K Number
K231585

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-06-30

(30 days)

Product Code
Regulation Number
888.3110
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vantage PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intra-operatively, and in guiding bone cutting. The Vantage PSI System is intended for use with Exactech's Vantage Total Ankle System and its cleared indications for use.

Device Description

3D System's Vantage PSI System is patient-specific guides created to fit the contours of the patient's distal tibial and proximal talar anatomy. The guides and models are designed and manufactured from patient imaging data (CT) and are made from biocompatible nylon. The surgical guides in combination with Exacted Vantage Total Ankle reusable instruments, facilitate the positioning of Vantage Total Ankle Implants. 3D System s Vantage PSI System produces a variety of patient specific outputs including surgical quides, anatomic models, and case reports.

AI/ML Overview

Here's an analysis of the acceptance criteria and supporting study for the Vantage PSI System, based on the provided FDA 510(k) summary:

  • This device is a patient-specific surgical guide system, not an AI device. Therefore, many of the typical AI/ML-specific questions regarding ground truth, expert consensus, and multi-reader studies are not applicable in the way they would be for an AI diagnostic or prognostic tool. The "performance" here relates to its accuracy in guiding surgical procedures compared to existing instrumentation.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Functional EquivalenceAbility to assist in positioning total ankle replacement components intra-operatively.Shown to be substantially equivalent to Exactech instrumentation.
Ability to guide bone cutting.Shown to be substantially equivalent to Exactech instrumentation.
AccuracyAccuracy in guiding surgical procedures.Demonstrated sufficient accuracy.
Manufacturing MaterialGuides and models made from biocompatible nylon.Manufactured using SLS technology with DuraForm® ProX PA (polyamide) material. (This meets the biocompatibility implied by use in surgery).
Device CompatibilityIntended for use with Exactech's Vantage Total Ankle System.Explicitly stated for use with Exactech's Vantage Total Ankle System.
Design Revisions (Impact)Additional fixation options, corner drill features, optional decoupled talus guide do not introduce technological differences.The changes "are not technologically different from the predicate Vantage PSI System."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated as a number of "patients" or "cases." The testing was described as "Cadaveric comparison testing." This implies a limited number of cadavers were used.
  • Data Provenance: Cadaveric. Specific country of origin is not mentioned. It is a prospective study in the sense that the test was conducted specifically for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This is not applicable in the traditional sense for this device. The "ground truth" for a surgical guide system is its ability to accurately direct cuts and component placement. The comparison was to an existing instrumentation system (Exactech instrumentation), which serves as the benchmark. The "experts" would be the surgeons or engineers performing the cadaveric tests and assessing the outcomes. Their qualifications are not specified in this summary.

4. Adjudication Method for the Test Set

  • Not explicitly stated. Given it's a comparison of mechanical accuracy and functionality on cadavers, it's likely measurements were taken and compared directly, rather than requiring a complex expert adjudication process for image interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • This is not applicable as the Vantage PSI System is a physical surgical guide system, not an AI/ML diagnostic or prognostic tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • This is not applicable as the Vantage PSI System is a physical surgical guide intended for human use in surgery. Its "performance" is inherently human-in-the-loop.

7. The Type of Ground Truth Used

  • The "ground truth" was established by comparing the subject device's guidance and resulting anatomical modifications (e.g., bone cuts, component placement) to those produced by Exactech's established instrumentation system. This comparison was done on cadavers, implying direct measurement or visual assessment of the outcome of the guided procedure.

8. The Sample Size for the Training Set

  • This is not applicable. The Vantage PSI System is a custom-manufactured guide based on a patient's individual CT scan. There isn't a "training set" in the machine learning sense for the device itself. The design principles and manufacturing process might be informed by historical data or engineering principles, but not a "training set" for an algorithm.

9. How the Ground Truth for the Training Set Was Established

  • This is not applicable for the same reasons as #8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 30, 2023

3D Systems, Inc. Ashley Dawson Director, Regulatory Affairs 5381 South Alkire Circle Littleton, Colorado 80127

Re: K231585

Trade/Device Name: Vantage PSI System Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: OYK Dated: May 31, 2023 Received: May 31, 2023

Dear Ashley Dawson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the name "Lixin Liu -S" in a simple, sans-serif font. The text is black and stands out against a light background. The name appears to be a title or heading, possibly indicating a person's name or a project title. The overall impression is clean and straightforward.

Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K231585

Device Name

Vantage PSI System

Indications for Use (Describe)

The Vantage PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intra-operatively, and in guiding bone cutting. The Vantage PSI System is intended for use with Exactech's Vantage Total Ankle System and its cleared indications for use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Contact Details

Prepared on: 2023-05-31 21 CFR 807.92(a)(1)

Applicant Name3D Systems, Inc.
Applicant Address5381 South Alkire Circle Littleton CO 80127 United States
Applicant Contact Telephone+1 803-326-3908
Applicant ContactDr. Ashley Dawson
Applicant Contact Emailashley.dawson@3dsystems.com

Device Name 21 CFR 807.92(a)(2)

Device Trade NameVantage PSI System
Common NameAnkle joint metal/polymer semi-constrained cemented prosthesis
Classification NameAnkle Arthroplasty Implantation System
Regulation Number888.3110
Product CodeOYK

Legally Marketed Predicate Devices 21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K193432Vantage PSI SystemOYK

Device Description Summary 21 CFR 807.92(a)(4)510(k) Summarv

3D System's Vantage PSI System is patient-specific guides created to fit the contours of the patient's distal tibial and proximal talar anatomy. The guides and models are designed and manufactured from patient imaging data (CT) and are made from biocompatible nylon. The surgical guides in combination with Exacted Vantage Total Ankle reusable instruments, facilitate the positioning of Vantage Total Ankle Implants. 3D System s Vantage PSI System produces a variety of patient specific outputs including surgical quides, anatomic models, and case reports.

Intended Use/Indications for Use

The Vantage PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intra-operatively, and in guiding bone cutting. The Vantage PSI System is intended for use with Exactech's Vantage Total Ankle System and its cleared indications for use.

Indications for Use Comparison

The indications for use is the same as the predicate device: The Vantage to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement intra-operatively, and in guiting bone cutting. The Vantage PSI System is intended for use with Exacted 's Vantage Total Ankle System and its cleared indications for use.

Technological Comparison

The principles of operation and technological characteristics are substantially equivalent between the subject Vantage PSI System and the predicate Vantage PSI System (K193432). Both the predicate and the subject devices are designed with CT-based methods to produce a patient specific instrument. Vantage PSI System Guides are manufactured using SLS technology with DuraForm®

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

{4}------------------------------------------------

K231585 Page 2 of 2

ProX PA (polyamide) material. Both the subject and predicate devices guide the placement of pins and talar anatomies for total ankle arthroplasties intended for use with Exactech's Vantage Total Ankle System and its cleared indications for use. The differences between the subject Vantage PSI System and the predicate Vantage PSI System are additional fixation options, addition of corner drill features, and an optional decoupled talus guide. The reason for the changes is to align with updates to the Exactech Vantage Total Ankle System. The changes included in the subject Vantage PSI System are not technologically different from the predicate Vantage PSI System.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Cadaveric comparison testing was performed between the subject and Exactech instrumentation. Accuracy and functionality were shown to be substantially equivalent across all devices and instruments.

§ 888.3110 Ankle joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.