K Number
K231834

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-04-04

(287 days)

Product Code
Regulation Number
882.5320
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VSP PEEK Cranial Implant is indicated for use to fill a bony void or defect area in the cranial skeleton of patients 21 years of age and older.

Device Description

The VSP PEEK Cranial Implants are non-load bearing, single use devices showing the following characteristics:
• The implants are designed individually for each patient to correct defects in cranial bone.

  • The implants are 3D printed using material extrusion technology.
  • The implants are manufactured using Polyetheretherketone (PEEK) implantable grade filament (i.e., Evonik VESTAKEEP® i4-3DF).
  • The VSP PEEK Cranial Implants are designed using CT imaging data and surgeon input.
  • · The implant size is ranging from 35 mm x 35 mm to 150 mm x 150 mm, and its thickness is constant, ranging from 3 mm to 6 mm with a nominal thickness of 3 mm.
  • If minor intra-operative adjustments are required, the implants can be modified by the surgeon with standard surgical burrs.
  • The implants are fixed to the native bone using commercially available cranial fixation screws and/or systems
  • The implants are provided non-sterile for sterilization prior to implantation.
AI/ML Overview

The provided text describes the 510(k) premarket notification for the "VSP PEEK Cranial Implant," which is a Class II device intended to fill bony voids or defects in the cranial skeleton. The information pertains to a medical device's regulatory clearance rather than an AI/ML-driven device, so many of the requested criteria related to AI acceptance, ground truth, and human reader studies are not applicable.

However, I can extract the relevant information regarding the device's acceptance criteria and the studies performed to demonstrate its performance.

Here's a breakdown based on the provided document:

1. Acceptance Criteria and Reported Device Performance (Table)

There are two main categories of acceptance criteria and performance reporting mentioned: Biocompatibility and Performance Bench Testing.

Criteria CategoryAcceptance CriteriaReported Device Performance/Results
Biocompatibility TestingCompliance with ISO-10993 ("Biological Evaluation of Medical Devices Part 1: Evaluation and Testing") and specific standards for each test.All tests found to be within acceptance criteria described in the standards.
Chemical Characterization & Toxicological Evaluation (ISO 10993-18 & 17)Acceptable Margin of Safety for all reported extractable substances.Acceptable Margin of Safety for all reported extractable substances.
Cytotoxicity (ISO 10993-5)Cell culture treated with test sample compared to control showing acceptable levels of dehydrogenase activity.Cell culture treated with test sample and compared dehydrogenase activity to control (Implied: met criteria for acceptance).
Sensitization (ISO 10993-10)Non-sensitizing.Non-sensitizing.
Irritation (ISO 10993-23)Non-irritating.Non-irritating.
Pyrogenicity (ISO 10993-11)Non-pyrogenic.Non-pyrogenic.
Acute Systemic Toxicity (ISO 10993-11)Non-toxic.Non-toxic.
Implantation effects (ISO 10993-6)No unexpected results.No unexpected results (28 & 90 day follow-up in rat calvarial bone).
Genotoxicity (ISO/TR 10993-33)Non-mutagenic for Reverse Mutation Assay and In vitro Mammalian Cell Gene Mutation Assay.Non-Mutagenic for both assays.
Performance Bench TestingAcceptable mechanical performance following sterilization or when used with common osteosynthesis fixation systems.All samples passed the acceptance criteria concerning static compression and dynamic impact testing.
Mechanical Performance TestingAll samples pass acceptance criteria (specific criteria not detailed but implied from results statement).All samples passed the acceptance criteria.
Anatomical Fit TestingAcceptable size accuracy and usability to surgeons.The size accuracy and usability of the VSP PEEK Cranial Implant was acceptable to surgeons.
Verification and Validation TestingCompliance with technical and biomechanical specifications.All samples compliant with specifications.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify exact sample sizes (e.g., number of implants or test specimens) for each test. It uses phrases like "All samples" and "Several implantations."
  • Data Provenance: The data is generated through laboratory bench testing and pre-clinical animal studies (rat model for implantation effects). The data origin is not specified by country, but it relates to the manufacturing and testing done by "3D Systems, Inc." located in Littleton, CO, USA. The studies are prospective in the sense that they were conducted for the purpose of this 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Experts

  • Experts: For the Anatomical Fit Testing, "surgeons" were involved in performing simulated surgical procedures and reviewing the results.
  • Qualifications: The general qualification is "surgeons." No specific number or detailed qualifications (e.g., years of experience, subspecialty) are provided in this summary.
  • Ground Truth: For the "anatomical fit," the ground truth was established by the subjective assessment of "surgeons" performing simulated implantations on anatomical models and completing a questionnaire.

