K Number
K182711

Validate with FDA (Live)

Date Cleared
2019-05-21

(236 days)

Product Code
Regulation Number
882.5320
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is intended to fill a bony void or defect area in a patient's specific cranial and craniofacial skeleton (orbital rim, zygoma, & adjacent bone).

Device Description

Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is an individually sized and shaped implantable prosthetic plates intended to fill a bony void or defect area in a specific patient's cranial and craniofacial skeleton (orbital rim, zygoma, & adjacent bone), ranging from 2 mm to 10 mm thick (typically 4mm thick based on the patient CT Scan imaging data) x 25mm to 250 mm wide x 25 mm to 250 mm long. A Customized Skull Implant (CSI) is intended to fill a bony void or defect area in a specific patient's skull whereas a Customized Craniofacial Implants (CCI) is intended to fill a bony void or defect area in a specific patient's facial region of the skull, excluding the Maxilla (upper jaw area surrounding the teeth only) and Mandible, which are both considered load bearing areas of the facial region of the skull. The size, asymmetrical shape, thickness, contour, and edge profile are design elements of the non-load bearing patient-specific Base Implant that are used to support the base implant in the bony void or defect area while providing for a "Precise Fit". The single patient use base implant is (1) fabricated from a billet block of natural implant grade Polyether ether ketone (PEEK) Thermoplastic Polymer using the patient's CT Scan imaging data, (2) provided clean but non-sterile for steam sterilization prior to implantation at a hospital or surgical site with neurosurgery capabilities, and (3) attached to the native bone using commercially available cranioplasty hardware and fasteners. The additional features listed below are additional design elements that may be added to the Base Implant and Base Implant with Perfusion Holes, as requested by the Physician. The Base Implant, as defined above (as previously cleared on K103582 and K121755).

AI/ML Overview

The provided document outlines the Kelyniam Global Inc. (KGI) Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) for which a 510(k) premarket notification (K182711) was submitted.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document does not present a formal table of acceptance criteria versus reported device performance for specific clinical or diagnostic metrics. Instead, the "Performance Data" section details the types of validations and tests conducted to demonstrate the device's substantial equivalence to predicate devices and its safety and effectiveness.

Here's a summary of the performance data and "accepted" findings:

Acceptance Criterion (Implicit)Reported Device Performance / Finding
Material Biocompatibility- Leveraged from previously cleared devices. - Risk assessment conducted to evaluate the impact of design changes on biocompatibility. - Result: Subject device meets biological safety requirements per ISO 10993-1 for permanently implanted devices with tissue/bone and cerebrospinal fluid contact.
Material Specification Compliance- Invibio PEEK Optima LT-1® and Evonik Vestakeep i4® Material Specifications and Certification of Analysis/Certification of Compliance reviewed against approved specifications. - Result: Accepted against approved specifications at each receipt as part of receiving inspection activities. Conforms to ASTM F2026-16 Standard Specification for Polyether ether ketone (PEEK) Polymers for Surgical Implant Applications.
Cleaning Validation- Conducted in accordance with ANSI/AAMI ST72:2011, USP <161>, USP <85>, EP 2.6.14, and JP 4.01. - Result: All samples passed the acceptance criteria.
Steam Sterilization Validation- Conducted in accordance with ANSI/AAMI/ISO 17665-1:2006 and ANSI/AAMI/ISO 14937:2009. - Result: All samples passed the acceptance criteria.
Mechanical Testing (for non-load bearing implants)- No industry-accepted standard exists. - A Mechanical Testing Validation Protocol was developed by KGI. - Result: All samples passed the acceptance criteria. (Specific criteria for this protocol are not detailed in the summary).
Ship Testing Validation- Conducted in accordance with FedEx Packaging Testing Under 150Lb (similar to ITSA-2A-2011). - Result: All samples passed the acceptance criteria.
MR Safety (Magnetic Resonance Compatibility)- Evaluation of magnetically induced displacement force and torque, radio frequency (RF) heating, and image artifacts. - Result: The implants are electrically nonconductive or a nonmagnetic item and pose no known hazards in all MR environments. KGI Customized Craniofacial (CCI) and Customized Skull (CSI) Implants are MR Safe.
Overall Safety and Effectiveness/Substantial Equivalence- Based on biocompatibility, steam sterilization, cleaning, mechanical, and ship testing, and comparison of indications for use, technology, and materials with predicate devices. - Result: Device is safe and effective and substantially equivalent to the predicate devices.

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of clinical data for performance evaluation in the way one might for an AI/diagnostic device. The studies performed are primarily bench testing and material characterization studies.

