K Number
K132412

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-10-10

(69 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Inertia™ MIS Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine (T1 to S2): severe spondylolisthesis (grades 3 and 4) of the L5-SI vertebra: degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; spinal stenosis; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis).

Device Description

The Inertia™ MIS Pedicle Screw System consists of rods, polyaxial screws and set screws. Rods are available in either straight or pre-contoured (curved) forms and in a variety of lengths. Polyaxial screws are available in a variety of diameter-length combinations. Set screws are used to fasten the rod and polyaxial screw. The purpose of this 510(k) is to add cannulated pedicle screws to the Inertia™ MIS Pedicle Screw System.

AI/ML Overview

The provided 510(k) summary for the Inertia™ MIS Pedicle Screw System describes a mechanical testing study to demonstrate substantial equivalence to a predicate device. This study focuses on engineering performance, not AI/ML-driven diagnostic or assistive capabilities. Therefore, many of the requested categories (e.g., sample size for test set, number of experts for ground truth, MRMC study, sample size for training set, etc.) are not applicable to this type of device and study.

Here's a breakdown of the available information:

Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Defined in the Design Control Activities Summary according to ASTM F1717 for static, dynamic compression bending, and static torsion.The results "demonstrate that the acceptance criteria defined in the Design Control Activities Summary were met."

Details of the study that proves the device meets the acceptance criteria:

  1. Sample size used for the test set and the data provenance: Not applicable. This was mechanical testing, not a study involving patient data. The "test set" would refer to the physical devices tested.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for mechanical testing is established by engineering standards (ASTM F1717) and measurement, not expert consensus.
  3. Adjudication method for the test set: Not applicable. This was mechanical testing against predefined engineering standards.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No. This device is a pedicle screw system, a physical implant. It does not involve AI or human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No. This is a physical medical device, not an algorithm.
  6. The type of ground truth used: Mechanical testing data compared against the performance requirements outlined in the ASTM F1717 standard and the company's "Design Control Activities Summary."
  7. The sample size for the training set: Not applicable. There is no "training set" as this is not an AI/ML device.
  8. How the ground truth for the training set was established: Not applicable.

Summary of the Study:

The study was a mechanical testing study conducted according to ASTM F1717. It involved testing the "worst case" subject Inertia™ MIS Pedicle Screw System and comparing its performance to the predicate Inertia™ Pedicle Screw System. The tests performed were:

  • Static compression bending
  • Dynamic compression bending
  • Static torsion

The results of these tests were reported to have met the acceptance criteria defined in the Design Control Activities Summary. This demonstrates the substantial equivalence of the new cannulated pedicle screws to the predicate device in terms of mechanical performance.

{0}------------------------------------------------

510(k) Summarv

Device Trade Name:InertiaTM MIS Pedicle Screw System
Manufacturer:Nexxt Spine, LLC14425 Bergen Blvd Suite BNoblesville, IN 46060
Contact:Eric LintulaDirector of Engineering and Regulatory AffairsPhone: 317.840.3816
Date Prepared:July 31, 2013
Classifications:21 CFR §888.3070, Pedicle screw spinal system.
Class:II
Product Codes:MNH, MNI, KWP

Indications For Use:

The Inertia™ MIS Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine (T1 to S2): severe spondylolisthesis (grades 3 and 4) of the L5-SI vertebra: degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; spinal stenosis; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis).

Device Description:

The Inertia™ MIS Pedicle Screw System consists of rods, polyaxial screws and set screws. Rods are available in either straight or pre-contoured (curved) forms and in a variety of lengths. Polyaxial screws are available in a variety of diameter-length combinations. Set screws are used to fasten the rod and polyaxial screw.

The purpose of this 510(k) is to add cannulated pedicle screws to the Inertia™ MIS Pedicle Screw System.

Predicate Device:

The subject Inertia™ MIS Pedicle Screw System is substantially equivalent to predicate Inertia™ Pedicle Screw System (K101278, K090984) with respect to indications, design, function, and materials.

{1}------------------------------------------------

Substantial Equivalence:

Mechanical testing (static, dynamic compression bending, static torsion) was performed according to ASTM F1717 on the worst case subject Inertia™ MIS Pedicle Screw System and predicate Inertia™ Pedicle Screw System. The results demonstrate that the acceptance criteria defined in the Design Control Activities Summary were met.

{2}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 10. 2013

Nexxt Spine, LLC % Mr. Justin Eggleton Musculoskeletal Clinical Regulatory Advisers, LLC 1331 H Street Northwest, 12" Floor Washington, District of Columbia 20005

Re: K132412

Trade/Device Name: Inertia™ MIS Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI. MNH. KWP Dated: September 11, 2013 Received: September 13, 2013

Dear Mr. Eggleton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA *s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Justin Eggleton

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-204 ) or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ering Keith

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation · Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) K132412

Device Name

Inertia(TM) MIS Pedicle Screw System

Indications for Use (Describe)

The Inertia™ MIS Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spinc (TI 10 S2): severe spondylolisthesis (grades 3 and 4) of the LS-S1 vertebra; degenerative spondylolishesis with objective evidence of neurologic impairment; facture; spinal stenosis; scolosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

Image /page/4/Figure/11 description: The image shows a document with the heading "FOR FDA USE ONLY". Below the heading is the text "Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)". The name "Ronald P. Jean -S" is written below this text, presumably as a signature.

N/A