4. Adjudication Method for the Test Set

  • For the anatomical fit, the document states "reviewed by each surgeon." This suggests individual assessments, but it doesn't specify a formal adjudication method (e.g., consensus, majority vote) if multiple surgeons were involved. It can be inferred that simple agreement among the involved surgeons was sufficient for the "acceptable" conclusion.
  • For other analytical and mechanical tests, adjudication would be based on predefined quantitative thresholds or qualitative observations against standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. This type of study is primarily relevant for diagnostic devices that involve interpretation by human readers (e.g., radiologists interpreting images). The VSP PEEK Cranial Implant is a structural implant, and its performance evaluation does not involve differential human interpretation of medical images.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable directly. This question primarily applies to AI/ML software devices. The "VSP PEEK Cranial Implant" is a physical device. While its design involves "CT imaging data and surgeon input," implying a design process that might be aided by software, the performance evaluation described here is for the physical implant itself, not a separate AI algorithm for image analysis or diagnosis. The "standalone" performance here refers to the device's intrinsic mechanical and biological properties.

7. The Type of Ground Truth Used

  • Bench Testing Standards & Expert Consensus/Subjective Assessment:
    • For biocompatibility, the ground truth is established by international standards (ISO 10993 series) and the measured physical/chemical properties or biological responses compared against the acceptance criteria defined by these standards.
    • For mechanical performance, the ground truth is established by literature research (as "there is no industry accepted standard governing mechanical testing for non-load bearing implantable prosthetic plates") and internal technical specifications, and the device's ability to withstand forces.
    • For anatomical fit, the "ground truth" is based on the subjective assessment and "acceptance" by the surgeons performing the simulated implantations.

8. The Sample Size for the Training Set

  • Not applicable. This device is a physical implant, not an AI/ML algorithm that requires a training set. The design of each implant is "individually for each patient" using their specific CT imaging data and surgeon input, which is a custom manufacturing process rather than a machine learning training process.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no AI/ML training set, there is no ground truth to establish for such a set.

{0}------------------------------------------------

April 4, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.

3D Systems, Inc. Ashley Dawson Director, Regulatory Affairs 5381 South Alkire Circle Littleton, Colorado 80127

Re: K231834

Trade/Device Name: VSP PEEK Cranial Implant Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: March 5, 2024 Received: March 5, 2024

Dear Ashley Dawson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows a digital signature. The signature is for Adam D. Pierce -S. The date of the signature is 2024.04.04, and the time is 16:48:51 -04'00'.

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological

{2}------------------------------------------------

and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231834

Device Name VSP PEEK Cranial Implant

Indications for Use (Describe)

The VSP PEEK Cranial Implant is indicated for use to fill a bony void or defect area in the cramial skeleton of patients 21 years of age and older.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY: K231834

Applicant
Name:3D Systems, Inc.
Address:5381 South Alkire CircleLittleton, CO 80127 USAPhone: +1 (720) 643-1001
Official Contact:Ashley Dawson, PhDDirector, Regulatory Affairs, Healthcare
Date Prepared:April 3rd, 2024
Device
Trade Name:VSP PEEK Cranial Implant
Common Name:Cranial Implant
Classification Name:Preformed alterable cranioplasty plate
Classification:Class II, 21 CFR 882.5320
Product Code:GWO

Predicate Device

Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI), Kelynium Global Inc. (K182711)

Device Description

The VSP PEEK Cranial Implants are non-load bearing, single use devices showing the following characteristics:

• The implants are designed individually for each patient to correct defects in cranial bone.

  • The implants are 3D printed using material extrusion technology.
  • The implants are manufactured using Polyetheretherketone (PEEK) implantable grade filament (i.e., Evonik VESTAKEEP® i4-3DF).

{5}------------------------------------------------

  • The VSP PEEK Cranial Implants are designed using CT imaging data and surgeon input.
  • · The implant size is ranging from 35 mm x 35 mm to 150 mm x 150 mm, and its thickness is constant, ranging from 3 mm to 6 mm with a nominal thickness of 3 mm.
  • If minor intra-operative adjustments are required, the implants can be modified by the surgeon with standard surgical burrs.
  • The implants are fixed to the native bone using commercially available cranial fixation screws and/or systems
  • The implants are provided non-sterile for sterilization prior to implantation.

Indications for Use Statement

The VSP PEEK Cranial Implant is indicated for use to fill a bony void or defect area in the cranial skeleton of patients 21 years of age and older.

Predicate Comparison Table

The VSP PEEK Cranial Implant and its predicate device have tunctionality and similar technical data as shown in the table below.

