K Number
K232027

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-04-04

(272 days)

Product Code
Regulation Number
864.7415
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CAPI 3 NEONAT Hb kit is intended as a qualitative screen for the detection of normal hemoglobins (F and A) and abnormal hemoglobins (S, C, E, D and Bart's) in blood from human new-born collected on filter paper. This analysis is performed by capillary electrophoresis with the CAPILLARYS 3 DBS instrument.

The CAPILLARYS 3 DBS instrument is a capillary electrophoresis based automated analyzer which performs a complete hemoglobin profile for the qualitative analysis of hemoglobins (A, F, S, C, D, E and Bart's). The assay is performed on the hemolysate of whole blood samples previously collected on filter paper.

The test result must be interpreted in conjunction with other biological and clinical findings. In case of abnormal hemoglobin presence, it should be confirmed by additional tests as per local recommendations. The device is intended for professional use only.

For In Vitro Diagnostic Use only.

Device Description

The CAPI 3 NEONAT Hb kit is intended for the detection of normal hemoglobins (F and A) and abnormal hemoglobins (S, C, E, D and Bart's) in blood from human new-born collected on filter paper. The resulting electrophoregrams are automatically evaluated for pattern abnormalities with identification of normal and pathological patterns.

The CAPILLARYS 3 DBS instrument uses the principle of capillary electrophoresis in free solution which is the most common form of capillary electrophoresis. With this technique, charged molecules are separated by their electrophoretic mobility in an alkaline buffer with a specific pH. Separation also occurs according to the electrolyte pH and electroosmotic flow.

The CAPILLARYS 3 DBS instrument has silica capillaries functioning in parallel allowing 12 simultaneous analyses. A sample dilution with hemolysing solution is prepared and injected by aspiration at the anodic end of the capillary. A high voltage protein separation is then performed and direct detection of the hemoglobins is made at the cathodic end of the capillary at 415 nm. which is the absorbance wavelength specific to hemoglobins. Before each run. the capillaries are washed with a wash solution and prepared for the next analysis with buffer.

The CAPILLARYS 3 DBS performs all sequences automatically to obtain a complete hemoglobin profile for the qualitative analysis of hemoglobins. The assay is performed on the hemolysate of whole blood samples previously collected on Guthrie filter paper and punched to obtain a paper circle.

By using alkaline pH buffer. normal and abnormal (or variant) hemoglobins are detected in the following order. from cathode to anode: C, A2, E, S, D, F, degraded F, A, degraded A and Bart's. Variants generated by the mutation of the y chain may appear in different zones of the electrophoretic pattern. The carbonic anhydrase is not visualized on the hemoglobin electrophoretic patterns with capillary electrophoresis.

AI/ML Overview

The provided text describes the analytical performance data for the CAPI 3 NEONAT Hb kit using the CAPILLARYS 3 DBS instrument. This device is intended for qualitative screening of normal and abnormal hemoglobins in newborn blood samples.

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are implicitly demonstrated through its performance in various analytical studies, primarily focusing on 100% identification pattern concordance and high positive/negative percent agreement (PPA/NPA) compared to a reference method.

Acceptance Criterion (Implicit)Reported Device Performance
Precision: Consistent identification of hemoglobin patterns across different capillaries, reagent lots, instruments, and days.Between capillaries, lots, instruments, and days: All tested samples (6 new-born whole blood samples with various patterns and 2 controls) showed 100% identification pattern concordance across all conditions. Total concordance (95% CI): - Between capillaries: 100% (96.2%; 100.0%) - Between lots: 100% (96.7%; 100.0%) - Between instruments: 100% (96.7%; 100.0%) - Between days: 100% (98.0%; 100.0%)
Analytical Specificity/Interference: No significant interference from common endogenous substances.No interference was detected from conjugated bilirubin (up to 40 mg/dL), unconjugated bilirubin (up to 40 mg/dL), and triglycerides (up to 1500 mg/dL).
Qualitative Method Comparison: High agreement with a commercially available capillary electrophoresis reference procedure.Internal Study: - Positive Percent Agreement (PPA): 100% (95% CI: 94.6%; 100.0%) - Negative Percent Agreement (NPA): 100.0% (95% CI: 94.9%; 100.0%) External Study No. 1 (USA): - PPA: 100.0% (95% CI: 98.1%; 100.0%) - NPA: 100.0% (95% CI: 98.2%; 100.0%) External Study No. 2 (Spain): - PPA: 100.0% (95% CI: 96.9%; 100.0%) - NPA: 100.0% (95% CI: 96.9%; 100.0%)

