K Number
K181534

Validate with FDA (Live)

Device Name
CIRRUS HD-OCT
Date Cleared
2019-02-15

(249 days)

Product Code
Regulation Number
886.1570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device intended for invivo viewing, axial cross-sectional, and three-dimensional imaging of anterior ocular structures. The device is indicated for visualizing and measuring and posterior ocular structures, including corneal epithelium, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head. The CIRRUS normative databases are quantitative tools indicated for the comparison of retinal nerve fiber layer thickness, macular thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of normal subjects.

CIRRUS' AngioPlex OCT Angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid. (Model 5000 only.)

CIRRUS HD-OCT is indicated as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.

Device Description

The CIRRUSTM HD-OCT is a computerized instrument that acquires and analyses crosssectional tomograms of anterior and posterior ocular structures (including cornea, retinal nerve fiber layer, macula, and optic disc). It employs non-invasive, non-contact, low-coherence interferometry to obtain these high-resolution images. Using this non-invasive optical technique, CIRRUS HD-OCT produces high-resolution cross-sectional tomograms of the eye without contacting the eye. It also produces images of the retina and layers of the retina from an en face perspective (i.e., as if looking directly in the eye).

The CIRRUS HD-OCT is offered in two models, Model 5000 and 500. In the CIRRUS HD-OCT Model 5000, the fundus camera is a line scanning ophthalmoscope. The CIRRUS HD-OCT Model 500 is similar to the Model 5000 except that it provides the fundus image using the OCT scanner only.

The acquired imaging data can be analyzed to provide thickness and area measurements of regions of interest to the clinician. The system uses acquired data to determine the fovea location or the optic disc location. Measurements can then be oriented using the fovea and/or optic disc locations. The patient's results can be compared to subjects without disease for measurements of RNFL thickness, neuroretinal rim area, average and vertical cup-to-disc area ratio, cup volume, macular thickness and ganglion cell plus inner plexiform layer thickness.

In addition to macular and optic disc cube scans, the CIRRUS HD-OCT 5000 also offers scans for OCT angiography imaging, a non-invasive approach with depth sectioning capability to visualize microvascular structures of the eye.

Anterior segment scans enable analysis of the anterior segment including Anterior Chamber Depth, Angle-to-Angle and automated measurement of the thickness of the cornea with the Pachymetry scan.

AI/ML Overview

The provided text describes the 510(k) summary for the Carl Zeiss Meditec, Inc. CIRRUS HD-OCT with Software Version 10. The main study detailed is for corneal epithelial thickness measurements.

Here's a breakdown of the acceptance criteria and study information:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the statistical analyses performed, primarily focusing on repeatability, reproducibility, and agreement with manual measurements. The performance is reported in terms of these statistical metrics rather than predefined thresholds for acceptance.

Corneal Epithelial Thickness Measurements (Pachymetry Scans)

MetricAcceptance Criteria (Implied)Reported Device Performance (Normal Eyes - Central Sector)Reported Device Performance (Pathology Eyes - Central Sector)
Repeatability SDLow standard deviation (SD) for repeated measurements.0.8 µm1.4 µm
Repeatability LimitLow limit (2.8 x Repeatability SD).2.2 µm4.0 µm
Repeatability CV%Low coefficient of variation.1.6%3.0%
Reproducibility SDLow standard deviation (SD) across different operators/devices.1.1 µm1.8 µm
Reproducibility LimitLow limit (2.8 x Reproducibility SD).3.2 µm5.1 µm
Reproducibility CV%Low coefficient of variation.2.3%3.8%
Automated vs. Manual Agreement (Deming Regression)Slope close to 1, Intercept close to 0.Slope: 0.88 (95% CI: 0.71, 1.04)Intercept: 4.69 (95% CI: -3.65, 13.02)Slope: 1.03 (95% CI: 0.95, 1.10)Intercept: -2.46 (95% CI: -6.35, 1.43)
Automated vs. Manual Agreement (Bland-Altman Limits of Agreement)Narrow limits of agreement around a mean difference close to 0.Mean Difference: -1.59 µm (SD 1.77)Lower LOA: -5.05 µmUpper LOA: 1.88 µmMean Difference: -1.17 µm (SD 2.98)Lower LOA: -7.00 µmUpper LOA: 4.67 µm

Note: The table provides data for the "Central" sector as an example. The document provides detailed results for 25 different sectors.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Normal Corneas (Group 1): 11 adult participants (one eligible eye per participant).
    • Keratoconus/Post-LASIK (Group 2): 12 participants (one eligible eye per participant).
  • Data Provenance: The document does not explicitly state the country of origin. It is a prospective clinical study specifically conducted to determine repeatability, reproducibility, and agreement.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Three masked graders.
  • Qualifications of Experts: Not explicitly stated beyond "masked graders."

4. Adjudication Method for the Test Set

The document states: "To generate manually marked corneal epithelial thickness measurements, three masked graders reviewed images and manually performed measurements in the 25 sectors." It does not specify an adjudication method like 2+1 or 3+1; it simply mentions that three graders performed the measurements. The comparison is made between the automated measurement and these manual measurements.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. The study focuses on the agreement between the device's automated measurements and manual measurements.

6. Standalone (Algorithm Only) Performance

Yes, a standalone performance assessment was done. The study specifically evaluated the "automated corneal epithelial thickness measurements" generated by the device's software. The comparison to manual measurements serves as a validation of this standalone algorithm's performance against human expert measurements.

