Neuro also supplies an optional head fixation frame that can be used with the ClearPoint Array System.

The ClearPoint System Array Workstation includes the following:

    1. ClearPoint Workstation Software (for trajectory planning and monitoring)
    1. Laptop Computer

The hardware components of the ClearPoint Array System are the SMARTFrame Array and accessories. Some of the accessories are common components with the ClearPoint System (K200079). They are all single use devices that are provided sterile. They include the following:

  • 1 SMARTFrame Array Pack
    • a. SMARTFrame Array (adjustable trajectory frame to guide and hold the neurosurgical tools, includes Probe Adapter and Tracker Rod)
    • b. SMARTFrame Array Scalp Mount Base (includes fiducials, titanium screws, and support pins)
    • Entry Point Locator C.
    • d. Targeting Stem
    • e. Centering Device
    • f. Dock
    • Device Lock (2 different diameters) q.
    • h. Screwdriver
    • i. 2.1-mm Guide Tube
    • j. Center Drill Guide
    • Offset Drill Guide k.
      1. 3.4-mm Drill Reducer Tube
    • m. Center Insertion Guide
    • n. Offset Insertion Guide
  • 2 SmartFrame Array Thumb Wheel Extension Set for the trajectory frame
  • 3 SmartFrame Array Guide Tube Kit
    • a. 1.7-mm Guide Tube
    • b. 2.5-mm Guide Tube and Device Lock
    • c. 3.2-mm Guide Tube and Device Lock

Common components to ClearPoint System include:

    1. SMARTGrid Pack (interacts with the Software to determine the desired location of the burr hole)
    • a. Marking Grid

{6}------------------------------------------------

  • b. Marking Tool
    1. Accessory pack
    • a. Peel away sheath
    • b. Stylet
      • c. Depth Stop
      • d. Ruler
    1. MRI Neuro Procedure Drape

Technological Characteristics

The purpose of the 510(k) is to modify the SmartFrame Array by:

    1. Material change and addition of small geometric features on the Adjustable Trajectory Frame (ATF) and Scalp Mount Base.
    1. Replacing the ATF's continuous X-Y adjustment with a discrete positional adjustment.
    1. Modification of the targeting cannula material and construction.
    1. Providing a passive guide and rod with the ATF that allows users to insert a surgical navigation system Probe or Universal Tracker and be held in the ATF.
    1. Minor updates to device packaging.
    1. Software modifications for:
    • New workflow options, user interface, image alignment methods. ●
    • Introduction of new software tools for volume of interest comparison. .
ClassificationClearPoint System (Predicate)K200079ClearPoint Array System (Subject)
21 C.F.R. § 882.456021 C.F.R. § 882.4560
Product codeHAWHAW
Intended UseThe ClearPoint® System is intended toprovide stereotactic guidance for theplacement and operation of instrumentsor devices during planning andoperation of neurological procedureswithin the MRI environment and inconjunction with MR imaging. TheClearPoint® Array System is intendedas an integral part of procedures thathave traditionally used stereotacticmethodology. These proceduresinclude biopsies, catheter and electrodeThe ClearPoint® Array System isintended to provide stereotacticguidance for the placement andoperation of instruments or devicesduring planning and operation ofneurological procedures within the MRIenvironment and in conjunction with MRimaging. The ClearPoint® Array Systemis intended as an integral part ofprocedures that have traditionally usedstereotactic methodology. Theseprocedures include biopsies, catheter
insertion including deep brainstimulation (DBS) lead placement. TheSystem is intended for use only with 1.5and 3.0 Tesla MRI scanners and MRConditional implants and devices.and electrode insertion including deepbrain stimulation (DBS) lead placement.The System is intended for use onlywith 1.5 and 3.0 Tesla MRI scannersand MR Conditional implants anddevices.
EnvironmentMRI SuiteMRI Suite or OR
SterilizationEO 10-6 SALEO 10-6 SAL
SMARTGrid PackMRI Planning Grid & Marking toolMRI Planning Grid & Marking tool
SMARTFrame PackSmartFrame XG Adjustable TrajectoryFrame with Targeting Cannula, SkullMount Base, Screwdriver, Dock, DeviceLock, Centering RingSMARTFrame Array AdjustableTrajectory Frame, Scalp Mount Base,Entry Point Locator, Targeting Stem,Centering Device, Dock, Device Lock,screwdriver, 2.1-mm Guide Tube,Center Drill Guide, Offset Drill Guide,3.4-mm Drill Reducer Tube, CenterInsertion Guide, Offset Insertion Guide
Hand ControllerThumbwheel ExtensionThumbwheel Extension
Accessory packPeel away sheath, Stylet, Depth stop,rulerPeel away sheath, Stylet, Depth stop,ruler
Drill GuidesDrill Guide, 3.4-mmDrill Guide, 4.5-mmDrill Guide, 6.0-mmDrill Guide, 4.5-mm (included inSMARTFrame Pack)
Targeting Cannula ID.083" (2.1 mm)No ID, Targeting Stem is completelyfluid-filled
Targeting CannulaMaterialUltem and PEEKPEEK
Guide Tube/DeviceGuide/Drill Guide IDDrill Guide ID: 4.5 mmDrill Guide ID: 6.0 mmDevice/Drill Guide ID: 3.4 mmDevice Guide ID: 3.2 mmDevice Guide ID: 2.5 mmDevice Guide ID: .093"Guide Tube ID: 1.7 mmDrill Guide ID: 4.5 mm (included inSmartFrame Pack)Drill Guide Tube ID: 3.4 mm (included inSmartFrame Pack)Guide Tube ID: 3.2 mmGuide Tube ID: 2.5 mmGuide Tube ID: 2.1 mmGuide Tube ID: 1.7 mm
PackagingSMARTFrame Pack: PETG Tray withsealed Tyvek Lid inside of a SealedMylar-Tyvek PouchSMARTFrame Array Pack: PETG Traywith sealed Tyvek Lid inside of a SealedMylar-Tyvek Pouch
Thumbwheel Extension: PETG Traysealed in a Double Mylar-Tyvek PouchSMARTFrame Array ThumbwheelExtension: PETG Tray sealed in aDouble Mylar-Tyvek Pouch
Guide Tubes and Drill Guides: SealedDouble Mylar-Tyvek PouchesSMARTFrame Array Guide Tube Kit:Sealed Double Mylar-Tyvek Pouches
Targeting Accuracy≤ 1.5 mm≤ 2.0 mm
SoftwareClearPoint 2.0.xClearPoint Array 1.0.x