Study Details Proving Acceptance Criteria

  1. Sample sizes used for the test set and the data provenance:

    • Precision Studies:

      • Identification Pattern Concordance: 8 different samples (6 newborn whole blood samples with FA, FAS, FAC, FAD, FAE, FA Bart's patterns, and 2 controls with AF, AFSC patterns).
        • Between capillaries: 12 analyses per sample (total 96 analyses).
        • Between lots: 6 analyses per sample (total 114 analyses).
        • Between instruments: 6 analyses per sample (total 114 analyses).
        • Between days: 10 analyses per sample (total 190 analyses).
      • Precision using controls (CLSI EP05-A3): 2 control samples (AF and AFSC patterns).
        • Within-laboratory: 480 analyses per control (total 960).
        • Between-lots: 90 analyses per control (total 180).
        • Between-instruments: 90 analyses per control (total 180).
      • Data Provenance: Not explicitly stated for precision samples' geographic origin, but samples include "new-born whole blood samples" and controls. The CLSI guidelines are international standards.
    • Analytical Specificity/Interference: Not explicitly stated how many samples were used, but the study evaluated common interfering factors at specified concentrations.

    • Qualitative Method Comparison:

      • Internal Study: 138 different newborn whole blood samples (71 normal and 67 pathological with variants S, C, D, E, and Bart's).
        • Data Provenance: Samples provided by 5 laboratories in France, Thailand, and Panama. This study was likely retrospective, using banked samples.
      • External Study No. 1: 411 different newborn whole blood samples (210 normal samples and 201 pathological samples with variants S, C, D, E, and Bart's).
        • Data Provenance: Samples analyzed in a laboratory based in the United States of America. This study was likely retrospective.
      • External Study No. 2: 240 different newborn whole blood samples (120 normal samples and 120 pathological samples with variants S, C, D, E, and Bart's).
        • Data Provenance: Samples analyzed in a laboratory based in Spain. This study was likely retrospective.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document implies that the ground truth for the test set (qualitative method comparison) was established by a "commercially available capillary electrophoresis procedure for hemoglobin analysis (reference)." It does not mention the use of human experts for ground truth establishment. This suggests the reference method itself serves as the "gold standard" for comparison, not human interpretation in this context.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • No human adjudication method is described. The comparison is directly between the candidate device and the "reference procedure."
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC or human-in-the-loop study involving human reader improvement with AI assistance was performed or reported. This device is an in vitro diagnostic (IVD) instrument that provides qualitative results directly, not an AI-assisted diagnostic imaging tool requiring human interpretation improvement studies.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance data presented (precision, analytical specificity, method comparison) reflects the standalone performance of the CAPI 3 NEONAT Hb kit and CAPILLARYS 3 DBS instrument, as it is an automated IVD device. The results are automatically evaluated for pattern abnormalities with identification of normal and pathological patterns.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the qualitative method comparison studies was established by a "commercially available capillary electrophoresis procedure for hemoglobin analysis (reference)." This is an established laboratory method serving as the analytical gold standard.
  7. The sample size for the training set:

    • This document describes analytical performance studies of a medical device, not a machine learning or AI algorithm development. Therefore, there is no mention of a "training set" in the context of machine learning. The device's underlying technology is capillary electrophoresis, and its performance is validated through traditional IVD analytical studies, not typically through machine learning training and testing paradigms that would involve a distinct "training set."
  8. How the ground truth for the training set was established:

    • As this is not an AI/ML device in the sense of requiring a "training set" for model development, this question is not applicable to the information provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 4, 2024

Sebia Inc. Karen Anderson Director of Regulatory 1705 Corporate Drive Suite 400 Norcross, Georgia 30093

Re: K232027

Trade/Device Name: CAPI 3 NEONAT Hb, CAPILLARYS 3 DBS Instrument Regulation Number: 21 CFR 864.7415 Regulation Name: Abnormal hemoglobin assay Regulatory Class: Class II Product Code: GKA Dated: March 4, 2024 Received: March 5, 2024

Dear Karen Anderson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min Wu-S

Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232027

Device Name CAPI 3 NEONAT Hb CAPILLARYS 3 DBS instrument

Indications for Use (Describe)

The CAPI 3 NEONAT Hb kit is intended as a qualitative screen for the detection of normal hemoglobins (F and A) and abnormal hemoglobins (S, C, E, D and Bart's) in blood from human new-born collected on filter paper. This analysis is performed by capillary electrophoresis with the CAPILLARYS 3 DBS instrument.

The CAPILLARYS 3 DBS instrument is a capillary electrophoresis based automated analyzer which performs a complete hemoglobin profile for the qualitative analysis of hemoglobins (A, F, S, C, D, E and Bart's). The assay is performed on the hemolysate of whole blood samples previously collected on filter paper.

The test result must be interpreted in conjunction with other biological and clinical findings. In case of abnormal hemoglobin presence, it should be confirmed by additional tests as per local recommendations. The device is intended for professional use only.

For In Vitro Diagnostic Use only.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY (Summary of Safety and Effectiveness)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Submitter NameSebia. Inc.
Address1705 Corporate Drive Suite 400 Norcross. Georgia30093. USA
ContactKaren Anderson. Dir of Regulatory. Sebia Inc.Phone: 1-800-835-6497Fax: 770-446-8511Email: karen.anderson@sebia-usa.comMatthew C Wagner. Scientific Affairs SpecialistPhone 1-800-835-6497. 3752Fax 770-446-8511Email: matthew.wagner@sebia-usa.com
Date PreparedApril 03, 2024
ManufacturingSebiaParc Technologique Léonard de VinciRue Léonard de Vinci.CP 8010 LISSES. 91008 EVRY Cedex FRANCEPhone: (33) 1 69 89 80 80Fax: (33) 1 69 89 78 78
Product NameCAPI 3 NEONAT Hb kit using the CAPILLARYS 3 DBSinstrument
Common NameHEMOGLOBINS BY CAPILLARY ELECTROPHORESIS
Product Regulation No.21 CFR Part 864.7415 – Abnormal Hemoglobin Assay21 CFR Part 864.7415 – Abnormal Hemoglobin Assay. Control(510(k) exempt. FDA docket 2017-N-1129/1610 publisheDecember 30. 2019; Class II (special control) devices)
Product CodesGKA - Abnormal Hemoglobin Quantitation – Class IIJCM - Control. Hemoglobin. Abnormal – Class II (special control)devices. 510(k) exempt
Device classification andPanel ClassificationClass II. Hematology
Establishment Registration No.8023024

{4}------------------------------------------------

1. DEVICE DESCRIPTIONS

The CAPI 3 NEONAT Hb kit is intended for the detection of normal hemoglobins (F and A) and abnormal hemoglobins (S, C, E, D and Bart's) in blood from human new-born collected on filter paper. The resulting electrophoregrams are automatically evaluated for pattern abnormalities with identification of normal and pathological patterns.

The CAPILLARYS 3 DBS instrument uses the principle of capillary electrophoresis in free solution which is the most common form of capillary electrophoresis. With this technique, charged molecules are separated by their electrophoretic mobility in an alkaline buffer with a specific pH. Separation also occurs according to the electrolyte pH and electroosmotic flow.