7. Type of Ground Truth Used

The ground truth for the corneal epithelial thickness measurements was established by expert manual measurements performed by three masked graders. This serves as a reference standard to which the automated measurements were compared.

8. Sample Size for the Training Set

The document does not provide information about the sample size used for the training set of the algorithm. This study focuses on the clinical evaluation of the device in its final form.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for the training set (if any) was established. The clinical evaluation described pertains to the performance validation of the already developed algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 15, 2019

Carl Zeiss Meditec, Inc. Mandy Ambrecht Senior Staff Regulatory Affairs Specialist 5160 Hacienda Drive Dublin, CA 94568

Re: K181534

Trade/Device Name: Cirrus Hd-oct Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO, IYO, ITX Dated: January 17, 2019 Received: January 18, 2019

Dear Mandy Ambrecht:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours,

Kesia Alexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Sincerely,

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181534

Device Name CIRRUS HD-OCT

Models 500, 5000

Indications for Use (Describe)

CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device intended for invivo viewing, axial cross-sectional, and three-dimensional imaging of anterior ocular structures. The device is indicated for visualizing and measuring and posterior ocular structures, including corneal epithelium, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head. The CIRRUS normative databases are quantitative tools indicated for the comparison of retinal nerve fiber layer thickness, macular thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of normal subjects.

CIRRUS' AngioPlex OCT Angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid. (Model 5000 only.)

CIRRUS HD-OCT is indicated as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY (per 21 CFR §807.92)

CIRRUS HD-OCT with Software Version 10

GENERAL INFORMATION

Manufacturer:Carl Zeiss Meditec, Inc.5160 Hacienda DriveDublin, California 94568(925) 557-4100 (phone)(925) 557-4259 (fax)Est. Reg. No. 2918630
Contact Person:Mandy AmbrechtSenior Staff Regulatory Affairs SpecialistCarl Zeiss Meditec, Inc.5160 Hacienda DriveDublin, California 94568(925) 557-4561 (phone)(925) 557-4259 (fax)
Date Summary Prepared:February 15, 2019
Classification name:Tomography, Optical Coherence; Ophthalmoscope
Classification:Class II (acc. 21 CFR 886.1570)
Product Code:OBO
Trade/Proprietary name:CIRRUS HD-OCT
Common or Usual name:Optical Coherence Tomographer (OCT)
Models:5000 / 500

PREDICATE DEVICE

Company:Carl Zeiss Meditec, Inc.
Primary Predicate Device:CIRRUS HD-OCT with Software Version 8 (K150977)
Company:Carl Zeiss Meditec, Inc.
Device:PLEX Elite 9000 SS-OCT (K161194)

{4}------------------------------------------------

REFERENCE DEVICE

Company:ArcScan, Inc.
Device:ArcScan Insight 100 (K153416)

INDICATIONS FOR USE (21 CFR §807.92(a)(5))

CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional, and three-dimensional imaging of anterior and posterior ocular structures. The device is indicated for visualizing and measuring anterior and posterior ocular structures, including cornea, corneal epithelium, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head. The CIRRUS normative databases are quantitative tools indicated for the comparison of retinal nerve fiber layer thickness, macular thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of normal subjects.

CIRRUS' AngioPlex OCT Angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid. (Model 5000 only.)

CIRRUS HD-OCT is indicated as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.

DEVICE DESCRIPTION (21 CFR §807.92(a)(4))

The CIRRUSTM HD-OCT is a computerized instrument that acquires and analyses crosssectional tomograms of anterior and posterior ocular structures (including cornea, retinal nerve fiber layer, macula, and optic disc). It employs non-invasive, non-contact, low-coherence interferometry to obtain these high-resolution images. Using this non-invasive optical technique, CIRRUS HD-OCT produces high-resolution cross-sectional tomograms of the eye without contacting the eye. It also produces images of the retina and layers of the retina from an en face perspective (i.e., as if looking directly in the eye).

The CIRRUS HD-OCT is offered in two models, Model 5000 and 500. In the CIRRUS HD-OCT Model 5000, the fundus camera is a line scanning ophthalmoscope. The CIRRUS HD-OCT Model 500 is similar to the Model 5000 except that it provides the fundus image using the OCT scanner only.

The acquired imaging data can be analyzed to provide thickness and area measurements of regions of interest to the clinician. The system uses acquired data to determine the fovea location or the optic disc location. Measurements can then be oriented using the fovea and/or optic disc locations. The patient's results can be compared to subjects without disease for measurements of RNFL thickness, neuroretinal rim area, average and vertical cup-to-disc area ratio, cup volume, macular thickness and ganglion cell plus inner plexiform layer thickness.

In addition to macular and optic disc cube scans, the CIRRUS HD-OCT 5000 also offers scans

{5}------------------------------------------------

for OCT angiography imaging, a non-invasive approach with depth sectioning capability to visualize microvascular structures of the eye.

Anterior segment scans enable analysis of the anterior segment including Anterior Chamber Depth, Angle-to-Angle and automated measurement of the thickness of the cornea with the Pachymetry scan.

NEW FEATURES - ANTERIOR SEGMENT

Epithelial Thickness Mapping

The Pachymetry scan analysis was updated to include an automated, quantitative measurement of the corneal epithelium.

