K Number
K170175

Validate with FDA (Live)

Device Name
ENDOSEAL MTA
Manufacturer
Date Cleared
2018-02-26

(403 days)

Product Code
Regulation Number
872.3820
Panel
Dental
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • Permanent obturation of the root canal following vital pulp-extirpation
  • Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.
Device Description

ENDOSEAL MTA is an endodontic sealer based on MTA, providing a biocompatible and effective root canal filling. It is premixed and pre-loaded in a syringe, which allows a complete filling of the entire root canal including accessory and lateral canals. The product is eugenol-free and will not impede adhesion inside the root canal.

AI/ML Overview

Below is a summary of the acceptance criteria and study information for the Endoseal MTA device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Test ParameterAcceptance Criteria (from referenced standards)Reported Device Performance
ISO 6876Visual, Capacity, Packaging, Flow, Setting time, Film Thickness, Solubility, RadiopacityAll test results met the preset test criteria.
Shelf Life TestVisual, Packaging, Setting time, Solubility (as per ISO 6876)All test results met the preset test criteria.
ISO 10993-5 (Cytotoxicity)Not explicitly stated, but implies compliance for biocompatibilityTest results supported substantial equivalence to predicate devices (near identical composition of products after setting with predicate devices, and similar cytotoxicity test results).
ISO 10993-10 (Maximization test for delayed hypersensitivity (LLNA))Not explicitly stated, but implies compliance for biocompatibilityTest results supported substantial equivalence to predicate devices.
ISO 10993-11 (Acute systemic toxicity)Not explicitly stated, but implies compliance for biocompatibilityTest results supported substantial equivalence to predicate devices.
ISO 10993-6 (Implantation)Not explicitly stated, but implies compliance for biocompatibilityTest results supported substantial equivalence to predicate devices.
ISO 10993-3 (AMES & Mammalian Erythrocyte Micronucleus)Not explicitly stated, but implies compliance for biocompatibilityTest results supported substantial equivalence to predicate devices.
Material CompositionComparable properties to predicate devices, demonstrating similar performance specifications.The main ingredients are similar to predicate devices. Despite some differences in other ingredients, biocompatibility and performance test results supported substantial equivalence.
Indications for UseSame as predicate devices.Same as predicate devices.
Shelf Life2 years2 years

2. Sample Size and Data Provenance

The provided text does not specify the sample sizes used for the test set in any of the performance or biocompatibility tests.
The data provenance is from non-clinical testing conducted by MARUCHI, as presented in the 510(k) submission. The country of origin of data is implicitly the Republic of Korea, where MARUCHI is located. The studies are retrospective in the sense that they are summary test results, not prospective trials.

3. Number of Experts and Qualifications for Ground Truth

This information is not applicable to this type of device submission. The "ground truth" for this medical device (a root canal filling resin) is established by adherence to recognized international standards (ISO) for material properties and biocompatibility, as well as by comparison to legally marketed predicate devices. The performance is assessed through laboratory testing, not through expert consensus on diagnostic images or patient outcomes in the same way an AI diagnostic device would be.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving expert review of data (e.g., imaging) where there might be disagreement in interpretations. For performance and biocompatibility testing of a material, the results are typically quantitative or qualitative based on established standard methods, and adjudication is not usually applied in this context.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted for this device. This type of study is relevant for diagnostic or AI-assisted devices where human readers interpret medical images or data. The Endoseal MTA is a material used in a dental procedure.

6. Standalone Performance Study

Yes, a standalone performance study was done in the sense that the device was subjected to a series of bench and biocompatibility tests without human intervention or interpretation of the material's properties by human operators with respect to its performance against the set standards. The tests were performed on the device itself to verify its physical, chemical, and biological properties.

7. Type of Ground Truth Used

The ground truth used for this device is based on established international standards (ISO 6876, ISO 10993 series) for the physical, chemical, and biological properties of root canal filling materials. The device's performance is measured against the criteria defined by these standards, and its safety and effectiveness are supported by demonstrating substantial equivalence to legally marketed predicate devices that also comply with these standards.

