(143 days)
iSchemaView's RAPID is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians.
The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of images. Data and images are acquired through DICOM compliant imaging devices.
The iSchemaView RAPID provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CT-P), CT Angiography (CTA), and MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT).
The DWI Module is used to visualize local water diffusion properties from the analysis of diffusion - weighted MRI data. The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
RAPID CT-Perfusion and RAPID MR-Perfusion can be used by physicians to aid in the selection of acute stroke patients. (with known occlusion of the intracranial internal carotid artery or proximal middle cerebral artery) for endovascular thrombectomy.
Instructions for use of contrast agents for this indication can be found in Appendix A of the User's Manual. Additional information for safe and effective drug use is available in product-specific iodinated CT and gadolinium-based MR contrast drug labeling.
In addition to the RAPID imaging criteria, patients must meet the clinical requirements for thrombectomy, as assessed by the physician, and have none of the following contraindications or exclusions.
Contraindications/Exclusions:
- · Bolus Quality: absent or inadequate bolus.
- · Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate.
- Presence of Hemorrhage
RAPID is a software package that provides for the visualization and study of changes in tissue using digital images captured by diagnostic imaging systems including CT (Computed Tomography) and MRI (Magnetic Image Resonance), as an aid to physician diagnosis. RAPID can be installed on a customer's Server or it can be accessed online as virtual system. It provides viewing, quantification, analysis and reporting capabilities.
RAPID works with the following types of (DICOM compliant) medical image data:
- CT (Computed Tomography)
- MRI (Magnetic Image Resonance) ●
RAPID acquires (DICOM compliant) medical image data from the following sources:
- DICOM file
- DICOM CD-R ●
- Network using DICOM protocol .
RAPID provides tools for performing the following types of analysis:
- . selection of acute stroke patients for endovascular thrombectomy
- volumetry of thresholded maps
- . time intensity plots for dynamic time courses
- measurement of mismatch between labeled volumes on co-registered image ● volumes
- . large vessel density
RAPID is a software-only device consisting of one or more RAPID Servers (dedicated or virtual and an iSchemaView Server). The RAPID Server is an image processing engine that connects to a hospital LAN, inside the Hospital Firewall. It can be a dedicated RAPID Server or a VM RAPID appliance, which is a virtualized RAPID Server that runs on a dedicated hospital server. Where available, the RAPID Server is placed logically in the demilitarized zone (DMZ) of the hospital's network to facilitate bidirectional secure connection between the (local) RAPID Server and the centralized iSchemaView Server.
RAPID is a software-only device designed to streamline medical image processing tasks that are time consuming and fatiguing in routine patient workup. RAPID is typically installed on a server running within a hospital's network and operates with minimal user interaction. Once the CT or MR data are acquired, the CT or MRI console operator selects RAPID as the target for the DICOM images, and then the operator selects which study/series data to be sent to RAPID. Based on the type of incoming DICOM data, RAPID will identify the data set as CT or MRI data, and determine the suitable processing module. RAPID is a toolbox of modules which support various analysis methods used in clinical practice today:
- . RAPID CTA: used to visualize large cerebral vessels and analyze hemispheric difference via contralateral comparison.
- RAPID MR DWI Module: used to visualize local water diffusion properties of tissue . from the analysis of diffusion-weighted MRI data.
- RAPID Dynamic Analysis Module: used for visualization and analysis of dynamic . imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
The iSchemaView Server is a dedicated server that provides a central repository for RAPID data. All iSchemaView Server data is stored on encrypted hard disks. It also provides a user interface for accessing RAPID data. It connects to a firewalled Data Center Network and has its own firewall for additional cyber/data security. The iSchemaView Server connects to one or more RAPID Servers via WAN. Available types of connection include VPN (Virtual Private Network - RFC2401 and RFC4301 Standards) Tunnel and SSH (Secure Shell).
I am sorry, but the provided text does not contain the specific information required to describe the acceptance criteria and the study that proves the device meets them in the requested format.