NEW FEATURES - POSTERIOR SEGMENT

Guided Progression Analysis (GPA) for Ganglion Cell with Inner Plexiform Layer

GPA was updated to follow changes to the thickness of the sum of the Ganglion Cell Layer + Inner Plexiform Layer (GCL/IPL). It compares Average GCL/IPL measurements from the Macular Cube 200x200 or 512x128 scan over time and determines if change over time has occurred that exceeds the test-retest variability. The analysis includes a chronological display of GCL/IPL thickness maps and thickness change maps, and a graphical plot of the Average GCL/IPL thickness as a function of patient age using the same layout as the graphical plots of RNFL average thickness measurements and Average Cup-to-Disc Ratio versus patient age.

Minimum Intensity (MinIP) Slab

The En Face analysis was expanded to include the Minimum Intensity Projection slab, referred to as "Min-IP". Minimum intensity projection (Min-IP) is an en face visualization of the retina that identifies and displays the minimum optical intensity found between the internal limiting membrane and retinal pigment epithelium.

OCT Angiography Analysis (AngioPlex)

OCT Angiography Analysis was updated to better image retinal vasculature by using both intensity and phase data (referred to as OMAG Complex reconstruction or OMAGS). An additional OCT angiography scan pattern was added to expand scanning from 6x6 to 8x8 mm. In addition, an option allows users the ability to edit laver segmentations for Angiography scans.

OCT Angiography (OCTA) Two Visit Comparison

The OCT Angiography Two Visit Comparison presents a side-by-side comparison of any two visits from a patient's history.

Smart HD Scans previously available for the Model 5000 are now available for the Model 500:

{6}------------------------------------------------

  • . HD 1 Line 100x1 - High-Definition Single Line Scan - This scan generates a single high definition scan at a depth of 2.0 mm by averaging 100 B-scans, each composed of 1024 A-scans. The scan can be positioned anywhere on the fundus image.
  • HD 21 Line: 21 High-Definition Scan This scan generates 21 high-definition . horizontal scan lines at a depth of 2.0 mm. The amount of B-scan averaging per line depends on the version of the instrument computer. The scan can be positioned anywhere on the fundus image.
  • . HD Cross: 10 High-Definition Scan - This scan generates five horizontal and five vertical high-definition scan lines at a depth of 2.0 mm. The scan can be positioned anywhere on the fundus image.
  • HD Radial: 12 High-Definition Radial Scans This scan generates 12 high-. definition radial scan lines at a depth of 2.0. Scan rotation and line spacing are fixed. The scan can be positioned anywhere on the fundus image.

PERFORMANCE DATA & SUMMARY OF VERIFICATION AND VALIDATION ACTIVITY (21 CFR §807.92(B))

Electrical Safety Testing

The CIRRUS HD-OCT was assessed for conformity with the relevant requirements of ANSI/AAMI 60601-1: 2005+ A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and was found to comply.

Electromagnetic Compatibility Testing

The CIRRUS HD-OCT was assessed for conformity with the relevant requirements of IEC 60601-1-2: Ed. 3 / 2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests and was found to comply.

System Level Verification

The system level testing verified that the device performed according to requirements.

DICOM Conformity Assessment

The DICOM Conformance Statement was provided.

Software Verification and Validation

Software documentation was provided in accordance with FDA's software guidance documents. The results of verification and validation testing demonstrate that the software performs in accordance with its established requirements and will therefore meet user needs and intended uses.

{7}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS (21 CFR §807.92(a)(6)) AND SUBSTANTIAL EQUIVALENCE (21 CFR §807.92(a)(3))

It is the opinion of Carl Zeiss Meditec, Inc. that the proposed device, CIRRUS HD-OCT with Software Version 10.0, is substantially equivalent to the CIRRUS HD-OCT with Software Version 8.0 (K150977). Both devices are indicated for in-vivo viewing, axial cross-sectional, and three-dimensional imaging and measurement of anterior ocular structures. Both devices offer CIRRUS OCT angiography which is indicated as an aid in the visualization of vascular structures of the retina and choroid.

The technological characteristics of the proposed product, CIRRUS HD-OCT with Software Version 10.0, and the predicate device, CIRRUS HD-OCT with Software Version 8.0 (K150977), regarding design, material, and energy source are the same. The principle of operation between the proposed CIRRUS HD-OCT with Software Version 10.0 and the primary predicate device, CIRRUS HD-OCT with Software Version 8.0, is identical. Both devices employ a non-invasive, non-contact low-coherence interferometry technique to generate crosssectional images of internal ocular tissue microstructures by measuring optical reflections from tissue. Both devices obtain high-resolution images using spectral domain optical coherence tomography. Both devices provide cross-sectional images of the anterior and posterior structures of the eye (i.e., retina, including the ganglion and retinal nerve fiber layers). Both devices can be used to image structures of the anterior segment by changing the focus of the OCT beam.

With the exception of an external keyboard, mouse and printer, both devices integrate all hardware components into a single unit, which includes the scan acquisition optics, the interferometer and spectrometer, and the system computer and video monitor.

Both devices use a beam of light, called a scanning super luminescent diode (SLD), to rapidly scan the eve in a non-contact manner to acquire very detailed cross-sectional images of various structures. Both devices utilize a central wavelength of 840 nanometers [0.8 microns (um)] primarily designed to scan the back of the eye such as the retina and the optic nerve.

The difference between the predicate device and the proposed device lies within the features offered by the software. In the anterior segment, whereas the predicate device offered automated measurement of the central corneal thickness, the proposed device offers automated measurement of the corneal epithelium thickness.

