K Number
K231895

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-12-04

(159 days)

Product Code
Regulation Number
868.5330
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Maxtec MaxBlend 2+p is intended to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life-supporting device or life sustaining device.

Device Description

The Maxtec MaxBlend 2+p is an oxygen delivery device which incorporates an air/oxygen blender, battery powered oxygen and pressure monitor, and an adjustable flowmeter, all in a single assembly. The integral air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms. It also allows the user to monitor pressure simultaneously using adjustable high and low alarm limits.

AI/ML Overview

The document describes the Maxtec MaxBlend 2+p, a medical device combining an air/oxygen blender, oxygen monitor, pressure monitor, and flowmeter. The 510(k) submission seeks substantial equivalence to existing predicate devices (Maxtec MaxBlend 2 - K161718 and MaxO2ME+p - K221734), particularly highlighting the addition of a pressure monitoring feature.

Here's an analysis of the acceptance criteria and the study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for several performance aspects are implicitly derived from the comparative tables (Table 1 and 2) where the subject device's specifications are listed and compared to the predicate devices. The "Reported Device Performance" for the subject device is simply its stated characteristics, which are claimed to meet or be identical to the predicate device specifications, thereby meeting the acceptance criteria based on substantial equivalence.

Feature / Performance AspectAcceptance Criteria (from predicate devices)Reported Device Performance (Maxtec MaxBlend 2+p)
Air/Oxygen Mixer Features
Gas Supply TypeAir / OxygenAir / Oxygen
Pressure30 to 75 psi30 to 75 psi
Mixed gas stability± 1% oxygen± 1% oxygen
Flow range of BlendersLow flow model – 0-30 Lpm; High flow model – 0-100 LpmLow flow model – 0-30 Lpm; High flow model – 0-70 Lpm (Primary Predicate) / 0-100 Lpm (Secondary Predicate)
Pressure supply differential alarmAir / oxygen pressure < 20 psi an alarm soundsAir / oxygen pressure < 20 psi an alarm sounds
Pressure Drop< 6 psi @ 50 psi< 6 psi @ 50 psi
Bleed flow3-13 Lpm at 50 psi depending upon model3-13 Lpm at 50 psi depending upon model
Oxygen Monitor Features
Oxygen measurement range0-100%0-100%
Total Accuracy±3% Actual oxygen level over full operating temperature range±3% Actual oxygen level over full operating temperature range
Response Time90% of final value in approx. 15 seconds at 23°C90% of final value in approx. 15 seconds at 23°C
Warm-up TimeNone requiredNone required
High alarm range16 – 100%16 – 100%
Operating Temperature15°C – 40°C (59°F – 104°F)15°C – 40°C (59°F – 104°F)
Storage Temperature-15°C – 50°C (5°F – 122°F)-15°C – 50°C (5°F – 122°F)
Humidity0-95% (non-condensing)0-95% (non-condensing)
Expected use-life of sensor1,500,000 O2% hours (approx. 2 years)1,500,000 O2% hours (approx. 2 years)
Flowmeter Features
Accuracy± 10% of indicated value when inlet pressure is 50 psi± 10% of indicated value when inlet pressure is 50 psi
Flow meter ranges0 – 3 Lpm, 0 – 15 Lpm, 0 – 30 Lpm, 0 – 70 Lpm (some predicates mention 0-100Lpm as well)0 – 3 Lpm, 0 – 15 Lpm, 0 – 30 Lpm, 0 – 70 Lpm
Pressure Monitor Features(Based on Secondary Predicate MaxO2 ME+p K221734)(Maxtec MaxBlend 2+p)
TechnologyMicroprocessor controlled deviceMicroprocessor controlled device
Pressure sensor typeSolid-state pressure transducerSolid-state pressure transducer
Pressure Measurement Range-15 to +60 cmH2O-15 to +60 cmH2O
Pressure Resolution1 cmH2O1 cmH2O
Display resolution0.5 cm H2O0.5 cm H2O
Total Accuracy± 1 cmH2O± 1 cmH2O
Low alarm range1 - 30 cmH2O1 - 30 cmH2O
High alarm range1 - 60 cmH2O1 - 60 cmH2O
Alarm delay3 seconds (pressure only)3 seconds (pressure only)
Zero calibrationYesYes
Operating Temperature15°C - 40°C, 0-95% RH15°C - 40°C, 0-95% RH
Storage Temperature-15°C - 50°C @ 95% RH-15°C - 50°C @ 95% RH
Atmospheric Pressure800 – 1013 mBars800 – 1013 mBars
Power requirements4x - AA alkaline batteries4x - AA alkaline batteries
Battery Life5000 hours5000 hours

