(197 days)
StrokeViewer Perfusion is an image processing software package intended to provide quantitative perfusion information in brain tissue for suspected ischemic stroke patients. It is to be used by medical imaging professionals who analyze dynamic perfusion studies, including but not limited to physicians such as neurologists, and radiologists.
The software is packed as a Docker container allowing installation on a standard "off-the-shelf" computer or a virtual platform. The software can be used to perform image viewing, processing, and analysis of brain CT perfusion images (CTP) images. Data and images are acquired through DICOM compliant imaging devices.
StrokeViewer Perfusion is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time, which are visualized as colored perfusion maps including flow-related parameters and tissue blood volume quantification.
Contraindications:
- Bolus Quality: absent or inadequate bolus.
- Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. .
- Presence of hemorrhage .
StrokeViewer Perfusion is an image processing application that runs on a standard "off-the-shelf" computer or a virtual platform, and can be used to perform image processing and analysis of CT perfusion images of the brain.
The software can receive, identify and extract DICOM information embedded in the imaging data. The output includes parametric maps related to tissue blood flow (perfusion) and tissue blood volume. Results of the analysis are exported as DICOM series and DICOM reports and can be sent to a preconfigured destination and can be reviewed on a compatible DICOM viewer Perfusion image analysis includes calculation of the following perfusion related parameters:
- . Cerebral Blood Flow (CBF)
- Cerebral Blood Volume (CBV)
- Mean Transit Time (MTT)
- Residue function time-to-peak (TMax) .
- . Arterial Input Function (AIF)
- Volume calculations of affected tissue based on Tmax and CBF abnormalities
The device description and overall fundamental scientific technology of the StrokeViewer Perfusion algorithm is equivalent to the predicate device (K182130) in addition to the listed reference devices within this submission.
Here's a summary of the acceptance criteria and study details for StrokeViewer Perfusion, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The available document does not provide a specific table of quantitative acceptance criteria and corresponding reported device performance values. It broadly states that "The results of algorithm performance testing showed that the device met performance goals and acceptance criteria."
However, it does describe the type of testing performed:
| Acceptance Criterion (Implicit) | Reported Device Performance (General) |
|---|---|
| Correlation with ground truth values on simulated datasets. | "The results of algorithm performance testing showed that the device met performance goals and acceptance criteria." |
| Functional performance and meeting design requirements. | "The results of software verification, and algorithmic testing demonstrate that StrokeViewer Perfusion meets all design requirements and specifications for its intended use." |
| Software validation (Unit, End-to-end, Reproducibility). | "The tests performed demonstrate that Perfusion in StrokeViewer performs as intended. All testing included within the respective sections of this submission uphold our belief that StrokeViewer is substantially equivalent to the predicate device (K182130)." |
| Adherence to risk management and consensus standards. | "All risk controls were identified, implemented, and mitigated according to hazards identified. All testing related to risk controls support the acceptance criteria outlined for our software requirement specifications." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document mentions "simulated datasets (Kudo digital phantom)" for non-clinical bench performance testing. It does not specify the exact number of simulated cases or provide a specific sample size for a "test set" in the context of clinical image data.
- Data Provenance: The testing was conducted using "simulated datasets (Kudo digital phantom) generated using simulating tracer kinetic theory." This implies the data is synthetic/simulated, not derived from real patients. Therefore, information regarding "country of origin" or "retrospective/prospective" is not applicable in this context.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Since the testing was performed on "simulated datasets (Kudo digital phantom)" where "ground truth values were calculated" based on tracer kinetic theory, the ground truth was established by the design of the simulation, not by human experts interpreting medical images. Therefore, no human experts were involved in establishing the ground truth for this specific non-clinical test set.
4. Adjudication Method for the Test Set
Not applicable, as ground truth was established by simulation design/calculation, not human interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was it done? No. The document explicitly states: "Clinical data was not required for this submission of the StrokeViewer Perfusion. We believe the safety and effectiveness of the proposed device was appropriately tested with non-clinical validation."
- Effect Size of Human Readers Improvement: Not applicable, as no MRMC study was conducted.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, the primary performance evaluation described is a standalone algorithmic assessment using simulated data: "Non-clinical bench performance testing was performed to assess Perfusion algorithm performance on simulated datasets (Kudo digital phantom) generated using simulating tracer kinetic theory. Correlations between the output of the StrokeViewer Perfusion device and the ground truth values were calculated."
7. Type of Ground Truth Used
The ground truth used for the non-clinical bench testing was derived from simulated tracer kinetic theory / calculated ground truth values within the Kudo digital phantom.
