K Number
K231480

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-07-28

(67 days)

Product Code
Regulation Number
872.3820
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Bright MTA Sealer Plus is used for filling root canals.

Device Description

Bright MTA Sealer Plus is a ready-to-use, injectable paste-like material for root canal filling, which is hardened and obturated after being injected into the root canal space. The product based on calcium silicate exhibits excellent biocompatibility as well as a low film thickness suitable for easy penetration of lateral and accessory canals.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for a medical device called "Bright MTA Sealer Plus." However, it is not an AI-driven device; it is a root canal filling material. Therefore, some of the requested information, such as details on AI acceptance criteria, expert ground truthing, MRMC studies, or training/test sets for an AI model, are not applicable and thus not present in the document.

The document primarily focuses on the biocompatibility and mechanical performance of the "Bright MTA Sealer Plus" in comparison to a predicate device.

Here's a breakdown of the available information based on your request, with an explanation for elements that are not applicable:


Device Name: Bright MTA Sealer Plus

Device Type: Root canal filling material (Non-AI device)


1. Table of Acceptance Criteria and Reported Device Performance

The closest information available is in the "Mechanical testing" section:

No.ItemsStandard & MethodAcceptance CriteriaResult
1Visual testISO 4049 Bare eyesNo alien substance and suitable for using the productNo alien substance and suitable for using the product
2Capacity testISO 4049 Weight differenceStandard weight ± 5%1.50 %
3Package testISO 4049 Bare eyesNo damages, cracksThe package was completely sealed, and there were no damages, cracks.
4Extraneous matter testISO 4049 Bare eyesNo extraneous matterNo Extraneous Matter
5Flow testEN ISO 6876: 2012 5.2Diameter ≥ 17mm27mm
6Setting time testEN ISO 6876: 2012 5.4≤ 360 minWithin 360min
7Film thickness testEN ISO 6876: 2012 5.5≤ 50 μm30 μm
8Radio-opacity testEN ISO 6876: 2012 5.7More than 3mm4.6mm
9Solubility testEN ISO 6876: 2012 5.6≤ 3%0.2%

In addition, Biocompatibility testing was performed with the following acceptance criteria and evaluation:

No.TestStandard & MethodAcceptance criteriaEvaluation
1CytotoxicityEN ISO 10993-5 Agar diffusion assayNon cytotoxic (Scale 0)Scale 0 (Non cytotoxic)
2Oral mucosal irritationEN ISO 10993-23Irritation index 0Irritation index 0
3Skin SensitizationEN ISO 10993-10 GPMTSensitization score and rate 0Sensitization score and rate 0
4Acute systemic toxicityEN ISO 10993-11 Single doseNo Acute systemic toxicityNo Acute systemic toxicity
5Systemic toxicityEN ISO 10993-11 Pyrogen testNo abnormal signs and deadNo abnormal signs and dead
6GenotoxicityEN ISO 10993-3 Back mutation.No back mutation regardless of the presence or absence of a metabolic activation systemNo back mutation
Chromosomal aberrationNo chromosomal aberration in CHL/IU cellsNo chromosomal aberration in CHL/IU cells
7ImplantationEN ISO 10993-6 ImplantationBiocompatibleBiocompatible
8Sub-chronic toxicityEN ISO 10993-11 Subchronic toxicityNo Subchronic toxicityNo Subchronic toxicity
9Chronic toxicityEN ISO 10993-11 Chronic toxicityNo chronic toxicityNo chronic toxicity
10CarcinogenicityEN ISO 10993-3 CarcinogenicityNo CarcinogenicityNo Carcinogenicity

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated in terms of number of samples/units tested for each mechanical or biocompatibility test, but the tests were performed on the device itself.
  • Data Provenance: The document implies in-house testing performed by or for Genoss Co., Ltd. The country of origin for the company is South Korea. The studies are assumed to be prospective tests on newly manufactured samples of the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This is not an AI device that requires expert-established ground truth for image interpretation or diagnosis. The ground truth for this device's performance is derived from standardized physical and biological material testing.

4. Adjudication method for the test set

  • Not Applicable. As this is not an AI device involving human interpretation, there is no need for an adjudication method for a test set. The results are based on objective measurements from standardized tests.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

  • No. An MRMC study is relevant for AI systems that assist human readers in tasks like radiological interpretation. This device is a material, not an AI system, so an MRMC study would not be performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an algorithm or AI system. Its performance is inherent to its physical and chemical properties.