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states:

  • Non-clinical Testing: "We performed a number of tests to demonstrate that the proposed device performed as intended." It then lists several international standards that the device met. It does not specify a sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective). The testing appears to be primarily related to engineering and performance validation against standards rather than clinical data.
  • Clinical Testing: "No human clinical testing was performed."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Since "No human clinical testing was performed," and the primary testing involved adherence to international standards and non-clinical performance, there is no mention of experts establishing ground truth for a clinical test set. The "ground truth" for the non-clinical testing would be the defined requirements and specifications within the cited standards.

4. Adjudication Method for the Test Set

Not applicable, as no human clinical test set is mentioned, and the testing focuses on compliance with technical standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a medical gas mixer and monitor, not an AI-assisted diagnostic or interpretation tool that would involve human readers or image analysis.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is a hardware product with integrated monitoring capabilities. Its "standalone" performance is measured through its adherence to the technical specifications and international standards listed under "Non-clinical Testing." The document states: "In all cases the subject device met the standard requirements." This implicitly confirms standalone performance against established benchmarks.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the performance specifications (e.g., oxygen accuracy, pressure accuracy, flow rates), the ground truth is based on metrological standards and calibrated equipment used in accordance with the listed international standards (e.g., ISO 80601-2-55, ISO 11195). For biocompatibility, the ground truth is established by adherence to ISO 10993 and ISO 18562 standards.

8. The Sample Size for the Training Set

Not applicable. This is a hardware medical device with integrated sensors and controls, not a machine learning or AI algorithm that requires a "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG' are written in blue, followed by 'ADMINISTRATION' in a smaller font size.

December 4, 2023

Maxtec, LLC % Paul Dryden Consultant ProMedic Consulting LLC 131 Bav Point Dr NE Saint Petersburg, Florida 33704

Re: K231895

Trade/Device Name: Maxtec MaxBlend2+p Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: Class II Product Code: BZR, CCL, CAP Dated: October 31, 2023 Received: November 1, 2023

Dear Paul Dryden:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}------------------------------------------------

See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bradley Q. Quinn -S

Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231895

Device Name

Maxtec MaxBlend 2+p

Indications for Use (Describe)

The Maxtec MaxBlend 2+p is intended to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healtheare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required.

This is not intended as a life-supporting device or life sustaining device.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

1-Dec-23

Date Prepared:

Maxtec, Inc. 2305 South 1070 West Salt Lake City, UT 84119 Tel - 1-732-244-0010 x1151

Official Contact:Rebecca Hudson - QARA
Submission Correspondent:Paul DrydenProMedic, LLC131 Bay Point Dr NESt. Petersburg, FL 33704
Proprietary or Trade Name:Maxtec MaxBlend2+p
Classification CFR:21 CFR 868.5330
Classification Code:BZR
Classification Name:Mixer, Breathing Gases, Anesthesia Inhalation
Classification CFR:21 CFR 868.1720
Classification Code:CCL
Classification Name:Analyzer, Gas, Oxygen, Gaseous-Phase
Classification CFR:21 CFR 868.2600
Classification Code:CAP
Classification Name:Monitor, Airway Pressure
Predicate Device:Maxtec MaxBlend 2 - K161718
Secondary Predicate:MaxO2ME+p -K221734

Device Description:

The Maxtec MaxBlend 2+p is an oxygen delivery device which incorporates an air/oxygen blender, battery powered oxygen and pressure monitor, and an adjustable flowmeter, all in a single assembly. The integral air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms. It also allows the user to monitor pressure simultaneously using adjustable high and low alarm limits.

Indications for Use:

The Maxtec MaxBlend 2+p is designed to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e.. hospital, sub-acute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required.

This is not intended as a life-supporting device or life sustaining device.

Patient Population:

Infant, pediatric, and adult patients for MaxBlend 2+p.

{5}------------------------------------------------

Environments of use:

Hospital, sub-acute, and nursing-care facilities

We present the proposed device vs. the predicates in the Tables below.