The CAPILLARYS 3 DBS instrument has silica capillaries functioning in parallel allowing 12 simultaneous analyses. A sample dilution with hemolysing solution is prepared and injected by aspiration at the anodic end of the capillary. A high voltage protein separation is then performed and direct detection of the hemoglobins is made at the cathodic end of the capillary at 415 nm. which is the absorbance wavelength specific to hemoglobins. Before each run. the capillaries are washed with a wash solution and prepared for the next analysis with buffer.

The CAPILLARYS 3 DBS performs all sequences automatically to obtain a complete hemoglobin profile for the qualitative analysis of hemoglobins. The assay is performed on the hemolysate of whole blood samples previously collected on Guthrie filter paper and punched to obtain a paper circle.

By using alkaline pH buffer. normal and abnormal (or variant) hemoglobins are detected in the following order. from cathode to anode: C, A2, E, S, D, F, degraded F, A, degraded A and Bart's. Variants generated by the mutation of the y chain may appear in different zones of the electrophoretic pattern. The carbonic anhydrase is not visualized on the hemoglobin electrophoretic patterns with capillary electrophoresis.

2. REAGENTS

REAGENTS AND MATERIALS SUPPLIED IN THE CAPI 3 NEONAT Hb KIT

CAPI 3 NEONAT Hb KIT includes:PN 2506
CAPILLARYS HEMOGLOBIN(E) BUFFER(ready to use)3 vials, 700 mL each
Filters5 filters
CAPILLARYS 3 DBS MICROPLATES96 well-microplates (with bar codes)1 pack of 10 microplates

{5}------------------------------------------------

REAGENTS/SUPPLIES REQUIRED BUT NOT SUPPLIED IN THE CAPI 3 NEONAT Hb KIT

SUPPLIESSEBIA PRODUCT NUMBER
CAPI 3 NEONAT HEMOLYSING SOLUTION KIT(ready to use)(PN 2563), 2 vials of 70 mL each and2 pierceable caps
CAPI 3 NEONAT Hb AF CONTROL(lyophilized)(PN 4757), 2 vials and one boardingcartridge
CAPILLARYS 3 CAPICLEAN(concentrated solution)(PN 2060), 1 vial of 25 mL and 1pierceable cap
CAPILLARYS 3 WASH SOLUTION(stock solution)(PN 2062), 1 vial of 75 mL

OPITIONAL SUPPLIES

SUPPLIESSEBIA PRODUCT NUMBER
Hb AFSC CONTROL(lyophilized)(PN 4792) 1 vial of 0.5 mL

3. INDICATIONS FOR USE

INTENDED USE STATEMENT

The CAPI 3 NEONAT Hb kit is intended as a qualitative screen for the detection of normal hemoglobins (F and A) and abnormal hemoglobins (S, C, E, D and Bart's) in blood from human new-born collected on filter paper. This analysis is performed by capillary electrophoresis with the CAPILLARYS 3 DBS instrument.

The CAPILLARYS 3 DBS instrument is a capillary electrophoresis based automated analyzer which performs a complete hemoglobin profile for the qualitative analysis of hemoglobins (A. F. S. C, D, E and Bart's). The assay is performed on the hemolysate of whole blood samples previously collected on filter paper.

The test result must be interpreted in conjunction with other biological and clinical findings. In case of abnormal hemoglobin presence, it should be confirmed by additional tests as per local recommendations. The device is intended for professional use only.

For In Vitro Diagnostic Use only.

Predicate Device NamePredicate Device510(k) numberRegulation No.Vendor
CAPILLARYS NEONAT HbK091283February 22. 201021 CFR Part 864.7415 – AbnormalHemoglobin AssaySebia

4. SUBSTANTIAL EQUIVALENCE INFORMATION

{6}------------------------------------------------

5. COMPARISON WITH PREDICATE DEVICE

SIMILARITIES (Table A).