Subject DeviceVSP PEEK CranialImplantPredicate DeviceCustomized CraniofacialImplant (CCI) and CustomizedSkull Implant (CSI)Remark
Manufacturer3D Systems, Inc.Kelyniam Global Inc. (KGI)
510K numberK231834K182711
Indicationsfor UseThe VSP PEEK CranialImplant is indicated foruse to fill a bony void ordefect area in thecranial skeleton ofpatients 21 years of ageor older.The Customized CraniofacialImplant (CCI) and CustomizedSkull Implant (CSI) is intendedto fill a bony void or defectarea in a patient's specificcranial and craniofacialskeleton (orbital rim, zygoma, &adjacent bone).Indications of use for VSPPEEK Cranial Implant isequivalent to thepredicate subject deviceCustomized Skull Implant(CSI)
MaterialEvonik VESTAKEEP® i43DFInvibio Inc PEEK-Optima LT- 1®and Evonik Vestakeep i4®Equivalent
TechnicalspecificationsPlate - sized to eachpatient using CT scandataPlate - sized to each patientusing CT scan dataSame
SterilizationProvided Non-sterileProvided Non-sterileSame

{6}------------------------------------------------

Product codeGWO (882.5320, Preformed Alterable Cranioplasty Plate)GWO (882.5320, Preformed Alterable Cranioplasty Plate)Same
ClassificationClass IIClass IISame

Brief Summary of Performance Testing

The impact of the differences in the manufacturing process on the device performances are studied in detail. The performance testing of the VSP PEEK Cranial Implant includes the following tests to evaluate the substantial equivalence of the subject device to the predicate device:

  • Biocompatibility Testing ●
  • Performance Bench Testing: Mechanics
  • Performance Bench Testing: Anatomical fit

Biocompatibility Testing

The VSP PEEK Cranial Implant was biologically evaluated in order to verify compliance according to the technical specifications and standards of ISO-10993 ("Biological Evaluation of Medical Devices Part 1: Evaluation and Testing").

Biocompatibility tests were conducted according to the above-mentioned standards and the results obtained from test specimens were found to be within the acceptance criteria described in the standards.

TestResultsConclusion
ChemicalCharacterizationand ToxicologicalEvaluation(ISO 10993-18 andISO 10993-17)Analysis of extractables byHS-GC-MS, GC-MS, LC-MS,and ICP-MS followingexhaustive extractionAcceptable Margin of Safety for allreported extractable substances
Cytotoxicity(ISO 10993-5)Cell culture treated with testsample and compareddehydrogenase activity tocontrol
Sensitization(ISO 10993-10)In vivo Maximization TestNon-sensitizing

{7}------------------------------------------------

Irritation(ISO 10993-23)In vivo intracutaneousreactivity testNon-irritating
Pyrogenicity(ISO 10993-11)In vivo pyrogenicity testNon-pyrogenic
Acute SystemicToxicity(ISO 10993-11)Clinical examinationfollowing injection in mousemodelNon-toxic
Implantation effects(ISO 10993-6)Implantation within ratcalvarial bone 28 day follow-up and 90 day follow-upNo unexpected results
Genotoxicity(ISO/TR 10993-33)Reverse Mutation AssayNon-Mutagenic
Genotoxicity(ISO/TR 10993-33)In vitro Mammalian Cell GeneMutation AssayNon-Mutagenic

Performance Bench Testing

Various performance-bench tests regarding mechanics, including static testing and impact testing, were performed on the VSP PEEK Cranial Implant. The testing protocol was set up based on literature research, since there is no industry accepted standard governing mechanical testing for non-load bearing implantable prosthetic plates. Results indicate that the VSP PEEK Cranial Implant as replacements for bone defects have acceptable mechanical performance following sterilization or when used in combination with common osteosynthesis fixation systems.

Additionally, the size accuracy and usability of the implant was tested performing an anatomical fit test. Simulated surgical procedures were performed on anatomical models on test bench by surgeons. Several implantations were performed using commercially available cranioplasty fixation systems and reviewed by each surgeon. Results indicate that the size accuracy and usability of the VSP PEEK Cranial Implant is acceptable to surgeons.

The VSP PEEK Cranial Implant was bench tested in order to verify compliance according to the technical and biomechanical specifications.

{8}------------------------------------------------

TestTest Method SummaryResults
MechanicalPerformanceTestingStatic compression testing anddynamic impact testing ofextreme geometries (includingscrew fixation testing)All samples passed theacceptance criteria
Anatomical FitTestingSize accuracy and usabilitysurgeon questionnairefollowing implantation onanatomical modelThe size accuracy and usabilityof the VSP PEEK Cranial Implantwas acceptable to surgeons
Verification andValidation TestingTechnical and BiomechanicalSpecificationsAll samples compliant withspecifications.

CONCLUSION

According to the comparison of the intended use/indications for use and the non-clinical performance testing, it is concluded that the subject device is substantially equivalent to the predicate device.

§ 882.5320 Preformed alterable cranioplasty plate.

(a)
Identification. A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.(b)
Classification. Class II (performance standards).