{6}------------------------------------------------

510(k) Summary Page 3 of 11 Table 1 Comparison of MaxBlend 2+p to Primary Predicate

AttributesSubject DevicePrimary Predicate
Maxtec MaxBlend 2+pMaxBlend 2 K161718
Product ClassificationPrimary - BZR Secondary - CCLNew Secondary - CAPPrimary - BZRSecondary - CCL
Indications for UseThe Maxtec MaxBlend 2+p is designed to provide acontinuous air/oxygen gas mixture and to continuouslymonitor the concentration of oxygen and pressure beingdelivered to infant, pediatric, and adult patients. It is arestricted medical device intended for use by qualified, trainedpersonnel, under the direction of a physician, in professionalhealthcare settings, i.e., hospital, sub-acute, and nursing-carefacilities where the delivery and monitoring of air/oxygenmixtures is required.This is not intended as a life-supporting device or lifesustaining device.The MaxBlend 2 and MaxBlend Lite are designed to providea continuous air/oxygen gas mixture and to continuouslymonitor the concentration of oxygen being delivered to infant,pediatric, and adult patients. It is a restricted medical deviceintended for use by qualified, trained personnel, under thedirection of a physician, in professional healthcare settings,i.e., hospital, sub-acute, and nursing-care facilities where thedelivery and monitoring of air/oxygen mixtures is required.This is not intended as a life supporting device.
New indicationPressure monitoringN/A See Table 2 for this feature
Environments of UseProfessional healthcare settings, i.e., hospital, sub-acute, andnursing-care facilities where the delivery and monitoring ofair/oxygen mixtures is required.Professional healthcare settings, i.e., hospital, sub-acute, andnursing-care facilities where the delivery and monitoring ofair/oxygen mixtures is required.
Not for use in MRI environmentsNot for use in MRI environments
Patient PopulationInfant, pediatric, and adult patientsInfant, pediatric, and adult patients
ComponentsOxygen monitoringAir/Oxygen BlenderFlowmeterBleedAdding pressure featureOxygen monitoringAir/Oxygen BlenderNo FlowmeterBleed
Air/Oxygen Mixer Features
BlenderBlender isCareFusion / Bird - K883038 orBioMed Devices - K925982Blender isCareFusion / Bird - K883038 orBioMed Devices - K925982
Gas Supply TypeAir / OxygenAir / Oxygen
Pressure30 to 75 psi30 to 75 psi
AttributesSubject DeviceMaxtec MaxBlend 2+pPrimary PredicateMaxBlend 2 K161718
Mixed gas stability+ 1% oxygen+ 1% oxygen
Flow range of BlendersLow flow model – 0-30 LpmHigh flow model – 0-70 LpmLow flow model – 0-30 LpmHigh flow model – 0-100 Lpm
Pressure supplydifferential alarmAir / oxygen pressure must be < 20 psi an alarm soundsAir / oxygen pressure must be < 20 psi an alarm sounds
Pressure Drop< 6 psi @ 50 psi< 6 psi @ 50 psi
Bleed flow3-13 Lpm at 50 psi depending upon model3-13 Lpm at 50 psi depending upon model
Oxygen Monitor Features
Oxygen measurementrange0-100%0-100%
Total Accuracy±3% Actual oxygen level over full operating temperaturerange±3% Actual oxygen level over full operating temperaturerange
Response Time90% of final value in approx. 15 seconds at 23°C90% of final value in approx. 15 seconds at 23°C
Warm-up TimeNone requiredNone required
High alarm range16 – 100%16 – 100%
Operating Temperature15°C – 40°C (59°F – 104°F)15°C – 40°C (59°F – 104°F)
Storage Temperature-15°C – 50°C (5°F – 122°F)-15°C – 50°C (5°F – 122°F)
Humidity0-95% (non-condensing)0-95% (non-condensing)
Expected use-life of sensor1,500,000 O2% hours (approx. 2 years)1,500,000 O2% hours (approx. 2 years)
Flowmeter Features
Accuracy+ 10% of indicated value when inlet pressure is 50 psi+ 10% of indicated value when inlet pressure is 50 psi
Flow meter ranges0 – 3 Lpm0 – 15 Lpm0 – 30 Lpm0 – 70 Lpm0 – 3 Lpm0 – 30 Lpm0 – 70 Lpm0 – 100 Lpm
AttributesSubject DevicePrimary Predicate
Maxtec MaxBlend 2+pMaxBlend 2 K161718
Pressure Monitor Features (adding this feature)