Kit Similarities
Table ACandidate DeviceCAPI 3 NEONAT HbCAPILLARYS 3 DBS instrumentPredicate DeviceCAPILLARYS NEONAT Hb(K091283)
Intended UseThe CAPI 3 NEONAT Hb kit is intended as a qualitative screen for the detection of normal hemoglobins (F and A) and abnormal hemoglobins (S, C, E, D and Bart's) in blood from human new-born collected on filter paper. This analysis is performed by capillary electrophoresis with the CAPILLARYS 3 DBS instrument.The CAPILLARYS 3 DBS instrument is a capillary electrophoresis based automated analyzer which performs a complete hemoglobin profile for the qualitative analysis of hemoglobins (A, F, S, C, D, E and Bart's). The assay is performed on the hemolysate of whole blood samples previously collected on filter paper.The test result must be interpreted in conjunction with other biological and clinical findings. In case of abnormal hemoglobin presence, it should be confirmed by additional tests as per local recommendations. The device is intended for professional use only.For In Vitro Diagnostic Use only.The CAPILLARYS NEONAT Hb kit is designed for the separation of the normal hemoglobins (F and A) in blood samples from human new-borns. and for the detection of the major hemoglobin variants (S, C, E, D and Bart's), by electrophoresis in alkaline buffer (pH 9.4) with the CAPILLARYS 2 system. The CAPILLARYS NEONAT Hb kit is designed for laboratory use.The CAPILLARYS 2 is an automated analyzer which performs a complete hemoglobin profile for the qualitative analysis of hemoglobins. The assay is performed on the hemolysate of whole blood samples previously collected on filter paper.For In Vitro Diagnostic Use only.
AnalyteHemoglobin detectionSame
MeasurementQualitativeSame
Sample MatrixBlood samples from human new-borns collected on filter paperSame
TechnologyCapillary electrophoresisSame
InstrumentWavelengthMeasurement415 nmSame
SoftwarePHORESISSame

{7}------------------------------------------------

REAGENT SIMILARITIES (Table B).

Reagent similarities
Table BCandidate DeviceCAPI 3 NEONAT HbCAPILLARYS 3 DBS instrumentPredicate DeviceCAPILLARYS NEONAT Hb(K091283)
BufferCAPILLARYS HEMOGLOBIN(E) bufferSame
HemolysingSolutionCAPILLARYS NNHb hemolysing solutionSame
ControlsHb AF ControlSame
Hb AFSC ControlSame

DIFFERENCES (Table C)

Kit Differences
Table CCandidate DeviceCAPI 3 NEONAT HbCAPILLARYS 3 DBS instrumentPredicate DeviceCAPILLARYS NEONAT Hb(K091283)
DilutionVessel96 well-microplatesDilution segments
Punch Size3.2 mm diameter3.8 mm diameter
InstrumentCAPILLARYS 3 DBSCAPILLARYS 2 SYSTEM

{8}------------------------------------------------

6. ANALYTICAL PERFORMANCE DATA

A) Precision

Between capillaries, between lots, between instruments, and between days precision.

Eight (8) different samples were analyzed using the CAPI 3 NEONAT Hb procedure used on the CAPILLARYS 3 DBS instrument. The samples included 6 new-born whole blood samples (with FA, FAS, FAC, FAD, FAE and FA Bart's patterns) and 2 controls.