The document is an FDA 510(k) summary for the iSchemaView RAPID device. It focuses on demonstrating substantial equivalence to a predicate device (K172477) rather than detailing a specific clinical performance study with acceptance criteria and results.
While it mentions performance validation testing, software verification and validation, and conformance to standards, it does not provide:
- A table of acceptance criteria and reported device performance.
- Sample sizes for a test set, data provenance, or the number/qualifications of experts for ground truth.
- Adjudication methods.
- Results of a multi-reader multi-case (MRMC) comparative effectiveness study, or a standalone algorithm study.
- The type of ground truth used in detail for a performance study.
- Training set sample size or how its ground truth was established.
The document highlights the intended use changes and technological characteristics to demonstrate equivalence, but not the specific performance metrics and studies that would prove it meets acceptance criteria for the expanded indication.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
iSchemaView, Inc. Jim Rosa VP Quality Assurance and Regulatory Affairs 433 Park Point Drive. Suite 220 GOLDEN. CO 80401
Re: K182130
Trade/Device Name: iSchemaView RAPID Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: November 28, 2018 Received: December 3, 2018
Dear Jim Rosa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
December 27, 2018
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K182130
Device Name iSchemaView RAPID
Indications for Use (Describe)
iSchemaView's RAPID is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians.
The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of images. Data and images are acquired through DICOM compliant imaging devices.
The iSchemaView RAPID provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CT-P), CT Angiography (CTA), and MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT).
The DWI Module is used to visualize local water diffusion properties from the analysis of diffusion - weighted MRI data. The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
RAPID CT-Perfusion and RAPID MR-Perfusion can be used by physicians to aid in the selection of acute stroke patients. (with known occlusion of the intracranial internal carotid artery or proximal middle cerebral artery) for endovascular thrombectomy.
Instructions for use of contrast agents for this indication can be found in Appendix A of the User's Manual. Additional information for safe and effective drug use is available in product-specific iodinated CT and gadolinium-based MR contrast drug labeling.
In addition to the RAPID imaging criteria, patients must meet the clinical requirements for thrombectomy, as assessed by the physician, and have none of the following contraindications or exclusions.
Contraindications/Exclusions:
- · Bolus Quality: absent or inadequate bolus.
- · Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate.
- · Presence of Hemorrhage
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
iSchemaView - Traditional 510(k) RAPID Section 5: 510(k) Summary
510(k) Summary
iSchemaView, Inc.'s RAPID
This document contains the 510(k) summary for the iSchemaView RAPID. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).
Applicant Name and Address:
| Name: | iSchemaView, Inc. |
|---|---|
| Address: | 433 Park Point Drive, Suite 220Golden, CO 80401 |
| Official Contact: | Jim RosaPhone: (303) 704-3374Email: rosa@ischemaview.com |
| Summary Preparation Date: | November 28, 2018 |
|---|---|
| --------------------------- | ------------------- |
Device Name and Classification:
| Trade Name: | iSchemaView RAPID | ||
|---|---|---|---|
| Common Name: | PACS – Picture Archiving CommunicationsSystem | ||
| Classification: | II | ||
| Product Code: | LLZ | ||
| Regulation No: | 21 C.F.R. §892.2050 | ||
| Classification Panel: | Radiology Devices |
Predicate Devices:
The iSchemaView RAPID is claimed to be substantially equivalent to the following legally marketed predicate device:
iSchemaView RAPID (K172477)
Previous FDA Submission:
iSchemaView RAPID (K121447) iSchemaView RAPID (K172477) iSchemaView Q-Submission Q172046-S001
Device Description:
RAPID is a software package that provides for the visualization and study of changes in tissue using digital images captured by diagnostic imaging systems including CT (Computed Tomography) and MRI (Magnetic Image Resonance), as an aid to physician diagnosis. RAPID can be installed on a customer's Server or it can be accessed online as virtual system. It provides viewing, quantification, analysis and reporting capabilities.