TechnologyMicroprocessor controlled deviceSee Table 2
Pressure sensor typeSolid-state pressure transducerSee Table 2
Displayed informationLow pressure alarm settingSee Table 2
High pressure alarm setting
Alarm silence indicator
Power supply indicator
Average (mean) pressure
Audible and visual alarm
Pressure Measurement-15 to +60 cmH2OSee Table 2
Range
Pressure Resolution1 cmH2OSee Table 2
Display resolution0.5 cm H2OSee Table 2
Total Accuracy+ 1 cmH2OSee Table 2
Low alarm range1 - 30 cmH2OSee Table 2
High alarm range1 - 60 cmH2OSee Table 2
Alarm delay3 seconds (pressure only)See Table 2
Zero calibrationYesSee Table 2
System interfaceDisposable tubing with filter and with or without NafionSee Table 2
connects to tee or face mask
Operating Temperature15°C - 40°C, 0-95% RHSee Table 2
Storage Temperature-15°C - 50°C @ 95% RHSee Table 2
Atmospheric Pressure800 – 1013 mBarsSee Table 2
Power requirements4x - AA alkaline batteriesSee Table 2
Battery Life5000 hoursSee Table 2
General Information
StandardsANSI/AAMI/ES 60601-1ANSI/AAMI/ES 60601-1
IEC 60601-1-2IEC 60601-1-2
IEC 60601-1-8IEC 60601-1-8
ISO 80601-2-55ISO 80601-2-55
ISO 11195
Biocompatibility /MaterialsIdentical materials to K161718 and K221734Externally communicating
in Gas PathwayTissue, Permanent duration
Identical to K161718 and K221734Materials of the Oxygen Sensor (K153659) and Blenders
(K883038 and K925982) are identical
AttributesSubject DeviceSecondary Predicate
Maxtec MaxBlend 2+pMaxO2 ME+p K221734
Product ClassificationPrimary – BZRPrimary - CCL
Secondary - CCLSecondary - CAP
New Secondary - CAP
Indications for UseThe Maxtec MaxBlend 2+p is designed to provide aThe MaxO2 ME+p is an oxygen monitor with integrated
continuous air/oxygen gas mixture and to continuouslypressure monitoring intended for continuous monitoring of
monitor the concentration of oxygen and pressure beingthe concentration of oxygen and pressure being delivered to
delivered to infant, pediatric, and adult patients. It is apatients ranging from newborns to adults. It can be used in the
restricted medical device intended for use by qualified, trainedhospital and sub-acute settings.
personnel, under the direction of a physician, in professional
healthcare settings, i.e., hospital, sub-acute, and nursing-care
facilities where the delivery and monitoring of air/oxygen
mixtures is required.
This is not intended as a life-supporting device or life
sustaining device.
New indicationAdding Pressure monitoringCleared for pressure monitoring
Environments of UseProfessional healthcare settings, i.e., hospital, sub-acute, andHospital and sub-acute settings
nursing-care facilities where the delivery and monitoring of
air/oxygen mixtures is required.
Not for use in MRI environmentsNot for use in MRI environments
Patient PopulationInfant, pediatric, and adult patientsNewborns to adults
ComponentsOxygen monitoringOxygen monitoring
Air/Oxygen Blender
Flowmeter
Bleed
Pressure monitoringPressure Monitoring
Pressure Monitoring Features (adding this feature)
TechnologyMicroprocessor controlled deviceMicroprocessor controlled device
Pressure sensor typeSolid-state pressure transducerSolid-state pressure transducer
Displayed informationLow pressure alarm settingLow pressure alarm setting
High pressure alarm settingHigh pressure alarm setting
Alarm silence indicatorAlarm silence indicator
Power supply indicatorPower supply indicator
AttributesSubject DeviceMaxtec MaxBlend 2+pSecondary PredicateMaxO2 ME+p K221734
Average (mean) pressureAudible and visual alarmAverage (mean) pressureAudible and visual alarm
Pressure MeasurementRange-15 to +60 cmH2O-15 to +60 cmH2O
Pressure Resolution1 cmH2O1 cmH2O
Display resolution0.5 cm H2O0.5 cm H2O
Total Accuracy+ 1 cmH2O+ 1 cmH2O
Low alarm range1 - 30 cmH2O1 - 30 cmH2O
High alarm range1 - 60 cmH2O1 - 60 cmH2O
Alarm delay3 seconds (pressure only)3 seconds (pressure only)
Zero calibrationYesYes