It is also the opinion of Carl Zeiss Meditec. Inc. that the proposed device. CIRRUS HD-OCT with Software Version 10.0, is substantially equivalent to the predicate device. PLEX Elite 9000 Swept-Source (SS) OCT (K161194). Both devices are intended for in-vivo viewing, axial crosssectional and three-dimensional imaging of posterior ocular structures. Both devices are computerized instruments that acquire cross-sectional tomograms of the posterior ocular structures (including cornea, retinal nerve fiber layer, macula, and optic disc). Both devices employ non-invasive, non-contact, low-coherence interferometry to obtain these highresolution images. Using this non-invasive optical technique, both devices produce highresolution cross-sectional tomograms of the eye without contacting the eye. Both devices produce images of the retina and layers of the retina from an en face perspective (i.e., as if

{8}------------------------------------------------

looking directly in the eye) and non-contrast angiographic imaging of the retinal microvasculature. On both devices, OCT Angiography images are processed to provide detailed images of ocular blood flow without the use of intravenous dyes to visualize microvasculature structures of the eye. The flow data that is generated by processing OCT images with OMAGc has the same resolution and axial and transverse extent as the OCT intensity and phase data, and therefore depth resolved flow images can be generated and displayed.

A difference between the proposed device and the predicate device is that the PLEX Elite 9000 is a Swept Source OCT (SS-OCT) operating at 1060 nm whereas the CIRRUS HD-OCT with Software Version 8.0 is a Spectral Domain OCT (SD-OCT) operating at 840 nm. It is also the opinion of CZMI that the CIRRUS HD-OCT with Software Version 10.0 is substantially equivalent to the reference device, ArcScan Insight 100 (K153416), with regard to the intended use. Both devices are indicated for imaging and measurement of ocular structures in the anterior segment. Both devices measure thickness of the cornea and cornea epithelium. Both devices measure anatomic structures such as the Anterior Chamber Depth and Angle-to-Angle Width.

A difference between the proposed device, and the previously cleared ArcScan Insight 100 is that the reference device uses very high frequency digital ultrasound whereas the CIRRUS HD-OCT uses spectral domain OCT. In addition, with the reference device, scans are performed under immersion whereas they are not with CIRRUS HD-OCT.

SUBSTANTIAL EQUIVALENCE

It is the opinion of Carl Zeiss Meditec, Incorporated that the CIRRUS HD-OCT with Software Version 10 is substantially equivalent to the CIRRUS HD-OCT with Software Version 8.0 (K150977) and PLEX Elite 9000 SS-OCT (K161194) and to the reference device, ArcScan Insight 100 (K153416). The indications for use for the CIRRUS HD-OCT with Software Version 10 is similar to the indications for the predicate devices and reference device cited in this application. A technological comparison and clinical testing demonstrate that the CIRRUS HD-OCT with Software Version 10 is functionally equivalent to the predicate devices.

Evaluation performed on the CIRRUS HD-OCT with Software Version 10 supports the indications for use statement and demonstrates that the device is substantially equivalent to the predicate devices and does not raise new questions regarding safety and effectiveness.

RISK MANAGEMENT

Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software means, user instructions, verification of requirements, and validation of the clinical workflow to ensure that the product meets its intended uses. Carl Zeiss Meditec, Inc. adheres to recognized and established industry practice and relevant international standards where indicated.

{9}------------------------------------------------

CLINICAL EVALUATION (21 CFR 807.92(b)(2))

Clinical data were collected and evaluated to support the indications for use statement for the CIRRUS HD-OCT with Software Version 10 and to demonstrate substantial equivalence to the CIRRUS HD-OCT with Retinal Nerve Fiber Layer (RNFL), Macular, Optic Nerve Head and Ganglion Cell Normative Databases with software version 8.0. The study is summarized below.

Corneal Epithelial Thickness Measurements (Pachymetry Scans)

A clinical study was conducted to determine the repeatability and reproducibility of corneal epithelial thickness measurements on the CIRRUS HD-OCT Model 5000 using the pachymetry scan analysis and to compare agreement between automated and manual measurements.

The study enrolled 11 adult participants with normal corneas (Group 1) and 12 participants (Group 2) with either keratoconus (five participants) or who had previously undergone laser insitu keratomileusis (LASIK; seven participants). To generate automated corneal epithelial thickness measurements, one eligible eve of each participant was scanned three times with the Pachymetry scan on each of three fixed CIRRUS HD-OCT device-operator combinations (Model 5000 with software version 10.0). The order of device-operator scanning was randomized. Scans were reviewed for adequate image quality after acquisition. Repeatability and reproducibility of the 25 sector thickness parameters were calculated using a crossed, randomeffects analysis of variance (ANOVA) model. To generate manually marked corneal epithelial thickness measurements, three masked graders reviewed images and manually performed measurements in the 25 sectors. The automated corneal epithelial measurement obtained from the first qualified scan from each eligible study eye was used for comparison to manual measurements from that study eye. Measurement was evaluated using Deming regression and Bland-Altman analyses.

Repeatability and reproducibility data for epithelial thickness measurements in all sectors is shown in Table 1 for Normal eyes and in Table 2 for eyes post-status LASIK or with keratoconus.

Results for Deming regression fits by Sector and Group are shown in Table 3. Deming regressions were carried out assuming equal variability in Manual and Algorithm measurements.