{4}------------------------------------------------
Section 5: 510(k) Summary
RAPID works with the following types of (DICOM compliant) medical image data:
- CT (Computed Tomography)
- MRI (Magnetic Image Resonance) ●
RAPID acquires (DICOM compliant) medical image data from the following sources:
- DICOM file
- DICOM CD-R ●
- Network using DICOM protocol .
RAPID provides tools for performing the following types of analysis:
- . selection of acute stroke patients for endovascular thrombectomy
- volumetry of thresholded maps
- . time intensity plots for dynamic time courses
- measurement of mismatch between labeled volumes on co-registered image ● volumes
- . large vessel density
RAPID is a software-only device consisting of one or more RAPID Servers (dedicated or virtual and an iSchemaView Server). The RAPID Server is an image processing engine that connects to a hospital LAN, inside the Hospital Firewall. It can be a dedicated RAPID Server or a VM RAPID appliance, which is a virtualized RAPID Server that runs on a dedicated hospital server. Where available, the RAPID Server is placed logically in the demilitarized zone (DMZ) of the hospital's network to facilitate bidirectional secure connection between the (local) RAPID Server and the centralized iSchemaView Server.
RAPID is a software-only device designed to streamline medical image processing tasks that are time consuming and fatiguing in routine patient workup. RAPID is typically installed on a server running within a hospital's network and operates with minimal user interaction. Once the CT or MR data are acquired, the CT or MRI console operator selects RAPID as the target for the DICOM images, and then the operator selects which study/series data to be sent to RAPID. Based on the type of incoming DICOM data, RAPID will identify the data set as CT or MRI data, and determine the suitable processing module. RAPID is a toolbox of modules which support various analysis methods used in clinical practice today:
- . RAPID CTA: used to visualize large cerebral vessels and analyze hemispheric difference via contralateral comparison.
- RAPID MR DWI Module: used to visualize local water diffusion properties of tissue . from the analysis of diffusion-weighted MRI data.
- RAPID Dynamic Analysis Module: used for visualization and analysis of dynamic . imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
The iSchemaView Server is a dedicated server that provides a central repository for RAPID data. All iSchemaView Server data is stored on encrypted hard disks. It also provides a user
{5}------------------------------------------------
Section 5: 510(k) Summary
interface for accessing RAPID data. It connects to a firewalled Data Center Network and has its own firewall for additional cyber/data security. The iSchemaView Server connects to one or more RAPID Servers via WAN. Available types of connection include VPN (Virtual Private Network - RFC2401 and RFC4301 Standards) Tunnel and SSH (Secure Shell).
Indications for Use:
iSchemaView's RAPID is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians.
The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of images. Data and images are acquired through DICOM compliant imaging devices.
The iSchemaView RAPID provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CT-P), CT Angiography (CTA), and MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT).
The DWI Module is used to visualize local water diffusion properties from the analysis of diffusion - weighted MRI data.
The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
RAPID CT-Perfusion and RAPID MR-Perfusion can be used by physicians to aid in the selection of acute stroke patients (with known occlusion of the intracranial internal carotid artery or proximal middle cerebral artery) for endovascular thrombectomy.
Instructions for use of contrast agents for this indication can be found in Appendix A of the User's Manual. Additional information for safe and effective drug use is available in productspecific iodinated CT and gadolinium-based MR contrast drug labeling.
In addition to the RAPID imaging criteria, patients must meet the clinical requirements for thrombectomy, as assessed by the physician, and have none of the following contraindications or exclusions.
Contraindications/Exclusions:
- Bolus Quality: absent or inadequate bolus.
- Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate.
- · Presence of Hemorrhage.
Technological Characteristics:
RAPID performs the following functions:
- processes DICOM images from multiple sources to provide visualization of changes ● of tissue perfusion, diffusion and change.
{6}------------------------------------------------
Section 5: 510(k) Summary
- receives DICOM images from external DICOM image providers (modalities (CT/MRI . Scanners), PACS and Workstations) and sends DICOM images to external image consumers.
- . processes requests, statuses and results, and references therein, which are stored in a searchable database
- processing status is available through a web browser using HTTP. HTML and PHP. .