  • Sample Size: The sample sizes for the various bench tests (e.g., cleaning, sterilization, mechanical, ship testing) are not explicitly stated in the summary. It generally mentions "all samples passed the acceptance criteria," implying that a sufficient number of units were tested as per the respective validation protocols.
  • Data Provenance: The data provenance is internal to the manufacturer (KGI) through their own validation protocols and testing, or from material suppliers (Invibio PEEK Optima LT-1® and Evonik Vestakeep i4®) for material specifications and certifications of analysis which are then reviewed by KGI. This data is generally considered prospective in the sense that the tests were conducted specifically to validate the device, rather than analyzed retrospectively from existing clinical datasets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the device is an implant (Customized Craniofacial/Skull Implant), not a diagnostic or AI device that requires expert interpretation for establishing ground truth in clinical cases. The "ground truth" for this device relates to its material properties, manufacturing quality, biomechanical integrity, and biocompatibility, which are verified through standardized testing and material certifications.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable for the type of device and studies performed. Adjudication methods are typically used in clinical studies, especially those involving human interpretation of diagnostic results, to reconcile discrepancies and establish a consensus ground truth. The studies cited are laboratory and bench tests relying on objective measurements and established standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a physical, custom-made implant, not an AI-powered diagnostic tool. Therefore, an MRMC study or an evaluation of human reader improvement with AI assistance is not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable. The device is an implant. No algorithm or standalone AI performance was evaluated.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this medical device is established through:

  • Standards Compliance: Adherence to recognized national and international standards for biocompatibility (ISO 10993-1), materials (ASTM F2026-16), cleaning (ANSI/AAMI ST72, USP <85>), and sterilization (ANSI/AAMI/ISO 17665-1, 14937).
  • Material Certifications: Certifications of analysis/compliance from raw material suppliers.
  • Engineering Specifications: Internal KGI specifications and validation protocols for mechanical testing and ship testing.
  • Objective Measurements: Laboratory measurements to confirm physical properties, absence of contaminants, and sterility.

No clinical outcomes data or pathology reports were used here as "ground truth" in this specific 510(k) summary, as the submission focused on demonstrating substantial equivalence through non-clinical testing. This is explicitly stated: "Clinical Testing was determined not applicable for Kelyniam Global Inc. Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI)."

8. The sample size for the training set

This is not applicable. The device is a physical implant and not an AI or machine learning model that requires a training set.

9. How the ground truth for the training set was established

This is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

May 21, 2019

Kelyniam Global Inc. Eric Boyea Quality Director 97 River Road. Suite A Canton, Connecticut 06019

Re: K182711

Trade/Device Name: Customized Craniofacial Implant (CCI), Customized Skull Implant (CSI) Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: April 15, 2019 Received: April 16, 2019

Dear Eric Boyea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew Krueger, M.S.E. Assistant Director THT5A1: Neurosurgical Devices DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182711

Device Name Customized Craniofacial Implant (CCI) Customized Skull Implant (CSI)

Indications for Use (Describe)

The Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is intended to fill a bony void or defect area in a patient's specific cranial and craniofacial skeleton (orbital rim, zygoma, & adjacent bone).