Results for Bland-Altman Limits of Agreement calculations by Sector and Group are shown in Table 4. For each Sector, and overall, the limits of Agreement were calculated as: mean +/- 1.96 * SD, where "mean" is the mean of the differences between Algorithm and Manual results, and SD is the standard deviation. Confidence intervals for the LOAs were calculated using the approximation from Bland and Altman (1999).

{10}------------------------------------------------

(Normal Eyes)
RepeatabilityReproducibility
ParameterMean(μm)SD(μm)Limit(µm)Mean(µm)SD(μm)Limit(μm)Mean(μm)
Central49.60.82.21.61.13.22.3
Inner Nasal48.91.02.92.11.33.62.6
Inner SuperoNasal47.81.33.72.81.54.13.1
Inner Superior47.21.33.62.71.54.13.1
Inner SuperoTemporal47.11.43.92.91.54.13.1
Inner Temporal47.51.23.52.61.54.13.1
Inner InferoTemporal48.81.33.62.71.44.02.9
Inner Inferior50.31.13.12.21.43.92.8
Inner InferoNasal50.11.02.71.91.23.32.4
Middle Nasal47.31.02.72.01.13.02.3
Middle SuperoNasal46.51.54.33.31.64.53.4
Middle Superior44.81.64.43.51.64.53.6
Middle SuperoTemporal45.11.54.13.21.54.13.2
Middle Temporal45.41.33.72.91.44.03.1
Middle InferoTemporal47.41.13.12.41.23.22.4
Middle Inferior48.91.23.42.51.33.62.7
Middle InferoNasal48.61.13.12.31.23.22.4
Outer Nasal47.11.02.92.21.23.32.5
Outer SuperoNasal45.21.85.04.01.85.04.0
Outer Superior41.91.85.24.41.95.34.5
Outer SuperoTemporal43.81.74.83.91.85.14.1
Outer Temporal44.81.23.32.61.33.62.8
Outer InferoTemporal47.31.43.93.01.64.53.4
Outer Inferior48.01.43.92.91.54.13.1
Outer InferoNasal47.61.44.03.01.54.13.1

Table 1 Repeatability and Reproducibility of OCT Epithelial Thickness

All statistics are estimated from two-way random-effect ANOVA model with random effects operator/device, eye and interaction between operator/device and eye.

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance.

Reproducibility SD = Square root of the sum of the operator/device variance, the interaction variance and the residual variance.

Repeatability limit = 2.8 x Repeatability SD.

Reproducibility limit = 2.8 x Reproducibility SD.

Repeatability CV% = (Repeatability SD)/Mean x 100%.

Reproducibility CV% = (Reproducibility SD)/Mean x 100%.

{11}------------------------------------------------

(Keratoconus or Post-LASIK Eyes)
RepeatabilityReproducibility
ParameterMean(μm)SD(μm)Limit(μm)CV(%)SD(μm)Limit(μm)CV(%)
Central47.71.44.03.01.85.13.8
Inner Nasal49.11.64.43.22.15.94.3
Inner SuperoNasal48.61.54.33.11.95.44.0
Inner Superior47.71.74.83.62.15.94.5
Inner SuperoTemporal46.71.85.23.92.36.44.9
Inner Temporal46.51.95.44.22.36.44.9
Inner InferoTemporal47.32.15.84.42.57.05.3
Inner Inferior48.21.95.44.02.67.25.4
Inner InferoNasal49.01.85.03.62.36.44.7
Middle Nasal47.41.74.73.62.67.25.4
Middle SuperoNasal46.51.64.63.52.36.44.9
Middle Superior44.91.74.93.92.15.94.7
Middle SuperoTemporal44.81.95.44.32.46.75.4
Middle Temporal45.82.56.95.42.67.45.8
Middle InferoTemporal46.62.77.45.73.28.96.8
Middle Inferior46.62.46.85.23.610.07.7
Middle InferoNasal47.72.05.74.33.08.36.2
Outer Nasal46.91.85.13.92.87.85.9
Outer SuperoNasal45.81.84.93.82.26.24.9
Outer Superior42.92.05.64.72.05.74.7
Outer SuperoTemporal41.91.74.84.12.26.35.3
Outer Temporal43.82.87.96.43.08.46.8
Outer InferoTemporal46.03.08.56.63.59.97.7
Outer Inferior46.02.67.25.64.713.210.3
Outer InferoNasal46.52.46.75.13.39.47.2

Table 2 Repeatability and Reproducibility of OCT Epithelial Thickness (Keratoconus or Post-LASIK Ev

All statistics are estimated from two-way random-effect ANOVA model with random effects operator/device, eye and interaction between operator/device and eye.

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance.

Reproducibility SD = Square root of the sum of the operator/device variance, the interaction variance and the residual variance.

Repeatability limit = 2.8 x Repeatability SD.

Reproducibility limit = 2.8 x Reproducibility SD.

Repeatability CV% = (Repeatability SD)/Mean x 100%.

Reproducibility CV% = (Reproducibility SD)/Mean x 100%.

