K Number
K231760

Validate with FDA (Live)

Device Name
HP-OCT (HP-OCT)
Manufacturer
Date Cleared
2024-03-08

(266 days)

Product Code
Regulation Number
886.1850
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HP-OCT™ instrument is a non-contact ophthalmic imaging and analysis device. It is indicated for visualization of the posterior segment of the eye.

Device Description

The Cylite Hyperparallel Optical Coherence Tomographer (HP-OCT) is an ophthalmic OCT system, intended to assist ophthalmic professionals in the diagnosis of ophthalmic conditions via non-contact imaging of the posterior segment. OCT imaging is accomplished in the HP-OCT instrument using a micro-lens array to separate the illumination from a super-luminescent diode (SLD) source into an array of parallel beamlets. The beamlet array is then projected simultaneously onto the eye. Each of the individual beamlets is equivalent to a single beam of an SD-OCT (Spectral Domain OCT) system. Each beamlet generates a spectral interferogram, equivalent to a single SD-OCT interferogram, which is processed to produce an individual A-scan. A set of 1008 parallel, SD-OCT type A-scans is thus acquired for each frame of the CMOS sensor, and multiple frames can be acquired to create dense volume images.

The HP-OCT instrument is supplied with an exchangeable lens which supports OCT imaging of the posterior segment.

The HP-OCT instrument is operated from the Cylite Focus softwarethat runs from a Windows 10 personal computer (PC) supplied by the user. The PC hosts the HP-OCT Capture and Focus Review software and connects to the HP-OCT instrument via a USB 3.0/3.1 cable. From the PC/software, the user employs the PC monitor, keyboard and mouse to interact with the HP-OCT instrument to perform instrument alignment, preview images, capture images, save or load image or patient data, review images, and manage patient data.

AI/ML Overview

The Cylite HP-OCT is an ophthalmic OCT system intended for non-contact imaging and analysis of ocular structures, specifically for visualization of the posterior segment of the eye. Its substantial equivalence to the predicate device, the Carl Zeiss Meditec, Inc. Cirrus HD-OCT 5000 (K181534), was evaluated through both non-clinical and clinical performance testing.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state acceptance criteria in a quantitative format for either the non-clinical or clinical performance. Instead, it describes the assessment of various performance aspects and concludes with a qualitative statement of "favorable clinical performance profile that supports a determination of substantial equivalence."

However, based on the comparative effectiveness study, we can infer the acceptance criterion was likely non-inferiority or comparability in the ability to identify anatomical structures and retinal abnormalities.

Acceptance Criteria (Inferred from Clinical Study)Reported Device Performance (HP-OCT vs. Cirrus HD-OCT 5000)
Ability to identify anatomical structures and key retinal abnormalities should be comparable to or non-inferior to the predicate device (Cirrus HD-OCT 5000) for qualitative diagnosis."The grading results of overall SGS support a determination of substantial equivalence for the two devices in Normal and Retinal Disease eyes." The study demonstrated a "favorable clinical performance profile" for the identification of anatomic structures and key retinal abnormalities.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set): A total of 83 eyes (43 normal eyes, 40 retinal diseased eyes) were enrolled in the clinical study.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical data. It mentions the study was conducted at three sites, implying prospective data collection for this specific study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Three independent graders were used.
  • Qualifications of Experts: The specific qualifications (e.g., years of experience, subspecialty) of these graders are not provided in the document. They are identified as "image graders."