System interfaceDisposable tubing with filter and with or without Nafionconnects to tee or face maskDisposable tubing with filter and with or without Nafionconnects to tee or face mask
Operating Temperature15°C - 40°C, 0-95% RH15°C - 40°C, 0-95% RH
Storage Temperature-15°C - 50°C @ 95% RH-15°C - 50°C @ 95% RH
Atmospheric Pressure800 – 1013 mBars800 – 1013 mBars
Power requirements4x - AA alkaline batteries4x - AA alkaline batteries
Battery Life5000 hours5000 hours
Oxygen Monitor Features
Oxygen measurementrange0-100%0-100%
Oxygen Resolution0.1%0.1%
Accuracy and Linearity±1% of full scale at constant temperature, RH and pressurewhen calibrated at fill scale±1% of full scale at constant temperature, RH and pressurewhen calibrated at fill scale
Total Accuracy±3% Actual oxygen level over full operating temperaturerange±3% Actual oxygen level over full operating temperaturerange
Response Time90% of final value in approx. 15 seconds at 23°C90% of final value in approx. 15 seconds at 23°C
Warm-up TimeNone requiredNone required
High alarm range16 - 100%16 - 100%
Operating Temperature15°C - 40°C (59°F – 104°F)15°C - 40°C (59°F – 104°F)
Storage Temperature-15°C - 50°C (5°F - 122°F)-15°C - 50°C (5°F - 122°F)
Atmospheric Pressure800 – 1013 mBars800 – 1013 mBars
Humidity0-95% (non-condensing)0-95% (non-condensing)
Power requirements4x - AA Alkaline batteries4x - AA Alkaline batteries
AttributesSubject DeviceSecondary Predicate
Maxtec MaxBlend 2+pMaxO2 ME+p K221734
Battery LifeApprox. 5000 hours, typical useApprox. 5000 hours, typical use
Low Battery IndicationsBattery indicator on LCD displayBattery indicator on LCD display
Sensor TypeMaxtec MAX-550E galvanic fuel cellMaxtec MAX-550E galvanic fuel cell
Expected use-life of sensor1,500,000 O2% hours (approx.. 2 years)1,500,000 O2% hours (approx.. 2 years)
Alarm SystemsHigh/Low alarms, flashing yellow LEDsHigh/Low alarms, flashing yellow LEDs
Nominal 975 Hz audio buzzer (IEC 60601-1-8)Nominal 975 Hz audio buzzer (IEC 60601-1-8)
Low Oxygen Alarm Range15% - 99% (>1% lower than high alarm)15% - 99% (>1% lower than high alarm)
High Oxygen Alarm Range16% - 100% (>1% higher than low alarm)16% - 100% (>1% higher than low alarm)
AccuracyExact to display alarm valueExact to display alarm value
AccessoriesDiverterDiverter
Tee adapter (15 mmm x 22 mm fittings)Tee adapter (15 mmm x 22 mm fittings)
Pressure monitoring linePressure monitoring line
Mounting bracketsMounting brackets
DC power adapterDC power adapter
Air/Oxygen Mixer Features
BlenderBlender isNot applicable see Table 1
CareFusion / Bird - K883038 or
BioMed Devices - K925982
Gas Supply TypeAir / OxygenNot applicable see Table 1
Pressure30 to 75 psiNot applicable see Table 1
% Oxygen Control21 - 100%Not applicable see Table 1
Accuracy + 3%
Mixed gas stability+ 1% oxygenNot applicable see Table 1
Flow range of BlendersLow flow model - 0-30 LpmNot applicable see Table 1
High flow model - 0-100 Lpm
Pressure supplyAir / oxygen pressure must be < 20 psi an alarm soundsNot applicable see Table 1
differential alarm
Pressure Drop< 6 psi @ 50 psiNot applicable see Table 1
Bleed flow3-13 Lpm at 50 psi depending upon modelNot applicable see Table 1
Flowmeter Features
Accuracy+ 10% of indicated value when inlet pressure is 50 psiNot applicable see Table 1
Flow meter ranges0 – 3 LpmNot applicable see Table 1
0 - 15 Lpm
0 - 30 Lpm
0 - 70 Lpm
AttributesSubject DeviceMaxtec MaxBlend 2+pSecondary PredicateMaxO2 ME+p K221734
General Information
StandardsANSI/AAMI/ES 60601-1IEC 60601-1-2 with RFIDIEC 60601-1-8ISO 80601-2-55ISO 11195ANSI/AAMI/ES 60601-1IEC 60601-1-2AIM RFIDIEC 60601-1-8ISO 80601-2-55
BiocompatibilityMaterials in Gas PathwayIdentical to K161718 and K221734Materials of the Oxygen Sensor (K153659) and PressureMonitoring (K221734)Materials of the Oxygen Sensor (K153659) and PressureMonitoring (K221734)Materials tested to applicable ISO 10993-1 and ISO 18562-1requirements