8. Sample Size for the Training Set

This information is not applicable. This device is a medical material, not an AI or machine learning algorithm that requires a training set. The "training set" concept is not relevant for evaluating the performance of a physical material.

9. How the Ground Truth for the Training Set was Established

This information is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

February 26, 2018

Maruchi Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc 800 Roosevelt STE 417 Irvine, California 92620

Re: K170175

Trade/Device Name: Endoseal MTA Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: January 17, 2018 Received: January 26, 2018

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mary S. Runner -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K170175

Device Name

ENDOSEAL MTA

Indications for Use (Describe)

· Permanent obturation of the root canal following vital pulp-extirpation

• Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.

Type of Use (Select one or both, as applicable)
V D---------------------------------------------------------------------------------------------------------------------------------------------------------------------------O -- TL- O- -- L--- II-- II-- IO4 OFF OOL O I----

|| | Prescription Use (Part 21 CFR 801 Subpart D)

|__| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

(K170175)

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: 01/17/2018__

Applicant / Submitter: 1.

MARUCHI 2-208, 42-10, Taejanggongdan-gil, Wonju-si, Gangwon-do, 26311, Republic of Korea Tel: +82-33-734-0330 Fax:+82-33-746-2804

2. Submission Correspondent:

Priscilla Chung LK Consulting Group USA, Inc. 800 Roosevelt STE 417 Irvine CA 92620 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com

3. Device:

Proprietary Name:ENDOSEAL MTA
Common Name:Root Filling Material
Classification Name:Root Canal Filling Resin
Classification:Class II, 21 CFR 872.3820
Classification Product Code:KIF

4. Predicate Device:

ENDOSEAL (K133054) by MARUCHI MTA Fillapex (K113568) by Angelus Industria de Produtos Odontologicos IROOT SP (K080917) by Innovative BioCeramix Inc.

{4}------------------------------------------------

5. Device Description:

ENDOSEAL MTA is an endodontic sealer based on MTA, providing a biocompatible and effective root canal filling. It is premixed and pre-loaded in a syringe, which allows a complete filling of the entire root canal including accessory and lateral canals. The product is eugenol-free and will not impede adhesion inside the root canal.

6. Indications for Use

  • Permanent obturation of the root canal following vital pulp-extirpation
  • · Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.

7. Performance Data(Non-Clinical):

The following properties were tested based on the referenced standard. All the test results met the preset test criteria.

  • = ISO 6876 Visual, Capacity, Packaging, Flow, Setting time, Film Thickness, Solubility and Radiopacity
  • = Shelf Life Test ISO 6876 (Visual, Packaging, Setting time, Solubility)
  • ISO 10993-5 Cytotoxicity
  • ISO 10993-10 Maximization test for delayed hypersensitivity (LLNA)
  • ISO 10993-11 Acute systemic toxicity
  • ISO 10993-6 Implantation
  • ISO 10993-3 AMES & Mammalian Erythrocyte Micronucleus

{5}------------------------------------------------

8. Substantial Equivalence

Proposed DevicePredicate DevicePredicate DevicePredicate DeviceDiscuss/Justify the
(K170175)(K133054)(K113568)(K080917)Differences
Trade NameENDOSEAL MTAENDOSEALMTA FillapexIROOT SP
DeviceENDOSEAL MTA isThe raw materials ofMTA Fillapex is aiRoot SP Root CanalEquivalent
Descriptionan endodontic sealerthe ENDOSEAL aremineral trioxideSealer(IRoot SP) is
based on MTA,Natural Pure Cement,aggregate(MTA) andpremixed ready-to-use
providing a root canalZirconium dioxide,resin root canal sealerinjectable white
filling. It is premixedandused duringhydraulic cement
and pre-loaded in aCitric acid anhydrous.endodontic treatmentpaste developed for
syringe, which allowsIt is prepared as ato permanently fill thepermanent root canal
a complete filling ofmixture of powdercanal systemfilling and sealing
the entire root canaland water, and it isfollowingapplications. iRoot SP
including accessoryused in a puttydebridement andis an insoluble.
and lateral canals.form which graduallydisinfection. Itradiopaque and
hardens in the oralconsists of twoaluminum-free
environment.component pastes thatmaterial based on a
are combined in a dualcalcium silicate
barrel syringe for easecomposition, which
of dispensing andrequires the presence
consistent dosage.of water to set and
harden. iRoot SP is
packaged in a pre-
loaded syringe and is
supplied with
disposable Intra Canal
Tips.
Common NameRoot Filling MaterialRoot Filling MaterialRoot Filling MaterialRoot Filling MaterialEquivalent
ClassificationRoot Canal FillingRoot Canal FillingRoot Canal FillingRoot Canal FillingEquivalent
NameResinResinResinResin
ClassClass IIClass IIClass IIClass IIEquivalent
Product CodeKIFKIFKIFKIFEquivalent
Indications▪ Permanent■ Repair of perforationMTA Fillapex is a▪ PermanentEquivalent
for Useobturation of the root▪ Root canal fillingroot canal sealerobturation of the root