4. Adjudication Method for the Test Set

The document states that "For each image grader attempted to identify 9 anatomic structures and 11 abnormalities which were pre-defined and scored 1 (if identifiable) or 0 (if unidentifiable) for each structure/abnormality." It also mentions "inter-grader difference for HP-OCT for each structure and abnormity." However, it does not explicitly describe an adjudication method (like 2+1 or 3+1) to resolve discrepancies or establish a single ground truth from the three graders' individual assessments for either structures or abnormalities. It appears each grader's scores were compared directly between devices.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • MRMC Study: Yes, a comparative effectiveness study involving multiple readers (three independent graders) and multiple cases (83 eyes) was performed, comparing the HP-OCT to the predicate device.
  • Effect Size of Human Reader Improvement with AI: This information is not applicable as the HP-OCT is presented as an imaging device, not an AI-assisted diagnostic tool for human readers. The study compares the imaging performance of the device itself to a predicate, not the improvement of human readers using the device with AI versus without AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Standalone Performance: The document describes the device as a "non-contact ophthalmic imaging and analysis device" indicated for "visualization." The clinical study evaluated the device's ability to generate images that aid qualitative diagnosis by human graders. The results are based on human interpretation of the images produced by the device. Therefore, a standalone algorithm-only performance without human-in-the-loop is not explicitly described or indicated as the primary assessment. The device itself is not
    stated to perform automated diagnoses.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The ground truth for the clinical study was established by the identification and scoring of pre-defined anatomical structures and abnormalities by three independent expert graders based on the images generated by both the HP-OCT and the predicate device. This is a form of expert consensus/interpretation of imaging data. There is no mention of pathology, outcomes data, or other independent gold standards.

8. The Sample Size for the Training Set

The document does not provide information on a training set sample size. The clinical study described is a performance validation study, not a study describing the development and training of an algorithm within the HP-OCT device itself. The HP-OCT is an imaging device, and while it uses software (Cylite Focus software), the provided information focuses on its image acquisition capabilities rather than an AI/ML component that would require a distinct training set for diagnostic purposes.

9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned for an AI/ML diagnostic component, this information is not applicable.

{0}------------------------------------------------

March 8, 2024

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and features a stylized image of a human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

Cylite Pty Ltd % Lena Sattler President Orasi Consulting 226 1st Street Bonita Springs, Florida 34134

Re: K231760

Trade/Device Name: HP-OCT Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO Dated: February 3, 2024 Received: February 5, 2024

Dear Lena Sattler:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231760

Device Name HP-OCT

Indications for Use (Describe)

The HP-OCT™ instrument is a non-contact ophthalmic imaging and analysis device. It is indicated for visualization of the posterior segment of the eye.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cylite, a company that appears to be in the technology or software space. The logo features the word "cylite" in a modern, sans-serif font. The "y" in "cylite" is stylized with a teal color, while the rest of the letters are in a dark purple color.

510(k) Summary

1. General Information

Manufacturer:Cylite Pty Ltd300 Wellington RoadMulgrave Victoria 3170Australia+61-3-95812601
Sponsor Contact Person:Arnold OuyangVP Quality & Regulatory300 Wellington RoadMulgrave Victoria 3170Australia
Official Correspondent:Lena SattlerPresident, Orasi Consulting226 1st streetBonita Springs, FL 34134
Date Prepared:March 5, 2024
Device Information:
Classification Name:Trade/Propriety Name:Common Name:Regulation Number:Device class:Product Code:Tomography, Optical CoherenceHP-OCT™Optical Coherence Tomography21 CFR 886.1570IIOBO

2. Predicate Device

Company:Carl Zeiss Meditec, Inc.
Device:Cirrus HD-OCT 5000 (K181534)

The subject device and predicate device are both intended for non-contact imaging and analysis of ocular structures. Both devices are indicated for posterior segment visualization. The additional indications for use (IFU) offered by the predicate device do not change the intended use.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Cylite. The logo consists of the word "cylite" in a sans-serif font. The letters are mostly black, but the "l" is a teal color. The logo has a modern and clean design.

3. Device Description

The Cylite Hyperparallel Optical Coherence Tomographer (HP-OCT) is an ophthalmic OCT system, intended to assist ophthalmic professionals in the diagnosis of ophthalmic conditions via non-contact imaging of the posterior segment. OCT imaging is accomplished in the HP-OCT instrument using a micro-lens array to separate the illumination from a super-luminescent diode (SLD) source into an array of parallel beamlets. The beamlet array is then projected simultaneously onto the eye. Each of the individual beamlets is equivalent to a single beam of an SD-OCT (Spectral Domain OCT) system. Each beamlet generates a spectral interferogram, equivalent to a single SD-OCT interferogram, which is processed to produce an individual A-scan. A set of 1008 parallel, SD-OCT type A-scans is thus acquired for each frame of the CMOS sensor, and multiple frames can be acquired to create dense volume images.