{12}------------------------------------------------

Table 3 Results of Deming Regression Analysis Stratified by Sector and Study Group

SectorGroupIntercept (SE)μmInt [95% CI]μmSlope (SE)no unitsSlope [95% CI]no units
Sector_01Normal4.69 (4.25)[-3.65, 13.02]0.88 (0.08)[0.71, 1.04]
Pathology-2.46 (1.98)[-6.35, 1.43]1.03 (0.04)[0.95, 1.10]
Sector_02Normal-9.71 (3.47)[-16.52, -2.90]1.16 (0.07)[1.02, 1.29]
Pathology1.09 (1.16)[-1.18, 3.35]0.94 (0.02)[0.90, 0.99]
Sector_03Normal-3.13 (2.95)[-8.91, 2.65]1.02 (0.06)[0.91, 1.13]
Pathology-2.87 (1.45)[-5.70, -0.03]1.01 (0.03)[0.95, 1.06]
Sector_04Normal-11.09 (4.48)[-19.87, -2.31]1.18 (0.09)[1.01, 1.35]
Pathology-1.52 (1.28)[-4.02, 0.99]0.99 (0.02)[0.94, 1.04]
Sector_05Normal-5.11 (3.33)[-11.64, 1.42]1.05 (0.06)[0.93, 1.18]
Pathology-2.91 (1.21)[-5.28, -0.55]1.01 (0.02)[0.97, 1.06]
Sector_06Normal2.65 (3.00)[-3.23, 8.54]0.90 (0.06)[0.79, 1.02]
Pathology-2.26 (1.71)[-5.61, 1.10]1.00 (0.03)[0.94, 1.07]
Sector_07Normal0.18 (3.23)[-6.15, 6.51]0.96 (0.06)[0.84, 1.09]
Pathology-0.27 (1.03)[-2.30, 1.76]0.96 (0.02)[0.92, 1.00]
Sector_08Normal3.79 (3.35)[-2.78, 10.35]0.89 (0.07)[0.76, 1.02]
Pathology-16.11 (3.49)[-22.95, -9.26]1.30 (0.07)[1.16, 1.43]
Sector_09Normal-1.51 (3.04)[-7.47, 4.45]0.98 (0.06)[0.87, 1.10]
Pathology-7.65 (2.67)[-12.89, -2.41]1.12 (0.05)[1.02, 1.23]
Sector_10Normal1.34 (3.77)[-6.06, 8.73]0.92 (0.08)[0.77, 1.08]
Pathology-6.39 (1.53)[-9.38, -3.40]1.07 (0.03)[1.01, 1.13]
Sector_11Normal-27.06 (5.06)[-36.98, -17.13]1.47 (0.10)[1.27, 1.66]
Pathology-4.73 (1.60)[-7.86, -1.60]1.05 (0.03)[0.98, 1.11]
Sector_12Normal-12.70 (4.37)[-21.26, -4.14]1.19 (0.08)[1.03, 1.36]
Pathology-1.75 (2.00)[-5.68, 2.18]0.98 (0.04)[0.90, 1.06]
Sector_13Normal-11.56 (4.80)[-20.96, -2.15]1.17 (0.09)[0.99, 1.36]
Pathology1.11 (1.12)[-1.08, 3.31]0.93 (0.02)[0.88, 0.97]
Sector_14Normal-6.34 (3.56)[-13.31, 0.64]1.08 (0.07)[0.94, 1.22]
Pathology-8.46 (1.92)[-12.21, -4.70]1.11 (0.04)[1.04, 1.18]
Sector_15Normal3.42 (3.57)[-3.57, 10.41]0.87 (0.07)[0.73, 1.02]
Pathology-0.60 (1.10)[-2.75, 1.55]0.96 (0.02)[0.92, 1.00]
Sector_16Normal-3.43 (2.88)[-9.08, 2.22]1.02 (0.06)[0.91, 1.14]
Pathology-9.25 (2.99)[-15.10, -3.39]1.14 (0.06)[1.02, 1.26]
Sector_17Normal-2.52 (3.10)[-8.60, 3.55]1.00 (0.06)[0.88, 1.12]
Pathology-8.33 (2.67)[-13.57, -3.09]1.12 (0.05)[1.02, 1.22]
Sector_18Normal-12.47 (3.22)[-18.79, -6.15]1.22 (0.07)[1.09, 1.35]
Intercept (SE)Int [95% CI]Slope (SE)Slope [95% CI]
SectorGroupμmμmno unitsno units
Sector_19Pathology4.44 (1.43)[1.65, 7.24]0.86 (0.03)[0.81, 0.92]
Normal3.18 (4.83)[-6.30, 12.65]0.90 (0.10)[0.71, 1.09]
Sector_20Pathology-2.91 (3.80)[-10.37, 4.54]1.02 (0.08)[0.87, 1.17]
Normal-5.80 (4.60)[-14.82, 3.22]1.07 (0.09)[0.89, 1.25]
Sector_21Pathology-1.25 (3.07)[-7.27, 4.77]0.98 (0.06)[0.86, 1.11]
Normal-10.72 (7.16)[-24.75, 3.31]1.17 (0.14)[0.88, 1.45]
Sector_22Pathology-1.05 (2.20)[-5.37, 3.26]1.00 (0.05)[0.91, 1.09]
Normal-1.92 (2.59)[-6.99, 3.15]1.00 (0.05)[0.90, 1.10]
Sector_23Pathology-2.11 (2.38)[-6.76, 2.55]1.00 (0.05)[0.90, 1.09]
Normal-6.47 (3.10)[-12.55, -0.39]1.07 (0.07)[0.95, 1.20]
Sector_24Pathology3.59 (1.32)[1.00, 6.18]0.87 (0.03)[0.81, 0.93]
Normal-5.77 (2.51)[-10.70, -0.85]1.06 (0.06)[0.95, 1.17]
Sector_25Pathology-0.27 (1.86)[-3.91, 3.37]0.96 (0.04)[0.88, 1.04]
Normal-4.91 (3.25)[-11.27, 1.45]1.07 (0.07)[0.94, 1.20]
All CombinedPathology1.91 (1.65)[-1.32, 5.14]0.92 (0.03)[0.85, 0.99]
All CombinedNormal-5.92 (0.61)[-7.11, -4.73]1.07 (0.01)[1.05, 1.10]
All CombinedPathology-2.54 (0.36)[-3.25, -1.83]1.01 (0.01)[0.99, 1.02]
All CombinedAll Combined-3.25 (0.30)[-3.83, -2.66]1.02 (0.01)[1.01, 1.03]