8. Sample Size for the Training Set
The document does not explicitly state the sample size of the training set used for the StrokeViewer Perfusion algorithm. The focus of this 510(k) summary is on the validation testing rather than the development or training process.
9. How the Ground Truth for the Training Set Was Established
The document does not provide details on how the ground truth for the training set was established. Given the nature of the validation testing using simulated data, it's plausible that similar simulated or engineered data with known ground truths were used for training, but this is not stated.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
NICo-Lab B.V. % Veronica Padharia Principal Regulatory Affairs Specialist Paascheuvelweg 25 Amsterdam, 1105BP Netherlands
Re: K231570
December 14, 2023
Trade/Device Name: StrokeViewer Perfusion Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: November 21, 2023 Received: November 21, 2023
Dear Veronica Padharia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K231570
Device Name StrokeViewer Perfusion
Indications for Use (Describe)
StrokeViewer Perfusion is an image processing software package intended to provide quantitative perfusion information in brain tissue for suspected ischemic stroke patients. It is to be used by medical imaging professionals who analyze dynamic perfusion studies, including but not limited to physicians such as neurologists, and radiologists.
The software is packed as a Docker container allowing installation on a standard "off-the-shelf" computer or a virtual platform. The software can be used to perform image viewing, processing, and analysis of brain CTP) images. Data and images are acquired through DICOM compliant imaging devices.
Stroke Viewer Perfusion is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time, which are visualized as colored perfusion maps including flow-related parameters and tissue blood volume quantification.
Contraindications:
- Bolus Quality: absent or inadequate bolus.
- Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate
- Presence of hemorrhage
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K231570
NICOLAB
510(k) Summary StrokeViewer Perfusion
[As required by section 21 CFR 807.92(c)]
NIC-230142-06-F
13-Dec-2023
{4}------------------------------------------------
NICOLAB
| Contents | |
|---|---|
| Contents | |
|---|---|
| 1 Identification of the Submitter | 3 |
| 2 Device Name and Classification | 3 |
| 3 Device Description | 3 |
| 4 Intended Purpose | 3 |
| 4.1 Indications for Use | 4 |
| 5 Comparison of Indications for Use Statements | 4 |
| 6 Comparison of Technological Characteristics with the Predicate Device | 4 |
| 7 Performance and safety | 5 |
| 7.1 Software Validation and Verification | 5 |
| 7.2 Performance testing | 5 |
| 7.3 Risk Management File | 5 |
| 7.4 Recognized Consensus Standards | 5 |
| 8 Conclusion | 6 |
{5}------------------------------------------------
K231570
Identification of the Submitter 1
| Importer/Distributor | NICo-Lab B.V. (Nicolab) | ||
|---|---|---|---|
| Establishment Registration Number | 3021915828 | ||
| Submitter Contact Information | Primary Contact | Secondary Contact | |
| Name | Veronica Padharia | Jorrit Boersen | |
| Title | Principal RA Specialist | RA Specialist | |
| vpadharia@nicolab.com | jboersen@nicolab.com | ||
| Phone Number | 630-917-1160 | +31(0)6 341 88 966 |
Device Name and Classification 2
| Subject Device | Predicate Device(K182130) | Reference Device (K200873) | |
|---|---|---|---|
| Device Name | StrokeViewer Perfusion | iSchemaView Rapid | StrokeViewer HALO |
| Proprietary TradeName | StrokeViewer Perfusion | Rapid | HALO (LVO Triaging) |
| Product Classification | LLZ | LLZ | QAS |
| Classification Name | System, ImageProcessing, Radiological | System, ImageProcessing,Radiological | Radiological Computer AidedTriage and Notification Software |
| Classification Panel | Radiology | Radiology | Radiology |
| CFR Section | 21 CFR §892.2050 | 21 CFR §892.2050 | 21 CFR §892.2080 |
| Device Class | Class II | Class II | Class II |
റ് Device Description
StrokeViewer Perfusion is an image processing application that runs on a standard "off-the-shelf" computer or a virtual platform, and can be used to perform image processing and analysis of CT perfusion images of the brain.
The software can receive, identify and extract DICOM information embedded in the imaging data. The output includes parametric maps related to tissue blood flow (perfusion) and tissue blood volume. Results of the analysis are exported as DICOM series and DICOM reports and can be sent to a preconfigured destination and can be reviewed on a compatible DICOM viewer Perfusion image analysis includes calculation of the following perfusion related parameters:
- . Cerebral Blood Flow (CBF)
- Cerebral Blood Volume (CBV)
- Mean Transit Time (MTT)
- Residue function time-to-peak (TMax) .