  • For the between capillaries precision, each sample was analyzed once on the 12 capillaries of i 1 instrument with 1 reagent lot, through 1 run.
  • For the between lots precision, for 3 days, same samples were analyzed once per run over 2 i runs (with a minimum of 2 hours between both runs), on 1 instrument with 1 reagent lot per day (total of 3 lots).
  • For the between instruments precision, within the same day, same samples were analyzed once i per run over 2 runs (with a minimum of 2 hours between both runs), on 3 instruments with 1 reagent lot.
  • For the between days precision, every day, for 5 days, same samples were analyzed once per run i over 2 runs (with a minimum of 2 hours between both runs) on 1 instrument with 1 reagent lot.
IdentificationpatternBetween capillariesBetween lotsBetween instrumentsBetween days
ResultTotal analysesResultTotal analysesResultTotal analysesResultTotal analyses
SampleNo. 1FA100%concordance12100%concordance6100%concordance6100%concordance10
SampleNo. 2FAS100%concordance12100%concordance6100%concordance6100%concordance10
SampleNo. 3FAC100%concordance12100%concordance6100%concordance6100%concordance10
SampleNo. 4FAD100%concordance12100%concordance6100%concordance6100%concordance10
SampleNo. 5FAE100%concordance12100%concordance6100%concordance6100%concordance10
SampleNo. 6FA Bart's100%concordance12100%concordance6100%concordance6100%concordance10
SampleNo. 7AFSC100%concordance12100%concordance6100%concordance6100%concordance10
SampleNo. 8FA100%concordance12100%concordance72100%concordance72100%concordance120
Total concordance(95% Cl. Wilsonscore method)100%(96.2%;100.0%)96100%(96.7%;100.0%)114100%(96.7%;100.0%)114100%(98.0%;100.0%)190

All samples gave 100 % identification pattern concordance between all capillaries, lots, instruments, and days.

{9}------------------------------------------------

Precision using controls as per CLSI EP05-A3 general study design

The precision of the CAPI 3 NEONAT Hb procedure performed with the CAPILLARYS 3 DBS instrument was evaluated in a study with a study design based on the Clinical and Laboratory Standards Institute (CLSI - USA) EP5-A3 guideline "Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition".

Two (2) different control samples (with FA and AFSC patterns) were analyzed using the CAPI 3 NEONAT Hb procedure used on the CAPILLARYS 3 DBS instrument.

  • For the within-laboratory precision, each sample was analyzed once on the 12 capillaries of 1 instrument with 1 lot of reagents, through 2 runs per day during 20 days.
ControlNMeanWithin-RunBetween-RunBetween-DayTotal
(%)SD% CVSD% CVSD% CVSD% CV
AF48069.20.400.60.250.40.490.70.681.0
AFSC48026.00.150.60.070.30.020.10.160.6

% Hb F

% Hb A

ControlNMean(%)Within-RunBetween-RunBetween-DayTotal
SD% CVSD% CVSD% CVSD% CV
AF48030.20.401.30.260.90.511.70.702.3
AFSC48046.60.300.60.230.50.000.00.380.8

% Hb S

ControlNMean(%)Within-RunBetween-RunBetween-DayTotal
SD% CVSD% CVSD% CVSD% CV
AFSC48016.60.140.80.110.70.000.00.171.0

% Hb C

ControlNMean(%)Within-RunBetween-RunBetween-DayTotal
SD% CVSD% CVSD% CVSD% CV
AFSC48010.70.080.70.060.60.040.40.111.0
  • For the between-lots precision, during 5 days, same samples were analyzed 3 times per run over 2 runs (with a minimum of 2 hours between both runs), on 1 instrument with 3 lots of reagents.

% Hb F

ControlNMean (%)Within-RunBetween-RunBetween-DayBetween-LotsTotal
SD% CVSD% CVSD% CVSD% CVSD% CV
AF9069.30.220.30.320.50.000.00.170.20.420.6
AFSC9025.80.100.40.050.20.040.20.060.20.140.5

{10}------------------------------------------------

% Hb A

ControlNMean (%)Within-RunBetween-RunBetween-DayBetween-LotsTotal
SD% CVSD% CVSD% CVSD% CVSD% CV
AF9030.00.210.70.341.10.000.00.160.50.431.4
AFSC9047.10.410.90.000.00.120.30.170.40.461.0