{13}------------------------------------------------

{14}------------------------------------------------

Table 4
Bland Altman Limits of Agreement stratified by Sector and Study Group
MeanDifference(SD)μηDifference[min,max]μηLimit ofAgreement (lower)Estimate (SE)[95% CI]μηLimit ofAgreement (upper)Estimate (SE)[95% CI]μη
SectorGroup
Sector_01Normal-1.59 (1.77)[-7.4, 2.8]-5.05 (0.24),[-5.5, -4.6]1.88 (0.24),[1.4, 2.3]
Pathology-1.17 (2.98)[-7.8, 14.7]-7.00 (0.38),[-7.8, -6.3]4.67 (0.38),[3.9, 5.4]
Sector_02Normal-1.83 (1.62)[-5.9, 2.2]-5.00 (0.22),[-5.4, -4.6]1.33 (0.22),[0.9, 1.8]
Pathology-1.86 (2.15)[-6.9, 11.6]-6.08 (0.27),[-6.6, -5.5]2.36 (0.27),[1.8, 2.9]
Sector_03Normal-1.94 (1.71)[-6.5, 1.9]-5.29 (0.23),[-5.7, -4.8]1.41 (0.23),[1.0, 1.9]
Pathology-2.48 (2.23)[-8.1, 5.2]-6.84 (0.28),[-7.4, -6.3]1.89 (0.28),[1.3, 2.4]
Sector_04Normal-1.70 (1.94)[-7.3, 2.3]-5.50 (0.26),[-6.0, -5.0]2.09 (0.26),[1.6, 2.6]
Pathology-1.93 (1.96)[-8.2, 2.0]-5.77 (0.25),[-6.3, -5.3]1.90 (0.25),[1.4, 2.4]
Sector_05Normal-2.49 (1.79)[-7.1, 1.4]-6.00 (0.24),[-6.5, -5.5]1.03 (0.24),[0.6, 1.5]
Pathology-2.24 (1.83)[-7.5, 2.9]-5.83 (0.23),[-6.3, -5.4]1.34 (0.23),[0.9, 1.8]
Sector_06Normal-2.45 (1.81)[-7.7, 2.0]-6.00 (0.24),[-6.5, -5.5]1.11 (0.24),[0.6, 1.6]
Pathology-2.04 (2.21)[-7.1, 6.4]-6.37 (0.28),[-6.9, -5.8]2.30 (0.28),[1.7, 2.9]
Sector_07Normal-1.64 (1.75)[-6.3, 3.1]-5.07 (0.23),[-5.5, -4.6]1.79 (0.23),[1.3, 2.3]
Pathology-2.46 (1.78)[-6.9, 2.5]-5.94 (0.23),[-6.4, -5.5]1.03 (0.23),[0.6, 1.5]
Sector_08Normal-1.80 (1.81)[-6.3, 1.8]-5.34 (0.24),[-5.8, -4.9]1.75 (0.24),[1.3, 2.2]
Pathology-1.22 (3.55)[-6.4, 17.1]-8.19 (0.45),[-9.1, -7.3]5.75 (0.45),[4.9, 6.6]
Sector_09Normal-2.29 (1.80)[-7.8, 1.8]-5.81 (0.24),[-6.3, -5.3]1.23 (0.24),[0.8, 1.7]
Pathology-1.50 (3.40)[-6.9, 14.7]-8.16 (0.43),[-9.0, -7.3]5.16 (0.43),[4.3, 6.0]
MeanDifference(SD)μmDifference[min,max]μmLimit ofAgreement (lower)Estimate (SE)[95% CI]μmLimit ofAgreement (upper)Estimate (SE)[95% CI]μm
SectorGroup
Sector_10Normal-2.38 (2.28)[-9.8, 2.7]-6.86 (0.31),[-7.5, -6.3]2.09 (0.31),[1.5, 2.7]
Pathology-2.94 (2.36)[-9.4, 3.6]-7.58 (0.31),[-8.2, -7.0]1.69 (0.31),[1.1, 2.3]
Sector_11Normal-3.10 (2.80)[-15.7, 3.8]-8.58 (0.39),[-9.3, -7.8]2.38 (0.39),[1.6, 3.1]
Pathology-2.42 (2.38)[-9.1, 6.7]-7.08 (0.32),[-7.7, -6.4]2.24 (0.32),[1.6, 2.9]
Sector_12Normal-2.73 (2.59)[-10.1, 4.0]-7.80 (0.35),[-8.5, -7.1]2.34 (0.35),[1.7, 3.0]
Pathology-2.72 (2.61)[-9.2, 8.2]-7.84 (0.36),[-8.6, -7.1]2.40 (0.36),[1.7, 3.1]
Sector_13Normal-2.70 (2.53)[-12.3, 4.2]-7.66 (0.34),[-8.3, -7.0]2.25 (0.34),[1.6, 2.9]
Pathology-2.65 (2.01)[-7.8, 1.8]-6.59 (0.27),[-7.1, -6.1]1.30 (0.27),[0.8, 1.8]
Sector_14Normal-2.52 (2.54)[-7.9, 3.0]-7.50 (0.34),[-8.2, -6.8]2.46 (0.34),[1.8, 3.1]
Pathology-2.92 (2.68)[-11.1, 7.2]-8.17 (0.34),[-8.8, -7.5]2.34 (0.34),[1.7, 3.0]