{7}------------------------------------------------

510(k) Summary Page 4 of 11

{8}------------------------------------------------

510(k) Summary Page 5 of 11

{9}------------------------------------------------

510(k) Summary Page 6 of 11

Table 2 Comparison of the proposed device vs. the Secondary Predicate MaxO2 ME+p

{10}------------------------------------------------

510(k) Summary Page 7 of 11

{11}------------------------------------------------

510(k) Summary Page 8 of 11

{12}------------------------------------------------

510(k) Summary Page 9 of 11

{13}------------------------------------------------

510(k) Summary Page 10 of 11

Substantial Equivalence Discussion and Conclusion

As discussed the only new feature is the pressure monitoring feature. All other features and performance are identical to the primary and secondary predicates, we will only discuss the pressure monitoring feature being added to the subject device.

Indications for Use

The subject device, Maxtec MaxBlend 2+p.is adding the cleared feature of measuring and monitoring pressure including the disposable accessory lines. This is similar to the secondary predicate K221734 – Maxtec – MaxO2 ME+p.

However, for product classification purposes we have compared to subject device to the primary predicate which has the same primary indications for use with the added pressure monitoring feature. The primary predicate, K161718 is the MaxBlend 2 which is an air/gas blender with oxygen monitoring.

Environment of Use

The subject device, MaxBlend 2+p has similar environments of use when compared to the primary predicate K161718 Maxtec - MaxBlend 2. There is no difference in the use environments and adding the pressure monitoring feature which has been cleared for the same use environments would raise different concerns of safety or effectiveness.

Population

The subject device, Maxtec MaxBlend 2+p has similar patient population as the primary predicate, K161718 Maxtec -MaxBlend 2. While the secondary predicate has been cleared with the pressure monitoring for newborn to adults, the subject device and its primary predicate have been cleared for infants, pediatrics and adults. The addition of the pressure monitoring feature remains within the same patient population as the secondary predicate.

Performance Specifications

The subject device, Maxtec MaxBlend 2+p, has similar performance specifications for the blender and oxygen monitoring features as the primary predicate, K161718 - Maxtec MaxBlend 2 and the pressure monitoring performance specifications as the secondary predicate, K221734 – Maxtec MaxO2 ME+p.

Non-clinical Testing

We performed a number of tests to demonstrate that the proposed device performed as intended. Testing includes:

  • AAMI ANSI ES 60601-1: 2005 +A1: 2012 Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance
  • IEC 60601-1-2 Edition 4.1 2020 Collateral standard: Electromagnetic Disturbances - Requirements and Tests
  • IEC 60601-1-8: 2012 Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
  • ISO 80601-2-55: 2018 Particular requirements for the basic safety and essential performance of respiratory gas monitors
  • ISO 11195:2018 Gas mixers for medical use

In all cases the subject device met the standard requirements. Biocompatibility

The only new patient contacting materials are the disposable pressure tubing which are identical to the secondary predicate, K221734 - Maxtec MaxO2 ME+p. There are no changes in materials or patient contact type. These materials were evaluated in K221734 and included:

  • ISO 10993-5:2009 Cytotoxicity ●
  • ISO 10993-10:2010 Sensitization and Irritation ●

{14}------------------------------------------------

510(k) Summary Page 11 of 11

  • ISO 10993-11:2017 - Material Mediated Pyrogenicity
  • ISO 10993-18:2020 Chemical Characterization with a Toxicological Risk Assessment ●
  • ISO 18562-2:2017 - Particulate Material
  • ISO 18562-3:2017 VOC with Toxicological Risk Assessment ●

The patient contacting materials were found to be safe for the intended population and indications for use.

Animal

No animal testing was performed.

Clinical

No human clinical testing was performed.

Conclusion

We have performed a comparison of specifications in the above tables and found the proposed addition of the pressure monitoring feature to the subject device to be equivalent to the predicates.

§ 868.5330 Breathing gas mixer.

(a)
Identification. A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.(b)
Classification. Class II (performance standards).