Intended Use

The INERTIA® CONNEXX™ Modular Pedicle Screw System is intended for immobilization and stabilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the INERTIA® CONNEXX™ Modular Pedicle Screw System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis and congenital scoliosis. Additionally, the system is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis / spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. The INERTIA® CONNEXX™ Modular Pedicle Screw System is to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Device Description

The INERTIA® CONNEXX™ Modular Pedicle Screw System consists of rods, pedicle screws, connectors and fasteners. Rods are available in either straight, or pre-contoured (curved) or revision forms in a variety of lengths. Pedicle screws are available in modular and non-modular polyaxial, and non-modular uniplanar designs having double lead standard or cortical/cancellous thread forms in a variety of diameter-length combinations. Connectors include offset, various wedding band options and transverse rod-to-rod. Set screws are used to fasten the components. The INERTIA® CONNEXX™ Modular Pedicle Screw System implants may be sold sterile or nonsterile.

AI/ML Overview

This document is a 510(k) Premarket Notification for a medical device called the "INERTIA® CONNEXX™ Modular Pedicle Screw System." It is a submission to the FDA to demonstrate substantial equivalence to existing legally marketed devices.

Based on the provided text, the device in question is a pedicle screw system for spinal immobilization and stabilization, not an AI/software-driven diagnostic tool. Therefore, the information typically requested in questions 1-9 (related to acceptance criteria for AI algorithms, sample sizes for training/test sets, expert adjudication, MRMC studies, etc.) is not applicable to this type of device submission.

The "Performance Data" section explicitly states:

  • "Mechanical testing of worst case INERTIA® CONNEXX™ constructs included dynamic compression bending according to ASTM F1717."
  • "The results demonstrate that INERTIA® CONNEXX™ Modular Pedicle Screw System performance is substantially equivalent to the predicate."

This indicates that the acceptance criteria and proof of performance are based on mechanical testing of the physical device to ensure it meets established ASTM standards and is comparable to previously cleared devices, rather than on the performance of a diagnostic algorithm or human-AI interaction.

Therefore, I cannot provide answers to questions 1-9 regarding acceptance criteria and study details for an AI-driven device, as this document describes a physical spinal implant.

{0}------------------------------------------------

July 20, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right, the FDA logo is displayed in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION". The logo is simple and professional, conveying the organization's role in regulating food and drugs.

Nexxt Spine LLC % Karen Warden President BackRoads Consulting Inc. PO Box 566 Chesterland, Ohio 44026

Re: K231799

Trade/Device Name: INERTIA® CONNEXX™ Modular Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: June 19, 2023 Received: June 20, 2023

Dear Karen Warden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

2

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image contains the name "Colin O'neill -S" on the left side of the image. On the right side of the image is the acronym "FDA" in a light blue color. The acronym is much larger than the name on the left side of the image. The background of the image is white.

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K231799

Device Name

INERTIA® CONNEXX™ Modular Pedicle Screw System

Indications for Use (Describe)

The INERTIA® CONNEXX™ Modular Pedicle Screw System is intended for immobilization and stabilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

When used for posterior non-cervical pedicle screw fixation in pediatic patients, the INERTIA® CONNEXX™ Modular Pedicle Screw System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis and congenital scoliosis. Additionally, the system is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. The INERTIA® CONNEXX™ Modular Pedicle Screw System is to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

510(k) Summary

Date:19 June 2023
Sponsor:Nexxt Spine, LLC14425 Bergen Blvd, Suite BNoblesville, IN 46060Office: 317.436.7801Fax: 317.245.2518
Sponsor Contact:Andy Elsbury, President
510(k) Contact:Karen E. Warden, PhDBackRoads Consulting Inc.PO Box 566Chesterland, OH 44026Office: 440.729.8457
Proposed Trade Name:INERTIA® CONNEXX™ Modular Pedicle Screw System
Common Name:Posterior pedicle screw system
Regulatory Class:Class II
Regulation Name:Thoracolumbosacral pedicle screw system
Regulation Number:21 CFR 888.3070
Product Code:NKB
Submission Purpose:The subject 510(k) adds connectors and revision rod options to theINERTIA® Pedicle Screw System.
Device Description:The INERTIA® CONNEXX™ Modular Pedicle Screw System consists ofrods, pedicle screws, connectors and fasteners. Rods are available in eitherstraight, or pre-contoured (curved) or revision forms in a variety of lengths.Pedicle screws are available in modular and non-modular polyaxial, andnon-modular uniplanar designs having double lead standard orcortical/cancellous thread forms in a variety of diameter-lengthcombinations. Connectors include offset, various wedding band options andtransverse rod-to-rod. Set screws are used to fasten the components.The INERTIA® CONNEXX™ Modular Pedicle Screw System implants maybe sold sterile or nonsterile.
Indications for Use:The INERTIA® CONNEXX™ Modular Pedicle Screw System is intended forimmobilization and stabilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally mature patients as an adjunct to fusion for thefollowing indications: degenerative disc disease (defined as back pain ofdiscogenic origin with degeneration of the disc confirmed by history andradiographic studies), spondylolisthesis, trauma (i.e., fracture ordislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, orlordosis), tumor, pseudarthrosis, and/or failed previous fusion.When used for posterior non-cervical pedicle screw fixation in pediatricpatients, the INERTIA® CONNEXX™ Modular Pedicle Screw Systemimplants are indicated as an adjunct to fusion to treat progressive spinaldeformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathicscoliosis, neuromuscular scoliosis and congenital scoliosis. Additionally, thesystem is intended to treat pediatric patients diagnosed with the followingconditions: spondylolisthesis / spondylolysis, fracture caused by tumorand/or trauma, pseudarthrosis, and/or failed previous fusion. The INERTIA®CONNEXX™ Modular Pedicle Screw System is to be used with autograftand/or allograft. Pediatric pedicle screw fixation is limited to a posteriorapproach.
Materials:INERTIA® CONNEXX™ implants are manufactured from Ti6Al4V ELItitanium alloy (ASTM F136) or cobalt chrome per ASTM F1537 (rods only).
Primary Predicate:INERTIA® Pedicle Screw and Deformity Correxxion™ System (Nexxt SpineLLC - K153453)
Additional Predicates:Astra Spinal System (SpineCraft, LLC - K223273), INERTIA® PedicleScrew System (Nexxt Spine, LLC - K090984, K101278, K132412,K141376, K221905)]
Performance Data:Mechanical testing of worst case INERTIA® CONNEXX™ constructsincluded dynamic compression bending according to ASTM F1717. Theresults demonstrate that INERTIA® CONNEXX™ Modular Pedicle ScrewSystem performance is substantially equivalent to the predicate.
TechnologicalCharacteristics:The INERTIA® CONNEXX™ devices possess the same technologicalcharacteristics as one or more of the predicate devices. These include:● performance (as described above),● basic design (rod and screw configuration),● material (titanium alloy) and● size (dimensions are comparable to those offered by the cleared devices).Therefore the fundamental scientific technology of the INERTIA®CONNEXX™ Modular Pedicle Screw System devices is the same aspreviously cleared devices.
Conclusion:The INERTIA® CONNEXX™ devices possesses the same intended use andtechnological characteristics as the predicate devices. Therefore theINERTIA® CONNEXX™ Modular Pedicle Screw System is substantiallyequivalent for its intended use.

{4}------------------------------------------------

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.