{6}------------------------------------------------

canal following vitalpulp-extirpation▪ Permanentobturation of the rootcanal followingremoval of infected ornecrotic pulp andplacement ofintracanal dressings.intended for thepermanent sealing ofroot canals and maybe used incombination with rootcanal obturationmaterials.canal following vitalpulp-extirpation▪ Permanentobturation of the rootcanal followingremoval of infected ornecrotic pulp andplacement ofintracanal dressings.iRoot SP is suitablefor use in the singlecone and lateralcondensationtechnique.
Shelf Life2 years2 years2 yearsEquivalent
StandardsISO 6876ISO 6876ISO 6876ISO 6876Equivalent
Basic ChemicalCompositionNatural pure cementZirconium dioxideBismuth trioxideBentonite ClayNatural pure cementZirconium dioxideCitric acid anhydrousDicalcium andtricalcium silicate,tricalcium aluminate,calcium oxide,Zirconium oxideCalcium silicatesCalcium phosphatemonobasicThe main ingredientsare similar but someother ingredients aredifferent. However,
n-Methyl-2-PyrrolidoneHypromellosepentaerythritolrosinate, n-ethyl-o,p-toluenesulfonamidesalicylate resin,bismuth oxide, fumedsilica, titaniumdioxideCalcium hydroxideFiller and thickeningagentsthe biocompatibilityand the performancetest results supportedthat the subject deviceis substantiallyequivalent to thepredicate devices.
Liquid FormulaPremixed typeMixing with distilledwaterMixing through a dualbarrel syringePremixed typeEquivalent
PackagingPre-loaded syringeVialDual barrel syringePre-loaded syringeEquivalent
SterileNon-sterileNon-sterileNon-sterileNon-sterileEquivalent

{7}------------------------------------------------

ENDOSEAL MTA has the same indications for use and the principle of operations as the predicate devices. It has similar physical and biocompatible properties, and demonstrates comparable performance specifications to the predicate devices. In addition, ENDOSEAL MTA has a comparable delivery system to IROOT SP.

Endoseal MTA and all of the predicate devices are equivalent as they all release calcium silicate oxide and calcium hydroxide by undergoing setting reactions when water reacts with calcium oxide. The reaction results in near identical composition of products after setting, as all of liquid components would have expelled during the setting reaction. Hence, all of the products resulted in similar test results including the cytotoxicity test.

The material compositions might slightly different from the predicate devices, however, despite of this difference, the test results of biocompatibility and performance tests show that the subject device is substantially equivalent to the predicate devices.

The bench and biocompatibility testing performed demonstrates that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that ENDOSEAL MTA is substantially equivalent to the predicate devices.

9. Conclusion:

Based on the testing results, MARUCHI concludes that the ENDOSEAL MTA is substantially equivalent to the predicate devices.

§ 872.3820 Root canal filling resin.

(a)
Identification. A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.(b)
Classification. (1) Class II if chloroform is not used as an ingredient in the device.(2) Class III if chloroform is used as an ingredient in the device.
(c)
Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to a root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976. Any other root canal filling resin shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.