- . Arterial Input Function (AIF)
- Volume calculations of affected tissue based on Tmax and CBF abnormalities
The device description and overall fundamental scientific technology of the StrokeViewer Perfusion algorithm is equivalent to the predicate device (K182130) in addition to the listed reference devices within this submission.
Intended Purpose ব
StrokeViewer Perfusion is an image processing software application that analyzes CT scans of the brain to provide quantitative perfusion in brain tissue for suspected ischemic stroke patients. Additionally, it performs calculations to quantify areas of perfusion in the brain.
StrokeViewer provides this information to the user to include it as additional data to interpret and analyze CTP scans of suspected stroke patients for the adult population. There is no change to the standard of care assessment of suspected stroke patients where the healthcare professional assesses the medical imaqes. StrokeViewer only provides additional imaging data and numerical data that may be taken into consideration by a healthcare professional. StrokeViewer is not intended to be used as a standalone diagnostic tool, and StrokeViewer output should always be interpreted in the context of clinical information about the patient rather than in isolation.
{6}------------------------------------------------
4.1 Indications for Use
StrokeViewer Perfusion is an image processing software package intended to provide quantitative perfusion information in brain tissue for suspected ischemic stroke patients. It is to be used by medical imaging professionals who analyze dynamic perfusion studies, including but not limited to physicians such as neurologists, and radiologists.
The software is packed as a Docker container allowinq installation on a standard "off-the-shelf" computer or a virtual platform. The software can be used to perform image viewing, processing, and analysis of brain CT perfusion images (CTP) images. Data and images are acquired through DICOM compliant imaging devices.
StrokeViewer Perfusion is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time, which are visualized as colored perfusion maps including flow-related parameters and tissue blood volume quantification.
Contraindications:
- Bolus Quality: absent or inadequate bolus.
- Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. .
- Presence of hemorrhage .
Comparison of Indications for Use Statements
Comparison of the indications for use statements between the subject and predicate cleared devices show that the statements are similar in classification of device, intended users, types of platform environments as well as the types of data required for input in order to acquire the identical types of output parameters for CT Perfusion.
The only difference between the devices is that StrokeViewer Perfusion currently does not support CTA or MR data in analysis of the Perfusion alqorithm. StrokeViewer Perfusion's indications for use statement is considered to be a subset of the indications for use that the predicate cleared device utilizes.
Given this difference, we have concluded that the differences in fundamental scientific technology do not affect the existing features of the StrokeViewer platform, the introduction of the StrokeViewer Perfusion algorithm to our existing platform, nor does it raise different questions of the safety and effectiveness of the StrokeViewer device, including the Perfusion algorithm.
Comparison of Technological Characteristics with the Predicate Device റ
As compared to the predicate device (K182130), StrokeViewer Perfusion works with Computed Tomography (CT) DICOM compliant medical image data. The algorithm analyzes dynamic imaging data acquired using CT perfusion scan protocol.
StrokeViewer Perfusion also similarly provides tools for performing the following types of analysis as compared to the predicate device:
- -Volumetry of threshold maps
- -Time-intensity plots for dynamic time courses
- -Measurement of mismatch between labeled volumes on co-registered image volumes
Like the predicate device, the StrokeViewer software consists of a variety of modules which support different analysis methods used in clinical practice. For this submission, the following is available with the introduction of the Perfusion algorithm:
- -StrokeViewer Dynamic Analysis: used to visualize and analyze dynamic imaging data that displays properties of changes in contrast over time. This includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume: CBF, CBV, MTT and Tmax.
The main difference between StrokeViewer Perfusion and the predicate cleared device, iSchemaView Rapid (K182130), is that StrokeViewer Perfusion does not analyze CTA or MR data.
The platform previously cleared as part of the reference device, StrokeViewer HALO (LVO Triaging, K200873), is now able to accommodate the deployment of the Perfusion algorithm. The previously cleared features remain unchanged. This addition/feature is optional.
StrokeViewer Perfusion's indications for use statement is considered to be similar to that of the predicate device. The minor differences in fundamental scientific technology do not affect the existing features of the StrokeViewer platform, the introduction of the StrokeViewer Perfusion algorithm to our existing platform, nor does it raise different questions of the safety and effectiveness of StrokeViewer Perfusion.