% Hb S

ControlNMean (%)Within-RunBetween-RunBetween-DayBetween-LotsTotal
SD% CVSD% CVSD% CVSD% CVSD% CV
AFSC9016.50.241.50.000.00.040.30.100.60.271.6

% Hb C

ControlNMean (%)Within-RunBetween-RunBetween-DayBetween-LotsTotal
SD% CVSD% CVSD% CVSD% CVSD% CV
AFSC9010.60.141.30.000.00.040.40.131.20.191.8
  • For the between-instruments precision, during 5 days, same samples were analyzed 3 times per run over 2 runs (with a minimum of 2 hours between both runs), on 3 instruments with 1 lot of reagents.

% Hb F

ControlNMean(%)Within-RunBetween-RunBetween-DayBetween-InstrumentsTotal
SD% CVSD% CVSD% CVSD% CVSD% CV
AF9069.40.330.50.430.60.000.00.450.60.701.0
AFSC9025.90.110.40.060.20.000.00.090.30.150.6

% Hb A

ControlNMean(%)Within-RunBetween-RunBetween-DayBetween-InstrumentsTotal
SD% CVSD% CVSD% CVSD% CVSD% CV
AF9029.90.331.10.451.50.000.00.441.50.712.4
AFSC9046.60.280.60.080.20.050.10.471.00.561.2

% Hb S

ControlNMean (%)Within-RunBetween-RunBetween-DayBetween-InstrumentsTotal
SD% CVSD% CVSD% CVSD% CVSD% CV
AFSC9016.70.150.90.000.00.050.30.231.40.281.7

{11}------------------------------------------------

% Hb C

ControlNMean (%)Within-RunBetween-RunBetween-DayBetween-InstrumentsTotal
SD% CVSD% CVSD% CVSD% CVSD% CV
AFSC9010.80.101.00.050.40.020.20.161.50.201.8

All samples gave 100 % identification pattern concordance between all analyses.

B) Analytical Specificity/Interference

Endogenous interference

The common interfering factors with the CAPI 3 NEONAT Hb procedure used on the CAPILLARYS 3 DBS instrument (conjugated bilirubin. unconjugated bilirubin and triglycerides) were evaluated in studies based on the Clinical Laboratory Standards Institute (CLSI - USA) EP7-A2 quideline "Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition" and the Clinical Laboratory Standards Institute (CLSI - USA) EP07. 3rd Edition guideline "Interference Testing in Clinical Chemistry".

The results are summarized below.

No interference with the CAPI 3 NEONAT Hb procedure used on the CAPILLARYS 3 DBS instrument was detected due to the new-born whole blood sample's high concentration of the following interfering factors tested at levels up to the concentrations listed below:

Endogenous interfering factorConcentration
Conjugated bilirubin40 mg/dL
Unconjugated bilirubin40 mg/dL
Triglycerides1500 mg/dL

C) Qualitative Method Comparison with Predicate Device

Qualitative method comparison - Internal study

The concordance study of the CAPI 3 NEONAT Hb procedure performed with the CAPILLARYS 3 DBS instrument was evaluated in a study based on the Clinical and Laboratory Standards Institute (CLSI - USA) EP12-A2 guideline "User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition". sections 9. Comparison of Methods and 10. Data Analysis.

NOTE: The analyzed samples were provided by 5 laboratories in France. Thailand and Panama. All the samples followed the same guidelines in regard to sample integrity and were prepared according to the instructions for use of each technique.

One hundred and thirty-eight (138) different new-born whole blood samples and 67 pathological samples with hemoglobin variants S. C. D. E and Bart's) were analyzed with the CAPI 3 NEONAT Hb procedure performed with the CAPILLARYS 3 DBS instrument and with a commercially available capillary electrophoresis procedure for hemoglobin analysis (reference).