- can send summary results to the user over email. For this, RAPID generally connects to the infrastructure of the medical partner (e.g., the hospital). In particular, RAPID uses a SMTP protocol with security extensions to provide secure emailing.
RAPID is available in the following configurations:
- Standard RAPID, which is installed directly on a customer's Linux-based server and . integrated with medical image processing software such as commercial PACS.
- . Virtual RAPID, wherein the user accesses RAPID online and uses it to process DICOM images otherwise available on his/her computer.
RAPID is a DICOM-compliant PACS software that provides comprehensive functionality to transfer, process, and display modality specific imaging data. RAPID runs on standard "offthe- shelf computer and networking hardware. RAPID is entirely independent from CT, MRI, or PACS platforms. It supports secure VPN (Virtual Private Network) networking or encapsulated Secure Shell (SSH), and seamlessly integrates into an existing radiological data network.
Clinical Characteristics:
The primary users of RAPID PACS software are medical imaging professionals who analyze tissue using CT or MRI images. The images generated by RAPID provide additional diagnostic information. which is derived from the temporal/diffusion/density features of the native CT or MRI images.
RAPID CT Perfusion and RAPID MRI can be used by physicians to select acute stroke patients for endovascular thrombectomy. The recommended selection criteria are listed in the table below. Patients must meet the clinical requirements for thrombectomy as assessed by the physician and have no contraindications or exclusions:
{7}------------------------------------------------
Section 5: 510(k) Summary
| KnownNeurologicBaseline(KNB)(Hrs) | Modality | RAPID Map | LesionVolume | DBF/DWIThreshold(rCBF/ADC) | Volume(ml) | TmaxThreshold(sec) | MismatchRatio* |
|---|---|---|---|---|---|---|---|
| MISMATCH BASED INDICATIONS | |||||||
| KNB ≤ 6 | CTP | Mismatch | CBF/Tmax | <0.3 | ≤ 70 | > 6 | ≥ 1.2 CBF |
| MRI | Mismatch | DWI/Tmax | <620 | ≤ 50 | > 6 | ≥ 1.8 DWI | |
| 6 < KNB ≤ 16 | CTP | Mismatch | CBF/Tmax | <0.3 | ≤ 70 | > 6 | ≥ 1.8 CBF |
| MRI | Mismatch | DWI/Tmax | <620 | ≤ 70 | > 6 | ≥ 1.8 DWI | |
| ISCHEMIC CORE RELATED INDICATIONS | |||||||
| Patient ≥ 80 yrs AND NIHSS ≥ 10 | |||||||
| 6 ≤ KNB ≤ 24 | CTP | Mismatch | CBF | <0.3 | ≤ 20 | NA | NA |
| MRI | Mismatch/DWI | DWI | <620 | ≤ 20 | NA | NA | |
| Patient < 80 yrs AND NIHSS 10-19 | |||||||
| 6 ≤ KNB ≤ 24 | CTP | Mismatch | CBF | <0.3 | ≤ 30 | NA | NA |
| MRI | Mismatch/DWI | DWI | <620 | ≤ 30 | NA | NA | |
| Patient ≤ 80 yrs AND NIHSS ≥ 20 | |||||||
| 6 ≤ KNB ≤ 24 | CTP | Mismatch | CBF | <0.3 | ≤ 50 | NA | NA |
| MRI | Mismatch/DWI | DWI | <620 | ≤ 50 | NA | NA | |
| *Note: A Mismatch volume ≥ 15ml is recommended |
Contraindications/Exclusions:
- Bolus Quality: absent or inadequate bolus. ●
- Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. ●
- Presence of Hemorrhage. ●
Performance Standards:
RAPID has been developed in conformance with the following standards, as applicable:
| EN ISO 14971:2012 | Application of Risk Management to Medical Devices |
|---|---|
| IEC 62304:2015 | Medical device software – Software lifecycle processes |
| IEC 62366:2015 | Application of Usability Engineering to Medical Devices |
| NEMA PS 3.1 - 3.20 | Digital Imaging and Communications in Medicine (DICOM) |
{8}------------------------------------------------
Section 5: 510(k) Summary
Performance Data:
RAPID complies with DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20.