The HP-OCT instrument is supplied with an exchangeable lens which supports OCT imaging of the posterior segment.

The HP-OCT instrument is operated from the Cylite Focus softwarethat runs from a Windows 10 personal computer (PC) supplied by the user. The PC hosts the HP-OCT Capture and Focus Review software and connects to the HP-OCT instrument via a USB 3.0/3.1 cable. From the PC/software, the user employs the PC monitor, keyboard and mouse to interact with the HP-OCT instrument to perform instrument alignment, preview images, capture images, save or load image or patient data, review images, and manage patient data.

4. Indications for Use

The HP-OCT™ instrument is a non-contact ophthalmic imaging and analysis device. It is indicated for visualization of the posterior segment of the eye.

5. Technological Comparison

The HP-OCT instrument and the predicate device Cirrus HD-OCT 5000 are both mains powered and examine the ocular structures by projecting measurement and/or illumination lights onto the eye and then collecting/analyzing the returning signals.

Both devices employ the same SLD light source with a wavelength spectrum centered at 840nm to generate an A-scan based on the same SD-OCT (Spectral Domain OCT) principle. Cirrus HD-OCT 5000 scans a single beam onto different locations over a short period of time to generate the A-scans that constitute a B-scan or C-scan, while the HP-OCT instrument projects and scans a grid of beamlets onto the eye surface at the same time and constructs C-scans from the A-scans generated by the individual beamlets. Both devices utilize the chin/head rest to facilitate positioning the patient's head for optical imaging. The body contact profiles are the same. No sterilization is required.

Other differences in the device technological characteristics include the following.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Cylite. The logo is in lowercase letters and the letters are connected. The "y" is teal and the rest of the letters are dark purple. The logo is simple and modern.

  • Although both devices employ LED for fixation/alignment illumination, the subject and . predicate devices use different LEDs with different wavelengths. These differences in these LED wavelengths do not raise different questions of safety and effectiveness.
  • While the range of technical specifications (e.g. acquisition rate, scan/imaging range/coverage, . resolutions) of the subjective HP-OCT instrument are not exactly the same as the predicate device, the differences in ranges do not raise different questions of safety and effectiveness.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Cylite. The logo is in a sans-serif font and is purple except for a green line that is part of the "y" and a green line that is part of the "i". The logo is simple and modern.