{12}------------------------------------------------

The positive percent agreement (PPA), the neqative percent agreement (NPA), and their confidence intervals (95 % Cl by Wilson score method) of the CAPI 3 NEONAT Hb procedure compared to the reference procedure have been calculated. The results are listed below:

Positive Percent Agreement (%): 100% (94.6 %;100.0 %). Negative Percent Agreement (%): 100.0% (94.9 %;100.0 %).

Hb fractionsNumber of samples
FA71
FAS25
FS7
FCS1
FAC7
FAD2
FAE8
FA + Hb Bart's10
FAS + Hb Bart's6
AFSC1
Total138

Qualitative method comparison - External study No. 1

The concordance study of the CAPI 3 NEONAT Hb procedure performed with the CAPILLARYS 3 DBS instrument was evaluated in a study based on the Clinical and Laboratory Standards Institute (CLSI - USA) EP12-A2 quideline "User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition". sections 9. Comparison of Methods and 10. Data Analysis.

NOTE: The samples were analyzed in a laboratory based in the United States of America. All the samples followed the same guidelines in regard to sample integrity and were prepared according to the instructions for use of each technique.

Four hundred and eleven (411) different new-born whole blood samples (210 normal samples and 201 pathological samples with hemoglobin variants S. C. D. E and Bart's) were analyzed with the CAPI 3 NEONAT Hb procedure performed with the CAPILLARYS 3 DBS instrument and with a commercially available capillary electrophoresis procedure for hemoglobin analysis (reference).

The positive percent agreement (PPA). the negative percent (NPA). and their confidence intervals (95 % CI by Wilson score method) of the CAPI 3 NEONAT Hb procedure compared to the reference procedure have been calculated. The results are listed below:

Positive Percent Agreement (%): 100.0 % (98.1 %;100.0 %). Negative Percent Agreement (%): 100.0 % (98.2 %;100.0 %).

{13}------------------------------------------------

Hb fractionsNumber of samples
FA210
FAS63
FAC40
FA + Hb Bart's35
FAD15
FAS + Hb Bart's12
FAE9
FS7
FSA7
FAX6
FCS2
AFS1
FSE1
FCX1
FE1
FAC + Hb Bart's1
Total411

Qualitative method comparison - External study No. 2

The concordance study of the CAPI 3 NEONAT Hb procedure performed with the CAPILLARYS 3 DBS instrument was evaluated in a study based on the Clinical and Laboratory Standards Institute (CLSI - USA) EP12-A2 guideline "User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition". sections 9. Comparison of Methods and 10. Data Analysis.

NOTE: The samples were analyzed in a laboratory based in Spain. All the samples followed the same guidelines in regard to sample integrity and were prepared according to the instructions for use of each technique.

Two hundred and forty (240) different new-born whole blood samples (120 normal samples and 120 pathological samples with hemoglobin variants S. C. D. E and Bart's) were analyzed with the CAPI 3 NEONAT Hb procedure performed with the CAPILLARYS 3 DBS instrument and with a commercially available capillary electrophoresis procedure for hemoglobin analysis (reference).

The positive percent agreement (PPA). the negative percent agreement (NPA), and their confidence intervals (95 % Cl by Wilson score method) of the CAPI 3 NEONAT Hb procedure compared to the reference procedure have been calculated. The results are listed below:

Positive Percent Agreement (%): 100.0 % (96.9 %;100.0 %). Negative Percent Agreement (%): 100.0 % (96.9 %;100.0 %).

{14}------------------------------------------------

Hb fractionsNumber of samples
FA120
FAS53
FS6
FSC1
FAC29
FAD5
FAE5
FAX5
FA + Hb Bart's13
FAS + Hb Bart's2
FAC + Hb Bart's1
Total240

7. CONCLUSION

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 864.7415 Abnormal hemoglobin assay.

(a)
Identification. An abnormal hemoglobin assay is a device consisting of the reagents, apparatus, instrumentation, and controls necessary to isolate and identify abnormal genetically determined hemoglobin types.(b)
Classification. Class II (special controls). A control intended for use with an abnormal hemoglobin assay is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.