Additionally, iSchemaView conducted extensive performance validation testing and software verification and validation testing of the RAPID system. This performance validation testing demonstrated that the RAPID system provides accurate representation of key processing parameters under a range of clinically relevant parameters and perturbations associated with the intended use of the software. Software performance, validation and verification testing demonstrated that the RAPID system met all design requirements and specifications.
Prescriptive Statement:
Caution: Federal law restricts this device to sale by or on the order of a physician.
Safety & Effectiveness:
RAPID has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2012 (risk management). The RAPID System performance has been validated through the use of phantoms.
Substantial Equivalence:
RAPID is as safe and effective as the previously cleared RAPID product (K172477). RAPID has a similar intended use with an added indication, technological characteristics and principles of operation as its predicate device. There are no technological differences between the RAPID and its predicate (K172477) and no new issues of safety or effectiveness have been identified. The testing during the 4.8 baseline release is provided as verification and validation. The testing conducted with RAPID confirms the software reliably processes and supports the analysis of CT and MRI medical images for tissue evaluation. Thus, the RAPID software is substantially equivalent. The claims have been expanded to include the use of RAPID in selection of acute stroke patients for endovascular thrombectomy, based on retrospective clinical data evaluation.
Visualization and quantitative results delivered by RAPID can be used, upon applicable clinical and technical training of the users, in the diagnostic and patient treatment process.
In version 4.8, RAPID provides the following:
- CT-Perfusion, MR-Perfusion and MR-Diffusion datasets, with the possibility to compute semi-quantitative (relative blood flow, volume, mean transit time, arterio-tissue delay time Tmax) and quantitative (apparent diffusion coefficient) parameters and compute volumes of tissue with parameters within certain ranges;
- -CT-Angiography datasets, where, after skull stripping, the system provides transverse, coronal and sagittal views of the intra-cranial vasculature, as well as hemispheric comparison of the blood vessel density in these datasets.
In version 4.8.3, RAPID provides the indications identified in version 4.8 plus the addition of a special case of CT-Perfusion and MR. RAPID provides guidance to physicians for acute
{9}------------------------------------------------
Section 5: 510(k) Summary
stroke patient selection for endovascular thrombectomy using specific parametric settings, as described in Table in paragraph 3.2.
The following table summarizes and compares data on the iSchemaView RAPID (K172477) to the RAPID (K182130) that is the subject of this Traditional 510(k) submission.
{10}------------------------------------------------
Section 5: 510(k) Summary
Substantial Equivalence Discussion:
| Parameter | RAPID (K182130) | RAPID (K172477) |
|---|---|---|
| Product Code | LLZ | LLZ |
| Regulation | 21 CFR §892.2050 | 21 CFR §892.2050 |
| Intended Use/ Indications forUse | iSchemaView's RAPID is an image processing software package tobe used by trained professionals, including but not limited tophysicians and medical technicians. | iSchemaView's RAPID is an image processing software package tobe used by trained professionals, including but not limited tophysicians and medical technicians. |
| The software runs on a standard off-the-shelf computer or a virtualplatform, such as VMware, and can be used to perform imageviewing, processing and analysis of images. Data and images areacquired through DICOM compliant imaging devices. | The software runs on a standard off-the-shelf computer or a virtualplatform, such as VMware, and can be used to perform imageviewing, processing and analysis of images. Data and images areacquired through DICOM compliant imaging devices. | |