Subject devicePredicate device
CharacteristicsCylite HP-OCTZeiss Cirrus 5000 HD-OCTComparison
510(k) numberK231760K181534N/A
Regulation21CFR 886.157021CFR 886.1570Same
ClassIIIISame
Intended useNon-contact imaging and analysisof ocular structuresNon-contact imaging and analysisof ocular structuresSame
Indications for useThe HP-OCT™ instrument is aCIRRUS™ HD-OCT is a non-Different - Both devices
(IFU)non-contact ophthalmic imagingand analysis device. It is indicatedfor visualization of the posteriorsegment of the eye.contact, high resolutiontomographic and biomicroscopicimaging device intended for in-vivo viewing, axial cross-sectional, and three-dimensionalimaging of anterior and posteriorocular structures. The device isindicated for visualizing andmeasuring anterior and posteriorocular structures, includingcornea, corneal epithelium,retina, retinal nerve fiber layer,ganglion cell plus inner plexiformlayer, macula, and optic nervehead. The CIRRUS normativedatabases are quantitative toolsindicated for the comparison ofretinal nerve fiber layerthickness, macular thickness,ganglion cell plus inner plexiformlayer thickness, and optic nervehead measurements to adatabase of normal subjects.CIRRUS' AngioPlex OCTAngiography is indicated as anaid in the visualization of vascularstructures of the retina andchoroid. (Model 5000 only.)CIRRUS HD-OCT is indicated as adiagnostic device to aid in thedetection and management ofocular diseases including,but not limited to, macular holes,cystoid macular edema, diabeticretinopathy, age-related maculardegeneration, and glaucoma.support posterior segmentimaging. The predicate deviceprovides automatedmeasurements of anterior andposterior ocular structures. Thesubject device does notprovide automatedmeasurements of any ocularparameters.
Imaging methodSpectral Domain OCT - C scanfrom multiple A scans generatedfrom a grid of beamlets beingsimultaneously projected ontothe eyeSpectral Domain OCT – B/C scanfrom multiple A scans generatedfrom the single beam beingscanned over the eye in a shortperiod of timeSimilar – The same OCToperating principle. DifferentA-scan acquisition methods donot raise different questions ofsafety or effectiveness.
Measurement lightsourceSuperluminescent Diode (SLD),840 nmSuperluminescent Diode (SLD),840 nmSame
Fixation/alignmentlight sourceLED - White (6500 K), Red (633nm), IR (940 nm)LED – 523 nm, 627 nmDifferent - Different LED lightwavelengths in auxiliaryfunctions (fixation andalignment) do not raisedifferent questions of safety oreffectiveness.
Acquisition rate>300K A-scans/second68k A-scans/secondDifferent - Higher acquisitionrate of the subject device doesnot raise different questions ofsafety or effectiveness.
CharacteristicsSubject deviceCylite HP-OCTPredicate deviceZeiss Cirrus 5000 HD-OCTComparison
Axial resolution9.0 µm (in tissue)5 µm (in tissue)1Different - Imaging resolutionsare not exactly same but thedifferences do not raisedifferent questions of safety oreffectiveness.
Lateral resolution(retina)19 µm15 µm1Different - Imaging resolutionsare not exactly same but thedifferences do not raisedifferent questions of safety oreffectiveness.
Retinal coverage (fieldof view)9.4 x 5.4 mm6 mm square grid1Different - Wider coverage ofthe subject device does notraise different questions ofsafety or effectiveness.
Patient contactingpartsChin and head restChin and head restSame
Skin contact duration(typical)< 24 hours< 24 hoursSame
ReusableYesYesSame
SterileNoNoSame
Rated voltage /frequency100 V to 240 V AC (± 10 %) / 50/60Hz100-120 V~ 50-60 Hz 6.3A220-240 V~ 50-60 Hz 3.15ASimilar
TransportationconditionsTemperature: 0°C to +45°CRelative Humidity: 10% to 95%Temperature: -40°C to +70°C,Relative humidity: 10% to 100%Similar – No impact on clinicalperformance or safety whenused in the recommendedconditions.

Table 1 - Summary of Comparison between HP-OCT and the Predicate Device

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Cylite. The logo is in lowercase letters and is a dark purple color, except for the "y" which is teal. The logo is simple and modern.

HP-OCT 510(k) Notification K231760

1 Cirrus HD-OCT 5000 user manual

{8}------------------------------------------------

6. Performance Test Summary

6.1 Non-clinical performance testing

Non-clinical performance bench tests have been conducted on the HP-OCT instrument to demonstrate the substantial equivalence to the predicate device, including,

  • · Optical imaging performances including resolution, field of view, axial range, signal-to-noise ratio and depth attenuation.
  • Optical safety tests and assessments in accordance with the following international and consensus standards
    • ISO 15004-1:2020 Ophthalmic instruments Fundamental requirements and test methods— Part 1: General requirements applicable to all ophthalmic instruments
    • ISO 15004-2:2007 Ophthalmic Instruments Fundamental requirements and test methods - Part 2: Light hazard protection
    • ANSI Z80.36:2021 American National Standard for Ophthalmics Light Hazard - Protection for Ophthalmic Instruments
  • Electrical, mechanical and thermal safety tests in accordance with the following international and consensus standards
    • i IEC 60601-1:2005/A1:2012 - Medical electrical equipment, Part 1: General requirements for basic safety and essential performance
    • ANSI/AAMI ES60601-1:2005/(R)2012 - Medical electrical equipment, Part 1: General requirements for basic safety and essential performance
  • Electromagnetic compatibility tests in accordance with the following consensus standard
    • IEC 60601-1-2:2014/A1:2020 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
  • Biocompatibility tests in accordance with the following international and consensus standards
    • ISO 10993-1:2018 Biological evaluation of medical devices Part 1 Evaluation and testing within a risk management process
    • ISO 10993-5:2009 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
    • i ISO 10993-10:2010 - Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
  • Software verification and validation tests as a "Moderate" level of concern in accordance with FDA's guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The development of the software also complied with the consensus standard IEC 62304:2015 -Medical Device Software life-cycle processes.
  • Human factors validation (usability summative evaluation) in accordance with FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices" and the consensus

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Cylite. The logo is in lowercase letters and the first two letters, 'cy', are in a dark purple color. The 'l' is in a teal color and is slightly angled to the left. The 'ite' is in a dark purple color and the 'i' is a vertical line that is slightly taller than the other letters.

standard IEC 62366-1:2020 - Medical devices - Part 1: Application of usability engineering to medical devices.