7. The type of ground truth used

  • The ground truth for this device's performance is based on objective measurements from internationally recognized standards (ISO and EN ISO) for dental materials and biocompatibility. For example, specific diameters for flow, time limits for setting, weight differences, and qualitative assessments of physical appearance and biological reactions are used as the "ground truth" against which the device's performance is measured. It's not expert consensus, pathology, or outcomes data in the traditional sense of diagnostic AI.

8. The sample size for the training set

  • Not Applicable. This is not an AI device that undergoes machine learning training.

9. How the ground truth for the training set was established

  • Not Applicable. As it's not an AI device, there is no training set or associated ground truth establishment process.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 28, 2023

Genoss Co., Ltd. Boram Choi Assistant Manager D-factory, 56, Changnyong-daero 256beon-gil, Yeongtong-gu Suwon-si. Gyeonggi-do 16229 SOUTH KOREA

Re: K231480

Trade/Device Name: Bright MTA Sealer Plus Regulatory Class: Class II Product Code: KIF Dated: April 5, 2023 Received: May 22, 2023

Dear Boram Choi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha -S

Michael Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231480

Device Name Bright MTA Sealer Plus

Indications for Use (Describe) Bright MTA Sealer Plus is used for filling root canals

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first three letters, "GEN", are in a dark green color. The last three letters, "OSS", are in a lighter green color.

510(k) Summary

02/05/2023

1. Company

Submitter
NameGENOSS Co., Ltd.
AddressHead office: 1F Gyeonggi R&DB center, 105 Gwanggyo-ro, Yeongtong-gu,Suwon-si, Gyeonggi-do, 16229, Republic of KoreaFactory: D-factory, 56, Changnyong-daero 256beon-gil, Yeongtong-gu,Suwon-si, Gyeonggi-do, 16256, Republic of Korea
Phone/Fax+82-70-7098-6352/ +82-31-888-5595
Contact personBoram Choi / RAbrchoi@genoss.com
Summary Date02/05/2023

2. Device

Proprietary name:Bright MTA Sealer Plus
Common Name:Root filling material
Classification Name:Root canal filling resin
Regulation Number:21 CFR 872.3820
Product Code:KIF
Regulatory Class:Class II

3. Predicate Device

K170175 Endoseal MTA

4. Device Description

Bright MTA Sealer Plus is a ready-to-use, injectable paste-like material for root canal filling, which is hardened and obturated after being injected into the root canal space. The product based on calcium silicate exhibits excellent biocompatibility as well as a low film thickness suitable for easy penetration of lateral and accessory canals.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first three letters, "GEN", are in a dark green color. The last three letters, "OSS", are in a lighter green color.

5. Indication for use

Bright MTA Sealer Plus is used for filling root canals.

6. Technological Characteristics

Bright MTA Sealer Plus was compared with the predicate device 'ENDOSEAL MTA' in clinical, technical, biological view. The characteristics that differed from the predicate device were performed by gap analysis, which confirmed equivalence with the predicate device. Technological characteristics of Bright MTA Sealer Plus and ENDOSEAL MTA are as following;

Device nameSUBJECT DEVICEPREDICATED DEVICE
Bright MTA Sealer PlusEndoseal MTA(K170175)
ManufactureGenoss Co.,Ltd.Maruchi
ClassificationClass IIClass II
Product codeKIFKIF
ClinicalTarget populationHuman tooth root canalHuman tooth root canal
PurposeBright MTA Sealer Plus is used forfilling root canals.Materials used for root canal filling
Site of applicationroot canalroot canal
ClinicalPerformanceRepair of root resorptionRoot-end-fillingRepair of root resorptionRoot-end-filling
Principle ofoperationBright MTA Sealer Plus is an injectablepaste like material for root canal filling,which is hardened and obturated afterbeing injected into the root canal space.Endoseal MTA is an endodontic sealerbased on MTA, providing a root canalfilling.
TechnicalSolubility0.2 %0.7 %
Flow test27 mm21 mm
Setting timeWithin 360 min12.31 min
Film Thickness30 μm15 μm
Radio-opacity4.6 mm10.5 mm
B1010MaterialCaCO3, Al2O3, SiO2, NaHCO3(Calcium silicate)ZrO₂, Li₂Co₃, 1,3-PropandiolCaCO3, Al2O3, SiO2, Fe2O3(Natural pure cement)ZrO₂, Bi₂O₃, Bentonite Clay