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

KELYNIAM

K182711 510(k) Summary

Date Prepared:Purpose:20MAY2019
Purpose of theSubmission:The purpose for the Premarket Notification 510(k) is for authorization to add additional features to thecurrently cleared (devices) base implant or base implant with Prefusion Holes. The additional features do notalter the intended therapeutic use of the subject device, the technological characteristic nor do they affect thesafety and effectiveness of the subject devices relative to the predicates. The additional features are intendedto respond to requests by surgeons to facilitate the surgical application of the implant(s) and do not introduceany new technological issues. The base implant, as well as the proposed new features is fabricated from abillet block of natural implant grade Polyether ether ketone (PEEK) Thermoplastic Polymer. This is the samematerial currently used by the listed predicate device organizations.
Submitter Information:
Submitter:Kelyniam Global Inc. (KGI), 97 River Road, Suite A, Canton, Connecticut 06019Tel: (800) 280-8192Fax: (501) 641-2000
OfficialCorrespondent:Eric Boyea, Director of Quality & Regulatory AffairsToll Free: (800) 280-8192 x 701Tel: (860) 352-5736Email: eboyea@kelyniam.com
Device Information
Trade Name:• Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI)
Common Name:ClassificationInformation:• Customized Cranial Implant or Craniofacial/Skull Implant• Panel: Neurology / Product Code: GWO / Classification: Class II in accordance with 21 CFR 882.5320,Preformed Alterable Cranioplasty Plate
PredicateDevices:• Primary Predicate - K121153 Stryker PEEK Customized Cranial Implant Kit• Secondary Predicate - K053199 Synthes Patient Specific Cranial/Craniofacial Implant (PSCI)• Tertiary Predicate - K103582 Kelyniam Global Inc. (KGI) Customized Skull Implant (CSI)• Quaternary Predicate - K121755 Kelyniam Global Inc. (KGI) Customized Craniofacial Implant (CCI)
DeviceDescription (withincluded featurespreviouslycleared):Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is an individually sized and shapedimplantable prosthetic plates intended to fill a bony void or defect area in a specific patient's cranial andcraniofacial skeleton (orbital rim, zygoma, & adjacent bone), ranging from 2 mm to 10 mm thick (typically4mm thick based on the patient CT Scan imaging data) x 25mm to 250 mm wide x 25 mm to 250 mm long. ACustomized Skull Implant (CSI) is intended to fill a bony void or defect area in a specific patient's skull whereasa Customized Craniofacial Implants (CCI) is intended to fill a bony void or defect area in a specific patient'sfacial region of the skull, excluding the Maxilla (upper jaw area surrounding the teeth only) and Mandible,which are both considered load bearing areas of the facial region of the skull. The size, asymmetrical shape,thickness, contour, and edge profile are design elements of the non-load bearing patient-specific Base Implantthat are used to support the base implant in the bony void or defect area while providing for a "Precise Fit".The single patient use base implant is (1) fabricated from a billet block of natural implant grade Polyetherether ketone (PEEK) Thermoplastic Polymer using the patient's CT Scan imaging data, (2) provided clean butnon-sterile for steam sterilization prior to implantation at a hospital or surgical site with neurosurgerycapabilities, and (3) attached to the native bone using commercially available cranioplasty hardware andfasteners. The additional features listed below are additional design elements that may be added to the BaseImplant and Base Implant with Perfusion Holes, as requested by the Physician.The Base Implant, as defined above (as previously cleared on K103582 and K121755).
Device Information (Continued)
AdditionalFeatures(Subject of thissubmission):Image: Integrated Fixation System (IFS) or IFS Tabs
Image: A Temporal Cutback
Image: A Multi-Part Implant
Indications forUse:The Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is intended to fill a bony void ordefect area in a patient's specific cranial and craniofacial skeleton (orbital rim, zygoma, & adjacent bone).
Statement ofSubstantialEquivalence:The KGI subject devices are intended to fill a bony void or defect area in a patient's specific Cranial andcraniofacial skeleton and are substantially equivalent in indications for use, technology, and material toK121153 Stryker PEEK Customized Cranial Implant Kit (Primary), K053199 Synthes Patient SpecificCranial/Craniofacial Implant (PSCI) (Secondary), K103582 KGI Customized Skull Implant (CSI) (Tertiary), andK121755 KGI Customized Craniofacial Implant (CCI) (Quaternary).
Comparison ofTechnologicalCharacteristicswith thePredicateDevice(s):The KGI subject devices are manufactured from implant grade Polyether ether ketone Thermoplastic Polymermaterial, known hereafter as PEEK, sold non-sterile, and is intended to fill a bony void or defect area in a patient'sspecific Cranial and craniofacial skeleton. The additional features, as listed above, are incorporated in the BaseImplant during the design phase and CNC Machined from a billet block of PEEK. The additional features areintended to respond to requests by surgeons to facilitate the surgical application of the implant(s) and do not

Perfusion Holes are a design element that allow the passage of fluid to an organ or tissue (as previously cleared on K103582 and K121755).

{4}------------------------------------------------

KELYNIAM

K182711 510(k) Summary

introduce any new technological issues. See Device Comparison Chart below for details. 1

Subject Devices vs Predicate Devices Comparison Chart with included Additional Features