Sector_15Normal-2.70 (2.33)[-8.8, 2.6]-7.26 (0.31),[-7.9, -6.6]1.86 (0.31),[1.2, 2.5]
Pathology-2.50 (2.20)[-8.7, 3.3]-6.81 (0.28),[-7.4, -6.3]1.81 (0.28),[1.3, 2.4]
Sector_16Normal-2.26 (1.92)[-8.3, 1.4]-6.02 (0.26),[-6.5, -5.5]1.50 (0.26),[1.0, 2.0]
Pathology-2.65 (3.76)[-8.9, 14.0]-10.02 (0.48),[-11.0, -9.1]4.72 (0.48),[3.8, 5.7]
Sector_17Normal-2.53 (2.34)[-9.4, 5.8]-7.12 (0.31),[-7.7, -6.5]2.06 (0.31),[1.4, 2.7]
Pathology-2.35 (3.59)[-15.5,14.7]-9.39 (0.46),[-10.3, -8.5]4.69 (0.46),[3.8, 5.6]
Sector_18Normal-2.10 (2.29)[-8.3, 4.6]-6.58 (0.33),[-7.2, -5.9]2.39 (0.33),[1.7, 3.1]
Pathology-2.09 (2.72)[-16.7, 4.1]-7.42 (0.40),[-8.2, -6.6]3.24 (0.40),[2.4, 4.0]
Sector_19Normal-1.87 (2.01)[-6.1, 3.5]-5.82 (0.37),[-6.5, -5.1]2.08 (0.37),[1.3, 2.8]
MeanDifference(SD)Difference[min,max]Limit ofAgreement (lower)Estimate (SE)[95% CI]Limit ofAgreement (upper)Estimate (SE)[95% CI]
SectorGroup$\mu m$$\mu m$$\mu m$$\mu m$
Sector_20Pathology-2.00 (3.86)[-18.0, 5.7]-9.56 (0.66),[-10.9, -8.3]5.56 (0.66),[4.3, 6.9]
Normal-2.31 (2.59)[-9.0, 4.0]-7.38 (0.45),[-8.3, -6.5]2.77 (0.45),[1.9, 3.7]
Sector_21Pathology-2.01 (3.45)[-15.5, 7.0]-8.77 (0.65),[-10.1, -7.5]4.74 (0.65),[3.5, 6.0]
Normal-2.55 (2.84)[-13.6, 5.9]-8.12 (0.44),[-9.0, -7.3]3.02 (0.44),[2.2, 3.9]
Sector_22Pathology-1.10 (2.74)[-7.0, 5.5]-6.47 (0.47),[-7.4, -5.5]4.27 (0.47),[3.3, 5.2]
Normal-1.91 (2.28)[-8.2, 3.8]-6.38 (0.34),[-7.0, -5.7]2.56 (0.34),[1.9, 3.2]
Sector_23Pathology-2.26 (3.68)[-14.1, 8.3]-9.48 (0.53),[-10.5, -8.4]4.96 (0.53),[3.9, 6.0]
Normal-2.93 (2.22)[-10.0, 3.9]-7.27 (0.31),[-7.9, -6.7]1.41 (0.31),[0.8, 2.0]
Sector_24Pathology-2.19 (2.35)[-7.4, 4.2]-6.80 (0.32),[-7.4, -6.2]2.41 (0.32),[1.8, 3.0]
Normal-3.04 (2.20)[-9.3, 2.7]-7.34 (0.34),[-8.0, -6.7]1.26 (0.34),[0.6, 1.9]
Sector_25Pathology-2.24 (2.38)[-8.3, 8.4]-6.91 (0.34),[-7.6, -6.2]2.44 (0.34),[1.8, 3.1]
Normal-1.67 (2.36)[-7.6, 4.5]-6.30 (0.35),[-7.0, -5.6]2.96 (0.35),[2.3, 3.6]
AllCombinedPathology-1.74 (2.61)[-9.6, 4.3]-6.87 (0.36),[-7.6, -6.2]3.38 (0.36),[2.7, 4.1]
Normal-2.28 (2.21)[-15.7, 5.9]-6.62 (0.06),[-6.7, -6.5]2.06 (0.06),[1.9, 2.2]
AllCombinedPathology-2.17 (2.77)[-18.0,17.1]-7.60 (0.07),[-7.7, -7.4]3.26 (0.07),[3.1, 3.4]
AllCombinedAllCombined-2.22 (2.52)[-18.0,17.1]-7.15 (0.05),[-7.2, -7.1]2.71 (0.05),[2.6, 2.8]

{15}------------------------------------------------

{16}------------------------------------------------

SUMMARY (21 CFR §807.92(b)(3))

As described in this 510(k) Summary, all testing deemed necessary was conducted on the CIRRUS HD-OCT with Software Version 10 to ensure that the device is substantially equivalent to the predicate device for its intended use when used in accordance with its Instructions for Use.

§ 886.1570 Ophthalmoscope.

(a)
Identification. An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.(b)
Classification. Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.