Summary of Technological Characteristics of the Device Compared to the Predicate Device

{7}------------------------------------------------

{8}------------------------------------------------

Biocompatibility

All patient contacting components remain the same as the predicate ClearPoint System except for the SMARTFrame Array Base. For this reason, biocompatibility was only performed on the SMARTFrame Array Base. The SmartFrame Lite Base was found to be biocompatible per ISO 10993-1 as summarized below.

Test DescriptionResultsPass/Fail
CytotoxicityISO 10993-5The test article extract showed no evidence of causing cell lysis or toxicity. The test article extract met the requirements of the test since the grade was less than a grade 2 (mild reactivity).Pass
IrritationISO 10993-11There was no erythema and no edema observed on the skin of the animals treated with the Saline and Sesame Oil extracts. For both test article extracts, the Primary Irritation Index was calculated to be 0.0 and the irritation response was categorized as negligible.Pass
SensitizationISO 10993-11The test article extracts showed no evidence of causing delayed dermal contact sensitization in the guinea pig. The test article was not considered a sensitizer in the guinea pig maximization test.Pass
Systemic Toxicity ISO 10993-11There was no mortality or evidence of systemic toxicity form the extracts injected into mice. Each test article extract met the requirements of the study.Pass
Material MediatedPyrogenicityISO 10993-11The total rise of rabbit temperatures during the 3 hour observation period was within acceptable USP requirements. The test article met the requirements for the absence of pyrogens.Pass

Performance Data

The modifications implemented in ClearPoint Array System were conducted in conformance with the company's design control procedures. Design inputs provided the requirements for the respective product specifications. Design Verification and Validation was performed relative to these specifications and with acceptable results. These tests included verification of physical, performance, and safety requirements, as well as benchtop accuracy testing. The navigational accuracy validation values are presented in the table below.

PerformanceValidationPositional Error (mm)Angular Error (deg.)
ClearPointArray SystemMean(x,y,z)Std. Dev.99% CIMeanStd. Dev.99% CI
0.780.391.140.67°0.14°0.85°
1.520.461.94
-1.410.73-2.08

*Cl = confidence interval

{9}------------------------------------------------

Risk analysis was performed with mitigation of all identified risks to acceptable levels. The tests and risk analysis demonstrated that the ClearPont Array System functions as intended and is substantially equivalent to the legally marketed ClearPoint System.

Substantial Equivalence

The modifications implemented in ClearPoint Array System were conducted in conformance with the company's design control procedures. Performance testing established the substantial equivalence of the ClearPoint Array System to the predicate ClearPoint System, including design verification testing.

The ClearPoint Array System has the same intended use and similar indications for use, technological characteristics and principles of operation as the predicate ClearPoint System. The minor differences in the workflow and usability do not raise new and different questions of safety and effectiveness.

Conclusions

The ClearPoint Array System has the same intended use and similar indications for use, technological characteristics and principles of operation as the predicate ClearPoint System. The differences between the ClearPoint Array System and the predicate raise no new issues of safety and effectiveness. Thus, the ClearPoint Array System is substantially equivalent to the previously cleared ClearPoint System.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).