{7}------------------------------------------------
7 Performance and safety
7.1 Software Validation and Verification
Performance data has been provided in this submission in accordance to the following FDA Guidance Documents in order to support our conclusion of substantial equivalence to the predicate device:
- -"Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005
- -"Content of Premarket Submissions for Management of Cybersecurity in Medical Devices—Final Guidance", October 2, 2014
- -"Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices", September 6, 2017
The performance data included is in accordance with the general controls outlined for a Software as a Medical Device (SaMD).
Design, risk management, verification, and validation were carried out in compliance with CFR 21 Part 820 and the DICOM standard (NEMA PS 3.1-3.20). The results of software verification, and algorithmic testing demonstrate that StrokeViewer Perfusion meets all design requirements and specifications for its intended use.
The testing included within this submission upholds our statement for safety and effectiveness of the proposed device while supporting our claim of substantial equivalence to the predicate device, K182130.
7.2 Performance testing
Clinical data was not required for this submission of the StrokeViewer Perfusion. We believe the safety and effectiveness of the proposed device was appropriately tested with non-clinical validation.
Software testing data included in the submission is in accordance with controls outlined for Software as a Medical Device and comprises the following testing: Unit testing, End-to-end testing, and reproducibility testing.
Non-clinical bench performance testing was performed to assess Perfusion algorithm performance on simulated datasets (Kudo digital phantom) generated using simulating tracer kinetic theory. Correlations between the output of the StrokeViewer Perfusion device and the ground truth values were calculated. The results of algorithm performance testing showed that the device met performance goals and acceptance criteria.
The tests performed demonstrate that Perfusion in StrokeViewer performs as intended. All testing included within the respective sections of this submission uphold our belief that StrokeViewer is substantially equivalent to the predicate device (K182130).
7.3 Risk Management File
StrokeViewer Perfusion was tested according to the FDA quidance and recognized standards. ISO 14971:2019, Medical devices—Application of risk management to medical devices, was followed.
All risk controls were identified, implemented, and mitigated according to hazards identified. All testing related to risk controls support the acceptance criteria outlined for our software requirement specifications.
7.4 Recognized Consensus Standards
FDA Recognized Consensus Standards were followed when testing the Perfusion feature within StrokeViewer to uphold claims of safety and effectiveness of the intended purpose of the device:
| Date of Entry | Recognition Number | Standards Developing Organization | Standard Designation Number and Date | Title of Standard |
|---|---|---|---|---|
| 12/23/2019 | 5-125 | ISO | 14971:2019 | Medical devices - Applications of risk management to medical devices |
| 07/06/2020 | 5-129 | IEC | 62366-1:2015+AMD1:2020 (Consolidated text) | Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1 |
| 01/14/2019 | 13-79 | IEC | 62304:2006/A1:2016 | Medical device software - Software lifecycle processes (including Amendment 1 (2016)) |
| 12-19-2022 | 12-349 | NEMA | PS 3.1 - 3.20 2022d | Digital Imaging and Communications in Medicine (DICOM) Set |
{8}------------------------------------------------
| 12/20/2021 | 5-134 | ISO | 15223-1 Fourth edition 2021-07Medical devices - Symbols to be usedwith information to be supplied by themanufacturer - Part 1: Generalrequirements |
|---|---|---|---|
| 12/19/2022 | 5-135 | ISO | 20417 First edition2021-04 Correctedversion 2021-12Medical devices - Information to besupplied by the manufacturer |
| 08/21/2017 | 13-97 | IEC | 82304-1 Edition1.0 2016-10Health Software - Part 1: Generalrequirements for product safety |
Harmonized standards that are not declared in the FDA Recognized Standards Database but are still followed:
| Standard Designation Number | Title of Standard |
|---|---|
| ISO 13485:2016 | Medical Devices—Quality Management Systems—Requirements forRegulatory Purposes |
| ISO/IEC 27001:2022 | Information Security, Cybersecurity and privacy protection—Informationsecurity management systems—Requirements |
| NEN 7510-1:2017+A1:2020 | Health informatics—Information security management in healthcare—Part 1: Management System |
| NEN 7510-2:2017 | Health informatics—Information security management in healthcare—Part 2: Controls |
8 Conclusion
StrokeViewer Perfusion's indications for use are similar to the predicate device (K182130).
The fundamental technological characteristics between both devices are similar as well. All attributes of the tested software, when compared, uphold our claim of substantial equivalence.
Nicolab considers StrokeViewer Perfusion to demonstrate the same safety and efficacy to the commercially available predicate device. Therefore, we believe StrokeViewer Perfusion demonstrates substantial equivalence to the predicate device (K182130) and have included evidence to support this statement throughout the submission.
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).