| The iSchemaView RAPID provides both viewing and analysiscapabilities for functional and dynamic imaging datasets acquiredwith CT Perfusion (CT-P), CT Angiography (CTA), and MRIincluding a Diffusion Weighted MRI (DWI) Module and a DynamicAnalysis Module (dynamic contrast-enhanced imaging data for MRIand CT). | The iSchemaView RAPID provides both viewing and analysiscapabilities for functional and dynamic imaging datasets acquiredwith CT Perfusion (CT-P), CT Angiography (CTA), and MRIincluding a Diffusion Weighted MRI (DWI) Module and a DynamicAnalysis Module (dynamic contrast-enhanced imaging data for MRIand CT). | |
| The DWI Module is used to visualize local water diffusionproperties from the analysis of diffusion - weighted MRI data. | The DWI Module is used to visualize local water diffusionproperties from the analysis of diffusion - weighted MRI data. | |
| The Dynamic Analysis Module is used for visualization andanalysis of dynamic imaging data, showing properties of changesin contrast over time. This functionality includes calculation ofparameters related to tissue flow (perfusion) and tissue bloodvolume. | The Dynamic Analysis Module is used for visualization andanalysis of dynamic imaging data, showing properties ofchanges in contrast over time. This functionality includescalculation of parameters related to tissue flow (perfusion) andtissue blood volume. | |
| Change to K172477: | ||
| RAPID CT-Perfusion and RAPID MR-Perfusion can be used byphysicians to aid in the selection of acute stroke patients (withknown occlusion of the intracranial internal carotid artery orproximal middle cerebral artery) for endovascularthrombectomy. | ||
| Instructions for use of contrast agents for this indication can befound in Appendix A of the User's Manual. Additionalinformation for safe and effective drug use is available inproduct-specific iodinated CT and gadolinium-based MRcontrast drug labeling. | ||
| In addition to the RAPID imaging criteria, patients must meetthe clinical requirements for thrombectomy, as assessed by the | ||
| physician, and have none of the following contraindications or exclusions. | ||
| Contraindications/Exclusions: | ||
| • Bolus Quality: absent or inadequate bolus. | ||
| • Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. | ||
| • Presence of Hemorrhage. | ||
| PACS Functionality | ||
| Basic PACS Functions | View, process and analyze medical images. Performs standard PACS functions with respect to querying and listing. | Same |
| Computer Platform | Standard off-the-shelf PC workstation/serverVirtual platform such as VMware | Same |
| DICOM Compliance | Yes | Same |
| Functional Overview | RAPID is a software package that provides for the visualization and study of changes of tissue in digital images captured by CT and MRI. RAPID provides viewing and quantification. | Same |
| Data Acquisition | Acquires medical image data from DICOM compliant imaging devices and modalities | Same |
| Data/Image Types | Computed Tomography (CT) via DICOM Format | Same |
| Magnetic Image Resonance (MRI) via DICOM Format | Same | |
| Acquisition and Modalities Features | ||
| MRI | Diffusion Weighted Image (DWI)Dynamic Analysis tissue flow (perfusion) and tissue blood volume | Same |
| CT | CT Perfusion (CTP)CTA-large vessel density analysis | Same |
| Computed Parameter Maps | ||
| Diffusion MRI | Isotropic DWI (isoDWI)ADCTrace of diffusion tensor (Trace)Fractional Anisotropy (FA) and color FA | Same |
| Perfusion MRI and PerfusionCT | Cerebral blood flow (CBF)Cerebral blood volume (CBV)Mean transit time (MTT) | |
| Tissue residue function time to peak (Tmax) | Same | |
| Measurement Tools | ||
| MRI and CT Tools | Arterial input function (AIF)Venous output function(VOF) | Same |
| Time-course | Same | |
| Mask | Same | |
| Region of interest (ROI) and Volumetry | Same | |
| Volumetric comparison between 2 ROIs | Same | |
| Motion correction | Same | |
| Export perfusion and diffusion files to PACS andDICOM file systems | Same | |
| Acquire, transmit, process, and store medicalimages | Same |
{11}------------------------------------------------
Section 5: 510(k) Summary
{12}------------------------------------------------
Section 5: 510(k) Summary
{13}------------------------------------------------
Section 5: 510(k) Summary
Conclusion:
In conclusion, the iSchemaView RAPID is substantially equivalent in intended use, technological characteristics, safety, and performance characteristics to the legally marketed predicate device, RAPID (K172477).
N/A