{5}------------------------------------------------

Polyethylene glycol,Polar acidic ester of long chainalcoholsN-Methyl-2-Pyrrolidone,Hypromellose
Chemical SafetyBiocompatibleBiocompatible
SterileNon sterileNon sterile
Shelf-Life2 years2 years
PackagingPre-loaded syringePre-loaded syringe
Duration of ContactPermanent contact - contact exceeds30dPermanent contact - contact exceeds30d

7. Performance Data

Biocompatibility testing

Biocompatibility testing on the proposed Bright MTA Sealer has been completed. Requirements for biological evaluation of the proposed device were based on FDA recognized concensus standard of ISO10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and nonsensitizing to biological bone and tissues with its intended use. The following biocompatibility tests were completed:

No.TestStandard & MethodAcceptance criteriaEvaluation
1CytotoxicityEN ISO 10993-5Agar diffusion assayNon cytotoxic(Scale 0)Scale 0(Non cytotoxic)
2Oral mucosalirrittationEN ISO 10993-23Irritation index 0Irritation index 0
3Skin SensitizationEN ISO 10993-10GPMTSensitization scoreand rate 0Sensitization score andrate 0
4Acute systemictoxicityEN ISO 10993-11Single doseNo Acute systemictoxicityNo Acute systemictoxicity
5Systemic toxicityEN ISO 10993-11Pyrogen testNo abnormal signsand deadNo abnormal signs anddead
6GenotoxicityEN ISO 10993-3Back mutation.No back mutationregardless of theNo back mutation

{6}------------------------------------------------

Chromosomalaberrationpresence or absenceof a metabolicactivation system
No chromosomalaberration inCHL/IU cellsNo chromosomalaberration in CHL/IUcells
7ImplantationEN ISO 10993-6ImplantationBiocompatibleBiocompatible
8Sub-chronictoxicityEN ISO 10993-11Subchronic toxicityNo SubchronictoxicityNo Subchronictoxicity
9Chronic toxicityEN ISO 10993-11Chronic toxicityNo chronic toxicityNo chronic toxicity
10CarcinogenicityEN ISO 10993-3CarcinogenicityNo CarcinogenicityNo Carcinogenicity

Mechanical testing

The proposed Bright MTA Sealer Plus was evaluated using the following performance bench testing to confirm the performance characteristics:

No.ItemsStandard &MethodAcceptance CriteriaResult
1Visual testISO 4049Bare eyesNo alien substance andsuitable for using theproductNo alien substance andsuitable for using theproduct
2Capacity testISO 4049WeightdifferenceStandard weight $\pm$ ±5%1.50 %
3Package testISO 4049Bare eyesNo damages, cracksThe package wascompletely sealed, andthere were no damages.cracks.
4Extraneousmatter testISO 4049Bare eyesNo extraneous matterNo Extraneous Matter
5Flow testEN ISO 6876:2012 5.2Diameter ≥ 17mm27mm
6Setting time testEN ISO 6876:2012 5.4≤ 360 minWithin 360min
7Film thicknesstestEN ISO 6876:2012 5.5≤ 50 $\mu$ m30 $\mu$ m
8Radio-opacitytestEN ISO 6876:2012 5.7More than 3mm4.6mm
9Solubility testEN ISO 6876:2012 5.6≤ 3%0.2%

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first three letters, "GEN", are in a dark green color, while the last three letters, "OSS", are in a lighter green color. The letters are connected, forming a single word.

All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Bright MTA Sealer Plus met the established specifications necessary for consistent performance according to its intended use.

8. Conclusion

Based on the information provided in this premarket notification of GENOSS Co., Ltd. concluded that Bright MTA Sealer Plus is acceptable and safe, substantially equivalent to predicate device.

§ 872.3820 Root canal filling resin.

(a)
Identification. A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.(b)
Classification. (1) Class II if chloroform is not used as an ingredient in the device.(2) Class III if chloroform is used as an ingredient in the device.
(c)
Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to a root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976. Any other root canal filling resin shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.