6.2 Clinical Performance Testing

A clinical study was conducted at three sites to validate clinical performance of the HP-OCT. The clinical study was performed to investigate the reliability of the HP-OCT in generating retinal images to aid qualitative diagnosis as compared to the predicate device Cirrus HD-OCT 5000 (K181534).

The study recruited two groups of subjects, subjects with normal eyes and subjects with retinal diseased eyes into the study. The scan types employed were HP-OCT – Retina and Cirrus HD-OCT 5000 – Macular Cube.

Both eyes of the same subject could be enrolled in different eye groups if the corresponding eligibility criteria was satisfied. If both eyes of a subject were eligible for the same eye group, the study eye was randomly determined. A total of 83 eyes (43 normal eyes, 40 retina diseased eyes) were enrolled in this study. The age range was 19 to 66 years in Normal group with a mean of 36.5 (+/-12.6) years and a median of 33 years. The Retinal Disease group had a mean age of 69.9 (+/-12.2) years and a median age of 73.5 years, with the range between 25-85 years. Majority of the subjects were female in Normal (86%). In Retinal Disease group, the male subjects (53%) were slightly more than female ones (47%). White (49%) and Asian (49%) subjects were evenly distributed in the Normal group and a higher percentage of Whites was in the Retinal Disease group (77.5%).

Data Collection

The study employed three HP-OCT and three Cirrus HD-OCT 5000. Each of the three sites possessed one pair of the two devices, one HP-OCT and one Cirrus HD-OCT (two device units in total at a site) to perform the study.

All eligible eyes underwent non-contact ocular scans using the HP-OCT and Cirrus HD-OCT 5000 in a single site visit. The order of scan devices and scan types were randomized. All imaging activities for a subject were taken by one operator in a single visit. The scans which satisfied the imaging quality criteria according to the study protocol (and respective operator manuals) were deemed acceptable.

After imaging, the first acceptable retinal image acquired on the HP-OCT and the Cirrus HD-OCT 5000 for each eligible subject were sent to an image reading center for retinal image grading by three independent graders. The image graders were masked to subject identification, eye group allocation, eye examination results and each other's grading results. For each image grader attempted to identify 9 anatomic structures and 11 abnormalities which were pre-defined and scored 1 (if identifiable) or 0 (if unidentifiable) for each structure/abnormality.

Data Analysis

The subjective grade score (SGS) as the sum of individual scores was then calculated for structure (maximum 9) and abnormality (maximum 11), respectively. The analyses performed included the

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Cylite. The logo is in lowercase letters and the first letter 'c' is in black. The 'y' is split into two colors, with the left side in teal and the right side in black. The rest of the letters are in black.

mean, standard deviation, median, maximum, and minimum of SGS differences between HP-OCT and Cirrus HD-OCT 5000, two-sided Wilcoxon Signed-Rank test of SGS differences between devices and graders respectively, agreement rate of the same score for the two devices for each structure and abnormity, and the inter-grader difference for HP-OCT for each structure and abnormity. These analyses were performed by eye groups.

Results

No adverse effects were observed or reported.

For the identification of anatomic structures and key retinal abnormalities, the results of the clinical performance testing demonstrate a favorable clinical performance profile that supports a determination of substantial equivalence. The grading results of overall SGS support a determination of substantial equivalence for the two devices in Normal and Retinal Disease eyes.

7. Conclusion

Comparison of intended use and technological characteristics and evaluation of non-clinical testing and clinical performance data support the substantial equivalence of the HP-OCT subject device to the predicate device Cirrus HD-OCT 5000 for the indications for use.

§ 886.1850 AC-powered slitlamp biomicroscope.

(a)
Identification. An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.(b)
Classification. Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.