Subject DeviceSubject DevicePrimary PredicateSecondary PredicateTertiary PredicateQuaternary Predicate
ManufacturerKelyniam Global Inc. (KGI)StrykerSynthesKelyniam Global Inc. (KGI)
510(k) #K182711K121153K053199K103582K121755
DescriptionCustomized CraniofacialImplant (CCI)Customized SkullImplant (CCI)PEEK Customized CraniofacialImplant KitPatient Specific Cranial/Craniofacial Implant (PSCI)Customized SkullImplant (CSI)Customized CraniofacialImplant (CCI)
Indications for UseIs intended to fill a bony void or defectarea in a patient's specific cranial andcraniofacial skeleton (orbital rim, zygoma,& adjacent bone)Is intended to be used toreplace bony voids in thecranial/craniofacial skeletonIs intended to replacebony voids in the Cranial/craniofacial skeletonIs intended to fill a bony void or defectarea in a patient's specific cranial andcraniofacial skeleton
MaterialInvibio Inc PEEK-Optima LT-1® and EvonikVestakeep i4®Invibio Inc PEEK-Optima LT-1®CP Titanium and PEEKInvibio Inc PEEK-Optima LT-1®
TechnicalSpecificationPlate - Custom sized to each patient usingCT scan dataCustomized patient-specificimplant based on CT scan dataImplants arepreformed/pre-shaped tofit the anatomy of thepatientCustom sized to each patient using CT scandata
SterilizationProvided Non-sterileProvided Non-sterileProvided Non-sterileProvided Non-sterile
Product CodeGWO (882.5320, Preformed AlterableCranioplasty Plate)GWOGXNGXN (882.5330, Preformed Non-alterableCranioplasty Plate)
ClassificationClass IIClass IIClass IIClass II
Perfusion HolesXXX-XX
IFSXX----
Temporal CutbackXX----
Multi-Part ImplantXX-X--

{5}------------------------------------------------

KELYNIAM

K182711 510(k) Summary

Performance Data
PerformanceDataBiocompatibilityTesting:Performancedata has beenprovided insupport of thesubstantialequivalencedetermination• Biocompatibility was leveraged from previously cleared devices. A risk assessment was conducted to evaluatethe impact of design changes on the biocompatibility of subject device. The results demonstrate that thesubject device meets biological safety requirements per ISO 10993-1 for permanently implanted devices thathave tissue/bone and cerebrospinal fluid contact.
PerformanceTesting Bench:Performancedata has beenprovided insupport of theverification &validation testingconducted for adetermination ofsubstantiallyequivalence. Thesubject devicesare safe,effective, andsubstantiallyequivalent to thepredicatedevices.• Invibio PEEK Optima LT-1® and Evonik Vestakeep i4® Material Specifications and Certification ofAnalysis/Certification of Complianceo ASTM F2026-16 Standard Specification for Polyether ether ketone (PEEK) Polymers for Surgical ImplantApplications.o PEEK Certification of Analysis/Certification of Compliance are reviewed and accepted against approvedspecifications at each receipt as part of receiving inspection activities.• Cleaning Validation testing was conducted in accordance with the following:o ANSI/AAMI ST72:2011 Bacterial endotoxins, Test methods, routine monitoring, and alternatives to batchtesting.o USP <161>, USP <85>, EP 2.6.14, and JP 4.01. USP <85> Bacterial Endotoxin Test (BET).• Steam Sterilization Validation testing was conducted in accordance with the following:o ANSI/AAMI/ISO 17665-1:2006, Sterilization of health care products - Moist heat - Part 1 - Requirements forthe development, validation, routine control of a sterilization process for medical devices, Annex D.o ANSI/AAMI/ISO 14937:2009, Sterilization of health care products, general requirements forcharacterization of a sterilizing agent & development, validation, and routine control of a sterilizationprocess for medical devices, Annex D.• Mechanical Testing Validationo There is no industry accepted standard governing mechanical testing for non-load bearing implantableprosthetic plates. A Mechanical Testing Validation Protocol was developed by KGI to validate the subjectdevices.• Ship Testing Validation testing was conducted in accordance with the following:
Summary onNon-ClinicalTesting:o FedEx Packaging Testing Under 150Lb (Same as ITSA-2A-2011 Packaged-Products 150 lb (68 kg) or Less)The safety and compatibility of passive implants in the Magnetic Resonance (MR) environment concernmagnetically induced displacement force and torque, radio frequency (RF) heating, and image artifacts. Thestatic magnetic field of the MR system induces displacement forces and torques on magnetic materials: TheKelyniam Global Inc. (KGI) implants are electrically nonconductive or a nonmagnetic item and poses no knownhazards in all MR environments. The KGI Customized Craniofacial (CCI) and Customized Skull (CSI) Implants areMR Safe.
Summary onClinical Testing:Clinical Testing was determined not applicable for Kelyniam Global Inc. Customized Craniofacial Implant (CCI)and Customized Skull Implant (CSI).
Conclusion:The biocompatibility testing, steam sterilization testing, cleaning testing, mechanical testing, and ship testing ofthe subject devices with and without the proposed features was performed and the results of these tests aswell as additional product related validations (all samples passed the acceptance criteria) support the device issafe and effective and substantially equivalent to the predicate devices.

§ 882.5320 Preformed alterable cranioplasty plate.

(a)
Identification. A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.(b)
Classification. Class II (performance standards).